K Number
K182355

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-09-06

(372 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Secret RF Smartcure applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

Device Description

Secret RF's High Frequency(=Radio Frequency) includes the system main body, a handpices with singleuse micro-needle type electrodes, footswitch and an LCD touch screen control panel.

The HF energy is delivered to the target tissue using a handpiece and disposable tip(micro needle electrode tip), the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the HF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating the tissue. Using the micro needle tip, the Secret RF system creates heat within the target dermal tissue via micro-needles inserted from the tip.

The Secret RF is consists of ;

    1. Secret RF main unit (FDA cleared K170325)
    1. Bipolar type handpieces with Bipolar type micro-needle electrodes (FDA cleared K170325)
    1. Secret RF Smartcure applicator

The Secret RF Smartcure Applicator consists of :

  • Smartcure handpiece
  • Monopolar type micro-needle electrodes (MTR-AC-01 , MTR-AC-04, MTR-AC-27G, K3i)
AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Secret RF Smartcure Applicator" and details its substantial equivalence to predicate devices, but it does not contain information on acceptance criteria and a study proving the device meets those criteria in the context of clinical performance or effectiveness data (e.g., sensitivity, specificity, accuracy).

The document focuses on:

  • Regulatory Clearance: Affirming that the device is substantially equivalent to legally marketed predicate devices for dermatologic and general surgical procedures for electro-coagulation and hemostasis.
  • Technical Specifications: Comparing the proposed device's technical characteristics (e.g., frequency, max power, RF duration, electrode type) with a predicate device (SecretRF K170325) and two reference devices (AGNES K160469, EVRF K112334).
  • Performance Data (Non-Clinical/Pre-Clinical): Listing performed tests, which are primarily related to safety, electrical compatibility, and basic functionality:
    • Biocompatibility testing (ISO10993-1;2009)
    • Electrical safety and essential performance (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2: 2009)
    • Animal testing (micropig models for histological data on ablation and thermal damage depth and zone at various post-treatment intervals and intensities).

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance related to clinical outcomes or diagnostic accuracy, as this information is not present. The "performance data" listed refers to design verification and validation testing for safety and basic function, not clinical efficacy or diagnostic performance metrics.
  2. Sample size used for the test set and data provenance: Not mentioned, as no clinical effectiveness study is described.
  3. Number of experts and their qualifications for ground truth: Not applicable, as no ground truth for labeling or diagnosing clinical conditions is discussed.
  4. Adjudication method for the test set: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not conducted or reported. The document explicitly states the device's substantial equivalence is based on similarities to predicate devices and performance testing related to safety and functionality.
  6. Standalone performance (algorithm only): Not applicable, as this is a physical medical device, not an AI algorithm.
  7. Type of ground truth used: Not applicable in the context of clinical performance data. The animal study used histological data as an endpoint for depth and zone of ablation/thermal damage.
  8. Sample size for the training set: Not applicable, as this is not an AI/machine learning device.
  9. How ground truth for the training set was established: Not applicable.

However, based on the provided text, I can infer a type of "acceptance criteria" related to regulatory compliance and safety, and report the "performance" in terms of meeting these standards.

Inferred Acceptance Criteria and Reported Performance (based on regulatory and safety compliance):

Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance / Compliance
BiocompatibilityCompliance with ISO 10993-1:2009Patient contact components and materials are tested and validated.
Electrical Safety & PerformanceCompliance with IEC 60601-1 (General Requirements for basic safety and essential performance)Requirements of specified standards were fulfilled.
Electromagnetic CompatibilityCompliance with IEC 60601-1-2 (General Requirements for basic safety and essential performance - collateral standards: electromagnetic compatibility)Requirements of specified standards were fulfilled.
High Frequency Surgical Equipment SafetyCompliance with IEC 60601-2-2: 2009 (Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories)Requirements of specified standards were fulfilled.
Functional EquivalenceDemonstrating similar intended use, indications for use, and technological characteristics to predicate devices.Device has the same intended use and similar indications as predicate devices. The technology of the predicate devices is also the same. Minimal differences do not raise new safety/effectiveness issues.
Histological ResponseEvaluation of depth and zone of ablation and thermal damage in vivo.In vivo animal testing using micropig models conducted to obtain histological data immediately, 7 days, and 14 days post-treatment at low, mid, and high intensity. (Specific results not detailed in this summary, only that the testing was performed).

Study Description (as reported in the document):

The document references several types of studies performed as part of the 510(k) submission, primarily for demonstrating safety and substantial equivalence, rather than clinical efficacy or diagnostic performance.

