Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K203762

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-09-17

    (268 days)

    Product Code
    Regulation Number
    868.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K944400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use to assist verification of tube placement during endotracheal intubation. Use on intubated patients to detect approximate ranges of end-tidal CO2 when clinically significant.

    Device Description

    The subject device, when connected between an endotracheal tube and a breathing device, detects approximate ranges of end-tidal CO2 by color comparison in patients. The detector may be used during patient transport or in any location where intubations are performed. It may be used for up to 2 hours. The subject device is non sterile and is intended for single use.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Nellcor™ EASYCAP II Adult Colorimetric CO2 Detector and Nellcor™ PEDCICAP Pediatric Colorimetric CO2 Detector. It focuses on demonstrating substantial equivalence to a predicate device, not on proving device performance against specific acceptance criteria through a full-fledged study as would typically be done for a novel device or AI/ML-based diagnostic.

    Therefore, many of the requested elements (like a test set sample size, expert consensus for ground truth, MRMC study, effects size of human readers, training set details) are not applicable (N/A) because the submission relies on bench testing and biocompatibility testing to establish substantial equivalence by comparing the device's characteristics to a legally marketed predicate, rather than a clinical performance study with human subjects interpreting results.

    Here's how the available information maps to your request:


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission based on substantial equivalence, the "acceptance criteria" for the device's performance are primarily that its technological characteristics are comparable to a predicate device and that any differences do not raise new questions of safety or effectiveness. The reported "performance" is demonstrated through bench testing and biocompatibility.

    Criterion TypeAcceptance Criteria (Bench Test/Spec)Reported Device Performance (as per comparison table and summary)
    Intended UseSimilar to predicate: To detect approximate ranges of end-tidal CO2.Same as predicate.
    Indications for UseSimilar to predicate: Assist verification of tube placement during endotracheal/nasotracheal intubation; detect approximate ranges of end-tidal CO2 when clinically significant on intubated patients.Similar to predicate.
    Patient PopulationNellcor™ EASYCAP II: Patients > 15 kg (33 lb).Nellcor™ PEDCICAP: Patients 1-15 kg (2.2-33 lb). (Predicate: 1-15 kg). The adult device (EASYCAP II) has a different patient population than the predicate (Pedi-CAP pediatric).Nellcor™ EASYCAP II: Different (Adult). Justification: Intended for adult patients.Nellcor™ PEDCICAP: Same as predicate.
    Duration of Use<2 hoursSame as predicate.
    SterilizationNon-sterileSame as predicate.
    Shelf LifeValidated for longer shelf life than predicate.26 Months (Predicate: 22 Months). Justification: Subject device validated for longer shelf life.
    Standard Connections15 mm connections.Same as predicate.
    Dead Space (Volume)Nellcor™ EASYCAP II: Different from predicate (larger volume for adults). Acceptable if safety/effectiveness not impacted.Nellcor™ PEDCICAP: Similar to predicate.Nellcor™ EASYCAP II: 22cc (Predicate: 3cc). Justification: Intended for adult patient, difference does not impact safety/effectiveness.Nellcor™ PEDCICAP: Same as predicate.
    Resistance to FlowNellcor™ EASYCAP II: Different from predicate (higher resistance for adult breathing patterns). Acceptable if safety/effectiveness not impacted.Nellcor™ PEDCICAP: Similar to predicate.Nellcor™ EASYCAP II: 4.4 cm H2O ± 1.0 cm at 60 L/min (Predicate: 2.5 cm H2O ± 0.5 cm at 10L/min). Justification: Intended for adult patient, difference does not impact safety/effectiveness.Nellcor™ PEDCICAP: Same as predicate.
    Detected % CO2 Ranges/ColorsA: 0.03 < 0.5 %; B: 0.05 < 2 %; C: 2 - 5 %Same as predicate.
    BiocompatibilityCompliance with ISO 10993-1 and ISO 18562 series (cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity, volatile organic compounds, particulate matter).Met all biocompatibility requirements for intended use. Material differences (Filter, Device Body) evaluated to ISO 10993-1 & ISO 18562-1 and found safe/effective.
    Bench TestingDesigns meet performance specification and evaluation for: color change/duration of use, resistance to flow, internal volume (dead space), anti-fog, shelf-life, packaging stability.Performed, and designs demonstrated to meet performance specifications and evaluation (implied conformance as substantial equivalence was granted based on these tests).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated as a "test set" in the context of clinical data for performance assessment. The "testing" mentioned is bench testing (e.g., measuring resistance to flow, dead space, shelf life, color change properties in a controlled environment). The specific number of units tested during bench validation is not detailed in this summary.
    • Data Provenance: N/A. This is not a study using patient data; it relies on engineering and material testing.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

