Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K243529

    Validate with FDA (Live)

    Date Cleared
    2025-03-14

    (120 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the administration of fluids from a container into the patient's vascular system through a vascular access device.

    Device Description

    The proposed devices consist of Solution Administration Sets. These devices include Basic, Secondary, CONTINU-FLO solution sets, Stand-Alone devices and Chemotherapy devices (see Table 2 for a list of subject device set names per product family). They are single use disposable, non-pyrogenic, sterile devices intended for the administration of fluids from a container into the patient's vascular system.

    AI/ML Overview

    This is a 510(k) premarket notification for "Solution Administration Sets" by Baxter Healthcare Corporation. The document states that the devices are substantially equivalent to a predicate device (K203609 cleared on September 30, 2021).

    Here's the breakdown of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a specific table of acceptance criteria with corresponding performance values in the format usually seen for AI/ML devices. Instead, it describes general conformance to recognized standards and the positive outcomes of various tests.

    Acceptance Criteria (Standard / Test)Reported Device Performance
    ISO 80369-7: 2021 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications)All proposed devices were found to be in conformance with this standard. Performance testing included mechanical (tensile strength), pressure (burst, leakage, backflow, internal), stress cracking, resistance, torque, spike insertion/removal force, drop form accuracy, vacuum, pump/set integrity, DEHP content.
    ISO 8536-4 (Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed)Complete sets meet the performance requirements of this standard (mentioned in relation to priming volume and tubing types).
    ISO 10993-1 (Biological Evaluation of Medical Devices) / FDA-2013-D-0350 GuidanceThe proposed devices are biocompatible and appropriate for their intended use. Biocompatibility tests conducted: Cytotoxicity, Sensitization, Intracutaneous (Irritation) Reactivity, Acute Systemic Toxicity, 30 Day Systemic Repeat Dose Toxicity Study, Material Mediated Pyrogen, Hemolysis. All met acceptance criteria.
    USP <788> Particulate Matter in InjectionsFilter performance testing included particulate retention, integrity, air filter flow. Particulate matter testing met the USP Acceptance criteria.
    Microbial Ingress Testing (Baxter's testing strategy, per K223175)All test results met their acceptance criteria, demonstrating the absence of microbial ingress into the sterile fluid path during simulated clinical use, supporting appropriate design for intended use.
    ISO 11137-1: 2006 (Sterilization of health care products - Radiation - Part 1)Sterilization process established per this standard. Devices sterilized via radiation with a minimum Sterility Assurance Level (SAL) of 10-6.
    ISO 11137-2: 2013 (Sterilization of health care products - Radiation - Part 2)Minimum Sterilizing Dose (MSD) established and validated as per Method 1. Continued validity confirmed via periodic dose audit studies.
    ISO 11607-1: 2019 (Packaging for terminally sterilized medical devices - Part 1)Package verification testing performed per this standard (Simulated Distribution per ASTM D4169-22) and included visual (ASTM F1886), seal strength (ASTM F88), and bubble test (ASTM F2096-11). All met requirements.
    ASTM F1980-21 (Accelerated Aging of Sterile Barrier Systems and Medical Devices)2-year shelf-life confirmed via accelerated aging.

    Note: This submission is for a traditional medical device (solution administration sets), not an AI/ML device. Therefore, the questions related to AI/ML specific studies (sample size for test set, data provenance, number of experts for ground truth, adjudication, MRMC study, standalone performance, training set sample size, training set ground truth) are not applicable to this document. The "tests" described are standard engineering, biocompatibility, and sterilization validations for physical medical devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as this is a traditional medical device, not an AI/ML device. The testing described is bench testing and biocompatibility assessments, not a study involving patient data or a specific test set in the AI/ML context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is a traditional medical device, not an AI/ML device. Ground truth as typically defined for AI/ML models is not relevant here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is a traditional medical device, not an AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is a traditional medical device, not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a traditional medical device, not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as this is a traditional medical device, not an AI/ML device. Instead of "ground truth," the device relies on conformance to established international and national standards (ISO, ASTM, USP) and predefined acceptance criteria for various physical, chemical, and biological tests.

