K Number
K172544

Validate with FDA (Live)

Date Cleared
2017-09-22

(30 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion).

Device Description

The Non-DEHP IV Fat Emulsion Administration Sets product line consists of single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a needle or catheter inserted into a vein. The Non-DEHP IV Fat Emulsion Administration Sets are indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion). These sets are comprised of a drip chamber with a non-vented spike, non-DEHP polyvinyl chloride (PVC) tubing, and a two piece Luer lock. On all sets there is a fixture slide clamp and a regulating roller clamp. Configurations of these sets differ in drip chamber (60 drops/mL or 10 drops/mL). These devices are nonpyrogenic, sterile and can be used for gravity or pump infusion of I.V. fluids using Sigma Spectrum and other Baxter Infusion Pumps.

AI/ML Overview

This document describes the non-clinical performance testing of the "Non-DEHP IV Fat Emulsion Administration Sets" (K172544) by Baxter Healthcare Corporation.

1. Acceptance Criteria and Reported Device Performance

The provided document lists various performance tests undertaken for the device. The reported performance for all tests is that they "All tests met the acceptance criteria." The specific acceptance criteria are broadly defined as "Per Baxter Test Method" for each test.

Here's a table summarizing the tests and the reported performance:

TestAcceptance CriteriaReported Device Performance
Solvent Bond Tensile Strength TestPer Baxter Test Method.Met Acceptance Criteria
Solvent Bond Air Pressure TestPer Baxter Test Method.Met Acceptance Criteria
Continuous Bubble Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
Accumulated Bubble Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
Slide Clamp Shut-Off TestPer Baxter Test Method.Met Acceptance Criteria
Slide Clamp Shut-Off Test-Post 24 HourPer Baxter Test Method.Met Acceptance Criteria
Slide Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.Met Acceptance Criteria
Slide Clamp Flow Stability TestPer Baxter Test Method.Met Acceptance Criteria
Slide Clamp Tug Stability TestPer Baxter Test Method.Met Acceptance Criteria
Upstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
Downstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
Roller Clamp Force TestPer Baxter Test Method.Met Acceptance Criteria
Roller Clamp Shut-Off TestPer Baxter Test Method.Met Acceptance Criteria
Roller Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.Met Acceptance Criteria
Roller Clamp Flow Stability TestPer Baxter Test Method.Met Acceptance Criteria
Roller Clamp Tug Stability TestPer Baxter Test Method.Met Acceptance Criteria
Roller Clamp Flow Control TestPer Baxter Test Method.Met Acceptance Criteria
Administration Set Integrity Test After Maximum Fluid DeliveryPer Baxter Test MethodMet Acceptance Criteria
Visual Inspection For Structural Damage After 24 hours Exposure to LipidsPer Baxter Test MethodMet Acceptance Criteria
Flow Rate Accuracy (Pump Compatibility Test)Per Baxter Test MethodMet Acceptance Criteria
Non-DEHP Claim VerificationPer Baxter Test MethodMet Acceptance Criteria
Shelf Life (Aging Test)Not explicitly stated2 years
Sterility (SAL)Not explicitly stated10^-6 SAL
Package Verification (Visual Inspection, ASTM F88 Seal Strength, ASTM F2096 Bubble Leak)Not explicitly statedMet Acceptance Criteria
BiocompatibilityISO-10993-1, G95-1Met Requirements

2. Sample Size Used for the Test Set and Data Provenance

The document states that testing was performed on "subject device model code set configurations 2R1145 and 2R1146" or on "a worst case/representative sample."

  • Sample Size: The exact number of units or samples used for each test is not explicitly stated.
  • Data Provenance: The tests are described as bench tests conducted by Baxter Healthcare Corporation. The document does not specify the country of origin of the data or whether it was retrospective or prospective, but it implies prospective testing for the purpose of the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable as the evaluation involves non-clinical bench testing of physical device properties and performance, not a diagnostic or clinical assessment requiring expert-established ground truth.

4. Adjudication Method for the Test Set

This section is not applicable for non-clinical bench testing. The evaluation relies on specific test methods and their quantitative or qualitative outcomes.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices or software where human readers interpret medical images, often with and without AI assistance. The submitted device is an administration set, which does not involve human image interpretation in that context.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is an intravenous administration set, not an algorithm. The performance testing is for the physical device itself.

7. The Type of Ground Truth Used

For the non-clinical performance tests, the "ground truth" is established by defined engineering specifications, industry standards, and internal Baxter test methods. These methods determine the acceptable range or outcome for each performance characteristic (e.g., tensile strength, leak integrity, flow rate, biocompatibility).