    • N/A. Ground truth in the context of a diagnostic performance study (e.g., for AI/ML) is not relevant here. The device's function (colorimetric CO2 detection) is evaluated through its physical and chemical properties in bench tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Not applicable for a device cleared via substantial equivalence based on bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is not an AI/ML-based device. No MRMC study was performed or required.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • N/A. This is a hardware device (colorimetric detector). No algorithm or standalone performance study was conducted.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. For this type of device, "ground truth" would correspond to established engineering specifications and physical/chemical standards (e.g., a known CO2 concentration for color change, standard measurements for dead space, flow resistance). The summary indicates these were evaluated against those specifications.

    8. The sample size for the training set

    • N/A. This device does not involve a "training set" as it is not an AI/ML product.

    9. How the ground truth for the training set was established

    • N/A. Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083056

    Validate with FDA (Live)

    Device Name
    NEO-STAT CO2 <KG
    Manufacturer
    Date Cleared
    2009-02-23

    (132 days)

    Product Code
    Regulation Number
    868.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K031411,K944400,K024300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neo-StatCO2 <Kg is indicated to provide a semi-quantitative visualization of the CO2 in the patient airway. It is an adjunct in patient assessment to be used in conjunction with other methods to determine clinical signs and symptoms by or on the order of a physician. Intended for use with neonate and infant patients from 250g to 6 kg.

    Device Description

    Mercury Medical, Inc. Neo-StatCO2 <Kg is a colorimetric breath indicator for visualization of exhaled CO2. It is designed to connect between an endotracheal tube and a breathing device to help verify proper intubation. Exhaled gas passes through the indicator to detect approximate ranges of end-tidal CO2 by color comparison. Color change is rapid and the detector has the ability to show color changes with 1 ml tidal volume and up to a respiratory rate of 100 breaths per minute. The detector may be used during patient transport or in locations when intubations are performed. While the indicator functions at relative humidity (RH) of up to 100%, it is not recommended for use below 10% RH. It may be used for up to 24 hours.

    AI/ML Overview

    This document describes the Neo-StatCO2 device, a colorimetric breath indicator for CO2 that helps verify proper intubation. The submission aims to demonstrate substantial equivalence to predicate devices, particularly the Mercury Medical Mini-StatCO2 (K031411).

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document implicitly defines acceptance criteria by comparing the new device (Neo-StatCO2) to predicate devices in terms of its physical, performance, and operating characteristics. The primary new "performance" aspect is the expanded patient population, which required specific testing.

    Feature / Criteria (Derived from Predicate Comparison)Acceptance Criteria (Predicate)Reported Device Performance (Neo-StatCO2)
    Indications for Use"semi-quantitative visualization of the CO2 in the patient airway. Adjunct in patient assessment... in conjunction with other methods... by or on the order of a physician."Identical to predicate (Mercury Mini-StatCO2 K031411)
    Environment of UseHospital, sub-acute, pre-hospital & transportIdentical to predicate (Mercury Mini-StatCO2 K031411)
    Patient PopulationNeonate and Pediatric (Mini-StatCO2), Adult, Pediatric, Infant/Neonatal (Oridion MicroCap)Neonate and infant (250g to 6 kg) Expanded from predicate, similar to Oridion MicroCap.
    Weight5 g3 g
    Internal Volume3 cc1 cc
    Resistance to Flow1.23 cm H2O @ 10 Lpm; 8.8 cm H2O @ 30 Lpm1.86 cm H2O @ 10 Lpm; 15.6 cm H2O @ 30 Lpm
    Duration of CO2 DetectionUp to 24 hoursUp to 24 hours
    Operating Conditions-10°C to +50°C-10°C to +50°C
    Storage ConditionsColor change function after storage at 0°C and 60°CColor change function after storage at 0°C and 60°C
    Contraindications & WarningsDo not use for detection of hypercapnia, main-stem bronchial intubation, mouth-to-tube ventilation, oropharyngeal tube placement. Standard clinical assessment must be used.Same as predicate
    Effectiveness of color change for intended populationDemonstrated ability of color change for predicate's population.Performance testing to demonstrate ability of indicator to change color for the intended body weight, tidal volume and breath rate under conditions of use. (This is the specific study mentioned).

