Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K042206

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-10-18

    (63 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K894317

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To verify the performance of the HMS instrument and HMS Heparin Assay cartridges

    Device Description

    The Purple/Black Heparin Assay control is an in vitro diagnostic device. This control is intended to verify the performance of the HMS instrument and the Purple/Black HMS Heparin Assay cartridge. The cartridge is for use in the HMS Plus (Hemostasis Management System) instrument.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Purple/Black Heparin Assay Control." However, the document focuses on establishing substantial equivalence to a predicate device based on design, test results, and indications for use. It does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance, nor does it describe a study that explicitly proves the device meets specific acceptance criteria in the manner requested by the prompt.

    The document mentions "Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device," but it does not provide any details about this testing.

    Therefore, most of the requested information cannot be extracted from the provided text.

    Here is what can be inferred or stated as "not provided":

    Acceptance Criteria and Study Details

    Information CategoryDetails from Provided Text
    1. Acceptance Criteria and Reported Device PerformanceNot provided in the document. The summary states, "Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device," but no specific criteria or performance metrics are listed.
    2. Sample size and Data Provenance (Test Set)Not provided. The document refers to "Validation testing," but no details on sample size, specific test sets, or data provenance (country, retrospective/prospective) are given.
    3. Number and Qualifications of Experts (Test Set GT)Not applicable/Not provided. The document describes an in vitro diagnostic control for an assay, not an AI or imaging device requiring expert ground truth for interpretation. There is no mention of experts establishing ground truth for any test set.
    4. Adjudication Method (Test Set)Not applicable/Not provided. No adjudication method is mentioned as there is no expert-based ground truth establishment described.
    5. MRMC Comparative Effectiveness Study (AI vs. without AI)Not applicable. This device is an in vitro diagnostic control, not an AI-assisted diagnostic tool that would involve human readers. Therefore, an MRMC study is not relevant here.
    6. Standalone Algorithm-Only Performance StudyNot applicable. This is an in vitro diagnostic control, not a standalone algorithm. The "validation testing" likely assessed the control's stability, reproducibility, and expected values.
    7. Type of Ground Truth UsedNot applicable/Not explicitly stated for the "validation testing." For a control solution, the "ground truth" would typically be the established lot values (target range with known uncertainty) for the analytes it controls, determined through rigorous laboratory methods (e.g., reference methods, inter-laboratory studies). The document states the control verifies performance against the HMS instrument and cartridges, implying the "truth" is the expected performance within the system.
    8. Sample size for the training setNot applicable. As an in vitro diagnostic control for an assay, this device does not involve a "training set" in the context of machine learning.
    9. How ground truth for the training set was establishedNot applicable. As this device does not involve a "training set" in the context of machine learning, this question does not apply.

    Summary of the Study (as described in the 510(k) Summary):

    The document mentions "Validation testing" was performed to establish the performance characteristics of the modifications made to the Purple/Black Heparin Assay Control compared to the previously marketed device. The core of the submission is to demonstrate "substantial equivalence" to the predicate devices (Purple/Black HMS Control, K922031, and Hepcon® HMS Coagulation System and its Associated Cartridges and Controls, K894317). The primary difference noted is the "source of heparin."

    Conclusion drawn by the submitter: Medtronic demonstrated substantial equivalence based on "design, test results, and indications for use."

    However, no specific details on the "test results" or the criteria against which they were evaluated are provided in this summary. The provided text is a summary seeking FDA clearance based on substantial equivalence, not a detailed study report with specific performance metrics and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981815

    Validate with FDA (Live)

    Date Cleared
    1998-07-30

    (69 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K894317

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEPtrac Electronic Quality Control, manufactured by Medtronic Blood Management, is an interactive mechanical and software controlled verification cartridge for use with instruments that utilize the Hepcon Hemostasis Management cartridges. The purpose of the HEPtrac is to provide a means for performing quality control of the instruments in the clinical testing environment.

    Device Description

    The HEPtrac is an electronic control that has been developed to assist users of the Hepcon Hemostasis Management (HMS) Instrument (K894317) and associated cartridges and controls (K894317), to perform control testing as required under 42CFR493.1253. Sec. ( c ) "for all automated coagulation testing systems, the laboratory must include two levels control each eight hours of operation and each time a change in reagents occurs". The HEPtrac, electronic quality control is designed to minimize the use of liquid based controls by providing the operator a means to determine the operating condition of the instrument.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HEPtrac Electronic Quality Control device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly define quantitative acceptance criteria in the way one might expect for a modern AI/ML device (e.g., target specificity, sensitivity, or AUC). Instead, the performance is demonstrated by its ability to perform specific checks on the HMS instrument. The criteria are implicitly tied to the proper functioning of these checks.

    Acceptance Criteria (Implicit)Reported Device Performance
    Ability to check Flag release force"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 1. Flag release force" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
    Ability to check Reagent delivery"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 2. Reagent delivery" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
    Ability to check Flag lift wire height"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 3. Flag lift wire height" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
    Ability to check Flag sensors"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 4. Flag sensors" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
    Substantial equivalence to the predicate device (ACTtrac) in:The HEPtrac "is very similar to the ACTtrac electronic control for the ACT instrument in the following areas; intended use, operation, and function." The key difference is the number of channels (2 for ACTtrac vs. 6 for HEPtrac) and the addition of a "Flag release test" in HEPtrac. This implies that the device's performance is acceptable because it maintains the core functionalities of the predicate while enhancing capacity.
    - Control Test TypeBoth are "Electronic" (Table 1.1)
    - Reagent delivery checkBoth have "yes" (Table 1.1)
    - Flag Sensor checkBoth have "yes" (Table 1.1)
    - Clot detectionBoth have "yes" (Table 1.1)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a numerical sample size for the test set. It broadly states, "Laboratory testing was done to confirm the devices performance in relation to the predicate device."
    • Data Provenance: The testing was "Laboratory testing," implying it was conducted in a controlled environment. There is no information about the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Given the nature of the device (an electronic quality control for an instrument), the "ground truth" would likely be defined by engineering specifications and the instrument's known functional parameters rather than expert medical interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable given the device's function. The testing described focuses on functional verification against engineering parameters rather than subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an electronic quality control for a medical instrument, not an AI/ML diagnostic or assistive tool for human readers. Therefore, this type of study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself is an "electronic control," which operates without human "interpretation" of results in the traditional sense of a diagnostic algorithm. Its function is to verify the performance of the main instrument. The performance testing "to verify the HEPtrac's ability to check the condition of the HMS instrument" can be considered a standalone performance evaluation of the control device in its intended function. There is no "human-in-the-loop" aspect to the HEPtrac's function itself; it's a verification tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for this device's performance evaluation would be the engineering specifications and known functional parameters of the Hepcon Hemostasis Management (HMS) Instrument. The HEPtrac is designed to test if the HMS instrument is operating within its defined parameters for flag release force, reagent delivery, flag lift wire height, and flag sensors.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable in the context of this device. The HEPtrac is described as an "electronic control" and a "mechanical and software controlled verification cartridge." It does not appear to utilize machine learning or AI that would require a "training set" in the conventional sense. Its function is based on programmed logic and mechanical interactions to test an existing instrument.

    9. How the Ground Truth for the Training Set was Established

    As there is no indication of a "training set" for AI/ML, this question is not applicable. The device's "knowledge" of what constitutes correct instrument operation is built into its design and programming, based on the specifications of the HMS instrument it is designed to test.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1