K Number
K203762

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-09-17

(268 days)

Product Code
Regulation Number
868.1400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use to assist verification of tube placement during endotracheal intubation. Use on intubated patients to detect approximate ranges of end-tidal CO2 when clinically significant.

Device Description

The subject device, when connected between an endotracheal tube and a breathing device, detects approximate ranges of end-tidal CO2 by color comparison in patients. The detector may be used during patient transport or in any location where intubations are performed. It may be used for up to 2 hours. The subject device is non sterile and is intended for single use.

AI/ML Overview

The provided text describes the 510(k) summary for the Nellcor™ EASYCAP II Adult Colorimetric CO2 Detector and Nellcor™ PEDCICAP Pediatric Colorimetric CO2 Detector. It focuses on demonstrating substantial equivalence to a predicate device, not on proving device performance against specific acceptance criteria through a full-fledged study as would typically be done for a novel device or AI/ML-based diagnostic.

Therefore, many of the requested elements (like a test set sample size, expert consensus for ground truth, MRMC study, effects size of human readers, training set details) are not applicable (N/A) because the submission relies on bench testing and biocompatibility testing to establish substantial equivalence by comparing the device's characteristics to a legally marketed predicate, rather than a clinical performance study with human subjects interpreting results.

Here's how the available information maps to your request:


Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission based on substantial equivalence, the "acceptance criteria" for the device's performance are primarily that its technological characteristics are comparable to a predicate device and that any differences do not raise new questions of safety or effectiveness. The reported "performance" is demonstrated through bench testing and biocompatibility.

Criterion TypeAcceptance Criteria (Bench Test/Spec)Reported Device Performance (as per comparison table and summary)
Intended UseSimilar to predicate: To detect approximate ranges of end-tidal CO2.Same as predicate.
Indications for UseSimilar to predicate: Assist verification of tube placement during endotracheal/nasotracheal intubation; detect approximate ranges of end-tidal CO2 when clinically significant on intubated patients.Similar to predicate.
Patient PopulationNellcor™ EASYCAP II: Patients > 15 kg (33 lb).Nellcor™ PEDCICAP: Patients 1-15 kg (2.2-33 lb). (Predicate: 1-15 kg). The adult device (EASYCAP II) has a different patient population than the predicate (Pedi-CAP pediatric).Nellcor™ EASYCAP II: Different (Adult). Justification: Intended for adult patients.Nellcor™ PEDCICAP: Same as predicate.
Duration of Use<2 hoursSame as predicate.
SterilizationNon-sterileSame as predicate.
Shelf LifeValidated for longer shelf life than predicate.26 Months (Predicate: 22 Months). Justification: Subject device validated for longer shelf life.
Standard Connections15 mm connections.Same as predicate.
Dead Space (Volume)Nellcor™ EASYCAP II: Different from predicate (larger volume for adults). Acceptable if safety/effectiveness not impacted.Nellcor™ PEDCICAP: Similar to predicate.Nellcor™ EASYCAP II: 22cc (Predicate: 3cc). Justification: Intended for adult patient, difference does not impact safety/effectiveness.Nellcor™ PEDCICAP: Same as predicate.
Resistance to FlowNellcor™ EASYCAP II: Different from predicate (higher resistance for adult breathing patterns). Acceptable if safety/effectiveness not impacted.Nellcor™ PEDCICAP: Similar to predicate.Nellcor™ EASYCAP II: 4.4 cm H2O ± 1.0 cm at 60 L/min (Predicate: 2.5 cm H2O ± 0.5 cm at 10L/min). Justification: Intended for adult patient, difference does not impact safety/effectiveness.Nellcor™ PEDCICAP: Same as predicate.
Detected % CO2 Ranges/ColorsA: 0.03 < 0.5 %; B: 0.05 < 2 %; C: 2 - 5 %Same as predicate.
BiocompatibilityCompliance with ISO 10993-1 and ISO 18562 series (cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity, volatile organic compounds, particulate matter).Met all biocompatibility requirements for intended use. Material differences (Filter, Device Body) evaluated to ISO 10993-1 & ISO 18562-1 and found safe/effective.
Bench TestingDesigns meet performance specification and evaluation for: color change/duration of use, resistance to flow, internal volume (dead space), anti-fog, shelf-life, packaging stability.Performed, and designs demonstrated to meet performance specifications and evaluation (implied conformance as substantial equivalence was granted based on these tests).

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated as a "test set" in the context of clinical data for performance assessment. The "testing" mentioned is bench testing (e.g., measuring resistance to flow, dead space, shelf life, color change properties in a controlled environment). The specific number of units tested during bench validation is not detailed in this summary.
  • Data Provenance: N/A. This is not a study using patient data; it relies on engineering and material testing.

