Search Results
Found 142 results
510(k) Data Aggregation
(267 days)
Zhuhai, 519040
China
Re: K251788
Trade/Device Name: Extension tube
Regulation Number: 21 CFR 870.1650
Angiographic Injector And Syringe
Review Panel: Cardiovascular
Regulation Number: 21 CFR 870.1650
Product Name:** High Pressure Tubing
Review Panel: Cardiovascular
Regulation Number: 21 CFR 870.1650
Instruments Co., Ltd | / |
| Product code | DXT | DXT | Same |
| Device Classification | Class II, 21 CFR 870.1650
Angiographic Injector And Syringe | Class II, 21 CFR 870.1650 Angiographic Injector And Syringe | Same
The Extension tube is intended as a connecting line for injection of a contrast and saline during coronary angiography procedures.
The Extension tube is made up of connectors, male and/or female luer, caps and tubing, with the maximum injection pressure of 1200psi, as a connecting line for direct injection of contrast media or saline. It is single use and supplied sterile by Ethylene Oxide.
N/A
Ask a specific question about this device
(155 days)
System and MR Injection System Syringe Kit; MEDRAD ISI2 Module (ISI2)
Regulation Number: 21 CFR 870.1650
Classification Name: Injector and Syringe, Angiographic
Classification Regulation: 21 CFR 870.1650
Classification Name: Injector and Syringe, Angiographic
Classification Regulation: 21 CFR 870.1650
Classification Name:** Injector, Contrast Medium, Automatic
Classification Regulation: 21 CFR 870.1650
| 21 CFR 870.1650 | Same |
| Classification Product Code | DXT | DXT | Same |
| **Intended Use /
The MEDRAD® MRXperion MR Injection System is a syringe‐based fluid delivery system indicated for delivery of contrast media and saline during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners. Only trained healthcare professionals are intended to operate this device. For a complete list of compatible contrast agents for use with the MEDRAD MRXperion MR Injection System, refer to the MRXperion MR Injection System Operation Manual. The ISI2 Module is indicated for the specific purpose of allowing an injector to interface with an imaging scanner.
The MRXperion Injection System is a software-controlled, electromechanical medical device used for the administration of intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) procedures. Commonly referred to as an automated injection system, it is designed to allow a user to fill disposable syringes and then to perform an injection with a user-programmed volume and flow rate.
The MRXperion Injection System is comprised of a Scan Room Unit (SRU), Control Room Unit (CRU) and Single Use, Disposable Syringe Kit.
The SRU is located within the scan room whereas the control room unit is not. The SRU, via a wired connection, interfaces with the control room unit. The CRU, also known as the workstation, includes a graphical user interface for the injector. The operator can use the touchscreen display to manage protocols and initiate and execute injections. However, the SRU can be used to also initiate and execute injections.
The fluids are delivered from a single use only, sterile disposable set that includes one 65 mL syringe for contrast media and one 115 mL syringe for saline.
The injection system is intended to be used in an MR suite. The MR suite may be located in a mobile medical imaging trailer. The injection system is intended to be operated by personnel with training and experience in MR procedures and use of MR injection systems. Operators will consist of radiological technologists trained in the use of the equipment. This system is intended for use with the general patient population, including adults and pediatrics.
The MRXperion Injection System subject to this submission includes modified hardware and software as compared to the most recent clearance (K182276). The submission also includes an additional connectivity accessory, Imaging System Interface (ISI2) Module. The design changes do not involve any changes to the MRXperion MR Injection System Syringe Kit.
N/A
Ask a specific question about this device
(119 days)
ulricheasyINJECT Max 3 (XD 10150); ulricheasyINJECT Max 3 (XD 10180)
Regulation Number: 21 CFR 870.1650
:** Contrast Media Management System
Classification: Class II
Regulation Number: 21 CFR 870.1650
| 510(k) Number | K241850 |
| Classification | Class II |
| Regulation Number | 21 CFR 870.1650
injection of contrast media (CM) for high-pressure angiography. | |
| Product Codes | IZQ (21 CFR 870.1650
) | IZQ (21 CFR 870.1650) | Same |
| Environment of Use | MR Environment | MR Environment | Same
ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150)
ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150) is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in magnetic resonance (MR) applications.
ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150) is specifically indicated for use in MRI procedures for the delivery of the following contrast media:
- Gadobutrol Injection in single-dose (SD) container or Imaging Bulk Package (IBP)
- Gadopiclenol Injection in SD container or IBP
- Gadobenate dimeglumine Injection in SD container
- Gadoterate meglumine Injection in SD container
Easy-Click-Cassette – flex Max 2M and Easy-Click-Cassette – flex Max 3 are used for a maximum time of twenty-four (24) hours or a maximum of 96 bottles of contrast media, whichever comes first.
Use time expiration per SD container is a maximum of four (4) hours, unless otherwise stated by the contrast media labeling.
Use time expiration per IBP or saline container is a maximum of twenty-four (24) hours, unless otherwise stated by the media labeling.
Spike for MRI disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure.
ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150) is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast agent.
The ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150) is not intended for injection of contrast media (CM) for high-pressure angiography.
ulricheasyINJECT Max 3 (XD 10180)
ulricheasyINJECT Max 3 (XD 10180) is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in magnetic resonance (MR) applications.
ulricheasyINJECT Max 3 (XD 10180) is specifically indicated for use in MRI procedures for the delivery of the following contrast media:
- Gadobutrol Injection in single-dose (SD) container or Imaging Bulk Package (IBP)
- Gadopiclenol Injection in SD container or IBP
- Gadobenate dimeglumine Injection in SD container
- Gadoterate meglumine Injection in SD container
ulricheasyINJECT Max is a syringeless contrast media management system that is designed for the controlled, automatic venous administration of contrast media in conjunction with physiological saline solution to human subjects undergoing diagnostic examinations in Magnetic Resonance Imaging (MRI or PET MRI).
