510(k) Data Aggregation

    K Number
    K241418

    Validate with FDA (Live)

    Device Name
    OptoMonitor 3
    Manufacturer
    Date Cleared
    2025-02-12

    (268 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Quebec, QC G1P 4S3 Canada

    Re: K241418

    Trade/Device Name: OptoMonitor 3 Regulation Number: 21 CFR 870.2870
    Prepared: January 10th, 2025

    DEVICE

    Name of Device: OptoMonitor 3

    Regulation Number: 21 CFR 870.2870
    | 870.2870
    | 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OptoMonitor 3 is intended to measure cardiovascular blood pressure, including in heart chambers, coronary vessels and peripheral vessels, during interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve for the diagnosis and treatment of blood vessels and such as valve gradients during structural heart procedures.

    Device Description

    The proposed OptoMonitor 3 includes the display of ARi/TIARi adjunctive hemodynamic indicators when compared to the approved OptoMonitor 3 with a fully integrated TAVI software update cleared via K213854.

    AI/ML Overview

    The provided text is a 510(k) summary for the OptoMonitor 3 device. It describes a comparative analysis to a predicate device, focusing on the addition of ARi/TIARi adjunctive hemodynamic indicators. Unfortunately, the document does NOT contain a table of acceptance criteria and reported device performance directly tied to a specific study meeting those criteria. Instead, it describes general claims of meeting acceptance criteria for risk and functionality, and then details two analyses: a retrospective analysis of ARi/TIARi calculations and a clinical annotation study.

    Based on the provided text, here's a description of the acceptance criteria (inferred from the studies described) and the study that proves the device meets the acceptance criteria, as much as can be extracted:

    Overview of Device Performance and Acceptance Criteria (Inferred)

    The document primarily focuses on demonstrating the substantial equivalence of the new OptoMonitor 3 (with ARi/TIARi) to a previously cleared OptoMonitor 3 (predicate). The acceptance criteria are therefore implicitly related to showing that the new features (ARi/TIARi calculation) are accurate and that the device's original functions (pressure measurement) remain robust after the software update.

    Inferred Acceptance Criteria & Reported Device Performance

    Acceptance Criteria (Inferred)Reported Device Performance
    ARi/TIARi Calculation Accuracy: The device's calculation of ARi and TIARi should be statistically equivalent to established mathematical expressions.Retrospective analysis showed that "manual calculation of regurgitation indices using mathematical formulas given in the literature yields statistically equivalent results to the indices displayed by OpM3 TAVI. The correlation coefficients... as well as the Bland-Altman plots, indicate high levels of agreement for each of the four indices. For additional exploration, the Pearson correlations were found to be higher than 0.99 for all four indices."
    Pressure Measurement Accuracy (against expert annotation): The device's reported pressure values (Systolic LV, Systolic Ao, Diastolic Ao, Diastolic LV, LVEDP) should show high agreement with expert annotations.Clinical annotation study: Bland-Altman analysis was performed, showing the following Levels of Agreement (LoA) between OptoMonitor 3 and expert annotations: - Systolic LV: Upper 0.22, Lower -0.26 - Systolic Ao: Upper 0.15, Lower -0.17 - Diastolic Ao: Upper 0.23, Lower -0.11 - Diastolic LV: Upper 4.02, Lower -3.45 - LVEDP: Upper 4.77, Lower -2.42
    No New Questions of Safety and Effectiveness: The addition of ARi/TIARi should not introduce new safety concerns or compromise existing functionalities."No new questions of safety and effectiveness were identified during review of Risk Management documentation or execution of Verification and Validation activities." and "All acceptance criteria were met regarding risks and device functionality."

    Study Information

    The document describes two key analyses: a Retrospective Analysis for ARi/TIARi calculation, and an Annotation Study for pressure measurement accuracy.

    1. Retrospective Analysis (for ARi/TIARi Calculation)

    • Sample size: 30 pressure recordings from 10 unique patients, for a total of 150 beats.
    • Data Provenance: Clinically derived data recorded with the OptoMonitor from "existing pre and post market data sources." (Country of origin not specified, retrospective).
    • Number of experts used to establish ground truth: Not applicable – ground truth was established by "mathematical expressions given in the literature by Sinning et al. and Bugan and Kumar et al." and manual calculation (Microsoft Excel). No human experts were involved in establishing this specific ground truth.
    • Qualifications of experts (for ground truth): Not applicable.
    • Adjudication method for the test set: Not applicable, as ground truth was mathematical.
    • MRMC Comparative Effectiveness Study: No, this was an algorithmic comparison to mathematical ground truth.
    • Standalone Performance: Yes (algorithm's calculation vs. mathematical formula).
    • Type of Ground Truth: Mathematical expressions from published literature and manual calculation.
    • Sample size for training set: Not specified, implicitly zero for this specific evaluation as it's testing the implementation of known mathematical formulas. The device is a "currently marketed device" and only software changes are discussed, implying any core training would have occurred previously.
    • How the ground truth for training set was established: Not applicable for this specific evaluation.

    2. Clinical Annotation Study (for Pressure Measurement and Waveform Interpretation)

    • Sample size used for the test set: 420 waveforms in 29 patients.
    • Data Provenance: Retrospectively conducted annotation study. (Country of origin not specified, retrospective).
    • Number of experts used to establish the ground truth for the test set: "Experts panel" (number not specified, but plural implies more than one).
    • Qualifications of those experts: "Expert clinicians." Specific qualifications (e.g., years of experience, specialty) are not provided.
    • Adjudication method for the test set: Not explicitly stated, but "experts panel annotated" implies a consensus or independent annotation approach. No formal 2+1 or 3+1 method is mentioned.
    • MRMC Comparative Effectiveness Study: No, this was a comparison of the device's output to expert annotations, not an MRMC study comparing human readers with and without AI assistance.
    • Standalone Performance: Yes (algorithm's interpretation of tracings vs. expert annotation).
    • Type of Ground Truth: Expert consensus/annotations. The experts annotated "the systolic Ao, diastolic LV, diastolic Ao, and LVEDP pressures on the pressure tracings."
    • Sample size for the training set: Not specified. As with the first study, the device is marketed, suggesting prior development.
    • How the ground truth for the training set was established: Not specified.

    Additional Considerations from the Document:

    • Risk-Based Approach: The document states that the Risk Management File for the predicate device was reviewed, and one new risk ("indices measured under suboptimal conditions") was identified, evaluated as tolerable, and benefits outweigh risks. This implies internal acceptance criteria for risk management were met.
    • Verification and Validation Activities: It's stated that "All acceptance criteria were met regarding risks and device functionality" and that "No new questions of safety and effectiveness were identified during review of Risk Management documentation or execution of Verification and Validation activities." While the specific criteria aren't listed, this indicates a broader V&V process was conducted.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233904

    Validate with FDA (Live)

    Date Cleared
    2024-07-17

    (219 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K233904

    Trade/Device Name: ACIST RXi System; ACIST Navvus II Catheter Regulation Number: 21 CFR 870.2870
    Usual Name: Catheter tip pressure transducer Classification Name: Transducer, pressure, catheter tip (870.2870
    | DXO, OBI21 CFR 870.2870
    | OBI, DXO21 CFR 870.1200,21 CFR 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACIST RXi System is indicated for obtaining intravascular pressurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The ACIST Navvus II MicroCatheter is intended for use with the ACIST RXi System.

