Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K143599, K142396

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chemtrue® Multi-Panel Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Phencyclidine, Ecstasy, Methadone, Oxycodone, Propoxyphene , Tramadol and Trivyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

    The multi test panels can consist of any analytes listed above in any combination. Only one cut-off concentration will be included per analyte per device.

    The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Antidepressants.

    The Chemtrue® Drug Screen Fentanyl / Tramadol Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    The Chemtrue® Drug Screen Fentany! / Tramadol Dip Card Test detects and is calibrated against norfentanyl, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

    The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    The test is not intended to differentiate between drugs of abuse and prescription use of Fentanyl/ Tramadol. The test is for in vitro diagnostic use only.

    The Chemtrue® Drug Screen Fentanyl / Tramadol Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Norfentanyl 5 and Tramadol 100 drugs in human urine. It is an in vitro diagnostic device. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    The Chemtrue® Drug Screen Fentany1 / Tramadol Cup Test detects and is calibrated against norfentany1, the major metabolite of fentanyl in human urine. The test is available in Single and multi-panels.

    The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to the drug test result, particularly when preliminary positive result is indicated.

    The test is not intended to differentiate between drugs of abuse and prescription use of Fentany/ Tramadol. The test is for in vitro diagnostic use only.

    The Chemtrue® Multi-Panel Drug Screen Dip Card Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentany, Marijuana, Methadone, Morphine, Opiates, Oxycodone, Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

    The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    The tests are not intended to differentiate between drugs of abuse and prescription use of Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Benzodiazepines, Antidepressants.

    The Chemtrue® Multi-Panel Drug Screen Cup Test is a rapid lateral flow immunoassay for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Ecstasy, Norfentanyl, Marijuana, Methamphetamine, Morphine, Opiates, Oxycodone, Phencyclidine, Propoxyphene, Tramadol and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    The multi test panels can consist of any drug analytes listed above in any combination. Only one cutoff concentration will be included per analyte per device.

    The test provides only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    The tests are not intended to differentiate between drugs of abuse and prescription use of Tricyclic Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Antidepressants.

    Device Description

    The Chemtrue® Drug Screen Tests are colloidal gold-based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in Dip Card or Cup formats, as indicated by the test name.

    AI/ML Overview

    The provided text describes the performance characteristics of the Chemtrue® Drug Screen Fentanyl/Tramadol Cup Test and Dip Card Test. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative immunoassay diagnostic devices typically involve high levels of agreement with a reference method, especially around the cutoff concentration. While explicit acceptance criteria ("Pass/Fail" thresholds) are not directly stated in the provided text, the reported performance demonstrates very high agreement.

    Note: The tables below focus on the Fentanyl (FYL) and Tramadol (TML) tests as these are the new additions being evaluated in this submission. Previous analytes were cleared under older 510(k)s. The percentages under "Reported Device Performance" are calculated from the provided raw counts.

    Method Comparison (Accuracy) Studies vs. LC/MS Reference Method

    The agreement is calculated as (Number of correct results / Total number of samples in that category) * 100%.

    Analyte (Cutoff)Category (LC/MS Conc.)Acceptance Criteria (Implicit: High agreement)Reported Device Performance (Dip Card)Reported Device Performance (Cup Test)
    Norfentanyl (5 ng/mL)Positive (Test Positive)100% (35/35)100% (35/35)
    Near Cutoff Positive (Cutoff to 150% C/O)100% (6/6)100% (6/6)
    Positive (>150% C/O)100% (26/26)100% (26/26)
    Negative (Test Negative)93.4% (57/61)95% (58/61)
    No drug present100% (22/22)100% (22/22)
    <50% of C/O96.7% (29/30)96.7% (29/30)
    50% C/O to Cutoff85.7% (6/7)87.5% (7/8)
    Tramadol (100 ng/mL)Positive (Test Positive)100% (31/31)100% (31/31)
    Near Cutoff Positive (Cutoff to 150% C/O)100% (8/8)100% (8/8)
    Positive (>150% C/O)100% (23/23)100% (23/23)
    Negative (Test Negative)100% (52/52)100% (52/52)
    No drug present100% (20/20)100% (20/20)
    <50% of C/O100% (20/20)100% (20/20)
    50% C/O to Cutoff100% (12/12)100% (12/12)