The ulricheasyINJECT Max device consists of a terminal, injector, and tubing system. The injector is a mobile pedestal device that consists of an injector head and injector base with rechargeable battery. The tubing system is the only component that comes in contact with the patient and has indirect contact with the blood path of a patient for a limited duration (few minutes). The tubing system consists of the following three components:
- Spikes,
- Easy-Click-Cassette – flex
- Patient Tubing
The ulricheasyINJECT Max uses a peristaltic pump as part of the injector which is designed to transport the media fluid through the tubing system (spikes, cassette, and patient tubing). ulricheasyINJECT Max is intended to be used with the following components that are not supplied with the system:
- Saline containers,
- Single-dose contrast media bottles,
- Imaging Bulk Package (IBP) contrast media containers, and
- Cannula.
ulricheasyINJECT Max is equipped with multiple hardware and software controls that work together for the safe operation of the intended use of the system. Controls include air detectors to detect the presence of air in the tubing system without direct contact with the medium, pressure controls to manage and regulate pressure inside the tubing system, and check valves to prevent backflow of media and avoid retrograde contamination.
The ulricheasyINJECT Max is provided in three models:
- ulricheasyINJECT Max 2M (XD 10140),
- ulricheasyINJECT Max 3 (XD 10150), and
- ulricheasyINJECT Max 3 (XD 10180).
The Max 3 models have 3 media connection points: 1 NaCl and 2 Contrast Media connections. The Max 2M has 2 media connection points: 1 NaCl and 1 Contrast Media connections. Max 2M and Max 3 are technically identical except the different available media connection points.
N/A
Ask a specific question about this device
(196 days)
Re: K251295
Trade/Device Name: ulrichINJECT CT Motion (XD 8000)
Regulation Number: 21 CFR 870.1650
:** Contrast Media Management System
Classification: Class II
Regulation Number: 21 CFR 870.1650
| 510(k) Number | K210541 |
| Classification | Class II |
| Regulation Number | 21 CFR 870.1650
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications.
ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of the following contrast media:
- Iohexol Injection in single-dose (SD) container or Imaging Bulk Package (IBP)
- Iodixanol Injection in SD container
- Iopamidol Injection in IBP
Pump tubing – flex can be used with a maximum of 36 bottles of contrast media or a maximum time of twenty-four (24) hours, whichever comes first.
Use time expiration per SD container is a maximum of four (4) hours from initial puncture, unless otherwise stated by the contrast media labeling.
Use time expiration per IBP is limited to a maximum time stated by the specific contrast media labeling.
Use time expiration per saline container is a maximum of twenty four (24) hours from initial puncture.
Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The maximum time usage for the Spike for CT is twenty four (24) hours, or the maximum use time for the contrast media container indicated for use with ulrichINJECT CT motion, whichever comes first.
The Patient tubing must be discarded after each patient procedure.
SYNCopen is indicated for the specific purpose of allowing an injector to interface with a CT scanner.
RIS/PACS is indicated for the specific purpose of allowing an injector to interface with a Radiological Information System (RIS) and a Picture Archiving and Communications System (PACS).
ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast media.
ulrichINJECT CT motion is a syringeless contrast media management system that is designed for the controlled, automatic administration, on the venous side, of contrast media and saline, to human subjects undergoing diagnostic examinations in computed tomography (CT) applications.
The ulrichINJECT CT motion system consists of a terminal, injector, and tubing system. The tubing system is the only component that comes in contact with the patient and has indirect contact with the blood path of a patient for a limited duration (<24 hours). The tubing system consists of the following three (3) components:
- Spike for CT (CM/NaCl);
- Pump tubing – flex; and
- Patient Tubing.
ulrichINJECT CT motion uses a peristaltic pump as part of the injector which is designed to transport the media fluid through the tubing system (spikes, pump tubing – flex, and patient tubing). ulrichINJECT CT motion is intended to be used with the following components that are not supplied with the system:
- Saline containers,
- Single-dose contrast media containers,
- Imaging Bulk Package (IBP) contrast media containers, and
- Cannula.
ulrichINJECT CT motion is equipped with multiple hardware and software controls that work together for the safe operation of the intended use of the system. Controls include air detectors to detect the presence of air in the tubing system without direct contact with the medium, pressure controls to manage and regulate pressure inside the tubing system, and check valves to prevent backflow of media and avoid retrograde contamination.
The ulrichINJECT CT motion is provided in two versions:
- Mobile pedestal version
- Ceiling version
The mobile pedestal version consists of the injector head and the injector base with rechargeable battery. The ceiling version consists of the injector head, a fixed-height arm, and a movable arm.
N/A
Ask a specific question about this device
(29 days)
Line; MEDRAD Centargo Replacement Spike; MEDRAD ISI2 Module (ISI2)
Regulation Number: 21 CFR 870.1650
Classification Name:** Injector, Contrast Medium, Automatic
Classification Regulation: 21 CFR 870.1650
CENTARGO CT INJECTION SYSTEM- 510(K) SUMMARY
PAGE 2 OF 16
Classification Regulation: 21 CFR 870.1650
-----------------|------------|
| Class | II | II | Same |
| FDA Regulation Number | 21 CFR 870.1650
| 21 CFR 870.1650 | Same |
| Classification Product Code | IZQ | IZQ | Same |
| **Intended Use/
MEDRAD® Centargo CT Injection System is an automated contrast injection system that is indicated for the controlled, automatic administration, via the venous route, of a contrast agent and 0.9% Sodium Chloride Injection USP (saline) to patients undergoing examinations in computed tomography (CT).
