K Number
K231187

Validate with FDA (Live)

Date Cleared
2024-01-23

(272 days)

Product Code
Regulation Number
866.3982
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nano-Check™ COVID-19 Antigen Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of SARS-CoV-2 nucleoprotein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 4 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Nano-Check™ COVID-19 Antigen Test and followed up with a molecular test.

The test does not differentiate between SARS-CoV or SARS-CoV-2.

A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Performance characteristics for SARS-CoV-2 were established during the 2022-2023 SARS-CoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

Device Description

The Nano-Check™ COVID-19 Antigen Test is an immunochromatographic assay for detection of extracted SARS-CoV-2 nucleoprotein antigens in human anterior nasal swab specimens.

The assay kit consists of 20 test cassette devices, 20 reagent tubes containing extraction buffer for Set A or 20 empty reagent tubes and 20 ampules containing extraction buffer for Set B, 20 anterior nasal specimen collection swabs, one positive control swab, one negative control swab and an Instructions for Use/Quick Reference Instruction. An external positive control swab is used with recombinant SARS-CoV-2 nucleocapsid dried onto the swab and an external negative control swab contains blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the Nano-Check COVID-19 Antigen Test meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for clinical performance are implicitly derived from the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). While explicit numerical acceptance targets are not stated, the device's performance is presented in comparison to a molecular RT-PCR test.

MetricAcceptance Criteria (Implicit)Reported Device Performance
Positive Percent Agreement (PPA)Acceptable level for aiding in diagnosis of SARS-CoV-2 infections in symptomatic individuals83.67% (95% CI: 76.86-88.78%)
Negative Percent Agreement (NPA)Acceptable level for ruling out SARS-CoV-2 infections, with a recommendation for molecular confirmation99.62% (95% CI: 98.62-99.90%)

For analytical performance, acceptance criteria are generally "no interference" for cross-reactivity and endogenous/exogenous interference, and "100% agreement" for precision and reproducibility at moderate to high concentrations.

Metric (Analytical)Acceptance Criteria (Implicit)Reported Device Performance
Repeatability (MP & TN samples)100% agreement100% agreement
Repeatability (LP & HN samples)>95% agreement>95% agreement (95.8% for LP, 96.9% for HN)
Reproducibility (All samples across sites)No significant difference in performance between sitesAll samples tested generated no significant difference
Cross-Reactivity/Microbial InterferenceNo cross-reactivity/interference"No" for all 43 microorganisms/substances tested
Endogenous/Exogenous InterferenceNo interference with positive or negative results"No" for all 20 substances tested
Biotin InterferenceNo interference up to the concentration testedNo interference up to 3,500 ng/mL
External Controls Lot-to-Lot Reproducibility100% agreement with expected results100% agreement for all lots
Specimen Stability (2x LoD positive samples)Stable for recommended storage durationStable up to 24 hours at room temperature, 48 hours at 2°C-8°C
Limit of Detection (LoD)Established LoD for relevant variants1.95x10^2 TCID50/mL for Omicron (e.g.)
Analytical Reactivity (Inclusivity)100% detection at a certain viral concentration100% detection (5/5) at listed LoD for 15 SARS-CoV-2 isolates/variants
High Dose Hook EffectNo Hook Effect100% positive results across tested high concentrations

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Clinical Study Test Set): 670 subjects
  • Data Provenance: Multi-center, prospective clinical study in the U.S. in 2022-2023. At the time of the study, the SARS-CoV-2 Omicron variant was the predominant strain.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical study. The ground truth was established using an "FDA Cleared SARS-CoV-2 RT-PCR Test as comparator." RT-PCR tests are molecular diagnostic methods and are considered the gold standard for SARS-CoV-2 detection, generally not requiring expert interpretation for generating positive/negative results.

4. Adjudication Method for the Test Set

The document does not explicitly state an adjudication method (like 2+1, 3+1, none) for discordant results between the Nano-Check COVID-19 Antigen Test and the RT-PCR comparator. The results are presented directly from this comparison.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no mention of AI or human readers being assisted by AI in this context. This device is a visually-read, lateral flow immunochromatographic assay, meaning results are interpreted directly by the user looking at control and test lines.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a visually-read assay. There is no algorithm for standalone performance. The interpretation is directly by the user.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth used for the clinical study was an FDA Cleared SARS-CoV-2 RT-PCR Test. This is considered the molecular "gold standard" for detecting the presence of SARS-CoV-2 viral RNA.

8. The Sample Size for the Training Set

The document does not report a sample size for a training set. This is expected as the Nano-Check COVID-19 Antigen Test is a traditional immunoassay, not an AI/machine learning-based device that typically requires a training set. The "training" of such a device is in its biochemical formulation and physical design, not in data-driven algorithm optimization.

9. How the Ground Truth for the Training Set was Established

As no training set is mentioned or applicable for this type of device, this information is not provided.