    Device Description

    The current RXi system obtains intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The RXi measures intravascular pressure in a hyperemic state following administration of adenosine as fractional flow reserve (FFR). The proposed software update for the RXi system adds a diastolic pressure ratio (dPR), which measures intravascular pressure in a non-hyperemic (resting) state. Both current and proposed ACIST RXi Systems are used in conjunction with the Navvus Catheter.

    The proposed RXi System console containing embedded software that provides the main user interface. The system is used with the Navvus catheter which contains a pressure sensor for acquisition of pressure distal (Pd) to a lesion. The proximal aortic pressure (Pa) is acquired via an interface to a third-party hemodynamic system. The system is intended for use in catheterization and related cardiovascular specialty laboratories to compute and display fractional flow reserve (FFR) using hyperemic agents and/or nonhyperemic indices of diastolic pressure ratio (dPR) and PdPa for physiological assessment of ischemic stenotic lesions.

    Measurement of FFR requires simultaneously monitoring the blood pressures proximal and distal to a lesion while inducing hyperemia. dPR is a measure of the diastolic portion of the hemodynamic waveform and can be used by the physician to perform a physiologic assessment without inducing hyperemia in the patient.

    AI/ML Overview

    The provided text describes the ACIST RXi System and Navvus II MicroCatheter, with a focus on a software update to include a diastolic pressure ratio (dPR) modality. The information primarily relates to the substantial equivalence determination for this medical device, rather than a clinical study evaluating an AI device's performance against human readers. Therefore, many of the requested points, particularly those pertaining to AI device performance evaluation criteria (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set details), are not present in the provided document.

    However, I can extract information related to the device's performance and the study that demonstrated its substantial equivalence.

    Here's a summary of the available information based on the provided text, addressing your points where possible and noting where information is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the dPR functionality were primarily based on demonstrating agreement with an FDA-cleared reference device (Volcano iFR Modality) when compared to FFR measurements.

    Acceptance Criteria (Performance Goals) vs. Reference iFR compared to FFRReported Device Performance (RXi dPR compared to FFR, and dPR vs iFR)
    Diagnostic accuracy: Not explicitly stated as a goal for dPR vs FFR, but indirectly implied by seeking agreement with iFR.Accuracy of dPR (cutpoint 0.89) vs. FFR (cutpoint 0.80): 76.39%
    Sensitivity: Pre-determined performance goal of 90% (for dPR vs iFR comparison)Sensitivity for dPR vs iFR: 99.68% (Higher than pre-determined goal)
    Specificity: Pre-determined performance goal of 84% (for dPR vs iFR comparison)Specificity for dPR vs iFR: 88.92% (Higher than pre-determined goal)
    Agreement of diagnostic accuracy between dPR and iFR compared to FFR.Diagnostic accuracy of dPR compared to iFR: 93.89% (This confirms the agreement, supporting substantial equivalence)
    Zero Drift: < 7 mmHg over one hour< 7 mmHg over one hour
    Pressure Accuracy: 3 mmHg or 3% of reading3 mmHg or 3% of reading

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions "a dataset collected in a prospective clinical study." The exact number of patients or waveforms in this dataset is not specified.
    • Data Provenance: The dataset was collected in the "prospective CONTRAST clinical study," suggesting real-world clinical data. The country of origin is not specified.
    • Retrospective or Prospective: The study is explicitly stated as prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable/not available as the study did not involve human experts establishing ground truth for an AI device. The ground truth was established by another medical device (FFR measurements).

    4. Adjudication Method for the Test Set

    This information is not applicable/not available as the study did not involve human experts requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not done. This study demonstrates the performance of a new device modality (dPR calculation) against an existing device/modality (iFR) and a reference standard (FFR), not an AI algorithm assisting human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the diagnostic performance evaluation was essentially a standalone assessment of the dPR algorithm's output compared to iFR and FFR. It states: "Resting and hyperemic pressure readings from each recorded waveform were measured using the ACIST RXi System and Navvus Catheter, and in parallel, the Philips Verrata PLUS wire and Philips (Volcano) CORE Mobile system." This describes an evaluation of the device's output (dPR) against other device outputs (iFR, FFR).

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating substantial equivalence was Fractional Flow Reserve (FFR) measurements using a cutpoint of 0.80.

    8. The Sample Size for the Training Set

    This information is not applicable/not available. The dPR algorithm is a new calculation modality added to existing software, not a deep learning AI model that requires a separate training set. The calculations are based on known physiological principles.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not available for the same reason as point 8. The dPR calculation is based on established hemodynamic principles, not learned from a labeled training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240864

    Validate with FDA (Live)

    Device Name
    PacePro Wire
    Manufacturer
    Date Cleared
    2024-05-24

    (57 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | RegulationNumber | 870.1330, 870.3680 | 870.1330, 870.2870
    , 870.3680 | Same except for 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PacePro Wire is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures.

    Additionally, the PacePro Wire can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.

    Device Description

    The proposed PacePro Wire is a guidewire that includes functions for a structural wire and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.). Throughout the submission this will be referred to as either PacePro Wire or Opsens Pacing Wire (OPW). These two names are interchangeable and represent branding name vs. internal descriptive name.

    AI/ML Overview

    The provided text is a 510(k) summary for the PacePro Wire. It primarily focuses on demonstrating substantial equivalence to a predicate device (SavvyWire) and does not contain information about a study proving the device meets specific acceptance criteria in the context of an AI/human reader performance study. The device is a medical guidewire with temporary pacing capabilities, not an AI-powered diagnostic device.

    Therefore, I cannot provide the requested information for acceptance criteria and a study proving device performance in the context of AI/human reader studies, as this information is not present in the provided document.

    However, I can extract information related to the device's performance testing as described in the summary, which serves a similar purpose of demonstrating the device meets its design requirements and is safe and effective.

    Here's what can be extracted based on the provided text, reinterpreting "acceptance criteria" and "study" in the context of a medical device submission focused on substantial equivalence:

    1. Table of Acceptance Criteria (or Performance Tests) and Reported Device Performance

    The document details performance tests conducted to show the PacePro Wire's equivalence to its predicate. While not explicitly "acceptance criteria" for an AI algorithm's performance, these are the criteria for the device's physical and functional performance.

    CategoryAcceptance Criteria/TestReported Device Performance
    SterilizationAdherence to ISO 11135-1 (Ethylene Oxide sterilization)."The following standards are utilized to show equivalence of Sterilization...ISO 11135-1 (2014)..." (Implies successful adherence)
    PackagingAdherence to ISO 11607-1 (Packaging for terminally sterilized medical devices)."The following standards are utilized to show equivalence of...Packaging...ISO 11607-1 (2019) – Amendment 1:2023..." (Implies successful adherence)
    Shelf-lifeNo specific standard mentioned, but testing was performed."The following standards are utilized to show equivalence of...Shelf life:" (Implies successful outcome).
    BiocompatibilityAdherence to FDA Guidance ISO 10993-1, categorized as "External communicating device – circulating blood," limited exposure up to 24h."Biocompatibility evaluation was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1..." Testing completed: Cytotoxicity, Sensitization, Intracutaneous reactivity, Systemic toxicity (acute), Hemocompatibility, Pyrogenicity, SC5b-9 complement. (Implies successful testing for all listed)
    Design VerificationAdherence to FDA Guidance "Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff October 2019.""All Design Requirements were successfully verified for the PacePro Wire device."
    Mechanical TestingIncluded as part of Design Verification.Part of "All Design Requirements were successfully verified."
    Particulate TestingIncluded as part of Design Verification.Part of "All Design Requirements were successfully verified."
    Rapid Pacing TestingIncluded as part of Design Verification.Part of "All Design Requirements were successfully verified."
    Safety and EffectivenessNo new questions of safety and effectiveness identified after Verification and Validation."No new questions of safety and effectiveness were identified during execution of Verification and Validation activities."
    Pacing CapabilitiesEquivalent to predicate device."The pacing capabilities of the PacePro Wire is tested to be equivalent to that of the predicate device."

