Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K251561

    Validate with FDA (Live)

    Device Name
    Biograph Trinion
    Date Cleared
    2025-07-31

    (71 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K233650, K231833, K242300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siemens PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

    The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

    The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and/or PET diagnostic imaging.

    These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

    This system can be used for low dose lung cancer screening in high risk populations.*

    *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    Device Description

    Biograph Trinion PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. This system is designed for whole body oncology, neurology and cardiology examinations. Biograph Trinion PET/CT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

    Biograph Trinion VK20 software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

    Biograph PET/CT systems, which are the subject of this application, are substantially equivalent to the commercially available Biograph Trinion VK10 family of PET/CT systems (K233677). Differences compared to the commercially available Biograph Trinion systems include:

    • The commercially available SOMATOM go.All and go.Top systems with VB10 (K233650) software have been incorporated into the Biograph Trinion VK20 systems, including commercially available CT features.

    • Additional PET axial field of view (FoV) systems allowing for more scalability.

    • Additional patient communication and comfort features.

    • PET respiratory gating with an external gating device has been implemented.

    The Biograph Trinion models may also use the names Biograph Mission, Biograph Wonder, Biograph Ambition and Biograph Devotion for marketing purposes.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Biograph Trinion PET/CT system primarily focuses on demonstrating substantial equivalence to a predicate device and adherence to recognized performance standards. It indicates that "all performance testing met the predetermined acceptance values," but does not provide specific numerical acceptance criteria or reported device performance for an AI/algorithm component, nor does it detail a study proving the device meets AI-specific acceptance criteria. The context suggests the "performance testing" refers to general PET/CT system performance, not AI-driven diagnostic assistance.

    Therefore, many of the requested details, particularly those related to a standalone AI algorithm's performance, human-in-the-loop studies, dataset characteristics (sample size, provenance), and ground truth establishment methods for an AI component, are not available in the provided text.

    Based on the information available in the document, here's what can be extracted and inferred, with explicit notes where information is missing or not applicable in the context of an AI study.


    Acceptance Criteria and Reported Device Performance

    The document states that "all performance testing met the predetermined acceptance values." However, it does not specify what those acceptance values were or the precise reported performance metrics beyond this general statement. The tests conducted were primarily related to the physical performance of the PET/CT system as per NEMA NU 2:2024 and NEMA XR 25:2019 standards, not specifically an AI component for diagnostic aid.

    Table of Acceptance Criteria and Reported Device Performance (Based on available information for the PET/CT system):

    Performance Metric (PET/CT system)Acceptance Criteria (Stated as "predetermined acceptance values")Reported Device Performance
    Spatial ResolutionMet acceptance valuesMet acceptance values
    Scatter Fraction, Count Losses, and RandomsMet acceptance valuesMet acceptance values
    SensitivityMet acceptance valuesMet acceptance values
    Accuracy: Corrections for Count Losses and RandomsMet acceptance valuesMet acceptance values
    Image Quality, Accuracy of CorrectionsMet acceptance valuesMet acceptance values
    Time-of-Flight ResolutionMet acceptance valuesMet acceptance values
    PET-CT Coregistration AccuracyMet acceptance valuesMet acceptance values
    No AI-specific performance metrics detailedNot specified in documentNot specified in document

    Study Details (Focusing on AI-related aspects where applicable, and general system testing otherwise)

    1. Sample size used for the test set and the data provenance:

      • For System Performance (NEMA tests): The document does not specify a "test set" in terms of patient data. NEMA tests typically involve phantom studies rather than patient data. Thus, sample size and data provenance are not applicable in the traditional sense for these tests.
      • For AI Component: The document does not provide any information on a test set (patient cases, images) or data provenance (e.g., country of origin, retrospective/prospective) for validating an AI component for diagnostic assistance. The descriptions are entirely about the physical PET/CT system.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For System Performance: Ground truth for NEMA tests is established by physical measurements and calibration standards, not human experts.
      • For AI Component: This information is not provided in the document as there's no mention of an AI-driven diagnostic aid requiring expert-established ground truth.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • For System Performance: Not applicable.
      • For AI Component: This information is not provided in the document.
    4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • The document does not indicate that an MRMC study was performed for an AI component. The focus is on the substantial equivalence of the PET/CT hardware and software to a predicate device, and compliance with performance standards for the imaging system itself.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The document does not detail any standalone algorithm performance testing. The performance testing described is for the integrated PET/CT system's physical and functional characteristics.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For System Performance: Ground truth for NEMA tests involves physical phantoms and established measurement protocols.
      • For AI Component: This information is not provided in the document.
    7. The sample size for the training set:

      • This information is not provided in the document, as there is no mention of an AI model that undergoes a separate training process requiring a distinct training set.
    8. How the ground truth for the training set was established:

      • This information is not provided in the document, as there is no mention of an AI model's training set.