{0}------------------------------------------------

January 23, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Nano-Ditech Corporation James Chang QA/RA Director 259 Prospect Plains Rd, Bldg K Cranbury, New Jersey 08512

Re: K231187

Trade/Device Name: Nano-Check COVID-19 Antigen Test Regulation Number: 21 CFR 21 CFR 866.3982 Regulation Name: Simple Point-Of-Care Device To Directly Detect Sars-Cov-2 Viral Target From Clinical Specimens In Near-Patient Setting Regulatory Class: Class II Product Code: QVF Dated: January 10, 2024 Received: January 10, 2024

Dear James Chang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the name "Silke Schlottmann -S" in large font on the left side of the image. On the right side of the image, the text "Digitally signed by Silke Schlottmann -S" is present. Below this, the date and time "Date: 2024.01.23 15:46:32 -05'00'" is displayed, indicating a digital signature timestamp.

Silke Schlottmann, Ph.D. Deputy Assistant Director Bacteriology Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231187

Device Name

Nano-Check COVID-19 Antigen Test

Indications for Use (Describe)

The Nano-Check COVID-19 Antigen Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of SARS-CoV-2 nucleoprotein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 4 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice days with at least 48 hours between tests; or when tested once, and negative by the Nano-Check™ COVID-19 Antigen Test and followed up with a molecular test.

The test does not differentiate between SARS-CoV or SARS-CoV-2.

A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Performance characteristics for SARS-CoV-2 were established during the 2022-2023 SARSCoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

Type of Use ( Select one or both, as applicable )
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Nano-Check™ COVID-19 Antigen Test (K231187)

I. Submitter

Nano-Ditech Corporation 259 Prospect Plains Rd., Bldg. K Cranbury, NJ 08512 USA

Contact Person: James B Chang (jchang@nanoditech.com) Phone: 1-609-409-3300 Date Prepared: June 29, 2023

II. Device

  • A. Trade, Proprietary, and Established Name: Nano-Check™ COVID-19 Antigen Test (Common name: Rapid COVID-19 Antigen Test)
  • B. Product Code: OVF
  • C. Classification: Class II
  • D. Regulation Number: 21 CFR 866.3982
  • E. Classification Name: Simple Point-Of-Care Device to directly detect SARS-CoV-2 viral target from clinical specimens in Near-Patient setting
  • F. Panel: MI-Microbiology
  • G. Purpose for Submission: New device 510(k) clearance for the Nano-Check™ COVID-19 Antigen Test
  • H. Measurand: Nucleocapsid protein antigen from SARS-Coronavirus 2 (SARS-CoV-2)
  • I. Type of Test: Qualitative lateral flow immunoassay

III. Predicate Device

  • Trade Name: Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set A.
  • De Novo Number: DEN220039 (granted March 8, 2023) B.

{4}------------------------------------------------

Product Code: OVF C.

This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

IV. Indications for Use

A. Intended Use(s):

See Indications for Use below.

B. Indication(s) for Use:

The Nano-Check™ COVID-19 Antigen Test is a lateral flow immunochromatographic assay for the rapid, qualitative detection of SARS-CoV-2 nucleoprotein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 4 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Nano-Check™ COVID-19 Antigen Test and followed up with a molecular test.

The test does not differentiate between SARS-CoV or SARS-CoV-2.

A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Positive results do not rule out co-infection with other bacteria or viruses and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Performance characteristics for SARS-CoV-2 were established during the 2022-2023 SARS-CoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

C. Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

  • D. Special Instrument Requirements:
    N/A

V. Device/System Characteristics:

A. Device Description :

The Nano-Check™ COVID-19 Antigen Test is an immunochromatographic assay for detection of extracted SARS-CoV-2 nucleoprotein antigens in human anterior nasal swab specimens.

{5}------------------------------------------------

The assay kit consists of 20 test cassette devices, 20 reagent tubes containing extraction buffer for Set A or 20 empty reagent tubes and 20 ampules containing extraction buffer for Set B, 20 anterior nasal specimen collection swabs, one positive control swab, one negative control swab and an Instructions for Use/Quick Reference Instruction. An external positive control swab is used with recombinant SARS-CoV-2 nucleocapsid dried onto the swab and an external negative control swab contains blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.

B. Principles of Operations :

The Nano-Check™ COVID-19 Antigen Test is based on antigen-antibody binding technology.

The viral antigens are released from the specimen during the extraction step. Two drops of the extracted sample are added to the cassette that contains a testing strip. During the migration through the strip, the viral antigens bind to anti-SARS-CoV-2 antibodies conjugated to indicator particles.

The antigen-antibody complexes migrate through the test strip and are captured by reagents on the membrane to form test lines. There are a test line detecting SARS-CoV-2 antigens and a control line serving as the internal control for the assay.

The test is visually read 15 minutes after the sample is added to the cassette.