    8. The sample size for the training set

    Not applicable, as this is a traditional medical device, not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable, as this is a traditional medical device, not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220558

    Validate with FDA (Live)

    Date Cleared
    2022-11-17

    (262 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180739

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Applicable to Product Code 2N3383: For the administration of blood components or solutions from a container into the patient's vascular system through a vascular access device.

    Applicable to Product Code 2N3385: For the administration of blood components or solutions from a container into the patient's vascular system through a vasular access device. Only for use with Neonates and Pediatios. Not for use in Trauma situations.

    Device Description

    Baxter's IV Administration Sets (Blood Administration Sets) are single use, nonpyrogenic, sterile disposable devices intended for the administration of fluids from a container into the patient's vascular system. They can be used to administer solutions, blood, and blood products to patients.

    The proposed blood set configuration (Product Code 2N3385) consists of non-DEHP PVC (< 0.1% DEHP) tubing, blood chamber, 200 um Filter, Clearlink Luer Activated Valve (LAV), notch clamp, female Luer lock, dual anti-syphon valve, male Luer lock, and a filter vented cap for a male Luer lock. It can be used to administer solutions, blood, and blood products to the patient.

    The blood set (2N3385) will be used for syringe-based infusion. The clinician attaches the blood set to a blood bag and manually draws blood into a syringe. The blood is then delivered into the patient's vascular system through a vascular access device.

    AI/ML Overview

    The provided text describes a medical device, "Blood Administration Sets," and its substantial equivalence to a predicate device, but it does not contain information relevant to AI/ML device acceptance criteria or studies. The document is a 510(k) premarket notification for a traditional medical device (intravascular administration set), not an AI/ML device.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for an AI/ML device from this document.

    The document focuses on:

    • Device Description: Physical components, materials, and intended use as a blood administration set.
    • Technological Characteristics Comparison: A detailed table comparing the proposed device (2N3385) with a predicate device (2N3383), highlighting differences like length, priming volume, and specific components (e.g., spike, blood chamber, dual anti-siphon valve, Clearlink LAV).
    • Nonclinical Tests: Bench tests (e.g., Luer tests, tensile strength, leak tests, blood filter tests, spike tests, LAV tests, particulate matter, DEHP claim, blood compatibility, microbial ingress, shelf-life, shipping simulation) to evaluate functional performance and safety.
    • Biocompatibility: Assessments per ISO 10993-1.
    • Sterility: Validation of gamma radiation sterilization according to ISO 11137-2.
    • Shelf-Life: 3-year claim supported by aging testing.
    • Microbial Ingress Testing: Evaluations of potential entry points.

    All these tests are standard for conventional medical devices and do not involve AI/ML performance evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212262

    Validate with FDA (Live)

    Device Name
    PCA Syringe Sets
    Date Cleared
    2022-10-03

    (440 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180739

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the administration of fluids from a container into the patient's vascular system through a vascular access device. For use in adult populations only.

    Device Description

    The proposed devices are patient controlled analgesia (PCA) syringe sets. They are single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a vascular access device. They are non-pyrogenic, sterile devices that can be directly attached to a syringe.

    The PCA syringe sets consist of non-DEHP PVC tubing/bushing, female luer cap, antisiphon valve with female luer, on-off clamp(s), male luer lock, male luer cap, back check valve with female luer (2P3331 and 2P3332 only), and y-connector (2P3331 and 2P3332 only). They are used to administer analgesics from a syringe to the patient IV access device (2P3331 and 2P3332 only) or from a syringe to a primary administration set (2P3333 only); and are also used to administer fluids from a container at the y-type connector (2P3331 and 2P3332 only).