The system is intended for use by and under direct quasi-continuous supervision of trained healthcare professionals to assure the integrity of the contrast agent and the injection system, in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of a contrast agent.
For a complete list of compatible contrast agents for use with the MEDRAD® Centargo CT Injection System, refer to the Centargo CT Injection System Operation Manual.
The Day Set is used for a maximum time of twenty-four (24) hours, or for a maximum of 25 bottles of contrast agents, whichever comes first. The Patient Line must be discarded after each patient procedure. The Replacement Spike is for single-container use only and must be discarded with the fluid container. Contrast agent containers are to be discarded after their respective use times have expired or the Day Set use life has expired, whichever occurs first.
Saline should only be used to deliver multiple single doses to multiple patients when used with the MEDRAD® Centargo CT Injection System and the provided Saline Tag. Once the port of the saline container is punctured, it should not be removed from the work area during the entire period of use. Saline containers are to be discarded with the Day Set.
The ISI2 Module is indicated for the specific purpose of allowing an injector to interface with an imaging scanner.
The MEDRAD® Centargo CT Injection System (Centargo) is an automated contrast injection system intended to be used to inject intravenous contrast agents and saline into humans for diagnostic studies in CT applications. Centargo is intended for use with the approved contrast agent in a single-dose container and Imaging Bulk Package (IBP)
Centargo is based on well-established technologies for injection, using a piston-based electromechanical device. The system is designed to protect the patient against air injection by incorporating multiple sensors to detect air within the fluid pathway. The system is comprised of two main modules:
- Scan Room Unit – responsible for handling the contrast/saline and delivering injections
- Control Room Unit - controls, monitors, and communicates with the Scan Room Unit through wired or wireless connection
The Scan Room Unit is located within the scan room whereas the Control Room Unit is not. When communicating via wired or wireless connection, the Control Room Unit can remotely control the Scan Room Unit and initiate and execute injections. However, the Scan Room Unit can operate independently without the use of the Control Room Unit via the Scan Room Unit's graphical user interface (GUI).
The Scan Room Unit is available in two different configurations:
- Pedestal with battery and AC power
- Overhead mount with AC power only
The fluids are delivered from a multi-patient disposable set (Day Set) through a single use patient line (Patient Line). The Day Set is labeled for 24 hours or for a maximum of 25 bottles of contrast agents, whichever comes first. Fluid source spikes (Replacement Spike) are for single container use.
The system is intended to be used in a CT suite. It is intended to be operated by personnel with training and experience in CT procedures and use of CT injection systems. Operators will consist of radiological technologists trained in the use of the equipment. This system is intended for use on the general patient population, including adults and pediatrics.
N/A
Ask a specific question about this device
(154 days)
Shandong 276599 China
Re: K242143
Trade/Device Name: Angiography Injector Regulation Number: 21 CFR 870.1650
Common or Usual Name: | Angiographic Injector And Syringe |
| Regulation Number: | 21 CFR 870.1650
| 21 CFR 870.1650
| 21 CFR 870.1650
The Angiography Injector is intended to aspirate and inject contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body.
The Angiography Injector is intended to provide the function of aspirating and injecting contrast media in the interventional procedure. It can also be used to inject fluids into, or withdraw fluids from the body. The proposed device is intended for single use and is provided sterile using EO sterilization.
The Angiography Injector consists of six components: 1) Piston, 2) Plunger cap, 3) Push-button 4) Barrel, 5) Plunger and 6) Adapter (rotating or fixed). The proposed device is available in volume of 6mL, 8mL, 10mL and 12mL.
The primary package is blister package and plastic paper package. The blister package and plastic paper package all consist of PE film and Tyvek®2FS paper.
The provided text is a 510(k) summary for a medical device called "Angiography Injector." This document describes the device, its intended use, comparison to a predicate device, and the non-clinical testing performed to demonstrate substantial equivalence.
However, the 510(k) summary does not contain information related to a study proving the device meets specific acceptance criteria in the context of AI/ML performance, human reader studies (MRMC), or a detailed setup for establishing ground truth as typically observed in AI/ML medical device submissions.
The document discusses "acceptance criteria" only in the general sense of meeting design specifications and regulatory standards for a conventional angiography injector (e.g., biocompatibility testing results showing acceptance criteria met, performance tests meeting ISO 7886 requirements). It specifically states: "No clinical study is included in this submission."
Therefore, based on the provided text, I cannot answer the questions about:
- A table of acceptance criteria and reported device performance related to AI/ML or human reader studies.
- Sample sizes for test sets (for AI/ML).
- Data provenance.
- Number of experts and their qualifications for ground truth.
- Adjudication method for test sets.
- MRMC comparative effectiveness study or human reader improvement with AI.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for AI/ML.
- Sample size for the training set (for AI/ML).
- How ground truth for the training set was established.
This is because the device described is a physical medical instrument (an angiography injector and syringe), not an AI/ML-driven diagnostic or assistive software. The "performance testing" mentioned refers to engineering and safety bench tests (e.g., force to operate piston, freedom from air/liquid leakage, sterility, biocompatibility etc.), not diagnostic performance on medical images.
In summary, the provided document does not support a response to the specific questions regarding AI/ML device performance or clinical study methodologies.
