Search Results
Found 62 results
510(k) Data Aggregation
(264 days)
San Diego, CA 92121
Re: K251091
Trade/Device Name: Lipids
Regulation Number: 21 CFR 862.1175
| Name | Regulation Number | Product Code / Class |
|---|---|---|
| Cholesterol (total) test system | 862.1175 |
Lipids is part of the TruWellness Panel™ and is intended for use on the TruVerus™. Lipids (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Total Cholesterol (TChol) and Triglycerides (TRIG) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the TRIG determination, Very Low-Density Lipoprotein Cholesterol (VLDL) is calculated by the analyzer.
Lipids (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in the diagnosis and treatment of the following disorders in adults 18 years of age or older:
Total Cholesterol (TChol): Excess cholesterol in the blood and lipid and lipoprotein disorders.
Triglyceride (TRIG): Diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism; various endocrine disorders.
The TruSystem is an automated, integrated in vitro diagnostic platform consisting of the TruVerus™ and the TruWellness Panel™, a Single-Use Consumable Kit that includes a Disc and a Support Pack. Designed for point-of-care and clinical laboratory use, the system enables the simultaneous measurement of clinical chemistry, immunoassay, and hematology parameters from a lithium-heparinized venous whole blood sample in a single run. The TruSystem delivers quantitative results for routine clinical chemistry and immunoassay analytes as well as a complete blood count (CBC) with a 3-part differential, all without the need for specialized operating skills, external calibration, or complex infrastructure.
The TruVerus™ is a benchtop instrument that fully automates sample processing, assay execution, and result reporting. Its touchscreen interface allows operators to control workflows, initiate tests, and review results with minimal training. The Single-Use Consumable Kit and blood sample are loaded into the analyzer drawer, where all necessary processing takes place. Internally, the TruVerus™ integrates:
- A high-precision pipettor for automated sample and reagent handling.
- An onboard centrifuge to separate whole blood into plasma.
- A closed-loop thermal control system to maintain precise assay temperatures.
- A motion control system to fully automate sample processing.
- Dedicated detection modules for clinical chemistry, immunoassay, and hematology testing.
- A high-definition camera used to collect assay readings, image capture, and instrument quality control checks.
- An electronics board to manage individual module boards and associated firmware.
- An integrated computer running the instrument software, accessible via the touchscreen user interface.
The TruVerus™ is factory-calibrated and continuously monitors its performance using optical sensing and electronic feedback mechanisms. Every time the analyzer is powered on or a Single-Use Consumable Kit (TruWellness Panel™) is loaded, the system runs an automated self-test to verify that it remains within calibration and is functioning properly. Internal quality control (QC) checks occur throughout the testing process to ensure the integrity of the analyzer, Disc, and Support Pack. If any self-test or QC check fails to meet system specifications, the TruVerus™ will display an error code and next steps on the touchscreen interface.
The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process as well as analyze samples on the TruVerus™, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards of at the end of the run.
Lipids is part of the TruWellness Panel™ and is intended for use on the TruVerus™. Lipids (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Total Cholesterol (TChol) and Triglycerides (TRIG) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the TRIG determination, Very Low-Density Lipoprotein Cholesterol (VLDL) is calculated by the analyzer.
N/A
Ask a specific question about this device
(365 days)
Device Name: CHOLESTEROL: HDL-Cholesterol: LDL-Cholesterol: TRIGLYCERIDES Regulation Number: 21 CFR 862.1175
{5}------------------------------------------------
Classification Name(s):
| 21 CFR § 862.1175 |
|---|
| Classification |
CHOLESTEROL: Reagent kit intended for the quantitative determination of Cholesterol in human serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood, of lipid and lipoprotein metabolism disorders.
HDL-Cholesterol: Reagent kit intended for the quantitative determination of high-density lipoprotein in human serum. Measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.
LDL-Cholesterol: Reagent kit intended for the quantitative determination of low-density lipoprotein in human serum. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
TRIGLYCERIDES: Reagent kit intended for the quantitative determination of triglycerides (neutral fat) in human serum. Measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
CHOLESTEROL: The Cholesterol Oxidase peroxidase (CHOD-PAP) enzymatic method is used. The cholesterol esterase enzyme catalyzes the hydrolysis of cholesterol and free fatty and free fatty acids. Free cholesterol, including that originally present in the sample, is then oxidized by the enzyme cholesterol oxidase (CHOD) to cholest-4-en-3-one, by using molecular oxygen as the electron acceptor and concurrently producing hydrogen peroxide (H2O2). The H2O2 produced is then used in a subsequent chromogenic oxidative coupling reaction, catalyzed by the enzyme peroxidase, in the presence of a redox indicator system, which leads to the formation of a colored compound, absorbing light at 550 nm. The increase in absorbance is directly proportional to the cholesterol concentration in the sample.
HDL-Cholesterol: The Accelerator Selective Detergent method is applied. The determination of HDL-Cholesterol is based on the following reactions: LDL, VLDL, and chylomicrons are neutralized by the combined action of the enzymes Cholesterol Oxidase, Peroxidase, accelerators and N,N-bis-(4-sulfobutyl)-m-toluidine-disodium (DSBmT). HDL remaining in the sample is disrupted by the action of a selective detergent and cholesterol is converted to △4 Cholestenone by the enzymatic action of Cholesterol Esterase and Cholesterol Oxidase, with the subsequent production of H2O2, which reacts with DSBmT and 4-aminoantipyrine in the presence of Peroxidase to a colored complex that absorbs light at 590 nm. The absorbance measured is proportional to the concentration of HDL-Cholesterol in the sample.
LDL-Cholesterol: The Selective Detergent method is applied. The method is in a two-reagent format and depends on the properties of a unique detergent. The first detergent solubilizes only the non-LDL lipoprotein particles. The cholesterol released is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. The second detergent solubilizes the remaining LDL particles, and a chromogenic coupler allows for color formation. The enzyme reaction with LDL-Cholesterol in the presence of the coupler at 590 nm produces color that is proportional to the amount of LDL cholesterol present in the sample.
TRIGLYCERIDES: The enzymatic glycerol-3-phosphate-peroxidase (GPO-POD) method is used. The method enzymatically hydrolyzes by lipase to free fatty acids and glycerol is phosphorylated by adenosine triphosphate (ATP) with glycerokinase (GK) to produce glycerol-3-phosphate and adenosine diphosphate (ADP). Glycerol-3-phosphate-oxidase oxidizes glycerol-3-phosphate to dihydroxyacetone phosphate and H2O2. The catalytic action of peroxidase (POD) forms quinoneimine from H202, aminoantipyrine, and Dihydrate (N-Ethyl-N-(2hydroxy-3-sulfopropyl)-m-toluidine (TOOS). The absorption change at 550 nm is proportional to the triglycerides concentration in the sample.
Here's a breakdown of the acceptance criteria and the study information for the Medicon Hellas Cholesterol, HDL-Cholesterol, LDL-Cholesterol, and Triglycerides test systems, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by comparison to legally marketed predicate devices and alignment with clinical laboratory guidelines (CLSI). The document presents a clear comparison in the "Device Comparison Table" sections. For this summary, I'll focus on the key performance indicators for each analyte.
CHOLESTEROL
| Acceptance Criteria (Predicate: OLYMPUS CHOLESTEROL REAGENT (K925603)) | Reported Device Performance (Medicon Hellas CHOLESTEROL) |
|---|---|
| Method comparison (correlation to comparator): 1.000 | Method comparison (correlation to comparator): 0.9980 |
| Reportable range: 20 to 700 mg/dL | Reportable range: 20 to 700 mg/dL |
| Sensitivity LoD: 1 mg/dL (Predicate LoQ not defined) | Sensitivity LoD / LoQ: LoD 4.4 / LoQ 4.6 (mg/dL) |
| Precision (within run & total for all LVs): <= 3% | Precision (within run & total for all LVs): <= 4% |
| Endogenous Interferences: Hemoglobin: up to 500 mg/dL | Endogenous Interferences: Hemoglobin: up to 500 mg/dL |
| Endogenous Interferences: Triglycerides: up to 1,000 mg/dL | Endogenous Interferences: Triglycerides: up to 1,500 mg/dL |
| Calibration frequency: 30 days | Calibration frequency: 14 days |
| On-board stability: Not defined | On-board stability: 28 days |
| Specimen Type: Human serum, plasma and urine | Specimen Type: Human serum |
HDL-Cholesterol
| Acceptance Criteria (Predicate: DIRECT HDL (K981224)) | Reported Device Performance (Medicon Hellas HDL-Cholesterol) |
|---|---|
| Method comparison (correlation to comparator): 1.999 (Typo in document, likely 0.999) | Method comparison (correlation to comparator): 0.997 |
| Reportable range: 5.0 to 221 mg/dL | Reportable range: 6.0 to 200 mg/dL |
| Sensitivity LoD / LoQ: LoD 2.5 / LoQ 5.0 (mg/dL) | Sensitivity LoD / LoQ: LoD 3.0 / LoQ 5.8 (mg/dL) |
| Precision (within run & total for all LVs): <= 4.0% | Precision (within run & total for all LVs): <=4.0% |
| Endogenous Interferences: Hemoglobin: up to 2,000 mg/dL | Endogenous Interferences: Hemoglobin: up to 1,000 mg/dL |
| Endogenous Interferences: Triglycerides: MDL1 1,000mg/dL & MDL2 2,000mg/dL | Endogenous Interferences: Triglycerides: up to 1,500 mg/dL |
| Calibration frequency: 28 days | Calibration frequency: 28 days |
| On-board stability: Not defined | On-board stability: 28 days |
| Specimen Type: Human serum & plasma | Specimen Type: Human serum |
LDL-Cholesterol
| Acceptance Criteria (Predicate: DIRECT LDL (K981303)) | Reported Device Performance (Medicon Hellas LDL-Cholesterol) |
|---|---|
| Method comparison (correlation to comparator): 0.960 | Method comparison (correlation to comparator): 0.999 |
| Reportable range: 1 to 800 mg/dL | Reportable range: 3 to 800mg/dL |
| Sensitivity LoD / LoQ: < 10mg/dL | Sensitivity LoD / LoQ: LoD 2 / LoQ 3 mg/dL |
| Precision (within run & total for all LVs): < 4.0% | Precision (within run & total for all LVs): < 4.0% |
| Endogenous Interferences: Hemoglobin: up to 500mg/dL | Endogenous Interferences: Hemoglobin: up to 1,000mg/dL |
| Endogenous Interferences: Triglycerides: up to 1,293 mg/dL | Endogenous Interferences: Triglycerides: up to 1,500 mg/dL |
| Calibration frequency: 28 days | Calibration frequency: At new lot |
| On-board stability: Not defined | On-board stability: 28 days |
| Specimen Type: Human serum & plasma | Specimen Type: Human serum |
TRIGLYCERIDES
| Acceptance Criteria (Predicate: OLYMPUS TRIGLYCERIDE TEST SYSTEM (K063804)) | Reported Device Performance (Medicon Hellas TRIGLYCERIDES) |
|---|---|
| Method comparison (correlation to comparator): 0.999 | Method comparison (correlation to comparator): 0.999 |
| Reportable range: 10 to 1,000mg/dL | Reportable range: 10 to 1,000mg/dL |
| Sensitivity LoD / LoQ: < 0.31 / 5.0 mg/dL | Sensitivity LoD / LoQ: LoD 5.5 / LoQ 9.7 mg/dL |
| Precision (within run & total for all LVs): < 5.0% | Precision (within run & total for all LVs): < 4.0% |
| Endogenous Interferences: Hemoglobin: up to 500mg/dL | Endogenous Interferences: Hemoglobin: up to 400mg/dL |
| Calibration frequency: 30 days | Calibration frequency: 28 days |
| On-board stability: 30 days | On-board stability: 28 days |
| Specimen Type: Human serum, plasma & urine | Specimen Type: Human serum |
2. Sample size used for the test set and the data provenance
-
Accuracy (Method Comparisons):
- A minimum of 75 leftover specimens.