    OTC Lay-user Accuracy Studies vs. LC/MS Values

    Analyte (Cutoff)Category (LC/MS Conc.)Acceptance Criteria (Implicit: High agreement)Reported Device Performance (Dip Card)Reported Device Performance (Cup Test)
    Norfentanyl (5 ng/mL)Positive (Test Positive)100% (47/47)100% (45/45)
    Near Cutoff Positive (125% C/O)100% (24/24)100% (22/22)
    Positive (150% C/O)100% (23/23)100% (23/23)
    Negative (Test Negative)100% (139/139)100% (142/142)
    No Drug Present100% (93/93)100% (93/93)
    Negative (50% C/O)100% (23/23)100% (25/25)
    Near Cutoff Negative (75% C/O)100% (23/23)100% (24/24)
    Tramadol (100 ng/mL)Positive (Test Positive)100% (46/46)100% (47/47)
    Near Cutoff Positive (125% C/O)100% (23/23)100% (24/24)
    Positive (150% C/O)100% (23/23)100% (23/23)
    Negative (Test Negative)100% (141/141)100% (140/140)
    No Drug Present100% (93/93)100% (93/93)
    Negative (50% C/O)100% (24/24)100% (24/24)
    Near Cutoff Negative (75% C/O)100% (24/24)100% (23/23)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Method Comparison (Accuracy) Studies:

      • Sample Size: 176 clinical urine specimens (93 Fentanyl, 83 Tramadol).
      • Data Provenance: Clinical urine specimens, retrospective (since they were confirmed with LC/MS prior to testing with the device). The country of origin is not specified but implied to be within regular clinical testing networks.
    • OTC Lay-user Accuracy Studies:

      • Sample Size: 140 intended lay-users for each device format (Dip Card and Cup). Each lay-user received two blind-coded samples.
      • Data Provenance: Urine samples spiked with drugs at various concentrations (0, 50%, 75%, 125%, 150% of cutoff) into drug-free urine pools. This is a prospective study design for evaluating lay-user performance with controlled samples. The samples were confirmed with LC/MS/MS prior to the study. The study was conducted at three intended user sites (Shopping Mall, School, and Hotel), suggesting diverse user demographics.
    • Precision/Reproducibility Studies:

      • Sample Size: For each analyte (Norfentanyl and Tramadol), 5 replicates at 6 concentration levels (0%, -50%, -25%, Cutoff, +25%, +50% of the cutoff) were tested with 3 lots of each device format (Cup and Dip Card) by 3 operators over 5 consecutive days.
        • This totals to 5 replicates * 6 concentrations * 3 lots * 3 operators = 270 tests per analyte per device format.
        • Thus, for FYL Dip Card: 270 samples. FYL Cup Test: 270 samples. TML Dip Card: 270 samples. TML Cup Test: 270 samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Method Comparison (Accuracy) Studies: The ground truth was established by LC/MS (Liquid Chromatography/Mass Spectrometry), which is a highly accurate analytical chemistry technique considered the "gold standard" for drug concentration determination. The text states: "Each specimen was confirmed with LC/MS assay with the value assignment." and "The accuracy of the Chemtrue® Test devices were evaluated against ... clinical urine specimens. Each specimen was confirmed with LC/MS assay with the value assignment."

      • This is an instrument-based determination, not typically requiring human expert consensus in the same way as, for example, image interpretation. The "experts" here are the drug analysts who performed the LC/MS analysis and validated the results.
      • The study mentions "One drug analyst with one set of blind code in one device format - Dip Card / Cup were tested by one operator." This refers to the operators performing the rapid immunoassay tests, not the LC/MS ground truth determination.
    • OTC Lay-user Accuracy Studies: The ground truth was established by LC/MS/MS ("Concentrations By LC-MS/MS (ng/mL)") of the spiked urine samples. As above, this is an instrument-based method.

