K Number
K232377

Validate with FDA (Live)

Date Cleared
2024-04-19

(255 days)

Product Code
Regulation Number
866.3982
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs specimens from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test
The test does not differentiate between SARS-CoV and SARS-CoV-2.
A negative test result is presumptive and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.
Positive results do not rule out co-infection with other respiratory pathogens
Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

Device Description

The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay that uses highly sensitive monoclonal antibodies to detect nucleocapsid protein from SARS-CoV-2 virus in nasal swab collected specimens. The test strip is composed of the following components: sample pad, reagent pad, reaction membrane, and absorbing pad housed within a test cassette. The reagent pad contains colloidal-gold-conjugated monoclonal antibody that recognizes and binds to the nucleocapsid protein of SARS-CoV-2; the reaction membrane in the test line (T) contains the second antibody that recognizes another epitope of the nucleocapsid protein of SARS-CoV-2.
External quality controls are required but not included with the test kit and are sold separately as the COVID-19 Antigen Control Kit. The control swabs should be processed according to the Instructions for Use (IFU) and are intended to be used as quality control samples to demonstrate that the test is performing and is being performed correctly.

AI/ML Overview

The provided text describes the performance characteristics and acceptance criteria for the Healgen Rapid COVID-19 Antigen Test.

Here's an analysis to extract the requested information:

1. A table of acceptance criteria and the reported device performance:

The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds in a table. Instead, it presents the results of various performance studies. The implicit acceptance criteria are that the device performance should be clinically acceptable and comparable to existing devices for substantial equivalence.

Here's a table summarizing the reported device performance, which can be interpreted as having met the (implicit) acceptance criteria:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Clinical Performance:
Positive Percent Agreement (PPA)Clinically acceptable85.4% (95% CI: 79.1% - 90%)
Negative Percent Agreement (NPA)Clinically acceptable99.7% (95% CI: 98.9% - 99.9%)
Analytical Sensitivity (LoD):
LoD (SARS-CoV-2 USA-WA1/2020)Lowest detectable concentration5.75 x 10³ TCID₅₀/mL (2.875 x 10² TCID₅₀/swab)
LoD (WHO International Standard)Lowest detectable concentration250 IU/mL (12.5 IU/swab)
Precision/Reproducibility:≥ 90% agreementOverall:
Negative (NPA)99.2% (134/135)
Weak Positive (PPA)100.0% (135/135)
Positive (PPA)100.0% (135/135)
Inclusivity (for tested variants):Reliable detectionAll tested variants (Alpha, Beta, Delta, Gamma, Kappa, Omicron) showed positive results at various concentrations.
Cross-Reactivity/Interference:No cross-reactivity/interferenceNo cross-reactivity or microbial interference observed with tested organisms. No interference from tested endogenous/exogenous substances.
Hook Effect:No detectable hook effectNo detectable hook effect up to 5.75 × 10⁶ TCID₅₀/mL
External Controls Validation:100% agreement expected100% agreement for both positive and negative controls.
Specimen Stability:Stable for up to 4 hoursSamples stable for up to 4 hours at room temperature.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: 806 evaluable subjects.
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA, and the virus strain mentioned is "USA-WA1/2020," suggesting a U.S. context.
    • Retrospective/Prospective: Prospective. The clinical performance study collected samples "prospectively collected from symptomatic subjects between May 2022 and July 2022 at six clinical point of care sites."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The ground truth for the clinical test set was established using RT-PCR assays, not expert human readers. Therefore, the concept of "number of experts" for ground truth establishment isn't applicable in the context of this device. The RT-PCR assays are considered the gold standard for detecting SARS-CoV-2 viral nucleic acid.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The adjudication method used for the comparator (RT-PCR) was a "2 out of 3 rules." This means that if the first two FDA-cleared, highly sensitive RT-PCR assays had discordant results, a third FDA-cleared RT-PCR comparator assay was performed, and the final result was determined by the majority.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is a rapid antigen test read visually by users (healthcare professionals), not an AI-powered diagnostic imaging device that assists human readers. The study focuses on the diagnostic accuracy of the device itself compared to a molecular gold standard.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a standalone "algorithm" in the sense of a chemical diagnostic assay that produces a visual result. There is no separate "algorithm" being evaluated without human-in-the-loop, as the human (healthcare professional) is part of reading the visually interpreted result. The "performance characteristics" section details the standalone performance of the device (the antigen test) against the RT-PCR comparator, which represents the device's accuracy without a separate human interpretation study impacting the core result.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The type of ground truth used for the clinical performance test set was molecular SARS-CoV-2 assay results, specifically composite RT-PCR assays. This is a highly sensitive and specific laboratory-based "diagnostic truth" for the presence of the virus.

8. The sample size for the training set

The document does not mention a training set in the context of machine learning or AI models. This device is a lateral flow immunochromatographic assay, a chemical test, not a software or AI-based diagnostic tool that would typically involve a separate training phase with a distinct dataset. The "development" of such a test involves chemical formulation and design, not typically data training in the AI sense.

9. How the ground truth for the training set was established

As there is no mention of a training set for an AI/ML model, this question is not applicable to the Healgen Rapid COVID-19 Antigen Test.

{0}------------------------------------------------

April 19, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Healgen Scientific LLC % Jinjie Hu President and Principal Consultant Axteria Biomed Consulting 8040 Cobble Creek Circle Potomac, Maryland 20854

Re: K232377

Trade/Device Name: Healgen Rapid COVID-19 Antigen Test Regulation Number: 21 CFR 866.3982 Regulation Name: Simple Point-Of-Care Device To Directly Detect SARS-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings Regulatory Class: Class II Product Code: QVF Dated: March 19, 2024 Received: March 19, 2024

Dear Jinjie Hu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Silke Digitally signed by Silke Schlottmann -S Schlottmann -S Date: 2024.04.19 09:08:30 -04'00"

Silke Schlottmann, Ph.D. Deputy Assistant Director Bacteriology Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Healgen. The logo consists of a cluster of blue hexagons on the left, followed by the word "HEALGEN" in blue, bold letters. A blue arc extends from the top of the hexagons, curving over the word "HEALGEN".

HEALGEN RAPID COVID-19 ANTIGEN TEST INSTRUCTIONS FOR USE

REF GCCOV-502a-NA

Image /page/2/Picture/3 description: The image contains two boxes with text inside. The first box on the left contains the letters "IVD" in bold, indicating In Vitro Diagnostics. The second box on the right contains the symbol "Rx" followed by the word "Only", indicating that the product is available by prescription only.

INTENDED USE

The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs specimens from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test

The test does not differentiate between SARS-CoV and SARS-CoV-2.

A negative test result is presumptive and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.

Positive results do not rule out co-infection with other respiratory pathogens

Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

SUMMARY

Coronaviruses belong to the ß genus. Coronavirus disease 19 (COVID-19) is an acute respiratory infectious disease in patients infected by Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2). The virus incubation period typically ranges from 1 to 14 days, and symptoms may appear within 2 to 14 days following exposure. Possible symptoms include fever or chills, fatique, new loss of taste or smell shortness of breath or difficulty breathing cough nasal congestion or runny nose, sore throat, myalgia, nausea or vomiting, headache, and diarrhea.

This test is for detection of SARS-CoV-2 nucleocapsid protein antigen to aid in the diagnosis of SARS-CoV-2 infection. Antigen is generally detectable in upper respiratory specimens during the acute phase of infection.

PRINCIPLE

The Healgen Rapid COVID-19 Antigen Test uses antibodies to detect nucleocapsid protein from SARS-CoV-2 in direct nasal (anterior nares) swabs. If SARS-CoV-2 nucleocapsid antigen is present in the sample a line will form at the test line at "T", which means the test is positive. If SARS-CoV-2 nucleocapsid protein antigen is not in the sample a test line will not appear at "T" which means the test is negative. A control will always appear at "O" on a test that is functioning properly and when the test procedure has been followed.