  1. Biocompatibility Testing:

    • Description: Testing of patient contact components and materials.
    • Standard: ISO 10993-1:2009.
    • Data Provenance: Not specified, but generally performed in a lab setting.
    • Sample Size: Not specified (refers to materials, not patients).
    • Ground Truth: Adherence to the ISO standard.
  2. Non-Clinical (Engineering/Safety) Testing:

    • Description: Evaluation of the device's electrical safety, essential performance, and electromagnetic compatibility.
    • Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2: 2009.
    • Data Provenance: Not specified, but generally performed in a lab setting.
    • Sample Size: Not specified (typically tests on a device unit).
    • Ground Truth: Compliance with the respective IEC standards.
  3. Animal Testing:

    • Description: In vivo study to obtain histological data on depth and zone of ablation and thermal damage.
    • Species: Micropig models.
    • Purpose: To assess immediate, 7-day, and 14-day post-treatment effects.
    • Data Provenance: Animal study. Country of origin not specified.
    • Sample Size: Not specified (number of micropigs or treatment sites).
    • Ground Truth: Histological analysis of tissue samples.
    • Experts: Not specified, but typically includes pathologists.
    • Adjudication: Not specified.

Missing Information:

Crucially, the document explicitly states this is a substantial equivalence determination for a medical device. For such submissions, extensive clinical efficacy studies with predefined acceptance criteria for patient outcomes (like those you’d expect for an AI diagnostic device's sensitivity/specificity or a pharmaceutical drug's effectiveness) are often NOT required if the device is sufficiently similar to legally marketed predicate devices. The focus is on ensuring basic safety and functional equivalence.

Therefore, the requested details related to clinical performance metrics, ground truth establishment for a diagnostic output, training sets, or human-AI comparative effectiveness studies are not present in this regulatory summary because they were likely not part of the 510(k) submission for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

September 6, 2019

ILOODA Co., Ltd. % Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, Texas 77025

Re: K182355

Trade/Device Name: Secret RF Smartcure Applicator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, OUH Dated: July 29, 2019 Received: August 8, 2019

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182355

Device Name Secret RF Smartcure Applicator

Indications for Use (Describe)

Secret RF Smartcure applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K182355

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: 7/29/2019

I. SUBMITTER

Submitter's Name :ILOODA CO LTD.
Submitter's HQ Address:120, Jangan-ro 458 Beon-gil, Jangan-Gu,
Suwon-Si Gyeonggido, KOREA, 16200
Submitter's Telephone:+82-31- 210-1622
Contact person:Yun-Jung HA (yjha@ilooda.com) / RD Manager
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-583-8988

II. DEVICE NAME

Trade/proprietary name:Secret RF Smartcure Applicator
Common or Usual Name:Micro-needle Fractional RF
Regulation Name:Electrosurgical, cutting & coagulation device & accessories
Regulation Number:21 CFR 878.4400 (Product Code: GEI, OUH)
Regulatory Class:Class II
Prescription Use.

III. PREDICATE DEVICE

Trade/proprietary name:SecretRF
Common or Usual Name:Micro-needle Fractional RF
510K number:K170325
Regulation Name:Electrosurgical, cutting & coagulation device & accessories
Regulation Number:21 CFR 878.4400 (Product Code: GEI, OUH)

{4}------------------------------------------------

K182355

Regulatory Class:Class II
Prescription Use.

IV. REFERENCE DEVICE 1

Trade/proprietary name:AGNES
Common or Usual Name:RF Electrosurgical device
510K number:K160469
Regulation Name:Electrosurgical, cutting & coagulation device & accessories
Regulation Number:21 CFR 878.4400 (Product Code: GEI)21 CFR 878.5350 (Produce Code:KCW)
Regulatory Class:Class II

v. REFERENCE DEVICE 2

Trade/proprietary name:EVRF
510K number:K112334
Common or Usual Name:Electrosurgical, cutting & coagulation device & accessories
Regulation Number:21 CFR 878.4400 (Product Code: ONQ)
Regulatory Class:Class II

VI. DEVICE DESCRIPTION

Secret RF's High Frequency(=Radio Frequency) includes the system main body, a handpices with singleuse micro-needle type electrodes, footswitch and an LCD touch screen control panel.

The HF energy is delivered to the target tissue using a handpiece and disposable tip(micro needle electrode tip), the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the HF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating the tissue. Using the micro needle tip, the Secret RF system creates heat within the target dermal tissue via micro-needles inserted from the tip.

The Secret RF is consists of ;

    1. Secret RF main unit (FDA cleared K170325)
    1. Bipolar type handpieces with Bipolar type micro-needle electrodes (FDA cleared K170325)
    1. Secret RF Smartcure applicator

The Secret RF Smartcure Applicator consists of :

  • Smartcure handpiece
  • Monopolar type micro-needle electrodes

(MTR-AC-01 , MTR-AC-04, MTR-AC-27G, K3i)

{5}------------------------------------------------

VII. INDICATIONS FOR USE

Secret RF Smartcure Applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VIII. DEVICE

Secret RF's Smartcure applicator changes the treatment(operation) mode depending on the electrode to be connected.

Since the specifications for each operation mode are different, we provide an equivalent comparison table for each mode.

