Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K970725

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-05-09

    (70 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K952063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Opti-Plast® XL 5.5 F PTA Catheter is recommended for use in the Percutaneous Transluminal Angioplasty of the Iliac and Femoral vessels. This catheter is not for use in coronary arteries. The device is indicated for single use only and indicated not to be resterilized nor reused.

    Device Description

    The Vas-Cath Opti-Plast® XL 5.5 F PTA Catheter is a dual lumen catheter with a balloon mounted on its distal end. One lumen accomodates the insertion guidewire and the other provides a channel for inflation and deflation of the balloon with contrast media. The catheters were designed to be used in conjuction with a 0.035 inch diameter guidewire. There are two radiopaque marker bands secured on the catheter body to indicate the position of the balloon within the vasculature. The balloon inflates to the labelled diameter and length at a 5 atmosphere operating pressure.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Opti-Plast® XL 5.5F PTA Catheter, addressing the requested information:

    Analysis of Acceptance Criteria and Study for Opti-Plast® XL 5.5F PTA Catheter

    Based on the provided 510(k) summary, the device is a medical catheter and the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its physical and functional performance, demonstrating substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are implicitly defined by the characteristics of the predicate device (Opti-Plast® PTA Catheter K933483) and industry standards for such devices. The "reported device performance" refers to the results of the bench testing.

    FeatureAcceptance Criteria (Predicate Device K933483)Reported Device Performance (Opti-Plast® XL 5.5F PTA Catheter)
    Material/Design:
    Balloon MaterialNylonNylon
    Catheter Shaft MaterialNylonNylon
    RadiopaqueYesYes
    Shaft ConfigurationDouble lumenDouble lumen
    Marker BandsYes, TantalumYes, Tantalum
    Tip Length5 mm5 mm
    Tip Inner Diameter0.037 inch0.037 inch
    Guidewire Capability0.035 inch diameter0.035 inch diameter
    Functional/Dimensional:
    Inflated Balloon Diameters10 mm (closest size for comparison)12 mm
    Inflated Balloon Length4 cm4 cm
    Recommended Operating Pressure6 atmospheres (closest size for comparison)5 atmospheres
    Rated Burst Pressure8 atmospheres8 atmospheres
    Shaft Diameter5.5 French5.5 French
    Shaft Length75, 100, 120 cm75, 100, 120 cm
    Tested Characteristics (Bench Testing indicates substantial equivalence):
    Balloon Burst StrengthPerformance comparable to K933483Met standards, substantially equivalent to K933483
    Balloon DistensibilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
    Balloon Inflation/Deflation TimePerformance comparable to K933483Met standards, substantially equivalent to K933483
    Balloon FatiguePerformance comparable to K933483Met standards, substantially equivalent to K933483
    Joints and Material TestPerformance comparable to K933483Met standards, substantially equivalent to K933483
    Catheter Shaft and Balloon ProfilePerformance comparable to K933483Met standards, substantially equivalent to K933483
    Introducer Sheath CompatibilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
    Catheter Flow/Maximum PressurePerformance comparable to K933483Met standards, substantially equivalent to K933483
    Trackability/DeflatabilityPerformance comparable to K933483Met standards, substantially equivalent to K933483
    Tip TorquePerformance comparable to K933483Met standards, substantially equivalent to K933483

    Note: For the dimensional characteristics (e.g., balloon diameter, operating pressure), the Opti-Plast® XL is different from the specific 10mm predicate balloon but is being compared for its safety and effectiveness, likely falling within acceptable variations for its intended use or being equivalent to the Medi-tech® XXL™ Balloon Dilatation Catheter (K952063) for its indications. The key assertion is that the overall performance based on the bench tests is substantially equivalent.

    2. Sample Size for the Test Set and Data Provenance

    The provided text only explicitly mentions "Bench testing." This implies the "test set" consisted of physical samples of the Opti-Plast® XL 5.5F PTA Catheter undergoing various laboratory tests.

    • Sample Size: The document does not specify the exact number of catheters tested for each characteristic.
    • Data Provenance: The data provenance is laboratory-based bench testing of the manufactured device. It is not related to patient data or human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This question is not applicable to this type of device submission. For a medical device like a catheter, "ground truth" is established through engineering specifications, material standards, and performance metrics defined by regulatory bodies and industry best practices. There are no human experts determining a "ground truth" in the way one would for diagnostic imaging.

    4. Adjudication Method for the Test Set

    This question is not applicable as the "test set" involves physical device testing against predefined engineering and performance standards, not human interpretation or consensus. No adjudication method is described or implied.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of outputs (e.g., medical images, AI algorithms). The Opti-Plast® XL 5.5F PTA Catheter is an interventional device; its performance is assessed through bench testing, not human reading of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm-only) performance study was not done. This question is relevant for AI or software as a medical device (SaMD). The Opti-Plast® XL 5.5F PTA Catheter is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on:

    • Engineering specifications and design parameters: The device is designed to meet specific dimensions, material properties, and functional characteristics.
    • Industry standards: Performance characteristics (e.g., burst pressure, fatigue) are likely evaluated against recognized ASTM or ISO standards for angioplasty catheters.
    • Predicate device performance: The performance of the Opti-Plast® XL 5.5F PTA Catheter in bench tests is compared against the established performance of the legally marketed predicate device (Opti-Plast® PTA Catheter K933483) to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" in the context of a physical medical device submission like this. Training sets are used for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as no training set exists for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1