- For the specific analytes:
- CHOLESTEROL: 93 human serum samples
- HDL-Cholesterol: 141 human serum samples
- LDL-Cholesterol: 107 human serum samples
- TRIGLYCERIDES: 163 human serum samples
- Data Provenance: The document states "left-over specimens," implying retrospective use of clinical samples. The country of origin is not explicitly stated, but the company is Medicon Hellas, S.A. based in Greece, and testing was performed at the company premises.
-
Precision/Reproducibility:
- Three human serum pools for Cholesterol and Triglycerides.
- Two pools for HDL-Cholesterol.
- Four pools for LDL-Cholesterol.
- Each sample was tested for 20 testing days, two different runs, and two replicate measurements per run (morning and afternoon run), for a total of 80 results per level concentration (e.g., for Cholesterol, 3 pools * 80 results/pool = 240 results).
- Data Provenance: Human serum pools, likely prepared in-house or acquired for the study.
-
Linearity:
- 11 to 13 levels per analyte, prepared by dilution of a human serum pool.
- Each level was tested in 4 replicates.
- Data Provenance: Human serum pool.
-
Analytical Specificity / Interference:
- Serum pools at low and high levels of each analyte.
- Each measurement of the blank and the sample containing the interferent was repeated at least 5 times.
- Data Provenance: Serum pools.
-
Detection Limit:
- LoB: 5 blank serum samples measured in 4 replicates for 3 days (total of 60 measurements).
- LoD: 5 low-level samples measured in 4 replicates for 3 days (total of 60 measurements).
- LoQ: 10 samples that span the low end of linearity, measured 5 times each day for 3 days (total of 150 measurements).
- Data Provenance: Serum samples.
-
Stability and Calibration Frequency:
- Two fresh serum pools and two serum-based commercial controls.
- Measurements were repeated in triplicate at regular time points.
- Data Provenance: Serum pools and commercial controls.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The ground truth for performance studies like those described (method comparison, precision, linearity, interference, detection limits) for in vitro diagnostic (IVD) devices like these typically involves established reference methods or highly accurate comparative analyzers, rather than expert human interpretation of results. The document states that the performance of the Medicon Hellas reagents was compared with "comparator methods" (Beckman Coulter reagents on AU400, Abbott Diagnostics reagents on Architect c8000), which serve as the reference for ground truth in these types of analytical performance studies. The qualifications of the operators performing these studies are not specified.
4. Adjudication method for the test set
This information is not applicable and therefore not provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies where human interpretation of complex data (like medical images) is involved and a consensus is needed to establish ground truth. For quantitative chemical assays, the ground truth is established by the highly precise and accurate measurement of reference methods or predicate devices.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and therefore not provided. MRMC studies are specific to evaluating diagnostic devices where human readers interpret medical cases, often with and without AI assistance (e.g., radiology AI). The Medicon Hellas devices are in vitro diagnostic reagents for quantitative chemical measurements in serum, not image-based diagnostic tools that require human "readers" in the context of an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable in the traditional sense for these in vitro diagnostic reagents. These devices are chemical assays that produce a quantitative numerical output (e.g., cholesterol level in mg/dL). There isn't an "algorithm" making a diagnostic interpretation independently in the way AI software would. The device is the standalone measurement system. Its performance is evaluated independently through analytical studies (precision, linearity, accuracy against reference methods, etc.). The results are then read and interpreted by a human clinician.
7. The type of ground truth used
The ground truth for the analytical performance studies (precision, linearity, interference, detection limits, and method comparison) was established against:
- Reference Methods/Predicate Devices: For method comparison, the device's performance was compared against established comparator methods (Beckman Coulter reagents on AU400 and Abbott Diagnostics reagents on ABBOTT Architect c8000). The document explicitly states these as the comparators.
- A Priori Values/Established Standards: For linearity, precision, and detection limits, the ground truth is based on the known concentrations of prepared samples (e.g., serially diluted pools, spiked samples, blank serum) and statistical analysis according to CLSI guidelines.
- Traceability to Reference Methods/Materials: For Cholesterol and Triglycerides, traceability is to Gas-chromatography-isotope dilution mass spectrometry (GC-IDMS). For HDL-Cholesterol and LDL-Cholesterol, traceability is to the Abell-Kendall (AK) reference method.
8. The sample size for the training set
This information is not applicable and therefore not provided. These are chemical reagent kits, not machine learning (AI/ML) models that require a "training set" in the computational sense. The development of such reagents involves chemical and enzymatic research and optimization, often tested with various batches and concentrations of samples during R&D. The studies described in this document are for validation and verification of the final device, not for "training" an algorithm.
9. How the ground truth for the training set was established
As noted above, the concept of a "training set" with ground truth established in the AI/ML sense is not applicable to these chemical reagent devices. The "ground truth" for evaluating the analytical performance of the developed reagent kits is established through the reference methods and standardized protocols described in section 7.
Ask a specific question about this device
(344 days)
Class II | 21 CFR 862.1345 | Clinical Chemistry |
| CHH | Class I | 21 CFR 862.1175
The Cholestech LDX™ System is a small, portable analyzer and test cassette system is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes:
The Lipid Profile GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density Ilpoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.
The TC+HDL GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood.
The TC GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood.
The Lipid Profile cassette is for the quantitative determination of total cholesterol. HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.
The TC+HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood.
The TC cassette is for the quantitative determination of total cholesterol in whole blood.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Triglyceride measurements are used in the diagnosis and treatment with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The Cholestech LDX ™ system combines enzymatic methodology and solid-phase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to the Cholestech LDX™ cassette®.
The cassette is then placed into the Cholestech LDX™ Analyzer where a unique system on the cassette separates the plasma from the blood cells. A portion of the plasma flows to the right side of the cassette and is transferred to both the total cholesterol and triglyceride reaction pads. Simultaneously, plasma flows to the left side of the cassette where the low- and very low-density lipoproteins (LDL and VLDL) are precipitated with dextran sulfate (50,000 MW) and magnesium acetate precipitating reagent.The filtrate, containing both glucose and HDL cholesterol, is transferred to both the glucose and HDL cholesterol reaction pads.
The Cholestech LDX ™ Analyzer measures total cholesterol and HDL cholesterol by an enzymatic method based on the method formulation of Allain et al, and Roeschlau. Cholesterol esterase hydrolyzes the cholesterol esters in the filtrate or plasma to free cholesterol and the corresponding fatty acid. Cholesterol oxidase, in the presence of oxygen, oxidizes free cholesterol to cholest-4-ene-3-one and hydrogen peroxide. In a reaction catalyzed by horseradish peroxidase, the peroxide reacts with 4-Aminoantipyrine and N-ethyl-N-sulfohydroxypropyl-m-toluidine, sodium sale (TOOS) to form a purple-colored quinoneimine dye proportional to the total cholesterol and HDL cholesterol concentrations of the sample.
The analyzer measures triglycerides by an enzymatic method based on the hydrolysis of triglycerides by lipase to glycerol and free fatty acids. Glycerol, in a reaction catalyzed by glycerol kinase, is converted to glycerol-3-phosphate. In a third reaction, glycerol-3phosphate is oxidized by glycerol phosphate oxidase to dihydroxyacetone phosphate and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as for the total cholesterol and HDL cholesterol. Estimated LDL cholesterol and non-HDL cholesterol and a TC/HDL ratio are calculated using the measured values for TC, HDL, and Triglycerides.
The analyzer measures glucose by an enzymatic method that uses glucose oxidase to catalyze the oxidation of glucose to gluconolactone and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as that for total cholesterol, HDL cholesterol and triglycerides. The resultant color in all the reactions is measured by reflectance photometry.
A brown (magnetic) stripe on each cassette contains the calibration information required for the Cholestech LDX ™ Analyzer to convert the reflectance reading (% R) to the total cholesterol, HDL cholesterol, triglycerides and glucose concentrations.
The provided text is a 510(k) summary for the Cholestech LDX™ System and primarily discusses device modifications and comparison to a predicate device. It certifies that verification studies were performed as required by risk analysis and all acceptance criteria were met. However, it does not provide the specific details of the acceptance criteria or the reported device performance for these studies. It also does not contain information about the sample size, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or how ground truth was established for test and training sets.