Ask a specific question about this device
(131 days)
Patient Line; MEDRAD Centargo Replacement Spike; MEDRAD ISI2 Module (ISI2) Regulation Number: 21 CFR 870.1650
Classification Name: | Injector, Contrast Medium, Automatic |
| Classification Regulation: | 21 CFR 870.1650
Contrast Medium, Automatic |
| Classification Regulation: | 21 CFR 870.1650
and Syringe, Angiographic |
| Classification Regulation: | 21 CFR 870.1650
| 21 CFR 870.1650
MEDRAD® Centargo CT Injection System is an automated contrast injection system that is indicated for the controlled, automatic administration, on the venous side, of ULTRAVIST® (Iopromide), ISOVUE® (Iopamidol), OPTIRAY® (loversol), or OMNIPAQUE™ (lohexol) contrast media as supplied in an approved Imaging Bulk Package (IBP) presentation and 0.9% Sodium Chloride Injection USP (saline) to human subjects undergoing examinations in computed tomography (CT). The system is to be used by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast agent.
The Day Set is used for a maximum time of twenty-four (24) hours, or for a maximum of 25 bottles of contrast media, whichever comes first. The Patient Line must be discarded after each patient procedure. The Replacement Spike is for single-container use only and must be discarded with the fluid container. Contrast media containers are to be discarded after their respective use times have expired or the Day Set use life has expired, whichever occurs first.
Saline should only be used to deliver multiple single doses to multiple patients when used with the MEDRAD® Centargo CT Injection System and the provided Saline Tag. Once the port of the saline container is punctured, it should not be removed from the work area during the entire period of use. Saline containers are to be discarded with the Day Set.
The ISI2 Module is indicated for the specific purpose of allowing an injector to interface with an imaging scanner.
The MEDRAD® Centargo CT Injection System (Centargo) is an automated contrast injection system intended to be used to inject intravenous contrast media and saline into humans for diagnostic studies in CT applications. Centargo is intended for use with the following approved contrast media in an Imaging Bulk Package (IBP):
- . ULTRAVIST® (Iopromide) NDA 021425/S-034
- . ISOVUE® (Iopamidol) NDA 020327/S-023
- . OPTIRAY® (loversol) IBP NDA 020923/S-026
- . OMNIPAQUE™ (lohexol) NDA 020608/S-045
Centargo is based on well-established technologies for injection, using a piston-based electromechanical device. The system is designed to protect the patient against air injection by incorporating multiple sensors to detect air within the fluid pathway. The system is comprised of two main modules:
- Scan Room Unit – responsible for handling the contrast/saline and delivering injections
- . Control Room Unit - controls, monitors, and communicates with the Scan Room Unit through wired or wireless connection
The Scan Room Unit is located within the scan room whereas the Control Room Unit is not. When communicating via wired or wireless connection, the Control Room Unit can remotely control the Scan Room Unit and initiate and execute injections. However, the Scan Room Unit can operate independently without the use of the Control Room Unit via the Scan Room Unit's graphical user interface (GUI).
The Scan Room Unit is available in two different configurations:
- . Pedestal with battery and AC power
- Overhead mount with AC power only ●
The fluids are delivered from a multi-patient disposable set (Day Set) through a single use patient line (Patient Line). The Day Set is labeled for 24 hours or for a maximum of 25 bottles of contrast media, whichever comes first. Fluid source spikes (Replacement Spike) are for single container use.
The system is intended to be used in a CT suite. It is intended to be operated by personnel with training and experience in CT procedures and use of CT injection systems. Operators will consist of radiological technologists trained in the use of the equipment. This system is intended for use on the general patient population, including adults and pediatrics.
The provided text describes a 510(k) premarket notification for the MEDRAD® Centargo CT Injection System. This submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific clinical acceptance criteria for an AI/ML model for diagnostic purposes.
Therefore, the information required to populate a table of acceptance criteria for an AI model (like sensitivity, specificity, AUC) and details about study design for AI model performance (expert ground truth, MRMC studies, training set details) are not present in the provided document. The document describes a medical device, an angiographic injector, and the testing outlined is typical for hardware and software validation of such a device, including:
- Performance testing: Verification of injection volume, flow rate accuracy, pressure accuracy, air detection, heat maintenance, battery performance, occlusion detection, and protocol management.
- Software Verification and Validation Testing: Adherence to IEC 62304 and FDA guidance for device software.
- Biocompatibility Testing: According to ISO 10993-1.
- Chemical Compatibility Testing: With various contrast media.
- Cleaning and Disinfection Validation Testing:
- Sterility, Packaging, and Shelf-Life Testing: Adherence to ISO standards.
- Contamination Control Testing: Microbial ingress and cross-contamination studies.
- Human Factors/Usability Testing: According to IEC 62366-1 and FDA guidance.
- Reliability Testing:
- Electromagnetic Compatibility (EMC) and Electrical Safety Testing: Adherence to ANSI/AAMI IEC standards.
The document explicitly states "Animal Studies: Not Applicable" and "Clinical Testing: Not Applicable." This reinforces that the submission is about a device (injector) and its accessory components, not a diagnostic AI/ML algorithm that would undergo the rigorous clinical validation described in your prompt.
Based on the information available in the provided text, it is not possible to complete the requested table and answer the specific questions about AI/ML model acceptance criteria and study details. The document provides information relevant to the safety and effectiveness of an injection system, not an AI diagnostic tool.
