K Number
K212262

Validate with FDA (Live)

Device Name
PCA Syringe Sets
Date Cleared
2022-10-03

(440 days)

Product Code
Regulation Number
880.5440
Age Range
2 - 65
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the administration of fluids from a container into the patient's vascular system through a vascular access device. For use in adult populations only.

Device Description

The proposed devices are patient controlled analgesia (PCA) syringe sets. They are single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a vascular access device. They are non-pyrogenic, sterile devices that can be directly attached to a syringe.

The PCA syringe sets consist of non-DEHP PVC tubing/bushing, female luer cap, antisiphon valve with female luer, on-off clamp(s), male luer lock, male luer cap, back check valve with female luer (2P3331 and 2P3332 only), and y-connector (2P3331 and 2P3332 only). They are used to administer analgesics from a syringe to the patient IV access device (2P3331 and 2P3332 only) or from a syringe to a primary administration set (2P3333 only); and are also used to administer fluids from a container at the y-type connector (2P3331 and 2P3332 only).

AI/ML Overview

This document describes the acceptance criteria and the studies conducted to prove that Baxter Healthcare Corporation's PCA Syringe Sets meet these criteria, thereby demonstrating substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
Functional Performance (Bench Tests)All tests met acceptance criteria.
ISO 80369-7 Tests on Male Luer Lock ConnectorISO 80369-7:2016, Clause 5 (as applicable), and Clauses 6.1, 6.2, 6.3, 6.4, 6.5, 6.6Met acceptance criteria
ISO 80369-7 Tests on Female Luer Lock ConnectorISO 80369-7:2016, Clause 5 (as applicable), and Clauses 6.1, 6.2, 6.3, 6.4, 6.5, 6.6Met acceptance criteria
Tensile Strength TestBS EN ISO 8536-9:2015, Clause 5.3Met acceptance criteria
Leak TestBS EN ISO 8536-9:2015, Clause A.4Met acceptance criteria
Counter Flow TestBS ISO 8536-12:2007+A1:2013, Clause A.4Met acceptance criteria
Blocking Performance TestBS ISO 8536-12:2007+A1:2013, Clause A.6Met acceptance criteria
Opening Pressure Test (Back Check Valve and Anti-Siphon Valve)BS ISO 8536-12:2007+A1:2013, Clause A.7.1 and per Baxter Test MethodMet acceptance criteria
Particulate Matter TestUSP Chapter <788>Met acceptance criteria
Clamp Activation Force TestActivation force ≤ 50NMet acceptance criteria
Clamp Shut-Off TestBS EN ISO 8536-14:2018, Clause A.1Met acceptance criteria
Non-DEHP Claim VerificationPer Baxter Test Method (as tested in K161808)Met acceptance criteria
BiocompatibilityDevice met the designated ISO 10993-1 categorization and is biocompatible for intended use.Met acceptance criteria
CytotoxicityISO 10993-5Met acceptance criteria
SensitizationISO 10993-10Met acceptance criteria
Intracutaneous (Irritation) ReactivityISO 10993-10Met acceptance criteria
Systemic Toxicity (acute and repeat dose)ISO 10993-11Met acceptance criteria
Material Mediated PyrogenISO 10993-11Met acceptance criteria
GenotoxicityISO 10993-3Met acceptance criteria
HemolysisISO 10993-4Met acceptance criteria
SterilitySterility Assurance Level (SAL) of 10⁻⁶Met acceptance criteria (validated)
Bacterial endotoxins testsUSP <85>, Endotoxin limit of 20 EU/device per USP <161>Met acceptance criteria
Shelf Life3 (three) years of shelf lifeSupported (aging testing performed)
Microbial Ingress TestingAbsence of microbial ingress to the sterile fluid path (following Baxter's testing strategy, K180739)Met acceptance criteria

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the exact sample sizes used for each individual performance test (e.g., number of devices tested for leak, tensile strength, etc.). It generally states that "All tests met the acceptance criteria" and that "Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses."

