Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K250450

    Validate with FDA (Live)

    Date Cleared
    2025-09-19

    (213 days)

    Product Code
    Regulation Number
    888.3353
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    • Versafitcup CC TRIO

    The Versafitcup CC Trio and the Versafitcup CC Trio No-Hole are designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

    • Quadra-P and Quadra-P Collared

    The hip prosthesis Quadra-P and Quadra-P Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The Quadra-P Cemented is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • AMIStem-H, AMIStem-H Collared, AMIStem-H Proximal Coating, AMIStem-P and AMIStem-P Collared

    The hip prosthesis AMIStem-H, AMIStem-H collared, AMIStem-H Proximal Coating, AMIStem-P and AMIStem-P Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The AMIStem-C is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.

    Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • MiniMAX

    The MiniMax is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • Quadra-H and Quadra-R

    The hip prosthesis Quadra -S, Quadra-H and Quadra-R is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The hip prosthesis Quadra-C is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    The Quadra-C size 0 implant should not be implanted in patients with a mass of 65 kg or greater.

    • SMS and SMS Collared

    The hip prosthesis SMS and SMS Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • Mpres

    The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery.

    Total Hip Arthroplasty with the Mpres stem is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head.
    • Failure of previous hip surgery:
      • Conservative hip surgery,
      • Internal fixation,
      • Arthrodesis,
      • Hip resurfacing replacement.

    Partial hip arthroplasty with the Mpres stem is indicated in the following cases:

    • Acute traumatic fracture of the femoral head.
    • Avascular necrosis of the femoral head.
    • Primary pathology involving the femoral head but with a non-deformed acetabulum.

    • Mpact Acetabular Shell

    The Mpact implants are designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
    Device Description

    The aim of this submission is to seek clearance for the addition of a new coating supplier for the following coatings:

    • Titanium + Hydroxyapatite coating;
    • Hydroxyapatite coating;
    • Porous Titanium coating for Mpact devices; and
    • Titanium coating for Mpact T sizes devices.

    The already FDA cleared devices affected by the change are listed in Table 1 below.

    510kProduct nameSubstrate materialCoating
    K103352 K122911Versafitcup CC TRIOTi-6Al-4V (ASTM F136)Titanium + Hydroxyapatite
    K181254 K192827 K202730Quadra-P Quadra-P Collared Quadra-P Short NeckTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K161635 K173794AMIStem-H Proximal CoatingTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K173794 K192126AMIStem-P AMIStem-P Collared AMIStem-P Short NeckTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K170845 K192352MiniMAXTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K181693 K201673SMSTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K203041SMS CollaredTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K210263MpresTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K093944 K121011AMIStem-H AMIStem-H CollaredTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K082792 K093944 K121011Quadra-H Quadra-H Short NeckTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K082792Quadra-RTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K103721 K122641 K132879 K230011Mpact Acetabular ShellTi-6Al-4V (ASTM F136)Porous Titanium Porous Titanium / Titanium
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K210263

    Validate with FDA (Live)

    Date Cleared
    2021-09-22

    (233 days)

    Product Code
    Regulation Number
    888.3353
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K111046, K201673, K181693, K170845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery.

    Total Hip Arthroplasty with the Mpres stem is indicated in the following cases:
    · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia.
    · Avascular necrosis of the femoral head.

    • · Acute traumatic fracture of the femoral head.
    • Failure of previous hip surgery:
      o Conservative hip surgery,
      o Internal fixation,
      o Arthrodesis,
      o Hip resurfacing replacement.

    Partial hip arthroplasty with the Mpres stem is indicated in the following cases:

    • · Acute traumatic fracture of the femoral head.
    • · Avascular necrosis of the femoral head.
      · Primary pathology involving the femoral head but with a non-deformed acetabulum.
    Device Description

    The Mpres Neck Preserving Stem is a cementless short femoral stem intended to be used in total or partial hip arthroplasty for primary or revision surgery with minimally invasive hip replacement techniques. The Mpres Neck Preserving Stem is made of titanium alloy (Ti6Al7Nb) according to ISO 5832-11 and coated with Titanium plasma spray according to ASTM F1580 and Hydroxyapatite in compliance with ASTM F1185. It is available in 12 sizes (sizes 3-14) for each neck variation configuration: a standard 130° CCD angle or an high offset 123° CCD angle.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Mpres Neck Preserving Stem" hip prosthesis. This type of document is generally about establishing substantial equivalence to an existing predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria in the way a clinical trial for a new drug or a novel AI diagnostic would.