Ask a specific question about this device
(29 days)
Trade/Device Name: Disposable Syringe (Kit); Nemoto Disposable Syringe (Kit) Regulation Number: 21 CFR 870.1650
Syringe, Angiographic |
| Regulation Number(s): | 870.1650
|
| Regulation Number | 870.1650
| 870.1650
| 870.1650
100 mL/200 mL Syringe: The Disposable Syringe (Kit) (100 mL or 200 mL Syringe for the injection of contrast media orsaline for CT and MR imaging, for use on Nemoto injectors. 150 mL Syringe: The Disposable Syringe (Kit) (150 mL Syringe for the injection of contrast media or saline forangiography. for use on Nemoto injectors. J-Tube: J-Tube is to aid in transferring a media from its container (typically a vial or bottle) into the barrel of the Disposable Syringe (Kit). Extension Tubes: The Extension Tube is to provide a connection between the syringe barrel and the needle or catheter into the patient.
The "Disposable Syringe (Kit)" (Subject Device) is specifically designed for use with the Nemoto power injection systems marketed and sold in the USA using 510(k) references K133189, K071691, K173450, K092896, K091734. The Subject Device is a package provided to the end user that contains one or two plastic syringes, plastic tubing used for connecting syringe to needle or catheter (Extension Tubing) and a plastic J-shaped tube for transferring contrast media or physiological saline into the plastic syringe (J-Tube). The Subject Device, in accordance with the Indications for Use, may be used in conjunction with CT or MRI applications or in conjunction with angiographic applications. The key difference in the application as it relates to the Subject Device is the pressure required and the flow rates required for the Subject Device. The Subject Device when used for the CT and MRI applications must support a maximum pressure of 300PSI and a maximum flow rate of 10mL/second. The Subject Device when used for the angiographic applications must support a maximum pressure of 1200PSI and a maximum flow rate of 30ml/second. The maximum values related to the application are controlled by the Nemoto power injection system which are fully described in the aforementioned 510(k)s.
The provided text is a 510(k) summary for a medical device (Disposable Syringe (Kit)) and does not contain information about an AI/ML device. Therefore, the questions related to AI/ML device acceptance criteria and study details cannot be answered based on the provided input.
However, I can provide a summary of the non-clinical performance data presented for the Disposable Syringe (Kit), which demonstrates it meets its acceptance criteria based on established standards.
Non-Clinical Performance Data for Nemoto Disposable Syringe (Kit)
The Nemoto Disposable Syringe (Kit) underwent extensive non-clinical testing to demonstrate substantial equivalence to its predicate devices. The device passed numerous tests in accordance with internal requirements, national standards, and international standards.
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Specific Test / Standard | Reported Device Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity testing (ISO 10993-5) | Passed |
| Sensitization testing (ISO 10993-10) | Passed | |
| Irritation testing (ISO 10993-23) | Passed | |
| Acute systemic toxicity testing (ISO 10993-11) | Passed | |
| Hemolysis testing (ISO 10993-4, ASTM F 756) | Passed | |
| Pyrogenicity testing (ISO 10993-11, USP-NF 38 <151>) | Passed | |
| Sterility & Contamination | Particulate Contamination testing (USP <788>) | Passed |
| Endotoxin testing (USP-NF <85> and <161>) | Passed | |
| Sterilization validation (demonstrates SAL of 10-6) | Passed | |
| Ethylene Oxide Sterilization Residuals (ISO 10993-7) | Passed | |
| Syringe Mechanical Performance | Lubricant testing (modified ISO 7886-1) | Passed |
| Tolerance on Graduated Capacity (ISO 7886-1) | Passed | |
| Scale testing (ISO 7886-1) | Passed | |
| Numbering of Scales testing (ISO 7886-1) | Passed | |
| Overall length of scale to nominal capacity line (ISO 7886-1) | Passed | |
| Position of Scale testing (ISO 7886-1) | Passed | |
| Barrel testing (ISO 7886-1) | Passed | |
| Plunger Stopper/Plunger Assembly testing (ISO 7886-1) | Passed | |
| Nozzle Lumen testing (ISO 7886-1) | Passed | |
| Dead Space testing (ISO 7886-1) | Passed | |
| Freedom from Air past Plunger Stopper (ISO 7886-1) | Passed | |
| Fit of Plunger Stopper/Plunger in Barrel (ISO 7886-1) | Passed | |
| Liquid Leakage past Plunger Stopper (in-house pressure) | Passed | |
| Chemical testing (ISO 7886-1, ISO 8536-9) | Passed | |
| Extension Tubes & Connectors | Transparency testing (ISO 8536-9) | Passed |
| Tensile Strength testing (ISO 8536-9) | Passed | |
| Leakage testing (ISO 8536-9) | Passed | |
| Injection Needle testing (ISO 8536-9) | Passed | |
| Connection force (in-house testing) | Passed | |
| Positive Pressure Liquid Leakage (ISO 80369-7) | Passed | |
| Sub-Atmospheric Pressure Air Leakage (ISO 80369-7) | Passed | |
| Stress Cracking testing (ISO 80369-7) | Passed | |
| Resistance to Separation from Axial Load (ISO 80369-7) | Passed | |
| Resistance to Separation from Unscrewing (ISO 80369-7) | Passed | |
| Resistance to Overriding testing (ISO 80369-7) | Passed | |
| Overall Device Performance | Compatibility (in-house testing) | Passed |
| Shelf Life Testing (supports 5 years) | Passed | |
| Transportation Testing (ASTM F88/88M, etc.) | Passed | |
| Usability Engineering (IEC 62366-1) | Evaluated |
2. Sample Size for Test Set and Data Provenance:
The document does not specify the sample sizes used for each non-clinical test. The tests are general performance and materials tests, not based on patient data.