    Study Proving Device Meets Acceptance Criteria:

    The document explicitly states: "The only difference between the Mercury Mini-Stat CO2 is the expansion of the patient population to include patients with a body weight of 250 g to 6 Kg which we have demonstrated as being effective via comparative bench testing."

    This implies that the "comparative bench testing" was the study conducted to ensure the device's effectiveness for the expanded patient population (neonate and infant - 250g to 6 kg) under the specified tidal volumes and breath rates.

    Additional Information (Based on the Provided Text):

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified. The document mentions "comparative bench testing" but does not give specific numbers for the test set used in this bench testing.
    • Data provenance: The testing was described as "bench testing," suggesting a laboratory environment. There is no information about the country of origin or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the testing was "bench testing." Ground truth would likely be established by objective measurements of CO2, tidal volume, and breath rate, rather than expert interpretation of the device's color change.

    4. Adjudication method for the test set:

    • Not applicable for bench testing. The evaluation would be based on direct measurement outcomes rather than human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is a colorimetric indicator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The "comparative bench testing" would be a standalone performance evaluation of the device's ability to change color under specific simulated conditions, independent of human interpretation for the physical manifestation of the color change itself. The human-in-the-loop aspect refers to a clinician interpreting the color output. The bench testing would confirm the device produces the correct color change.

    7. The type of ground truth used:

    • For the "comparative bench testing," the ground truth would be precise, known values of CO2 concentration, tidal volume, and respiratory rate, simulated in a controlled environment. The device's color change would then be observed and compared against these known conditions to verify its accuracy.

    8. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI/machine learning algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established:

    • Not applicable as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042572

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-11-05

    (45 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K894053, K944400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Used to assist verification of tube location during oral or nasal placement of a gastric tube in adults.

    Device Description

    The Kendall CO2nfirm NOW™ CO2Detector is a disposable device that is used during the oral or nasal placement of a gastric tube to preliminarily determine if the tube is properly placed by detection of CO2

    AI/ML Overview

    Based on the provided 510(k) summary for the Kendall C02nfirm NOW™ C02 Detector, here's an analysis of the acceptance criteria and supporting study information:

    The Kendall C02nfirm NOW™ C02 Detector is a disposable device used to assist in the verification of gastric tube placement by detecting CO2.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly list quantitative acceptance criteria or specific performance metrics for the Kendall C02nfirm NOW™ CO2 Detector. Instead, it relies on demonstrating equivalence to predicate devices and meeting industry standards.