Proposed devicePredicate Device
Model nameSecretRFSmartcure Applicator (Model:MTR-AC-01, MTR-AC-04, MTR-AC-27G)SecretRF
ManufacturerILOODA CO.,LTDILOODA CO.,LTD
510(k)numberK182355K170325
Intended useSecretRF Smartcure Applicator isintended for use in dermatologicand general surgical procedures forelectro-coagulation and hemostasis.SecretRF is intended for use indermatologic and general surgicalprocedures for electro-coagulationand hemostasis.Same
Main unitSecretRFSecretRF
Connected handpieceMonopolarBipolar handpieceReferencedevice #1
Output energy typeHigh frequencyHigh frequencySame
Electrode typeMonopolar micro needleBipolar micro needleReferencedevice #2
Frequency2MHz ± 10%2MHz ± 10%Same
Max powerMax 45W at 500ΩMax 25W at 500ΩDifferent
RF duration(ON/OFF RF TIME)Continuous,100 ms ~ 3000 ms50 ms ~ 950 msDifferent
Treatment time10~15min(recommended)10~15min(recommended)Same

Table 1. Technical Comparison for the SC mode

Proposed deviceReference device 1Remark
Model nameSecret RFSmartcure Applicator (Model: MTR-AC-01 , MTR-AC-04, MTR-AC-27GAGENS
ManufacturerILOODA CO.,LTDGowoonsesangCosmetics Co., Ltd.
510(k)numberK182355K160469

{6}------------------------------------------------

Proposed deviceReference device 1Remark
Intended useSecret RF SmartcureApplicator isintended for use in dermatologicand general surgical proceduresfor electro-coagulation andhemostasis.AGNES is indicated for use indermatological and general surgicalprocedures for electrocoagulationand hemostasis.Same
Main unitSecret RFAGENSSame
Connected handpieceMonopolar handpieceMonopolar handpieceSame
Output energy typeHigh frequencyHigh frequencySame
Electrode typeMonopolar micro needle typeMonopolar micro needle typeSame
Frequency2MHz ± 10%1MHz
Max powerMax 45W at 500ΩMax 46 W at 200 Ω
RF duration(ON/OFF RF TIME)Continuous,100 ms ~ 3000 ms50ms ~ 2000ms
Treatment time10~15min(recommended)Unknown
ConnectedelectrodesNeedleLength0.8/1.25/1.5/ 2.0 mm0.8/1.25/1.5/ 2.0 mmSame
Thickness0.2mm0.2mmSame
MaterialsStainless steel 304,Insulating Coating:Teflon PTFEStainless steel 304,Insulating Coating:Teflon PTFESame
Single UseSingle UseSingle UseSame
SterilizationEO gasEO gasSame

Table 2. Technical Comparison for the VC mode

Proposed deviceReference device 2Remark
Model nameSecret RFSmartcure Applicator (Model K3i)EVRF
ManufacturerILOODA CO.,LTDF Care System NV
510(k)numberK182355K130283
Intended useSecret RF Smartcure Applicator isintended for use in dermatologicandgeneral surgical procedures forelectro-coagulation and hemostasis.The EVRF System is intendedfor the epilation and for thetreatment of lower limb spidervein or telangiectasia bythermocoagulation.Same
Output energytypeHigh frequencyHigh frequencySame
Main unitSecret RFEVRFSame
Electrode typeMonopolarMonopolarSame
Frequency4MHz ± 10%4MHz ± 10%Same
Max powerMax 18W at 500ΩMax 25W at 500ΩSame
RF duration(ON/OFF RF TIME)10ms~900ms50 ms ~ 950 msSame
Treatment time10~15min10~15minSame

{7}------------------------------------------------

Proposed deviceReference device 2Remark
(recommended)(recommended)
Connected electrodesCleared FDANeedles are purchased from Ballet
(Recommended K883892)Technologies, Ltd (K883892)

IX. PERFORMANCE DATA

The following performance data was provided in support of the substantial equivalence determination.

Biocompatibility testing:

The patient contact components and materials are tested and validated according to ISO10993-1;2009.

Non Clinical testing:

IEC 60601-1 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance. The requirements of specified standards were fulfilled.

IEC 60601-1-2 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility.

IEC 60601-2-2: 2009 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. The requirements of specified standards were fulfilled.

Animal testing :

In vivo animal testing using micropig models was also conducted to obtain histological data of values for depth and zone of ablation and thermal damage immediately post treatment: 7 days post treatment; and 14 days post treatment.

The treatment was performed at the intensity(power) low, mid, high.

X. CONCLUSIONS

Secret 87 Smartcure applicator with electrode device has the same intended use and similar indications as its predicate devices. The technology of the predicate devices is also the same.

The envelope of power and frequency of the submitted Secret 88 Smartcure applicator with electrodes are covered by the envelopes of its predicate devices.

Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing.

{8}------------------------------------------------

K182355

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Ilooda Co, Ltd. concludes that SECRET RF Smart Applicator is substantially equivalent in comparison SECRET RF (K170325), the predicate device, while it is similar to AGENS and EVRF, the reference devices, as described herein.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.