510(k) Data Aggregation

    K Number
    K250065

    Validate with FDA (Live)

    Date Cleared
    2025-07-08

    (179 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201685

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The DENSITY, DENZA, DENSITY Noir generates radiofrequency (RF) energy by means of high RF at 6.78MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. In detail, the main body uses the "reverse thermal gradient" principle to deliver RF energy, and thus, heat is generated and selective coagulation occurs while cooling the epidermis, resulting in denaturation and contraction of collagen fibers.

    The devices are an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis.

    It consists of the following components:

    • Electrosurgical Unit - Main body
    • Handpiece
    • electrode tips
    • Neutral electrode pad and neutral electrode pad cable, cleared under K201685
    • Foot switch
    • Power cord
    AI/ML Overview

    This document is a 510(k) clearance letter for an electrosurgical device, not a study report for an AI device. As such, most of the requested information regarding AI device acceptance criteria, study details, and performance metrics (like effect sizes for AI assistance, standalone performance, training sets, etc.) is not present in the provided text.

    The device described, "DENSITY; DENZA; DENSITY Noir," is an electrosurgical cutting and coagulation device. The clearance is based on substantial equivalence to a predicate device (K230663), implying that its performance is expected to be similar to legally marketed devices based on established standards and non-clinical testing.

    Here's a breakdown of the information that can be extracted from the provided text, along with an explanation of why other requested information is not available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for performance that would be found in a typical AI device study (e.g., sensitivity, specificity, AUC thresholds). Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device through adherence to recognized standards and non-clinical performance testing.

    Acceptance Criteria (Implied by Substantial Equivalence and Standards Adherence):

    Criterion TypeDescriptionReported Device Performance/Verification Method
    Safety & Performance StandardsCompliance with recognized national and international safety and performance standards for electrosurgical devices.Evaluated in accordance with: - IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 (Basic safety and essential performance) - IEC 60601-1-2:2014 (Electromagnetic disturbances) - IEC 60601-2-2 (HF surgical equipment particular requirements) - IEC 60601-1-6:2013 (Usability)
    SoftwareSoftware verified and validated to a moderate level of concern, adhering to software lifecycle processes.Evaluated according to IEC 62304:2006.
    Usability/Human FactorsApplication of usability engineering to medical devices.Evaluated in accordance with IEC 62366:2008 based on Human Factor Engineering.
    BiocompatibilityMaterials in contact with patients are biocompatible.Documented in reference to ISO 10993-1:2009, ISO 10993-5:2009, and 10993-10:2010.
    Bench Testing PerformanceDevice operates safely and within predefined design specifications, demonstrating intended function for electrocoagulation and hemostasis.Bench testing conducted to assure safe operation and adherence to design specifications.
    Ex Vivo Thermal TestingPerformance in tissue coagulation and hemostasis, particularly regarding thermal effects.Ex Vivo testing conducted on three types of tissue under GLP Thermal testing in accordance with FDA guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery".
    Risk ManagementRisks identified, analyzed, and controlled.Risk analysis conducted based on ISO 14971:2019.
    Substantial EquivalenceDevice is as safe and effective as a legally marketed predicate device, with no new questions of safety or effectiveness raised by differences.Compared to predicate device K230663. All key parameters (output energy type, user interface, operating frequency, max power, electrode tip types, temperature range, impedance, coolant solution, communication) were found to be substantially equivalent. Minor differences in electrode tip styles and number of tips were deemed not to affect performance or safety.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as this letter describes non-clinical testing for an electrosurgical device, not a clinical study or AI performance evaluation. The "test set" here refers to the materials and conditions used for bench and ex-vivo testing. The document states:

    • "Ex Vivo testing conducted on three types of tissue". No specific sample size (e.g., number of tissue samples) or provenance is mentioned beyond "under GLP Thermal testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. Ground truth for an AI device is typically established by expert interpretation of medical images or data. For an electrosurgical device, performance is evaluated against engineering specifications, recognized safety standards, and physiological responses during ex-vivo testing, not by expert interpretation generating a "ground truth" in the same sense as an AI diagnostic.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods are used in studies involving human interpretation or labeling, especially for AI algorithm ground truth establishment. This is not relevant for the type of non-clinical testing described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. An MRMC study is relevant for evaluating the impact of AI assistance on human performance, typically in diagnostic imaging. The device in question is a standalone electrosurgical instrument and does not involve AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. This device is a manual electrosurgical instrument, not an AI algorithm. Therefore, "standalone algorithm performance" is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical testing:

    • Bench testing: Ground truth is defined by the device's technical specifications and engineering design (e.g., power output within tolerance, correct operating frequency).
    • Ex Vivo testing: Ground truth would be the expected physiological effect (e.g., desired thermal coagulation) confirmed through histological examination or other scientific methods in the tissue samples. This is implied by "Ex Vivo testing conducted... under GLP Thermal testing."

