Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242815

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-17

    (90 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K083302, K180865

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    POCKET TENS (EP-300) which is home used device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.

    Device Description

    This device powered by CR2032 Lithium 3V battery is a self-adhesive TENS with 15 adjustable intensity levels falling into the electro-stimulation device category. It includes pocket storage rack for electrodes and TENS device. It designed for shoulder, back, upper extremities (arm) and lower extremities (leg) pain relief. Users only need to place the electrode gel pads on the sides of pain point and then power on the device, and adjust the pulse intensity to a proper level to start a 20 minute treatment session. For optimum result, rest 30 minutes between sessions.

    TENS refers to the electrical stimulation of nerves through the skin which is a non-pharmacological method of pain relief. Any symptoms that could be relieved using TENS must be checked by patient's general practitioner who will also give patient instruction on how to carry out a TENS self-treatment regime.

    This device is only compatible with the 75 mm x 48 mm gel pads which are the OTC medical device cleared by FDA under K083302 or K180865, and come with the device.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the POCKET TENS (EP-300) device.

    Here's the information extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document states that the POCKET TENS (EP-300) device underwent a series of safety and performance tests. It doesn't explicitly list specific quantitative acceptance criteria values next to the device performance for most parameters, but generally states that "All the test results demonstrate the subject device, POCKET TENS (EP-300), meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate devices."

    However, a comparison table (Section 12) between the subject device and two predicate devices provides various specifications. This comparison indirectly implies the acceptance criteria are met if the subject device's specifications are comparable or within acceptable ranges of the predicate devices and comply with relevant standards.

    Here's a summary of the stated performance and comparison points, implying acceptance criteria are met by equivalence or compliance:

    Acceptance Criteria (Implied)Reported Device Performance (POCKET TENS (EP-300))
    Intended UseTemporary relief of pain associated with sore and aching muscles in the shoulder, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
    FDA product codeNUH
    Prescription or OTCOTC (Identical to predicate devices)
    Power source(s)CR2032 Lithium 3V (Identical to Predicate device #2)
    Function and designElectrical stimulation (Identical to predicate devices)
    Number of output modes (TENS)1 (Identical to Predicate device #2)
    Number of output channelsSingle Channel (Identical to Predicate device #2)
    Method of channel isolationN/A (Identical to Predicate device #2)
    Regulated current or regulated voltageRegulated voltage (Identical to predicate devices)
    Software/Firmware/Microprocessor Control?Yes (Identical to predicate devices)
    Automatic overload trip?Yes (Identical to predicate devices)
    Automatic no-load trip?Yes (Identical to predicate devices)
    Automatic shut off?Yes (Identical to predicate devices)
    Patient override control?Yes (Identical to predicate devices)
    Indicator display (On/off status, Low battery, Voltage/current level)No (Identical to Predicate device #2 for all three)
    Timer range (minutes)20 (Identical to Predicate device #2)
    Compliance with voluntary standardsISO 10993-1, 10993-5, 10993-10, 10993-23, IEC 62304, ISO 14971, ISO TR 24971, IEC 60601-1-2, IEC TR 60601-4-2, ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-2-10, IEC 60086-4, IEC 60601-1-6, ANSI/AAMI/IEC 62366-1 (Complies with the latest standards)
    Compliance with 21 CFR 898?Yes (Identical to predicate devices)
    WeightMain device (including a set of pad): 35.6 g; Gel pads (2 pieces): 9.4 g; Battery: 3.0 g
    DimensionsMain device: 66.8 × 35.0 × 10.6 mm; Pad: 99.5 × 62.0 × 1.0 mm; Wire: 391.0 mm
    Housing materials and constructionABS plastic enclosure (Identical to predicate devices)
    WaveformBiphasic, Symmetrical (Identical to Predicate device #1)
    ShapeRectangular (Identical to predicate devices)
    Maximum output voltage (@ 500 Ω, 2 kΩ, 10 kΩ)50V for both phases, peak of 25V for each phase; 92V; 101V (Refer to Section 13 for substantial equivalence)
    Maximum output current (@ 500 Ω, 2 kΩ, 10 kΩ)100mA peak-to-peak, 50mA peak; 37.5mA; 10.1mA (Refer to Section 13 for substantial equivalence)
    Pulse Width (μs)50mA x 0.4mSec = 20 μC (Refer to Section 13 for substantial equivalence)
    Frequency (Hz)4, 6 and 50 (Refer to Section 13 for substantial equivalence)
    Net charge (μC per pulse @ 500Ω)0 (Refer to Section 13 for substantial equivalence)
    Maximum phase charge (μC @ 500Ω)20 (Refer to Section 13 for substantial equivalence)
    Maximum average current (mA @ 500Ω)3.46 (Refer to Section 13 for substantial equivalence)
    Electrode conductive surface area (cm²)31.04 cm² × 2 (=62.08 cm²) (Refer to Section 13 for substantial equivalence)
    Maximum current density (mA/cm² @ 500Ω)1.61080 (Refer to Section 13 for substantial equivalence)
    Maximum power density (W/cm² @ 500Ω)0.0801 (Refer to Section 13 for substantial equivalence)
    Burst Mode (Pulse per burst, Burst per second, Burst duration, Duty cycle, Average DC current)N/A (Identical to Predicate device #1 where applicable, otherwise N/A for those parameters)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly mention a "test set" in the context of clinical data. It states, "No clinical test data was used to support the decision of substantial equivalence."