3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

  • N/A. Ground truth in the context of a diagnostic performance study (e.g., for AI/ML) is not relevant here. The device's function (colorimetric CO2 detection) is evaluated through its physical and chemical properties in bench tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • N/A. Not applicable for a device cleared via substantial equivalence based on bench testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • N/A. This is not an AI/ML-based device. No MRMC study was performed or required.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • N/A. This is a hardware device (colorimetric detector). No algorithm or standalone performance study was conducted.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • N/A. For this type of device, "ground truth" would correspond to established engineering specifications and physical/chemical standards (e.g., a known CO2 concentration for color change, standard measurements for dead space, flow resistance). The summary indicates these were evaluated against those specifications.

8. The sample size for the training set

  • N/A. This device does not involve a "training set" as it is not an AI/ML product.

9. How the ground truth for the training set was established

  • N/A. Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Covidien Ashley Johnston Regulatory Affairs Manager 6135 Gunbarrel Ave Boulder, Colorado 80301

Re: K203762

Trade/Device Name: Nellcor EASYCAP II Adult Colorimetric CO2 Detector, Nellcor PEDCICAP Pediatric Colorimetric CO2 Detector Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: August 19, 2021 Received: August 20, 2021

Dear Ashley Johnston:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203762

Device Name

Nellcor™ EASYCAP II Adult Colorimetric CO2 Detector

Nellcor™ PEDICAP Pediatric Colorimetric CO2 Detector

Indications for Use (Describe)

Use to assist verification of tube placement during endotracheal intubation. Use on intubated patients to detect approximate ranges of end-tidal CO2 when clinically significant.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white cross in the center, followed by the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, and the colors are clean and professional.

510(k) SUMMARY K203762

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a 510(k) Summary for the Nellcor™ EASYCAP II Colorimetric CO2 Detector and Nellcor™ Pediatric Colorimetric CO2 Detector.

Covidien LLC

I. SUBMITTER

Submitted By:

6135 Gunbarrel AvenueBoulder, CO 80301
Date:21 December 2020
Contact Person:Ashley JohnstonRegulatory Affairs Manager720-440-1057

DEVICE II.

III.

DEVICE
Proprietary Name:Nellcor™ EASYCAP II Adult Colorimetric CO2 DetectorNellcor™ PEDCICAP Pediatric Colorimetric CO₂ Detector
Common Name:Carbon dioxide gas analyzer
Device Classification Regulation:21 CFR 868.1400
Device Product Code & Panel:CCK / Anesthesiology (73)
Class:II
Predicate Device
Predicate Devices:Pedi-CAP™ Pediatric End-Tidal CO₂ Detector

(K944400)

IV. Device Description

The subject device, when connected between an endotracheal tube and a breathing device, detects approximate ranges of end-tidal CO2 by color comparison in patients. The detector may be used during patient transport or in any location where intubations are performed. It may be used for up to 2 hours.

The subject device is non sterile and is intended for single use.

The EASYCAP II is used in adult patients weighing more than 15 kg (33 lb), and the PEDICAP is used in pediatric patients weighing 1 - 15 kg (2.2 - 33 lb).

V. Indications for Use/Intended Use

Use to assist verification of tube placement during endotracheal intubation. Use on intubated patients to detect approximate ranges of end-tidal CO2 when clinically significant.

COVIDIEN LLC 15 HAMPSHIRE STREET MANSFIELD, MA 02048

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a grid pattern in lighter shades of blue on the left, and the word "COVIDIEN" in blue, sans-serif font on the right. The logo is simple and modern, and the colors are calming and professional.

VI. Technological Characteristics Comparison

The subject device has similar indications for use, design and functionality to the predicate device. A comparison table is provided below.