    • Precision/Reproducibility Studies: The ground truth for the spiked urine controls was defined by their known concentrations, which were "LC/MS confirmed concentrations."


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • For the Method Comparison (Accuracy) Studies and OTC Lay-user Accuracy Studies, no human adjudication method (like 2+1 or 3+1) is described for establishing the ground truth. The ground truth (LC/MS or LC/MS/MS results) is considered definitive. The rapid immunoassay device results are compared directly to this definitive ground truth.
      • For the actual reading of the rapid immunoassay devices, 4 operators performed the tests for the method comparison study, but their readings were individual comparisons to the LC/MS reference, not adjudicated against each other.
      • For the OTC study, "intended lay-users" performed the interpretation, and their results were compared to the LC/MS/MS values.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done.
    • This device is a rapid lateral flow immunoassay for drug detection, not an AI-powered diagnostic system requiring human interpretation or assistance from AI, so this type of study is not applicable. The immunoassay itself provides a visual result (line/no line) that lay users interpret.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This question is not applicable as the device is a rapid lateral flow immunoassay, which is a standalone point-of-care test, not an algorithm. The "performance" is the chemical reaction and visual read-out of the test strip/cup. The "human-in-the-loop" is the user interpreting the visual result. However, the studies evaluating its performance involved human users (operators for method comparison, lay-users for OTC study) interpreting the visual output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The primary ground truth used for all performance studies (accuracy, precision, OTC accuracy) was analytical confirmation by LC/MS or LC/MS/MS. This is a highly precise and quantitative laboratory method considered definitive for identifying and quantifying drug metabolites in urine.

    8. The Sample Size for the Training Set

    • No training set is described because this device is a rapid lateral flow immunoassay, not a machine learning or AI-based algorithm that requires a training set. The device operates based on chemical reactions and antigen-antibody binding.

    9. How the Ground Truth for the Training Set was Established

    • This question is not applicable as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153192

    Validate with FDA (Live)

    Date Cleared
    2016-04-18

    (167 days)

    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K142396, K143599

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chemitrue® Multi-Panel Drug Screen Cup Tests are rapid lateral flow immunossays for the qualitative detection of Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methamphetamine, Morphine, Phencyclidine, Ecstasy, Methadone, Propoxyphene and Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
    AmphetamineAMPd-Amphetamine300
    AmphetamineAMPd-Amphetamine500
    AmphetamineAMPd-Amphetamine1000
    BarbituratesBARSecobarbital/Pentobarbital200
    BarbituratesBARSecobarbital/Pentobarbital300
    BenzodiazepinesBZOOxazepam200
    BenzodiazepinesBZOOxazepam300
    BuprenorphineBUPBuprenorphine10
    CocaineCOCBenzoylecgonine150
    CocaineCOCBenzoylecgonine300
    EcstasyMDMAd,l-Methylenedioxymethamphetamine500
    MethamphetamineMAMPd-Methamphetamine300
    MethamphetamineMAMPd-Methamphetamine500
    MethamphetamineMAMPd-Methamphetamine1000
    MarijuanaTHC11-nor-Δ9-THC-9-COOH50
    MethadoneMTDMethadone300
    OpiatesOPIMorphine2000
    OxycodoneOXYOxycodone100
    PhencyclidinePCPPhencyclidine25
    PropoxyphenePPXPropoxyphene300
    TricyclicAntidepressantsTCANortriptyline1000

    The multi test panels can consist of up to fourteen (14) of the above issed analytes in any combination. Only one cutoff concentration will be included per analyte per device. The tests are intended for prescription and Over-The-Counter (OTC) use.