3. Number of Experts and Qualifications for Ground Truth:
Not applicable. These are non-clinical engineering and laboratory tests, not clinical studies requiring expert ground truth for interpretation.
4. Adjudication Method:
Not applicable. This is not a clinical study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This is not an AI-assisted diagnostic device.
6. Standalone (Algorithm Only) Performance Study:
Not applicable. This is a physical medical device.
7. Type of Ground Truth Used:
For the non-clinical tests, the "ground truth" is defined by the specific parameters and thresholds set by the relevant ISO standards (e.g., ISO 10993, ISO 7886-1, ISO 8536-9, ISO 80369-7) and internal requirements. These standards specify acceptable limits for various properties like cytotoxicity, leakage, tensile strength, and dimensions.
8. Sample Size for Training Set:
Not applicable. This is a physical medical device, not an AI/ML system requiring a training set.
9. How Ground Truth for Training Set Was Established:
Not applicable.
Ask a specific question about this device
(60 days)
10140); ulricheasyINJECT Max 3 (XD 10150); ulricheasyINJECT Max 3 (XD 10180) Regulation Number: 21 CFR 870.1650
10150); ulricheasyINJECT Max 3 (XD 10180)Contrast Media Management SystemClass II21 CFR 870.1650
| K233737 |
| Regulation Number | 21 CFR 870.1650
|
| Product Codes | IZQ(21 CFR 870.1650
| IZQ(21 CFR 870.1650
ulricheasyINJECT Max 2M (XD 10140) / 3 (XD 10150) is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in magnetic resonance (MR) applications.
ulricheasyINJECT Max 2M / 3 (XD 10140 / XD 10150) is specifically indicated for use in MRI procedures for the delivery of Clariscan (Gadoterate Meglumine) Injection - GE Healthcare Inc., Gadavist (gadobutrol) Injection - Bayer HealthCare Pharmaceuticals Inc., VUEWAY™ (gadopiclenol) – Bracco Diagnostics, Inc., MultiHance (gadobenate dimeglumine) - Bracco Diagnostics, Inc., and Gadobutrol Injection - Fresenius Kabi AG, contrast media as supplied in approved single dose bottles and Gadavist (gadobutrol) Injection - Bayer HealthCare Pharmaceuticals Inc. and Gadobutrol Injection - Fresenius Kabi AG, contrast media as supplied in approved Imaging Bulk Packages (IBPs).
ulricheasyINJECT Max 3 (XD 10180) is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in magnetic resonance (MR) applications.
ulricheasyINJECT Max 3 (XD 10180) is specifically indicated for use in MRI procedures for the delivery of VUEWAY™ (gadopiclenol) Injection - Bracco Diagnostics, Inc., MultiHance (gadobenate dimeglumine) - Bracco Diagnostics, Inc., Clariscan™ (Gadoterate Meglumine) Injection - GE Healthcare Inc., Dotarem® (gadoterate meglumine) Injection -Guerbet, LLC, Gadavist™ (gadobutrol) Injection - Bayer HealthCare Pharmaceuticals Inc., and Gadobutrol Injection -Fresenius Kabi AG, contrast media as supplied in approved single dose vials and Gadavist™ (gadobutrol) Injection -Bayer HealthCare Pharmaceuticals Inc. and Gadobutrol Iniection - Fresenius Kabi AG. contrast media as supplied in approved Imaging Bulk Packages (IBPs).
ulricheasyINJECT Max is a syringeless contrast media management system that is designed for the controlled, automatic venous administration of contrast media in conjunction with physiological saline solution to human subjects undergoing diagnostic examinations in Magnetic Resonance Imaging (MRI or PET MRI).
The ulricheasyINJECT Max device consists of a terminal, injector, and tubing system. The injector is a mobile pedestal device that consists of an injector base with rechargeable battery. The tubing system is the only component that comes in contact with the patient and has indirect contact with the blood path of a patient for a limited duration (few minutes). The tubing system consists of the following three components:
- Spikes,
- Easy-Click-Cassette flex
- Patient Tubing
The ulricheasyINJECT Max uses a peristaltic pump as part of the injector which is designed to transport the media fluid through the tubing system (spikes, cassette, and patient tubing). ulricheasyINJECT Max is intended to be used with the following components that are not supplied with the system:
- Saline containers.
- Single-dose contrast media bottles,
- IBP contrast media containers, and
- Cannula.
ulricheasyINJECT Max is equipped with multiple hardware and software controls that work together for the safe operation of the intended use of the system. Controls include air detectors to detect the presence of air in the tubing system without direct contact with the medium, pressure controls to manage and regulate pressure inside the tubing system, and check valves to prevent backflow of media and avoid retrograde contamination.
The ulricheasyINJECT Max is provided in three models:
- ulricheasyINJECT Max 2M (XD 10140),
- ulricheasylNJECT Max 3 (XD 10150), and
- ulricheasylNJECT Max 3 (XD 10180).
The Max 3 models have 3 media connection points: 1 NaCl and 2 Contrast Media connections. The Max 2M has 2 media connection points: 1 NaCl and 1 Contrast Media connections. Max 2M and Max 3 are technically identical except the different available media connection points.
The provided text describes a 510(k) premarket notification for the ulricheasyINJECT Max contrast media management system, which primarily references a previously cleared predicate device (K233737) for most of its acceptance criteria and supporting studies. The information available focuses on the differences between the current device and its predicate.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of "acceptance criteria" with quantitative targets alongside "reported device performance" in a separate column. Instead, it describes various non-clinical tests conducted for the predicate device (K233737) and states that the subject device (ulricheasyINJECT Max) supports a broader range of contrast media, with additional testing performed to address this difference.