    8. The sample size for the training set

    This information is not applicable/not provided. This device is an electrosurgical instrument, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231287

    Validate with FDA (Live)

    Device Name
    POTENZA
    Date Cleared
    2024-11-05

    (551 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201685

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The POTENZA is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The POTENZA is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis. The POTENZA generates radiofrequency (RF) energy by means of high RF at 1MHz or 2MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. The POTENZA has two operating modes: monopolar mode and bipolar mode. The NEW POTENZA consists of the following components: Electrosurgical Unit - Main body, handpieces (Predicate Device have Motor, AC handpieces), electrode tips for the handpiece – Add three non-invasive tips connect to S handpiece, Neutral electrode pad and neutral electrode pad cable, cleared under K201685, Handpiece stand, Foot switch, Power cord.

    AI/ML Overview

    This 510(k) summary describes a device (POTENZA) that is substantially equivalent to a previously cleared predicate device. The key changes are the addition of a new S handpiece and three new non-invasive electrode tips. The provided text primarily focuses on demonstrating the safety and effectiveness of these additions through non-clinical testing rather than complex AI performance studies. As such, many of the requested points regarding AI acceptance criteria, MRMC studies, and large-scale data sets are not directly applicable or available in this specific document.

    However, based on the provided text, we can describe the acceptance criteria and study that proves the device meets them as they relate to the changes in the device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For the specific changes (new S handpiece and non-invasive tips), the acceptance criteria are generally focused on safety and functionality demonstrating equivalence to the predicate device.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from text)Reported Device Performance (as stated or inferred)
    BiocompatibilityAll four newly added electrode tips must pass biocompatibility tests (Cytotoxicity, Sensitization, Skin Irritation) as per ISO 10993 standards for devices with limited skin contact (<24 hours).Pass: All specified tests (Cytotoxicity, Sensitization: Guinea Pig Maximization Test, Skin Irritation Test) were passed for the non-invasive electrode tips as per ISO 10993-05:2009 and ISO 10993-10:2010.
    Electrical Safety & EMCThe POTENZA system (including new components) must conform to recognized consensus standards for electrical safety and electromagnetic compatibility (EMC), specifically IEC 60601-2-2:2017 and IEC 60601-1-2:2014.Conformity: Electrical safety, EMC, device-related electrical safety for high frequency, and usability were conducted according to IEC 60601-2-2:2017 and IEC 60601-1-2:2014. Implies successful conformance.
    Bench Testing / PerformanceThe device (with new handpiece and tips) must operate safely and within predefined design specifications, including accurate output power (monopolar/bipolar, 1MHz/2MHz), correct frequency, stable power fluctuation, negative output protection, accurate impedance measurement, HO count accuracy, motor speed, and safe failure mechanisms.Conformity: Bench testing was conducted to assure safe operation and conformance to design specifications. Tested parameters included (and implicitly passed): Output accuracy (Monopolar/Bipolar at 1MHz and 2MHz), Frequency, Power fluctuation, Negative output protection, Impedance measurement accuracy/range, HO count accuracy, Safety test of various warnings/failsafe mechanisms, Motor speed level.
    Ex Vivo TestingMust demonstrate appropriate electrocoagulation and hemostasis characteristics on various tissue types (Liver, Kidney, Muscle) under GLP, and thermal testing in accordance with FDA guidance.Conformity: Ex Vivo testing was conducted on three types of tissue (Liver, Kidney and Muscle) under GLP. Thermal testing was done in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014. Implies successful outcomes for these tests.
    Impact on Safety & EffectivenessThe addition of new S handpiece with three electrode tips must not introduce new or different questions of safety or effectiveness.No New Questions: Verification and validation activities confirmed that the additions do not introduce new or different questions of safety or effectiveness. The new tip is explicitly stated as being "safer than the invasive type" due to being non-invasive.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify numerical sample sizes for most performance tests (e.g., number of units tested for electrical safety, number of times functions were tested). For biocompatibility, it refers to standard ISO test protocols, which imply specific sample sizes for each test type (e.g., Guinea Pig Maximization Test typically involves a certain number of animals). For Ex Vivo testing, it mentions "three types of tissue (Liver, Kidney and Muscle)," but not the quantity of each.
    • Data Provenance: The tests are described as non-clinical (bench and ex vivo) and conducted by Jeisys. There is no information regarding country of origin for data samples (as it's not a clinical study involving patients), or if the testing was retrospective or prospective. It is implied to be prospective testing specifically for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document as it pertains to AI/clinical ground truth establishment, which is not the focus of this 510(k) submission for an electrosurgical device modification. The "ground truth" here is based on engineering specifications and recognized consensus standards (e.g., ISO, IEC).