    The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GCMS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodizepines, Barbiturates, Buprenorphine, Oxycodone, Propoxyphene and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

    Device Description

    The Chemtrue® Drug Screen Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in Dip Card or Cup formats, as indicated by the test name.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Chemtrue Multi-Panel Drug Screen Dip Card/Cup Tests

    This document describes the acceptance criteria and the study that demonstrates the Chemtrue Multi-Panel Drug Screen Dip Card/Cup Tests meet these criteria. The device is a rapid lateral flow immunoassay for the qualitative detection of various drugs of abuse in human urine, intended for prescription and Over-The-Counter (OTC) use.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device are implicitly demonstrated through the performance characteristics presented in the 510(k) summary. These include:

    • Precision (Reproducibility): Consistent results at and around the cut-off concentrations across different operators and lots. The exact acceptance percentage for concordance at cut-off is not explicitly stated as a numerical criterion but is demonstrated by the reported distribution of positive/negative results.
    • Specificity (Cross-Reactivity): Detect the target analytes and related compounds at specified concentrations, while exhibiting minimal cross-reactivity with non-target substances. Acceptance is qualitative based on the reported cross-reactivity percentages.
    • Interference: No interference from common endogenous compounds or non-structurally related compounds. Acceptance is qualitative, showing "no interference" for a wide range of substances at tested concentrations.
    • Effect of Urine pH and Specific Gravity: Test performance should not be affected by variations in urine pH and specific gravity within specified ranges. Acceptance is demonstrated by stating these factors "do not affect the test performance."
    • Stability: Maintain performance over a specified shelf-life. Acceptance is a 2-year shelf life at 2°C to 30°C.
    • Accuracy (Method Comparison): High agreement with a confirmed reference method (GC/MS). The acceptance criterion for accuracy is ≥ 99% agreement between the Chemtrue® Drug Screen test device and GC/MS values.
    • Lay-user Accuracy and Usability (for OTC): High accuracy when performed by lay-users and ease of understanding instructions. The acceptance criterion is demonstrated by ≥ 99% agreement with GC/MS values for lay-user studies and ≥ 96% of lay users easily following instructions.

    Below is a summary of the key performance criteria and the reported device performance. Note that for precision and cross-reactivity, specific numerical acceptance thresholds were not explicitly stated as distinct criteria, but the reported performance data in the tables indicate successful demonstration of these characteristics.

    Table 1: Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criterion (Implicit where not numerical)Reported Device Performance
    Precision (Reproducibility)Consistent qualitative results (+/-) at negative, 50%, 75%, cut-off, 125%, and 150% of cut-off across operators and lots.At Cutoff (e.g., AMP 300 ng/mL Dip Card): 16 positive, 14 negative out of 30. At 125% & 150% of Cutoff: 30 positive out of 30. At 50% & 75% of Cutoff & Negative: 0 positive, 30 negative out of 30. Similar distributions for other analytes and formats, indicating consistent qualitative detection around the cutoff.
    Specificity (Cross-Reactivity)Detect target analytes and related compounds; minimal cross-reactivity with non-target substances.Reported cross-reactivity percentages for various related compounds (e.g., d-Amphetamine 100% at 500 ng/mL, d,l-Amphetamine 62.5% at 800 ng/mL for AMP 500). Non-structurally related compounds found "not to cross-react" at 100 µg/mL.
    InterferenceNo interference from 103 specified potential interferents (endogenous and non-structurally related compounds) at tested concentrations."It was found not to cross-react when tested at concentrations of 100 µg/mL at ±25% of the drug cut-off concentrations." Listed 103 compounds explicitly stated as "do not interfere."
    Effect of Urine pHTest performance unaffected within pH range of 2.0 to 9.0 at ±25% of the drug cut-off concentrations."The testing results demonstrate that the urine pH ranges from 2.0 to 9.0 at ±25% of the drug cut-off concentrations do not affect the test performance."
    Effect of Specific GravityTest performance unaffected within SG ranges of 1.001, 1.015, 1.020, 1.025, and 1.030 at ±25% of the drug cut-off concentrations."The specific gravity (SG) ranges of 1.001, 1.015, 1.020, 1.025 and 1.030 at ±25% of the drug cut-off concentrations do not affect the test results."
    Stability (Shelf Life)Support a 2-year shelf-life at 2℃ to 30℃."The stability study results support two (2) years shelf-life of the products at (2℃ to 30℃). The real time stability study is still on going."
    Accuracy (Method Comparison)≥ 99% agreement with GC/MS reference method.Reported: ≥ 97.6% to 100% agreement with GC/MS for individual analytes in Method Comparison Study (Tables 6a & 6b). "The results demonstrate that the agreement between the Chemtrue® Drug Screen test device and GC/MS values is ≥ 99%." (This combines all analytes for both Dip Card and Cup).
    OTC Lay-user Accuracy≥ 99% agreement with GC/MS reference method when performed by lay-users.Reported: 99% to 100% agreement with GC/MS for individual analytes in OTC Accuracy studies (Tables 7a & 7b). "The results demonstrate that the agreement between the Chemtrue® Drug Screen test device and GC/MS values is ≥ 99%." (This combines all analytes for both Dip Card and Cup).
    OTC Lay-user Usability≥ 96% of lay users can easily follow instructions to perform the test and interpret results, with a Flesch-Kincaid reading level of 7th grade."The results demonstrate that ≥ 96% of the lay users can easily follow the instructions to perform the test and interpret the results. A Flesch-Kincaid reading analysis supports a 7th grade reading level."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly delineate a "test set" in the context of the training/testing split common in AI/ML studies, as this device is a lateral flow immunoassay. Instead, it refers to various validation studies.