The following table summarizes the performance characteristics and states of verification for the device and its predicate:
| Characteristic | Acceptance Criteria (Implied) | Reported Device Performance (as stated in the document) |
|---|---|---|
| Software | Conformance with established performance criteria. | Software verification and validation performed as part of K233737, and repeated for software updates for the current submission. |
| EMC / Electrical Safety | Adherence to IEC 60601-1:2005, AMD 1:2012, AMD 2:2020 and ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012. | Testing performed in accordance with the specified standards as part of K233737. |
| Sterilization Validation | Sterility assurance level of 10-6. | Validated to a sterility assurance level of 10-6 as part of K233737 (Ethylene Oxide). |
| Shelf Life & Transport | Packaging integrity and functional stability over time and through transport. | Real-time and accelerated aging studies performed as part of K233737. Transport validation performed on tubing system, injector, and terminal as part of K233737. |
| Chemical Compatibility | Material compatibility with specified contrast media. | Additional Chemical Compatibility testing performed for the current submission to support new contrast media (Gadobutrol single dose bottle, Gadavist™ (gadobutrol) IBP, Gadobutrol IBP). Previous testing done for K233737. "The safety and effectiveness of the ulricheasyINJECT Max has been confirmed through contamination control testing, chemical compatibility testing, and extractables and leachables testing." |
| Contamination Control | Ability to prevent microbial ingress and cross-contamination; residuals within defined limits after rinsing. | Microbial ingress study demonstrated ability to prevent ingress during use. Cross-contamination study demonstrated effectiveness in preventing contamination. Rinsing study demonstrated residuals within defined limits. (All performed as part of K233737). |
| Biocompatibility | Compliance with ISO 10993-1. | Verification results indicated compliance with ISO 10993-1 as part of K233737 for indirect patient contact materials. |
| Performance – Bench | Conformance to predetermined specifications and applicable standards (ISO 8536-4 for applicable requirements). | Test and verification results indicated that the ulricheasyINJECT Max tubing system conforms to its predetermined specifications and the applicable standards (performed as part of K233737). |
| Extractables & Simulation | Leachable compounds within acceptable limits. | Testing included extractables and simulation testing for leachable compounds as part of K233737. |
| Human Factors / Usability | Safe and effective for use by intended users. | Usability study performed as part of K233737 to confirm that the ulricheasyINJECT Max is safe and effective for use by its intended users. |
| Flow Rate Accuracy | ± 5% | ± 5% (Same as predicate, implied validation via K233737 non-clinical testing). |
| Volume Accuracy | ± 5% (for 10-400 mL) | ± 5% (for 10-400 mL of contrast media; Same as predicate, implied validation via K233737 non-clinical testing). |
| Technical Detection Limit of Air in Tubing | 0.05 mL | 0.05 mL (Same as predicate, implied validation via K233737 non-clinical testing). |
| Air Detector Alarm Limit | 1 mL | 1 mL (Same as predicate, implied validation via K233737 non-clinical testing). |
| Occlusion Detection Alarm Limit | 203 PSI | 203 PSI (Same as predicate, implied validation via K233737 non-clinical testing). |
2. Sample size used for the test set and the data provenance
- Sample Size: The document does not specify exact sample sizes for each test set. It mentions "studies" and "testing" without providing numerical details for the number of units or data points tested in non-clinical assessments.
- Data Provenance: All non-clinical testing (Software, EMC/Electrical Safety, Sterilization Validation, Shelf Life and Transport Validation, Contamination Control and Rinsing, Biocompatibility, Performance – Bench, Extractables and Simulation, Human Factors / Usability) was performed as part of K233737, indicating it was conducted for the predicate device. Additional chemical compatibility testing was conducted for the current submission (K241850) to support new contrast media. The origin of the data (e.g., country) is not specified beyond being part of ulrich GmbH & Co. KG's submission, implying internal company testing. Studies are described as "non-clinical" and "bench," which means they are not human clinical trials.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The document discusses non-clinical bench testing, not image-based diagnostic performance involving expert readers establishing ground truth. The "Human Factors / Usability" study confirms the device's safety and effectiveness for its intended users (trained healthcare professionals), but it's not about expert clinical interpretation for a diagnostic algorithm.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This type of adjudication method is used for establishing ground truth in diagnostic studies involving multiple human readers, which is not described in this document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a contrast media management system, not an AI-powered diagnostic imaging tool. Therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a physical medical device (an injector) with software controls, not an algorithm meant for standalone diagnostic performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For non-clinical testing, "ground truth" generally refers to established scientific principles, engineering specifications, recognized industry standards (e.g., ISO, IEC), and predetermined specifications for the device's performance. For biocompatibility, it's compliance with ISO 10993-1. For sterilization, it's achieving a specified Sterility Assurance Level. For performance metrics like flow rate and volume accuracy, it's meeting the ±5% specification.
8. The sample size for the training set
Not applicable. The document refers to a physical medical device and its controls, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set for a machine learning model is mentioned.
Ask a specific question about this device
(122 days)
Name: Single-use Sterile High-pressure Angiographic Syringes and Accessories Regulation Number: 21 CFR 870.1650
: Injector And Syringe, Angiographic Classification: II Product Code: DXT Regulation Number: 21 CFR 870.1650
| |
| RegulationNumber | CFR 870.1650
| CFR 870.1650
Single-use Sterile High-pressure Angiographic Syringe and Accessories are intended for the injection of contrast media or saline, they shall be used with U.S. legally marketed angiographic injectors.