Therefore, based solely on the provided text, I cannot fulfill most of the requested information regarding the study that proves the device meets the acceptance criteria. The document states that such studies were done and met acceptance criteria, but omits the specifics.
Here's what can be inferred or stated from the provided text, and what is missing:
Table of Acceptance Criteria and Reported Device Performance
Information Not Available in the Text: The document explicitly states, "Verification studies were performed as required by risk analysis and all acceptance criteria were met." However, it does not list the specific acceptance criteria (e.g., specific accuracy thresholds, precision ranges, etc.) or the detailed reported device performance (e.g., actual measured accuracy, precision values, etc.) from these studies. The modification pertains to updating the performance claim related to conjugated and unconjugated Bilirubin interference. While it mentions that less than 10% interference was seen at specified levels for various substances, this is a general statement from the predicate device's limitations, not a specific acceptance criterion for the current modification or the exact performance data achieved.
2. Sample Size Used for the Test Set and the Data Provenance
Information Not Available in the Text: The document states that "verification studies" were performed, but it does not specify the sample size (e.g., number of patients, number of samples) used for any test set or the provenance of the data (e.g., country of origin, retrospective or prospective nature of the data collection).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Information Not Available in the Text: The document details changes to an in vitro diagnostic (IVD) device for measuring cholesterol, triglycerides, and glucose. For IVD devices, ground truth is typically established by reference laboratory methods, not by human experts interpreting images or clinical cases. Therefore, the concept of "experts" as in radiologists or pathologists establishing ground truth is not applicable here. Even if it were (e.g., for method comparison studies requiring expert clinical correlation), the document does not mention any role for experts in establishing ground truth.
4. Adjudication Method for the Test Set
Information Not Available in the Text: Since the ground truth for an IVD device is generally established using reference methods (as opposed to human interpretation needing adjudication), an adjudication method as typically used in AI studies of imaging (e.g., 2+1, 3+1) is not applicable or described in this document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Information Not Applicable/Available in the Text: This is an in vitro diagnostic (IVD) device, not an AI-assisted diagnostic imaging device. Therefore, MRMC studies comparing human readers with and without AI assistance are not relevant to this type of device and are not mentioned in the documentation.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Information Not Applicable/Available in the Text: The Cholestech LDX™ System is a chemical analyzer, not an AI algorithm. Its performance is inherent to the device's enzymatic and solid-phase technology. The concept of "standalone algorithm performance" without human-in-the-loop is not directly applicable in the same way it would be for a software-as-a-medical-device (SaMD) that processes and interprets data for human review. The document describes the device's direct measurement capabilities.
7. The Type of Ground Truth Used
Inferred from Text: For an in vitro diagnostic device measuring analytes (cholesterol, HDL, triglycerides, glucose), the ground truth is typically established by reference laboratory methods (e.g., highly accurate and precise methods run on core laboratory instruments). While the document does not explicitly state "reference laboratory comparison" for ground truth, the context of an IVD device submission, especially one measuring these specific analytes, strongly implies this method.
8. The Sample Size for the Training Set
Information Not Applicable/Available in the Text: This is a chemical analyzer, not a machine learning or AI-based device that requires a "training set" in the computational sense. The device's operation is based on established enzymatic and chemical reactions, not on data-driven learning. Therefore, there is no "training set" in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established
Information Not Applicable/Available in the Text: As noted above, there is no "training set" for this type of IVD device in the context of AI/ML. The device's calibration and performance are based on chemical principles and validation studies, not on learning from a training dataset.
Ask a specific question about this device
(568 days)
Longford, Ireland
Re: K203597
Trade/Device Name: Cholesterol2 Regulation Number: 21 CFR 862.1175 Regulation
Classification Name: Enzymatic esterase--oxidase, cholesterol Governing Regulation Number: 21 CFR §862.1175
The Cholesterol2 assay is used for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. The Cholesterol2 assay is to be used as an aid in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
The Cholesterol2 assay is an automated clinical chemistry assay for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. Cholesterol esters are enzymatically hydrolyzed by cholesterol esterase to cholesterol and free fatty acids. Free cholesterol, including that originally present, is then oxidized by cholesterol oxidase to cholest-4-ene-3-one and hydrogen peroxide. The hydrogen peroxide oxidatively couples with N,N-Bis(4-sulfobutyl)-3-methylaniline (TODB) and 4-aminoantipyrine to form a chromophore (quinoneimine dye) which is quantitated at 604 nm.
The provided text describes the Abbott Cholesterol2 assay, an in vitro diagnostic device for quantifying cholesterol in human serum or plasma.
Here's an analysis of the acceptance criteria and study data:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria for each performance characteristic. Instead, it presents the results of various studies as proof of device performance. The table below summarizes the reported performance, which implicitly serves as the "met" criteria.
| Performance Characteristic | Reported Device Performance (Cholesterol2) |
|---|---|
| Analytical Measuring Interval (AMI) | 5-748 mg/dL |
| Extended Measuring Interval (EMI) | 748-2992 mg/dL |
| Reportable Interval | 2-2992 mg/dL |
| Precision | |
| Control Level 1 (251 mg/dL) | SD: 1.9 mg/dL (Within-Run), 2.6-3.1 mg/dL (Within-Laboratory); %CV: 0.7% (Within-Run), 1.0-1.2% (Within-Laboratory) |
| Control Level 2 (106 mg/dL) | SD: 1.0 mg/dL (Within-Run), 1.3-1.7 mg/dL (Within-Laboratory); %CV: 1.0% (Within-Run), 1.2-1.6% (Within-Laboratory) |
| Panel A (21 mg/dL) | SD: 0.6 mg/dL (Within-Run), 0.7-0.8 mg/dL (Within-Laboratory); %CV: 3.0% (Within-Run), 3.2-4.1% (Within-Laboratory) |
| Panel B (237 mg/dL) | SD: 2.8 mg/dL (Within-Run), 3.7-4.9 mg/dL (Within-Laboratory); %CV: 1.2% (Within-Run), 1.5-2.0% (Within-Laboratory) |
| Panel C (718 mg/dL) | SD: 6.4 mg/dL (Within-Run), 4.6-6.9 mg/dL (Within-Laboratory); %CV: 0.9% (Within-Run), 0.7-1.0% (Within-Laboratory) |
| Limit of Blank (LoB) | 0 mg/dL |
| Limit of Detection (LoD) | 2 mg/dL |
| Limit of Quantitation (LoQ) | 5 mg/dL (at 20% CV maximum allowable precision) |
| Linearity | Linear across the analytical measuring interval of 5 to 748 mg/dL |
| Interference (Endogenous) | |
| Conjugated Bilirubin (7 mg/dL) | No significant interference (within ± 10%) |
| Unconjugated Bilirubin (11 mg/dL) | No significant interference (within ± 10%) |
| Hemoglobin (1000 mg/dL) | No significant interference (within ± 10%) |
| Total Protein (15 g/dL) | No significant interference (within ± 10%) |
| Conjugated Bilirubin (40 mg/dL) | Interference: -39% at 150 mg/dL analyte, -31% at 220 mg/dL analyte |
| Unconjugated Bilirubin (16 mg/dL) | Interference: -11% at 150 mg/dL analyte |
| Interference (Exogenous) | |
| Acetaminophen (160 mg/L) | No significant interference |
| Acetylcysteine (150 mg/L) | No significant interference |
| Acetylsalicylic acid (30 mg/L) | No significant interference |
| Aminoantipyrine (40 mg/L) | No significant interference |
| Ampicillin-Na (80 mg/L) | No significant interference |
| Biotin (4250 ng/mL) | No significant interference |
| Ca-dobesilate (60 mg/L) | No significant interference |
| Cefotaxime (53 mg/dL) | No significant interference |
| Cefoxitin (6600 mg/L) | No significant interference |
| Cyclosporine (2 mg/L) | No significant interference |
| Desacetylcefotaxime (6 mg/dL) | No significant interference |
| Dipyrone (100 mg/L) | No significant interference |
| Dobutamine (0.2 mg/dL) | No significant interference |
| Doxycycline (20 mg/L) | No significant interference |
| Ibuprofen (220 mg/L) | No significant interference |
| Intralipid (1050 mg/dL) | No significant interference |
| Levodopa (8 mg/L) | No significant interference |
| Methotrexate (140 mg/dL) | No significant interference |
| Metronidazole (130 mg/L) | No significant interference |
| Methylaminoantipyrine (40 mg/L) | No significant interference |
| Methyldopa (20 mg/L) | No significant interference |
| N-Acetyl-p-benzoquinone (NAPQI) (20 mg/L) | No significant interference |
| Phenylbutazone (330 mg/L) | No significant interference |
| Phenytoin (6 mg/dL) | No significant interference |
| Rifampicin (50 mg/L) | No significant interference |
| Sodium heparin (4 U/mL) | No significant interference |
| Sulpiride (15 mg/L) | No significant interference |
| Theophylline (60 mg/L) | No significant interference |
| Ascorbic acid (60 mg/L) | Interference: -10% at 150 mg/dL analyte |
| Intralipid (2000 mg/dL) | Interference: -27% at 150 mg/dL analyte, -22% at 220 mg/dL analyte |
| Methyldopa (30 mg/L) | Interference: -14% at 150 mg/dL analyte |
| Method Comparison (vs. Predicate) | |
| Serum (n=138) | Correlation Coefficient: 1.00; Intercept: 0.41; Slope: 0.98 (Range: 7-684 mg/dL) |
| Tube Type Suitability | Acceptable for Serum, Serum separator, Lithium heparin, Lithium heparin separator, Sodium heparin tubes. |
| Dilution Verification | Automated dilution protocol (1:5.97) and manual dilution procedure (1:4) evaluated. (Performance details not provided in summary). |
| Traceability | Traceable to National Reference System for Cholesterol (Abell-Kendall reference method in a CDC-Certified CRMLN). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Precision Study: 2 controls and 3 human serum panels were tested. Each sample was tested in duplicate, twice per day for 20 days. This means 80 measurements per sample (2 duplicates x 2 times/day x 20 days).