    • Precision (Reproducibility) Studies:
      • Sample Size: 30 samples per concentration level (Negative, 50% of cutoff, 75% of cutoff, Cutoff, 125% of cutoff, 150% of cutoff) for each analyte and device format. This equates to 180 samples per analyte/format combination (e.g., AMP Dip Card Test: Cutoff: 300 ng/mL used 180 samples).
      • Data Provenance: Not specified, but the samples were "GC/MS confirmed drug spiked urine controls." This implies controlled laboratory generation rather than real clinical samples. No country of origin is mentioned. These are prospective, controlled experiments due to the spiked nature.
    • Specificity Studies:
      • Sample Size: Not explicitly stated as a number of "samples." Various related compounds were tested for cross-reactivity at different concentrations, and 103 potential interferents were tested at a concentration of 100 µg/mL at ±25% of the drug cut-off concentrations.
      • Data Provenance: Not specified, likely laboratory-generated samples with controlled spiking of compounds. Prospective.
    • Interference Studies:
      • Sample Size: 103 potential interferents tested. Each was tested with "one lot each of the test device format." The implied sample count would be (number of interferents) * (number of device formats).
      • Data Provenance: Laboratory-generated, spiked samples. Prospective.
    • Effect of Urine pH and Specific Gravity Studies:
      • Sample Size: Not explicitly stated, but tests were performed within specified pH and SG ranges at ±25% of the drug cut-off concentrations.
      • Data Provenance: Laboratory-generated, controlled urine samples. Prospective.
    • Accuracy (Method Comparison) Studies:
      • Sample Size: "On average of 85 clinical specimens for each drug test and a total of 685 samples were tested" for the analytes specifically included in this submission (AMP300/500, BAR200/BZO200, COC150, MET300/500 and PPX).
      • Data Provenance: "blind-labeled clinical specimen correlation study." This indicates real human urine samples, but the country of origin is not specified. These are retrospective analyses as they compare the device's results to pre-existing or independently confirmed GC/MS values.
    • OTC Lay-user Accuracy Studies:
      • Sample Size: "One hundred (130) intended lay-users participated in the evaluation for each of the device format (Dip Card and Cup)." The number of tests performed is much higher, as each lay-user was given "up to two (2) random blind labeled samples" with concentrations covering negative, 50%, 75%, 125%, 150%, and 200% of the cutoff.
      • Data Provenance: "GC/MS confirmed urine samples," which were "spiking drugs into drug-free urine pool." This suggests laboratory-controlled, spiked samples rather than true "clinical" samples from drug users. Prospective experiment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • For Precision, Specificity, Interference, pH/SG, OTC Lay-user Studies: The ground truth was established by GC/MS confirmation of drug-spiked urine samples. This method is considered the gold standard for drug detection and quantification in urine. Therefore, the "experts" are the technicians/analysts operating the GC/MS equipment, whose qualifications are implicitly high-level laboratory personnel but not explicitly stated (e.g., clinical chemists, toxicologists).
    • For Accuracy (Method Comparison) Studies: The ground truth was also established by GC/MS values. These were "blind-labeled clinical specimen" samples. Again, the experts are the GC/MS laboratory personnel. The document states, "Three operators performed the testing," but this refers to the Chemtrue device operators, not necessarily the GC/MS operators who established ground truth.