The proposed device is intended for the injection of contrast media or saline. It includes disposable syringes, connection tube, J shape tube, and spike.
Syringe: The syringes are intended to be used with an U.S. legally marketed angiographic injector.
Compatibility is shown in the Table 1 Compatibility between Syringe and Injectors. (Note: the newly added and the difference between K211564 and the proposed device is highlighted in Red.)
Connection tube: it is used to connect the syringe and the catheter. The tubes (extension tube) are also available in various configurations, which are Type B (used with single shot syringe, highlighted in Red mark), renamed "Straight tube" to "Type B" and added a new model (400204) connecting tube in the Table. The product structure, manufacturing process, product materials have not changed. Type Y and type T tube are used with dual shots syringe. The pressure specification for connection tube is provided in Table 2 Pressure Specifications for Connection Tube. (Note: the newly added and the difference between K211564 and proposed device is highlighted in Red)
J shape tube: it is used to draw contrast media/saline into the syringe barrel before installation.
Spike: it is used to draw contrast media/saline into the syringe barrel before the syringe installation. The pressure specification for the spike is provided in the following table as Table 3 Pressure Specifications for Spike (Note: this model is only used to distinguish whether it is made with or without DEHP). The product materials, manufacturing process and product structure have not changed. Spikes not made with DEHP are marked with Red. The new additions of spike models are: 700204- 1, 700204-2,700205-1, 700205-2, 700206, 700207-1, 700207-2)
This document is an FDA 510(k) summary for a medical device, specifically Single-use Sterile High-pressure Angiographic Syringes and Accessories. It outlines the comparison between the proposed device and a legally marketed predicate device to establish substantial equivalence.
Important Note: This document describes a medical device clearance process for physical hardware (syringes and accessories), not an AI/Software as a Medical Device (SaMD). Therefore, many of the requested categories related to AI/MRMC studies, expert consensus on images, ground truth for training/test sets, etc., are not applicable to this type of submission. The 'acceptance criteria' here refer to meeting engineering and biocompatibility standards, and 'proving the device meets the acceptance criteria' is done through non-clinical laboratory testing.
Here's an attempt to answer your questions based on the provided text, while acknowledging that many are not relevant for this device type:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria are primarily defined by adherence to recognized international and ASTM/USP standards for medical devices and the ability to withstand specified pressures. The reported performance confirms compliance with these standards and pressure specifications.
| Acceptance Criteria Category | Specific Acceptance Criteria (from standards/specifications) | Reported Device Performance/Compliance |
|---|---|---|
| Biocompatibility | ISO 10993-4, -5, -7, -10 standards; USP-NF <151> Pyrogen Test; ASTM F756 Hemolysis Assessment | No Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No Hemolysis |
| Material Safety | Ethylene Oxide sterilization residuals (ISO 10993-7) | Complies with ISO 10993-7 |
| Sterility & Packaging | Seal strength (ASTM F88/F88M-23), Seal leaks (ASTM F1929-15), Bacterial Endotoxin Limit (USP-NF <85>) | Demonstrated compliance with specified standards |
| Syringe Performance | ISO 7886-1:2017 (manual), ISO 7886-2:2020 (power-driven) Including: Lubricant weight, Graduated capacity tolerance, overall length of scale, syringe dimension, nozzle lumen, dead space, liquid leakage, air leakage, flow characteristics. | Complies with ISO 7886 standards; Meets performance for listed characteristics. |
| Connection Integrity | Small-bore connectors (ISO 80369-7:2021) | Complies with ISO 80369-7 |
| Pressure Withstanding | Syringes: 300psi, 400psi, 1200psi (depending on model) Connection tube: 400psi, 1200psi (depending on model) Spike: 400psi | Each device meets performance under maximum sustained pressure specifications. Test reports indicated compliance. |
| Shelf Life | 5-year stability | 5-year shelf life |
| Compatibility | Compatibility with U.S. legally marketed angiographic injectors (listed in Table 1) | Compatibility tests passed, demonstrating device meets specified performance requirements. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document states "Nonclinical tests were conducted to verify that the proposed device met all design specifications." and refers to compliance with ISO, ASTM, and USP standards. However, specific sample sizes for each non-clinical test are not disclosed in this summary. The data provenance is internal testing conducted by Shenzhen Boon Medical Supply Co., Ltd. (China). This would be prospective, laboratory-based testing of manufactured units, not retrospective analysis of patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not Applicable. This is a hardware medical device; no "ground truth" based on expert medical opinion (like for imaging diagnosis) is established for the purpose of these non-clinical tests. The "truth" is whether the device meets engineering specifications and safety standards as measured by laboratory equipment.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not Applicable. As this is a hardware device undergoing non-clinical engineering and biological safety tests, there is no adjudication process involving multiple human reviewers for a "test set" in the context of diagnostic accuracy. Test results are objective measurements against defined standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable. This device is a physical syringe and accessories, not an AI software or system that assists human readers in diagnosis. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not Applicable. This is not an algorithm or software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" in this context is defined by international, national, and industry-recognized standards and specifications (e.g., ISO, ASTM, USP) for medical device performance, safety, and biocompatibility. Compliance is determined by objective laboratory measurements and tests against these established benchmarks.
8. The sample size for the training set
Not Applicable. As this is a hardware device, there is no "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
Not Applicable. There is no training set.
Ask a specific question about this device