    Intended Use

    The Rebar™ Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.

    Device Description

    The Rebar" Micro Catheter is an endhole, single-lumen catheter designed to be introduced over a steerable guidewire into the vasculature. The proximal end of the catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter has a semi-rigid proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. Single or dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated to increase lubricity.

    AI/ML Overview

    The provided text does not describe a study involving a device that uses AI or machine learning. Instead, it describes a 510(k) premarket notification for a medical device called the "Rebar™ Micro Catheter." This submission is focused on demonstrating substantial equivalence to an existing predicate device based on material changes (e.g., to the hub material and adhesive) and associated non-clinical bench testing.

    Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and a study proving an AI/ML device meets them, as the provided document pertains to a traditional, non-AI medical device clearance.

    The document discusses:

    • Device: Rebar™ Micro Catheter (a vascular microcatheter)
    • Purpose: Delivery of interventional devices or contrast media into the vasculature.
    • Regulatory Pathway: 510(k) premarket notification, aiming for substantial equivalence to a previously cleared Rebar™ Micro Catheter (K093750).
    • Changes: Primarily changes to the material of the catheter hub (from Polypropylene to Trogamid) and the adhesive used.
    • Proof of Equivalence: Non-clinical bench testing and biocompatibility data, rather than clinical trials or AI/ML performance studies.

    The acceptance criteria listed are for physical and functional properties of the catheter (e.g., hub integrity, pressurization, tensile strength) and not related to AI/ML performance metrics like accuracy, sensitivity, or specificity. There is no mention of a test set, ground truth acquisition involving experts, MRMC studies, or training sets, as these concepts are not applicable to the type of device and study described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202318

    Validate with FDA (Live)

    Date Cleared
    2020-09-14

    (28 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K182101

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Marathon™ Flow Directed Micro Catheter is intended to access the peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic agents such as embolization materials and of diagnostic materials such as contrast media.

    Device Description

    The Marathon™ Flow Directed Micro Catheters are single-lumen, endhole catheters designed for the sub selective infusion of physician-specified therapeutic agents such as embolization materials and diagnostic materials such as contrast media in tortuous, distal vessels. The catheter has a semi-rigid proximal shaft and a highly flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The proximal end of the catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter has a radiopaque marker at the distal end to facilitate fluoroscopic visualization. The outer surfaces of the catheter are coated to increase lubricity. Microcatheter may be used with stylet, guidewire or introducer sheath to increase the rigidity of the distal section during introduction into the guiding catheter.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Marathon™ Flow Directed Micro Catheter. The document focuses on demonstrating substantial equivalence to a predicate device through technical comparisons and bench testing.

    However, the provided text DOES NOT contain information regarding:

    • AI/Algorithm performance: This document is for a physical medical device, not an AI or algorithmic diagnostic tool.
    • Acceptance criteria for an AI model: The acceptance criteria listed are for physical characteristics and performance of a catheter.
    • Sample size for a test set or training set for an AI model: No AI model is being evaluated.
    • Data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone performance: These are all concepts relevant to the evaluation of AI/ML-based medical devices or diagnostic algorithms, which are not the subject of this document.
    • Effect size of human reader improvement with AI assistance: Not applicable.
    • Type of ground truth used (pathology, outcomes data, etc.): Not applicable in this context.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance in the context of an AI/ML model, nor can I answer the specific questions about AI model testing that were asked.

    The document instead focuses on demonstrating the substantial equivalence of the Marathon™ Flow Directed Micro Catheter to a predicate device (K093750) through bench testing and biocompatibility assessments.

    Here's the relevant information about the device's acceptance criteria and proven performance based on the provided text, re-framed to address the questions where applicable, but noting when the question is not relevant to this document.

    1. A table of acceptance criteria and the reported device performance (for the physical device, not AI):

    Bench Testing CategoryAcceptance Criteria (Test Method Summary)Reported Device Performance (Summary of Results)
    Biocompatibility
    CytotoxicityTest article extract should show no evidence of causing cell lysis or toxicity; should be non-cytotoxic.The test article met the requirements of the test since the test article was non-cytotoxic.
    HemocompatibilityTest article should show no evidence of color and be free of particulates; should be non-hemolytic.The test article met the test requirements and was non-hemolytic.
    Performance Data - Bench
    Dimension - Usable LengthMarathon™ Flow Directed Micro Catheter should measure 165 ± 2.5cm.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for usable length.
    Dynamic BurstEvaluated per ISO 10555-1 2014/A1:2017 Annex G.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for dynamic burst.
    Static BurstEvaluated per ISO 10555-1 2014/A1:2017 Annex F.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for static burst.
    Static Burst Post Plug & PushEvaluated per ISO 10555-1 2014/A1:2017 Annex F.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for static burst, post plug and push.
    Hub TensileEvaluated per ISO 10555-1 2013/A1:2017 Annex B.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for hub tensile.
    Hub Tensile Post Plug & PushEvaluated per ISO 10555-1 2014/A1:2017 Annex F.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for hub tensile post plug and push.
    DeadspaceMarathon™ Flow Directed Micro Catheter deadspace should be ≤ 0.27 ml and ≥ 0.23 ml, without Syringe adapter.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for deadspace.
    Visual InspectionMarathon™ Flow Directed Micro Catheter Hub should be clear and free from defects and crazing.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for visual inspection.
    Standard Luer Hub RequirementsEvaluated per ISO 80369-7:2016 and ISO 80369-20:2015.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for standard luer hub requirements.
    Hub Air LeakEvaluated per ISO 10555-1 2014/A1:2017 Annex D.The Marathon™ Flow Directed Micro Catheter met the acceptance criteria for hub air leak.

    2. Sample size used for the test set and the data provenance:

    • This document is for a physical medical device and discusses bench testing, not an AI model's test set.
    • Sample Size for Bench Testing:
      • Specific sample sizes for each bench test are not explicitly stated in this summary table, only the results. However, typical regulatory submissions for physical devices involve testing a statistically representative number of units.
    • Data Provenance: Not applicable in the terms used for AI. The data originates from internal laboratory bench testing conducted by the manufacturer (Micro Therapeutics, Inc. d/b/a ev3 Neurovascular).
    • Retrospective/Prospective: Not applicable in this context. These are lab tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is evaluated through engineering bench tests and biocompatibility testing, not through expert-labeled diagnostic data for an AI model.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is for a physical device, not an AI test set requiring human adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document is about a physical medical device (catheter), not an AI system that assists human readers. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the biocompatibility tests, the "ground truth" is established by standard biological assays and the observation of cellular and blood responses against established safety thresholds.
    • For the bench performance tests, the "ground truth" is the physical properties and performance characteristics of the catheter as measured against predefined engineering specifications and international standards (e.g., ISO 10555-1, ISO 80369). These are objective measurements.

    8. The sample size for the training set:

    • Not applicable. There is no AI model requiring a training set described here.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no AI model or training set described here.