- Lower Limits of Measurement Study: n ≥ 60 replicates for LoB, LoD, and LoQ determinations. They used low-analyte level samples and zero-analyte samples.
- Linearity Study: The number of samples for the linearity study is not explicitly stated, but it covered the range of 5 to 748 mg/dL.
- Interference Studies: Each endogenous substance was tested at 2 analyte levels (approximately 150 mg/dL and 220 mg/dL). Exogenous substances were tested at various specified interferent levels. The number of samples for each interferent is not provided.
- Method Comparison Study: 138 serum samples were used.
- Tube Type Study: Samples were collected from a minimum of 40 donors.
- Dilution Verification: 8 human serum samples.
Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the context of medical device regulatory submission, these are typically prospective studies conducted in a controlled laboratory environment. The "human serum panels" and "human serum samples" imply human-derived biological samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device is an in vitro diagnostic (IVD) chemistry assay. The concept of "experts establishing ground truth" as it applies to image interpretation or clinical diagnosis by medical professionals (like radiologists) does not directly apply here in the same way.
For IVDs like this, the "ground truth" or reference values are established through:
- Reference methods: The Cholesterol2 reagent is certified to be traceable to the National Reference System for cholesterol, against the Abell-Kendall reference method in a CDC-Certified Cholesterol Reference Method Laboratory Network (CRMLN). The Abell-Kendall method is considered the gold standard for cholesterol measurement.
- Analytically Validated Methods: For values outside the AMI but within the EMI, samples were "value assigned using the analytically validated method."
- Known concentrations: For studies like linearity, spiked samples with known concentrations are used.
Therefore, the "experts" are the methodologists and laboratory professionals overseeing and validating the reference methods and the analytical validation processes. No specific number or qualification of clinical experts (e.g., radiologists) is relevant for establishing the ground truth for a quantitative chemistry assay.
This is a standalone performance evaluation of the assay itself, demonstrating its analytical accuracy, precision, and robustness.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The concept of "adjudication" (e.g., 2+1, 3+1 where multiple human readers agree or have a tie-breaker by an expert) is not applicable to this type of quantitative diagnostic assay. The results are numerical values generated by the automated instrument and reagents. Deviations or discrepancies would be resolved through re-testing, quality control, or investigation into instrument or reagent issues, rather than human expert adjudication of a "diagnosis."
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging device or an AI assistant for human readers. Its output is a quantitative measurement of cholesterol, not an interpretation that requires human "reading" or decision support.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, a standalone performance evaluation of the device (Cholesterol2 assay on the ARCHITECT c8000 System) was done. The studies described (reportable interval, precision, lower limits of measurement, linearity, interference, method comparison, tube type, dilution verification) all evaluate the analytical performance of the assay and instrument directly, without human interpretation as part of the primary outcome measure. The output is a numerical concentration of cholesterol.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth used for this quantitative assay primarily relies on:
- Reference Methods: Specifically, the Abell-Kendall reference method, which is considered the gold standard for cholesterol measurement and is used in CDC-Certified Cholesterol Reference Method Laboratory Networks (CRMLN). The device's traceability to this method is explicitly stated.
- Analytically Validated Methods: For verifying values in the extended measuring interval.
- Known Spiked Concentrations: For studies such as linearity and dilution verification, where samples are prepared with precisely known concentrations.
This is an analytical ground truth, not a clinical ground truth derived from pathology or patient outcomes.
8. The sample size for the training set
This document does not describe a typical "training set" in the context of machine learning or AI. This is a chemistry assay that uses reagents and enzymatic reactions, not an algorithm that is "trained" on data. Therefore, the concept of a training set as used in AI development is not applicable here. The assay's analytical characteristics are determined through standard laboratory validation studies.
9. How the ground truth for the training set was established
As explained above, there is no "training set" in the AI sense for this device. The analytical accuracy and reliability are established through comparisons to certified reference methods and known standard concentrations, as described in point 7.
Ask a specific question about this device
(787 days)
| 21 CFR 862.1435 Ketones (nonquantitative) test system |
| | 21 CFR 862.1175
The FORA ADVANCED GD40 Glucose, ß-Ketone and Cholesterol Monitoring System consists of the FORA ADVANCED GD40 meter, the FORA ADVANCED GD40 Blood Glucose strips, the FORA ADVANCED GD40 B-Ketone strips, and the FORA ADVANCED GD40 Total Cholesterol strips.
The FORA ADVANCED GD40 Glucose, B-Ketone and Cholesterol Monitoring System is intended for the quantitative measurement of glucose, beta-hydroxybutyrate (8-ketone), and cholesterol in fresh capillary whole blood from the finger. This system is intended for single-patient home use and should not be intended for use by patients with diabetes. It is only for use outside the body (in vitro diagnostic use).
Glucose and {}-ketone measurements are used as an aid to monitor the effectiveness of a diabetes control program. Glucose measurements should not be used for the diagnosis of or screening for diabetes.
Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Cholesterol should be measured at the frequency recommended by your healthcare provider.
FORA ADVANCED GD40 Glucose, ß-Ketone and Cholesterol Monitoring System includes the FORA ADVANCED GD40 Meter, analyte-specific test strips (FORA ADVANCED GD40 Blood Glucose Test Strips, FORA ADVANCED GD40 Blood Cholesterol Test Strips, and FORA ADVANCED GD40 Blood Ketone Test Strips) and control solutions (FORA Glucose Control Solutions, B-Ketone Control Solutions and Total Cholesterol Control Solutions).
The glucose/9-ketone test strips and control solutions utilized in the FORA ADVANCED GD40 Glucose, B-Ketone and Cholesterol Monitoring System are the same as the predicate, FORA ADVANCED GD40 Blood Glucose and ß-Ketone Monitoring System (K161738).
The provided text describes the FORA ADVANCED GD40 Glucose, β-Ketone and Cholesterol Monitoring System, a device for measuring glucose, beta-hydroxybutyrate (β-ketone), and cholesterol in fresh capillary whole blood. It is a 510(k) summary, therefore, it focuses on demonstrating substantial equivalence to predicate devices rather than a detailed clinical study for a novel device.
Based on the provided information, here's a breakdown of the acceptance criteria and the study conducted to prove the device meets these criteria:
Key Takeaway: The document primarily focuses on non-clinical and clinical "studies" to demonstrate performance characteristics and compare them to predicate devices, rather than a single, large-scale comparative effectiveness study with human readers (MRMC). The "acceptance criteria" are implied by the performance characteristics tested and similarity to the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as numerical targets in a table within the provided text. Instead, the document discusses "performance criteria were met" and that the device "met design specifications and requirements." The "acceptance criteria" are implied by meeting the established ranges and performance levels comparable to the predicate devices.
However, the "performance" is reported through the summary of testing. Since this is a 510(k) submission, the primary "performance" shown is that the device is substantially equivalent to the predicate devices. The listed similarities and differences in characteristics serve as the de facto "reported device performance."
| Performance Characteristic/Acceptance Criterion (Implied) | Device Performance (FORA ADVANCED GD40) | Notes |
|---|---|---|
| Glucose Measurement | ||
| Intended Use | Same as Predicate: quantitative measurement in fresh capillary whole blood from the finger to monitor effectiveness of diabetes control program. Not for diagnosis/screening. | Demonstrated equivalence. |
| Setting | Over-the-counter Home use. | Same as Predicate. |
| Test Range | 20 – 600 mg/dL. | Same as Predicate. |
| Sample Type | Fingertip Capillary whole blood. | Same as Predicate. |
| Sample Volume | 0.9 μl. | Same as Predicate. |
| Test Time | 5 sec. | Same as Predicate. |
| Calibration | Automatic. | Same as Predicate. |
| Operating Principle | Enzymatic (Glucose Dehydrogenase). | Same as Predicate. |
| Data storage | 1000 Results with Date/Time. | Same as Predicate. |
| Weight | 71 g. | Same as Predicate. |
| Non-Clinical Performance (Precision, Linearity, Interference, Flex, Disinfection, Robustness) | Met design specifications and requirements. | Confirmed via design verification and validation testing. Specific numerical results are not provided in this summary. |
| Clinical Performance (System Accuracy, Operation, Ease of Use) | Confirmed. | Confirmed via user evaluation. Specific numerical results or study methodology (e.g., sample size, metrics) not provided in this summary. |
| β-Ketone Measurement | ||
| Intended Use | Same as Predicate: quantitative measurement in fresh capillary whole blood from the finger to monitor effectiveness of diabetes control program. Not for diagnosis/screening. | Demonstrated equivalence. |
| Setting | Over-the-counter Home use. | Same as Predicate. |
| Test Range | 20 – 600 mg/dL (same as glucose listed in table, but typical ketone ranges are different, likely copied from glucose for tabular similarity, the actual range for β-Ketone would be specified in detailed report). | Needs clarification as this range is for glucose. β-Ketone measurements are based on B-hydroxybutyrate dehydrogenase. However, the document clearly states the β-Ketone test system is 'Class I, meets the limitations of exemption 21 CFR 862.9(c)(5)', implying minimal testing required for equivalence assuming the core technology (test strips and control solutions) are the same as the predicate (K161738), as stated in Device Description section E. |
| Non-Clinical Performance | Met design specifications and requirements. | Confirmed via design verification and validation testing. Specific numerical results are not provided in this summary. The glucose/ß-ketone test strips and control solutions are stated to be "the same as the predicate," which simplifies the testing needed to demonstrate substantial equivalence for these analytes. |
| Clinical Performance | Confirmed. | Confirmed via user evaluation. |
| Cholesterol Measurement | ||
| Intended Use | Same as Predicate: diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism. | Demonstrated equivalence. |
| Setting | Over-the-counter Home use. (Predicate: Over-the-counter Home use and Prescription use). | Difference noted: Proposed device is only OTC Home use, while predicate was also Prescription use. This is a narrowing of intended use, which is generally acceptable for substantial equivalence. |
| Test Range | 100 – 400 mg/dL. | Same as Predicate. |
| Sample Type | Fingertip Capillary whole blood. | Same as Predicate. |
| Sample Volume | 5 μl. (Predicate: 10 μl) | Difference noted: Lower sample volume on the proposed device. This is typically viewed as an improvement, but would have been supported by specific performance data in the full submission. |
| Test Time | 60 sec. (Predicate: < 2 min) | Difference noted: Faster test time. This is typically viewed as an improvement, supported by specific performance data. |
| Calibration | Same as Predicate: Calibration strip. | Demonstrated equivalence. |
| Operating Principle | Enzymatic (Cholesterol esterase to free cholesterol, then oxidized by cholesterol oxidase). | Same as Predicate. |
| Data storage | 1000 Results with Date/Time. (Predicate: 200 Results) | Difference noted: Significantly more data storage. This is an improvement. |
| Weight | 71 g. (Predicate: 145 g) | Difference noted: Significantly lighter. This is an improvement. |
| Non-Clinical Performance | Met design specifications and requirements. | Confirmed via design verification and validation testing. Specific numerical results for precision, linearity, interference, flex studies, disinfection, cleaning, robustness, and software validation are not provided, but the conclusion states they were met. |
| Clinical Performance | Confirmed. | Confirmed via user evaluation. |
| General Device Performance | ||
| Electrical/Mechanical Safety | Met specifications. | Confirmed via testing. |
| Software Validation (Moderate Level of Concern) | Performed and met requirements as per FDA Guidance. |
2. Sample Size and Data Provenance
- Test Set Sample Size: The document does not specify the sample sizes for the "non-clinical" or "clinical" studies. It only states that "Non-clinical and clinical studies were conducted to test, verify and validate the performance."
