    K251289

    Validate with FDA (Live)

    Date Cleared
    2025-10-22

    (180 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K251289**
    Trade/Device Name: WELLlife COVID-19 Antigen Test Rx
    Regulation Number: 21 CFR 866.3982
    clinical specimens in near-patient settings
    Product Code: QVF
    Regulation Number: 21 CFR 866.3982
    --|--------|-------------------|
    | | | characteristics may vary. |
    | Regulation Number | 21 CFR 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WELLlife COVID-19 Antigen Test Rx is a visually read lateral flow immunoassay test intended for the qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the WELLlife COVID-19 Antigen Test Rx and followed with a molecular test.

    A negative test result is presumptive, and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

    Positive results do not rule out co-infection with other respiratory pathogens and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Performance characteristics for SARS-CoV-2 were established from April 2023 to February 2024 when SARS-CoV-2 Omicron was dominant. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

    Device Description

    The WELLlife COVID-19 Antigen Test Rx is a lateral flow immunoassay intended for qualitative detection of nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19 within the first five (5) days of symptom onset. Results are for the identification of SARS-CoV-2 nucleocapsid protein antigen. The test cassette in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with two lines: a test line (T line) and a control line (C line). The device is for in vitro diagnostic use only. The device is for prescription use only.

    The WELLlife COVID-19 Antigen Test Rx consists of the following components:

    • Test Cassette
    • Tube (pre-filled extraction buffer)
    • Swab
    • Tube Holder
    • Quick Reference Instructions (QRI)
    • Instructions for Use (IFU)
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243872

    Validate with FDA (Live)

    Date Cleared
    2025-06-16

    (181 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K243872**
    Trade/Device Name: BD Veritor System for SARS-CoV-2
    Regulation Number: 21 CFR 866.3982
    Clinical Specimens In Near-Patient Settings

    Regulatory Information:

    • Regulation section: 866.3982
      specimens in near-patient settings (Reclassification order for DEN220039 and special controls under 21 CFR 866.3982
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD Veritor™ System for SARS-CoV-2 is a chromatographic digital immunoassay for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic). The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the BD Veritor™ System for SARS-CoV-2 and followed up with a molecular test.

    A negative test result is presumptive and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

    Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Performance characteristics for SARS-CoV-2 were established between April 2024 and August 2024 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. Performance characteristics may vary with newly emerging SARS-CoV-2 virus variants.

    Device Description

    The BD Veritor™ System for SARS-CoV-2 is a rapid (approximately 15 minutes) chromatographic digital immunoassay for the direct detection of the presence or absence of SARS-CoV-2 antigens in anterior nasal swab specimens taken from patients with signs and symptoms of upper respiratory infection (i.e., symptomatic) who are suspected of COVID-19 by their healthcare provider. The test is intended for use with an opto-electronic interpretation instrument, the BD Veritor™ Plus Analyzer Instrument and is not interpreted visually.

    • When specimens are processed and added to the test device, SARS‑CoV‑2 antigens present in the specimen bind to biotinylated antibodies and antibodies conjugated to detector particles in the test strip.
    • The biotinylated antibody‑antigen‑conjugate complexes migrate across the test strip to the reaction area and are captured by a line of streptavidin bound on the membrane.
    • A positive result is determined by the BD Veritor™ Plus Analyzer when antigen‑conjugate is deposited at the Test "T" position and a control conjugate is deposited at the Control "C" position on the assay device.
    • The instrument analyzes and corrects for non‑specific binding and detects positives not recognized by the unaided eye to provide an objective result.

    Procedures to evaluate test devices depend on the BD Veritor™ Plus Analyzer workflow configuration chosen. In Analyze Now mode, the instrument evaluates assay devices after manual timing of their development. In Walk Away mode, devices are inserted immediately after application of the specimen, and timing of assay development and analysis is automated. Additionally, connection of a BD Veritor™ Plus Analyzer to a printer or IT system is possible if desired. Additional result documentation capabilities are possible with the integration of a BD Veritor™ barcode scanning enabled module.

    The Analyzer uses a proprietary algorithm that subtracts the nonspecific signal at the negative control line from the signal present at the test line. If the resultant test line signal is above a preselected cutoff, the specimen is scored as positive. If the resultant test line signal is below or equal to the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ Plus Analyzer to correctly interpret test results that cannot be scored visually because the human eye is unable to accurately perform the subtraction of the nonspecific signal. The Analyzer measures the amount of light reflected from various zones along the assay strip. The measurement of the assay background zone is an important factor during the test interpretation as the reflectance value is compared to that of the control and test zones.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary describe the BD Veritor System for SARS-CoV-2. Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets in the document. However, based on the clinical study results and FDA clearance, the implicit acceptance criteria for clinical performance are related to the confidence intervals for Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). The reported device performance is presented below:

    Performance MetricImplicit Acceptance Criteria (based on 95% C.I. reported)Reported Device Performance
    PPALower bound of 95% C.I. > 77.2%84.0% (C.I.: 77.2%, 89.1%)
    NPALower bound of 95% C.I. > 99.0%99.7% (C.I.: 99.0%, 99.9%)

    Note: The document does not explicitly state numerical acceptance thresholds for PPA and NPA (e.g., "PPA must be > 80%"). Therefore, the "Implicit Acceptance Criteria" are inferred from the demonstrated performance and the fact that the device received clearance. The FDA typically evaluates these metrics within acceptable ranges for diagnostic tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 1,032 direct anterior nasal swabs.
    • Data Provenance: The samples were prospectively collected from individual symptomatic patients across 15 geographically diverse areas across the United States between April and August 2024.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts used to establish the ground truth. The ground truth was established by an FDA-cleared SARS-CoV-2 RT-PCR test. For the false positive/negative re-testing, it broadly states "a second RT-PCR method," implying multiple tests might have been performed to confirm results without specifying expert involvement in interpreting these specific results beyond the RT-PCR outcome itself.

    4. Adjudication Method for the Test Set

    The primary ground truth for the clinical study was established by an FDA-cleared SARS-CoV-2 RT-PCR test without explicit mention of expert adjudication for every case. However, there was a form of adjudication for discordant results:

    • False Positive Adjudication: The three BD Veritor System for SARS-CoV-2 false positive results were retested with a second RT-PCR method and were confirmed negative. This suggests a method where initial discrepancies against the reference method were independently verified.
    • False Negative Adjudication: The 23 BD Veritor System for SARS-CoV-2 false negative results were retested with a second RT-PCR method in which 14 were confirmed positive and 9 were negative.