    Instead, the evaluation appears to be based on non-clinical testing, compliance with recognized standards, and comparison with predicate devices. For these non-clinical tests (e.g., biocompatibility, software validation, EMC, electrical safety, performance verification, usability), the sample sizes are not specified, nor is the country of origin or whether they were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical test data was used and therefore no need for experts to establish ground truth for a clinical test set. The substantial equivalence relies on technical comparison and compliance with standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test data was used and therefore no adjudication was necessary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief, not an AI-assisted diagnostic or interpretative tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical TENS unit, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical test data was used. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles outlined in the referenced national and international standards (e.g., ISO, IEC, ANSI/AAMI standards, FDA guidance documents).

    8. The sample size for the training set

    Not applicable, as the device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as the device is not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220997

    Validate with FDA (Live)

    Date Cleared
    2022-09-01

    (150 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180865

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TENS:

    [RS-18, RS-28, RS-38 models]: For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.

    [RS-18, RS-38 models]: For temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medications.

    [RS-18, RS-28 models]: It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.

    EMS:

    [RS-18. RS-28]: To stimulate healthy muscles in order to improve and facilitate muscle performance.

    Environments of Use: Clinics, hospital and home environments (over-the-counter) Patient Population: Adult

    Device Description

    "Wireless Electrical Stimulator" is a battery-powered wireless electrical stimulator designed and intended to provide Transcutaneous Electrical Nerve Stimulation (TENS) and Electrical Muscular Stimulation (EMS) for the use of non-pharmacological pain relief and improve muscle performance and recovery.

    The "Wireless Electrical Stimulator" includes a remote control, stimulator pod(s) and several accessories. Both the remote control and the stimulator pod are powered by a 3.7V lithium-ion polymer battery. The remote control has a 1.77" TFT LCD screen and 4 buttons to control multiple stimulator pod(s) at the same time. Each stimulator pod has two output channels, and will perform electrical stimulation according to the parameter settings sent by the remote control. The stimulator pods can also be used independently via its own 3 buttons based on preset parameter.

    The accessories include self-adhesive electrodes, kneelelbow brace, and the USB Type-C charging cable. The adhesive electrodes for "Wireless Electrical Stimulator" are connected with the Stimulator Pod(s) via magnetic snap and lead wire. There are replaceable hydrogel pads on the self-adhesive electrode pads. Users can replace the hydrogel when it exceeds its useful life, but does not need to discard the entire electrode. To apply electrical stimulation to body joints (e.g. knees and elbows), users can easily place electrodes on specific body parts using the Knee/elbow brace. The remote control and stimulator pods are the same for all models. The 3 models have different accessories to match their specific usage instructions.

    AI/ML Overview

    The provided text is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) / Electrical Muscle Stimulator (EMS) device. It details the device's characteristics, comparison to predicates, and performance testing for FDA clearance.

    However, the provided text DOES NOT contain the type of AI/ML-specific acceptance criteria and study information requested in the prompt. The device is a medical stimulator, not an AI/ML-driven diagnostic or prognostic tool. Therefore, information like "multi-reader multi-case (MRMC) comparative effectiveness study," "standalone (i.e. algorithm only without human-in-the-loop performance)," "ground truth established by experts/pathology," or sample sizes for training/test sets for AI models are not applicable to this document.

    The document focuses on demonstrating substantial equivalence to legally marketed predicate devices through comparisons of indications for use, technological characteristics, and various non-clinical performance data (biocompatibility, electrical safety, EMC, software verification, human factors, bench testing).

    Therefore, I cannot populate the requested table or answer the specific questions related to AI/ML acceptance criteria and studies based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213802

    Validate with FDA (Live)

    Date Cleared
    2022-08-26

    (263 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180865, K180943

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StimOn™ Pain Relief System (GM2439) is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain.

    Device Description

    StimOn™ Pain Relief System (GM2439) is a hand-held stimulator designed to pass electrical signals via surface electrodes through the skin to the underlying nerves. It may be used as a TENS device to aid the blocking of pain signals traveling to the brain.