MATERIALS SUPPLIED

  • 20 Test Cassettes
  • 20 Sterile Swabs 20 Pre-filled buffer tubes with flip-top
  • 2 Tube Holders
  • 1 Instructions for Use
  • 1 Quick Reference Guide

MATERIALS REQUIRED BUT NOT PROVIDED

Timer

External Controls: COVID-19 Antigen Control Kit (GCCOV(Ag)-PN10/GCCOV(Ag)-PN20)

WARNINGS AND PRECAUTIONS

    1. For prescription use only
    1. For in vitro diagnostic use only
    1. Read all instructions carefully before performing the test. Failure to follow the instructions may result in inaccurate results.
    1. Serial testing should be performed in symptomatic individuals with negative results at least twice over three days (with 48 hours between tests). You may need to purchase additional tests to perform this serial (repeat) testing or follow up testing with a molecular test.
    1. Do not use the kit past its expiration date.
  • Do not use if any of the test kit contents or packaging is damaged.
  • Swabs, tubes, and test devices are for single use only. Do not re-use.
    1. Do not interchange or mix components from different kit lots. Testing should only be performed using the swabs provided within the kit. Do not touch the swab tip.
    1. To obtain accurate results, do not use visually bloody or overly viscous samples.
    1. Use appropriate precautions in the collection, handling, storage, and disposal of patient samples and used kit contents.
    1. Wear appropriate personal protection equipment and gloves when running each test and handling patient specimens. Change gloves between handling of specimens.
    1. Dispose of used contents as biohazardous wastes in accordance with federal, state, and local requirements.
    1. Do not open the Test Cassette until you are ready to perform the test. Once opened, the test cassette should be used within one hour.
    1. Do not read test results before 15 minutes or after 20 minutes. Results read before 15 minutes or after 20 minutes may lead to a false positive, false neqative, or invalid results.
    1. Do not use the kit to evaluate patient specimens if either the positive control swab or negative control swab fails to give the expected results.
    1. Do not move test device during result development.

STORAGE

    1. The kit can be stored at room temperature or refrigerated (2-30°C/36-86°F) in the original sealed pouch.
  • Do not freeze any of the test kit components.
    1. Keep away from direct sunlight.
  • Do not use the test device and reagents after the expiration date. 4. Unused test devices that have been outside of the sealed pouch for more
  • than one hour should be discarded. 6. Close the kit box and secure contents when not in use.

SPECIMEN COLLECTION

Acceptable specimen type for testing is direct anterior nasal swab specimen. Inadequate specimen collection, improper specimen handling and/or transport may vield false results.

Samples should be tested as soon as possible after collection. Based on data generated with the Healgen Rapid COVID-19 Antigen Test, nasal swabs are stable for up to 4 hours at room temperature but should be stored in a sterile container if not tested immediately.

  • Open the swab pouch by peeling the cover back. Hold the plastic stick end 1. of the swab and remove from pouch. Do not touch the swab tip.
    Image /page/2/Figure/51 description: The image shows two steps of a process. In the first step, a hand is opening a package containing a swab. An arrow indicates the direction of the opening. In the second step, the hand is holding the swab, which has been removed from the package.

  • To collect the anterior nasal swab sample, tilt the patient's head back 70 degrees and insert the swab tip into nostril, at about 1/2 to 3/4 inches deep. For children, young children, swab should not be inserted more than 1/2 inch. Rotate the swab along the inside walls of nostril in a complete circle at least 5 times. Remove swab from nostril and use the same swab to repeat in the other nostril.
    Image /page/2/Figure/53 description: The image shows a diagram of a nasal swab being inserted into a nostril. The swab is inserted and rotated in a circular motion 5 times. The image shows the process being repeated on the other nostril.

TEST PROCEDURE

Check expiration on each individual test package or outer box before using.

The test should be performed at room temperature (15-30°C). Allow the test materials to reach room temperature prior to use. Do not open the Test Cassette until you are ready to use.

  • While maintaining tube upright, remove seal from tube. Avoid spilling liquid.
    Image /page/2/Figure/58 description: The image shows an illustration of a swab being inserted into a tube opening. The swab is being inserted into the tube at an angle, as indicated by the blue arrow. The text below the illustration reads "2. Insert swab tip in tube opening."

  • Grab swab shaft and mix well by rolling swab in a circular motion at least 6 times. Press the swab tip against the bottom and sides of tube while rolling swab.
    Image /page/2/Figure/60 description: The image shows a diagram of a hand holding a test tube with a swab inside. The swab is being swirled in the test tube, and there is a label that says "6 circles" with an arrow indicating the swirling motion. The test tube is held in a rack with a red container next to it.

  • While holding the tube in your hand place the swab's tip near the top of the tube. Gently bend the swab until it snaps at the breakpoint.
    Image /page/2/Figure/62 description: The image shows a diagram of a swab being inserted into a tube. The first image shows the swab being inserted into the tube, with a dotted circle around the top of the tube. The text "Breakpoint 1 inch from swab tip" is next to the tube. The second image shows the swab broken off inside the tube.

  • Hold tube upright. Close the flip cap of the tube tightly to ensure there is a tight fit so it does not leak. Do not remove red cap or swab tip from tube.
    Image /page/2/Figure/64 description: The image shows a diagram of how to fold a tube holder. The first image shows a tube with a red cap and a blue arrow pointing down. The second image shows the tube with the red cap closed. The text below the image describes how to fold the tube holder by folding the vertical flaps toward the center and holding the horizontal flap down and inward.

Image /page/2/Figure/65 description: The image shows a diagram of a box being assembled. The left side of the image shows the box in its unfolded state, with arrows indicating how to fold the flaps. The right side of the image shows the box in its assembled state, with holes on the top.

  • Allow the swab to sit in buffer tube for at least 1 minute, but no longer than 7 30 minutes immediately after specimen collection.
    Image /page/2/Figure/67 description: The image shows a diagram of a small bottle being inserted into a hole in a block. The block has six holes in total, each with a gray circle inside. A blue arrow points downward toward the bottle, indicating the direction of insertion. The bottle has a red cap and a clear body.

  • Open the test cassette pouch by tearing the area circled below. Place the test cassette on a flat surface
    Image /page/2/Figure/69 description: The image shows a diagram of a process. On the left side of the image, there is a container with a green section labeled "Test" and a blue section. An arrow points from the container to a card with a similar color scheme. The card has a blue section on the top left and a green section in the middle.

  • Gently remove red cap from tip. Ensure tip remains secure within tube opening.
    If the test is
    negative, do
    not collect.
    NOTE: Do not
    moisten the
    swab with any
    softener.

NOTE: Negative results are presumptive and may be confirmed with a molecular assay, if necessary, for patient management. Individuals with symptoms of COVID-19 and initial negative results should be tested again after 48 hours or followed up with a molecular test. INVALID

Image /page/2/Picture/74 description: The image shows a diagram of a medical device, possibly a vial or container with a connected cap and a red component. An arrow indicates a rotational movement of the cap, suggesting a process of opening or closing the device. The red component is attached to the side of the vial, and the cap is connected to the top. The diagram is simple and illustrative, likely part of instructions for using the medical device.

Image /page/2/Figure/75 description: The image shows two steps for removing a swab from its packaging. In the first step, a hand is shown opening the packaging by pulling down on the top. An arrow indicates the direction of the pulling motion. The second step shows a hand holding the swab after it has been removed from the packaging.

Image /page/2/Figure/79 description: The image shows a diagram of a test tube with a swab inside. In the first image, the swab is inside the test tube, and there is a blue arrow pointing upwards. In the second image, the swab has been removed from the test tube, and there is a blue arrow pointing to the right.

  1. Invert the tube and gently squeeze, from the middle of the tube, to add 4 drops of solution into the sample well. labeled as "S" on the test cassette. Start timer for 15 minutes.

Image /page/2/Figure/81 description: The image shows a COVID-19 antigen test being performed. A hand is holding a tube and dispensing liquid into the sample well of the test device. An arrow points to a clock with the time set to 15 minutes, indicating the waiting time for the test results.

  1. After 15 minutes read the test results visually in the result window, labeled as "C" and "T" on the test cassette. Do not read result before 15 minutes or after 20 minutes.

INTERPRETATION OF RESULTS

POSITIVE

Image /page/2/Figure/85 description: The image shows two COVID-19 antigen rapid tests. The test on the left shows two red lines, one at the control line (C) and one at the test line (T), indicating a positive result. The test on the right shows a red line at the control line (C) and a faint red line at the test line (T), also indicating a positive result, although the intensity of the test line is weaker.

If the Control (C) line and the Test (T) line are visible, the test is positive. Any visible faint red or pink test (T) line with a visible control (C) line should be read as positive. Repeat testing is not needed for individuals with a positive result.

NEGATIVE

Image /page/2/Figure/88 description: The image shows a COVID-19 rapid antigen test. The test has a control line (C) visible, but no test line (T). The presence of only the control line indicates a negative test result.

If the Control (C) line is visible, but the Test (T) line is not visible, the test is negative. A negative test result indicates that the virus that causes COVID-19 was not detected in the sample

Image /page/2/Figure/91 description: The image shows a COVID-19 rapid antigen test. The test has a white plastic casing with a small rectangular window in the center. Inside the window, a single red line is visible at the bottom, near the 'T' marking, indicating a positive test result.

Image /page/2/Picture/92 description: The image shows a close-up of a COVID-19 test strip. The strip has labels "COVID-19 Ag" at the top. Below this, there are two indicators labeled "C" and "T". The image appears to be a diagram or illustration of the test strip rather than a photograph of a physical test.

If a control (C) line is not visible, the test is not valid. Invalid tests should be repeated with a new test.

