510(k) Data Aggregation

    K Number
    K220298

    Validate with FDA (Live)

    Device Name
    CerroZone Mobile
    Manufacturer
    Date Cleared
    2022-07-01

    (149 days)

    Product Code
    Regulation Number
    880.5045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K203189

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.

    Device Description

    The CerroZone Mobile is a medical recirculating air cleaner that uses multiple inactivation processes consisting of reactive oxidizing species (ozone) and ultraviolet radiation to inactivate bacteria and viruses in the air. The CerroZone Mobile may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mobile's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 220 cubic feet of air per minute (CFM), or 2.36 air changes per hour (ACH) in a standard 5,600 cu. Ft room. This process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction of airborne particles in a standard 579 cu. Ft room is achieved in 30 minutes or less. A 5.67 LOG, or 99.9998%, reduction is achieved in 45 minutes or less.

    AI/ML Overview

    The CerroZone Mobile device is a medical recirculating air cleaner designed to filter out and inactivate airborne bacteria and viruses. The following information summarizes its acceptance criteria and the studies conducted.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    MS2 bacteriophage aerosolization in a sealed environmental bioaerosol chamberTo evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized MS2 bacteriophage by a combination of entrainment and destructionGreater than 4 log reduction (99.99%)Average net log reduction for MS2 bacteriophage: 5.67 ± 0.19 / 60 minutes
    Methicillin Resistant Staphylococcus epidermidis (MRSE) aerosolization in a single-pass flow-through chamber connected to the CerroZone MobileTo evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized bacteria by a combination of entrainment and destructionGreater than 3 log reduction (99.9%) in single-pass testingDemonstrated effectiveness against Staphylococcus epidermidis
    Ozone EmissionsTo assess ozone emissionsEmissions within the maximum acceptable level of ozone given in 21 CFR 801.415 (0.050 ppm)Emits between 0.000 and 0.010 ppm of Ozone with a measured absolute max of 0.036 ppm. All emissions fall within the acceptable level.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for the biological aerosolization tests (MS2 bacteriophage and MRSE). The studies appear to be laboratory-based non-clinical tests rather than human subject or field trials, so terms like "country of origin" and "retrospective/prospective" might not be strictly applicable in the same way they would be for clinical studies. The context implies these were controlled experiments conducted to assess device performance under specific conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The "ground truth" for these non-clinical tests would have been established by the direct measurement of reduction in viable microorganisms using standard microbiological techniques, rather than expert interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests involved direct measurement of microbial reduction, not subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The CerroZone Mobile is a medical device for air cleaning, not an interpretative diagnostic tool, so such a study would not be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    These were standalone performance tests of the device itself (algorithm not applicable), evaluating its ability to reduce airborne pathogens directly.

    7. The Type of Ground Truth Used:

    The ground truth used for the efficacy studies was based on quantitative laboratory measurements of microbial viability (e.g., plaque assays for bacteriophages, colony counts for bacteria) before and after exposure to the device. For ozone emissions, the ground truth was direct quantitative measurement of ozone concentration.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical air cleaner, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.