510(k) Data Aggregation

    K Number
    K161739

    Validate with FDA (Live)

    Date Cleared
    2016-12-19

    (178 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120028, K962804, K991134

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses:
      Spherical and Aspherical
      Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Toric
      Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
      Multifocal
      Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.
    2. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses:
      Spherical and Aspherical
      Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Toric
      Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
      Multifocal
      Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
      The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
    Device Description

    Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are available as spherical, aspherical, toric and multifocal lenses. The model illuminated with high water (58 %). These hydrogel lens materials are random copolymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which were crosslinked with glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane ethylene e trimethacrylate (TMPTMA) via photo-polymerization. These lenses are tinted blue using C.I. reactive blue 19 to make them more visible for handling. These lenses contain UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of theses lenses are less than 5% in the UVB range of 280-315 nm and less than 50% in the UVA range of 316-380nm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution with Tween 80, Sodium Hyaluronate and Polyethylene Glycol.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses.

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a full, detailed study report with all raw data. Therefore, some information, especially regarding detailed sample sizes for specific tests (like each biocompatibility test), expert qualifications, and adjudication methods, is not explicitly provided in the summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Predicate)Reported Device Performance (Aquamax Etafilcon A)
    Technological Characteristics
    Production MethodCast-Molded (K120028, K962804, K991134)Cast-Molded
    USAN NameEtafilcon A (K120028, K962804, K991134)Etafilcon A
    Material ClassificationGroup 4 (High Water Ionic) (K120028, K962804, K991134)Group 4 (High Water Ionic)
    Water Content (%)58% (K120028, K962804, K991134)58%
    Refractive Index1.402 (K120028), 1.40 (K962804, K991134)1.402
    Oxygen Permeability (Dk) (x 10⁻¹¹ (cm²/sec)(ml O₂/ml-mmHg) @ 35°C) (edge corrected)19.73 (K120028), 26 (K962804), 26.3 (K991134)19.73
    Percent Transmittance (% T) at 593nm> 95% (K120028), > 85% (K962804, K991134)> 95%
    Percent Transmittance (% T) at 380-315nm (UVA)< 50% (K120028), < 30% (K962804, K991134)< 50%
    Percent Transmittance (% T) at 315-280nm (UVB)< 5% (K120028, K962804, K991134)< 5%
    Lens DesignSpherical, Aspherical (K120028); Spherical, Toric, Multifocal, Toric Multifocal (K962804); Spherical, Bifocal, Toric, Toric Bifocal (K991134)Spherical, Aspherical, Toric, Multifocal
    Packaging SolutionBorate buffered saline (K120028), N/A (K962804, K991134)Borate buffered saline (with Tween 80, Sodium hyaluronate and Polyethylene Glycol)
    BiocompatibilityNegative responses for standard testsNegative responses for all tests
    MicrobiologyMinimum SAL of 10⁻⁵Steam sterilization validated to deliver minimum SAL of 10⁻⁵
    Shelf-life SterilityStereo through expiration date claimedLenses remained sterile for 5 years (accelerated conditions)
    Bacteriostatic ValidationMicroorganisms killed under tested conditions vs. controlSteam sterilizer effective, tested microorganisms killed
    LeachabilityNo leachable monomers and additive residues at detection levelsNo leachable monomers and additive residues at detection levels
    pH and Osmolality of Packaging SolutionWithin normal range for commercial soft lens (Consistent with K120028)Within normal range, consistent with K120028
    Manufacturing Verification (for Toric & Spherical Multifocal)Meet prescribed specifications with established tolerances for diameter, power, and base curve (per 1994 FDA Guidance)Verification studies conducted and ensured lenses met specifications/tolerances

    2. Sample Sizes Used for Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated for most tests in quantitative terms (e.g., number of lenses, number of animals). The document indicates "All tests were conducted" and "Studies were conducted to determine leachable materials," implying standard testing protocols were followed, which would inherently involve specific sample sizes per test type (e.g., for cytotoxicity, sensitization, ocular irritation, material property measurements) but these numbers are not detailed in this summary.
    • Data Provenance: The studies were conducted by Pegavision Corporation (Taiwan). The report implies that these were internal studies to support the 510(k) submission. Therefore, it is retrospective in the sense that the data was generated to support the application, but the tests themselves would have been conducted prospectively on the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • This information is not provided in the summary. For non-clinical tests (biocompatibility, material characteristics, microbiology), "ground truth" is typically established by laboratory standards and validated methods rather than expert consensus on individual observations. For manufacturing verification, engineers and quality control personnel would assess against specifications.