LIMITATIONS

    1. This test is only for the detection of proteins from SARS-CoV-2, not for any other viruses or pathogens.
    1. The test is intended to only be used with direct anterior nasal swabs and is not validated for use with swabs in viral transport media.
    1. This test is only for use with individuals who show symptoms of upper respiratory infection within 6 days of symptom onset.
    • Serial testing should be performed in individuals with negative results at least twice over three days (with 48 hours between tests) for symptomatic individuals.
    • This test is not for use in at-home test settings.
    1. This test is read visually and has not been validated for use by those with impaired vision or color-impaired vision.
    1. This test should not be used beyond the expiration date listed on the packaging. Use of expired tests can lead to incorrect results.
    1. Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe procedures in any one of these steps can lead to incorrect results.
    1. Test results should be interpreted in conjunction with other clinical and laboratory information available to the healthcare provider. Results from the Page 1 of 2

{3}------------------------------------------------

device should be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.

    1. A negative test result does not preclude the possibility of infection with othe bacteria or viruses
    1. A negative test result may occur if the level of antigen in the sample is below the detection limit of the test.
    1. All COVID-19 antigen test negative results are presumptive and confirmation with a molecular assav may be necessary.
    1. There is a higher chance of false negative results with antigen tests than with laboratory-based molecular tests due to the sensitivity of the test technology. This means that there is a higher chance this test will give a false negative result in an individual with COVID-19 as compared to a molecular test, especially in samples with low viral load.
    1. Positive test results do not rule out co-infections with other non-SARS viral or bacterial pathogens.
    1. Positive test results do not differentiate between SARS-CoV and SARS-CoV-
    1. False positive test results are more likely when prevalence of upper respiratory infection is low in the community.
    1. This test detects both viable (live) and non-viable SARS-CoV-2 virus. Test performance depends on the amount of virus (antigens) in the sample and may or may not correlate with viral culture results performed on the same sample. A false-negative test result may occur if the level of viral antigen in a sample is below the detection limit of the test or if the sample was collected or transported improperly.
    1. There is a risk of erroneous results (i.e., false negatives) due to the presence of novel, emerging respiratory viral variants (e.g., specific strains or isolates).
    1. The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between May 2022 and July 2022. The clinical performance has not been established in all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation. Performance at the time of testing may vary denending on the variants circulating including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time.
    1. Based on sequence analysis, a potential for cross-reactivity between the SARS-CoV-2 test and HKU1 exists. Wet testing for HKU1 coronavirus was not conducted and therefore, cross-reactivity between SARS-CoV-2 and HKU1 coronavirus cannot be ruled out. measures)

PERFORMANCE CHARACTERISTICS

Clinical Performance

Clinical performance of the Healgen Rapid COVID-19 Antigen Test was established with 806 anterior nasal swab (ANS) samples prospectively collected from symptomatic subjects between May 2022 and July 2022 at six clinical point of care sites. Samples (ANS for the investigational device and nasopharyngeal (NP) swab for the comparator) were collected from sequentially enrolled subjects presenting with symptoms of upper respiratory infection within 6 days of onset of symptoms. Results obtained with the Healgen Rapid COVID-19 Antigen Test were compared to a composite comparator method of two FDA cleared, highly sensitive RT-PCR assays. A third FDA cleared RT-PCR comparator assay was performed on comparator samples with discordant results between the first and second RT-PCR tests. The final result for the comparator was determined using a 2 out of 3 rules. Testing was performed by operators who had no prior experience in the laboratory and were representative of the intended users Operators used only the QRI to conduct testing without comprehensive training provided.

Of the 806 subjects, 164 were confirmed positive by the comparator and 642 were confirmed negative. The positive percent agreement (PPA) was 85.4% and the neqative percent agreement (NPA) was 99.7% (see tables below).

Table 1. Healgen Rapid COVID-19 Antigen Test Results versus RT-PCR Composite Comparator for Symptomatic Patients

Healgen Rapid COVID-19 Antigen TestComposite ComparatorTotal
PositiveNegative
Positive1402142
Negative24640664
Total164642806

Positive Percent Agreement (PPA) = (140/164) = 85.4% (95% Cl: 79.1% - 90%) Negative Percent Agreement (NPA) = (640/642) = 99.7% (95% Cl: 98.9% - 99.9%)

Table 2. Patient Demographics

Subject AgeGroupNumber ofsamplestestedAntigenPositivesCompositeComparatorPositives% Positive(byComparator)
2-13 years ofage75111317.3%
14-24 yearsof age118171916.1%
25-64 yearsof age5149111021.4%
≥ 65 years ofage99212222.2%
Total80614016420.3%

Table 3. Clinical Performance in Subjects on Different Symptomatic Days

Days PostCOVID-19SymptomsNumber ofsamplestestedAntigenPositivesCompositeComparatorPositivesPPA (%)
Day 02633100.0%
Day 192141687.5%
Day 2213364678.3%
Day 3219313979.5%
Day 4150272993.1%
Day 575171989.5%
Day 6311212100.0%
Total80614016485.4%

Analytical Sensitivity: Limit of Detection (LoD)

LoD studies determine the lowest detectable concentration of SARS-CoV-2 at which approximately 95% of all (true positive) replicates test positive. Heat inactivated SARS- CoV-2 virus, isolate 2019-nCoV/USA-WA1/2020, was spiked into negative clinical matrix composed of pooled nasopharyngeal swab specimens in PBS/saline and serially diluted. A preliminary LoD test was performed by spiking 50 µL of each diluted sample onto the sample collection swab head in triplicate. The confirmatory LoD test was performed at the selected preliminary I oD concentration and at concentrations above and below the preliminary LoD with an additional 20 replicates on the Healgen Rapid COVID-19 Antigen Test. The Limit of Detection is 5.75 x 103 TCID50/mL (2.875 x 102 TCID50/swab).

Furthermore, the LoD was established using the 1* WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in real clinical matrix of pooled swab specimens. Initially, a preliminary LoD was performed in range finding studies, and a confirmatory LoD test was then conducted to confirm the preliminary LoD concentration and additional dilutions bracketing the preliminary concentration. It was determined that the LoD of the 1st WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) for the Healgen Rapid COVID-19 Antigen Test was determined to be 250 IU/mL (12.5 IU/swab).

Inclusivity

Inclusivity testing was conducted with the currently available commercial stock strains of SARS-CoV-2 Alpha, Beta, Gamma, Delta, Kappa, and Omicron variants

Table 4. Results of Inclusivity Testing

SARS-CoV-2 VariantLowest Concentration with 9/9Positive Results (TCID50/mL)
B.1.1.7 (Alpha)$1.00×10^2$
B.1.351 (Beta)$3.83×10^2$
B.1.617.2 (Delta)$1.10×10^2$
P1 (Gamma)$6.30×10^2$
B.1.617.1 (Kappa)$1.90×10^2$
B.1.1.529 (Omicron)$2.51×10^2$

Analytical Specificity: Cross Reactivity (Exclusivity) and Microbial Interference

Various microorganisms were evaluated for cross-reactivity and microbial interference by wet testing with the Healgen Rapid COVID-19 Antigen Test. Human coronavirus HKU1 was not tested for cross-reactivity due to a lack of availability. The samples were tested in triplicate, and no cross-reactivity and no microbial interference were observed Microbial concentrations and results are outlined in Table 5.

Table 5. Results of Microorganisms Tested for Cross-Reactivity and Microbial Interference

MicroorganismConcentration TestedCross-ReactivityInterference
Human coronavirus 229E8.00 × 105 TCID50/mLNoNo
Human coronavirus OC437.00 x 106 TCID50/mLNoNo
Human coronavirus NL632.93 x 104 TCID50/mLNoNo
MERS-coronavirus7.0 × 105 TCID50/mLNoNo
Adenovirus 212.39 x 106 TCID50/mLNoNo
Adenovirus 101.14 x 106 TCID50/mLNoNo
Human Metapneumovirus3.95 x 105 TCID50/mLNoNo
Parainfluenza virus Type 12.23 x 106 TCID50/mLNoNo
Parainfluenza virus Type 22.23 x 105 TCID50/mLNoNo
Parainfluenza virus Type 34.00 x 106 TCID50/mLNoNo
Parainfluenza virus Type 4a4.00 x 103 TCID50/mLNoNo
Influenza virus, Type A (H1N1)4.00 x 108 CEID50/mLNoNo
Influenza virus, Type A (H3N2)7.00 x 105 TCID50/mLNoNo
Influenza virus, Type B7.00 x 105 TCID50/mLNoNo
Enterovirus 682.23 x 106 TCID50/mLNoNo
MicroorganismConcentration TestedCross-ReactivityInterference
Enterovirus 71$4.00 \times 10^7$ TCID50/mLNoNo
Respiratory syncytial virus$2.23 \times 10^6$ TCID50/mLNoNo
Rhinovirus 60$8.00 \times 10^5$ TCID50/mLNoNo
Haemophilus influenzae$1.74 \times 10^8$ CFU/mLNoNo
Streptococcus pneumoniae$3.35 \times 10^8$ CFU/mLNoNo
Streptococcus pyogenes$5.98 \times 10^8$ CFU/mLNoNo
Candida albicans$1.19 \times 10^8$ CFU/mLNoNo
Bordetella pertussis$4.90 \times 10^9$ CFU/mLNoNo
Mycoplasma pneumoniae$6.75 \times 10^7$ CCU/mLNoNo
Chlamydia pneumoniae$4.25 \times 10^7$ CFU/mLNoNo
Legionella pneumophila$9.20 \times 10^9$ CFU/mLNoNo
Staphylococcus aureus$5.0 \times 10^6$ CFU/mLNoNo
Staphylococcus epidermidis$1.75 \times 10^8$ CFU/mLNoNo
Pooled human nasal washN/ANoNo