The data provenance is implied to be from internal laboratory testing conducted by Baxter Healthcare Corporation. No information is provided regarding the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of these tests (bench tests, biocompatibility, sterility, shelf life), they are typically prospective studies performed in a controlled laboratory environment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. The studies described are primarily engineering performance tests, biocompatibility assessments, and sterility validations, which rely on industry standards (ISO, BS EN, USP) and predefined acceptance criteria rather than expert-established ground truth in the context of clinical interpretation or diagnosis.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable. The tests performed are objective measurements and validations against specified standards, not subjective assessments requiring adjudication by multiple experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device (PCA Syringe Sets) is a medical administration set, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers or AI assistance are irrelevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

This is not applicable. The device is a physical medical administration set, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance of the device is established by adherence to recognized national and international engineering and medical device standards, such as ISO 80369-7, BS EN ISO 8536-9, BS ISO 8536-12, BS EN ISO 8536-14, USP Chapters <788> and <85>, USP <161>, ISO 10993 series, and ANSI/AAMI/ISO 11137 series. These standards define the acceptable range of performance for specific physical and biological characteristics of the device.

8. The sample size for the training set:

This is not applicable. The device is a physical product and not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:

This is not applicable because there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 3, 2022

Baxter Healthcare Corporation Meaghan Bonn Principal Specialist, Regulatory Affairs 25212 West Illinois Route 120 Round Lake, Illinois 60073

Re: K212262

Trade/Device Name: PCA Syringe Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA Dated: September 1, 2022 Received: September 2, 2022

Dear Meaghan Bonn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Wolloscheck, Ph.D. For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212262

Device Name PCA Syringe Sets

Indications for Use (Describe)

For the administration of fluids from a container into the patient's vascular system through a vascular access device. For use in adult populations only.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be italicized. The word is the logo for the healthcare company Baxter International.

K212262 510(k) Summary

DATE:

October 3, 2022

OWNER:

Baxter Healthcare Corporation 25212 W. IL Route 120 Round Lake, Illinois 60073

CONTACT PERSON:

Meaghan Bonn Principal Specialist, Regulatory Affairs 25212 W. IL Route 120 Round Lake, Illinois 60073 Telephone: (224) 270 6470 Fax: (224) 270 4119

IDENTIFICATION OF THE DEVICE:

Trade/Device Name: PCA Syringe Sets Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name: Set, Administration, Intravascular Regulatory Class: Class II Product Code: FPA

Code #Device Description
2P3331PCA Mini-Volume Syringe Set with Anti-Siphon Valveand Y-Type Connector with Back Check Valve, 74" (188 cm), Vol 1.2 mLImage: [Device Description Image]1: Female Luer Cap2: Anti-Siphon Valve w/ Female Luer3: Back Check Valve w/ Female Luer4: On-Off Clamp5: Y-Connector6: Male Luer Lock7: Male Luer Cap8: Non-DEHP Tube Bushing

Table 1. Proposed PCA Syringe Sets Configurations

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and has a slightly italicized appearance. The word is the logo for the healthcare company Baxter International.

Code #Device Description
9,10,11: Non-DEHP Tube
2P3332PCA Mini-Volume Syringe Set with Anti-Siphon Valve and Y-Type Connector with Back Check Valve, 98" (249 cm), Vol 1.5 mLImage: [PCA Mini-Volume Syringe Set with Anti-Siphon Valve and Y-Type Connector with Back Check Valve, 98" (249 cm), Vol 1.5 mL]1: Female Luer Cap2: Anti-Siphon Valve w/Female Luer3: Back Check Valve w/Female Luer4: On-Off Clamp5: Y-Connector6: Male Luer Lock7: Male Luer Cap8: Non-DEHP Tube Bushing9,10,11: Non-DEHP Tube
2P3333PCA Mini-Volume Syringe Set with Anti-Siphon Valve, 92" (234 cm), Vol 1.3 mLImage: [PCA Mini-Volume Syringe Set with Anti-Siphon Valve, 92" (234 cm), Vol 1.3 mL]1: Female Luer Cap2: Anti-Siphon Valve w/Female Luer3: On-Off Clamp4: Male Luer Lock5: Male Luer Cap6: Non-DEHP Tube Bushing7: Non-DEHP Tube