Results Interpretation:

  • · The presence of two lines in the cassette, one for SARS-CoV-2 next to the letter "Ag" and one control line next to the letter "C", indicates SARS-CoV-2 positive result.
  • · The presence of the control line in the cassette next to the letter "C" without the test line indicates a negative result.
  • · The absence of a "C" line, regardless of whether the test line is present or absent, represents an invalid result.

VI. Comparison of Technological Characteristics with the Predicate Device

The table below compares the Nano-Check™ COVID-19 Antigen Test to the predicate devices of Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set.

Similarities and Differences
ItemsCandidatePredicate (DEN220039)
Device TradeNameNano-Check™ COVID-19 Antigen TestSofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set
The Nano-Check™ COVID-19 AntigenTest is a lateral flowimmunochromatographic assay for therapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigensdirectly in anterior nasal swab specimensfrom individuals with signs andsymptoms of upper respiratory infection(i.e., symptomatic) when testing is startedwithin 4 days of symptom onset.The Sofia 2 SARS Antigen+ FIA is alateral flow immunofluorescent sandwichassay that is used with the Sofia 2instrument for the rapid, qualitativedetection of SARS-CoV-2 nucleocapsidprotein antigens directly in anterior nasalswab specimens from individuals withsigns and symptoms of upper respiratoryinfection (i.e., symptomatic) when testingis started within 6 days of symptomonset.
Intended Use/Indication ForUseThe test is intended for use as an aid inthe diagnosis of SARS-CoV-2 infections(COVID-19) in symptomatic individualswhen either: tested at least twice overthree days with at least 48 hours betweentests; or when tested once, and negativeby the Nano-Check™ COVID-19Antigen Test and followed up with amolecular test.The test is intended for use as an aid inthe diagnosis of SARS-CoV-2 infections(COVID-19) in symptomatic individualswhen tested at least twice over three dayswith at least 48 hours between tests.
The test does not differentiate betweenSARS-CoV and SARS-CoV-2.The test does not differentiate betweenSARS-CoV and SARS-CoV-2.
A negative test result is presumptive, andit is recommended these results beconfirmed by a molecular SARS-CoV-2assay. Positive results do not rule out co-infection with other bacteria or virusesand should not be used as the sole basisfor diagnosis, treatment, or other patientmanagement decisions.A negative test result is presumptive, andit is recommended these results beconfirmed by a molecular SARS-CoV-2assay. Negative results do not precludeSARS-CoV-2 infections and should notbe used as the sole basis for treatment orother patient management decisions.Positive results do not rule out co-infection with other respiratorypathogens.
Performance characteristics for SARS-CoV-2 were established during the 2022-2023 SARS-CoV-2 pandemic whenSARS-CoV-2 Omicron was thepredominant SARS-CoV-2 variant incirculation. When other SARS-CoV-2virus variant are emerging, performancecharacteristics may vary.Performance characteristics for SARS-CoV-2 were established during the 2021-2022 SARS-CoV-2 pandemic whenSARS-CoV-2 Omicron was thepredominant SARS-CoV-2 variant incirculation. When other SARS-CoV-2virus variant are emerging, performancecharacteristics may vary.This test is intended for prescription useonly and can be used in Point-of-Caresettings.
Sample TypeAnterior nasal swab
Test ResultQualitativeSame
Test TargetNucleocapsid protein of SARS-CoV-2Same
Test Design/PrincipleLateral flow immunochromatographicSame
Test Time15 minutesSame
ReadingResultVisual determination of the presence orabsence of colored line(s) for the testline(s) and a colored line on the test stripindicate the presence of SARS-CoV-2Sofia 2 scans the test strip and measuresthe fluorescent signal by processing theresults using method specific algorithms.Sofia 2 displays the test results (Positive,Negative, or Invalid) on the screen.

Table 1. Comparison with Predicate(s)

{6}------------------------------------------------

{7}------------------------------------------------

VII. Performance Data

A. Analytical Performance

    1. Precision/Reproducibility:
  • a) Precision and Repeatability

The repeatability study was undertaken by two operators with each operator conducting 24 test runs over a 12-days period, resulting in total 96 replicates per level. The study employed blind sample panels comprising true negative (TN), high negative (HN, 0.1× LoD), low positive (LP, 1× LoD), and moderate positive (MP, 4× LoD) samples. The precision testing was conducted according to the Package Insert and the qualitative results are summarized in Table 2 below.

Table 2. Precision and Repeatability Study-Summary Result

RunTrue negativeHigh NegativeLow PositiveModerate Positive
12121212
Run 124/2424/2424/2423/2424/2423/2424/2424/24
Run 224/2424/2423/2423/2422/2423/2424/2424/24
Total48/4848/4847/4846/4846/4846/4848/4848/48
%Agreement100%100%97.9%95.8%95.8%95.8%100%100%
95% CI[92.6 -100%][92.6 -100%][89.1 -99.6%][86.0 -98.9%][86.0 -98.9%][86.0 -98.9%][92.6 -100%][92.6 -100%]
%Agreement100%96.9%95.8%100%
95% CI96.2 - 100%91.2 - 98.9%89.8 - 98.4%96.2 - 100%

The study results demonstrated a 100% agreement between the test results and sample properties for MP samples and TN samples. Additionally, there was over 95% agreement for LP samples and HN samples.