    AI/ML Overview

    This document describes the acceptance criteria and the studies conducted to prove that Baxter Healthcare Corporation's PCA Syringe Sets meet these criteria, thereby demonstrating substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Functional Performance (Bench Tests)All tests met acceptance criteria.
    ISO 80369-7 Tests on Male Luer Lock ConnectorISO 80369-7:2016, Clause 5 (as applicable), and Clauses 6.1, 6.2, 6.3, 6.4, 6.5, 6.6Met acceptance criteria
    ISO 80369-7 Tests on Female Luer Lock ConnectorISO 80369-7:2016, Clause 5 (as applicable), and Clauses 6.1, 6.2, 6.3, 6.4, 6.5, 6.6Met acceptance criteria
    Tensile Strength TestBS EN ISO 8536-9:2015, Clause 5.3Met acceptance criteria
    Leak TestBS EN ISO 8536-9:2015, Clause A.4Met acceptance criteria
    Counter Flow TestBS ISO 8536-12:2007+A1:2013, Clause A.4Met acceptance criteria
    Blocking Performance TestBS ISO 8536-12:2007+A1:2013, Clause A.6Met acceptance criteria
    Opening Pressure Test (Back Check Valve and Anti-Siphon Valve)BS ISO 8536-12:2007+A1:2013, Clause A.7.1 and per Baxter Test MethodMet acceptance criteria
    Particulate Matter TestUSP Chapter <788>Met acceptance criteria
    Clamp Activation Force TestActivation force ≤ 50NMet acceptance criteria
    Clamp Shut-Off TestBS EN ISO 8536-14:2018, Clause A.1Met acceptance criteria
    Non-DEHP Claim VerificationPer Baxter Test Method (as tested in K161808)Met acceptance criteria
    BiocompatibilityDevice met the designated ISO 10993-1 categorization and is biocompatible for intended use.Met acceptance criteria
    CytotoxicityISO 10993-5Met acceptance criteria
    SensitizationISO 10993-10Met acceptance criteria
    Intracutaneous (Irritation) ReactivityISO 10993-10Met acceptance criteria
    Systemic Toxicity (acute and repeat dose)ISO 10993-11Met acceptance criteria
    Material Mediated PyrogenISO 10993-11Met acceptance criteria
    GenotoxicityISO 10993-3Met acceptance criteria
    HemolysisISO 10993-4Met acceptance criteria
    SterilitySterility Assurance Level (SAL) of 10⁻⁶Met acceptance criteria (validated)
    Bacterial endotoxins testsUSP <85>, Endotoxin limit of 20 EU/device per USP <161>Met acceptance criteria
    Shelf Life3 (three) years of shelf lifeSupported (aging testing performed)
    Microbial Ingress TestingAbsence of microbial ingress to the sterile fluid path (following Baxter's testing strategy, K180739)Met acceptance criteria

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the exact sample sizes used for each individual performance test (e.g., number of devices tested for leak, tensile strength, etc.). It generally states that "All tests met the acceptance criteria" and that "Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses."

    The data provenance is implied to be from internal laboratory testing conducted by Baxter Healthcare Corporation. No information is provided regarding the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of these tests (bench tests, biocompatibility, sterility, shelf life), they are typically prospective studies performed in a controlled laboratory environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The studies described are primarily engineering performance tests, biocompatibility assessments, and sterility validations, which rely on industry standards (ISO, BS EN, USP) and predefined acceptance criteria rather than expert-established ground truth in the context of clinical interpretation or diagnosis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable. The tests performed are objective measurements and validations against specified standards, not subjective assessments requiring adjudication by multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device (PCA Syringe Sets) is a medical administration set, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers or AI assistance are irrelevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This is not applicable. The device is a physical medical administration set, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance of the device is established by adherence to recognized national and international engineering and medical device standards, such as ISO 80369-7, BS EN ISO 8536-9, BS ISO 8536-12, BS EN ISO 8536-14, USP Chapters <788> and <85>, USP <161>, ISO 10993 series, and ANSI/AAMI/ISO 11137 series. These standards define the acceptable range of performance for specific physical and biological characteristics of the device.