Table 1. Proposed PCA Syringe Sets Configurations

PREDICATE DEVICE:

DeviceCompanyPredicate 510(k)ClearanceDate
Clearlink Luer ActivatedValve, Clearlink System Non-DEHP Catheter Extension SetsBaxter HealthcareCorporationK112893 (model2N8374)October 18,2011

REASON FOR SUBMISSION:

The basis for this premarket notification is the intent to market patient controlled analgesia (PCA) syringe sets. The proposed devices in this submission are single-use, disposable devices, intended for the administration of fluids from a container into the patient's vascular system through a vascular access device.

DESCRIPTION OF THE DEVICE:

The proposed devices are patient controlled analgesia (PCA) syringe sets. They are single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a vascular access device. They are non-pyrogenic, sterile devices that can be directly attached to a syringe.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue, sans-serif font. The letters are slightly slanted to the right, giving the word a dynamic appearance. The word is likely a logo or brand name.

The PCA syringe sets consist of non-DEHP PVC tubing/bushing, female luer cap, antisiphon valve with female luer, on-off clamp(s), male luer lock, male luer cap, back check valve with female luer (2P3331 and 2P3332 only), and y-connector (2P3331 and 2P3332 only). They are used to administer analgesics from a syringe to the patient IV access device (2P3331 and 2P3332 only) or from a syringe to a primary administration set (2P3333 only); and are also used to administer fluids from a container at the y-type connector (2P3331 and 2P3332 only).

The anti-siphon valve and back check valve components add specific functionalities to facilitate the administration of fluid to the patient's vascular device, as described below:

  • Anti-siphon valve: reduces the risk of any inadvertent free flow of solution to the . patient. The higher opening pressure does not allow administration of flow into the patient's vascular access device, unless the stipulated pressure is applied. Another safety feature is that the valve does not allow any backflow. This prevention of backflow is achieved through the valve's ability to withstand a counterflow pressure.
  • Back check valve: does not allow any backflow. This prevention of backflow is . achieved through the valve's ability to withstand a counterflow pressure.

INDICATIONS FOR USE:

For the administration of fluids from a container into the patient's vascular system through a vascular access device. For use in adult populations only.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed devices are substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K112893 on October 18, 2011. The intended use and function of the proposed devices is equivalent to the predicate device.

Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is a dark blue color. The letters are slightly slanted to the right, giving the word a dynamic appearance.