    4. Adjudication Method for the Test Set

    • This information is not applicable in the context of these non-clinical, objective tests. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where subjective assessments (e.g., image interpretation, patient-reported outcomes) from multiple readers or clinicians require reconciliation for ground truth establishment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is a physical medical device (contact lenses), not an AI/software device that assists human readers. Therefore, this type of study is completely irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone performance study was not done.
    • As established above, this is a physical medical device, not an algorithm or AI.

    7. The Type of Ground Truth Used

    • For the non-clinical tests described:
      • Material properties: Measured values compared against established standards and values of predicate devices.
      • Biocompatibility: Negative responses (absence of toxicity, sensitization, irritation) based on standardized in vitro and in vivo models.
      • Microbiology/Sterility: Achievement of a specified Sterility Assurance Level (SAL) confirmed by validation studies using microbiological indicators and established protocols.
      • Leachability: Absence of detectable levels of specific substances using analytical chemistry methods.
      • Manufacturing Verification: Conformance to engineering specifications and tolerances for physical dimensions (diameter, power, base curve).

    8. The Sample Size for the Training Set

    • Not applicable. This device is a manufactured product, not an AI model requiring a training set. The term "training set" is typically associated with machine learning and AI algorithms.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8.

    Summary of the Study per the Document:

    The study presented in this 510(k) summary is a series of non-clinical, laboratory-based tests designed to demonstrate the safety and effectiveness of the Aquamax (Etafilcon A) contact lenses and their substantial equivalence to legally marketed predicate devices. The "study" encompasses:

    • Technological Characteristics Comparison: Direct comparison of material properties and design specifications with three predicate devices (K120028, K962804, K991134). The current device largely matches the most relevant predicate (K120028) for key properties like oxygen permeability and percentage transmission, and is in acceptable ranges for others.
    • Biocompatibility Testing: According to standard cytotoxicity, maximization sensitization, and ocular irritation tests.
    • Microbiology and Sterilization Validation: Validation of the steam sterilization process to ensure a minimum SAL of 10⁻⁵ and demonstration of sterility throughout the shelf life.
    • Bacteriostatic Validation: Testing of the steam sterilizer's effectiveness in killing microorganisms.
    • Leachability Studies: To confirm the absence of leachable monomers and additive residues.
    • Manufacturing Verification Studies: For the new toric and multifocal lens designs, ensuring they meet specified tolerances for key parameters like diameter, power, and base curve.

    The overarching "study" proves adherence to the "Premarket Notification 510(K) Guidance Document for Class IV Contact Lenses" (May 1994) and establishes substantial equivalence based on these non-clinical tests. Clinical studies were deemed "not required" due to the well-documented safety and effectiveness of etafilcon A lenses and the similarity to predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150385

    Validate with FDA (Live)

    Date Cleared
    2015-04-30

    (71 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K962804, K991134, K120028

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

    NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The lenses are intended for single-use disposable wear.

    Device Description

    The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.

    AI/ML Overview

    The provided text is a 510(k) summary for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. This document does not describe a study that uses AI or machine learning to determine performance against acceptance criteria. Instead, it describes non-clinical tests to demonstrate the safety and effectiveness of the contact lenses.

    Therefore, many of the requested elements (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable as they relate to AI/ML device evaluations.

    However, I can extract the acceptance criteria and reported device performance from the non-clinical studies described in the document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are based on demonstrating substantial equivalence to a predicate device (K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses) and compliance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses. Where specific numerical acceptance criteria are not explicitly stated, the equivalence to the predicate device serves as the implicit acceptance criterion.

    CharacteristicAcceptance Criteria (Implicitly: Equivalent to Predicate K140025 or regulatory standards)Reported Device Performance (NaturalVue (etafilcon A))Meets Acceptance Criteria?
    Material Properties
    Production MethodCast-Molded (per predicate)Cast-MoldedYes
    USAN Nameetafilcon A (per predicate)etafilcon AYes
    Material ClassificationGroup 4 high water ionic (per predicate)Group 4 high water ionicYes
    Water Content (%)58% (per predicate)58%Yes
    Refractive Index1.4023 (per predicate)1.4023Yes
    Oxygen Permeability19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°C (per predicate)19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°CYes
    % Transmittance at 593nm> 95% (per predicate)> 95%Yes
    % Transmittance at 380-315nm< 30% (per predicate)< 30%Yes
    % Transmittance at 315-280nm< 5% (per predicate)< 5%Yes
    BiocompatibilityNegative responses for cytotoxicity, maximization, and ocular irritation (per predicate). Validity of blister packaging (passing standard extraction tests).Negative responses for all tests. Blister packaging validity demonstrated.Yes
    MicrobiologyMinimum SAL of 10-9 for steam sterilization. Sterility maintained through expiration.Validated to deliver SAL of 10-9. Shelf-life stability supports sterility.Yes
    Bacteriostatic ValidationUniformity of temperature in sterilizer; Tested microorganisms killed.Demonstrated uniformity of temperature; Tested microorganisms killed.Yes
    Shelf-LifeLenses remain sterile and packages remain tight for claimed expiration.Lenses remained sterilized with no microbial growth for 4 years (accelerated conditions). Seal remained tight for 4 years (accelerated conditions).Yes
    LeachabilityNo leachable monomers and additive residues at detection levels.No leachable monomers and additive residues at detection levels.Yes

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of an AI/ML device. For non-clinical material property testing, the sample sizes are not explicitly stated. The biocompatibility, microbiology, and leachability studies were conducted on the predicate device (K140025) and applied to the current device due to substantial equivalence, so the specifics of their sample sizes are not in this document.