8. The Sample Size for the Training Set

This section is not applicable as the device is a physical medical device, not a machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established

This section is not applicable as the device is a physical medical device, not a machine learning algorithm.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 22, 2017

Baxter Healthcare Corporation Jill Laack Regulatory Affairs Specialist 32650 N. Wilson Road Round Lake, Illinois 60073

Re: K172544

Trade/Device Name: Non-DEHP IV Fat Emulsion Administration Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: August 21, 2017 Received: August 23, 2017

Dear Jill Laack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/0/Picture/10 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K172544

Device Name Non-DEHP IV Fat Emulsion Administration Sets

Indications for Use (Describe)

Indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right.

K172544

Section 5. 510(k) Summary

September 22, 2017

OWNER:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015

CONTACT PERSON:

Jill M. Laack Regulatory Specialist, Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073 Telephone: (224) 270-2686 Fax: (224) 270-4119

IDENTIFICATION OF THE DEVICE:

Common Name: Intravascular Administration Sets Trade Name: Non-DEHP IV Fat Emulsion Administration Sets Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name: Set, Administration, Intravascular Regulatory Class: Class II Product Code: FPA

Code NumberName
2C1145Non-DEHP IV Fat Emulsion Administration Set86" (2.2 m) Approximately 10 drops/mL
2C1146Non-DEHP IV Fat Emulsion Administration Set86" (2.2 m) Approximately 60 drops/mL
2R1145Non-DEHP IV Fat Emulsion Administration Set86" (2.2 m) Approximately 10 drops/mL
2R1146Non-DEHP IV Fat Emulsion Administration Set86" (2.2 m) Approximately 60 drops/mL

Table 1.Device Model Code(s) Non-DEHP IV Fat Emulsion Administration Sets

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is likely a logo or brand name, given its prominent display.

PREDICATE DEVICE:

DeviceCompanyPredicate 510(k)Clearance Date
Low PlasticizerSolubility PVCAdministration SetTravenol Laboratories,S.A.1K791496April 8, 1980

Table 2. Predicate Device

1 Travenol Laboratories, S.A. became Baxter Healthcare Corporation in 1988

Table 3. Reference 510(k)

DeviceCompanyReference 510(k)Clearance Date
Solution Set With 1.2Micron Air EliminatingFilterBaxter HealthcareCorporationK161808June 21, 2017

DESCRIPTION OF THE DEVICE:

The Non-DEHP IV Fat Emulsion Administration Sets product line consists of single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a needle or catheter inserted into a vein. The Non-DEHP IV Fat Emulsion Administration Sets are indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion). These sets are comprised of a drip chamber with a non-vented spike, non-DEHP polyvinyl chloride (PVC) tubing, and a two piece Luer lock. On all sets there is a fixture slide clamp and a regulating roller clamp. Configurations of these sets differ in drip chamber (60 drops/mL or 10 drops/mL). These devices are nonpyrogenic, sterile and can be used for gravity or pump infusion of I.V. fluids using Sigma Spectrum and other Baxter Infusion Pumps. The Non-DEHP IV Fat Emulsion Administration Sets were previously cleared on April 8, 1980 under 510(k) premarket notification "Low Plasticizer Solubility PVC Administration Sets" (K791496).

The basis for this premarket notification is modifications to the Non-DEHP IV Fat Emulsion Administration Sets product line. The modifications consist of the use of a Non-DEHP PVC tubing material that uses Di (2-Ethylhexyl) terephthalate as a plasticizer, conversion to sterile packaging, and the addition of an expiration date. The proposed Non-DEHP PVC tubing material is currently used in Baxter's Solution Sets with 1.2 Micron Air Eliminating Filter and has been previously cleared under 510(k) premarket notification K161808 (cleared June 21, 2017).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The overall impression is of a clean and professional logo.

The product labels are being updated to include compatible infusion pump device references, the addition of the indications for use statement, the addition of a latex statement, the addition of an expiration date, a change to the sterility claim, and to implement other clarifying information and modifications to comply with Baxter's labeling standards and other labeling enhancements.

See Table 4 for a summary of the proposed changes to the codes that are part of this premarket notification.