FeaturesPredicate DeviceCleared underK112893 (Model2N8374)Proposed DevicesAssessment of Differences
Intended UseFor use with avascular access devicefor the administrationof drugs andsolutions. TheClearlink LuerActivated Valve is anin-line injection site,which can beconnected to thestandard male Lueradapters (e.g.,syringes or sets) forcontinuous orintermittent fluidadministration or thewithdrawal of fluid.For theadministration offluids from acontainer into thepatient's vascularsystem through avascular accessdevice.SameMinor rewording of theIntended Use statement hasbeen made to better alignwith 21 CFR 880.5440 andfor the purpose ofstreamlining the informationprovided to the user.The general purpose of thedevice and its functionremain unchanged.The minor rewording of theIntended Use statement doesnot raise different questionsof safety and effectiveness.
Indications for UseFor use with avascular access devicefor the administrationof drugs andsolutions. TheClearlink LuerActivated Valve is anin-line injection site,which can beconnected to thestandard male Lueradapters (e.g.,syringes or sets) forcontinuous orintermittent fluidadministration or thewithdrawal of fluid.For theadministration offluids from acontainer into thepatient's vascularsystem through avascular accessdevice. For use inadult populationsonly.SameMinor rewording of theIntended Use statement hasbeen made to better alignwith 21 CFR 880.5440 andfor the purpose ofstreamlining the informationprovided to the user.This minor modificationdoes not alter the disease orcondition the device willdiagnose, treat, prevent,cure/mitigate, or the patientpopulation for which thedevice is intended to be used.In addition, the minorrewording does not reflect adifferent anatomical sitefrom which a disease state orpopulation may be inferred.The minor rewording of theIntended Use statement doesnot raise different questionsof safety and effectiveness.
Regulation Number21 CFR 880.5440SameN/A
Product CodeFPASameN/A
FeaturesPredicate DeviceCleared underK112893 (Model2N8374)Proposed DevicesAssessment of Differences
SterileGamma RadiationSameN/A
Sterility AssuranceLevel (SAL)$10^{-6}$SameN/A
Non-PyrogenicYesSameN/A
Single UseYesSameN/A
Length8.2" (21 cm)74" (188 cm) - 98" (249 cm)The predicatedevice comparatively has ashorter length than all of theproposed devices. Theproposed devices offersets with different lengths,allowing the clinician toselect the appropriate one forthe intended therapy. Fordifferent therapies, cliniciansmight require sets withlonger lengths, to ensuredelivery of solutions withoutexcessive manipulation ofthe set.Design control activitieshave been conducted andhave confirmed that thedifferent technologicalcharacteristics of theproposed devices do notraise different questions ofsafety and effectiveness.
Tubing Inner / OuterDiameter0.039" (0.99 mm) /0.089" (2.26 mm)0.0315" (0.8 mm) /0.0866" (2.2 mm)The proposed devices tubinginner / outer diameters arenominally less than thepredicate device tubing inner/ outer diameter. Designcontrol activities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed device doesnot raise different questionsof safety and effectiveness.
Priming Volume0.5 mL1.2 mL - 1.5 mLDue to longer tubing lengths,the priming volumes of the
FeaturesPredicate DeviceCleared underK112893 (Model2N8374)Proposed DevicesAssessment of Differences
proposed devices are largerthan the predicate device.Design control activitieshave been conducted andhave confirmed that thedifferent technologicalcharacteristics of theproposed device does notraise different questions ofsafety and effectiveness.
Fluid Path Components/Materials
Anti-Siphon ValveNot ApplicablePolymethylmethacrylate(Acrylic)[Female Luer andTube Port]Silicone(Membrane)(2P3331, 2P3332,2P3333)The predicate device doesnot have an anti-siphonvalve. Design controlactivities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed devices donot raise different questionsof safety and effectiveness
Back Check ValveNot ApplicablePolymethylmethacrylate(Acrylic)[Female Luerand Tube Port]Silicone(Membrane)(2P3331, 2P3332)The predicate device doesnot have a back checkvalve. Design controlactivities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed devices donot raise different questionsof safety and effectiveness
TubingPolyvinyl Chloride(2N8374)Same(2P3331, 2P3332,2P3333)N/A
Tube BushingNot ApplicablePolyvinyl ChlorideThe predicate device doesnot have a tubebushing. Design controlactivities have been
FeaturesPredicate DeviceCleared underK112893 (Model2N8374)Proposed DevicesAssessment of Differences
Y-ConnectorNot ApplicableAcrylonitrileButadiene Styrene(2P3331, 2P3332)The predicate device doesnot have a y-connector. Design controlactivities have beenconducted and haveconfirmed that thedifferent technological characteristics of the proposeddevices do not raise differentquestions of safety andeffectiveness.
Male Luer LockAcrylonitrileButadiene Styrene(2N8374)Same(2P3331, 2P3332,2P3333)N/A
Male Luer CapPolypropylene(2N8374)High DensityPolyethylene (Cap)Hydrophobic Filter,Acrylic (W/Non-WovenNylon Substrate)(FilterMembrane)(2P3331, 2P3332,2P3333)The predicate device doesnot have the same type ofmaterial formulation. Designcontrol activities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed devices doesnot raise different questionsof safety and effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a company logo. The word is slightly slanted to the right.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is in blue color. The letters are closely spaced together.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Baxter" in a stylized, sans-serif font. The word is rendered in a vibrant blue color. The font appears bold and slightly italicized, giving it a dynamic and modern look.

DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is the primary focus of the image and is positioned horizontally.