    Data Provenance: The document refers to the predicate device (K140025), which was submitted by Pegavision Corporation, Taiwan. The current device is from Visioneering Technologies, Inc., and the non-clinical tests were performed to demonstrate equivalence. It's a non-clinical, retrospective comparison to previously established data and standards.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML device, and no "ground truth" established by experts for a test set is mentioned.


    4. Adjudication method for the test set

    Not applicable. No test set requiring expert adjudication is described for this type of device and study.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    Not applicable. This is not an AI/ML device, and thus, no MRMC study was conducted.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device or an algorithm.


    7. The type of ground truth used

    Not applicable in the context of AI/ML. For the non-clinical tests, "ground truth" would be the established scientific and regulatory standards for contact lens material properties, biocompatibility, sterility, and leachability. For example:

    • Material Properties: Direct measurements using laboratory techniques.
    • Biocompatibility: Standardized in vitro and in vivo (animal, if applicable, though not specified here) tests.
    • Microbiology: Validation of sterilization processes and sterility testing.
    • Leachability: Analytical chemistry techniques.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."


    9. How the ground truth for the training set was established

    Not applicable. No training set is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140025

    Validate with FDA (Live)

    Date Cleared
    2014-10-15

    (282 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K962804, K991134

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
      Spherical and Aspheric: Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Toric: Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
      Multifocal: Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Multifocal Toric: Aquamax (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.
      Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.
    2. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement (Replacement Recommended Every 2 Weeks):
      Spherical and Aspheric: Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Toric: Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
      Multifocal: Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Multifocal Toric: Aquamax (Etafilcon A) Multifocal Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.
      Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. Pegavision recommends up to two week replacement for Etafilcon A lenses, or replacement as recommended by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
    3. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
      Spherical and Aspheric: NaturalVue (Etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Toric: NaturalVue (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
      Multifocal: NaturalVue (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
      Multifocal Toric: NaturalVue (Etafilcon A)Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.
      NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.
    Device Description

    The Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and Natural/ue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are clear or visibility-tinted, containing a UV blocker and are available in a spherical, aspherical multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from Etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% Etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380nm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.

    AI/ML Overview

    This document is a 510(k) premarket notification for three types of soft contact lenses: Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the acceptance criteria and the study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate devices based on material properties, biocompatibility, microbiology, bacteriostatic validation, and leachability. There are no explicit quantitative acceptance criteria stated beyond "complying with the requirement of FDA Group 4" for sterilization. Differences in oxygen permeability and percent transmittance are evaluated against the predicate devices.

    Criteria/PropertyAcceptance Criteria (Implied by Predicate)Proposed Device (Aquamax/NaturalVue Etafilcon A) Reported Performance
    Material Properties
    Production MethodCast-Molded (for K120028 & K962804 predicates)Cast-Molded
    USAN NameEtafilcon A (for K120028 & K962804 predicates)Etafilcon A
    Material ClassificationGroup 4 high water ionic (for K120028 & K962804 predicates)Group 4 high water ionic
    Water Content (%)58% (for K120028 & K962804 predicates)58%
    Refractive Index1.4023 (for K120028 predicate)1.4023
    Oxygen Permeability (Dk) @ 35°CApprox. $19.73 \times 10^{-11}$ (K120028) to $21.4 \times 10^{-11}$ (K962804)$19.73 \times 10^{-11}$ (cm²/sec)(ml O₂/ml-mmHg)
    Percent Transmittance % T at 593nm> 95% (for K120028 & K013445 predicates), > 85% (for K962804 predicate)> 95%
    % T at 380-315nm (UVA)< 30% (for K120028 & K962804 predicates)< 30%
    % T at 315-280nm (UVB)< 5% (for K120028 & K962804 predicates)< 5%
    BiocompatibilityNegative responses for cytotoxicity, maximization, and ocular irritationNegative responses recorded for all tests
    Microbiology (Sterilization)Minimum SAL of 10-6 (FDA Group 4 requirement)Validated to deliver a minimum SAL of 10-6
    Bacteriostatic ValidationEffective killing of microorganisms, sterility maintained (5 years)Steam sterilizer effective, no microbial growth for 5 years
    LeachabilityNo leachable monomers and additive residues at detection levelsNo leachable monomers and additive residues at detection levels

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a clinical study with a specific sample size. The evaluation relies heavily on non-clinical tests comparing the device's properties to legally marketed predicate devices.