    This indicates a process of re-testing or confirmation for discordant results, which serves as a form of adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance (or similar comparative effectiveness of human readers with vs. without the device) was not explicitly mentioned or described in the provided document. The BD Veritor System for SARS-CoV-2 uses an instrument (BD Veritor™ Plus Analyzer) for interpretation, replacing visual interpretation with an automated read. The comparison is between the device's performance and a reference RT-PCR, not between human readers with and without assistance from the device.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone study was done. The entire clinical performance study (Table 9 and 11) is a standalone study, as it evaluates the performance of the BD Veritor System for SARS-CoV-2 (algorithm/device only) compared to a reference RT-PCR without human interpretation of the lateral flow assay itself. The BD Veritor™ Plus Analyzer instrument is explicitly stated to read and interpret the results, and the device "is not interpreted visually."

    7. Type of Ground Truth Used

    The ground truth used for the clinical study was an FDA-cleared SARS-CoV-2 RT-PCR test (molecular test results).

    8. Sample Size for the Training Set

    The document does not specify a sample size for a training set. This submission is for a device, and the analytical and clinical studies described are for validation of the device's performance, not for developing or training an AI/ML algorithm in the context of a typical AI/ML development pipeline. The device uses a "proprietary algorithm" for signal subtraction and interpretation, but it's not presented as a machine learning model that requires a distinct training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    Since no specific training set and its ground truth establishment are discussed in the context of AI/ML model training, this information is not applicable/provided based on the document. The "proprietary algorithm" for the instrument is described in terms of processing reflectance data and applying a preselected cutoff, and its development process (including any data used for internal calibration or parameter setting) is not detailed here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250273

    Validate with FDA (Live)

    Date Cleared
    2025-06-13

    (134 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K250273
    Trade/Device Name: BinaxNOW COVID-19 Ag Card
    Regulation Number: 21 CFR 866.3982
    Sars-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings
    Regulation Number: 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BinaxNOW COVID-19 Ag Card is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS-CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory tract infection (i.e., symptomatic). The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the BinaxNOW COVID-19 Ag Card and followed up with a molecular test.

    A negative test is presumptive and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

    Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Device Description

    The BinaxNOW COVID-19 Ag Card is an immunochromatographic membrane assay that uses antibodies to detect SARS-CoV-2 nucleocapsid protein from anterior nasal swab specimens. SARS-CoV-2 specific antibodies and a control antibody are immobilized onto a membrane support as two distinct lines and combined with other reagents/pads to construct a test strip. This test strip and a well to hold the swab specimen are mounted on opposite sides of a cardboard, book-shaped hinged test card.

    To perform the test, an anterior nasal swab specimen is collected from the patient, 6 drops of extraction reagent from a dropper bottle are added to the top hole of the swab well. The patient sample is inserted into the test card through the bottom hole of the swab well, and firmly pushed upwards until the swab tip is visible through the top hole. The swab is rotated 3 times clockwise and the card is closed, bringing the extracted sample into contact with the test strip. Test results are interpreted visually at 15 minutes based on the presence or absence of visually detectable pink/purple colored lines. Results should not be read after 30 minutes.

    AI/ML Overview

    The provided document is a 510(k) summary for the BinaxNOW COVID-19 Ag Card. It does not describe a study proving a device meets acceptance criteria in the manner typically associated with AI/ML-driven medical devices, which would involve measures like sensitivity, specificity, or AUC against a ground truth, often with human readers involved (MRMC studies).

    Instead, this document describes the validation of an immunochromatographic assay (a rapid antigen test) for COVID-19. The "acceptance criteria" here are typically performance targets for analytical and clinical characteristics (e.g., Limit of Detection, cross-reactivity, Positive Percent Agreement, Negative Percent Agreement). The "study" refers to the analytical and clinical studies conducted to demonstrate these performance characteristics.

    Therefore, the following response will interpret "acceptance criteria" as the performance benchmarks for a diagnostic assay and describe the validation studies for the BinaxNOW COVID-19 Ag Card based on the provided text.

    Here's a breakdown of the information requested, interpreted in the context of a rapid antigen test (not an AI/ML device):


    Acceptance Criteria and Device Performance for BinaxNOW COVID-19 Ag Card

    The BinaxNOW COVID-19 Ag Card is a lateral flow immunochromatographic assay, not an AI/ML diagnostic device. Therefore, the "acceptance criteria" are based on the analytical and clinical performance characteristics typical for such an in-vitro diagnostic (IVD) device, rather than metrics like AUC, sensitivity/specificity of an AI algorithm, or human reader improvement with AI assistance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric/StudyPerformance Target (Implicit/Typical for IVDs)Reported Device Performance
    Analytical PerformanceLimit of Detection (LOD)Lowest virus concentration detected ≥ 95% of the time (e.g., 19/20 replicates positive)USA-WA1/2020: 3.5 x 10³ TCID50/mL (70 TCID/swab) B.1.1.529 (Omicron): 1.6 x 10³ TCID50/mL (32.06 TCID/swab) WHO International Standard (NIBSC 21/368): 375 IU/mL (7.5 IU/swab), with 100% detection (20/20) at this concentration.
    Analytical Reactivity (Inclusivity)Detection of various SARS-CoV-2 strains at specified concentrations (all 5 replicates positive for a given concentration)Detected 19 different SARS-CoV-2 variants (Alpha, Beta, Delta, Gamma, Iota, Italy-INMI1, Kappa, Zeta, Omicron variants including BA.2.3, BA.2.12.1, BA.2.75.5, BA.4.6, BA.5, BA.5.5, BF.5, BF.7, BQ.1, BQ.1.1, XBB, JN.1) at concentrations ranging from 8.75 x 10² TCID50/mL to 5.60 x 10⁴ TCID50/mL (or IU/mL for JN.1).
    Analytical Specificity (Cross Reactivity) & Microbial InterferenceNo cross-reactivity or interference with common respiratory pathogens/commensals.No cross-reactivity or interference observed with 28 tested microorganisms (9 bacteria, 17 viruses, 1 yeast, pooled human nasal wash, and 4 Coronavirus HKU1 clinical specimens). In silico analysis for P. jirovecii showed very low potential for cross-reactivity. Possible susceptibility to SARS-CoV (due to homology) noted, but deemed low clinical likelihood.
    High Dose Hook EffectNo hook effect at high viral concentrations.No high dose hook effect observed up to 1.4 x 10⁶ TCID50/mL.
    Interfering SubstancesNo interference from specified endogenous or exogenous substances (e.g., common nasal medications, blood, mucin).No effect on test performance found at specified concentrations for 25 substances (e.g., throat lozenges, various nasal sprays, hand sanitizer, blood, mucin).
    Reproducibility/Near the Cut OffHigh agreement across sites for negative, low, moderate positive, and high negative samples.Moderate Positive: 100% (135/135) overall agreement (95% CI: 97.2%-100.0%). Low Positive: 94.1% (127/135) overall agreement (95% CI: 88.7%–97.0%). High Negative: 99.2% (132/133) overall agreement (95% CI: 95.9%-99.9%). True Negative: 99.3% (134/135) overall agreement (95% CI: 95.9%-99.9%).
    Clinical PerformancePositive Percent Agreement (PPA)High PPA against a molecular comparator (RT-PCR) in symptomatic individuals.Overall (Combined Studies): 86.9% (186/214) with 95% CI: 81.7%, 90.8% (within 5 days symptom onset). Original Study: 81.6% (71/87) with 95% CI: 72.2%, 88.4%. Omicron Study: 90.6% (115/127) with 95% CI: 84.2%, 94.5%.
    Negative Percent Agreement (NPA)High NPA against a molecular comparator (RT-PCR) in symptomatic individuals.Overall (Combined Studies): 98.5% (384/390) with 95% CI: 96.7%, 99.3% (within 5 days symptom onset). Original Study: 98.6% (205/208) with 95% CI: 95.8%, 99.5%. Omicron Study: 98.4% (179/182) with 95% CI: 95.3%, 99.4%.
    Performance by Days Post Symptom Onset (DPSO)Performance maintained within the specified window.PPA ranged across DPSO: - Day 0: 69.23% (Omicron Study) - Day 1: 94.12% (Original), 88.24% (Omicron) - Day 2: 73.33% (Original), 97.22% (Omicron) - Day 3: 76.00% (Original), 100.00% (Omicron) - Day 4: 88.89% (Original), 66.67% (Omicron) - Day 5: 100.00% (Original), 100.00% (Omicron)
    Invalid RateLow invalid rate.0.68% overall (5/730).
    User/Environmental FactorsFlex Studies (Robustness)Device performs accurately under various usage and environmental conditions.Demonstrated robustness to usage variation and environmental factors. Identified that direct exposure of test strip to wet cleaning solutions or excessive glove powder may cause erroneous results, leading to specific instructions for use.

