Found 2 results

510(k) Data Aggregation

    K Number
    K231263

    Validate with FDA (Live)

    Date Cleared
    2024-01-05

    (248 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    everX Posterior K153127, Cal LC K212457, Tetric Evoceram K042819, Opallis Flow K201707, Evadyne K111666

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fusion Core DC Flo is indicated for use as core build up material on vital teeth, sealing root canal openings, cavity floor liner, post cementation, restorative material, and as a dentin replacement material.

    Fusion Flo is indicated for direct filling of class V (cervical caries, root erosions, and wedge shaped defects), anterior restorations( Class III and IV), small posterior restorations, repair of composite/ceramic veneers, and blocking out of undercuts.

    Fusion Flo SE is indicated for direct filling of class V (cervical caries, root erosion, wedge shaped defects), anterior restorations (Class III and IV), small posterior restorations,

    Fusion I-Seal is indicated for Class I and Class II restorations, as a liner under composite restorations, and as a pit and fissure sealant.

    Magma NT is indicated for direct restoration in anterior and posterior teeth classes I, III, IV and V, and porcelain/ composite repairs.

    Device Description

    Fusion Core DC Flo is dual-curing, radiopaque, flowable restoration for core build-up composite supplied in an auto-mix delivery system.

    Fusion Flo is a flowable composite of medium viscosity used for fillings in anterior teeth and small fillings in posterior teeth. Fusion Flo can be used separately or with Magma-NT composites.

    Fusion Flo SE is a self-etch, self-adhesive flowable composite. The flowable composite is ready for use and doesn't require the preparatory steps of etching, priming and bonding. Fusion Flo SE can be used separately or with Magma-NT composites.

    Fusion I-Seal is a single component, light cured flowable composite.

    Magma-NT is a visible light cured, radio opaque highly filled hybrid composite restorative material for anterior and posterior restoration of all teeth.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of several dental composite resins from Prevest Denpro Limited (Fusion Core DC Flo, Fusion Flo SE, Fusion I-Seal, and Magna NT) as part of a 510(k) premarket notification to the FDA. The study aims to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting the requirements of ISO 4049:2019 and ISO 13116:2014, with specific values mentioned for some parameters. The reported device performance indicates that "All test results met the criteria in standards."

    Device PropertyAcceptance Criteria (Standard / Value)Reported Device Performance
    Fusion Core DC FloBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥50 MPa), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (maximum of 40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³), Working Time (1:25), Setting Time (3:20).
    Fusion FloBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (1.00 mm opaque / 1.5 mm others minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Fusion Flo SEBased on ISO 4049:2019 & ISO 13116:2014, USP 971 (for pH)Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (1.00 mm opaque / 1.5 mm others minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (<40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Fusion I-SealBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 25 MPa), Depth of Cure (1.5 mm minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Magma NTBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (2.59 mm opaque / 2.89 mm non-opaque), Radio Opacity (3mm compared with aluminum wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Shelf Life (all composites)Testing protocols based on ISO 4049:2019Shelf life of 3 years (except Fusion Core DC Flo: 2 years). Determinations use the same testing protocols and met the criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the test sets for each specific test (e.g., flexural strength, water sorption). It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. The manufacturer, Prevest Denpro Limited, is based in India, so it's possible the testing was conducted there, but this is not confirmed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable to this type of device and study. The "ground truth" for dental composite resins is established through objective physical and chemical testing against international standards (ISO standards), rather than expert interpretation of images or clinical outcomes that require human assessment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the "truth" is determined by objective laboratory measurements against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This study concerns the physical and chemical properties of dental composite resins, not an AI-powered diagnostic device requiring evaluation of human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for evaluating these dental composite resins is defined by international standard specifications for physical and chemical properties as outlined in ISO 4049:2019 (Dentistry - Polymer-based restorative materials) and ISO 13116:2014 (Dentistry - Dental impression materials - Radiopacity). For Fusion Flo SE, USP 971 was also referenced for pH testing. The "truth" is whether the measured properties of the device meet the quantitative limits set by these standards.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI device, there is no training set or ground truth established in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231696

    Validate with FDA (Live)

    Date Cleared
    2023-08-11

    (60 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K133318, K070538, K212457, K060684, K210804, K992985

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fusion Bond 5 is indicated for direct composite or compomer restorations, adhesive cementation, and composite repair.

    Fusion Bond 7 is indicated for bonding of composites to tooth structure, core build up, and adhesive cementation of crown & bridges, including inlays and onlays.

    Fusion Bond DC is indicated for direct light cure composite or compomer restorations, core build up, and adhesive cementation of crown & bridges, including inlays and onlays.