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for each test. For physical device testing, sample sizes typically refer to the number of devices tested for each characteristic (e.g., tensile strength, flexibility, etc.). This detail is not provided in a summary document like a 510(k).
    • Data Provenance: The tests are conducted by the manufacturer, Opsens Inc. The data provenance would be internal laboratory testing and third-party lab testing as required for standards like ISO 10993. It's prospective testing conducted specifically for this submission. Country of origin for data is likely Canada (where Opsens Inc. is located) or certified testing labs they utilize elsewhere.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The device is a physical guidewire, not a diagnostic AI system requiring expert human interpretation as ground truth. The "ground truth" for this device's performance is established by objective engineering and biological tests against established standards.

    4. Adjudication method for the test set

    Not applicable, as it's not an AI/human reader study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI-powered device.

    7. The type of ground truth used

    The "ground truth" for this medical device is based on:

    • Compliance with established international and national standards (e.g., ISO 11135-1, ISO 11607-1, ISO 10993-1, FDA Guidance documents for guidewires).
    • Engineering specifications and functional requirements of the device.
    • Biocompatibility testing results for material safety.

    8. The sample size for the training set

    Not applicable. This is not an AI-powered device with a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI-powered device with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222216

    Validate with FDA (Live)

    Date Cleared
    2022-12-22

    (150 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    92614

    Re: K222216

    Trade/Device Name: TruWave Disposable Pressure Transducer Regulation Number: 21 CFR 870.2870
    |
    | RegulationNumber/RegulationName | 21 CFR 870.2870
    |
    | Regulation Number | 21 CFR §870.2870
    | 21 CFR §870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

    Device Description

    The TruWave Disposable Pressure Transducer is a sterile, single-use device that is used to monitor intravascular, intracranial, and intrauterine pressures. The pressure transducer has a straight, flow-through design across the pressure sensor, and is available with or without the integral flush device. The pressure sensor is a pressure sensitive silicon chip with two electrodes for excitation voltage and two electrodes for signal output. A transparent fluid path with an integral stopcock at one end and an integral flush device (either 3mL or 30mL) at the other end encloses the sensor. The enclosure of the TruWave Disposable Pressure Transducer has a pathway for air to enter the housing and acts as a vent. A disposable cable (available in 10-inch/25 cm and 48-inch/120 cm lengths) attached to the pressure transducer housing interfaces with an Edwards Lifesciences reusable cable that is specifically wired for the monitor being used. The TruWave Disposable Pressure Transducer can be mounted on the patient's arm using an arm strap or it may be mounted on an IV pole in a holder. The TruWave Disposable Pressure Transducer may be a component in various pressure monitoring kits or systems.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the TruWave Disposable Pressure Transducer. The information primarily focuses on demonstrating substantial equivalence to a predicate device through various non-clinical tests.

    However, the question asks for details related to acceptance criteria, device performance, and study information that would typically be found in a clinical study report or a more detailed performance study. The provided document does not contain a table of acceptance criteria with reported device performance metrics in a quantitative sense, nor does it detail a clinical study with human patients, ground truth establishment by experts, or MRMC studies.

    Instead, the document states: "All device acceptance criteria were met. Results of non-clinical testing show that the subject device TruWave Disposable Pressure Transducer meets its intended use and demonstrate that the device is as safe, as effective, and performs as well as the predicate device." It then lists types of non-clinical testing performed.

    Therefore,Based on the provided text, the following information can be extracted regarding acceptance criteria and the study:

    1. A table of acceptance criteria and the reported device performance:

    The document states "All device acceptance criteria were met" and that the non-clinical testing demonstrated the device "meets its intended use and demonstrate that the device is as safe, as effective, and performs as well as the predicate device." However, a specific table outlining quantitative acceptance criteria and their corresponding reported device performance values is not provided in the excerpt. The text mentions several types of testing performed to demonstrate equivalence and safety/effectiveness.

    Acceptance Criteria CategoryReported Device Performance
    Biocompatibility (ISO 10993-1:2018)Compliance demonstrated.
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2 standards.
    Electromagnetic Immunity (EMI)Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2 standards.
    Electrical Safety (ANSI/AAMI BP22:1994, IEC 60601-2-34:2011)Defibrillator challenge, dielectric strength, liquid ingress, and leakage current tests performed and met.
    Device Performance (ANSI/AAMI BP22:1994, IEC 60601-2-34:2011)Accuracy, excitation/signal impedance, light sensitivity, symmetry, and overpressure tests conducted and met.
    Magnetic Resonance (MR) Safety (ASTM F2503-20)Performed and met standards.
    Mechanical Testing (ANSI/AAMI BP22:1994, Edwards' design requirements)Performed and met requirements.
    Usability/Human FactorsEvaluated by healthcare provider users. Output implied to be satisfactory as acceptance criteria were met.

    2. Sample size(s) used for the test set and the data provenance:

    The document refers to "non-clinical testing" and does not specify sample sizes in terms of number of devices or data points for each test category. No information on data provenance (e.g., country of origin, retrospective or prospective) is provided, as these are non-clinical (laboratory/bench) tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided. The testing described is non-clinical performance and safety testing against established standards, not a clinical study requiring expert consensus for ground truth. The "usability/human factors" evaluation was performed by "healthcare provider users," but without details on the number or specific qualifications.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable and not provided, as it pertains to clinical studies with expert reviewers and not the non-clinical testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. The device is a physical pressure transducer, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided, as the device is a physical pressure transducer, not an algorithm.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" relates to compliance with established international and national standards (e.g., ISO 10993-1, IEC 60601 series, ANSI/AAMI BP22, ASTM F2503) and the manufacturer's internal design requirements. This is based on standardized measurement and testing protocols, not expert consensus, pathology, or outcomes data in a clinical sense.

    8. The sample size for the training set:

    This information is not applicable and not provided. The described tests are for device verification and validation, not for training a model or algorithm.

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided, as there is no "training set" in the context of the reported non-clinical device testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213854

    Validate with FDA (Live)

    Device Name
    SavvyWire
    Manufacturer
    Date Cleared
    2022-09-14

    (278 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | RegulationNumber | 870.1330,870.2870
    | 870.2870
    | 870.1330,870.2870
    | 870.2870
    | 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SavvyWire™ is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures, while measuring the pressure within the heart allowing calculation of hemodynamic parameters. Additionally, the Savvy Wire™ can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.

    The OptoMonitor 3 is intended to measure cardiovascular blood pressure, including in heart chambers, coronary vessels and peripheral vessels, during interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve for the diagnosis and treatment of blood vessels and such as valve gradients during structural heart procedures.