{8}------------------------------------------------

b) Reproducibility

The reproducibility study was designed to evaluate site-to-site, operator, and level-tolevel variability and demonstrate that the Nano-Check™ COVID-19 Antigen Test can be performed consistently and correctly. This study was conducted at 4 distinct sites, each with two (2) or three (3) different operators, testing three (3) lest lots, using a coded panel of contrived samples consisting of a true negative (TN), a high negative sample (HN, 0.1x LoD), a low positive (LP, 1.0 x LoD) and a medium positive (MP, 3.0 x LoD). A total of 7 assigned operators at 3 POC sites and 3 operators in the internal site each tested using 60 coded samples (TN: 15, HN: 15, LP: 15 and MP:15 samples, respectively) over five nonconsecutive days. The summary of results is presented in Table 3 below.

SampleSite 1Agreement(Count)Site 2Agreement(Count)Site 3Agreement(Count)Site 4Agreement(Count)All SitesAgreement(Count)95% CI
TrueNegative100%(45/45)100%(30/30)100%(30/30)100%(45/45)100%(150/150)97.5-100.0
HighNegative100%(45/45)100%(30/30)100%(30/30)100%(45/45)100%(150/150)97.5-100.0
LowPositive93.3%(42/45)100%(30/30)100%(30/30)100%(45/45)95%(147/150)94.3-99.3
MediumPositive100%(45/45)100%(30/30)100%(30/30)100%(45/45)100%(150/150)97.5-100.0

Table 3. Reproducibility Study- Summary Result

All samples tested in the reproducibility study generated no significant difference in performance between sites.

2. Linearity:

This section is not applicable as this test device is a qualitative assay.

    1. Cross-reactivity (Analytical Specificity):
  • a) Cross-reactivity and Microbial Interference Study

Cross-Reactivity and Interference study of the Nano-Check™ COVID-19 Antigen Test was conducted with 43 potential cross-reactive or interfering pathogens of bacteria (10), fungi (1), virus (31), and negative matrix (1). Each microorganism was tested in three (3) replicates in the absence or presence of 2xLoD of SARS-CoV-2 Omicron Variant (hCoV-19/USA/MD-HP20874/2021). The concentrations of potentially interfering microorganisms tested and the results from the crossreactivity study are presented in Table 4 below.