In-silico analysis was conducted for SARS-coronavirus Human coronavirus HKU1, Mycobacterium tuberculosis, Pneumocystis jirovecii (PJP) and MERS coronavirus

No sequence was found with significant homology with N protein sequence in both Mycobacterium tuberculosis and Pneumocvstis iirovecii genomes. suggesting no cross-reactivity would not occur; however, cross-reactivity cannot be ruled out

Homologous N protein sequences were identified in SARS-coronavirus. Human coronavirus HKU1, and MERS coronavirus; therefore, cross-reactivity cannot be ruled out. The N protein sequence of SARS coronavirus shares 79.01%- 97.61% sequence identity indicating that cross-reactivity is likely.

Endogenous Interfering Substances

To assess endogenous interference, potentially interfering substances which may be present in respiratory samples were tested to determine if interference may occur on the Healgen Rapid COVID-19 Antigen Test. A summary of the results is shown in Table 6.

Table 6. Results of Potential Interfering Substances

SubstanceConcen-trationCross-ReactivityInterference
Whole Blood4%NoNo
Human Leukocytes$1 \times 10^7$cells/mLNoNo
Mucin0.5%NoNo
Chloraseptic (Menthol/Benzocaine)3 mg/mLNoNo
Naso GEL (NeilMed)5% v/vNoNo
CVS Nasal Drops (Phenylephrine)15% v/vNoNo
Afrin (Oxymetazoline)15% v/vNoNo
CVS Nasal Spray (Cromolyn)15% v/vNoNo
Zicam5% v/vNoNo
Homeopathic (Alkalol)15% v/vNoNo
Sore Throat Phenol Spray15% v/vNoNo
Tobramycin4 $\mu$ g/mLNoNo
Mupirocin10 mg/mLNoNo
Fluticasone Propionate15% v/vNoNo
Tamiflu (Oseltamivir Phosphate)5 mg/mLNoNo

Hook Effect

The Hook Effect study tested up to 5.75 × 10° TCIDso/mL of heat-inactivated SARS-CoV-2 (USA-WA1/2020) on the Healgen Rapid COVID-19 Antigen Test. The test results show that there was no detectable hook effect even for the highest concentrated virus sample.

Precision/Reproducibility

A reproducibility study was performed to evaluate reproducibility of the Healgen Rapid COVID-19 Antigen Test. The study was performed at three external. CLIAwaived testing sites consisting of three replicates each of positive (prepared at 3x LoD), low positive (prepared at 1x LoD), and negative samples tested by three (3) untrained operators over 5 days, i.e., 3 replicates × 3 operators × 3 sites × 5 days= 135 replicates per concentration and a total of 405 data points collected. Three (3) test lots were used in this study, so lot-to-lot variability was also assessed. Fifty (50) µL of the prepared sample were applied to kit swabs, shipped and stored frozen at -20°C until testing. The results were ≥ 90% agreement between expected and read result within run, by lot, by operator, by day, between sites and overall

Tot

Sit

Image /page/3/Figure/48 description: The image shows a stack of boxes with letters inside. The letters are L and U. The boxes are black and white.

Image /page/3/Figure/49 description: The image contains the letters 'TEC' in large, bold, black font. A blue line is underneath the letters. The bottom of the image is cut off.

NegativeWeak PositivePositive
SiteCorrectReads/TotalPPACorrectReads/TotalPPACorrectReads/TotalPPA
144/4597.8%45/45100.0%45/45100.0%
245/45100.0%45/45100.0%45/45100.0%
345/45100.0%45/45100.0%45/45100.0%
Total134/13599.2%135/135100.0%135/135100.0%

Table 7. Results of Multisite Precision Study (Reproducibility)

REFERENCES

  1. Julien Favresse, Constant Gillot, Maxime Oliveria, Julie Cadrobbi, Marc Elsen. Christine Eucher. Kim Laffineur. Catherine Rosseels. Sandrine Van Eeckhoudt, Jean-Baptiste Nicolas, Laure Moiremont, Jean-Michael Dogné and Jonathan Douxfils. Head-to-Head Comparison of Rapid and Automated Antigen Detection Tests for the Diagnosis of SARS- CoV-2 Infection J. Clin. Med. 2021. 10. 265.
    1. Ignacio Torres, Sandrine Poujois, Eliseo Albert, Gabriela Álverez, Javier Colomina and David Navarro. Point-of-care evaluation of a rapid antigen test for diagnosis of SARS- CoV-2 infection in symptomatic and asymptomatic individuals February 11, 2021
    1. Public Health England. SARS-CoV-2 lateral flow antigen tests: evaluation of VOC1 (Kent UK) and VOC2 (South Africa) Published12 February 2021

INDEX OF SYMBOLS

Image: Do not reuse symbolDo not reuseImage: See instruction for use symbolSee Instructionfor UseImage: Expiration date symbolExpiration Date
Image: Tests per kit symbolTests per KitImage: Temperature symbolStore Between2-30°C (36-86°F)Image: Keep dry symbolKeep Dry
LOTBatch NumberREFCatalog#Image: Keep away from sunlight symbolKeep Away fromSunlight
UDIUnique DeviceIdentifierIVDFor in vitrodiagnostic use onlyImage: Manufacturer symbolManufacturer
Image: Prescription only symbolPrescriptionOnly

TECHNICAL SUPPORT

For questions or technical support, please contact the Technical Support Number 866-982-3818 or send email to Customer.Support@healgen.com

Image /page/3/Picture/61 description: The image shows a black silhouette of a building or structure. The structure has a flat roof with a rectangular extension on the right side. The left side of the structure has a jagged, saw-tooth pattern, creating a series of triangular peaks. The silhouette is simple and lacks any internal details or textures.

Healgen Scientific, LLC Address: 3818 Fuqua Street, Houston, TX 77047, USA. Tel: +1 713-733-8088 Website: www.healgen.com

Revision Date: 2024-04-19 C-04112024-A1-IVD-D4-V3

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Healgen. The logo consists of a blue hexagon shape on the left, with the text "HEALGEN" in blue, block letters to the right of the hexagon. There is a blue arc above and below the hexagon and text. There is a registered trademark symbol to the right of the word "HEALGEN".

HEALGEN RAPID COVID-19 ANTIGEN TEST

HEALTHCARE PROVIDER QUICK REFERENCE GUIDE

For in vitro diagnostic use For prescription use

IMPORTANT

· Read all instructions carefully before performing the test. Failure to follow the instructions may result in inaccurate test results. Refer to "Instructions for Use" for additional information.

• The test is intended for use with nasal (anterior nares) swab specimens from individuals: suspected of COVID-19 by their healthcare provider within the first six (6) days of symptom onset.

· Negative results are presumptive and should be confirmed with a molecular test, if necessary.

• Serial testing should be performed in individuals with negative results at least twice over three days (with 48 hours between tests) for symptomatic individuals.

• All test materials must be at room temperature (15°C - 30°C) before use. Do not open the Test Cassette until you are ready to perform the test.

· Wear appropriate personal protective equipment when performing this test.

· Check expiration date of each test device before use. Do not use any test past the expiration date. Refer to test device pouch or outer box labeling for expiration date.

• External Controls may be used to demonstrate that the reagents and assay procedure perform properly. Refer to the COVID-19 Antigen Control Kit for additional information.

Image /page/4/Figure/13 description: The image shows the materials needed for a COVID-19 antigen test. The materials include a test cassette pouch, a swab, a tube holder, a tube with liquid, and a timer. The swab has a breakpoint 1 inch from the swab tip. The timer is not included.

SPECIMEN COLLECTION See reverse side of page for TEST PROCEDURE

OPEN SWAB

Open the swab pouch by peeling the cover back. Hold the plastic stick end of the swab and remove from pouch. Do not touch the swab tip.