    Device Description

    The proposed SavvyWire™ is a new Catheter guidewire that includes functions for a structural wire, a pressure wire, and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.). Throughout the submission this will be referred to as either SavvyWire™ or Opsens Structural Wire (OSW). These two names are interchangeable and represent branding name vs. internal descriptive name. SavvyWire™ is intended to be used with the OptoMonitor 3 (K202943), and ideally with the OpM3-DU TAVI, which is based on the approved OptoMonitor 3 (K202943), with a fully integrated TAVI software update.

    AI/ML Overview

    The provided text describes the performance data and substantial equivalence arguments for the SavvyWire™ and OptoMonitor 3 with TAVI, rather than detailing specific acceptance criteria and the results of a study designed to prove the device meets those criteria in a quantitative table. The document focuses on demonstrating substantial equivalence to existing predicate devices through various tests and studies.

    Based on the provided text, a direct table of acceptance criteria and reported device performance with specific quantitative thresholds and met/not met status cannot be fully constructed as this information is not presented in that format. However, I can infer the general nature of the acceptance criteria from the tests conducted and the stated conclusions.

    Here's a breakdown of the requested information based on the document, with inferred acceptance criteria where explicit quantitative values are not given:


    Inferred Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Test Goals)Reported Device Performance (as stated in the document)
    Functional Equivalence (Overall System)The SavvyWire™ and OptoMonitor 3 with TAVI are shown to perform in a manner equivalent to the predicate devices with the same intended use. All substantial equivalence requirements were met.
    Safety and Effectiveness (Risk Management)"All acceptance criteria were met regarding risks and device functionality." "No new questions of safety and effectiveness were identified during review of Risk Management documentation or execution of Verification and Validation activities." "The risks related to all applicable hazards which were identified for the SavvyWire™ have been reduced to the acceptable level by mitigation...all residual risks post-mitigation have been deemed acceptable for this design." The device is shown to be at least as safe and effective as the predicate device.
    Sterilization, Packaging, Shelf-lifeEquivalence to specified ISO and ASTM standards (ISO 11135-1, ISO 11607-1, ASTM F1980-07) was utilized. Devices subjected to accelerated aging equivalent to 2 years ("TO" = no aging) were tested. (Specific results are not detailed, but adherence to standards implies meeting criteria).
    BiocompatibilityEvaluation conducted in accordance with FDA guidance Use of International Standard ISO 10993-1. All listed tests (Cytotoxicity, Sensitization, Intracutaneous reactivity, Systemic toxicity (acute), Hemocompatibility, Pyrogenicity, SC5b-9 complement) were completed. (Specific results are not detailed, but completion implies meeting criteria for classification as "External communicating device – circulating blood" with "limited exposure").
    Electrical Safety & EMCThe OptoMonitor 3 with TAVI Software complies with applicable requirements of IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-34 when used with SavvyWire™. Certain applicable portions of ISO 14708 are also applied. (Compliance implies meeting criteria).
    Mechanical Performance (e.g., Guide Wire Properties)Design Verification was conducted based on FDA Guidance Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling. All Design Requirements were verified. "Verification testing includes: Rapid Pacing Testing, Mechanical Testing, Functional Testing, FOIC Testing." "Design Specification Verification and ISO 11070 Compliance were completed by documentation check." (Meeting design requirements and compliance implies success).
    Pacing Functionality (Equivalence to Bipolar Pacing Catheters)The pacing of the SavvyWire™ (unipolar) was "tested to be equivalent to that of the reference device, Pacel™ Bipolar Pacing Catheters in an animal study comparing the two devices and clinical performance is also confirmed via a clinical study." (Successful equivalence demonstration). Animal performance testing "validated device functionality and to provide comparison data to the predicate devices." Clinical study demonstrated "use of the SavvyWire™ guidewire in the LV for rapid ventricular pacing...is substantially equivalent to other marketed devices."
    Pressure Measurement Performance (Accuracy, Range, Drift, etc. for OptoMonitor 3)Pressure Range: -30 to 300 mmHg. Pressure Accuracy: +/- 1 mmHg plus +/- 1% of reading (pressure range -30 to 50 mmHg) or +/- 3% of reading (pressure range 50 to 300 mmHg). Thermal Zero Shift: <0.3 mmHg/deg C. Zero Drift: <1 mmHg/h. (These are performance specifications listed for the OptoMonitor 3, which is updated for TAVI use, implying they are the acceptance criteria for its pressure measurement functionality and are met). The OptoMonitor 3 (with TAVI) maintains the same pressure measurement specifications as the predicate OptoMonitor 3.
    Guidewire Introduction & Positioning (for TAVR)Clinically confirmed in a single-arm study: "use of the SavvyWire™ guidewire in the LV for...appropriate advancement and positioning of the THV system during TAVR with both balloon expandable and self-expandable valves is substantially equivalent to other marketed devices."

    Here's the rest of the requested information:

    2. Sample sizes used for the test set and the data provenance:

    • Clinical Study (Test Set for Clinical Performance):

      • Sample Size: 20 patients. Twenty-one (21) devices were used (one per patient except the first patient needed two).
      • Data Provenance: Single-arm, two-center, prospective study. The document does not specify the country of origin for the clinical study data, but it is an FDA submission, implying compliance with US regulatory standards for clinical data. Given the context of the company (Opsens Inc. in Quebec, Canada) and a US consultant, the study might be multicenter, potentially including sites in the US or Canada.
    • Animal Studies:

      • Sample Size: Not specified quantitatively.
      • Data Provenance: Animal performance testing (prospective).
    • Design Verification (Mechanical, Electrical, etc.):

      • Sample Size: Not specified quantitatively, but tests were conducted. "Design Specification Verification and ISO 11070 Compliance were completed by documentation check and therefore did not use any SavvyWire™ product." implying some tests were document-based, while others (mechanical, functional, FOIC, rapid pacing) involved physical devices.
      • Data Provenance: Laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not explicitly state the number or qualifications of experts used to establish ground truth for the clinical or animal studies.
    • For a clinical study involving TAVR procedures, it is implied that the procedures were performed and outcomes assessed by qualified interventional cardiologists or cardiac surgeons. The "ground truth" for success in TAVR procedures and pacing would typically be established by established clinical protocols and expert medical judgment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document does not specify any formal adjudication method (e.g., 2+1, 3+1 for imaging or clinical endpoints) for the clinical study. It describes a single-arm clinical study for TAVR procedures.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC or AI-assisted human reader study was conducted or described. This device is a physical medical device (guidewire and monitor), not an AI-powered diagnostic imaging tool that would typically involve human-in-the-loop performance studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The device is a physical guidewire with pressure sensing and pacing capabilities, used with a monitor. The "performance" described is the device's functional and safety performance, not an algorithmic diagnostic output. Therefore, the concept of "standalone (algorithm only)" doesn't directly apply here in the typical sense of AI/software performance. However, aspects like "Pressure Accuracy," "Thermal Zero Shift," and "Zero Drift" for the OptoMonitor 3 are standalone device performance metrics, independent of human interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Clinical Study: The ground truth for the clinical performance (e.g., successful pacing, successful THV advancement/positioning) was established by clinical observation and outcomes during actual TAVR procedures in patients. This falls under outcomes data and clinical observation/judgment by the treating physicians.
    • Animal Study: Ground truth was based on direct observation of device functionality in animal models, comparison to predicate devices, and findings from necropsy and histopathology.
    • Design Verification: Ground truth for mechanical, electrical, and biocompatibility tests derived from engineering specifications, established medical device standards (ISO, ASTM, IEC), and laboratory test results.