    Renew MDP is indicated for bonding of dual cure, light cure or self cure composite or compomer restorations to tooth structure. treatment of hypersensitive teeth, and intraoral repairs of fractured restorations.

    Renew Universal is indicated for direct bonding of light-cured compomers to tooth structure, bonding of dual-cured core build up composites to tooth structure as long as these materials are light-cured. intraoral repair of composite, metal-based and zirconia /alumina-based restorations, intraoral repair of ceramic restorations in combination with a silane coupling agent, treatment of hypersensitive teeth, and cavity sealing as a pretreatment for indirect restorations.

    Device Description

    Prevest Denpro Tooth Bonding Agents attach different types of restorations to teeth.

    Fusion Bond 5 is a light curing single component bonding agent. It primes and bonds in one step to attach restorations of composites, copomers, amalgams, and porcelains to tooth structures. Tooth structures should have a 37% phosphoric acid etchant application of Fusion Bond 5.

    Fusion Bond 7 is a light curing single component bonding agent. It etches, primes, and bonds in one step to attach restorations of composites, copomers, amalgams, and porcelains to tooth structures.

    Fusion Bond DC is a dual cure bonding direct and indirect restorations. Enamel and dentin surfaces should be conditioned with 37% phosphoric acid etchant before application of Fusion Bond DC. It is a two component adhesive with bonding agent and activator.

    Renew MDP is a light curing single component bondines etching, priming and bonding in one bottle. It is an ethanol-water based dental adhesive containing 10-Methacryloyloxydececyl Dihydrogen Phosphate (10-MDP) a functional monomer which helps in bonding to dentin and cut and un-cut enamel. Renew MDP works with light-cured and dual-cured composite materials.

    Renew Universal is a light curing single component bondines etching, priming and bonding. Renew Universal contains a combination of functional monomers such as 10-MDP and 4-META. Renew Universal works with light-cured and dual-cured composite materials.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a series of dental bonding agents manufactured by Prevest Denpro Limited. The purpose of the 510(k) is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove clinical efficacy through a comparative effectiveness study with human readers and AI assistance. Therefore, many of the requested criteria often associated with AI/ML medical devices are not applicable to this submission.

    Based on the provided text, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria are established by the ISO 29022:2013 standard, specifically for "Dentistry - Adhesive - Notched-edge shear bond strength test." The reported device performance is stated as "Complies with ISO 29022" for both shear bond strength and curing time, with specific values provided as well.

    Device NameAcceptance Criteria (from ISO 29022:2013)Reported Device Performance (Shear Bond Strength)Reported Device Performance (Curing Time)
    Fusion Bond 5Complies with ISO 29022 (Shear Bond Strength)10 MPa20-30 sec
    Complies with ISO 29022 (Curing Time)
    Fusion Bond 7Complies with ISO 29022 (Shear Bond Strength)25 MPa20-30 sec
    Complies with ISO 29022 (Curing Time)
    Fusion Bond DCComplies with ISO 29022 (Shear Bond Strength)19 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)
    Renew MDPComplies with ISO 29022 (Shear Bond Strength)21 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)
    Renew UniversalComplies with ISO 29022 (Shear Bond Strength)21 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)

    Note: The document explicitly states "All test results met the criteria in standard." The specific numeric values (e.g., 10 MPa for Fusion Bond 5) are the reported performance, which is implicitly accepted as meeting the standard's criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for each test. It states "Fusion Bond 5, Fusion Bond 7, Fusion Bond DC, Renew MDP and Renew Universal were tested for appearance, shear bond strength and curing time according to protocols based on ISO 29022:2013."

    The data provenance is from Prevest Denpro Limited, located in Bari Brahmana, Jammu 181133 India. The tests are presented as part of a regulatory submission, implying they were prospective tests conducted specifically for this 510(k).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is a physical testing submission, not a diagnostic imaging AI/ML device. Therefore, the concept of "experts establishing ground truth for a test set" in the context of medical image interpretation (e.g., radiologists) is not applicable. The ground truth is established by the physical properties measured according to an international standard (ISO 29022:2013).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus among human readers for image interpretation. This submission involves physical, objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This is not an AI/ML medical device where human reader performance with and without AI assistance is relevant. This is a dental material.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm or AI component in these dental bonding agents. The "performance" is the physical properties of the material.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the device's performance is based on objective physical measurements conducted according to the ISO 29022:2013 standard for dental adhesives, specifically for shear bond strength and curing time. In essence, the ground truth is the scientific and engineering standard.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of machine learning for this type of device. The products are dental materials, and their formulation and manufacturing are based on established chemical and material science principles, not on iterative training data.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set (in the ML sense), there is no ground truth to establish for it. The "ground truth" for the development of these materials would stem from principles of chemistry, material science, and prior knowledge in dental adhesive development.