    2. Sample Sizes and Data Provenance (Clinical Studies)

    • Clinical Test Set Sample Size:
      • Study 1 (Original): 295 evaluable subjects.
      • Study 2 (Omicron): 309 evaluable subjects.
      • Combined Clinical Data: 604 evaluable nasal swabs from symptomatic patients (within 5 days of symptom onset).
    • Data Provenance: Clinical studies were conducted within the United States.
      • Study 1: November 2020 through March 2021 (when Delta and Omicron were dominant).
      • Study 2: February 2022 to July 2022 (when Omicron and its variants were prevalent).
    • Retrospective/Prospective: Both clinical studies were prospective.

    3. Number of Experts and Qualifications for Ground Truth for Test Set

    This type of diagnostic device (lateral flow immunoassay) does not typically utilize human experts in the same way an AI/ML device would for image interpretation or clinical diagnosis. For the BinaxNOW COVID-19 Ag Card, the ground truth for the clinical studies was established by a comparator molecular test (RT-PCR). The experts involved would be the laboratory personnel performing and interpreting the RT-PCR assays. Their specific qualifications are not detailed in this summary but are implicitly assumed to be standard for clinical laboratory professionals performing EUA-authorized RT-PCR tests.

    4. Adjudication Method for the Test Set

    Not applicable in the typical sense for an AI/ML study involving human interpretation. The comparator method (RT-PCR) serves as the reference standard. The document mentions for the serial testing study's composite comparator method that in cases of discordant RT-PCR results, a third RT-PCR test was performed, and the final result based on majority rule.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is a rapid antigen test, not an AI/ML system where human readers would interpret results "with vs. without AI assistance." The test is visually read by the user, and its performance is assessed against a molecular gold standard.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    This question is not applicable in the context of this device. The BinaxNOW COVID-19 Ag Card is a manually read, qualitative visual assay. There is no AI algorithm to evaluate for standalone performance. The "performance" tables provided in the document (PPA and NPA) essentially represent the "standalone" performance of the rapid antigen test itself when interpreted visually.

    7. Type of Ground Truth Used

    • For Clinical Studies: The primary ground truth for clinical performance (PPA, NPA) was an FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assay for the detection of SARS-CoV-2.
    • For Serial Testing Study: A composite comparator method was used, involving at least two highly sensitive EUA RT-PCRs. If discordant, a third RT-PCR was performed, and the final result was based on majority rule.
    • For Analytical Studies: Ground truth was established by known concentrations of heat-inactivated SARS-CoV-2 virus or WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) for LoD and inclusivity studies, and known presence/absence of specific microorganisms for cross-reactivity.

    8. Sample Size for the Training Set

    This information is not applicable for this type of IVD device. The BinaxNOW COVID-19 Ag Card is a laboratory-developed lateral flow assay, not an AI/ML model that is 'trained' on data. Its 'training' is the fundamental assay development and optimization process, not a computational training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for this device type. The manufacturing process and quality control of the reagents and test strip govern its 'performance' characteristics, which are then analytically and clinically validated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243346

    Validate with FDA (Live)

    Date Cleared
    2025-04-11

    (165 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    0 - 18
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: cobas liat SARS-CoV-2 v2 nucleic acid test
    Regulation Number: 21 CFR 866.3982
    v2 nucleic acid test |
    | Common Name | cobas® liat SARS-CoV-2 v2 |
    | Classification | 21 CFR 866.3982
    SARS-CoV-2 for use on the cobas® Liat® System (K223783) |
    |---|---|---|
    | Regulation Name | 21 CFR 866.3982
    SPECIAL CONTROLS/STANDARDS/GUIDANCE REFERENCED

    Class II Special Controls as per 21 CFR 866.3982.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is an automated real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the qualitative detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens collected from individuals exhibiting signs and symptoms of respiratory tract infection (i.e., symptomatic). Additionally, this test is intended to be used with nasal and nasopharyngeal swab specimens collected from individuals without signs and symptoms of COVID-19 (i.e., asymptomatic).

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical and epidemiological information and laboratory findings. SARS-CoV-2 RNA is generally detectable in nasal swab and nasopharyngeal swab specimens during the acute phase of infection.

    Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out co-infection with other microorganisms. Negative results do not preclude SARS-CoV-2 infection. Negative results must be combined with clinical observations, patient history, and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    A negative result from an asymptomatic individual is presumptive. Additionally, a negative result obtained with a nasal or nasopharyngeal swab collected from an asymptomatic individual should be followed up by testing at least twice over three days with at least 48 hours between tests.

    Device Description

    The cobas® liat SARS-CoV-2 v2 nucleic acid test is performed on the cobas® liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2. An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.

    AI/ML Overview

    The provided text is a 510(k) Clearance Letter for a medical device which does not include information about AI/ML models. Therefore, it's not possible to extract the information you requested about Acceptance Criteria and a study proving an AI/ML device meets those criteria.

    The device described, the "cobas liat SARS-CoV-2 v2 nucleic acid test," is an in vitro diagnostic (IVD) device based on real-time RT-PCR technology. It directly detects viral targets and its performance is evaluated through analytical and clinical studies common for IVDs, not against AI/ML performance metrics like sensitivity, specificity, MRMC studies, or multi-reader reviews.