    AI/ML Overview

    The document describes the StimOn™ Pain Relief System (GM2439) and its substantial equivalence to predicate devices, but it does not contain specific acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) for an AI/algorithm-based device, nor does it detail a study proving the device meets such criteria in terms of AI performance.

    Instead, the document focuses on the regulatory clearance of a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief, primarily through comparison with existing predicate devices and compliance with relevant electrical safety and performance standards. The "performance testing" mentioned refers to these safety and electrical performance standards, not to clinical efficacy measured by AI performance metrics.

    Therefore, many of the requested items regarding acceptance criteria and study details for an AI-based device's performance cannot be extracted from this document.

    Here's the information that can be extracted, and where the document does not provide the requested information:

    1. A table of acceptance criteria and the reported device performance:

    The document does not specify quantified acceptance criteria for AI performance (e.g., sensitivity > X%, specificity > Y%). Instead, it discusses the device's technical characteristics and how they compare to predicate devices. The "performance testing" section refers to compliance with electrical safety and performance standards, not AI accuracy.

    CharacteristicAcceptance Criteria (Not explicitly stated for AI performance)Reported Device Performance (from "Table 1. Comparison table" and subsequent descriptions)
    Intended UseSame as predicate devicesSymptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical or post-traumatic pain.
    Product CodeGZJGZJ
    Power SupplyNot explicitly stated as acceptance criteria, but compared to predicatesOne 3.7V, 125mAh rechargeable Lithium polymer battery
    Treatment Timer MaximumNot explicitly stated as acceptance criteria, but compared to predicates15 minutes
    Number of Output ModesNot explicitly stated as acceptance criteria1 (RF Mode)
    Number of WaveformsNot explicitly stated as acceptance criteria1
    WaveformNot explicitly stated as acceptance criteriaBiphasic (Sine Wave)
    RF Stimulation ModeNot explicitly stated as acceptance criteriaTranscutaneous Pulsed Radio-frequency (PRF) mode
    RF FrequencyNot explicitly stated as acceptance criteria500 kHz fixed
    Burst TimeNot explicitly stated as acceptance criteria25ms
    Burst FrequencyNot explicitly stated as acceptance criteria2Hz
    Electrode ContactNot explicitly stated as acceptance criteriaHydrogel pad
    Load ImpedanceNot explicitly stated as acceptance criteria, but stated as a shutdown threshold1 k Ohm, Over this load, the system will automatically shut off.
    Maximum Output VoltageNot explicitly stated as acceptance criteria7.4 V+/-10% (6.6 V +/- 20% @@500Ω with Electrode Pad)
    Maximum Output CurrentNot explicitly stated as acceptance criteria14.8 mA (Peak)@500Ω (13.2 mA (Peak)@500Ω with Electrode Pad)
    Maximum Average Current (Over primary phase)Not explicitly stated as acceptance criteria9.43 mA (Peak)@500Ω (8.41mA(Peak)@500Ω with Electrode Pad)
    Maximum Average Power Density (Smallest electrode area) “P”Must be within value range of predicates112.81 µW/mm² @500Ohm with Electrode
    Maximum Current (RMS) Density (mA/mm²)Not explicitly stated as acceptance criteria0.0242 mA /mm² @500Ω with Electrode
    Net ChargeNot explicitly stated as acceptance criteriaThe theoretical value is 0.
    Maximum average phase chargeNot explicitly stated as acceptance criteria0.00942μC
    Electrical Safety & PerformanceCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ANSI/AAMI NS4Device underwent testing and demonstrated compliance (implied by substantial equivalence).
    BiocompatibilityMeet standards for electrodes leveraged from K180865Leveraged from cleared electrodes (K180865).
    UsabilityCompliance with IEC 62366: Medical devices Part 1: Application of usability engineering to medical devicesDevice complies with IEC 62366 and is similar to K180943 (BioWaveGO) cleared for OTC use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided. The document states "No Human Clinical Performance Testing have been required" for this application. The performance testing refers to electrical and safety standards, not clinical performance data from a test set of patients/cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. No clinical test set requiring expert ground truth was used for assessing the device's efficacy, as per the submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable / Not provided. This device is a Transcutaneous Electrical Nerve Stimulator, not an AI-assisted diagnostic or interpretative device that augments human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable / Not provided. This device is hardware for pain relief, not an algorithm/AI system in the sense of a diagnostic or predictive tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable / Not provided. No clinical ground truth was established for "performance" as it relates to AI or clinical efficacy. The "performance testing" was for electrical safety, electromagnetic compatibility, and biocompatibility, which rely on technical standards and laboratory measurements.

    8. The sample size for the training set

    • Not applicable / Not provided. Not an AI/algorithm-based device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. Not an AI/algorithm-based device requiring a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1