Image /page/4/Picture/17 description: The image shows a two-step process. In the first step, a hand is shown opening a test kit with a swab inside, indicated by a blue arrow showing the direction of opening. The second step, separated by a blue arrow, shows a hand holding the swab that was removed from the test kit.

SWAB BOTH NOSTRILS

To collect the anterior nasal swab sample, tilt the patient's head back 70 degrees and insert the swab tip into one nostril of the patient, at about 1/2 to 3/4 inches deep. For children, young children, swab should not be inserted more than 1/2 inch. Rotate the swab along the inside walls of nostril in a complete circle at least 5 times. Remove swab from the nostril and use the same swab to repeat in another nostril.

Image /page/4/Figure/20 description: The image shows a diagram of a nasal swab being inserted into a nostril. The swab is inserted and rotated in a circular motion 5 times. An arrow points to a second diagram of the same process being repeated in the other nostril.

While maintaining tube upright, remove seal from tube. Avoid spilling liquid.

Image /page/4/Figure/22 description: The image shows two steps in a process. The first step, labeled "4 PLACE SWAB IN TUBE", shows a diagram of a tube with a swab being inserted into the opening. The second step, labeled "5 MIX WITH SWAB", shows a diagram of the swab fully inserted into the tube.

Grab swab shaft and mix well by rolling swab in a circular motion at least 6 times. Press the swab tip against the bottom and sides of tube while rolling swab.

Image /page/4/Figure/24 description: The image shows a diagram of a hand holding a test tube with a swab inside. The text "6 circles" is written near the top of the test tube, indicating that the swab should be rotated six times. The text "BREAK SWAB SHAFT" is written at the bottom of the image, indicating that the swab shaft should be broken.

6

While holding the tube in your hand, place the swab's tip near the top of the tube. Gently bend the swab until it snaps at the breakpoint.

Image /page/4/Figure/26 description: The image shows a diagram of a tube with a swab being inserted into it. The text "Breakpoint 1 inch from swab tip" is visible, indicating the location where the swab should be broken. The image also includes the text "CLOSE TUBE", suggesting that the tube should be closed after the swab is inserted.

Hold tube upright. Close the flip cap of the tube tightly to ensure there is a tight fit so it does not leak. Do not remove red cap or swab tip from tube.

Image /page/4/Figure/28 description: The image shows a diagram of a test tube with a red cap. On the left, the red cap is being placed on the test tube, and there is a blue arrow indicating the direction of movement. On the right, the red cap is securely placed on the test tube. There is a blue arrow in the middle of the image.

Revision Date: 2024-4-18 C-04122024-A1-IVD-D2-V1

IVD | Ronly

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Healgen. The logo consists of three blue hexagons arranged in a cluster on the left, followed by the word "HEALGEN" in blue, block letters. A blue arc extends from below the hexagons, curving upwards and over the word "HEALGEN". A small "®" symbol is located to the upper right of the word "HEALGEN".

TEST PROCEDURE See reverse side of page for SPECIMEN COLLECTION

FOLD TUBE HOLDER

Fold tube holder by folding vertical flaps towards center (A). Hold vertical flaps in place and fold horizontal flap down and inward (B).

Image /page/5/Figure/4 description: The image shows a diagram of a box being assembled. On the left, the box is partially assembled with flaps labeled A and B. An arrow points from the partially assembled box to a fully assembled box on the right.

PLACE TUBE IN TUBE HOLDER

Allow the swab to sit in buffer tube for at least 1 minute, but no longer than 30 minutes immediately after specimen collection.

Image /page/5/Picture/7 description: In the image, a white rectangular block with six circular holes is depicted. A clear plastic cup with a red lid and a straw is inserted into one of the holes on the top left of the block. A blue arrow points downwards towards the cup, suggesting a direction or action related to the cup and the block. The image is a simple illustration, possibly indicating a process or instruction involving the cup and the block.

ਵ OPEN TEST DEVICE

Open the test cassette pouch by tearing the area circled below. Place the test device on a flat surface.

Image /page/5/Picture/10 description: The image shows a diagram of a test being taken. On the left side of the image, there is a box with a green and blue test inside. An arrow points from the box to a test being taken. The test is being taken by a machine.

REMOVE CAP

Gently remove red cap from tip. Ensure tip remains secure within tube opening.

Image /page/5/Picture/13 description: The image shows a diagram of a test tube with a red cap and a swab inside. There is a blue arrow pointing upwards on the left side of the test tube. There is also a blue arrow pointing to the right, indicating a process or sequence.

ADD 4 DROPS OF SAMPLE

Invert the tube and gently squeeze, from the middle of the tube, to add 4 drops of solution into the sample well, labeled as "S" on the test device.

Start timer for 15 minutes.

5

Image /page/5/Picture/17 description: In the image, a hand is holding a dropper above a COVID-19 test. The dropper is positioned directly over the sample well of the test, and a drop of liquid is about to fall into the well. The test itself has markings labeled 'C', 'T', and 'S', indicating the control line, test line, and sample well, respectively. An arrow is pointing to the right of the image.

Image /page/5/Picture/18 description: The image shows an analog clock with the time set to 15 minutes. The clock face is white with black tick marks, and the area from the 12 to the 3 is shaded in blue. Below the clock is the text "15 minutes" in blue.

READ TEST RESULT

After 15 minutes find result window, labeled as "C" and "T" on the test device.

Do not read result after 20 minutes.

Below are results examples for positive, negative and invalid results.

External positive and negative control swabs are supplied separately as "COVID-19 Antigen Control Kit". These controls provide additional quality control material to assess that the test kit reagents perform as expected. Process the controls in the same manner as a clinical sample swab, and conduct the assay as described in "Test Procedure" section.

Used test materials should be discarded as biohazard waste according to Federal, State and local regulatory requirements.

RESULTS INTERPRETATION

POSITIVE

If the Control (C) line and the Test (T) line are visible, the test is positive. Any visible faint red or pink test (T) line with a visible control (C) line should be read as positive. Repeat testing is not needed for individuals with a positive result.

Image /page/5/Figure/28 description: The image shows a COVID-19 rapid antigen test cassette. The test has two red lines, one at the control line (C) and one at the test line (T), indicating a positive result. The test cassette is white with black borders and blue text.

Image /page/5/Picture/29 description: The image shows a COVID-19 rapid antigen test. The test has two lines, labeled 'C' and 'T'. There is a red line at the 'C' position and a gray line at the 'T' position, indicating a positive result.

NEGATIVE

If the Control (C) line is visible, but the Test (T) line is not visible, the test is negative. A negative test result indicates that the virus that causes COVID-19 was not detected in the sample.

NOTE: Negative results are presumptive and may be confirmed with a molecular assay, if necessary, for patient management. Individuals with symptoms of COVID-19 and initial negative results should be tested again after 48 hours or followed up with a molecular test.

Image /page/5/Figure/33 description: The image shows a COVID-19 antigen test with a single red line at the top of the test window. The test window is labeled with the letters 'C' and 'T' on the right side. The red line is located at the 'C' position, indicating a negative test result. The test is labeled 'COVID-19 Ag' on the left side of the test window.

INVALID

If a control (C) line is not visible, the test is not valid. Invalid tests should be repeated with a new test.

Image /page/5/Figure/36 description: The image shows a COVID-19 antigen test. The test has a window with a single red line at the bottom, indicating a positive result. The letters 'C' and 'T' are visible on the right side of the test, likely representing the control and test lines, respectively.

Image /page/5/Figure/37 description: The image shows a COVID-19 test strip. The test strip has a window in the middle. The letters C and T are on the right side of the test strip.

TECHNICAL SUPPORT

For questions or technical support, please contact the Technical Support Number 866-982-3818 or send email to Customer.Support@healgen.com.

Prescription

Use Only

R Only

DEX OF SYMBO

Image /page/5/Picture/41 description: The image shows a white box with the letters "IVD" in bold black font. To the right of the box, there is text that reads "For in vitro diagnostic use only". The text is in a smaller font size than the letters "IVD".

Image /page/5/Picture/42 description: The image shows a close-up of a COVID-19 test strip. The test strip has two labeled regions, 'C' and 'T', which likely stand for 'Control' and 'Test', respectively. The 'C' region shows a visible line, while the 'T' region does not, suggesting a negative test result. The test strip is part of a COVID-19 Ag test, as indicated by the text on the strip.

Healgen Scientific, LLC. 3818 Fuqua Street Houston, TX, 77047 Phone: 713-733-8088 www.healgen.com

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Healgen. The logo consists of three blue hexagons connected together on the left side of the word "HEALGEN" in blue font. There is a blue arc above and below the word "HEALGEN" and the hexagons. There is a registered trademark symbol to the right of the word "HEALGEN".

COVID-19 Antigen Control Kit

REF: GCCOV(Ag)-PN10 or GCCOV(Ag)-PN20

Configurations: (PN10: 5 positive swabs. 5 negative swabs) (PN20: 10 positive swabs. 10 negative swabs)

For in vitro diagnostic use only. For prescription use only. For professional use only.