    Here's a breakdown of why your requested information isn't present in this document:

    • No AI/ML Component: The document describes a traditional RT-PCR assay. There is no mention of algorithms, machine learning, deep learning, or any AI component.
    • Performance Metrics Differ: The performance metrics provided (Limit of Detection, Inclusivity, Cross-reactivity, Reproducibility, Positive Percent Agreement, Negative Percent Agreement) are standard for IVD assays. They are not analogous to metrics used for evaluating AI/ML models (e.g., AUC, F1-score, accuracy in image classification, or diagnostic improvement from AI-assistance).
    • No Human Reader Interaction: Since it's an automated lab test, there's no "human reader" (like a radiologist) involved in interpreting the device's output in a way that an AI would assist. The output is qualitative (Detected/Not Detected).
    • No Ground Truth Experts in the AI Sense: Ground truth for this IVD is established by a "comparator" (another FDA-cleared NAAT) and clinical/epidemiological information, not by multiple human experts reviewing AI outputs or images.
    • No Training/Test Set Split for AI: The "test set" and "training set" concepts described in your request are fundamental to AI/ML model development and validation. For this IVD, there's a "clinical performance evaluation" using prospective and retrospective samples, which serves as the validation dataset, but it's not structured as a training/test split for an AI.

    Therefore, I cannot provide the requested table and detailed points because the provided document does not pertain to an AI/ML device.

    If you have a document describing an AI/ML medical device, please provide that, and I can attempt to extract the relevant information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233358

    Validate with FDA (Live)

    Date Cleared
    2024-08-23

    (329 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Washington 98109

    Re: K233358

    Trade/Device Name: SCoV-2 Ag Detect Rapid Test Regulation Number: 21 CFR 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SCOV-2 Ag Detect Rapid Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory tract infection (i.e., symptomatic). This test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the SCoV-2 Ag Detect Rapid Test and followed up with a molecular test.

    A negative test result is presumptive, and does not preclude SARS-CoV-2 infection; it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

    Positive results do not rule out co-infection with other bacteria or viruses and should not be used as tor diagnosis, treatment, or other patient management decisions.

    Performance characteristics for SARS-CoV-2 were established during the SARS-CoV-2 pandemic from April 2021- April 2024. The dominant variant in circulation varied throughout this period. Performance characteristics may vary with newly emerging SARS-CoV-2 virus variants.

    Device Description

    The SCOV-2 Ag Detect Rapid Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of the SARS CoV-2 nucleocapsid protein antigen directly in anterior nasal swab specimens.

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for the "SCoV-2 Ag Detect Rapid Test." It primarily focuses on the regulatory aspects of the device's clearance and does not contain detailed information about the acceptance criteria, specific study design, or performance metrics from the validation study.

    Therefore, I cannot extract the requested information (table of acceptance criteria, sample size, expert qualifications, etc.) from this document. The letter itself confirms the device's substantial equivalence to legally marketed predicates but does not include the performance data that supported this determination.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232377

    Validate with FDA (Live)

    Date Cleared
    2024-04-19

    (255 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    20854

    Re: K232377

    Trade/Device Name: Healgen Rapid COVID-19 Antigen Test Regulation Number: 21 CFR 866.3982
    To Directly Detect SARS-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings 21 CFR 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs specimens from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test
    The test does not differentiate between SARS-CoV and SARS-CoV-2.
    A negative test result is presumptive and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.
    Positive results do not rule out co-infection with other respiratory pathogens
    Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

    Device Description

    The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay that uses highly sensitive monoclonal antibodies to detect nucleocapsid protein from SARS-CoV-2 virus in nasal swab collected specimens. The test strip is composed of the following components: sample pad, reagent pad, reaction membrane, and absorbing pad housed within a test cassette. The reagent pad contains colloidal-gold-conjugated monoclonal antibody that recognizes and binds to the nucleocapsid protein of SARS-CoV-2; the reaction membrane in the test line (T) contains the second antibody that recognizes another epitope of the nucleocapsid protein of SARS-CoV-2.
    External quality controls are required but not included with the test kit and are sold separately as the COVID-19 Antigen Control Kit. The control swabs should be processed according to the Instructions for Use (IFU) and are intended to be used as quality control samples to demonstrate that the test is performing and is being performed correctly.

    AI/ML Overview

    The provided text describes the performance characteristics and acceptance criteria for the Healgen Rapid COVID-19 Antigen Test.

    Here's an analysis to extract the requested information:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds in a table. Instead, it presents the results of various performance studies. The implicit acceptance criteria are that the device performance should be clinically acceptable and comparable to existing devices for substantial equivalence.

    Here's a table summarizing the reported device performance, which can be interpreted as having met the (implicit) acceptance criteria:

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    Clinical Performance:
    Positive Percent Agreement (PPA)Clinically acceptable85.4% (95% CI: 79.1% - 90%)
    Negative Percent Agreement (NPA)Clinically acceptable99.7% (95% CI: 98.9% - 99.9%)
    Analytical Sensitivity (LoD):
    LoD (SARS-CoV-2 USA-WA1/2020)Lowest detectable concentration5.75 x 10³ TCID₅₀/mL (2.875 x 10² TCID₅₀/swab)
    LoD (WHO International Standard)Lowest detectable concentration250 IU/mL (12.5 IU/swab)
    Precision/Reproducibility:≥ 90% agreementOverall:
    Negative (NPA)99.2% (134/135)
    Weak Positive (PPA)100.0% (135/135)
    Positive (PPA)100.0% (135/135)
    Inclusivity (for tested variants):Reliable detectionAll tested variants (Alpha, Beta, Delta, Gamma, Kappa, Omicron) showed positive results at various concentrations.
    Cross-Reactivity/Interference:No cross-reactivity/interferenceNo cross-reactivity or microbial interference observed with tested organisms. No interference from tested endogenous/exogenous substances.
    Hook Effect:No detectable hook effectNo detectable hook effect up to 5.75 × 10⁶ TCID₅₀/mL
    External Controls Validation:100% agreement expected100% agreement for both positive and negative controls.
    Specimen Stability:Stable for up to 4 hoursSamples stable for up to 4 hours at room temperature.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: 806 evaluable subjects.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA, and the virus strain mentioned is "USA-WA1/2020," suggesting a U.S. context.
      • Retrospective/Prospective: Prospective. The clinical performance study collected samples "prospectively collected from symptomatic subjects between May 2022 and July 2022 at six clinical point of care sites."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The ground truth for the clinical test set was established using RT-PCR assays, not expert human readers. Therefore, the concept of "number of experts" for ground truth establishment isn't applicable in the context of this device. The RT-PCR assays are considered the gold standard for detecting SARS-CoV-2 viral nucleic acid.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The adjudication method used for the comparator (RT-PCR) was a "2 out of 3 rules." This means that if the first two FDA-cleared, highly sensitive RT-PCR assays had discordant results, a third FDA-cleared RT-PCR comparator assay was performed, and the final result was determined by the majority.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is a rapid antigen test read visually by users (healthcare professionals), not an AI-powered diagnostic imaging device that assists human readers. The study focuses on the diagnostic accuracy of the device itself compared to a molecular gold standard.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone "algorithm" in the sense of a chemical diagnostic assay that produces a visual result. There is no separate "algorithm" being evaluated without human-in-the-loop, as the human (healthcare professional) is part of reading the visually interpreted result. The "performance characteristics" section details the standalone performance of the device (the antigen test) against the RT-PCR comparator, which represents the device's accuracy without a separate human interpretation study impacting the core result.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The type of ground truth used for the clinical performance test set was molecular SARS-CoV-2 assay results, specifically composite RT-PCR assays. This is a highly sensitive and specific laboratory-based "diagnostic truth" for the presence of the virus.