- Data Provenance: The document does not explicitly state the country of origin for the data or whether the studies were retrospective or prospective. Given the submitter's address (TaiDoc Technology Corporation, New Taipei City, Taiwan), it's highly probable the studies were conducted in Taiwan or involved participants from that region. The studies are described as "verification and validation testing" and "user evaluation," which suggests prospective data collection for the purpose of regulatory submission.
3. Number of Experts and Qualifications for Ground Truth
- This information is not provided in the summary. Since the device is a direct measurement system (glucose, ketone, cholesterol) and relies on chemical reactions rather than interpretation of medical images or complex diagnostic algorithms, the "ground truth" would typically be established by laboratory reference methods, not by expert consensus of clinicians/radiologists.
- Therefore, the concept of "experts" in the context of establishing ground truth for this type of device (an in-vitro diagnostic) would refer to the use of highly accurate laboratory equipment and trained laboratory personnel following established protocols, rather than medical specialists.
4. Adjudication Method (for test set)
- This is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are used in studies where human readers (often radiologists) interpret medical images and their interpretations need to be reconciled to establish ground truth or compare to AI performance. For an in-vitro diagnostic device that measures chemical concentrations, the ground truth comes from laboratory reference methods, not subjective human interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done (or at least not described in this summary). This type of study is relevant for imaging AI devices where the effect of AI assistance on human reader performance is evaluated.
- This device is an in-vitro diagnostic for direct chemical measurement, not an imaging AI. Therefore, an MRMC study is not relevant for demonstrating its performance or substantial equivalence.
6. Standalone (Algorithm Only) Performance
- No, not specified as a separate study type in the context of imaging AI.
- However, the "Non-Clinical Testing Summary" mentions "functional performance tests (precision, linearity, interference, flex studies)." These are essentially "standalone" performance evaluations of the device's accuracy and robustness under various conditions, independent of a human user's direct interpretive "performance" in the way an AI algorithm's standalone performance might be assessed. The "clinical testing" included a "user evaluation" to confirm accuracy and ease of use in a real-world scenario.
7. Type of Ground Truth Used
The ground truth used for this device would be laboratory reference methods.
- The test principle explicitly states that the device is "calibrated to display the equivalent of plasma glucose values to allow comparison of results with laboratory methods."
- The non-clinical testing included "precision, linearity, interference, flex studies," all of which require comparison to a highly accurate reference method to establish the true concentration of the analytes (glucose, β-ketone, cholesterol).
- For the β-ketone and cholesterol measurements, the process involves enzymatic reactions that are proportional to the amount of the substance in the sample, which implies quantitative comparison to a known concentration.
8. Sample Size for the Training Set
- This information is not applicable and not provided. This device is a traditional in-vitro diagnostic, not a machine learning or AI-based device that would require a "training set" in the sense of AI model development. Its calibration and function are based on electrochemical biosensor technology and established chemical principles.
9. How the Ground Truth for the Training Set was Established
- This is not applicable as there is no "training set" in the context of an AI/ML algorithm. The device's operational parameters would be established through calibration against known standards and validation against reference laboratory methods during its development and manufacturing process.
Ask a specific question about this device
(810 days)
- |
| СНН | Class I, meets thelimitation of exemption 21CFR 862.9(c)(4) | 21 CFR 862.1175 - |
| CHH | Class I, meets thelimitation of exemption 21CFR 862.9(c)(4) | 21 CFR 862.1175
The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
A Chol/HDL ratio is calculated by the CardioChek Plus analyzer.
The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
A Chol/HDL ratio is calculated by the CardioChek Plus analyzer.
The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Lipid Panel test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Plus Analyzer.
The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels eGlu test strips) is for the quantitative determination of glucose in venous whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The CardioChek Plus Home Test System (consisting of the CardioChek Plus Home analyzer and CardioChek Plus Home Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The CardioChek Plus Home Test System (consisting of the CardioChek Plus Home analyzer and CardioChek Plus Home Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
A Chol/HDL ratio is calculated by the CardioChek Plus Home analyzer
The CardioChek Plus Home Test System (consisting of the CardioChek Plus Home analyzer and CardioChek Plus Home Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
A Chol/HDL ratio is calculated by the CardioChek Plus Home analyzer.
The CardioChek Plus Home Test System (consisting of the CardioChek Plus Home analyzer and CardioChek Plus Home Lipid Panel test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.
· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Plus Home Analyzer.
The CardioChek Plus Home Test System (consisting of the CardioChek Plus Home analyzer and CardioChek Plus Home eGlu test strips) is for the quantitative determination of glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The CardioChek Plus and CardioChek Plus Home analyzers are professional (Rx) use (CardioChek Plus) and home use/OTC (CardioChek Plus Home) in vitro diagnostic systems to measure various analytes in capillary fingerstick whole blood for both home and professional use and in venous whole blood for professional use only for glucose, cholesterol, HDL cholesterol, and triglycerides. The system includes a small analyzer and test strips. The analyzers utilize PTS Diagnostics brands of dry strip chemistry test strips. The test strips are single-use and utilize one of two types of technologies: reflectance photometry and amperometric/electrochemical technology. The test strips are used with the CardioChek Plus and CardioChek Plus Home analyzers to measure total cholesterol, HDL cholesterol, triglycerides, and glucose in whole blood. The test strips utilize enzymatic methods on dry colorimetric test strips that are read by reflectance photometry or amperometric/electrochemical test strips that measure the current produced when blood is applied to the test strip. These test strips are for in vitro diagnostic use only. The analyzer has software that converts the reflectance or current produced into an analyte concentration by comparing the reading to a lot-specific calibration curve that is programmed into a EEPROM MEMo chip that is inserted into the analyzer. Each vial of test strips includes a lot-specific MEMo chip, thus eliminating any need for the user to calibrate the system. The analyzer is powered by 4 AA alkaline batteries.
The supplied document is a 510(k) premarket notification for a modification to an existing device, the CardioChek Plus Test System and CardioChek Plus Home Test System. This notification focuses on a design change to the battery compartment of the analyzer case to prevent overheating due to incorrect battery insertion. Therefore, the information provided primarily addresses the modification and its impact, rather than a comprehensive study on the overall device performance against acceptance criteria for its intended use (measurement of cholesterol, HDL, triglycerides, and glucose).
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of acceptance criteria for the analytical performance of the CardioChek Plus Test System (e.g., accuracy, precision for cholesterol, glucose, etc.) nor does it report the device performance against such criteria. This is because the submission is for a modification to an already cleared device, K140068, which likely established those initial performance criteria. The current submission's "testing" section is solely focused on the battery compartment modification.
| Acceptance Criteria (for the battery compartment modification) | Reported Device Performance |
|---|---|
| Prevention of contact between negative battery terminal and positive terminal of analyzer case when batteries are accidentally inserted with polarity reversed. | New design prevented contact between negative battery terminal and positive terminal of case when batteries are accidentally inserted with polarity reversed. The modification eliminated any potential concerns associated with incorrect insertion of the batteries. |
| Elimination of potential for battery overheating if batteries are accidentally inserted with polarity reversed. | The modification prevented any battery overheating. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a distinct "test set sample size" for the battery compartment modification study in terms of number of devices or number of tests. It broadly states "Testing of the modified battery compartment of the case showed that the new design prevented the negative terminal...". The nature of this testing would likely involve physical examination and insertion attempts, rather than a large-scale clinical study.
Data Provenance: Not specified, but implied to be from the manufacturer's internal testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable to the type of modification described. Ground truth in this context would likely be engineering verification that the physical design change addresses the identified safety issue. No external experts or their qualifications are mentioned for this specific modification testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This type of engineering verification for a physical modification does not typically involve adjudication methods like those used in clinical or image-based studies.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a diagnostic test system for quantitative determination of analytes, not an AI-assisted diagnostic imaging or interpretation tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a physical diagnostic system, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the battery compartment modification, the "ground truth" would be the successful physical prevention of contact between the incorrectly inserted negative battery terminal and the positive terminal of the case, and the absence of battery overheating, verified through engineering tests. This is a technical, hardware-level verification, not based on clinical "expert consensus, pathology, or outcomes data."
For the original performance of the analytes (cholesterol, glucose, etc.), the ground truth would likely have been established during the clearance of the predicate device (K140068) using reference methods or clinical laboratory analyzers. However, this document does not detail those studies.
8. The sample size for the training set
Not applicable. This is a hardware modification for an existing device, not an AI or algorithm-based device that would require a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set is relevant to this submission.