    8. The sample size for the training set:

    • This device is not an AI/machine learning model, so there is no "training set" in the sense of data used to train an algorithm. The development of such a device involves engineering design, prototyping, and testing, rather than data-driven machine learning.

    9. How the ground truth for the training set was established:

    • As above, there is no "training set" in the context of an AI/ML model for this device. Ground truth for the underlying engineering design values would be based on established physics, biomedical engineering principles, and performance requirements derived from clinical needs and predicate device capabilities.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210235

    Validate with FDA (Live)

    Date Cleared
    2021-04-15

    (77 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | CLASSIFICATION: | 870.1330 Catheter guide wire / DQX870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Verrata PLUS pressure guide wire is indicated for use to measure pressure in blood vessels, including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. It can also be used to guide the positioning of a balloon dilatation cather interventional devices. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

    Device Description

    The Verrata PLUS pressure guide wire (hereafter referred to as the "pressure guide wire") is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The Verrata PLUS guide wire measures pressure when used with the SmartMap, ComboMap, s5 Series, CORE , and IntraSight series of systems. This pressure guide wire will not operate if connected to any other imaging system. The pressure guide wire has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm or 300 cm and in straight or pre-shaped tips. The pressure guide wire is packaged attached to the connector with a torque device to facilitate navigation through the vasculature.

    AI/ML Overview

    The provided document does not contain all the requested information for acceptance criteria and the study that proves the device meets them. This document is a 510(k) summary for the Verrata PLUS Pressure Guide Wire and focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed report of a clinical study with specific acceptance criteria and detailed performance statistics.

    Here is the information that can be extracted or inferred from the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "All device acceptance criteria were met" and that the device "meets its intended use." However, it does not provide a specific table of quantitative acceptance criteria values or the corresponding reported device performance values. It lists types of performance testing completed.

    Acceptance Criteria CategoryReported Device Performance
    EMC and Electrical SafetyMet (implied)
    Design VerificationMet (implied)
    Shelf LifeMet (implied)
    Drift PerformanceMet (implied)
    System VerificationMet (implied)
    Sensor AccuracySAME as predicate (K161887)
    Sensor DriftSAME as predicate (K161887)
    Sensor Operating RangeSAME as predicate (K161887)

    The comparison table on page 4-5 explicitly states that for "Sensor" attributes, the "Accuracy," "Drift," and "Operating Range" are "SAME" as the predicate device (Verrata PLUS Pressure Guide Wire, K161887). This implies that the acceptance criteria for these sensor-related performance metrics are identical to those established for the predicate device, and the subject device achieved performance equivalent to the predicate.

    2. Sample Size for Test Set and Data Provenance:

    The document does not specify the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily bench and laboratory-based design verification and system verification, not human clinical trials.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The document describes performance testing, likely against established engineering and technical standards, rather than expert-established ground truth from clinical data.

    4. Adjudication Method for Test Set:

    Not applicable. There is no indication of an adjudication method for a test set, as no clinical study or expert review for ground truth establishment is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study is not mentioned or implied. The device is a pressure guide wire, not an imaging or diagnostic algorithm that would typically involve human readers.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. The device is a physical medical instrument (pressure guide wire) meant for use in conjunction with other systems and interventional procedures, not a standalone algorithm.

    7. Type of Ground Truth Used:

    The "ground truth" for the performance testing cited would be based on established engineering specifications, calibration standards, and industry-recognized testing methodologies for medical devices, particularly pressure sensors and guide wires. This is not explicitly stated but is inferred from the types of tests listed (EMC, Electrical Safety, Design Verification, Shelf Life, Drift, System Verification).

    8. Sample Size for Training Set:

    Not applicable. This device is a physical instrument, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201720

    Validate with FDA (Live)

    Date Cleared
    2021-02-18

    (240 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Catheter Tip Pressure Transducer/ Catheter Guide Wire Product Code: DXO, DQX Regulation Number: 21 CFR 870.2870
    |
    | Product Code/Regulation Number | DXO, DQX21 CFR 870.2870
    | DXO, DQX21 CFR 870.2870
    | DXO, DQX21 CFR 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LiPPS™ Intravascular Pressure Sensing System is intended to be used in an environment like cathlab and for use in blood vessels, including coronary and peripheral vessels, to measure intravascular pressure during angiography and/or interventional procedures. Pressurements are obtained to provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel diseases.

    Device Description

    The subject device LiPPSTM Intravascular Pressure Sensing System contains H2000 LiPPS™ Analyzer and LiPPS™ Wire Pressure Guide Wire.

    The LiPPS™ Wire Pressure Guide Wire is used together with the H2000 LiPPS™ Analyzer in order to measure intravascular pressure. The LiPPS™ Wire contains an optical sensor and an optical fiber to induce the blood pressure and transmit signal. The LiPPS™ Wire can be connected to the analyzer through the companied handle (with an optical cable and an optical connector). The LiPPS™ Analyzer is an electronic signal processing and display units which can process signals received from the LiPPSTM Wire to display intravascular blood pressure and fractional flow reserve (FFR) values, and various connection cables. LiPPS™ Wire and companied handle (with an optical cable and an optical connector) are sterile, single-use devices. The LiPPS™ Analyzer and its accessories are reusable.

    AI/ML Overview

    The provided text is a 510(k) Summary for the LiPPS™ Intravascular Pressure Sensing System, which details the device's characteristics and its comparison to predicate devices to demonstrate substantial equivalence. This document does not describe acceptance criteria for an AI/ML-driven device or a study involving human readers and AI assistance. It focuses on the non-clinical performance and substantial equivalence of a physical medical device (an intravascular pressure sensing system) to existing predicate devices.

    Therefore, many of the requested items regarding AI/ML device performance, MRMC studies, ground truth establishment, sample sizes for AI training/testing, and expert adjudication are not applicable to the information provided in this document.

    However, I can extract information related to the device's non-clinical performance that is analogous to "acceptance criteria" and "device performance" for a medical device that measures physical parameters.

    Based on the provided text, here's what can be extracted and what cannot:


    LiPPS™ Intravascular Pressure Sensing System Acceptance Criteria and Performance (Non-Clinical)

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines performance characteristics of the LiPPS™ Intravascular Pressure Sensing System and compares them to predicate devices to demonstrate substantial equivalence. For a device like this, the "acceptance criteria" are implied by the specifications listed and the successful completion of standard-based testing.