{9}------------------------------------------------

Microorganism (Strain)ConcentrationTestedPositive Sample(# Positive/# Tested)Negative Sample(# Positive/# Tested)Cross-Reactivity/MicrobialInterference
Human Coronavirus, 229E$1.0\times10^5$TCID50/mL3/30/3No
Human Coronavirus, OC43$4.45\times10^4$TCID50/mL3/30/3No
Coronavirus, NL63$1.0\times10^5$TCID50/mL3/30/3No
MERS-CoV, EMC/2012,$1.0\times10^5$TCID50/mL3/30/3No
SARS-CoV, Urbani$1.0\times10^5$PFU/mL3/30/3No
Adenovirus Type 2, C$1.0\times10^5$TCID/mL3/30/3No
Human Adenovirus 1$1.0\times10^5$TCID50/mL3/30/3No
Human Mastadenovirus B, Type7$1.0\times10^5$TCID50/mL3/30/3No
Human Metapneumovirus, TN/83-1211$1.0\times10^5$TCID50/mL3/30/3No
Human Parainfluenza Virus 1,HPIV1/FRA/29221106/2009$1.0\times10^5$TCID50/mL3/30/3No
Human Parainfluenza Virus 2,Greer$1.0\times10^5$TCID50/mL3/30/3No
Human Parainfluenza Virus 3,NIH 47885$1.0\times10^5$TCID50/mL3/30/3No
Human Parainfluenza Virus 4B,19503$1.0\times10^5$TCID50/mL3/30/3No
Influenza A Virus/NewCaledonia/20/1999 (H1N1)$1.0\times10^5$CEID50/mL3/30/3No
Influenza A Virus/A/SanDiego/1/2009 (H1N1) pdm09$1.0\times10^5$TCID50/mL3/30/3No
Influenza A Virus/A/Victoria/361/2011 (H3N2)$1.0\times10^5$CEID50/mL3/30/3No
Influenza A Virus/A/Wisconsin/67/2005 (H3N2)$1.0\times10^5$CEID50/mL3/30/3No
Influenza B Virus/B/Brisbane/60/2008 (Victoria Lineage)$1.0\times10^5$CEID50/mL3/30/3No
Influenza B Virus/B/Texas/06/2011 (Yamagata Lineage)$1.0\times10^5$CEID50/mL3/30/3No
Influenza B Virus/B/GL/1739/54$1.0\times10^5$CEID50/mL3/30/3No
Enterovirus 71 (EV-71), MP4$1.0\times10^5$TCID50/mL3/30/3No
Enterovirus Species D Type 68,USA/2018-23087$1.0\times10^5$TCID50/mL3/30/3No
Human Respiratory SyncytialVirus, A/Long$1.0\times10^5$PFU/mL3/30/3No
Human Respiratory SyncytialVirus A/2001/2-20$1.0\times10^5$TCID50/mL3/30/3No
Human Respiratory SyncytialVirus B/PR-348-00 MISC082$1.0\times10^5$TCID50/mL3/30/3No
Human Respiratory SyncytialVirus A/1998/12-21$1.0\times10^5$TCID50/mL3/30/3No
Rhinovirus 20, 15-CV19$1.0\times10^5$TCID50/mL3/30/3No
Measles Virus, Edmonston$1.7\times10^4$TCID50/mL3/30/3No
Mumps Virus, MuV/Iowa.US/2006$1.0\times10^5$TCID50/mL3/30/3No
Coxsackie Virus B4$1\times10^5$U/mL TCID503/30/3No
Epstein-Barr (Strain: B95-8)$1.0\times10^5$cp/mL3/30/3No
Haemophilus influenzae,(Type b; Eagan)$1.0\times10^6$CFU/mL3/30/3No
Streptococcus pneumoniae,(Strain Z022, 19F)$1.0\times10^6$CFU/mL3/30/3No
Streptococcus pyogenesStrain Bruno$1.0\times10^6$CFU/mL3/30/3No
Streptococcus salivarius$1.0\times10^6$CFU/mL3/30/3No
Candida albiccans, Z006$1.0\times10^6$CFU /mL3/30/3No
Bordetella pertussis$1.0\times10^6$CFU /mL3/30/3No
Mycoplasma pneumoniae$1.0\times10^6$CFU /mL3/30/3No
Chlamydophila pneumoniae$1.0\times10^6$IFU/mL3/30/3No
Legionella pneumophila subsp.pneumophila$1.0\times10^6$CFU /mL3/30/3No
Staphylococcus aureus(Strain MRSA; COL)$1.0\times10^6$CFU/mL3/30/3No
Staphylococcus epidermidis(Strain MRSE; RP62A)$1.0\times10^6$CFU/mL3/30/3No
Pooled human nasal washN/A3/30/3No
Coronavirus HKU1 was not tested for cross-reactivity due to lack of availability. Twenty (20) clinicalspecimens containing Coronavirus HKU1 were tested, and all resulted as negative, additional cross-reactivity wet testing was not required.

Table 4. Results of Cross-Reactivity/Microbial Interference

{10}------------------------------------------------

reactivity wet testing was not required.

{11}------------------------------------------------

The study results showed that the Nano-Check™ COVID-19 Antigen Test would not react with nor be interfered by listed pathogens, high prevalence disease agents, and normal or pathogenic flora that are reasonably likely to be encountered in the clinical specimen.

b) Endogenous/Exogenous Interference

The Endogenous/Exogenous interference study of the Nano-Check™ COVID-19 Antigen Test was conducted using medically relevant endogenous and exogenous interferents. The SARS-CoV-2 Omicron Variant (hCoV-19/USA/MD-HP20874/2021) was diluted in negative clinical matrix to 3x LoD and the potentially interfering substance was spiked to the concentration presented in Table 5 below. The effects of the substances were evaluated by the agreement with the expected negative or positive results. Three (3) replicates were tested for each sample prepared.

SubstancesActive ingredientsInterferent Conc.Interference
Beclomethasone dipropionateBeclomethasonedipropionate$5.04×10^0$ µg/mLNo
Budesonide Nasal SprayBudesonide15% v/vNo
DexamethasoneDexamethasone$1.20×10^1$ µg/mLNo
FlunisolideFlunisolide$8.70×10^1$ µg/mLNo
Hand Sanitizer GelEthyl alcohol 70%1% v/vNo
Hand Soap Fresh Breeze ScentN/A10% v/vNo
Histaminum 30CHistaminum hydrochloricum,HPUS5% w/vNo
Homeopathic Allergy Nasal SprayN/A15% v/vNo
MolnupiravirMolnupiravir$4.20×10^4$ µg/mLNo
Mucin (Bovine submaxillaryGlands, Type I-S)Mucin protein2.5 mg/mLNo
MupirocinMupirocin$1.50×10^0$ µg/mLNo
Nasacort Allergy 24HRTriamcinolone acetonide15% v/vNo
Nasal Allergy ReliefCardiospermun, Galphimiaglauca, Luffa operculateSabadilla15% v/vNo
Nasal DropsXylometazoline15% v/vNo
Nasal Spray 1Oxymetazoline15% v/vNo
Nasal Spray 2Cromolyn sodium15% v/vNo
Nasal Spray 3Fluticasone propionate15% v/vNo
Nasal Spray 4Phenylephrine HCl15% v/vNo
NasoGEL (Gel Spray)Sodium Hyaluronate,Allantoin, Sodium chloride,Methylparaben,Propylparaben15% v/vNo