    8. The sample size for the training set

    The document does not mention a training set in the context of machine learning or AI models. This device is a lateral flow immunochromatographic assay, a chemical test, not a software or AI-based diagnostic tool that would typically involve a separate training phase with a distinct dataset. The "development" of such a test involves chemical formulation and design, not typically data training in the AI sense.

    9. How the ground truth for the training set was established

    As there is no mention of a training set for an AI/ML model, this question is not applicable to the Healgen Rapid COVID-19 Antigen Test.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231187

    Validate with FDA (Live)

    Date Cleared
    2024-01-23

    (272 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K231187

    Trade/Device Name: Nano-Check COVID-19 Antigen Test Regulation Number: 21 CFR 21 CFR 866.3982
    Regulation Number: 21 CFR 866.3982

    • E.
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nano-Check™ COVID-19 Antigen Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of SARS-CoV-2 nucleoprotein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 4 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Nano-Check™ COVID-19 Antigen Test and followed up with a molecular test.

    The test does not differentiate between SARS-CoV or SARS-CoV-2.

    A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Performance characteristics for SARS-CoV-2 were established during the 2022-2023 SARS-CoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

    Device Description

    The Nano-Check™ COVID-19 Antigen Test is an immunochromatographic assay for detection of extracted SARS-CoV-2 nucleoprotein antigens in human anterior nasal swab specimens.

    The assay kit consists of 20 test cassette devices, 20 reagent tubes containing extraction buffer for Set A or 20 empty reagent tubes and 20 ampules containing extraction buffer for Set B, 20 anterior nasal specimen collection swabs, one positive control swab, one negative control swab and an Instructions for Use/Quick Reference Instruction. An external positive control swab is used with recombinant SARS-CoV-2 nucleocapsid dried onto the swab and an external negative control swab contains blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the Nano-Check COVID-19 Antigen Test meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for clinical performance are implicitly derived from the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). While explicit numerical acceptance targets are not stated, the device's performance is presented in comparison to a molecular RT-PCR test.

    MetricAcceptance Criteria (Implicit)Reported Device Performance
    Positive Percent Agreement (PPA)Acceptable level for aiding in diagnosis of SARS-CoV-2 infections in symptomatic individuals83.67% (95% CI: 76.86-88.78%)
    Negative Percent Agreement (NPA)Acceptable level for ruling out SARS-CoV-2 infections, with a recommendation for molecular confirmation99.62% (95% CI: 98.62-99.90%)

    For analytical performance, acceptance criteria are generally "no interference" for cross-reactivity and endogenous/exogenous interference, and "100% agreement" for precision and reproducibility at moderate to high concentrations.

    Metric (Analytical)Acceptance Criteria (Implicit)Reported Device Performance
    Repeatability (MP & TN samples)100% agreement100% agreement
    Repeatability (LP & HN samples)>95% agreement>95% agreement (95.8% for LP, 96.9% for HN)
    Reproducibility (All samples across sites)No significant difference in performance between sitesAll samples tested generated no significant difference
    Cross-Reactivity/Microbial InterferenceNo cross-reactivity/interference"No" for all 43 microorganisms/substances tested
    Endogenous/Exogenous InterferenceNo interference with positive or negative results"No" for all 20 substances tested
    Biotin InterferenceNo interference up to the concentration testedNo interference up to 3,500 ng/mL
    External Controls Lot-to-Lot Reproducibility100% agreement with expected results100% agreement for all lots
    Specimen Stability (2x LoD positive samples)Stable for recommended storage durationStable up to 24 hours at room temperature, 48 hours at 2°C-8°C
    Limit of Detection (LoD)Established LoD for relevant variants1.95x10^2 TCID50/mL for Omicron (e.g.)
    Analytical Reactivity (Inclusivity)100% detection at a certain viral concentration100% detection (5/5) at listed LoD for 15 SARS-CoV-2 isolates/variants
    High Dose Hook EffectNo Hook Effect100% positive results across tested high concentrations

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study Test Set): 670 subjects
    • Data Provenance: Multi-center, prospective clinical study in the U.S. in 2022-2023. At the time of the study, the SARS-CoV-2 Omicron variant was the predominant strain.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical study. The ground truth was established using an "FDA Cleared SARS-CoV-2 RT-PCR Test as comparator." RT-PCR tests are molecular diagnostic methods and are considered the gold standard for SARS-CoV-2 detection, generally not requiring expert interpretation for generating positive/negative results.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (like 2+1, 3+1, none) for discordant results between the Nano-Check COVID-19 Antigen Test and the RT-PCR comparator. The results are presented directly from this comparison.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study and no mention of AI or human readers being assisted by AI in this context. This device is a visually-read, lateral flow immunochromatographic assay, meaning results are interpreted directly by the user looking at control and test lines.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a visually-read assay. There is no algorithm for standalone performance. The interpretation is directly by the user.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth used for the clinical study was an FDA Cleared SARS-CoV-2 RT-PCR Test. This is considered the molecular "gold standard" for detecting the presence of SARS-CoV-2 viral RNA.

    8. The Sample Size for the Training Set

    The document does not report a sample size for a training set. This is expected as the Nano-Check COVID-19 Antigen Test is a traditional immunoassay, not an AI/machine learning-based device that typically requires a training set. The "training" of such a device is in its biochemical formulation and physical design, not in data-driven algorithm optimization.

    9. How the Ground Truth for the Training Set was Established

    As no training set is mentioned or applicable for this type of device, this information is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233688

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-13

    (26 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Sofia 2 SARS Antigen+ FIA; Sofia 2 SARS Antigen+ FIA Control Swab Set Regulation Number: 21 CFR 866.3982
    viral targetsfrom clinical specimens in near-patient settings |
    | Regulation Number | 21 CFR 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sofia 2 SARS Antigen+ FlA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia 2 instrument for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 6 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Sofia 2 SARS Antigen+ FIA and followed up with a molecular test.

    The test does not differentiate between SARS-CoV and SARS-CoV-2.

    A neqative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay.