Code and DescriptionChange Summary
2C1145 - Non-DEHP IV Fat EmulsionAdministration Set 86" (2.2 m) Approximately 10drops/mL1. Label Changes2. Conversion to sterile packaging3. Addition of Expiration Date
2C1146 - Non-DEHP IV Fat EmulsionAdministration Set 86" (2.2 m) Approximately 60drops/mL1. Label Changes2. Conversion to sterile packaging3. Addition of Expiration Date
2R1145 - Non-DEHP IV Fat EmulsionAdministration Set 86" (2.2 m) Approximately 10drops/mL1. Label Changes2. Conversion to sterile packaging3. Addition of Expiration Date4. Use of Non-DEHP PVC TubingMaterial that uses Di (2-Ethylhexyl) terephthalate as aplasticizer
2R1146 - Non-DEHP IV Fat EmulsionAdministration Set 86" (2.2 m) Approximately 60drops/mL1. Label Changes2. Conversion to sterile packaging3. Addition of Expiration Date4. Use of Non-DEHP PVCTubing Material that uses Di (2-Ethylhexyl) terephthalate as aplasticizer

Table 4. Summary of Proposed Changes per Device Model Code

These modifications do not impact the intended use or the fundamental technology of the devices.

See Table 5 for the subject device model code set configuration that will utilize the proposed Non-DEHP PVC tubing material. Table 6 contains the remaining subject device model code set configurations that are part of this premarket notification.

Table 5. Set Configuration Using Proposed Non-DEHP PVC Tubing Material

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of forward motion.

Image /page/6/Figure/1 description: The image shows a medical device called a Non-DEHP IV Fat Emulsion Administration Set. The set is 86 inches (2.2 m) long and delivers approximately 60 drops/mL. The image also contains the code 2R1146.

Table 6. Set Configuration of Current Codes

Image /page/6/Figure/3 description: The image shows a table with two rows describing medical devices. The first row describes a Non-DEHP IV Fat Emulsion Administration Set 86" (2.2 m) that dispenses approximately 10 drops/mL, with a code number of 2C1145. The second row describes a similar device, also a Non-DEHP IV Fat Emulsion Administration Set 86" (2.2 m), but this one dispenses approximately 60 drops/mL and has a code number of 2C1146.

INDICATIONS FOR USE:

Indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion).

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed devices have equivalent technological characteristics as Baxter's current legally marketed device cleared under 510(k) premarket notification K791496 (cleared April 8, 1980).

DEVICE COMPARISON TABLE:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is likely a logo or brand name.

Table 7 provides a comparison of the (predicate) device and the proposed devices. This comparison identifies similarities and differences of the proposed devices to the current legally marketed devices to which substantial equivalency is claimed.

FeaturesPredicate Device(Cleared underK791496)Proposed Devices(2C1145, 2C1146, 2R1145,2R1146)Assessment of Differences
Indicationsfor useThe IV Fat EmulsionAdministration Sets areindicated for use whenlow plasticizer solubilityis desired (e.g., foradministration of highlipid content solutionssuch as parenteral fatemulsion).The IV Fat EmulsionAdministration Sets areindicated for use when lowplasticizer solubility isdesired (e.g., foradministration of high lipidcontent solutions such asparenteral fat emulsion).Same
SterileYesYesSame
Non-PyrogenicYesYesSame
Single UseYesYesSame
PumpCompatibilityNot identified insubmissionYes, compatible withSIGMA Spectrum and otherBaxter infusion pumps.Pump compatibility was notidentified in the originalsubmission for the PredicateDevice. Proposed codes havebeen tested for pumpcompatibility.
Materials
Non-ventedSpikeNot identified insubmissionAcrylonitrile ButadieneStyreneMaterials were not identifiedin the original submission forthe Predicate Device. Thematerial has been previouslycleared for the same intendeduse. All materials that arecurrently used have gonethrough design controlactivities which confirmedthere was no impact to thesafety or effectiveness of thedevice due to any changes ofmaterial.
FeaturesPredicate Device(Cleared underK791496)Proposed Devices(2C1145, 2C1146, 2R1145,2R1146)Assessment of Differences
DripChamberNot identified insubmissionPolyvinyl ChlorideMaterials were not identifiedin the original submission forthe Predicate Device. Thematerial has been previouslycleared for the same intendeduse. All materials that arecurrently used have gonethrough design controlactivities which confirmedthere was no impact to thesafety or effectiveness of thedevice due to any changes ofmaterial.
Non-DEHPTubingNot identified insubmissionMaterial: Polyvinyl ChloridePlasticizer: Tris (2-Ethylhexyl) Trimellitate(TOTM) (For codes 2C1145,2C1146)Plasticizer: Di (2-Ethylhexl)Terephthalate (DEHT) (Forcodes 2R1145, 2R1146)Materials were not identifiedin the original submission forthe Predicate Device. Thematerial has been previouslycleared for the same intendeduse. All materials that arecurrently used have gonethrough design controlactivities which confirmedthere was no impact to thesafety or effectiveness of thedevice due to any changes ofmaterial.
CannulaNot identified insubmissionStainless Steel 304Materials were not identifiedin the original submission forthe Predicate Device. Thematerial has been previouslycleared for the same intendeduse. Materials that arecurrently used have gonethrough design controlactivities which confirmedthere was no impact to thesafety or effectiveness of thedevice due to any changes ofmaterial.
FeaturesPredicate Device(Cleared underK791496)Proposed Devices(2C1145, 2C1146, 2R1145,2R1146)Assessment of Differences
Two PieceLuer BodyNot identified insubmissionAcrylonitrile ButadieneStyreneMaterials were not identifiedin the original submission forthe Predicate Device. Thematerial has been previouslycleared for the same intendeduse. All materials that arecurrently used have gonethrough design controlactivities which confirmedthere was no impact to thesafety or effectiveness of thedevice due to any changes ofmaterial.