Device Comparison Table: Subject Device, Predicate Device
Predicate DeviceSubject Device Comparison
FeaturesNellcor Pedi-CAP™ Pediatric End-Tidal CO2 Detector (K944400)Nellcor™ EASYCAP II ColorimetricCO2 DetectorNellcor™ PEDICAP ColorimetricCO2 Detector
Intended UseTo detect approximate ranges of end-tidal CO2Same as predicateSame as predicate
Indications forUseIntended for use duringendotracheal or nasotrachealintubation to assist verification oftube placement.It is also intended for use onintubated patients where measuringapproximate ranges of end-tidalCO2 may be clinically significantand other more precise methods arenot feasible or available.SimilarUse to assist verification of tubeplacement during endotrachealor nasotracheal intubation.Use on intubated patients to detectapproximate ranges of end-tidal CO2when clinically significant.SimilarUse to assist verification oftube placement duringendotracheal or nasotrachealintubation.Use on intubated patients to detectapproximate ranges of end-tidal CO2when clinically significant.
PatientPopulationPatients weighing 1–15 kg (2.2-33 lb).DifferentPatients weighing more than 15 kg (33 lb).Justification for difference:The subject is used for adult patients.Same as predicate
UseSingle patientSame as predicateSame as predicate
FeaturesPredicate DeviceSubject Device Comparison
Nellcor Pedi-CAP™ Pediatric End-Tidal CO2 Detector (K944400)Nellcor™ EASYCAP II ColorimetricCO2 DetectorNellcor™ PEDICAP ColorimetricCO2 Detector
Duration of Use<2 hoursSame as predicateSame as predicate
Features, Specification and Performance
SterilizationNon-sterileSame as predicateSame as predicate
Shelf Life22 MonthsDifferentDifferent
26 Months26 Months
Justification for difference:The subject device was validated forlonger shelf life.Justification for difference:The subject device was validated forlonger shelf life.
Standard 15 mmconnectionsYesSame as predicateSame as predicate
Dead Space(Volume)3ccDifferent22ccJustification for difference:The subject device is intended for adultpatient, therefore has different deadspace in comparison to the predicatedevice.The difference in the dead space(volume) does not impact the safety andeffectiveness of the device.Same as predicate
FeaturesPredicate DeviceSubject Device Comparison
Nellcor Pedi-CAPTM Pediatric End-Tidal CO2 Detector (K944400)NellcorTM EASYCAP II ColorimetricCO2 DetectorNellcorTM PEDICAP ColorimetricCO2 Detector
Resistance to Flow$2.5 cm H_2O \pm 0.5 cm$ at 10L/minDifferent$4.4 cm H_2O \pm 1.0 cm$ at 60 L/minJustification for difference:The subject device is intended for adultpatient, therefore has differentresistance to flow in comparison to thepredicate device. The difference in theresistance to flow does not impact thesafety and effectiveness of the device.Same as predicate
Detected % CO2Ranges andColorsA: 0.03 < 0.5 %B: 0.05 < 2 %C: 2 - 5 %Same as predicateSame as predicate
Materials
Device CoverAcrylic copolymerSame as predicateSame as predicate
pH IndicatorPaperpH-sensitive indicator paperSame as predicateSame as predicate
Device SupportPolypropyleneSame as predicateSame as predicate
FeaturesPredicate DeviceSubject Device Comparison
Nellcor Pedi-CAP™ Pediatric End-Tidal CO2 Detector (K944400)Nellcor™ EASYCAP II ColorimetricCO2 DetectorNellcor™ PEDICAP ColorimetricCO2 Detector
FilterPolypropyleneDifferentPolypropylene and acrylicJustification fordifference:The material has been evaluated to ISO10993-1 & ISO 18562-1 and found tobe safe and effective.DifferentPolypropylene and acrylic Justification fordifference:The material has been evaluated to ISO10993-1 & ISO 18562-1and found tobe safe and effective.
Device BodyAcrylic copolymerSimilarAcrylic polymerJustification fordifference:This material has been evaluated toISO 10993-1& ISO 18562-1 andfound to be safe and effective.Same as predicate

COVIDIEN LLC 15 HAMPSHIRE STREET MANSFIELD, MA 02048

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the word "COVIDIEN" in a bold, sans-serif font. To the left of the word is a blue square divided into smaller squares of varying shades of blue. The word "COVIDIEN" is also in blue, and the overall design appears to be a logo.

Device Comparison Table: Subject Device, Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "COVIDIEN" in a bold, sans-serif font. To the left of the word is a blue square logo. The logo is made up of smaller squares of varying shades of blue.

Device Comparison Table: Subiect Device, Predicate Device

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a lighter blue horizontal stripe in the middle, followed by the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, with a focus on the company name.

Device Comparison Table: Subject Device, Predicate Device

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square made up of smaller squares of varying shades of blue. To the right of the square is the word "COVIDIEN" in blue sans-serif font.

Substantial Equivalence - Summary of Performance Testing

The following performance testing were performed in support of the substantial equivalence determination:

Biocompatibility testing

The biocompatibility evaluation was conducted in accordance with FDA Guidance "Use of International Standard ISO-10993, 'Bibogical Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" June 16, 2016, ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process' and ISO 18562-1 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process".

The biocompatibility tests for external communicating with tissue/bone/dentin with limited (less than 24 hours) include: cytotoxicity, sensitization, acute systemic toxicity, and material-mediated pvrogenicity, as well as an assessment to demonstrate compliance to the ISO 18562 standard series, including volatile organic compounds and particulate matter testing. The subject device met all biocompatibility requirements for its intended use.

Bench Testing

Performance bench testing including color change/duration of use, resistance to flow, internal volume(dead space), anti-fog, shelf-life, packaging stability and biocompatibility were performed on the subject devices to verify the designs meet performance specification and evaluation.

Substantial Equivalence - Clinical Evidence

N/A - Clinical evidence was not necessary to show substantial equivalence.

Substantial Equivalence - Conclusions

The subject device and the predicate device have the same intended use and the differences in technological features do not raise different questions of safety and effectiveness. From the evidence presented in the Premarket Notification, the subject device can be considered substantially equivalent.

§ 868.1400 Carbon dioxide gas analyzer.

(a)
Identification. A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.(b)
Classification. Class II (performance standards).