    Acceptance Criteria (Explicitly Stated)Reported Device Performance
    Equivalence to predicate devices (Easy Cap® CO2 detector, Pedi-Cap® CO2 detector) in technological characteristics, construction, materials, detector chemistry, and sensitivity.The proposed device has the same technological characteristics as the predicate devices. "Both the proposed and predicate devices function by colorimetric detection of CO2 in air. The construction and materials of the devices, as well as the detector chemistry and sensitivity are identical between the proposed and predicate devices."
    Adherence to industry accepted criteria for such devices as defined in EN1615:2000 and ISO 594, Parts 1&2."Testing was conducted to demonstrate that the design of the proposed device... met the industry accepted criteria for such devices as defined in EN1615:2000 and ISO 594, Parts 1&2." (No specific performance data against these standards is provided in the summary.)
    Modified inlet/outlet ports to accommodate new intended use (gastric tube placement)."The inlet and outlet ports of the proposed device have been modified to accommodate the new intended use." (Implied successful modification, no specific performance data given.)
    Modified device printing to replace quantitative scale with qualitative (yes/no) scale."The device printing has been modified to replace the quantitative scale with a qualitative (yes/no) scale." (Implied successful modification, no specific performance data given.)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide details on a specific "test set" in terms of subject or patient data. The non-clinical testing described focuses on demonstrating design equivalence to predicate devices and adherence to industry standards for device construction and function. It does not mention a clinical study involving human subjects or patient data. Therefore, there's no information on sample size, country of origin, or whether it was retrospective or prospective for a clinical test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Since no clinical test set with human subjects or patient data is described, there is no mention of experts being used to establish ground truth in this context. The evaluation appears to be purely technical and comparative against existing devices and standards.

    4. Adjudication Method for the Test Set

    Given the absence of a clinical test set and expert review for ground truth, no adjudication method is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. The submission focuses on substantial equivalence to predicate devices, not on comparing human reader performance with and without AI assistance. The device is a diagnostic tool, not an AI-powered image analysis system.

    6. Standalone Performance Study

    The summary describes "Nonclinical Testing" conducted to demonstrate design equivalence and adherence to industry standards. This can be considered a form of standalone testing in the sense that the device itself was tested for its physical and functional characteristics. However, it's not a standalone clinical performance study in the way it's typically understood for diagnostic algorithms (e.g., performance metrics like sensitivity, specificity on a clinical dataset). The testing described is more about the intrinsic properties of the device (colorimetric detection, materials, construction) rather than its diagnostic accuracy in a clinical setting.

    7. Type of Ground Truth Used

    For the nonclinical testing, the "ground truth" implicitly used would be:

    • Predicate Device Performance: The established functionality and performance of the Easy Cap® CO2 detector and Pedi-Cap® CO2 detector.
    • Industry Standards: The requirements and specifications outlined in EN1615:2000 and ISO 594, Parts 1&2.
      The summary does not mention pathology, outcomes data, or expert consensus in a clinical context as ground truth.

    8. Sample Size for the Training Set

    No training set is mentioned. This device is a passive, colorimetric CO2 detector, not a machine learning or AI-driven system that requires a "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML model, this question is not applicable. The device's function is based on a well-understood chemical principle, not on learned patterns from a dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991953

    Validate with FDA (Live)

    Date Cleared
    1999-07-01

    (30 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K894053, K944400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable manual resuscitator with CO2 detection for use with patients requiring manual ventilatory support. The CO2 detector is located on the expiratory port. It may be supplied separately and attached to any resuscitator expiratory port. The CO2 detector assists in verification of the endotracheal tube placing during endotracheal or nasotracheal intubation. The CO2 detector detects approximate ranges of CO2 by color comparison. It is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as in the hospital, transport, mobile or home settings.

    Device Description

    The EMS Disposable Manual Resuscitator is a bag-valve-mask device with the capability of delivering supplemental oxygen. It incorporates a feature CO2 detection.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EMS Disposable Manual Resuscitator with CO2 Detection, based on the provided text:

    Preamble:

    It's important to note that this document is a 510(k) submission, not a full clinical study report. The primary goal of a 510(k) is to demonstrate substantial equivalence to a predicate device, rather than proving novel efficacy or conducting extensive new performance studies. Therefore, the "study" referred to here is primarily a comparison to existing, legally marketed devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the attributes of the predicate devices. The "reported device performance" is the statement of equivalence to these attributes.