Table 5. Results of Endogenous/Exogenous Interference Study

{12}------------------------------------------------

Nasonex 24 hr AllergyMometasone furoatemonohydrate15% v/vNo
Oseltamivir PhosphateOseltamivir Phosphate$3.99\times10^{-1}$ µg/mLNo
Human NeutrophilsN/A$5\times10^{6}$ cells/mLNo
RemdesivirRemdesivir$2.40\times10^{3}$ µg/mLNo
Sore Throat(Oral Pain Reliever spay)Phenol, Menthol15% v/vNo
Sore throat LozengesBenzocaine, Menthol15% w/vNo
TobramycinTobramycin$3.30\times10^{1}$ µg/mLNo
Whole BloodN/A2.5%No
ZICAM® Oral mistZincum aceticum,Zincum gluconicum15% v/vNo
Zinc LozengesZincum gluconicum5% w/vNo

The study results showed no interference with positive or negative results for the Nano-Check™ COVID-19 Antigen Test at the concentrations tested in the presented study.

c) Biotin Interference

The samples containing 3 x LoD concentrations of SARS-CoV2 (2019-nCoV/USA-WA1/2020) at different concentrations of biotin were tested in 5 replications. The Biotin Interference Study for Nano-Check™ COVID-19 Antigen Test showed that there was no interference up to 3,500 ng/mL of biotin in the samples so no limitation in the labeling regarding the biotin interference was necessary.

4. Assay Reportable Range:

This section is not applicable as this test device is a qualitative assay.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

a) Internal Quality Control

The Nano-Check™ COVID-19 Antigen Test contains built in internal assay control. The appearance of the control line on the test ensures that sufficient flow of the sample occurred during the assay.

b) External Quality Controls

The Nano-Check™ COVID-19 Antigen Test kit contains one positive external control swab and one negative external control swab that allow for monitoring of the performance of the assay. The positive

{13}------------------------------------------------

control swab contains recombinant SARS-CoV-2 nucleocapsid and negative control swab contains blank Universal Viral Transport media.

c) External Controls Lot-to-Lot Reproducibility

The lot-to-lot reproducibility for the external positive and negative control swabs was evaluated using three lots of external controls with three operators. The results from the study are summarized in Table 6 below.

External controlLot No.Test result (Test count / Total replication)
Positive countNegative countAgreement
PositiveExternal Control0804-CN-2B01-P10 / 100 / 10100%
0804-CN-2D29-P10 / 100 / 10100%
0804-CN-2E04-P10 / 100 / 10100%
NegativeExternal Control0804-CN-2B01-N0 / 1010 / 10100%
0804-CN-2D29-N0 / 1010 / 10100%
0804-CN-2E04-N0 / 1010 / 10100%
Table 6. External Controls Lot-to-Lot Reproducibility Study Results

All lots of external controls generated 100% agreement with the expected results.

d) Specimen Stability

Two test samples were prepared for the specimen stability study: a negative sample (consisting of pooled human nasal wash with no analyte) and a low positive sample (containing a diluted SARS-CoV-2 Omicron Variant, hCoV-19/USA/MD-HP20874/2021, sample at 2 x LoD in the negative pooled human nasal wash). Swabs were spiked with 50 uL of each sample and subjected to two different temperature conditions: room temperature (23.5℃), high room temperature (30 ℃) and 2℃ to 8℃. The samples were exposed for various time intervals including 0 hours, 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, and 48 hours. Subsequently, the exposed sample swabs were tested using 5 replicates for each exposure time, as shown in the table 7 below.

Specimen StorageTemperatureTestedSampleSpecimen Storage Time (No. of Positive/No. of Total Test)
0 hour1 hour2 hours4 hours8 hours24 hours48 hours
23.5°C2x LoD5/55/55/55/55/55/53/5
Negative0/50/50/50/50/50/50/5
30°C2x LoD5/55/55/55/55/55/52/5
Negative0/50/50/50/50/50/50/5
2°C to 8°C2x LoD5/55/55/55/55/55/55/5

Table 7. Specimen Stability Results

{14}------------------------------------------------

Negative0/50/50/50/50/50/50/5
-----------------------------------------------

The results showed that the SARS-CoV-2 positive sample swabs were stable up 24 hours at room temperature and up to 48 hours at 2°C to 8°C. Therefore, freshly collected specimen swabs are recommended to be processed no later than one hour after specimen collection at room temperature or before 48 hours when stored at 2℃ to 8℃.