Instructions for use must be carefully followed when performing the test. Failure to follow the instructions may result in inaccurate results.

INTENDED USE

The COVID-19 Antigen Control Kit is a ready-to-use external control kit for use with the Healgen® Rapid COVID-19 Antigen Test to ensure that the reagents and materials are working properly and that the test procedure is correctly performed.

CONTENTS

The COVID-19 Antigen negative control swab is composed of negative control buffer dried onto a swab, with a blue shaft, containing a preservative.

The COVID-19 Antigen positive control swab is composed of a SARS-CoV-2 recombinant antigen extract dried onto a swab, with a red shaft, containing a preservative.

The preservative is 0.1% sodium azide.

The COVID-19 Antigen Control swabs are designed to be used to verify proper test procedure and performance of the Healgen® Rapid COVID-19 Antigen Test.

PRECAUTIONS

  • · For in vitro diagnostic use only.
  • · For professional use only.
  • · Do not use after expiration date.
  • · The COVID-19 Antigen Control Kit is only for use with the Healgen® Rapid COVID-19 Antigen Test. They have not been validated for use with other tests.
  • · The swab should remain in the sealed pouch until use.
  • · Do not use the swab if pouch is damaged.
  • Do not re-use any contents in the product. .
  • · Follow Good Laboratory Practices, wear protective clothing and use disposal disposable gloves when working with these controls. Do not eat, drink or smoke in the area.
  • · All the specimens should be considered potentially hazardous and handled in the same manner as an infectious agent.
  • · The control swab and test device should be discarded in a proper biohazard container after testing.

STORAGE AND STABILITY

Store as packaged in the sealed pouch either at refrigerated or room temperatures (2-30°C/36-86°F). The swab is stable through the expiration date printed on the sealed pouch. The swab must remain in the sealed pouch until use. Do not freeze.

MATERIALS

MATERIALS PROVIDEDMATERIALS REQUIRED BUTNOT PROVIDED
- 5 or 10 Negative control swab(blue)- 5 or 10 Positive control swab(red)- Instructions for use- Disposable gloves- Timer- Healgen® Rapid COVID-19Antigen Test

CONTROL KIT TEST PROCEDURE

Allow the test materials to reach room temperature (15-30°C/59-86°F) prior to testing. Do not open pouches until ready to perform the test.

For full instructions for use on how to use these controls, please refer to Healgen® Rapid COVID-19 Antigen Test Healthcare Provider Instructions for Use (IFU). The swab samples should be tested in the same manner as patient swab specimens.

We recommend controls be run once for:

{7}------------------------------------------------

  • Each new kit lot,

  • Each new operator,

  • Each new shipment,

  • As required by site quality control procedures and in accordance with local, state and federal regulations or accreditation requirements.

If either or both external quality controls give invalid results, repeat the external controls with new swabs on the Healgen® Rapid COVID-19 Antigen Test.

Image /page/7/Figure/5 description: The image shows three different test results. The first test result is positive, with two dark lines appearing in the C and T regions. The second test result is negative, with only one dark line appearing in the C region. The third test result is invalid, with a dark line appearing in the T region but not in the C region.

POSITIVE:

If the Control (C) line and the Test (T) line are visible, the test is positive. Any visible faint red or pink test (T) line with a visible control (C) line should be read as positive. Repeat testing is not needed for individuals with a positive result.

NEGATIVE:

If the Control (C) line is visible, but the Test (T) line is not visible, the test is negative. A negative test result indicates that the virus that causes COVID-19 was not detected in the sample.

NOTE: Negative results are presumptive and may be confirmed with a molecular assay, if necessary, for patient management. Individuals with symptoms of COVID-19 and initial negative results should be tested again after 48 hours or followed up with a molecular test.

INVALID:

If the control line (C) is not visible within the result window after performing the test, the result is invalid. Some causes of invalid results include failure to correctly follow directions or the test was used beyond the expiration date. It is recommended that the control swab be re- tested using a new test device.

Header
Add

Healgen Scientific, LLC. Address: 3818 Fuqua Street, Houston, TX 77047, USA. Tel: +1 713-733-8088 Website: www.healgen.com

Revision Date: 2024-04-19

{8}------------------------------------------------

510(k) Summary:

Submitter:Healgen Scientific LLC3818 Fuqua StreetHouston, TX 77047Telephone: 713-733-8088FAX: 713-733-8848
Contact person:Jinjie HuPresident and Principal ConsultantAxteria BioMed Consulting Inc.Tel: 301-814-4985Email: jinjie.hu@axteriabiomed.com
Secondary contact:Bingliang Fang, PhDChief Scientific OfficerHealgen Scientific LLCTel: 713-733-8088

Date prepared: April 19, 2024

A. 510(k) Number:

K232377

B. Purpose for Submission:

New 510(k) device clearance for the Healgen Rapid COVID-19 Antigen Test

Email: bingliang.fang@healgen.com

C. Measurand:

SARS-Coronavirus 2 (SARS-CoV-2) Nucleocapsid Protein Antigen

D. Type of Test:

Qualitative lateral flow immunoassay

E. Device

Proprietary and Established Name: Common Name: Classification Name:

Regulation Number: Classification:

Healgen Rapid COVID-19 Antigen Test Healgen Rapid COVID-19 Antigen Test Simple Point-Of-Care Device To Directly Detect SARS-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings 21 CFR 866.3982 Class II

{9}------------------------------------------------

Product Code: Panel: Predicate Device: QVF

Microbiology Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set (DEN220039; granted March 8, 2023)

F. Intended Use:

    1. Intended use(s):
      The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test, and followed with a molecular test.

The test does not differentiate between SARS-CoV and SARS-CoV-2.

A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.

Positive results do not rule out co-infection with other respiratory pathogens.

Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.

2. Indication(s) for use:

Same as the intended use

3. Special condition for use statement(s):

For prescription use only

For in vitro diagnostic use only

    1. Special instrument requirements:
      None

G. Device Description:

The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay that uses highly sensitive monoclonal antibodies to detect nucleocapsid protein from SARS-CoV-2 virus in nasal swab collected specimens. The test strip is composed of the following components: sample

{10}------------------------------------------------

pad, reagent pad, reaction membrane, and absorbing pad housed within a test cassette. The reagent pad contains colloidal-gold-conjugated monoclonal antibody that recognizes and binds to the nucleocapsid protein of SARS-CoV-2; the reaction membrane in the test line (T) contains the second antibody that recognizes another epitope of the nucleocapsid protein of SARS-CoV-2.

External quality controls are required but not included with the test kit and are sold separately as the COVID-19 Antigen Control Kit. The control swabs should be processed according to the Instructions for Use (IFU) and are intended to be used as quality control samples to demonstrate that the test is performing and is being performed correctly.

H. Test Principle:

The Rapid COVID-19 Antigen Test uses immunochromatographic-based technology and highly sensitive antibodies in a lateral flow design to detect nucleocapsid protein from SARS-COV-2 in direct nasal (anterior nares) swabs.

To begin the test, nasal swab specimens are collected by a health care professional (HCP). Following sample collection, the nasal swab is transferred to the prefilled extraction tube to lyse the sample and solubilize the viral nucleoproteins. Four (4) drops of the lysed sample are loaded onto the test cassette sample well, and the test result is read by the HCP after 15 minutes. If SARS-CoV-2 nucleocapsid antigen is present in the sample, a complex would form between the colloidal-goldmonoclonal antibody and the viral antigen, which is then captured by the second anti-SARS-2 antibody coated on the test line region (T).

The sample migrates along the test strip across 2 distinct areas: the test line and the procedural control line. If SARS-CoV-2 viral antigens are present, they will be captured by antibodies and bound to the test line of the test strip, and a colored line will be visible in the test window of the cassette at the test line at the "T" site, showing a positive result. If SARS-CoV-2 nucleocapsid protein antigen is absent in the sample, a test line would not appear at the "T" site, indicating a negative result. A control line would always appear at the "C" site in a test that functions normally and when the test procedure is followed. If no line is seen at the "C" site after 20 minutes, then the result is invalid and should be repeated with new test materials.