    4. Adjudication Method for the Test Set

    This information is not applicable/provided as it relates to expert review of data (e.g., medical images for AI), which is not detailed for this type of device submission. Performance is based on objective measurements against engineering specifications and standard test protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical performance testing was performed." MRMC studies are typically for evaluating the impact of AI on human reader performance, which is not relevant here as there is no AI component or clinical reader assessment being evaluated.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is an electrosurgical unit, not an algorithm. There is no standalone algorithm performance described.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance validation is based on:

    • Engineering Specifications: Predefined design parameters for power output, frequency, impedance, motor speed, etc.
    • Recognized Consensus Standards: International standards like ISO 10993 (biocompatibility) and IEC 60601 series (electrical safety, EMC).
    • FDA Guidance Documents: Bench and thermal testing were conducted in accordance with FDA guidance for electrosurgical devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set, there is no ground truth establishment for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230663

    Validate with FDA (Live)

    Device Name
    Density
    Date Cleared
    2023-08-14

    (157 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K201685

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The DENSITY generates radiofrequency (RF) energy by means of high RF at 6.78MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. In detail, the main body uses the "reverse thermal gradient" principle to deliver RF energy, and thus, heat is generated and selective coaqulation occurs while cooling the epidermis, resulting in denaturation and contraction of collagen fibers.

    The DENSITY is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis.

    The DENSITY consists of the following components:
    •Electrosurgical Unit - Main body
    •Handpiece
    •6 different electrode tips
    •Neutral electrode pad and neutral electrode pad cable, cleared under K201685
    •Foot switch
    •Power cord

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the medical device "DENSITY", an electrosurgical unit for dermatologic and general surgical procedures. It seeks substantial equivalence to a predicate device, the Thermage FLX System.

    Based on the provided text, the device DENSITY is an electrosurgical cutting and coagulation device. The document does not describe an AI/ML medical device or provide information on acceptance criteria and studies related to AI/ML performance metrics such as accuracy, precision, recall, or human reader improvement with AI assistance. Instead, it focuses on demonstrating substantial equivalence based on technical characteristics and safety standards for an electrosurgical device.

    Therefore, many of the requested items related to AI/ML device performance and testing (e.g., acceptance criteria for AI, sample sizes for AI test sets, expert ground truth, MRMC studies, standalone AI performance) are not applicable or cannot be extracted from this document.

    However, I can extract the information pertinent to the device's substantial equivalence and non-clinical performance data.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance:

    The document does not specify "acceptance criteria" in terms of performance metrics like sensitivity or specificity for a diagnostic/AI device. Instead, "acceptance" appears to be based on demonstrating substantial equivalence to a predicate device through conformity to recognized standards and comparable technical specifications.

    Here's a table summarizing the comparison with the predicate device (Thermage FLX System K170758), which serves as the basis for demonstrating equivalence in performance and safety for this type of device:

    Feature/Acceptance Criteria (Implied Comparability)Proposed Device (DENSITY) Performance/CharacteristicPredicate Device (Thermage FLX System K170758) Performance/Characteristic
    Indications for Use"DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.""The radiofrequency-energy only delivery components of the Thermage FLX System are indicated for use in: • Dermatologic and general surgical procedures for electrocoagulation and hemostasis;"
    Output Energy TypeRadio FrequencyRadio Frequency
    User InterfaceColor Touch PanelColor Touch Panel
    Operating Frequency6.78MHz6.78MHz
    Max Power400 W400W
    Electrode Tip Types3 types (0.25 cm², 4 cm²)4 types (0.25 cm², 3.0 cm², 4 cm², 16cm²)
    Coolant SolutionCryogenCryogen
    Temperature Range65~75°C65~75°C
    RF Time50 ~ 800 ms50 ~ 1000 ms
    Impedance75 - 400 Ω75 - 400 Ω
    Communication to GeneratorLCD touch screenLCD touch screen
    Style of Electrode TipMonopolar, BipolarMonopolar
    Number of Active Electrodes per Applicator- I-Tip(Around eye) I, F-Tip(Face): 1ea- I-Tip II(Around eye), F-Tip II(Face): 2ea- B-Tip(Body), B-Tip(Body)II: 4ea- EYE Tip 0.25: 1ea- TOTAL Tip 3.0: 1ea- NEW TOTAL TIP 4.0: 1ea- BODY TIP 16.0: 4ea
    Coolant Control ParametersType of gas: 1234ze; Method of gas control: Handpiece solenoid control operation; Mouth valve specifications: 12V; Solenoid valve specifications: 1.8V 24.8 BARType of gas: 1234ze; Method of gas control: Handpiece solenoid control operation; Mouth valve specifications: 12V; Solenoid valve specifications: 1.8V 24.8 BAR

