Search Results
Found 2 results
510(k) Data Aggregation
(389 days)
AnapnoGuard 100 Respiratory Guard System is intended for airway management by oral/nasal intubation while providing continuous endotracheal cuff pressure control using non-invasive measurement and monitoring of carbon dioxide concentration in the subglottic space and evacuation of secretions from above the endotracheal tube's cuff.
AnapnoGuard 100 Respiratory Guard System is comprised of the following three main components: The AnapnoGuard endotracheal tube (ETT) with inflatable cuff (FDA cleared under K093126). The AnapnoGuard 100 Respiratory Guard System interconnection harness of tubes, connecting the ETT to the AnapnoGuard 100 control unit The AnapnoGuard 100 Respiratory Guard System control unit which consists of the following main modules: Host computer (PC) Microcontroller (MCU) Suction module (regulator and flow potency meter): including . a set of valves and pipes controlling the secretions suction/evacuation from above the ETT cuff. Rinsing module: Pumps saline to rinse the Suction and Vent/ ● CO2 lumens. CO2 analyzer module: including CO2 analyzer assembly, ● pump, valve and flow filter which sucks air from the subglottic space above the ETT cuff into the CO2 analyzer. . Cuff pressure module: includes two pressure gauges which monitor cuff pressure, a miniature air pump and two valves. Pneumatic module: valves, pipes and filters . Connectors panel for connecting the interconnection harness . (ETT), vacuum, trap bottle, rinsing fluid and filters. Operation buttons panel and navigation wheel . I/O communication panel . Display monitor . AnapnoGuard 100 Respiratory Guard System, including its three components monitors leak between the endotracheal tube's cuff and the trachea by measuring the Carbon Dioxide levels in the subglottic area above the cuff through a dedicated lumen in the endotracheal tube. Detection of a high level of Carbon Dioxide is an objective indicator for a leak (improper sealing of the trachea by the endotracheal tube cuff). The system continuously monitors and adjusts the cuff pressure to prevent a leak at minimum possible pressure (all within pressure limits preset by the user). Preventing a leak reduces the likelihood of aspiration of secretions from the upper airways into the lungs and increases the likelihood for no loss of ventilation and delivery of anesthetic and nebulized drugs into the lungs. Keeping the cuff pressure as low as possible reduces the mechanical pressure of the cuff on the tracheal tissue throughout the intubation period. The system also performs evacuation of secretions from above the endotracheal tube's cuff through a dedicated lumen at the dorsal side of the endotracheal tube.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The document describes several performance criteria derived from a comparison with predicate devices and specific performance bench tests. The clinical study primarily focuses on the effectiveness of the AnapnoGuard 100 Respiratory Guard System in optimizing cuff pressure and reducing CO2 leakage.
Table 1: Acceptance Criteria and Reported Device Performance
Feature/Metric | Acceptance Criteria (from predicate/bench tests) | Reported Device Performance (from clinical study/bench tests) |
---|---|---|
Cuff Pressure Control | ||
Minimum measured pressure | 15 mmHg (PYTON predicate) | 0 mmHg (AnapnoGuard 100) |
Maximal cuff pressure | 22 mmHg (PYTON predicate) | 33 mmHg (AnapnoGuard 100) |
Control Accuracy | ± 0.73 mmHg (PYTON predicate) | ± 0.1 mmHg (AnapnoGuard 100) |
Recording Accuracy | ± 0.73 mmHg (PYTON predicate) | ± 0.1 mmHg (AnapnoGuard 100) |
Pressure drop Alarm time | N/A (PYTON predicate) | 0.2 Sec (AnapnoGuard 100) |
Cuff Pressure Safety Test | Maintain pressures precision and safety boundaries | Validated: Design maintains cuff pressures precision and safety boundaries. |
Pressure Maintenance Comparison | To maintain constant pressure (Tracoe) | Compared favorably: Ability to maintain constant pressure compared to Tracoe Pressure Regulator demonstrated. |
CO2 Leakage / Measurement | ||
CO2 Sensor Precision Test | N/A | Test performed. (Specific precision not quantified in this summary, but device claims ±0.1 mmHg control/recording accuracy.) |
Overall duration and level of around ETT cuff leakage (CO2 Area under the Curve - AUC) | Non-inferiority to standard of care/reduction in leakage | Study Group: 0.09±0.04 |
Control Group: 0.22±0.32 | ||
(Statistically significant reduction in study group, p twice the control group (mean ratio Study/Control = 2.03, P |
Ask a specific question about this device
(132 days)
The Neo-StatCO2
Mercury Medical, Inc. Neo-StatCO2
This document describes the Neo-StatCO2 device, a colorimetric breath indicator for CO2 that helps verify proper intubation. The submission aims to demonstrate substantial equivalence to predicate devices, particularly the Mercury Medical Mini-StatCO2 (K031411).