    Positive results do not rule out co-infection with bacteria or other viruses and should not be used as the sole basis for treatment or other patient management decisions.

    Performanc characteristics for SARS-CoV-2 were established during the 2021-2022 SARS-CoV-2 pandemic when SARS CoV 2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variant are emerging, performance characteristics may vary.

    This test is intended for prescription use only and can be used in Point-of-Care settings.

    Sofia 2 SARS Antigen+ FIA Control Swab Set:

    The Sofia 2 SARS Antigen+FIA Control Swabs are intended to be used as quality control samples with the Sofia 2 SARS Antigen+ FIA and are representative of positive and negative test samples. These Controls may be used to demonstrate that the reagents and assay procedure perform properly.

    Device Description

    The Sofia 2 SARS Antigen+FIA is based upon a lateral flow technology that employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from the SARS-CoV-2 virus in human anterior nasal swab specimens.

    The patient sample is placed in the Reagent Tube, during which time the virus particles in the sample are disrupted, exposing internal viral nucleoproteins. After disruption, the sample is dispensed into the Test Cassette sample well. The Test Strip is composed of the following biochemical components dried and immobilized onto the nitrocellulose membrane: 1) sample pad that receives the specimen: 2) a label pad that contains detection fluorescent micro-particles, coated with monoclonal antibodies that are specific for SARS-CoV-2 nucleocapsid antigen; 3) embedded monoclonal antibodies specific for SARS-CoV-2 nucleocapsid antigen to capture the antigen-microparticle complex at the tocation. The sample pad facilitates migration of the sample fluid across the nitrocellulose strip into the absorbent pad (See Figure 4-1 in attachment). The test strip also contains a desiccant that does not participate in the assay but serves as a stabilizing agent during storage.

    Sample is applied to in the sample well and migrates through a test strip, then passes through the test and control lines. If SARS-CoV-2 viral antigen is present, they will be bound by the fluorescent microparticles io the label pad reqion, forming an antigen-microparticle complex.

    The test line is coated with monoclonal antibodies that are specific to SARS-CoV-2 nucleocapsid antigen and is intended to capture the antigen-microparticle complex. If SARS-CoV-2 viral antigen is not present, the fluorescent microparticles will not be trapped by the capture antibodies nor detected by Sofia 2.

    The Sofia 2 SARS Antigen+ FIA employs antibody tagged microparticles dyed with a fluorescent compound, to be detected and read by the Sofia 2 reader instrument. The Sofia 2 analyzers automatically scan/image the test strip, collect and analyze the fluorescence data, and report the result as either positive, negative, or invalid.

    Additionally, the Sofia 2 Antigen+ FIA utilizes a reference line for the Sofia 2 reader (to locate the test line and negative control line) and a procedural control (to assess for sample presence and adequate sample flow). No colored lines will be visible in the test window of the fluorescent assay cassette, thereby preventing visual interpretation of the test results. The operator must use the Sofia 2 analyzer to obtain a test result.

    The Sofia 2 SARS Antigen+ FIA Control Swabs are intended to be used as quality control samples representative of positive and neqative test samples, to demonstrate that the reagents are that the assay procedure is correctly perform.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the Sofia 2 SARS Antigen+ FIA. It primarily details the device's indications for use, its description, and a comparison to a legally marketed predicate device.

    Here's an analysis of the provided information to address your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in terms of performance metrics (e.g., sensitivity, specificity thresholds) for the currently reviewed device (K233688). This submission (K233688) is for a modification to the labeling of an already cleared device (DEN220039). The core performance characteristics were established during the clearance of the predicate device (DEN220039).

    The "510(k) Summary" section explicitly states under "Non-Clinical and/or Clinical Tests Summary & Conclusions": "Not applicable." This indicates that no new performance studies were conducted or provided with this specific submission because the design, material, chemical composition, and principle of operation are unchanged from the predicate device. The changes are limited to a minor revision of the intended use statement on labeling.

    Therefore, performance data to directly prove acceptance criteria for this specific submission is not provided in this document. The document refers to the established performance characteristics from the predicate device, stating: "Performanc characteristics for SARS-CoV-2 were established during the 2021-2022 SARS-CoV-2 pandemic when SARS CoV 2 Omicron was the predominant SARS-CoV-2 variant in circulation."

    If we were to infer the implied acceptance criteria from the predicate device's clearance, it would be that the device performs "as safe, as effective, and performs as well as the predicate device." However, no specific numerical performance metrics are listed in this document.

    2. Sample Size Used for the Test Set and Data Provenance

    As stated above, no new performance studies were conducted for this specific submission (K233688). The document explicitly states "Not applicable" for clinical tests. Therefore, there is no sample size for an additional test set, nor new data provenance provided here. Performance characteristics were established during the 2021-2022 SARS-CoV-2 pandemic for the predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Since no new studies were conducted, this information is not provided in the document.

    4. Adjudication Method for the Test Set

    Since no new studies were conducted, this information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No MRMC study is mentioned or implied for this submission. The device is an in vitro diagnostic (IVD) rapid antigen test read by an instrument (Sofia 2 analyzer), not typically involving multiple human readers for interpretation in the same way an imaging AI would. The Sofia 2 analyzer automatically scans, collects, analyzes fluorescence data, and reports results as positive, negative, or invalid, preventing visual interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this device inherently operates in a "standalone" fashion as an algorithm-only device in terms of result interpretation. The Sofia 2 analyzer processes the test strip by detecting and reading fluorescent microparticles, analyzing the fluorescence data, and automatically reporting the result without human interpretation of the signal on the test strip. The human interaction is limited to collecting the sample, performing the assay steps, and reading the final result from the instrument.

    7. The Type of Ground Truth Used

    The document indicates the test is intended as an aid in diagnosis, and negative test results are presumptive and "recommended these results be confirmed by a molecular SARS-CoV-2 assay." This strongly implies that the ground truth for establishing the device's performance characteristics (for the predicate device) would have been molecular SARS-CoV-2 assays (e.g., RT-PCR), which are considered the gold standard for SARS-CoV-2 detection.

    8. The Sample Size for the Training Set

    No information about a training set or its sample size is provided in this document, as no new performance studies were conducted for this submission. The initial development and training (if applicable, for the Sofia 2 analyzer's signal processing algorithm) would have occurred prior to the predicate device's clearance.