    Therefore, the prompt's request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical performance metrics like sensitivity, specificity, or reader improvement with AI assistance, does not directly apply to this document's content.

    The document focuses on demonstrating that the Mpres Neck Preserving Stem is substantially equivalent to legally marketed predicate devices based on technological characteristics and non-clinical performance data. It explicitly states that "No clinical studies were conducted."

    However, I can extract information related to the performance data and the types of studies that were conducted to support the substantial equivalence claim.

    Here's an adaptation of the requested information based on the provided text, highlighting what is available and what is not:

    1. A table of acceptance criteria and the reported device performance

    Since this is a substantial equivalence submission for a mechanical implant, performance is demonstrated through biomechanical testing and material characterization against established standards, rather than clinical performance metrics typically associated with diagnostic devices. The acceptance criteria are implicit in meeting the requirements of the cited ISO/ASTM standards.

    Performance CharacteristicAcceptance Criteria (Implicit from Standards)Reported Device Performance
    Design ValidationAdequate performance in simulated use (e.g., sawbone, cadaveric evaluation)Mpres Sawbone Design validation report, Plastic Stem Impactor, cadaveric evaluation completed.
    Pull-Off TestMeet requirements of ASTM F2009-00:2005, ISO 5832-9:2007, ISO 5832-12:1996, ISO 7206-10:2003Test Report No.: 167.090722.10.1309 (EndoLab) and CeramTec AG Test Reports 3128, 3300.
    Range of Motion (ROM)Meet requirements of EN ISO 21535:2009Mpres evaluation of the ROM completed.
    Neck and Shaft FatigueMeet requirements of ISO 7206-6:2013, ISO 7206-4:2010, ASTM F2068-15, ASTM F2996-13, ISO 5832-11EndoLab Test Report 167 201005 10 3284-part1-rev0, Accentus Test Report OTC357, Straight Cantilever Test Report.
    Coating ValidationMeet requirements of ISO 5832-11Mpres Coating validation rationale completed.
    Fretting CorrosionMeet requirements of ISO 5832-11 and ISO 21534Mpres femoral stems Fretting Corrosion Rationale completed.
    PyrogenicityMeet requirements of European Pharmacopoeia §2.6.14 (LAL test) and USP chapter <151>Bacterial endotoxin test (LAL test) and Pyrogen test completed. Devices are not labeled as non-pyrogenic.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated in terms of specific numbers of devices tested for each non-clinical study. The reports cited (e.g., "Endolab Test Report," "CeramTec AG Test Reports," "Accentus Test Report") would contain this detail, but it's not present in this summary. For "cadaveric evaluation," the number of cadavers is not given.
    • Data Provenance: The testing appears to have been conducted by various external and internal labs (e.g., EndoLab, CeramTec AG, Accentus). The company, Medacta International SA, is based in Switzerland, and Medacta USA is in Tennessee, USA. The standards cited are international (ISO, ASTM, EN) and US (USP, European Pharmacopoeia). This indicates a varied provenance tied to international standards for medical device testing.
    • Retrospective or Prospective: These non-clinical performance tests are "prospective" in the sense that they are specifically performed for the purpose of this submission and validate the current design. They are not observational studies on existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This concept does not apply to this type of device submission. There is no "ground truth" established by experts in the context of diagnosing or interpreting images, as this is a hip implant, not an AI diagnostic device. The "ground truth" for these tests is the adherence to established engineering and material science standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This concept does not apply. Adjudication methods are relevant for clinical studies or studies involving expert interpretation, which were not performed here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This is designed for AI diagnostic aids, not for a hip implant. The document explicitly states: "No clinical studies were conducted."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept does not apply. A hip implant is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" here is the performance against established engineering and material science standards (e.g., ISO, ASTM, EN, USP). For example, a fatigue test must demonstrate that the device can withstand a certain number of cycles at a specific load without failure, as defined by the standard.