    • Data Provenance: The document explicitly states that "Aquamax (Etaflicon A) Soft (Hydrophilic) Contact Lenses from the previous FDA cleared K120028 did not require clinical studies as the USAN name and manufacturing processes are the same as the above-mentioned ACUVUE predicate (and reference) devices." This indicates a reliance on the historical performance and existing regulatory clearances of similar devices.
    • Non-Clinical Tests: These tests would involve laboratory samples of the contact lenses. The specific number of contact lenses or batches tested for each non-clinical criterion (e.g., water content, Dk, transmittance, biocompatibility) is not provided. The provenance of these laboratory samples (e.g., country of manufacture) is implicitly Taiwan, where Pegavision Corporation is located. These would be retrospective in the sense that they are conducted on manufactured devices for regulatory submission, rather than a prospective clinical trial on human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as there was no "test set" in the context of clinical data requiring expert review for ground truth. The evaluation is based on non-clinical laboratory testing and equivalence to predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This submission is for a medical device (contact lenses), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a contact lens and not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance is established by:

    • Laboratory measurements/standards: For material properties (water content, Dk, transmittance, refractive index), biocompatibility (standard cytotoxicity, maximization, ocular irritation tests), microbiology (sterilization SAL), bacteriostatic validation, and leachability. These are objective, quantifiable measurements against established scientific and regulatory standards.
    • Demonstrated equivalence to predicate devices: The "ground truth" for the overall safety and effectiveness is largely based on the fact that the proposed devices are substantially equivalent to previously cleared devices (K120028, K013445, K962804) which have established safety and effectiveness profiles through prior regulatory review and historical clinical use.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is not an AI/machine learning model.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K994324

    Validate with FDA (Live)

    Date Cleared
    2000-02-23

    (63 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K991134

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

    The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

    The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

    The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.

    ACUVUE Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The device description does not change from that cleared under K991134.

    AI/ML Overview

    The provided 510(k) submission (K994324) for the ACUVUE (etafilcon A) soft contact lenses is a labeling modification submission, not a submission for a new device requiring a comprehensive performance study against acceptance criteria.

    The purpose of this 510(k) is to revise the wear schedule in the labeling to include a statement about the reduced risk of giant papillary conjunctivitis (GPC) with frequent replacement.

    Therefore, many of the typical elements of a performance study required for novel device clearance are not present or applicable in this document.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This 510(k) submission does not present a table of acceptance criteria or new device performance data in the way a new device would. The core claim is a labeling modification supported by existing clinical literature, not new performance metrics for the device itself.

    The "acceptance criteria" in this context would be the FDA's acceptance of the scientific literature supporting the labeling change.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (for the labeling claim): The submission states that the labeling modification is supported by "clinical studies published in the July 1999 CLAO Journal (see Attachment A)." Attachment A is not provided in the document. Therefore, the sample size and data provenance of these specific studies are not detailed in this 510(k) submission.
    • Data Provenance: The document does not specify the country of origin of the data mentioned in the CLAO Journal. It states the studies were "published," implying they are retrospective in terms of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the 510(k) submission. As the evidence relies on published clinical studies, the details about expert involvement in establishing ground truth would typically be found within those external publications, not in this regulatory submission.

    4. Adjudication Method for the Test Set

    This information is not provided in the 510(k) submission for the same reasons as above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done, nor would it be relevant for a contact lens labeling modification based on GPC risk. This type of study is more common for diagnostic imaging or AI-assisted diagnostic devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    N/A. This device is a contact lens; there is no algorithm or human-in-the-loop performance component.

    7. The Type of Ground Truth Used

    The "ground truth" for the labeling modification regarding reduced GPC risk would be derived from the clinical outcomes data reported in the published studies cited (July 1999 CLAO Journal). This would include observations and diagnoses of GPC in patients using contact lenses with different replacement schedules.

    8. The Sample Size for the Training Set

    N/A. This device is a contact lens. There is no training set in the context of machine learning or algorithms. The "training data" for general contact lens safety and efficacy would have been the extensive clinical trials conducted for the original device clearance (K991134).

    9. How the Ground Truth for the Training Set Was Established

    N/A. See point 8. The ground truth for the original ACUVUE lens (K991134) would have been established through standard clinical trial methodologies, including patient examinations, objective measurements, and subjective patient feedback, reviewed and interpreted by qualified clinicians, to demonstrate the lens's safety and effectiveness for its intended use.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1