6. Detection Limit:

The Limit of Detection (LoD) was determined with gamma-irradiated SARS-CoV-2 (isolate USA-WA1/2020), heat-inactivated SARS-CoV2 Omicron (isolate hCoV-19/USA/MD-HP20874/2021) and UV inactivated SARS-CoV-2 Delta (isolate USA/PHC658/2021). Contrived samples were prepared by spiking the strain into pooled negative nasal wash matrix. A preliminary LoD was determined by spiking 50 µL of serially diluted sample onto swab heads and tested using the Nano-Check™ COVID-19 Antigen Test. The preliminary LoD initially determined by testing two-fold serial dilution series of 3 replicates was confirmed by testing in 20 replicates. Based on the testing procedure for this study the LoD was determined to be 7.00×102 TCID56/mL (USA-WA1/2020), 1.95×102 TCIDso/mL (hCoV-19/USA/MD-HP20874/2021), and 5.21×102 TCID50/mL (USA/PHC 658/2021), equating to 3.50×101 TCIDso/swab, 9.75×100 TCIDso/swab and 2.61×101 TCIDs0 swab, respectively

SARS-CoV-2VariantStrainLoD(TCID50/mL)Positive Rate (%)(# Positive/# Tested)
Wild2019-nCoV/USA-WA1/2020$7.00\times10^2$95% (19/20)
OmicronhCoV-19/USA/MD-HP20874/2021$1.95\times10^2$95% (19/20)
DeltaUSA/PHC658/2021$5.21\times10^2$95% (19/20)

Table 8. Limit of Detection (LoD)

Furthermore, the LoD was established using the 1st WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in real clinical matrix of nasal fluid. Initially, a preliminary LoD test was performed by spiking 50 uL of each diluted sample collection swab head in three replicates. Following this, a confirmatory LoD test with 17 additional replicates was conducted at the preliminary LoD concentration, a total of 20 replications. To determine LoD concentration, additional dilutions were investigated bracketing the confirmed LoD concentration, with each dilution level undergoing 20 replications. It was determined that the LoD of the 18' WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in nasal matrix was confirmed to be 889 IU/mL (44 IU/swab).

7. Analytical Reactivity (Inclusivity):

{15}------------------------------------------------

A total of 15 SARS-CoV-2 isolates were tested in the inclusivity study. 10-fold dilutions of virus stocks were prepared in pooled negative nasal wash matrix and 50 µL were pipetted onto the swabs. The swabs were tested in 5 replicates in this study. The lowest dilution of each strain that resulted in 100% detection (5/5) is presented in Table 9 below.

Lineage/VariantStrainLoD# Positive/# Tested
WildUSA-WA1/2020$4.79 \times 10^4$ TCID50/mL5/5
B.1.1.7(Alpha)USA/CA_CDC_5574/2020$6.77 \times 10^6$ genome equivalency/mL5/5
England/204820464/2020$7.19 \times 10^3$ TCID50/mL5/5
USA/CA_CDC_5574/2020$2.39 \times 10^4$ TCID50/mL5/5
B.1.351(Beta)hCoV-19/USA/MD-HP01542/2021$7.20 \times 10^4$ genome equivalency/mL5/5
USA/MD-HP01542/2021$3.80 \times 10^6$ genome equivalency/mL5/5
South Africa/KRISP-K005325/2020$1.90 \times 10^4$ TCID50/mL5/5
P.1(Gamma)Japan/TY7-503/2021$1.58 \times 10^4$ TCID50/mL5/5
USA/NY-Wadsworth-21033899-01/2021$7.85 \times 10^3$ TCID50/mL5/5
B.1.617.2(Delta)hCoV-19/USA/MD-HP05285/2021$7.20 \times 10^7$ genome equivalency/mL5/5
USA/PHC658/2021$5.21 \times 10^2$ TCID50/mL5/5
USA/MD-HP05285/2021$5.00 \times 10^3$ genome copies/mL5/5
B.1.617.1(Kappa)USA/CA-Stanford-15_S02/2021$8.48 \times 10^4$ TCID50/mL5/5
B.1.1.529(Omicron)hCoV-19/USA/MD-HP20874/2021$1.95 \times 10^2$ TCID50/mL5/5
USA/GA-EHC-2811C/2021$2.11 \times 10^4$ genome copies/mL5/5

Table 9. Analytical Reactivity

8. High Dose Hook Effect:

High-dose hook effect study was conducted to determine if a hook effect would be observed at high concentrations of the analyte (i.e., a false negative at high concentrations of SARS-CoV-2) in a testing sample. The series of five concentrations were prepared in pooled negative nasal wash matrix and tested in three replicates on Nano-Check™ COVID-19 Antigen Test, between and including 1x LoD (7.0 x 102 TCIDso'mL) and to its stock concentration of 2.8 x 106 TCIDso'mL. All spiked samples were 100% positive, as expected, at all tested concentrations. The Nano-Check™ COVID-19 Antigen Test did not display a Hook Effect for high concentrations of SARS-CoV-2 tested herein.