Ask a specific question about this device
(27 days)
Trade/Device Name: VITROS XT Chemistry Products TRIG-CHOL Slides Regulation Number: 21 CFR 862.1175
Triglyceride testsystem | ClinicalChemistry |
| CHOL | CHH | Class I* | 21 CFR 862.1175
Rx Only For in vitro diagnostic use only
The TRIG test within the VITROS XT Chemistry Products TRIG-CHOL Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver diseases involving lipid metabolism, or various endocrine disorders.
The CHOL test within the VITROS XT Chemistry Products TRIG-CHOL Slides quantitatively measure cholesterol (CHOL) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.
The new device, the VITROS XT Chemistry Products TRIG-CHOL Slide is a single device that contains both a TRIG test and a CHOL test multilayered, analytical elements coated on a polyester support separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur and test results are generated for each analyte independently of the other analyte.
To perform the TRIG test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. The Triton X-100 surfactant in the spreading layer aids in dissociating the triglycerides from lipoprotein complexes present in the sample. The triglyceride molecules are then hydrolyzed by lipase to yield glycerol and fatty acids. Glycerol diffuses to the reagent layer, where it is phosphorylated by glycerol kinase in the presence of adenosine triphosphate (ATP). In the presence of L-a-glycerolphosphate oxidase, L-α-glycerophosphate is then oxidized to dihydrox vacetone phosphate and hydrogen peroxide. The final reaction involves the oxidation of a leuco dye by hydrogen peroxide, catalyzed by peroxidase, to produce a dye. The density of the dye formed is proportional to the triglyceride concentration present in the sample and is measured by reflectance spectrophotometry.
To perform the CHOL test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. The Triton X-100 (TX100) surfactant in the spreading layer aids in dissociating the cholesterol and cholesterol esters from lipoprotein complexes present in the sample. Hydrolysis of the cholesterol esters to cholesterol is catalyzed by cholesterol ester hydrolase. Free cholesterol is then oxidized in the presence of cholesterol oxidase to form cholestenone and hydrogen peroxide. Finally, hydrogen peroxide oxidizes a leuco dye in the presence of peroxidase to generate a colored dye. The density of dye formed is proportional to the cholesterol concentration present in the sample and is measured by reflectance spectrophotometry.
The provided text describes the analytical performance of the VITROS XT Chemistry Products TRIG-CHOL Slides, an in vitro diagnostic device for quantitatively measuring triglyceride (TRIG) and cholesterol (CHOL) concentrations in serum and plasma.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of "acceptance criteria" but rather reports the analytical performance of the new device and shows its comparison to the predicate devices. The implicit acceptance criteria are that the new device performs at least as well as, and is substantially equivalent to, the predicate devices. The reported performance metrics are detailed in the tables for method comparison, precision, detection capability, and linearity.
TRIG Test
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance (VITROS XT TRIG-CHOL Slides) |
|---|---|---|
| Method Comparison | Substantially equivalent to predicate device (VITROS TRIG Slides) | Slope: 0.99, Intercept: 1.49, Correlation Coefficient: 1.000 |
| Precision (CV%) | Acceptable variability for clinical use | Pool 1: 1.3%, Native Pool: 1.5%, Control 1: 0.7%, Control 2: 0.8%, Pool 2: 0.9%, Pool 3: 0.9% |
| Limit of Quantitation (LoQ) | Clinically relevant LoQ for TRIG | 10 mg/dL (Criteria: %CV < 20%) |
| Linearity | Linear within claimed measuring range | Linear Range: 8.0 - 542.8 mg/dL (Claimed: 10 - 525 mg/dL) |
CHOL Test
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance (VITROS XT TRIG-CHOL Slides) |
|---|---|---|
| Method Comparison | Substantially equivalent to predicate device (VITROS CHOL Slides) | Slope: 0.97, Intercept: 0.09, Correlation Coefficient: 0.999 |
| Precision (CV%) | Acceptable variability for clinical use | Pool 1: 2.1%, Control 1: 1.9%, Native Pool: 1.3%, Pool 2: 1.5%, Control 2: 1.6%, Pool 3: 1.5% |
| Limit of Quantitation (LoQ) | Clinically relevant LoQ for CHOL | 50 mg/dL (Criteria: %CV < 9%) |
| Linearity | Linear within claimed measuring range | Linear Range: 27 - 358 mg/dL (Claimed: 50 - 325 mg/dL) |
2. Sample sizes used for the test set and the data provenance
- Method Comparison: 148 serum samples were used for both TRIG and CHOL tests.
- Precision: 80 observations (2 replicates per run, 2 runs per day over 20 days) for serum samples (patient pools and quality control materials).
- Detection Capability (LoQ): 180 determinations for both TRIG and CHOL.
- Linearity: Eighteen proportionally related admixtures of low and high test fluids, each tested in quadruplicate.
The data provenance is not explicitly stated in terms of country of origin or whether the samples were retrospective or prospective. It implies the use of patient samples, but specific details are absent.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This document describes the analytical performance of an in vitro diagnostic device, not a device that relies on expert human interpretation for its output (like an AI imaging device). Therefore, there is no mention of "experts used to establish the ground truth" in the way it would apply to a clinical imaging study or a study validating human performance. The "ground truth" for this type of device is established by its quantitative measurements against established analytical standards and reference methods/predicate devices.
4. Adjudication method for the test set
Not applicable. As noted above, this is an analytical performance study for an in vitro diagnostic device, not a human reader study requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an analytical performance study of an in vitro diagnostic device, not an AI-assisted diagnostic tool that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, this entire study represents a "standalone" performance evaluation of the device (VITROS XT Chemistry Products TRIG-CHOL Slides on the VITROS XT 7600 Integrated System) without human intervention in the result generation. The device quantitatively measures the analytes.
7. The type of ground truth used
The ground truth for this in vitro diagnostic device is established through:
- Comparison to a legally marketed predicate device: The VITROS Chemistry Products TRIG Slides and VITROS Chemistry Products CHOL Slides. This implies that the predicate devices serve as the established reference standard for performance.
- Established analytical methods and materials: CLSI protocols (EP09c, EP05-A3, EP17-A2, EP06-A, EP07-03) are referenced, which dictate the methodology for evaluating analytical performance parameters.
- Quality Control Materials and reference pools: Used in precision and linearity studies.
- Clinically established guidelines: NCEP guidelines are referenced for expected values/classification of TRIG and CHOL, indicating alignment with clinical understanding of these analytes.
8. The sample size for the training set
This document describes pre-market validation studies for a diagnostic test kit and instrument system. It does not refer to a "training set" in the context of machine learning or AI. The development of such a device involves internal optimization and development work, but the data presented in this 510(k) summary are for the validation of the finalized product.
9. How the ground truth for the training set was established
Not applicable, as this refers to a diagnostic test kit and instrument's analytical validation, not an AI or machine learning model that would involve a "training set" with ground truth established through expert annotation.
Ask a specific question about this device
(81 days)
Kit | | |
| Classification Names: | Cholesterol (Total) Test System (21 CFR 862.1175
The PTS Professional Chemistry Kit is a medical device convenience kit that quantitatively measures the percent of glycated hemoglobin (%Alc), total cholesterol, high density lipoprotein (HDL), triglycerides, and glucose in capillary (fingerstick) or venous whole blood.
· Cholesterol measurements are used in the diagnosis and treatments of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
• The HbA1C test provides quantitative measurement of the percent of glycated hemoglobin levels. The test is for professional use to monitor glycemic control in people with diabetes.
The PTS Professional Chemistry Kit combines devices that are already cleared and packages them together for the convenience of the user.
There are two CardioChek® Analyzers: the CardioChek® PA Analyzer and the CardioChek® Plus Analyzer. These analyzers are part of portable test systems. The CardioChek® PA test system and the CardioChek® Plus test system consist of their respective analyzer and PTS Panels® Test Strips with a lotspecific MEMo® Chip.
The CardioChek® PA and Plus test systems are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and are intended for multiple patient uses in professional healthcare settings. These systems are for in vitro diagnostic use only.
The PTS Panels® Test Strips consist of varying amounts of test strips packed into a vial and a lot-specific MEMo® Chip. PTS Panels® Test Strips include reflectance and electrochemical test strips. The reflectance assays include quantitative determination of glucose, total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides. The electrochemical assay is for glucose measurement. The reflectance test strips are available with one, two, and three assays and come in various combinations of assays. The included PTS Panels® Test Strips include: PTS Panels® Lipid Panel Test Strips, PTS Panels® Chol+HDL+Glu Test Strips, PTS Panels® Chol+HDL Test Strips, PTS Panels® Chol+Clu Test Strips, PTS Panels® eGlu Test Strips, PTS Panel® Glucose Test Strips, and CardioChek Plus Smart Bundle™ Pack. The PTS Panels® Test Strips are single-use.
PTS Panels® Controls contain chemicals that react with test strips to produce color or an electrical current depending on the test strips used. Controls should be run to verify the test system performance, when results are questionable, to comply with a facility's quality control requirements, or as required by local accrediting and regulatory bodies. They contain two vials of control solution that have two levels of each analyte. Controls included in this kit are the PTS Panels® HDL Cholesterol Controls and PTS Panels® Multi-Chemistry Controls.
The PTS Collect™ Capillary Tubes are a glass and plastic tube capable of dispensing an exact volume of blood from the capillary tube to a test strip. The tubes are available with volumes of 15, 20, 30, and 40 uL.
The A1CNow + test provides quantitative measurement of glycated hemoglobin (%A1C) levels in capillary (fingerstick) or venous whole blood samples. A1CNow®+ consists of three components: a semi-disposable plastic-encased device (the analyzer), a plastic cartridge enclosing dry reagent strips, and a shaker kit. The units of the analyzer are pre-programmed on the analyzer and are available in either percentage of glycated hemoglobin (%A1C) or millimoles per mole (mmol/mol). The analyzers are lot specific and intended to be discarded once the provided cartridges are consumed.
The Unistik® 3 Pre-set Single Use Safety Lancets are sterile, single-use, auto-disabling lancets to perform finger sticks to obtain capillary blood.