    8. The sample size for the training set

    This concept does not apply. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    This concept does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201673

    Validate with FDA (Live)

    Date Cleared
    2020-09-18

    (91 days)

    Product Code
    Regulation Number
    888.3353
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

    · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

    · Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.

    Device Description

    The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery.

    The SMS implants are comprised of the following products:

    • SMS Cementless Solid Standard Stem (available in 11 sizes from size 3 to 13); and ●
    • . SMS Cementless Solid Lateralized Stem (available in 11 sizes - from size 3 to 13).

    Both are available on the US market via the clearance - K181693.

    Concerning the new sizes of the solid version: 1V2, 2V2, size 14, and size 15; both STD and LAT versions have been introduced as a product range extension to the stems currently marketed as the SMS product line (K181693).

    The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), MiniMAX (K170845), and SMS (K181693).

    The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).

    The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).

    AI/ML Overview

    This is a 510(k) premarket notification for the "SMS Cementless Stem" (K201673), a hip prosthesis. The document details the device, its intended use, and a comparison to predicate devices, along with performance data.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document outlines various non-clinical performance tests conducted for the SMS Cementless Stem. However, it does not provide specific acceptance criteria values or the quantitative results of these tests in a table format. It only states that the testing was conducted "according to written protocols with acceptance criteria that were based on standards."

    Here's what can be inferred for a table, but actual criteria values and results are missing in the document:

    Test (Standard)Acceptance Criteria (Not provided explicitly, generally "meet standard requirements")Reported Device Performance (Not provided explicitly, implied "met criteria")
    Range of Motion (EN ISO 21535:2009)Based on standard EN ISO 21535:2009Conducted
    Fatigue Testing - Stemmed Femoral Components (ISO 7206-4:2010 + AME 1:2016)Based on standard ISO 7206-4Conducted
    Fatigue Testing - Head and Neck Region of Stemmed Femoral Components (ISO 7206-6:2013)Based on standard ISO 7206-6Conducted
    Pull-off Force Testing - Taper Connections (ASTM F2009-00 (Reapproved 2011))Based on standard ASTM F2009-00Conducted
    Coating Tests - Hydroxyapatite (ISO 13779-1:2008)Based on standard ISO 13779-1Conducted
    Coating Tests - Tension Testing of Calcium Phosphate and Metal Coatings (ASTM F1147-99)Based on standard ASTM F1147-99Conducted
    Pyrogenicity - Bacterial Endotoxin Test (European Pharmacopoeia §2.6.14 / USP chapter <85>)Based on European Pharmacopoeia §2.6.14 / USP chapter <85>Conducted
    Pyrogenicity - Pyrogen Test (USP chapter <151>)Based on USP chapter <151>Conducted

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical studies were conducted." Therefore, there is no test set in the context of human clinical data or patient outcomes. The "performance data" refers to non-clinical mechanical and material tests. The sample sizes for these mechanical tests are not provided in this summary. Data provenance is not applicable here as no human data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable as no clinical studies were conducted and no ground truth in the clinical sense was established. The "ground truth" for the non-clinical tests would be the specifications and requirements of the referenced ISO and ASTM standards.

    4. Adjudication Method for the Test Set

    Not applicable as no clinical studies were conducted which would require adjudication of expert opinions or outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    No MRMC study was mentioned or performed, as "No clinical studies were conducted." This device is a mechanical implant, not an AI or diagnostic tool that would typically undergo MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hip implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the non-clinical performance and coating tests, the "ground truth" implicitly refers to the
    requirements and specifications outlined in the referenced international standards (ISO and ASTM) and the European Pharmacopoeia/USP chapters. The device is expected to meet these pre-defined engineering and material standards. This is a form of engineering/material specification compliance as ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This device is a mechanical implant, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of medical device's approval.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1