Device & PredicateDevice(s):Healgen Rapid COVID-19 Antigen Test(K232377) (Device)Sofia 2 SARS Antigen+ FIA,Sofia 2 SARS Antigen+ FIA ControlSwab Set(DEN220039) (Predicate)
Intended Use/Indicationsfor UseThe Healgen Rapid COVID-19 AntigenTest is a lateral flowimmunochromatographic assay intendedfor the rapid, qualitative detection ofSARS-CoV-2 nucleocapsid proteinantigens directly in anterior nasal swabsfrom individuals with signs and symptomsof upper respiratory infection within theThe Sofia 2 SARS Antigen+ FIA is alateral flow immunofluorescent sandwichassay that is used with the Sofia 2instrument for the rapid, qualitativedetection of SARS-CoV-2 nucleocapsidprotein antigens directly in anterior nasalswab specimens from individuals withsigns and symptoms of upper respiratory
first six (6) days of symptom onset. Thetest is intended for use as an aid in thediagnosis of SARS-CoV-2 infections(COVID-19) in symptomatic individualswhen tested at least twice over three dayswith at least 48 hours between tests; orwhen tested once, and negative by theHealgen Rapid COVID-19 Antigen Test,and followed with a molecular test.The test does not differentiate betweenSARS-CoV and SARS-CoV-2.A negative test result is presumptive, andit is recommended these results beconfirmed by a molecular SARS-CoV-2assay. Negative results do not precludeSARS-CoV-2 infections and should notbe used as the sole basis for treatment orother patient management decisions.Positive results do not rule out co-infection with other respiratory pathogens.Performance characteristics for SARS-CoV-2 were established from May 2022to July 2022 when SARS-CoV-2 Omicronwas the predominant SARS-CoV-2variant in circulation. When other SARS-CoV-2 virus variants are emerging,performance characteristics may vary.infection (i.e., symptomatic) when testingis started within 6 days of symptom onset.The test is intended for use as an aid inthe diagnosis of SARS-CoV-2 infections(COVID-19) in symptomatic individualswhen tested at least twice over three dayswith at least 48 hours between tests.The test does not differentiate betweenSARS-CoV and SARS-CoV-2.A negative test result is presumptive, andit is recommended these results beconfirmed by a molecular SARS-CoV-2assay. Negative results do not precludeSARS-CoV-2 infections and should notbe used as the sole basis for treatment orother patient management decisions.Positive results do not rule out co-infection with other respiratorypathogens.Performance characteristics for SARS-CoV-2 were established during the 2021-2022 SARS-CoV-2 pandemic whenSARS-CoV-2 Omicron was thepredominant SARS-CoV-2 variant incirculation. When other SARS-CoV-2virus variant are emerging, performancecharacteristics may vary.This test is intended for prescription useonly and can be used in Point-of-Caresettings.
SIMILARITIES
Intended Use PopulationSymptomatic individuals within 6 daysDPSOSame
Target AnalyteNucleocapsid protein antigen from SARS-CoV-2Same
Sample TypeAnterior nasal swab specimenSame
Test Time15 minutes after the extracted sample isadded to the test cassette sample wellSame
Results ReportedQualitative (positive, negative, invalid)Same
Serial TestingYesSame
Quality ControlsBuilt-in procedural control (control lineregion (C)) in the test strip and externalquality controls (separately packaged kit).Built-in procedural control in the test stripand external quality controls (included inkit).
DIFFERENCES
Test Design/ PrincipleLateral flow immunochromatographicassayLateral flow immunofluorescencesandwich assay
InstrumentationNoneSofia 2 Instrument
Result InterpretationVisually read - visual interpretation of thepresence or absence of colored line(s) onthe control and test line(s) of the test stripis used to determine Positive, Negative, orInvalid results.Instrument read - the Sofia 2 instrumentscans the test strip and measures thefluorescent signal by processing theresults using method specific algorithms.Sofia 2 displays the test results (Positive,Negative, or Invalid) on the screen.
Reagent Storage
Store at room temperature or refrigerate,36 to 86°F (2 to 30°C).Store the kit at room temperature, 59°F to86°F (15°C to 30°C).

I. Substantial Equivalence Information

{11}------------------------------------------------

{12}------------------------------------------------

Healgen Rapid COVID-19 Antigen Test Healgen Scientific LLC

J. Performance Characteristics

1. Analytical Performance:

a. Precision/Reproducibility:

A reproducibility study was performed at three (3) external testing sites, each site with three (3) operators. Each of the three (3) operators per site tested a panel of samples in replicates of three (3) over the course of five (5) days. Panel members consisted of inactive SARS-CoV-2 virus added into negative matrix at each of three different levels representing negative, low positive (1x LoD) and moderate positive (3x LoD) samples. Three (3) test lots were used in this study, so lot-to-lot variability was also assessed. Results are shown in the tables below.

NegativeWeak PositivePositive
SiteCorrectReads/TotalNPACorrectRead /TotalPPACorrectReads/TotalPPA
144/4597.8%45/45100.0%45/45100.0%
245/45100.0%45/45100.0%45/45100.0%
345/45100.0%45/45100.0%45/45100.0%
Total134/13599.2%135/135100.0%135/135100.0%

Table 1: Summary Results of Multisite Precision Study

NegativeWeak PositivePositive
LotCorrectReads/TotalNPACorrectRead/TotalPPACorrectReads/TotalPPA
145/45100.0%45/45100.0%45/45100.0%
244/4597.8%45/45100.0%45/45100.0%
345/45100.0%45/45100.0%45/45100.0%
Total134/13599.2%135/135100.0%135/135100.0%

Table 2: Summary Results of Lot-to-Lot Precision

b. Limit of Detection

For the limit of detection study, serially diluted samples were tested using three lots of test devices to determine the limit of detection (LoD), which is determined to be the lowest concentration of SARS-CoV-2 that the Healgen Rapid COVID-19 Antigen Test can reliably detect 95% of the time. A preliminary LoD was first established using 3 replicates per lot of 1:500, 1:1,000, and 1:4,000 serial dilution preparations of SARS-CoV-2 (USA-WA1/2020) on the Healgen Rapid

{13}------------------------------------------------

COVID-19 Antigen Test. Thereafter, the LoD was confirmed using 20 replicates per lot of 1:1,000. 1:2,000, and 1:4,000 serial dilution preparations. The LoD of the Healgen Rapid COVID-19 Antigen Test was determined as 5.75 x 102 TCIDso/mL at the 1:2000 dilution. The results are summarized in the table below.

DilutionAnalyteConcentration(TCID50/mL)TCID50/swab# Positive% Positive
Range Finding
1:500$2.3\times10^4$11.5 x 1029/9100.0%
1:1000$1.15\times10^4$5.75 x 1029/9100.0%
1:2000$5.75\times10^3$2.875 x 1029/9100.0%
1:4000$2.875\times10^3$1.438 x 1024/944.4%
Confirmatory
1:1000$1.15\times10^4$5.75 x 10260/60100.0%
1:2000$5.75\times10^3$2.875 x 10258/6096.7%
1:4000$2.875\times10^3$1.438 x 10251/6085.0%

Table 3: Summary Results for the Limit of Detection Study

A study was also performed to determine the LoD of the candidate device with the WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in real clinical matrix. The results are shown below, and the LoD for the WHO International Standard antigen was determined to be 250 IU.

DilutionAnalyteConcentration(IU/mL)IU/swab# Positive% Positive
1:405002520/20100.0%
1:8025012.520/20100.0%
1:1601256.2513/2065.0%

Table 4. Results of LoD Studv with WHO SARS-CoV-2 Standard

c. Linearity/assay reportable range:

Not applicable. This is a qualitative assay.

d. High dose hook effect study:

The hook effect study was conducted to determine whether the presence of a high concentration of the SARS-CoV-2 virus (USA-WA1/2020) in a sample would result in a negative test result, a phenomenon known as the hook effect. A heat-inactivated SARS-CoV-2 virus sample was used in

{14}------------------------------------------------

this study. The highest concentration tested was 5.75 × 106 TCID50/mL, and no high-dose hook effect was observed.

e. Inclusivity:

A total of six COVID-19 variants (Delta, Beta, Alpha, Omicron, P1 and Kappa) were tested to determine if the test can reliably detect these variants. A 3-dilution series for each lineage was prepared, and each dilution was tested in replicates of three (3) operators using three (3) kit lots. All variants gave positive results.

SARS-CoV-2VariantSublineageLowest Concentration Tested Positivefor All 3 Replicates in 3 Lots
DeltaB.1.617.2$1.10\times10^2$ TCID50/mL
BetaB.1.351$3.83\times10^2$ TCID50/mL
AlphaB.1.1.71.0 x 10² TCID50/mL
OmicronB.1.1.5292.51 x 10² TCID50/mL
GammaP16.30 x 10² TCID50/mL
KappaB.1.617.11.90 x 10² TCID50/mL

Table 5: Summary of Inclusivity Results

Analytical specificity: f.

Cross Reactivity and Microbial Interference:

The microbial cross reactivity and interference studies were performed to determine whether microorganisms that may be present in respiratory samples will cross-react or interfere with device performance.

Three replicates each of 1:1 dilutions of microorganism stock prepared in negative matrix were tested in the absence and presence of 2x LoD inactivated SARS-CoV-2 virus (USA-WA1/2020) on 3 test lots. Fifty (50) µL of the prepared sample was applied to kit swab and eluted into the extraction buffer tube. No cross-reactivity or microbial interference was observed with the organisms tested.