The PTS Connect™ Dock and PTS Connect® ProLink are included to transfer data from the CardioChek® Plus Analyzer, CardioChek® PA Analyzer, and the A1CNow®+ to computer systems. The CardioChek® Printer interfaces with the CardioChek® Plus and CardioChek® PA Analyzer to print results onto paper and self-adhesive labels.
The PTS Professional Chemistry Kit contents are customizable according to user needs. Individual kits may not contain all the devices described in this submission.
The provided text describes the 510(k) premarket notification for the "PTS Professional Chemistry Kit." However, it does not contain a study demonstrating the device meets specific acceptance criteria in the format requested.
Instead, the document states that the "PTS Professional Chemistry Kit" is a convenience kit combining previously cleared devices. Therefore, the performance criteria and studies for each individual component refer to their original 510(k) clearances.
The available information is as follows:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a specific table of acceptance criteria for the kit itself. Instead, it refers to the performance established for its individual components in their respective prior 510(k) clearances. The "Similarities" table indicates the analytes measured and the intended use are the same as the predicate devices.
| Analyte | Acceptance Criteria (from previous clearances) | Reported Device Performance (from previous clearances) |
|---|---|---|
| Glycated Hemoglobin (%A1C) | Performance established in K090413 | Precision, linearity, limit of detection, analytical specificity, and method comparison established. |
| Total Cholesterol | Performance established in K140068, K151545, K162282, K071507, K041750, K014099 | Precision, linearity, limit of detection, analytical specificity, and method comparison established. |
| High Density Lipoprotein (HDL) | Performance established in K140068, K151545, K162282, K071507, K041750, K014099 | Precision, linearity, limit of detection, analytical specificity, and method comparison established. |
| Triglycerides | Performance established in K140068, K151545, K162282, K071507, K041750, K014099 | Precision, linearity, limit of detection, analytical specificity, and method comparison established. |
| Glucose | Performance established in K140068, K151545, K162282, K071507, K041750, K014099 | Precision, linearity, limit of detection, analytical specificity, and method comparison established. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the given document for the "PTS Professional Chemistry Kit." It would be found within the detailed clinical and non-clinical testing reports for the individual predicate devices (K140068, K090413, K151545, K162282, K071507, K041750, K014099).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the given document. It is relevant for image-based diagnostic AI, not typically for chemical test kits like this. The ground truth for chemical analytes is established through laboratory reference methods.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable and not provided in the given document. Adjudication methods are typically used in clinical studies involving interpretation of results, often for imaging or subjective assessments, not for quantitative chemical measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable and not provided in the given document. MRMC studies are relevant for human-in-the-loop AI systems that assist human readers in interpreting complex data (e.g., radiology images). This device is a quantitative chemical test kit.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device measures chemical analytes directly. The performance described (precision, linearity, limit of detection, analytical specificity, method comparison) for the predicate devices would inherently be "standalone" performance for the instrument measuring the analytes. The "PTS Professional Chemistry Kit" itself is a collection of these standalone devices.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For chemical measurement devices, the ground truth is typically established using reference laboratory methods (e.g., standardized assays on high-precision lab equipment). The document mentions "method comparison" studies, implying comparison against such reference methods.
8. The sample size for the training set:
This information is not provided in the given document. For traditional chemical test kits, a "training set" in the machine learning sense is not typically used. Instead, methods are developed and validated through extensive analytical testing and calibration curves.
9. How the ground truth for the training set was established:
As mentioned, a "training set" as understood in AI/ML is not directly applicable here. For the development and validation of these chemical assays, the "ground truth" (i.e., true concentration values) would be established using reference laboratory methods for calibrating the devices and evaluating their accuracy against known standards.
Ask a specific question about this device
(147 days)
Morrisville, NC 27560
Re: K181373
Trade/Device Name: Extended Lipid Panel Assay Regulation Number: 21 CFR 862.1175
Names:
Triglyceride test system, 21 CFR 862.1705, Product Code CDT Cholesterol test system 21 CFR 862.1175
The Extended Lipid Panel Assay is an in vitro diagnostic test for quantitative determination of Total Cholesterol, High Density Lipoprotein Cholesterol, and Triglycerides in human serum and Apolipoprotein B in human serum. Values for Total Cholesterol, High Density Lipoprotein Cholesterol, Triglycerides and Apolipoprotein B are calculated by the Vantera® Clinical Analyzer.
· Total Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood, lipid and lipoprotein metabolism disorders.
· High Density Lipoprotein Cholesterol measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
· Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
· Apolipoprotein B measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis.
The Extended Lipid Panel Assay involves the acquisition of a 400 MHz proton NMR spectrum of serum or plasma, passing the spectral information through a Partial Least Squares (PLS) regression model, and deriving analyte concentrations from the spectrum based on the trained PLS model. The proton NMR spectrum of serum and plasma is replete with information from the lipids packaged in lipoproteins. The spectrum consists of multiple proton signals emanating from the TG, cholesteryl esters and free cholesterol present in chylomicrons, VLDL, LDL and HDL, out of which the methylene and methyl proton signals are the most abundant. NMR spectra were recorded for several hundred to several thousand representative serum specimens for which the TG, TC, HDL-C and ApoB were chemically measured. Using a PLS regression routine, the spectral information in the combined methylene and methyl region (0.56 - 1.40 ppm) was trained against the chemical measurements where the information is connected through latent variables. Cross-validation was performed with PRESS statistics to optimize the regression model with an appropriate number of latent variables. Once trained with sufficient number of specimens, for any test specimen spectrum, the spectral information is then converted into lipid or ApoB concentrations through the optimum number of 24 to 27 latent variables for which the regression coefficients were known from the predictor matrix.
The Extended Lipid Panel Assay is an in vitro diagnostic test for the quantitative determination of Total Cholesterol (TC), High Density Lipoprotein Cholesterol (HDL-C), Triglycerides (TG), and Apolipoprotein B (ApoB) in human serum and plasma. The device uses Nuclear Magnetic Resonance (NMR) technology to derive analyte concentrations.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are demonstrated through various analytical performance studies. The document does not explicitly state pre-defined acceptance criteria in a dedicated table with pass/fail results against specific thresholds for each performance metric, but rather presents the results of these studies. Based on the provided performance data, the implicit acceptance criteria would be for the device to show good analytical sensitivity, precision, linearity, and minimal interference, and to demonstrate substantial equivalence to predicate devices through method comparison studies.
Below is a table summarizing the reported device performance, which implicitly demonstrates it meets the necessary criteria for "substantial equivalence" based on the FDA's assessment in the 510(k) clearance.
| Performance Metric | Analyte | Reported Device Performance | Implicit Acceptance Standard (Inferred from predicates/regulatory guidance for clinical chemistry tests) |
|---|---|---|---|
| Analytical Sensitivity (LoQ) | TC | 24 mg/dL | Lowest concentration measurable with acceptable precision and accuracy (implicitly deemed acceptable by FDA) |
| TG | 15 mg/dL | Lowest concentration measurable with acceptable precision and accuracy | |
| HDL-C | 13 mg/dL | Lowest concentration measurable with acceptable precision and accuracy | |
| ApoB | 18 mg/dL | Lowest concentration measurable with acceptable precision and accuracy | |
| Within-run Precision (%CV) (n=20) | TC | 0.9 - 1.6% | Low %CV indicating high within-run precision (typically < 5% or < 10% depending on analyte/concentration) |
| TG | 1.0 - 1.0% | Low %CV indicating high within-run precision | |
| HDL-C | 1.3 - 2.4% | Low %CV indicating high within-run precision | |
| ApoB | 1.1 - 1.2% | Low %CV indicating high within-run precision | |
| Within-Laboratory Precision (%CV) (n=80) | TC | 1.1 - 1.6% | Low %CV indicating high within-laboratory precision (typically < 5% or < 10%) |
| TG | 1.0 - 1.4% | Low %CV indicating high within-laboratory precision | |
| HDL-C | 1.4 - 2.8% | Low %CV indicating high within-laboratory precision | |
| ApoB | 1.9 - 2.4% | Low %CV indicating high within-laboratory precision | |
| Reproducibility (%CV) | TC | 2.1 - 3.6% | Low %CV across multiple sites/runs (typically < 10%) |
| TG | 2.1 - 3.4% | Low %CV across multiple sites/runs | |
| HDL-C | 2.1 - 4.6% | Low %CV across multiple sites/runs | |
| ApoB | 3.7 - 6.7% | Low %CV across multiple sites/runs | |
| Linearity (R²) | TC | 1.000 | R² close to 1 indicates excellent linearity over the measuring range |
| TG | 1.000 | R² close to 1 indicates excellent linearity over the measuring range | |
| HDL-C | 1.000 | R² close to 1 indicates excellent linearity over the measuring range | |
| ApoB | 0.999 | R² close to 1 indicates excellent linearity over the measuring range | |
| Linearity (Measuring Range) | TC | 66-868 mg/dL | Broad medically relevant measuring range |
| TG | 35-950 mg/dL | Broad medically relevant measuring range | |
| HDL-C | 14-152 mg/dL | Broad medically relevant measuring range | |
| ApoB | 35-366 mg/dL | Broad medically relevant measuring range | |
| Interfering Substances | Endogenous/Exogenous | No significant interference observed for numerous agents at specified concentrations. | No clinically significant bias from common interferents. |
| Method Comparison (Correlation with Predicate) | TC | r = 0.994 | High correlation (r > 0.975 typically sought) and Deming regression slope/intercept close to 1 and 0, respectively. |
| TG | r = 0.9805 (implied from y= -3.858 + 0.9805x and visual) | High correlation | |
| HDL-C | r = 0.985 | High correlation | |
| ApoB | r = 0.980 | High correlation |
2. Sample sizes used for the test set and the data provenance
- Analytical Sensitivity (LoQ): Not specified how "LoQ" was determined by sample size, but indicates "lowest concentration measurable with acceptable precision and accuracy."
- Assay Precision (Within-run and Within-laboratory): 20 replicates for within-run and 80 replicates (for n=80, it means 40 total runs over 20 days with 2 replicates per run on 1 instrument) for within-laboratory precision, using three patient serum pools (low, medium, high). The provenance is "patient serum pools."