    8. The sample size for the training set:

    This is not applicable. The device is a physical product and not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable because there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203609

    Validate with FDA (Live)

    Date Cleared
    2021-09-30

    (294 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the administration of fluids from a container into the patient's vascular system through a vascular access device.

    Device Description

    The proposed devices consist of Intravascular Administration Sets. These devices include Basic, Secondary, and CONTINU-FLO solution sets. They are single use disposable, non-pyrogenic, sterile devices intended for the administration of fluids from a container into the patient's vascular system.

    AI/ML Overview

    The provided text is a 510(k) summary for Baxter Healthcare Corporation's Intravascular Administration Sets (K203609). It details the substantial equivalence determination for this medical device to a predicate device (Solution Administration Sets, K112893).

    However, the summary does not contain the specific information requested in the prompt regarding acceptance criteria, device performance, sample sizes, expert involvement, ground truth establishment, or multi-reader multi-case studies.

    The document primarily focuses on:

    • Device Description: What the Intravascular Administration Sets are, their components, and how they function.
    • Predicate Device Comparison: A detailed table comparing the technological characteristics of the proposed device to the predicate device, highlighting differences in length, priming volume, dimensions, and material components.
    • Discussion of Differences: Explanations for each technological difference and assurance that these differences do not raise new questions of safety or effectiveness.
    • Nonclinical Tests: A list of bench tests (Luer tests, particulate matter, non-DEHP claim, solvent bond tests, flow rate tests, pump compatibility), biocompatibility tests (cytotoxicity, sensitization, toxicity, hemolysis), and sterility tests (barrier packaging, fluid path, shelf life, sterilization dose establishment, pyrogen testing, microbial ingress). It states that all test results meet their acceptance criteria.

    Therefore, I cannot populate the table or provide detailed answers to questions 1-9 as the necessary information is not present in the provided document. The document states that "All test results meet their acceptance criteria," but it does not define what those criteria are or report specific performance metrics against those criteria. It also does not discuss any studies involving human readers or expert consensus for ground truth.

    Here's what I can infer from the document, though it falls short of the requested detail:

    • Acceptance Criteria & Reported Performance: The document states, "All test results meet their acceptance criteria." This implies that acceptance criteria were established for each of the listed bench, biocompatibility, and sterility tests. However, the specific quantitative acceptance criteria (e.g., maximum allowable particulate matter, minimum burst pressure, flow rate accuracy range) and the reported device performance (e.g., actual particulate count, measured burst pressure, achieved flow rate accuracy) are not provided.
    • Sample Sizes: The document does not specify sample sizes used for any of the tests.
    • Data Provenance: The tests are described as "bench tests," "biocompatibility," and "sterility" tests conducted by the manufacturer (Baxter Healthcare Corporation). The data would therefore be prospective, internal testing data. No country of origin for data is specified beyond the manufacturer's location (Round Lake, Illinois).
    • Experts for Ground Truth / Adjudication / MRMC Study / Standalone Performance: This section of the prompt is highly relevant for AI/ML device clearances (e.g., software as a medical device). This document describes a traditional Class II medical device (intravascular administration sets). There is no mention of AI/ML components, human readers, expert panels, or comparative effectiveness studies in the context of diagnostic or interpretive performance. Therefore, questions regarding these aspects are not applicable to the content provided.
    • Type of Ground Truth: For this type of physical device, "ground truth" would be established by the physical and chemical properties and performance characteristics measured in the listed bench, biocompatibility, and sterility tests, compared against established engineering standards (e.g., ISO, USP, ASTM) and internal specifications. There is no subjective human interpretation or diagnostic outcome data involved.
    • Training Set Sample Size / Ground Truth Establishment (for AI/ML): These questions are entirely irrelevant to this device and the provided document, as it is not an AI/ML product. The "training set" here would metaphorically be the design and manufacturing processes refined over time.

    In summary, the provided document is a regulatory submission for a physical medical device, not an AI/enabled one. Therefore, many of the questions asked, particularly those related to data sets, expert involvement, and reader studies, are not applicable to the content provided.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1