Table 6: Cross-Reactivity and Microbial Interference Results

MicroorganismConcentration TestedCross-ReactivityInterference
Human coronavirus 229E$8.00 \times 10^5$ TCID50/mLNoNo
Human coronavirus OC43$7.00 \times 10^6$ TCID50/mLNoNo
Human coronavirus NL63$2.93 \times 10^4$ TCID50/mLNoNo
MERS-coronavirus$7.0 \times 10^5$ TCID50/mLNoNo
Adenovirus 21$2.39 \times 10^6$ TCID50/mLNoNo

{15}------------------------------------------------

MicroorganismConcentration TestedCross-ReactivityInterference
Adenovirus 101.14 × 106 TCID50/mLNoNo
Human Metapneumovirus3.95 × 105 TCID50/mLNoNo
Parainfluenza virus Type 12.23 × 106 TCID50/mLNoNo
Parainfluenza virus Type 22.23 × 105 TCID50/mLNoNo
Parainfluenza virus Type 34.00 × 106 TCID50/mLNoNo
Parainfluenza virus Type 4a7.0 × 104 TCID50/mLNoNo
Influenza virus, Type A(H1N1)4.0 × 108 CEID50/mLNoNo
Influenza virus, Type A(H3N2)7.00 × 105 TCID50/mLNoNo
Influenza virus, Type B7.00 × 105 TCID50/mLNoNo
Enterovirus 682.23 × 106 TCID50/mLNoNo
Enterovirus 714.00 × 107 TCID50/mLNoNo
Respiratory syncytial virus2.23 × 106 TCID50/mLNoNo
Rhinovirus 608.00 × 105 TCID50/mLNoNo
Haemophilus influenzae1.74 × 108 CFU/mLNoNo
Streptococcus pneumoniae3.35 × 108 CFU/mLNoNo
Streptococcus pyogenes5.98 × 108 CFU/mLNoNo
Candida albicans1.19 × 108 CFU/mLNoNo
Bordetella pertussis4.90 × 109 CFU/mLNoNo
Mycoplasma pneumoniae6.75 × 107 CCU/mLNoNo
Chlamydia pneumoniae4.25 × 107 CFU/mLNoNo
Legionella pneumophila9.20 × 109 CFU/mLNoNo
Staphylococcus aureus5.0 × 106 CFU/mLNoNo
Staphylococcus epidermidis1.75 × 108 CFU/mLNoNo
Pooled human nasal washN/ANoNo

Healgen Rapid COVID-19 Antigen Test Healgen Scientific LLC

In-silico analysis was conducted for SARS-coronavirus, Human coronavirus HKU1, Mycobacterium tuberculosis, Pneumocystis jirovecii (PJP) and, for comparison, MERS coronavirus. The N protein sequence (GenBank ID: UUL70282.1) derived from an Omicron variant of SARS-CoV-2, belonging to the sublineages BA.1, BA.1.1, BA.2 (Accession Number OP160218) was used for the in-silico analysis.

There was no sequence with significant homology with this N protein sequence in both Mycobacterium tuberculosis and Pneumocystis jirovecii genomes, suggesting that cross-reactivity would not occur. Homologous sequences (N proteins) were identified in SARS-coronavirus, Human coronavirus HKU1, and MERS coronavirus. The N protein of Human coronavirus HKU1 shares 41.18% to 49.00% homology or identity with that of the SARS-CoV-2 Omicron N protein sequence whereas that of MERS coronavirus shares 52.87% to 54.04% identity. Homology is relatively low: however, cross-reactivity cannot be ruled out. The N protein sequence of SARS coronavirus shares 79.01% to 97.61% sequence identity with that of SARS CoV-2 Omicron variant indicating that cross-reactivity is likely. Since cross reactivity with SARS-CoV. MERS-CoV and HKU1 cannot be ruled out for SARS-CoV-2, the intended use states that it cannot distinguish SARS-CoV-2 and SARS-CoV.

{16}------------------------------------------------

Endogenous/Exogenous Interference:

Substances which may be present in respiratory samples were tested to determine if interference may occur on the Healgen Rapid COVID-19 Antigen Test in an endogenous/exogenous interference study. The potential interfering substances were tested with negative samples and low positive samples (2x LoD) SARS-CoV-2 (USA-WA1/2020) in triplicate, with three lots of tests at the listed concentrations. The results showed that none of the tested endogenous/exogenous substances interfered with the test at the tested concentrations.

SubstanceConcentrationCross-ReactivityInterference
Whole Blood4%NoNo
Human Leukocytes$1 \times 10^7$ cells/mLNoNo
Mucin0.5%NoNo
Chloraseptic (Menthol/Benzocaine)3 mg/mLNoNo
Naso GEL (NeilMed)5% v/vNoNo
CVS Nasal Drops (Phenylephrine)15% v/vNoNo
Afrin (Oxymetazoline)15% v/vNoNo
CVS Nasal Spray (Cromolyn)15% v/vNoNo
Zicam5% v/vNoNo
Homeopathic (Alkalol)15%NoNo
Sore Throat Phenol Spray15% v/vNoNo
Hand soap1%NoNo
Hand sanitizer1%NoNo
Tobramycin4 µg/mLNoNo
Mupirocin10 mg/mLNoNo
Fluticasone Propionate15% v/vNoNo
Tamiflu (Oseltamivir Phosphate)5 mg/mLNoNo

Table 7: Summary of Results for Potential Interfering Substances

g. Traceability, Stability and Expected Values (Controls, Calibrators, or Methods):

External Controls:

The Healgen Rapid COVID-19 Antigen Test is required to be run with external quality controls that are sold separately as the COVID-19 Antigen Control Kit. Three (3) external control lots were evaluated as follows: 4 replicates per lot of positive or negative external control swabs were tested on 3 different test kit lots for a total of 36 positive or negative external control results. All results were as expected.

ControlControlLot #TestLot #n# ofNeg# ofPos%Agreement
NegativeControl Swab11440100
2440100
3440100

Table 3: Summary Results of External Controls Validation

{17}------------------------------------------------

ControlControlLot #TestLot #n# ofNeg# ofPos%Agreement
21440100
2440100
3440100
31440100
2440100
3440100
Positive ControlSwab11404100
2404100
3404100
21404
2404100
3404100
31404
2404100
3404100

Healgen Rapid COVID-19 Antigen Test Healgen Scientific LLC

Specimen Stability:

The objective of the specimen stability study is to demonstrate that test performance is not impacted by using specimens that are not immediately tested after collection. Three (3) replicates each of low positive (prepared at 2x LoD) and negative samples were prepared and incubated at room temperature for 0, 2, 4 and 6 hours and then tested on three (3) test lots. The study results showed that the samples were stable for up to 4 hours when stored at room temperature.

2. Clinical Performance:

Clinical performance of the Healgen Rapid COVID-19 Antigen Test was evaluated in a multi-center, prospective study conducted from May 2022 to July 2022 using 32 untrained operators at six different CLIA-waived sites. Enrolled subjects were symptomatic subjects within six (6) days post symptom onset (DPSO) and exhibiting symptoms at the time of collection. A total of 806 evaluable subjects were collected and tested.

Two swabs were collected, the 1st was the anterior nares (AN) swab, it was tested on the Healgen Rapid COVID-19 Antigen device, and the 2nd was nasopharyngeal (NP) swab, it was collected into 3 mL of viral transport media (VTM) and tested on the composite reference method. For comparator testing, samples were tested on two highly sensitive RT-PCR comparator assays; testing on a third comparator would be performed only if discordant results were observed between the first two RT-PCR tests. The comparator result for a sample was determined based on a 2 out of 3 rule. Results obtained on the candidate device were compared to the comparator method to determine a positive percent agreement (PPA) of 85.4% and a negative percent agreement (NPA) of 99.7% for the symptomatic cohort.

{18}------------------------------------------------

Table 9: Summary of Clinical Performance

Healgen Rapid COVID-19Composite RT-PCR Comparator
Antigen TestPositiveNegativeTotal
Positive1402142
Negative24640664
Total164642806

PPA: 85.4% (140/164, 95% CI: 79.1% - 90.0%)

NPA: 99.7% (640/642, 95% CI: 98.9% - 99.9%)

Table 4: Clinical Performance Based on DPSO

Days of COVID-19SymptomsNumber ofSubject SamplesTestedAntigen PositivesRT-PCRPositivesPPA (%)
Day 02633100.0%
Day 192141687.5%
Day 2213364678.3%
Day 3219313979.5%
Day 4150272993.1%
Day 575171989.5%
Day 6311212100.0%

3. Clinical Cut-off:

Not applicable

4. Expected Value/Reference Range:

Not applicable

K. Proposed Labeling

The labeling supports a finding of substantial equivalence for the candidate device.

L. Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

N/A