    9. How the Ground Truth for the Training Set Was Established

    Since no information on a training set is provided, the method for establishing its ground truth is also not mentioned. It is likely that for the predicate device, if machine learning was involved in interpreting the fluorescent signal, the training data's ground truth would have been established against molecular SARS-CoV-2 assays, similar to the test set ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223783

    Validate with FDA (Live)

    Date Cleared
    2023-12-04

    (353 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: cobas SARS-CoV-2 Nucleic acid test for use on the cobas Liat System Regulation Number: 21 CFR 866.3982
    Targets fromClinical Specimens in Near-Patient Settings |
    | Regulation Number | 21 CFR 866.3982
    ----------------------------------------------------------|
    | Regulation Name | 21 CFR 866.3982
    | 21 CFR 866.3982
    SPECIAL CONTROLS/STANDARDS/GUIDANCE REFERENCED

    Class II Special Controls as per 21 CFR 866.3982.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® SARS-CoV-2 Nucleic acid test for use on the cobas® SARS-CoV-2) is an automated, realtime reverse transcriptase polymerase chain reaction (RT-PCR) test intended for the rapid in vitro qualitative detection of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in anterior nasal (nasal) and nasopharyngeal swab specimens collected from individuals with signs and symptoms of respiratory tract infection (i.e., symptomatic). Additionally, this test is intended to be used with nasal and necimens collected from individuals without signs and symptoms suspected of COVID-19 (i.e., asymptomatic).

    The cobas® SARS-CoV-2 performed on the cobas® Liat® System is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. SARS-CoV-2 RNA is generally detectable in nasal and nasopharyngeal swab specimens during the acute phase of infection.

    Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out co-infection with other microorganisms.

    A negative result from an asymptomatic individual is presumptive. Additionally, a negative result obtained with a nasal swab collected from an asymptomatic patient should be followed up by testing at least twice over three days with at least 48 hours between tests.

    Negative results do not preclude SARS-CoV-2 infection.

    The results of this test should not be used as the sole basis for diagnosis, treatment management decisions.

    This test is intended for prescription use only and can be used in Point-of-Care settings.

    Device Description

    The cobas® SARS-CoV-2 Nucleic acid test for use on the cobas® Liat® System (cobas® SARS-CoV-2) uses real-time reverse transcriptase polymerase chain reaction (RT-PCR) technology to rapidly (approximately 20 minutes) detect SARS-CoV-2 virus from nasopharyngeal and nasal swabs. The automation, small footprint, and easy-to-use interface of the cobas® Liat® System enable performance of this test to occur at the POC or in a clinical laboratory setting.

    The cobas® SARS-CoV-2 assay is performed on the cobas® Liat® Analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time RT-PCR assays. The assay targets both the ORF1 a/b nonstructural region and nucleocapsid protein gene that are unique to SARS-CoV-2. An Internal Process Control (IPC) is also included. The IPC is present to control for adequate processing of the target virus through steps of sample purification, nucleic acid amplification, and to monitor the presence of inhibitors in the RT-PCR processes.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study findings for the "cobas SARS-CoV-2 Nucleic acid test for use on the cobas Liat System" based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the clinical performance evaluation results, specifically the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with a composite comparator method. While explicit acceptance criteria values are not stated (e.g., "PPA must be >X%"), the device's performance is presented to demonstrate its effectiveness.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (NPS - Symptomatic)Reported Device Performance (NPS - Asymptomatic)Reported Device Performance (NS - Symptomatic)Reported Device Performance (NS - Asymptomatic)
    Positive Percent Agreement (PPA)High PPA (e.g., >90%)95.4% (95% CI: 90.4% - 97.9%)96.3% (95% CI: 87.5% - 99.0%)96.3% (95% CI: 91.6% - 98.4%)90.0% (95% CI: 78.6% - 95.7%)
    Negative Percent Agreement (NPA)High NPA (e.g., >95%)99.5% (95% CI: 98.4% - 99.8%)99.6% (95% CI: 99.0% - 99.8%)99.8% (95% CI: 99.0% - 100.0%)99.9% (95% CI: 99.5% - 100.0%)

    Study Information

    1. Sample sizes used for the test set and data provenance:

      • Clinical Performance Evaluation (Test Set):
        • Evaluable NPS samples: 1874 (fresh and frozen)
          • Symptomatic: 673 NPS
          • Asymptomatic: 1201 NPS (413 suspected, 788 no suspicion)
        • Evaluable NS samples: 1872 (fresh and frozen, including 1 indeterminate result)
          • Symptomatic: 674 NS
          • Asymptomatic: 1198 NS (411 suspected, 787 no suspicion)
        • Frozen Samples (included in above totals): 23 positive and 23 negative NPS specimens, and 23 positive and 23 negative NS specimens.
        • Data Provenance: Prospectively collected clinical specimens from February-June 2022 (fresh samples) and earlier during the COVID-19 pandemic (March-June 2021) for frozen samples. The document does not specify the country of origin, but it implies collection within "10 point-of-care healthcare facilities" without further geographic detail.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The ground truth for the test set was established using a composite comparator method consisting of "three highly sensitive FDA-EUA-authorized laboratory-based RT-PCR assays."
      • The document does not specify the number or qualifications of human experts involved in establishing this ground truth. The ground truth relies on the performance of these reference RT-PCR assays.
    3. Adjudication method for the test set:

      • The text describes a "composite comparator method" using "three highly sensitive FDA-EUA-authorized laboratory-based RT-PCR assays." This implies a form of consensus or a predefined algorithm for combining the results of these three assays to establish the definitive positive or negative status. However, the specific adjudication rules (e.g., "2 out of 3 positive for overall positive") are not explicitly described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving human readers or AI assistance is described. This study focuses on the standalone performance of the cobas SARS-CoV-2 test.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done. The entire clinical performance evaluation section (Section 7) describes the performance of the "cobas SARS-CoV-2 test" (the device/algorithm) against a composite comparator method (ground truth) without human intervention in the result interpretation of the device. The device itself is an automated RT-PCR test.
    6. The type of ground truth used:

      • The ground truth used was a composite comparator method based on the results from "three highly sensitive FDA-EUA-authorized laboratory-based RT-PCR assays." This is a form of laboratory reference standard.
    7. The sample size for the training set:

      • The document describes a clinical performance evaluation (test set) and analytical studies. It does not provide information about the sample size of a specific "training set" for the device, as it is a molecular diagnostic test rather than a machine learning algorithm that typically undergoes explicit training on large datasets in the way an AI image analysis product would. The development (implied "training" in a broader sense for assay optimization) would have used various samples, but these are not explicitly detailed as a distinct training set in the provided text.
    8. How the ground truth for the training set was established:

      • As no explicit "training set" with established ground truth is described in the context of typical AI/ML development, this question cannot be answered from the provided text. The assay development would involve extensive analytical validation using characterized viral samples and clinical specimens validated by reference methods, which broadly contribute to optimizing the assay's performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221925

    Validate with FDA (Live)

    Date Cleared
    2023-08-10

    (405 days)

    Product Code
    Regulation Number
    866.3982
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Scarborough, Maine 04074

    Re: K221925

    Trade/Device Name: ID NOW COVID-19 2.0 Regulation Number: 21 CFR 866.3982
    NAME ID NOW COVID-19 2.0

    COMMON NAME ID NOW COVID, ID NOW COVID-19

    CLASSIFICATION NAME

    21 CFR 866.3982
    |
    | Regulation Number/Name | 21 CFR 866.3982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ID NOW COVID-19 2.0 performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology (NAAT) intended for the qualitative detection of nucleic acid from SARS-CoV-2 in direct anterior nasal (nasal) or nasopharyngeal swabs from individuals with signs and symptoms of respiratory tract infection. ID NOW COVID-19 2.0 performed on the ID NOW Instrument is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. SARS-CoV-2 RNA is generally detectable in nasal and nasopharyngeal swab specimens during the acute phase of infection.

    Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not preclude co-infection with bacteria or other viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    A negative test result is presumptive, and it is recommended these results be confirmed by another molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. This test is intended for prescription use only and can be used in Pointof-Care settings.

    Device Description

    ID NOW COVID-19 2.0 is a rapid, instrument-based isothermal test for the qualitative detection of nucleic acid from SARS-CoV-2 viral RNA in direct nasal or nasopharyngeal swabs. The ID NOW COVID-19 2.0 System utilizes isothermal nucleic acid amplification technology and is comprised of:

    • Sample Receiver single use, disposable containing the elution buffer.
    • Test Base single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet.
    • Transfer Cartridge - single use, disposable for transfer of the eluted sample to the Test Base
    • Patient Swabs sterile anterior nasal swabs (foam) for anterior nasal swab collection and for use as a Negative Control Swab
    • Positive Control Swab - single use, to ensure that test reagents are working properly and that the test is correctly performed, and
    • ID NOW Instrument

    The reaction tubes in the Test Base contain lyophilized reagents required for amplification of the target nucleic acid and an internal control. ID NOW COVID-19 2.outilizes a pair of templates (similar to primers) for the specific amplification of RNA from SARS-CoV-2 and a fluorescently labeled molecular beacon designed to specifically identify the amplified nucleic acid targets. ID NOW COVID-19 2.0 is performed within the confinement of the Test Base, and no other part of the ID NOW Instrument has contact with the sample during the amplification process. This minimizes the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the ID NOW Instrument. The sample is added to the Sample Receiver and transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating viral lysis and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the ID NOW Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the ID NOW Instrument by the operator either manually or using barcode scanner. Data can be retrieved and downloaded by the operator at any time after testing. An external Universal Printer can be attached via USB to the ID NOW Instrument to print test results.

    AI/ML Overview

    The provided text describes the performance of the ID NOW COVID-19 2.0 device, which is an in vitro diagnostic test for SARS-CoV-2.

    Here's an analysis addressing your questions, keeping in mind that this is a diagnostic test and not an AI-assisted imaging device, so some questions (like MRMC studies) are not applicable.

    1. Table of Acceptance Criteria and Reported Device Performance

    For diagnostic tests, "acceptance criteria" are typically defined by performance targets for sensitivity (Positive Agreement) and specificity (Negative Agreement). These are assessed against a highly accurate comparator method (Ground Truth).

    Acceptance Criteria (Implied by High Performance and Clinical Utility for a Molecular Test) and Reported Device Performance:

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Positive Agreement (Sensitivity)High (e.g., >85-90%)91.7% (95% CI: 87.8% - 94.4%)
    Negative Agreement (Specificity)Very High (e.g., >95%)98.4% (95% CI: 97.1% - 99.1%)
    Analytical Sensitivity (LoD)Low (detect SARS-CoV-2 at low concentrations)500 copies/swab (20 copies/reaction)
    Analytical Reactivity (Inclusivity)Detect a wide range of SARS-CoV-2 strains including variants of concernAll tested strains detected at specified concentrations (various Omicron sub-variants were tested)
    Analytical Specificity (Cross-Reactivity)No cross-reactivity with common respiratory pathogens or commensalsAll 38 tested microorganisms (24 viruses, 12 bacteria, 2 yeasts) found negative; In-silico analysis predicted no cross-reactivity
    Microbial InterferenceNo interference from other common respiratory pathogensNo effect on test performance when SARS-CoV-2 was present with other pathogens
    Interfering SubstancesNo interference from common interfering substances (nasal sprays, blood, etc.)No effect on test performance by specified substances at tested concentrations
    ReproducibilityConsistent results across sites, operators, and daysHigh agreement: Moderate positive 98.1%, Low positive 96.3%, High negative 89.6%, True negative 99.6%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Test Set: 914 specimens (460 anterior nasal swabs and 454 nasopharyngeal swabs).
    • Data Provenance:
      • Country of Origin: United States (US). The study was conducted at twenty-one (21) US sites.
      • Retrospective or Prospective: Prospective clinical study. Specimens were collected from individuals showing signs and symptoms of upper respiratory infection at the time of enrollment.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This is a diagnostic test based on molecular detection, not an imaging device requiring expert reader interpretation for ground truth.

    • Number of "Experts": Not applicable in the context of human readers for establishing ground truth for individual cases. The ground truth was established by a composite comparator method using three (3) FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assays. This method is considered the "gold standard" for SARS-CoV-2 detection.
    • Qualifications of Experts: N/A for individual case ground truth. The RT-PCR assays are performed by laboratory personnel following established protocols.

    4. Adjudication Method for the Test Set

    The ground truth was established by a composite comparator method using three (3) FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assays. This implies that the consensus or outcome of these highly sensitive and specific molecular tests served as the ground truth. The document does not describe a human adjudication process, as the ground truth is derived from objective molecular testing results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • Was it done?: No, an MRMC study was not conducted for this device.
    • Effect Size of Human Readers Improvement with AI vs. without AI assistance: This question is not applicable as the ID NOW COVID-19 2.0 is a standalone molecular diagnostic test, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Was it done?: Yes, the clinical performance study evaluated the standalone performance of the ID NOW COVID-19 2.0 device. The results are automatically reported by the ID NOW Instrument (see Device Description), meaning there is no human interpretation of the raw signal or algorithm output before a result is given. Results are "Positive," "Negative," or "Invalid."

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the clinical study was established using a composite comparator method consisting of three (3) FDA Emergency Use Authorized real-time Polymerase Chain Reaction (RT-PCR) assays for the detection of SARS-CoV-2. This is considered a molecular gold standard or definitive diagnostic method.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size for a "training set" in the context of machine learning model development. This is a traditional IVD device, not an AI/ML diagnostic. The analytical studies (LoD, Inclusivity, Cross-Reactivity, Interference) and clinical study are part of the validation process, not "training" of an AI model.

    • Training Set Size: Not applicable/not explicitly stated as this is not an AI/ML device that undergoes a discrete training phase with labeled data in the same way. The development and optimization of the assay itself would involve iterative testing, but this is different from an AI "training set."

    9. How the Ground Truth for the Training Set was Established

    • How Ground Truth for Training was Established: Not applicable as this is a molecular diagnostic test, not an AI/ML device requiring a distinct training set with established ground truth labels in the typical AI sense. The assay's analytical characteristics and performance are established through experiments and clinical trials.