    Key takeaway on performance: The document states: "There are no significant differences between DENSITY and the predicate devices (K170758) that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics." "Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing."

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: "No clinical performance testing was performed." The evaluation relies on non-clinical performance data and comparison to the predicate device's established safety and effectiveness.
    • Data Provenance: Not applicable as no clinical test set data was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no clinical test set requiring expert ground truth was performed for this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as no clinical test set requiring adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable as no clinical ground truth was established for this submission. The "ground truth" for demonstrating substantial equivalence is based on established technical standards and the performance of the predicate device.

    8. The sample size for the training set:

    • Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable as this is not an AI/ML device.

    Summary of Device Acceptance (for this traditional medical device):

    The acceptance of the DENSITY device for marketing is based on demonstrating substantial equivalence to an already legally marketed predicate device (Thermage FLX System K170758). This is achieved through:

    • Non-clinical performance data:
      • Evaluation against recognized consensus standards for electrical medical equipment such as IEC 60601-1 (basic safety and essential performance), IEC 60601-1-2 (electromagnetic disturbances), IEC 60601-2-2 (high frequency surgical equipment), IEC 60601-1-6 (usability).
      • Software verification and validation in accordance withIEC 62304 (moderate level of concern) and IEC 62366 (usability engineering/Human Factor Engineering).
      • Biocompatibility testing documented according to ISO 10993 standards.
      • Risk analysis conducted based on ISO 14971.
    • Comparison of technical characteristics: As detailed in the table above, demonstrating analogous specifications (e.g., operating frequency, max power, temperature range, impedance, coolant type) to the predicate device.

    The document explicitly states: "No clinical performance testing was performed." This means the device's safety and effectiveness for its intended use were demonstrated through the non-clinical testing and the comparison to a legally marketed predicate, rather than through clinical trials with human subjects.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210084

    Validate with FDA (Live)

    Device Name
    SILKRO
    Manufacturer
    Date Cleared
    2022-06-03

    (507 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K182355, K201685

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for use in dermatologic and general surgical procedures for electro-coagulation.

    Device Description

    This device is an instrument used for tissue coagulation by means of high-frequency current. It consists of the main unit, Four handpieces, GP Cable, foot switch, LCD touchscreen, ground pad and power cable. The SILKRO has two operating modes: monopolar mode and bipolar mode.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and reported device performance from a clinical study for the SILKRO device. Instead, it focuses on non-clinical tests and a comparison of technological characteristics with predicate devices to demonstrate substantial equivalence for regulatory clearance.

    Specifically, the document lists:

    • Non-clinical tests submitted (Section 7): These are primarily related to electrical safety, electromagnetic compatibility, usability, risk management, software validation, and biocompatibility, conducted according to various IEC and ISO standards.
    • Ex vivo animal testing (Section 7): This was conducted to evaluate the thermal effect of the handpieces on tissue, observing depth and zone of coagulation and thermal damage.
    • Biocompatibility testing (Section 8): Confirmed that materials met biocompatibility requirements for skin contact, following ISO 10993 standards.
    • Sterilization and shelf-life testing (Section 9): Demonstrated a sterility assurance level and a three-year shelf-life through accelerated aging tests.

    Therefore, I cannot provide the requested information for the following points as they are not present in the given text:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample size used for the test set and the data provenance.
    3. Number of experts used to establish the ground truth for the test set and their qualifications.
    4. Adjudication method for the test set.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of how much human readers improve with AI vs without AI assistance.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The document's conclusion (Section 11) states that "Results of performance testing demonstrated substantial equivalence of the subject device to the predicate," but it refers to the non-clinical tests mentioned earlier, not a clinical study involving human patients or a complex diagnostic performance evaluation.