- Reproducibility: Five levels of serum panels were tested for 5 days, 6 runs per day, 2 replicates per run at 3 sites. This results in 5 (levels) * 5 (days) * 6 (runs/day) * 2 (replicates/run) * 3 (sites) = 900 measurements per analyte. The provenance is from "serum panels," presumably patient-derived.
- Linearity: "Reference serum pools were prepared from patient specimens with low to high values" and "mean values from analysis of four replicates of each pool" were used. The exact number of 'pools' created for each analyte is not explicitly stated but implies multiple points across the range. The provenance is "patient specimens."
- Interfering Substances: "samples with spiked concentrations of interferent" were used. The number of samples is not specified, but it covered "Eight endogenous agents and thirty drugs."
- Method Comparison:
- TC: n=281 pooled serum samples
- TG: n=270 pooled serum samples
- HDL-C: n=15575 (This number seems exceptionally high compared to other n values; it might be a typo or represent cumulative data points over various experiments) pooled serum samples
- ApoB: n=266 pooled serum samples
The data provenance for method comparison is "pooled serum samples across the reportable range" of the device. All studies appear to be retrospective using banked or prepared samples, as typical for analytical performance studies of this nature. The country of origin is not specified but is presumably the US given the FDA submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This device is a clinical chemistry assay for quantifying specific analytes (TC, TG, HDL-C, ApoB) rather than an imaging or diagnostic AI device that interprets qualitative information and requires human expert consensus for ground truth.
- Ground Truth for Method Comparison: The "ground truth" for the method comparison studies was established by predicate clinical chemistry devices (biochemical assays like enzymatic colorimetric assay or nephelometric immunoassay). These are established methods that serve as the comparative standard. No human experts were involved in establishing the ground truth for the values themselves, as it's a quantitative measurement.
- Qualifications of Experts: Not applicable in the context of establishing ground truth for quantitative chemical measurements. However, the development, validation, and regulatory submission would have been overseen by qualified scientists and regulatory affairs professionals.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for qualitative or semi-quantitative diagnostic devices where human interpretation might differ, and an expert panel resolves discrepancies to establish ground truth. For quantitative clinical chemistry assays, the reference method (predicate device in this case) provides the "adjudicated" ground truth.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. MRMC studies are designed for evaluating diagnostic devices that involve human interpretation of images or other qualitative data, often with AI assistance, to assess the impact of AI on reader performance. This device is a standalone clinical chemistry analyzer that provides quantitative measurements; it does not involve human "readers" or AI assistance in a diagnostic interpretation workflow.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, this device is inherently a standalone algorithm. The Extended Lipid Panel Assay is described as utilizing "passing the spectral information through a Partial Least Squares (PLS) regression model, and deriving analyte concentrations from the spectrum based on the trained PLS model." This process is entirely automated and does not involve human intervention in the result generation once the sample is run on the Vantera® Clinical Analyzer. The performance studies (precision, linearity, method comparison) directly evaluate this standalone algorithmic performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for evaluating the Extended Lipid Panel Assay was based on measurements obtained from legally marketed predicate clinical chemistry devices. These predicate devices are established methods for quantifying the specific lipid and lipoprotein analytes (TC, TG, HDL-C, ApoB).
8. The sample size for the training set
For the Partial Least Squares (PLS) regression model: "NMR spectra were recorded for several hundred to several thousand representative serum specimens for which the TG, TC, HDL-C and ApoB were chemically measured." The exact number is not precisely stated but indicates a substantial training dataset.
9. How the ground truth for the training set was established
The ground truth for the training set was established by chemical measurements (presumably using standard, often wet-chemistry or enzymatic reference methods) for TG, TC, HDL-C, and ApoB on the "several hundred to several thousand representative serum specimens." These chemical measurements served as the target values for the PLS regression model to learn the relationship between the NMR spectra and the analyte concentrations.
Ask a specific question about this device
(30 days)
92121
Re: K180504
Trade/Device Name: Mission Lipid Panel Monitoring System Regulation Number: 21 CFR 862.1175
The Mission® Lipid Panel Monitoring System is intended for the quantitative determination of Total Cholesterol. High Density Lipoprotein Cholesterol, and Triglycerides in human capillary whole blood from the fingerin venous whole blood. The Mission® Lipid Panel Monitoring System consisting of the Mission® Lipid Panel Meter, Mission® Lipid Panel Test Cartridges, Mission® Cholesterol Optical Verifier, and Mission® Cholesterol Control Solution, and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto disabling lancing devices. This system is for in vitro diagnostic use only.
Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. HDL (High Density Lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism or various endocrine disorders.
An estimated value for Low Density Lipoprotein Cholesterol is calculated by the Mission Cholesterol Pro Meter and is reported only when Triglycerides are ≤400 mg/dL.
Not Found
The document provided (K180504) is a 510(k) premarket notification for the Mission Lipid Panel Monitoring System. Based on the content, this device is a quantitative assay system for measuring Total Cholesterol, HDL-C, and Triglycerides, primarily used in professional healthcare settings. It is NOT an AI/ML powered device, nor does it involve image analysis by experts, or require MRMC studies.
Therefore, many of the requested criteria (like number of experts, adjudication methods, multi-reader multi-case studies, effect size of human improvement with AI, training set details, etc.) are not applicable to this type of medical device and the information is not present in the provided document.
However, I can extract the acceptance criteria and performance data relevant to a clinical chemistry device, which typically involves analytical accuracy and precision.
Here's an attempt to answer the request based only on the provided document, acknowledging the limitations for an AI/ML context:
The Mission Lipid Panel Monitoring System is a quantitative in vitro diagnostic device. The documentation provided focuses on clinical chemistry performance, not AI/ML algorithm performance. As such, many of the acceptance criteria and study details commonly associated with AI/ML devices (e.g., number of experts, MRMC studies, training set details, ground truth for image analysis) are not relevant or provided for this device type.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document (K180504) is a 510(k) clearance letter and an Indications for Use statement. It does not contain the detailed performance data or the specific acceptance criteria tables that would typically be part of the predicate comparison or detailed validation studies submitted to the FDA. The letter only states that the device has been reviewed and determined to be substantially equivalent to legally marketed predicate devices.
To illustrate what would typically be found for a quantitative diagnostic like this, and based on common regulatory expectations for lipid panels, performance would involve metrics like:
| Performance Metric | Acceptance Criteria (Illustrative - Not explicitly stated in the provided document, but typical for lipid panels) | Reported Device Performance (Illustrative - Not explicitly stated in the provided document) |
|---|---|---|
| Accuracy (Bias) | For each analyte (Total Cholesterol, HDL-C, Triglycerides), often assessed against a reference method (e.g., CDC-certified laboratory correlation). Criteria might be expressed as a percentage of bias or absolute bias within clinically acceptable limits for different concentration ranges. | e.g., Total Cholesterol: Bias within ±5% or ±X mg/dL compared to reference method. |
| Precision (Reproducibility) | Within-run, between-run, and total precision (CV% or SD) for each analyte at different concentration levels (low, medium, high). Criteria based on CLIA or professional guidelines. | e.g., HDL-C: CV% < 5% at clinically relevant concentrations. |
| Linearity/Reportable Range | The device should demonstrate linear response across the entire claimed measurement range for each analyte. | Demonstrated linearity across the stated measurement ranges for each analyte. |
| Interference | No significant interference from common endogenous substances (e.g., hemoglobin, bilirubin, triglycerides for cholesterol measurements) or exogenous substances (common medications). | Showed no significant interference from tested substances within specified limits. |
| Method Comparison | Strong correlation (e.g., R-squared > 0.95) and acceptable agreement (e.g., Bland-Altman plots) with a predicate device or established reference method. | Strong correlation and agreement with predicate device/reference method. |
| User Performance/Lay User Study (if applicable, though this is professional use) | For professional use, demonstrates acceptable performance across different users or sites. | Performance consistent across different operators in a professional setting. |
Important Note: The provided document does not contain any of these specific performance values or acceptance criteria. It is a regulatory clearance letter, not a summary of the validation study.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size for any test set or the data provenance (e.g., country of origin, retrospective/prospective). Such details would be found in the more detailed 510(k) submission or a summary of safety and effectiveness, neither of which is part of this provided extract.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. This device measures chemical analytes (Total Cholesterol, HDL-C, Triglycerides). The "ground truth" for such measurements is typically established by reference laboratory methods (e.g., enzymatic assays, gas chromatography-mass spectrometry (GC-MS), or high-performance liquid chromatography (HPLC)) traceable to international standards (e.g., CDC Lipid Standardization Program). It does not involve human expert interpretation of images or complex diagnostic reasoning.
4. Adjudication Method for the Test Set
Not applicable, as ground truth is established by objective laboratory methods, not expert consensus requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
Not applicable. MRMC studies are relevant for imaging devices where human readers interpret medical images. This device is a quantitative diagnostic instrument that measures chemical concentrations.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an analytical instrument; its "performance" is its ability to accurately and precisely measure analytes. While there's an "algorithm" for calculating LDL-C (Friedewald formula), its performance is assessed through the accuracy of the underlying direct measurements (TC, HDL-C, TG) and comparison of the calculated LDL-C to a reference method, not typically in a "standalone algorithm" context as understood for image analysis AI.
7. The Type of Ground Truth Used
The ground truth for a device like the Mission Lipid Panel Monitoring System would be established by:
- Reference Laboratory Methods: Highly accurate and precise laboratory methods (e.g., those traceable to the CDC Lipid Standardization Program or other certified reference methods) for Total Cholesterol, HDL-C, and Triglycerides.
- Certified Reference Materials: Use of materials with known, established concentrations of the analytes.
8. The Sample Size for the Training Set
Not applicable in the context of AI/ML training. This device is likely based on established electrochemical or photometric principles, not on a machine learning model "trained" on a dataset of patient samples in the AI sense. Performance is validated through analytical studies on patient samples and quality control materials.
9. How the Ground Truth for the Training Set was Established
Not applicable. As noted above, this is not an AI/ML device that requires a "training set" in that context. Its analytical method is based on established chemical principles. Validation is done against reference methods and materials, as described in point 7.