    4. Adjudication Method for the Test Set

    • For Precision Studies: The ground truth for the spiked samples was known. Qualitative results (+/-) from the device were compared against the expected positive/negative based on the known concentration relative to the cutoff. Discordant results are implicitly handled by the quantitative reporting in the tables (e.g., at cutoff, some were positive, some negative, reflecting the expected variability around the cutoff).
    • For Accuracy (Method Comparison) Studies: The ground truth was the GC/MS value. This method serves as the definitive reference, so no adjudication among multiple experts for ground truth was performed; it was a direct comparison to the GC/MS result. The results describe "discordant results" where the device's reading didn't match the GC/MS, and these were "confirmed at the drug cutoff level with the GC/MS concentrations," which is a further verification against the gold standard.
    • For OTC Lay-user Accuracy Studies: Similar to the accuracy studies, the ground truth was the GC/MS values from the spiked samples.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. MRMC studies are typically used to assess the improvement in human reader performance (e.g., radiologists interpreting images) with and without AI assistance. The Chemtrue device is a rapid diagnostic test (lateral flow immunoassay) that provides a direct qualitative result, not an AI system designed to assist human interpretation of complex data. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the device essentially operates in a standalone manner. The Chemtrue Multi-Panel Drug Screen Dip Card/Cup Tests are standalone devices that generate a qualitative (positive/negative) result without human intervention in the result generation process itself. Humans are "in the loop" for running the test and interpreting the visual lines, but the device's chemical reactions provide the "algorithm's" output.

    The validation studies (precision, specificity, interference, pH/SG, and the majority of the accuracy studies) demonstrate this standalone performance. The "OTC Lay-user Accuracy Studies" did involve human interpretation of the visual lines by lay-users, evaluating their ability to correctly use and interpret the standalone device.


    7. The Type of Ground Truth Used

    The primary ground truth used throughout the studies is Gas Chromatography / Mass Spectrometry (GC/MS). For all accuracy studies, spiked samples and clinical specimens were confirmed by GC/MS. This is explicitly stated: "GC/MS or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods." Given the context of drug screening devices, GC/MS is widely considered the definitive reference method for identifying and quantifying drugs and their metabolites in biological matrices.


    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable to traditional lateral flow immunoassay devices like the Chemtrue Multi-Panel Drug Screen tests. These are not machine learning or AI models that require training data to develop their "algorithm." Their mechanism is based on specific antigen-antibody reactions. Therefore, there is no "training set" in the computational sense. The device's components (antibodies, drug-protein conjugates) are developed and optimized through laboratory research and development, but this is a different process than "training" a dataset.


    9. How the Ground Truth for the Training Set was Established

    As explained in point 8, there is no "training set" in the context of this device. The development of the assay relies on established biochemical principles and extensive R&D to select and optimize antibodies and reagents. The "ground truth" for the development and optimization of the assay components would involve controlled laboratory experiments using known concentrations of analytes and interferents, confirmed by reference methods like GC/MS, to ensure the assay's sensitivity and specificity are appropriate to achieve the desired cut-offs. This is part of the extensive biochemical and quality control processes inherent in developing such diagnostic tests.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1