Performance Data:

The following bench tests (Table 4) were conducted to evaluate the functional performance of the proposed devices:

TestAcceptance Criteria
ISO 80369-7 Tests on Male Luer Lock ConnectorISO 80369-7:2016, Clause 5 (as applicable),
ISO 80369-7:2016, Clause 6.1,
ISO 80369-7:2016, Clause 6.2,
ISO 80369-7:2016, Clause 6.3,
ISO 80369-7:2016, Clause 6.4,
ISO 80369-7:2016, Clause 6.5,
ISO 80369-7:2016, Clause 6.6
ISO 80369-7 Tests on Female Luer Lock ConnectorISO 80369-7: 2016, Clause 5 (as applicable),
ISO 80369-7:2016, Clause 6.1,
ISO 80369-7:2016, Clause 6.2,
ISO 80369-7:2016, Clause 6.3,
ISO 80369-7:2016, Clause 6.4,
ISO 80369-7:2016, Clause 6.5,
ISO 80369-7:2016, Clause 6.6
Tensile Strength TestBS EN ISO 8536-9:2015, Clause 5.3
Leak TestBS EN ISO 8536-9:2015, Clause A.4
Counter Flow TestBS ISO 8536-12:2007+A1:2013, Clause A.4
Blocking Performance TestBS ISO 8536-12:2007+A1:2013, Clause A.6
Opening Pressure Test (Back Check Valve and Anti-Siphon Valve)BS ISO 8536-12:2007+A1:2013, ClauseA.7.1 and per Baxter Test Method
Particulate Matter TestUSP Chapter $<788>$
Clamp Activation Force TestActivation force $≤$ 50N
Clamp Shut-Off TestBS EN ISO 8536-14:2018, Clause A.1
Non-DEHP Claim VerificationPer Baxter Test Method (as tested inK161808)
Table 4. Performance Data
-------------------------------

All tests met the acceptance criteria.

Biocompatibility:

Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged contact duration, external communicating device, indirect blood path, and FDA-2013-D-0350 Guidance for Industry & FDA Staff,

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is colored in a bright blue. The letters are slightly slanted to the right, giving the word a dynamic appearance.

Use of ISO 10993-1, "Biological evaluation and medical devices - Part 1: Evaluation and testing within a risk management process, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". Biocompatibility assessments include the following assays:

  • Cytotoxicity ISO 10993-5 ●
  • Sensitization ISO 10993-10
  • Intracutaneous (Irritation) Reactivity ISO 10993-10
  • Systemic Toxicity (acute and repeat dose) ISO 10993-11
  • Material Mediated Pyrogen ISO 10993-11
  • Genotoxicity ISO 10993-3
  • Hemolysis ISO 10993-4

Based upon the results the device met the designated ISO 10993-1 categorization and is biocompatible and appropriate for the intended use.

Sterility:

The proposed device is sterilized with gamma radiation. The product is in the bioburden (sub) category "General Sets Labeled 'Sterile''. The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, "Sterilization of health care products - Radiation-Part 2: Establishing the sterilization dose." The dose setting method used includes, but is not limited to, Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. Bacterial endotoxins tests were conducted in conformance to USP <85>. The endotoxin limit is 20 EU/device per USP <161>. In addition, routine periodic pre-sterilization bioburden testing is performed for each (sub) category. The sterilization process for the proposed devices was established in accordance with ANSI/AAMI/ISO 11137-1. "Sterilization of health care products-Radiation-Part 1; Requirements for development, validation and routine control of a sterilization process for medical devices."

Shelf Life:

Baxter has performed aging testing to support a shelf-life claim of 3 (three) years.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is in blue and appears to be a logo. The letters are closely spaced, giving the word a solid, unified appearance.

Microbial Ingress Testing:

Baxter has conducted testing on all the potential points of microbial entry into the sterile fluid pathway of the proposed devices. The potential microbial entry points consist of male and female Luer connector sites. The Luer connector sites were tested following Baxter's testing strategy (as previously cleared under K180739 (cleared on May 28, 2019)) of challenging the connections during simulated clinical use to ensure the absence of microbial ingress to the sterile fluid path. All test results met their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.

CONCLUSION:

The non-clinical data demonstrate that the subject devices are substantially equivalent and perform comparably to the predicate device that is legally marketed for the same intended use.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.