    AttributeEMS DMR with CO₂ (Modification)Predicate Devices (EMS DMR K912203B, K924610A, NPB DMR Plus with CO₂ K973419)Acceptance Criteria (implied by predicate)Reported Device Performance
    Use
    Intended to provide manual ventilatory supportYesYesYesYes
    Intended to assist in verification of tube placement by expired CO₂ detectionYesYes (for CO₂-enabled predicates)YesYes
    Used in hospitals, home, transport, mobile settingsYesYesYesYes
    Single Patient UseYesYesYesYes
    Design
    CO₂ detector placed on expiratory port of manual resuscitatorYesYes (for CO₂-enabled predicates)YesYes
    CO₂ detection by color comparisonYesYes (for CO₂-enabled predicates)YesYes
    CO₂ detector good for up to 2 hoursYesYes (for CO₂-enabled predicates)YesYes
    Can be replaced if neededYesYes (for CO₂-enabled predicates)YesYes
    Materials
    Materials in CO₂ detection media exactly the same as predicateYesYes (for CO₂-enabled predicates)YesYes
    All other materials exactly the same as predicateYesYes (for non-CO₂ enabled predicates)YesYes
    Packaging
    Provided clean, non-sterileYesYesYesYes
    Performance Standards / Specifications
    None applicable under Section 514YesYesYesYes

    Summary of Device Performance:

    The document explicitly states: "There are no differences between the proposed modification, that of adding CO2 detection capabilities to the exhalation port of the manual resuscitator, and the predicates." This statement implies that the EMS Disposable Manual Resuscitator with CO₂ Detection meets all the acceptance criteria by being functionally and materially identical or equivalent to the predicate devices with CO₂ detection capabilities, and other attributes identical to other predicates.


    Regarding "The Study That Proves the Device Meets the Acceptance Criteria":

    The provided text doesn't describe a traditional "study" with a test set, experts, and ground truth in the way one might expect for a novel device's performance evaluation. Instead, the "proof" for a 510(k) submission primarily relies on:

    • Comparison to legally marketed predicate devices: The core argument is that the device is substantially equivalent to existing devices, meaning it has the same intended use, similar technological characteristics, and raises no new questions of safety or effectiveness.
    • Verification of attributes: The table explicitly compares attributes of the new device to predicates, indicating "Yes" for matching characteristics.
    • Statement of no differences: The crucial statement "There are no differences between the proposed modification... and the predicates" serves as the primary evidence.

    Given this context, the following points address your specific questions:

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: Not applicable in the context of a 510(k) substantial equivalence submission based on predicate device comparison. There isn't a "test set" of patient data or device performance data gathered for this specific submission beyond the functional demonstration inherent in comparing specifications.
    • Data Provenance: Not applicable. The "data" here is the descriptive comparison of the device's design, materials, and intended use against legally marketed predicate devices (EMS DMR K912203B, K924610A, and Nellcor Puritan Bennett DMR2 Plus with CO2 Detection K973419, and Nellcor Puritan Bennett EasyCap CO₂ Detector K894053 and K944400). This is a retrospective comparison of device specifications/features to already approved devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. Ground truth, in the sense of expert consensus on patient data or clinical outcomes, is not established for this type of submission. The "ground truth" is effectively the established safety and effectiveness of the predicate devices as determined by their prior FDA clearances.

    4. Adjudication Method for the Test Set

    • Not applicable. There is no "adjudication method" in the context of this 510(k) submission. The comparison is a direct assessment of device characteristics against pre-defined market standards (the predicate devices).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not applicable. This device is a manual medical device (resuscitator with CO₂ detection), not an AI-powered diagnostic or interpretive tool. Therefore, MRMC studies and concepts of human reader improvement with AI assistance are irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a manual medical device, not an algorithm. Its function inherently involves human operation.

    7. The Type of Ground Truth Used

    • Implicit "Ground Truth": The "ground truth" is the established safety and effectiveness of the predicate devices (EMS Disposable Manual Resuscitator K912203B, K924610A, and Nellcor Puritan Bennett DMR2 Plus Manual Resuscitator with CO2 Detection K973419, and Nellcor Puritan Bennett EasyCap CO₂ Detector K894053 and K944400). The current device is deemed safe and effective because its characteristics are substantially equivalent to these already approved devices.

    8. The Sample Size for the Training Set

    • Not applicable. This is not a machine learning or AI device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1