    Acceptance Criteria (Implied/Standard-Based)Reported Device Performance (LiPPS™ Intravascular Pressure Sensing System)
    For H2000 LiPPS™ Analyzer:
    Compliance with IEC 60601-1:2005+AMD1:2012 (Basic Safety & Essential Performance)"The test results demonstrated that the proposed device comply with the standard requirements."
    Compliance with IEC 60601-1-2:2014 (EMC)"The test results demonstrated that the proposed device comply with the standard requirements."
    Compliance with IEC 60601-2-34:2011 (Invasive Blood Pressure Monitoring Equipment)"The test results demonstrated that the proposed device comply with the standard requirements."
    Software Development and Validation (IEC 62304, FDA Guidance)"The software embedded in H2000 LiPPS™ Analyzer has been developed, documented and validated in accordance with industry standards and FDA guidance."
    Pressure Range (Analyzer)-30 to 300mmHg
    For LiPPS™ Wire Pressure Guide Wire:
    Sterilization (EO process)Validated to 10⁻⁶ SAL in accordance with ISO 11135.
    Compliance with ISO 11070:2014 (Sterile single-use intravascular introducers, dilators and guidewires)"The test results demonstrated that the proposed device comply with the standard requirements."
    Pressure Range (Guide Wire)-300 to 300mmHg
    Accuracy± 1mmHg plus ± 1% of reading (over -30 to 50mmHg) or ± 1mmHg plus ± 3% of reading (over 50 to 300 mmHg)
    Zero Thermal Effect< 0.3 mmHg / °C
    Zero Drift< 1 mmHg / h
    Biological Evaluation (ISO 10993-1 classification: External Communicating Device, Circulating Blood, Limited Duration)All relevant tests (Cytotoxicity, Sensitization, Intracutaneous, Systemic Toxicity, Pyrogen, Hemolysis, Partial Thromboplastin Time, In Vivo Thromboresistance, Complement Activation, Platelet Leukocyte Count) were performed. Results demonstrated compliance.
    Physical/Mechanical Properties (Dimensions, Tensile Strength, Torque Strength, Torqueability, Coating Integrity, Particulate, Lubricity, Corrosion Resistance, Kink Resistance, Fracture, Flexing, Tip flexibility, Radiopacity, Optical contrast of Interferogram, Connection/Disconnection)Tested per ISO 11070:2014. "The test results demonstrated that the proposed device comply with the standard requirements."

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Sample Size: Not explicitly stated as a numerical count of patients or cases. For a medical device like this, "sample size" typically refers to the number of devices or components subjected to specific engineering and biocompatibility tests. These are bench tests and in-vitro/in-vivo biological evaluations.
    • Data Provenance: The studies are non-clinical (bench testing, in-vitro, and animal studies for biological evaluation). The document does not specify a country of origin for the data (beyond the manufacturer's location in Beijing, China), nor does it describe trials as retrospective or prospective (as these are not clinical trials).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Not Applicable. This is a non-clinical submission for a physical medical device. "Ground truth" in the context of expert consensus is relevant for diagnostic AI/ML devices where human interpretation provides the reference standard. For this device, ground truth is established through standardized measurements and validated test methodologies.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As there are no human readers or diagnostic interpretations involved, there's no need for an adjudication method. Performance is measured against predefined engineering specifications and regulatory standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This document describes a physical medical device for intravascular pressure sensing, not an AI/ML diagnostic tool meant to assist human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. While the device has "embedded software," its primary function is direct physical measurement (pressure sensing), not image interpretation or diagnosis via an algorithm that could be evaluated in a standalone manner like an AI application. The software processes signals, displays data, and ensures device functionality, as opposed to making diagnostic inferences.

    7. The Type of Ground Truth Used:

    • For performance metrics (e.g., accuracy, pressure range, zero drift): Measured values against calibrated reference standards using laboratory equipment and controlled environments.
    • For biological safety: Results from standard biological compatibility tests (e.g., cytotoxicity, hemolysis, pyrogenicity) as defined by ISO 10993 series and USP standards.
    • For physical/mechanical properties: Measurements and observations against design specifications and ISO 11070 standard limits.

    8. The Sample Size for the Training Set:

    • Not Applicable. This document does not describe an AI/ML device that requires a "training set" in the machine learning sense. The "training" of the device refers to its design, engineering tolerances, and calibration, which are informed by scientific principles and material properties rather than iterative machine learning from a dataset.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. (See #8)
    Ask a Question

    Ask a specific question about this device

    K Number
    K202943

    Validate with FDA (Live)

    Device Name
    OptoMonitor 3
    Manufacturer
    Date Cleared
    2020-11-24

    (55 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Quebec, QC G1P 4S3 Canada

    Re: K202943

    Trade/Device Name: OptoMonitor 3 Regulation Number: 21 CFR 870.2870
    3

    Common or Usual Name: Pressure Monitor

    Classification name: Transducer, pressure, catheter tip (870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or any interventional procedures.

    Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel desease.

    Device Description

    The proposed OptoMonitor 3 and its components are considered accessories to Opsens OptoWire™ pressure guidewires and are intended for use with legally marketed pressure guidewires.

    The proposed OptoMonitor 3 includes an Optical Unit (OU), a Display Unit (DU), a Handle Unit (HU) and accessories (cables, power supply, etc).

    The device is a non-sterile, non-patient contact device.

    AI/ML Overview

    The provided text describes the 510(k) submission for the OptoMonitor 3 device, which is a pressure monitor used with OptoWire™ pressure guidewires to measure pressure in blood vessels. The submission focuses on demonstrating substantial equivalence to a previously cleared predicate device (OptoMonitor 3 cleared via K193620).

    Crucially, this document does not describe a study involving an AI/Machine Learning algorithm for diagnostic purposes, nor does it present acceptance criteria and performance data in the context of an AI-based system. Instead, it concerns a medical device that measures physiological pressure. The "performance data" section specifically refers to electrical safety, electromagnetic compatibility (EMC), and wireless coexistence testing, not diagnostic accuracy or efficacy.

    Therefore, many of the requested points related to AI/ML (e.g., ground truth, expert consensus, MRMC study, training data) are not applicable to the content of this document.

    However, I can extract the relevant information regarding acceptance criteria and performance testing for this specific device.

    Device Type: Medical Device - Catheter Tip Pressure Transducer (OptoMonitor 3)
    Intended Use: To measure pressure in blood vessels (coronary and peripheral) during diagnostic angiography and/or interventional procedures, providing hemodynamic information such as fractional flow reserve for diagnosis and treatment of blood vessel disease.

    Here's an attempt to answer the prompt based only on the provided text, recognizing that it's for a traditional medical device, not an AI product:


    Acceptance Criteria and Device Performance Study for OptoMonitor 3

    The OptoMonitor 3 is a pressure monitoring device, and the provided document is a 510(k) submission seeking substantial equivalence to a predicate device. The "study" described here is primarily a series of verification and validation (V&V) tests to confirm that changes in the new device version do not introduce new questions of safety and effectiveness, and that its performance remains comparable to the predicate. It is not a clinical study assessing diagnostic accuracy in the way an AI algorithm would be evaluated.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal "acceptance criteria" table with numerical targets in the same format as for an AI/ML diagnostic. Instead, it focuses on demonstrating that the performance characteristics of the new OptoMonitor 3 are "essentially the same" or "equivalent" to the predicate OptoMonitor 3 (K193620). The changes are primarily related to communication methods (Bluetooth) and display unit options, along with minor software updates. The "performance data" section details compliance with various electrical safety, EMC, and wireless coexistence standards.

    The table below summarizes the key performance characteristics compared to the predicate, implying that the acceptance criterion is "same" or "equivalent performance" to the legally marketed and cleared predicate device.

