Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251014

    Validate with FDA (Live)

    Device Name
    APRO 45 Catheter
    Manufacturer
    Date Cleared
    2025-05-01

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K234115

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 45 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 45 Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The APRO 45 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 45 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 45 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 45 Catheter is provided with an Introducer Sheath. The APRO 45 Catheter contains animal derived materials made from tallow derivatives.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the APRO 45 Catheter focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing. It does not present information related to an AI/ML-driven medical device, a multi-reader multi-case (MRMC) study, or acceptance criteria for algorithmic performance.

    Therefore, based solely on the provided text, it's not possible to fulfill most of the requested information points, as they pertain to the evaluation of AI/ML diagnostic or assistive devices. The document describes a physical medical device (catheter) and its non-clinical testing.

    However, I will extract and present the acceptance criteria and performance information that is available from the provided text, primarily from the "Performance Data" section.


    Acceptance Criteria and Device Performance for APRO 45 Catheter

    The APRO 45 Catheter is a physical medical device, not an AI/ML driven diagnostic or assistive device. The acceptance criteria and performance data provided in the FDA 510(k) summary pertain to the physical and biological characteristics of the catheter, supporting its substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.The APRO 45 Catheter met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 45 Catheter met the acceptance criteria.
    Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 45 Catheter met the acceptance criteria.
    Hub Air LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 45 Catheter met the acceptance criteria.
    Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 45 Catheter met the acceptance criteria.
    Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 45 Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 45 Catheter met the acceptance criteria.
    Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 45 Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 45 Catheter met the acceptance criteria.
    Hub and Shaft Tensile StrengthCatheter hub and shaft must meet tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
    Tip Tensile StrengthCatheter tip must meet tip tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
    3-Point BendCatheter 3-Point Bend force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to the predicate.The APRO 45 Catheter met the acceptance criteria.
    Simulated UseWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model and during simulated clot retrieval, the Catheter must meet functionality specifications including compatibility with a stent retriever.The APRO 45 Catheter met the acceptance criteria.
    Corrosion ResistanceCatheter shall show no signs of corrosion when tested in accordance with ISO 10555-1, Annex A.The APRO 45 Catheter met the acceptance criteria.
    Sterilization (Ethylene Oxide)Verified to a sterility assurance level of 1 x 10⁻⁶ in accordance with ISO 11135.Met acceptance criteria.
    Shelf-Life (Aging Studies)Subject device and packaging remain functional for the labeled expiration date; packaging integrity, seal strength, and device functionality met acceptance criteria.Met acceptance criteria. (1-year shelf life established)
    Biocompatibility: SensitizationDid not elicit a sensitization response.Non-sensitizing
    Biocompatibility: Irritation/Intracutaneous ReactivityNo evidence of irritation.Non-irritant
    Biocompatibility: CytotoxicityDid not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
    Biocompatibility: Hemolysis - IndirectNo significant differences between the test article extract and negative control.Non-hemolytic
    Biocompatibility: Hemolysis - DirectNo differences between the hemolytic index of the test article and negative control.Non-hemolytic
    Biocompatibility: Thrombogenicity (PTT)Average clotting time not significantly different from vehicle and negative control.Non-thrombogenic
    Biocompatibility: Thrombogenicity (Platelet Leukocyte Count)Test article performed similar to comparator and negative controls.Met acceptance criteria (performed similar)
    Biocompatibility: Thrombogenicity (Comparative Surface and Geometry Assessment)Similar surface morphology as the predicate device, no differences in roughness or defects.Subject device has similar surface morphology as predicate.
    Biocompatibility: SC5b9 Complement ActivationSc5b9 concentration statistically less than positive control and not statistically higher than negative control.Acceptable
    Biocompatibility: Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity.Non-toxic
    Biocompatibility: Material-Mediated PyrogenicityTotal rise in temperature of < 0.5ºC; determined to be nonpyrogenic.Non-pyrogenic

    Regarding the requested points specifically for AI/ML device evaluations:

    Due to the nature of the device (a physical catheter) described in the provided FDA 510(k) clearance letter and summary, the following information points are not applicable and are not found in the document:

    1. Sample size used for the test set and the data provenance: This relates to data for algorithmic evaluation. Not applicable for a physical catheter.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for a physical catheter.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for a physical catheter.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for a physical device is established through performance testing (bench, animal, biocompatibility) against engineering specifications and regulatory standards.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

    Summary of Study Type:

    The study conducted to support the APRO 45 Catheter's 510(k) clearance was primarily non-clinical bench testing, sterilization validation, shelf-life studies, and biocompatibility testing. Animal testing and human clinical data were noted as "not deemed necessary" for this particular submission based on the substantial equivalence argument to the predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1