Table 10. High Dose Hook Effect Test

{16}------------------------------------------------

Concentration of SARS-CoV-2 (TCID50/mL)# Positive/# Tested
$2.8\times10^6$3/3
$2.8\times10^5$3/3
$2.8\times10^4$3/3
$2.8\times10^3$3/3
$7.0\times10^2$3/3

9. Assay Cut-off:

Not Applicable

10. Carry-Over:

Carry-over contamination is not applicable to this test device as each sample uses an independent, new, single-use test cassette that is discarded after each run.

B. Comparison Studies:

1. Method Comparison with Predicate Device:

Not applicable. The performance of the Nano-Check™ COVID-19 Antigen Test was evaluated in a clinical study against an FDA-cleared molecular assay.

2. Matrix Comparison:

The Nano-Check™ COVID-19 Antigen Test is only intended for the qualitative detection of the nucleocapsid protein antigen from SARS-Co V-2 in direct anterior nasal swab specimens. As no other specimen or sample type is claimed herein.

C. Clinical Study:

The clinical performance of the Nano-Check™ COVID-19 Antigen Test was evaluated in a multicenter, prospective clinical study in the U.S. in 2022-2023, during a period when the SARS-CoV-2 Omicron variant was the predominant strain. The study only enrolled subjects with symptoms of respiratory infection consistent with a SARS-CoV-2 infection. A total of six hundred seventy (670) subjects were consecutively enrolled and tested by eleven (11) operators across four (4) different Point of Care (POC) CLIA-waived sites. Two anterior nasal swabs were collected from each study subject during the same visit. The first AN swab sample for the comparator method was collected by the operators using the swab included in a Universal Viral Transport from both sides of the nose. The collected AN swab was then added into the UVT media containing 1 mL media and stored in dry ice until it was shipped to a reference laboratory and tested with an FDA Cleared SARS-CoV-2 RT-PCR Test as comparator. The second AN swab sample was collected from both

{17}------------------------------------------------

sides of the nose using the provided swab and was tested immediately using the Nano-Check™ COVID-19 Antigen Test by an operator at the site.

There were 670 subjects evaluated, with 24.9% collected from patients aged 21 years or younger, 59.7% from patients aged 22-60 years, and 15.4% from patients aged 61 years or older. Among the subjects, 59.6% were from female patients, while 40.4% were from male patients. Results obtained with the Nano-Check™ COVID-19 Antigen Test were compared to the results obtained with the RT-PCR comparator test to determine clinical sensitivity and specificity. The study cohort included 21.1% low-positive samples. The Nano-Check™ COVID-19 Antigen Test demonstrated a Positive Percent Agreement of 83.67 % (123/147, 95% CI: 76.86-88.78%) and Negative Percent Agreement 99.62% (521/523, 95% CI: 98.62-99.90%).

Table 11. Clinical Performance of Nano-Check™ COVID-19 Antigen Test Compared to an RT-
PCR Comparator
Nano-Check™ COVID-19Antigen TestRT-PCR ComparatorTotal
PositiveNegative
Positive1232125
Negative24521545
Total147523670
Positive Percent Agreement (PPA) = 83.67% (123/147, 95% CI: 76.86-88.78%)
Negative Percent Agreement (NPA) = 99.62% (521/523, 95% CI: 98.62-99.90%)
Positivity in Study Cohort = 21.94% (147/670; 95% CI: 18.97-25.23%)

1. Clinical Sensitivity:

Please refer to Section VI.C (Clinical Studies) above for the clinical validation. The PPA for the test is 83.67% (123/147, 95% CI: 76.86-88.78%).

2. Clinical Specificity:

Please refer to Section VI.C (Clinical Studies) above for the clinical validation. The NPA for the test is 99.62% (521/523, 95% CI: 98.62-99.90%).

3. Serial Testing:

As a mitigation for low performance of the device at Day 0 of symptom onset., as indicated by these clinical study data and in other studies for test devices of a similar principle and design, the Intended Use for this test device (and associated Instructions for Use) include recommendations for repeat testing (i.e., test at least twice over three days with at least 48 hours between tests.). This mitigation is supported by data generated by the National Institutes for Health (NIH) and the University of Massachusetts Chan Medical School (in collaboration with the FDA) demonstrating that repeat testing over multiple days improves test performance and increases the likelihood that a COVID-19 antigen test will accurately detect an infection. These results have informed the FDA's general understanding that repeat testing after a negative result from a COVID-19 antigen

Page | 15

{18}------------------------------------------------

test reduces the risk of a false negative result. Please refer to the following studies for additional details:

D. Clinical Cut-Off:

Not Applicable since there is no clinical cutoff related to the presence of SARS-CoV-2 in patient samples.

E. Expected Value/Reference Range:

Patient Samples are expected to be negative for COVID-19 Antigen.

VIII. Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

N/A