    CharacteristicPredicate Device (K193620) PerformanceSubject Device (K202943) PerformanceDifferences (and Implied Acceptance: "Same/Equivalent")
    Intended UseTo measure pressure in blood vessels... for diagnosis and treatment of blood vessel disease.To measure pressure in blood vessels... for diagnosis and treatment of blood vessel disease.Same
    General Indication for UseDiagnostic computer... to compute and display various physiological parameters.Diagnostic computer... to compute and display various physiological parameters.Same
    FFR CapabilityYesYesSame
    Basis for FFR DeterminationRatio of whole heartbeats of Pd and PaRatio of whole heartbeats of Pd and PaSame
    Operating Temperature15°C to 30°C15°C to 30°CSame
    Operating Relative Humidity10% to 85% non-condensing10% to 85% non-condensingSame
    Operating Pressure70 to 106 kPa70 to 106 kPaSame
    Pressure Range-30 to 300 mmHg-30 to 300 mmHgSame
    Pressure Accuracy+/- 1 mmHg plus +/- 1% of reading (-30 to 50 mmHg) or +/- 3% of reading (50 to 300 mmHg)+/- 1 mmHg plus +/- 1% of reading (-30 to 50 mmHg) or +/- 3% of reading (50 to 300 mmHg)Same
    Zero Drift<1 mmHg/h<1 mmHg/hSame
    Electrical IsolationClass 1Class 1Same
    User InterfaceBedside: Touch screen; Control room: NoneBedside: Touch screen; Control room: YesSame, except for the addition of a control room.
    Auto-zeroingYesYesSame
    Real Time CurvesAortic instantaneous/mean pressure, distal instantaneous/mean pressure.Aortic instantaneous/mean pressure, distal instantaneous/mean pressure.Same
    Real Time Numerical ValuesMean aortic pressure, mean distal pressure, FFR, dPRMean aortic pressure, mean distal pressure, FFR, dPRSame
    Minimum Pd/Pa CursorYesYesSame
    Recording ValuesInstantaneous Pa, Pd and Pd/Pa; mean Pa; mean PdInstantaneous Pa, Pd and Pd/Pa; mean Pa; mean PdSame
    Display MonitorLCDLCDSame
    Display Unit Size15 inch15 inch, 10 inchEquivalent. Refer to Section 18 for test data.
    Bluetooth CommunicationCabled ethernet / serial between Optical and Display unitsBluetooth communication between Optical and Display units (cabled ethernet/serial backup)Equivalent. Refer to Section 17 for test data.
    Aortic InputLow Level (5µV/V/mmHg)Low Level (5µV/V/mmHg)Same
    Distal InputOptoWire (optical)OptoWire (optical)Same
    AUX InputHigh Level (100 mmHg/V)High Level (100 mmHg/V)Same
    Distal OutputLow Level (5µV/V/mmHg)Low Level (5µV/V/mmHg)Same
    Pressure GuidewireOpsens OptoWire™Opsens OptoWire™Same

    Performance Data provided to support substantial equivalence (Verification & Validation (V&V))

    • Electrical safety and electromagnetic compatibility (EMC) testing:
      • Compliance with IEC 60601-1:2012 (Consolidated text - edition 3.1)
      • Compliance with IEC60601-1-2:2007 (third edition)
      • Compliance with IEC60601-1-2:2014 (fourth edition)
    • Wireless coexistence testing:
      • Successfully tested per ANSI/IEEE C63.27:2017 in accordance with the FDA guidance (2013) Radio Frequency Wireless Technology in Medical Devices.
    • Risk Management Process: Evaluated changes, and "no new questions of safety and effectiveness were identified."
    • Verification and Validation activities: "No new questions of safety and effectiveness were identified during the execution of Verification and Validation activities."
    • Software Validation: Changes are validated in accordance with Opsens QMS, including code review, unit testing, system testing, and regression testing.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable in the context of clinical data for diagnostic accuracy. The testing described is primarily laboratory-based V&V testing (e.g., electrical safety, EMC, wireless coexistence), not a patient-based test set size.
    • Data Provenance: Not applicable. The "study" is a technical V&V assessment of the device hardware and software, not a collection of patient data.
    • Retrospective/Prospective: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not a study requiring expert-established ground truth for diagnostic purposes. The "ground truth" for the V&V tests are the established standards for electrical safety, EMC, and wireless communication, and the specifications of the predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted device, therefore an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm. Its performance is inherent in its measurement capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the AI/clinical ground truth sense. The "ground truth" for this device's performance relies on:
      • Engineering specifications and standards: Compliance with IEC 60601-1, IEC 60601-1-2, ANSI/IEEE C63.27, etc., which define acceptable performance for medical electrical equipment.
      • Predicate device performance: The previously cleared OptoMonitor 3 (K193620) serves as the benchmark for "equivalent" performance characteristics.

    8. The sample size for the training set:

    • Not applicable. The device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202543

    Validate with FDA (Live)

    Date Cleared
    2020-09-28

    (26 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | Product RegulationNumber and Name: | 870.1330 Wire, Guide, Catheter870.2870
    ProductRegulation:Number & Name | 870.1330, Catheter guidewire and870.2870
    | 870.1330, Catheter guidewire and870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniWire pressure guide wire is indicated for use to measure pressure in blood vessels, including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. It can also be used to facilitate the placement of catheters as well as other interventional devices in coronary and peripheral vessels. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

    Device Description

    The OmniWire pressure guide wire is a steerable guide wire with a pressure sensor mounted 3 cm proximal to the tip. The pressure guide wire measures pressure when used with the IntraSight and CORE Series of systems and is intended to be used in adult patients eligible for endovascular procedures. The pressure guide wire has a diameter of 0.014" (0.36 mm), a length of 185 cm and is available in straight or pre-shaped tips. It is coated with hydrophilic coating (39 cm length) on the distal portion to reduce surface friction and enhance lubricity. It is also coated with a hydrophobic coating (146 cm length) on the proximal portion to reduce surface friction. The pressure guide wire is packaged attached to the connector with an OmniWire-specific torque device to facilitate navigation through the vasculature.

    AI/ML Overview

    This document is a 510(k) summary for the OmniWire Pressure Guide Wire (K202543). It indicates that the device is substantially equivalent to a previously marketed predicate device (OmniWire Pressure Guide Wire, K192886).

    Based on the provided information, the device is a pressure guide wire, and the 510(k) is for demonstrating its substantial equivalence. The document explicitly states that no new clinical testing was completed, nor relied upon, in support of this Special 510(k) submission. Therefore, the information regarding acceptance criteria and studies demonstrating that the device meets these criteria is very limited to what is typically found in a clinical study report.

    However, the document does contain information about non-clinical testing (bench testing) that was conducted. This bench testing serves as the "study" proving the device meets certain acceptance criteria related to its design, electrical safety, electromagnetic compatibility, and aging.

    Here's a breakdown of the requested information based on the provided document:

    Acceptance Criteria and Reported Device Performance (as inferred from Non-Clinical Testing)

    Acceptance Criteria CategoryReported Device Performance (from "Summary of Non-Clinical Testing")
    System Design VerificationPerformed, indicated that the subject device performed as intended.
    Electrical SafetyEvaluated, indicated that the subject device performed as intended.
    Electromagnetic CompatibilityEvaluated, indicated that the subject device performed as intended.
    Aging StudiesEvaluated, indicated that the subject device performed as intended.
    Physical Properties"identical design, materials, physical properties, and performance specifications" to the predicate. Example: Wire Diameter: 0.0145", Wire Length: 185 cm, Sterility Assurance Level (SAL): 10^-6

    Note: The document states that the subject device and the predicate device have identical design, materials, physical properties, and performance specifications. This implies that the acceptance criteria for these physical and performance specifications are met if they are identical to the already cleared predicate device. Specific numerical acceptance criteria for these physical properties are not listed as distinct "acceptance criteria" but are stated as identical to the predicate's known specifications.

