K Number
K060684

Validate with FDA (Live)

Date Cleared
2006-04-24

(40 days)

Product Code
Regulation Number
872.3200
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bonding between dentin/enamel and composite filling materials Bonding between dentin/enamel and compomer filling materials Bonding mediator for fissure sealing Desensitization of hypersensitive areas of teeth Bonding between RelyX™ Fiber Post root posts and light-curing composite core build-up materials

Device Description

Adper Prompt/Adper Prompt L-Pop is classified as Resin tooth bonding agent (21 C.F.R. §872.3200) because it is a device intended to be painted on the interior of a prepared cavity of a tooth to improve retention of restorative materials (compomer and composite restorative material). Additionally, Adper Prompt/Adper Prompt L-Pop can be used to scal dentinal tubules of exposed root surfaces to prevent from dentinal hypersensitivity. Furthermore, as recent results show, Adper Prompt L-Pop is suited to bond RelyX™ Fiber Post, glassfiber-reinforced root canal posts by 3M ESPE, to light-curing composite core build-up materials.

AI/ML Overview

The provided text does not contain specific acceptance criteria or a study design to prove that the device meets those acceptance criteria in the format requested. The document is a 510(k) summary for a dental adhesive (Adper Promote/Adper Prompt L-Pop) seeking clearance for a new indication: bonding between RelyX™ Fiber Post root posts and light-curing composite core build-up materials.

Instead of presenting quantitative acceptance criteria and a detailed study report, the document states:

  • "Performance and comparative testing of Adper Prompt/Adper Prompt L-Pop has been carried out."
  • "The results suggest that Adper Prompt/Adper Prompt L-Pop is a suitable agent for bonding between RelyX Fiber Post root posts, by 3M ESPE, and light-curing composite core build-up materials."
  • "The data provided in this 510(k) submission shows that Adper Prompt/Adper Prompt L-Pop is substantially equivalent to the predicate devices."

The document focuses on substantiating substantial equivalence to a predicate device rather than providing a detailed performance study against predefined acceptance criteria for the new indication. Therefore, I cannot extract the information required for the table and bullet points as it is not present in the provided text.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

APR 2 4 2006

Submitter
Company:3M ESPE AG
Street:ESPE Platz
ZIP-Code, City:D-82229 Seefeld
Federal State:Bavaria
Country:Germany
Establishment Registration Number:9611385
Official Correspondent:Dr. Andreas Petermann, Manager Regulatory Affairs
Phone:011-49-8152-700 1395
Fax:011-49-8152-700 1869
E-mail:andreas.petermann@mmm.com
Date:March 09, 2006

Name of Device

Proprietary Name:Adper™ Prompt™
Adper™ Prompt™ L-Pop™
Classification NameResin tooth bonding agent is designated at
21 C.F.R. §872.3200 as a Class II device.
Common Name:Dental Adhesive

Predicate Devices:

Adper™ Prompt™, Adper™ Prompt™ L-Pop™ by 3M ESPE (K040857) Adper™ Single Bond Plus by 3M ESPE (K962785)

{1}------------------------------------------------

Description for the Premarket Notification

Adper Prompt/Adper Prompt L-Pop is classified as Resin tooth bonding agent (21 C.F.R. $872.3200) because it is a device intended to be painted on the interior of a prepared cavity of a tooth to improve retention of restorative materials (compomer and composite restorative material). Additionally, Adper Prompt/Adper Prompt L-Pop can be used to scal dentinal tubules of exposed root surfaces to prevent from dentinal hypersensitivity.

Furthermore, as recent results show, Adper Prompt L-Pop is suited to bond RelyX™ Fiber Post, glassfiber-reinforced root canal posts by 3M ESPE, to light-curing composite core build-up materials. 3M ESPE submits this 510(k) premarket notification to seck clearance for this new indication for use.

Performance and comparative testing of Adper Prompt/Adper Prompt L-Pop has been carried out. The results suggest that Adper Prompt/Adper Prompt L-Pop is a suitable agent for bonding between RelyX Fiber Post root posts, by 3M ESPE, and light-curing composite core build-up materials.

The chemical composition of Adper Prompt/Adper Prompt L-Pop remained unchanged in comparison to 510(k) K040857. Performance data for Adper Prompt/Adper Prompt L-Pop in the indications for use already cleared by 510(k) are, therefore, not subject of this 510(k) submission. The data provided in this 510(k) submission shows that Adper Prompt/Adper Prompt L-Pop is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 4 2006

Dr. Sabine Krischer Regulatory Affairs Specialist 3M ESPE AG ESPE Platz Seefeld, Bavaria D-82229 GERMANY

Re: K060684

Trade/Device Names: Adper Prompt and Adper Prompt L-Pop Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: March 09, 2006 Received: March 15, 2006

Dear Dr. Krischer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for usc stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Dr. Sabine Krischer

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Suze
Chiu S. Lin, PhD

Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

K660684

Device Name: Adper Prompt

Indications For Use:

Bonding between dentin/enamel and composite filling materials Bonding between dentin/enamel and compomer filling materials Bonding mediator for fissure sealing Desensitization of hypersensitive areas of teeth Bonding between RelyX™ Fiber Post root posts and light-curing composite core build-up materials

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

y, General Hospi

L.A. W.S

KILGORE

Page 1 of イ

§ 872.3200 Resin tooth bonding agent.

(a)
Identification. A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.(b)
Classification. Class II.