    Summary of Device and Performance Information from Document:

    The provided 510(k) clearance letter for the Biograph Trinion is for a PET/CT imaging system, not an AI-based diagnostic software. The "performance testing" described in the document pertains to the physical and functional aspects of the PET/CT scanner (e.g., spatial resolution, sensitivity, image quality) as measured against industry standards (NEMA NU 2:2024). The clearance is based on proving substantial equivalence to a predicate device and adherence to these well-established performance standards for imaging hardware.

    Therefore, the detailed questions regarding AI acceptance criteria, AI test set characteristics, human-in-the-loop studies, and AI ground truth establishment are not addressed in this document because the device being cleared is the imaging system itself, not an AI software component for image analysis or diagnostic support. The document implies that the system can be used for certain clinical applications (like lung cancer screening), but it doesn't describe an automated AI system within the device that requires separate clinical validation with reader studies or large patient datasets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251671

    Validate with FDA (Live)

    Date Cleared
    2025-07-03

    (34 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K231833, K223547, K230421

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

    The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

    The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and/or PET diagnostic imaging.

    These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging, and restaging of lesions, tumors, disease, and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders, and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

    This CT system can be used for low dose lung cancer screening in high risk populations. *

    • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365; 395-409) and subsequent literature, for further information.
    Device Description

    The Biograph Vision and Biograph mCT PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole-body oncology, neurology and cardiology examinations. The Biograph Vision and Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

    Biograph Vision and Biograph mCT software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

    The software for the Biograph Vision and Biograph mCT systems, which are the subject of this application, is substantially equivalent to the commercially available Biograph Vision and Biograph mCT software.

    • Somaris Software (cleared in K230421)
      • Upgrade to the latest revision of Somaris Software (Somaris/7 syngo CT VB30) with modified software features:
        • FAST Bolus
        • FAST 4D
        • FAST Applications (FAST Spine, FAST Planning)
        • Automatic Patient Instructions
        • Additional default exam protocols
        • Additional kV setting for Tin Filtration
    • PETsyngo software
      • SMART Image Framer (available for Vision 600 and X models only – cleared in K223547)
    • Updated computer hardware due to obsolescence issues (cleared in K230421). These changes do not affect system performance characteristics and have no impact on safety or effectiveness.

    The Biograph Vision may also use the names Biograph Vision Quantum and Peak for marketing purposes.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) clearance letter for Siemens Biograph Vision and mCT PET/CT Systems, focusing on acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the performance of the updated software (VG85) for the Siemens Biograph Vision and Biograph mCT PET/CT Systems, comparing it to the predicate device (VG80). The "Acceptance Criteria" for the subject device are explicitly stated as "Same" as the predicate device's performance values. This implies that the updated system must perform at least as well as the predicate device across all tested metrics.

    Performance Criteria (NEMA NU2-2018)Predicate Device Acceptance Values (K193248)Reported Device Performance (VG85)Meets Criteria?
    Resolution – Full Size
    Transverse Resolution FWHM @ 1 cm≤ 4.0 mm (Vision) / ≤ 4.7 mm (mCT)SamePass
    Transverse Resolution FWHM @ 10 cm≤ 4.8 mm (Vision) / ≤ 5.4 mm (mCT)SamePass
    Transverse Resolution FWHM @ 20 cm≤ 5.2 mm (Vision) / ≤ 6.3 mm (mCT)SamePass
    Axial Resolution FWHM @ 1 cm≤ 4.3 mm (Vision) / ≤ 4.9 mm (mCT)SamePass
    Axial Resolution FWHM @ 10 cm≤ 5.4 mm (Vision) / ≤ 6.5 mm (mCT)SamePass
    Axial Resolution FWHM @ 20 cm≤ 5.4 mm (Vision) / ≤ 8.8 mm (mCT)SamePass
    Count Rate / Scatter / Sensitivity
    Sensitivity @435 keV LLD≥ 8.0 cps/kBq (Vision 450) ≥ 15.0 cps/kBq (Vision 600) ≥ 5.0 cps/kBq – (mCT 3R) ≥ 9.4 cps/kBq – (mCT 4R)SamePass
    Count Rate peak NECR≥140 kcps @ ≤ 32 kBq/cc (Vision 450) ≥250 kcps @ ≤ 32 kBq/cc (Vision 600 and X) ≥95 kcps @ ≤ 30 kBq/cc (mCT 3R) ≥165 kcps @ ≤ 40 kBq/cc (mCT 4R)SamePass
    Count Rate peak trues≥600 kcps @ ≤ 56 kBq/cc (Vision 450) ≥1100 kcps @ ≤ 56 kBq/cc (Vision 600 and X) ≥350 kcps @ ≤ 46 kBq/cc (mCT 3R) ≥575 kcps @ ≤ 40 kBq/cc (mCT 4R)SamePass
    Scatter Fraction (435 keV LLD)≤43% @ Peak *<40% @ low activity (Vision) ≤40% @ Peak *<37% @ low activity (mCT)SamePass
    Co-Registration Accuracy≤ 5 mmSamePass
    Time of Flight Resolution at 5.3kBq/cc≤ 214 ps (Vision.X) ≤ 249 (Vision 450 and 600) ≤600 (mCT)SamePass
    Mean bias [%] at peak NEC[-6, 6]SamePass
    Image Quality – (% Contrast / Background Variability)
    10mm sphere≥ 55% / ≤ 10% (Vision) ≥ 10% / ≤ 10% (mCT)SamePass
    13mm sphere≥ 60% / ≤ 9% (Vision) ≥ 25% / ≤ 10% (mCT)SamePass
    17mm sphere≥ 65% / ≤ 8% (Vision) ≥ 40% / ≤ 10% (mCT)SamePass
    22mm sphere≥ 70% / ≤ 7% (Vision) ≥ 55% / ≤ 10% (mCT)SamePass
    28mm sphere≥ 75% / ≤ 6% (Vision) ≥ 60% / ≤ 10% (mCT)SamePass
    37mm sphere≥ 80% / ≤ 5% (Vision) ≥ 65% / ≤ 10% (mCT)SamePass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size" in terms of number of patient cases for the test set. Instead, it states:

    • Test Set: The testing was "PET Testing in accordance with NEMA NU2-2018." NEMA NU2-2018 is a standard for performance measurements of PET scanners, which involves phantom studies, not human patient data.
    • Data Provenance: The data provenance is from phantom studies conducted to NEMA NU2-2018 standards on the Siemens Biograph Vision and Biograph mCT systems. This is non-clinical testing. The document does not indicate country of origin or whether it was retrospective/prospective, as it's a controlled laboratory environment with phantoms.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable and therefore not provided in the document. Since the testing was conducted on phantoms according to NEMA NU2-2018 standards, the "ground truth" is defined by the known physical properties and configurations of the phantoms, as well as the standardized measurement protocols. There are no human experts involved in establishing a diagnostic "ground truth" for these types of physical performance measurements.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided in the document. Adjudication methods like "2+1" or "3+1" are used in studies where human readers are interpreting images and there's a need to resolve discrepancies in their evaluations to establish a consensus ground truth. Since this study involved non-clinical phantom testing, such adjudication was not necessary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the effect size of how much human readers improve with AI vs without AI assistance?

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not conducted for this submission." This submission focuses on the hardware and software's physical performance characteristics and its substantial equivalence to a predicate device, not on clinical utility or an AI's impact on human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The document describes the performance of the PET/CT system's hardware and software components. It's a "standalone" performance study in the sense that it measures the inherent physical and technical capabilities of the device (e.g., resolution, sensitivity, image quality) without human diagnostic interpretation being part of the measurement. However, it's critical to clarify that this is not a standalone AI algorithm performance study as typically understood in the context of diagnostic AI. The "Somaris Software" and "PETsyngo software" are operational control and image reconstruction software for the PET/CT system itself, not necessarily an "AI" in the diagnostic sense that would report findings for human review.

    7. The Type of Ground Truth Used

    The ground truth used for this testing was physical phantom data with known properties. The NEMA NU2-2018 standard specifies phantoms with defined sizes, activity concentrations, and geometric configurations, allowing for objective measurement of the scanner's performance parameters (e.g., resolution, sensitivity, image quality spheres).

    8. The Sample Size for the Training Set

    This information is not provided and is likely not applicable in the context of this 510(k) submission. The document describes updates to existing established software and hardware for a medical imaging device. While software development for such systems may involve extensive internal testing and validation, a "training set" with established ground truth as used for machine learning models is not typically a concept applied to the fundamental operational and reconstruction software of a PET/CT scanner being cleared under these circumstances. The emphasis is on meeting performance standards (NEMA NU2-2018) and demonstrating substantial equivalence.

    9. How the Ground Truth for the Training Set was Established

    As stated above, a "training set" in the machine learning sense is likely not applicable. For the development and verification of the system's operational and reconstruction software, the "ground truth" would have been established through methodologies typical for medical device software engineering, including:

    • Reference data/simulations: Utilizing known physical models, simulations, and calibrated reference data to test reconstruction algorithms and system functionality.
    • Engineering specifications: Ensuring the software's output aligns with predefined technical specifications and physics principles governing PET/CT image formation.
    • Phantom studies (internal development): Similar to the NEMA NU2-2018 testing, phantom studies would be used extensively during development to refine algorithms and establish performance characteristics.

    The specific details of the software's internal development and validation, including how any "training" (if applicable for specific algorithmic components, though not explicitly an AI for diagnosis) was performed, are not part of this public 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1