    Other Requested Information:

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for specific non-clinical tests. The document mentions "Bench testing was completed as part of design verification."
    • Data Provenance: Not specified, but generally, bench testing data for medical devices originates from the manufacturer's testing facilities (Volcano Corporation in this case). The document is a 510(k) submission, not a detailed test report. The general context suggests lab-based, pre-market non-clinical testing. It is retrospective in the sense that results are reported after testing is completed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k) submission. For non-clinical bench testing, "ground truth" is typically established by engineering specifications, calibration standards, and established testing methodologies (e.g., ASTM, ISO standards) rather than expert consensus on a test set. This type of analysis (expert review to establish ground truth) is more common in clinical studies, especially those involving image interpretation or diagnostic accuracy by human readers.

    4. Adjudication method for the test set:

    • Not applicable. See explanation for #3. Adjudication methods like 2+1 or 3+1 are primarily used in multi-reader studies to resolve discrepancies in expert interpretation, which is not the nature of the non-clinical bench testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. The document explicitly states: "No new clinical testing was completed, nor relied upon, in support of this Special 510(k) submission." This device is a physical medical device (pressure guide wire), not an AI-powered diagnostic tool, so an MRMC study comparing human performance with and without AI assistance would not be relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical medical device for measuring pressure, not a standalone algorithm.

    7. The type of ground truth used:

    • For the non-clinical testing: The ground truth is engineering specifications, calibrated instruments, and established industry test standards (e.g., for electrical safety, EMC, material properties). The device's performance was compared against these predefined physical and functional requirements.

    8. The sample size for the training set:

    • Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning. The "training" for such a device involves its design and manufacturing processes to meet specifications.

    9. How the ground truth for the training set was established:

    • Not applicable. See explanation for #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193620

    Validate with FDA (Live)

    Device Name
    OptoMonitor 3
    Manufacturer
    Date Cleared
    2020-06-18

    (175 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Technologique Quebec, G1P 4S3 Ca

    Re: K193620

    Trade/Device Name: OptoMonitor 3 Regulation Number: 21 CFR 870.2870
    3

    Common or Usual Name: Pressure Monitor

    Classification name: Transducer, pressure, catheter tip (870.2870
    | 870.2870
    | 870.2870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or any interventional procedures.

    Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel desease.

    Device Description

    The proposed OptoMonitor 3 is a new version of the OptoMonitor System. This device and its components are considered accessories to Opsens OptoWire™ pressure guidewires and are intended for use with legally marketed pressure guidewires.

    The proposed OptoMonitor 3 includes an Optical Unit (OU), a Display Unit (DU), a Handle Unit (HU) and accessories (cables, power supply, etc). These hardware components and device functionalities are equivalent to that of the previous generation OptoMonitor (K192340 (cleared on 12/12/2019).

    The device is a non-sterile, non-patient contact device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the OptoMonitor 3 device, which is an updated version of the OptoMonitor System. The submission aims to establish substantial equivalence to a predicate device (K192340).

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comparison table between the proposed OptoMonitor 3 and its predicate device, OptoMonitor (K192340), rather than explicitly listing acceptance criteria with reported performance for a new study. However, the "Technological Characteristics" section of the table effectively serves as a list of performance parameters with implied acceptance criteria being "Same" as the predicate device.

    Performance ParameterAcceptance Criteria (Implied: Same as Predicate)Reported Device Performance (OptoMonitor 3)
    Pressure Range-30 to 300 mmHg-30 to 300 mmHg
    Pressure Accuracy+/- 1 mmHg plus +/- 1% of reading (pressure range -30 to 50 mmHg) or +/- 3% of reading (pressure range 50 to 300 mmHg)+/- 1 mmHg plus +/- 1% of reading (pressure range -30 to 50 mmHg) or +/- 3% of reading (pressure range 50 to 300 mmHg)
    Thermal Zero Shift<0.3 mmHg/deg C<0.3 mmHg/deg C
    Zero Drift<1 mmHg/h<1 mmHg/h

    2. Sample Size for Test Set and Data Provenance

    The document states: "No animal studies or clinical investigations are included with this submission." and "Results from these tests mentioned above demonstrate that the technological and performance characteristics of the proposed OptoMonitor 3 is comparable to the predicate device". This implies that the substantial equivalence determination for performance was primarily based on bench testing and verification/validation activities of the device's hardware and software changes, rather than a clinical trial with a "test set" of patients.

    Therefore:

    • Sample Size for Test Set: Not applicable in the context of a clinical test set. Performance was evaluated through technical testing.
    • Data Provenance: Not applicable for clinical data. The data provenance would be internal laboratory testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since no clinical studies or human-in-the-loop evaluations were performed as part of this submission, there were no experts used to establish ground truth for a clinical test set. The validation focused on engineering and software verification.

    4. Adjudication Method for the Test Set

    Not applicable, as there was no clinical test set requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was performed or cited in this submission. The device is a pressure monitoring system, not an imaging interpretation or diagnostic aid that typically requires such studies.

    6. Standalone (Algorithm Only) Performance Study

    The document focuses on the entire OptoMonitor 3 system. While it mentions that "algorithms used for the FFR calculation remained unchanged from the predicate OptoMonitor," it does not detail a standalone performance study solely on the FFR algorithm. The performance data presented (pressure range, accuracy, etc.) relates to the entire device's ability to measure pressure.

    7. Type of Ground Truth Used

    For the engineering and software verification/validation, the ground truth would be established by:

    • Reference standards and calibrated instruments: For pressure measurements, the device's output would be compared against known, highly accurate reference pressure sources.
    • Pre-defined specifications and requirements: The software and hardware performance were validated against their design specifications.

    8. Sample Size for the Training Set

    Not applicable. This device is a measurement instrument, not an AI/ML model that undergoes a training phase with a "training set" of data in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set for an AI/ML model.

    Summary of Device Changes and Performance Rationale:

    The OptoMonitor 3 is an updated version of the predicate OptoMonitor. The key changes are related to:

    • Optical Unit: New casing, CPU, PCBs, validated software (algorithms for FFR calculation unchanged), addition of capability to receive aortic pressure signal from Cathlab Hemodynamic system, addition of low-level signal input, redesigned handle unit.
    • Display Unit: Use of commercially available all-in-one PC, new software to support additional hardware options, addition of possibility to connect to hospital DICOM system.
    • Aortic Input: Change from High Level (100 mmHg/V) to Low Level (5μV/V/mmHg) for the aortic input.
    • AUX Input: Addition of a High Level (100 mmHg/V) AUX input, which was not present on the predicate.

    The manufacturer argues substantial equivalence because:

    • Indications for Use are the same.
    • Technological characteristics are "essentially the same."
    • The "Substantial Equivalence Table" shows identical specifications for crucial performance parameters like pressure range, accuracy, thermal zero shift, and zero drift.
    • Software verification and validation testing were conducted.
    • Electrical safety and EMC testing confirmed compliance with relevant standards.
    • Risk Management processes evaluated the changes, and "no new questions of safety and effectiveness were identified."
    • The algorithms for FFR calculation remained unchanged.

    Therefore, the study supporting the device's acceptance criteria primarily involved bench testing and verification/validation activities against established engineering specifications and comparison to the predicate device's performance characteristics, without the need for animal or clinical studies.