1. Table of Acceptance Criteria and Reported Device Performance:
The document implicitly defines acceptance criteria by comparing the new device (Neo-StatCO2) to predicate devices in terms of its physical, performance, and operating characteristics. The primary new "performance" aspect is the expanded patient population, which required specific testing.
Feature / Criteria (Derived from Predicate Comparison) | Acceptance Criteria (Predicate) | Reported Device Performance (Neo-StatCO2) |
---|---|---|
Indications for Use | "semi-quantitative visualization of the CO2 in the patient airway. Adjunct in patient assessment... in conjunction with other methods... by or on the order of a physician." | Identical to predicate (Mercury Mini-StatCO2 K031411) |
Environment of Use | Hospital, sub-acute, pre-hospital & transport | Identical to predicate (Mercury Mini-StatCO2 K031411) |
Patient Population | Neonate and Pediatric (Mini-StatCO2), Adult, Pediatric, Infant/Neonatal (Oridion MicroCap) | Neonate and infant (250g to 6 kg) Expanded from predicate, similar to Oridion MicroCap. |
Weight | 5 g | 3 g |
Internal Volume | 3 cc | 1 cc |
Resistance to Flow | 1.23 cm H2O @ 10 Lpm; 8.8 cm H2O @ 30 Lpm | 1.86 cm H2O @ 10 Lpm; 15.6 cm H2O @ 30 Lpm |
Duration of CO2 Detection | Up to 24 hours | Up to 24 hours |
Operating Conditions | -10°C to +50°C | -10°C to +50°C |
Storage Conditions | Color change function after storage at 0°C and 60°C | Color change function after storage at 0°C and 60°C |
Contraindications & Warnings | Do not use for detection of hypercapnia, main-stem bronchial intubation, mouth-to-tube ventilation, oropharyngeal tube placement. Standard clinical assessment must be used. | Same as predicate |
Effectiveness of color change for intended population | Demonstrated ability of color change for predicate's population. | Performance testing to demonstrate ability of indicator to change color for the intended body weight, tidal volume and breath rate under conditions of use. (This is the specific study mentioned). |
Study Proving Device Meets Acceptance Criteria:
The document explicitly states: "The only difference between the Mercury Mini-Stat CO2 is the expansion of the patient population to include patients with a body weight of 250 g to 6 Kg which we have demonstrated as being effective via comparative bench testing."
This implies that the "comparative bench testing" was the study conducted to ensure the device's effectiveness for the expanded patient population (neonate and infant - 250g to 6 kg) under the specified tidal volumes and breath rates.
Additional Information (Based on the Provided Text):
2. Sample size used for the test set and the data provenance:
- Sample size: Not specified. The document mentions "comparative bench testing" but does not give specific numbers for the test set used in this bench testing.
- Data provenance: The testing was described as "bench testing," suggesting a laboratory environment. There is no information about the country of origin or whether it was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as the testing was "bench testing." Ground truth would likely be established by objective measurements of CO2, tidal volume, and breath rate, rather than expert interpretation of the device's color change.
4. Adjudication method for the test set:
- Not applicable for bench testing. The evaluation would be based on direct measurement outcomes rather than human adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a colorimetric indicator, not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, implicitly. The "comparative bench testing" would be a standalone performance evaluation of the device's ability to change color under specific simulated conditions, independent of human interpretation for the physical manifestation of the color change itself. The human-in-the-loop aspect refers to a clinician interpreting the color output. The bench testing would confirm the device produces the correct color change.
7. The type of ground truth used:
- For the "comparative bench testing," the ground truth would be precise, known values of CO2 concentration, tidal volume, and respiratory rate, simulated in a controlled environment. The device's color change would then be observed and compared against these known conditions to verify its accuracy.
8. The sample size for the training set:
- Not applicable. This is a physical device, not an AI/machine learning algorithm that requires a training set in the conventional sense.
9. How the ground truth for the training set was established:
- Not applicable as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1