Table 7. Device Comparison

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly angled, with the left side being lower than the right side. The background is white.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The overall impression is of a clean and professional logo.

DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet the acceptance criteria, and support that the proposed devices are appropriately designed for their intended use. All testing was performed on the configuration of the devices presented in this premarket notification.

Performance Data:

The following bench tests (Table 8) were performed on subject device model code set configurations 2R1145 and 2R1146 (Table 5) in order to mitigate risks such as leaks, under/over infusion, failure to shut-off flow, interruption of therapy, air infused into patient, exposure to toxic materials, exposure to non-sterile product, exposure to endotoxins or pyrogens, set is not sterile and product not functioning during shelflife. This testing that was performed for the proposed Non-DEHP PVC tubing material mitigated these risks so the devices are substantially equivalent to the predicate and can meet their intended use.

All testing described in this section was performed on codes that are listed in this submission or a worst case/representative sample. Worst case/representative samples were chosen because of having characteristics that are the same as the proposed device (e.g. -same components, same materials, same dimensions, and clamp/tubing interface).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slanting slightly to the right. The word is presented in a simple, clean design, with no additional graphics or elements surrounding it.

Table 8. Performance Testing Summary

TestAcceptance Criteria
Solvent Bond Tensile Strength TestPer Baxter Test Method.
Solvent Bond Air Pressure TestPer Baxter Test Method.
Continuous Bubble Test (Pump Compatibility Test)Per Baxter Test Method.
Accumulated Bubble Test (Pump Compatibility Test)Per Baxter Test Method.
Slide Clamp Shut-Off TestPer Baxter Test Method.
Slide Clamp Shut-Off Test-Post 24 HourPer Baxter Test Method.
Slide Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.
Slide Clamp Flow Stability TestPer Baxter Test Method.
Slide Clamp Tug Stability TestPer Baxter Test Method.
Upstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.
Downstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.
Roller Clamp Force TestPer Baxter Test Method.
Roller Clamp Shut-Off TestPer Baxter Test Method.
Roller Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.
Roller Clamp Flow Stability TestPer Baxter Test Method.
Roller Clamp Tug Stability TestPer Baxter Test Method.
Roller Clamp Flow Control TestPer Baxter Test Method.
Administration Set Integrity Test After Maximum Fluid DeliveryPer Baxter Test Method
Visual Inspection For Structural Damage After 24 hours Exposureto LipidsPer Baxter Test Method
Flow Rate Accuracy (Pump Compatibility Test)Per Baxter Test Method
Non-DEHP Claim VerificationPer Baxter Test Method

Note: All pump compatibility testing was performed with the Sigma Spectrum Pump All tests met the acceptance criteria.

Biocompatibility:

The following cleared 510k submissions: K161808, K153158, and K130245 have identical materials and similar intended uses compared to this Non-DEHP I.V. Fat Emulsion Administration Sets premarket notification. All the devices in these submissions have similar risks that were adequately mitigated through testing to show the devices are substantially equivalent to the predicate and meet their intended use.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "Baxter" in a stylized, bold, blue font. The font appears to be italicized, with the letters slanted to the right. The color of the text is a vibrant blue, which stands out against the white background. The word is easily readable and takes up a significant portion of the image.

Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path, and Blue Book Memorandum G95-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", as recommended in the IV Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]''.

The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device that is currently marketed for the same intended use.

Sterility

The Intravenous (IV) Fat Emulsion Administration Set product line is sterilized with radiation. The Minimum sterilizing Dose (MSD) required to provide a 10° Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO ISO 11137-2, "Sterilization of health care products- Radiation-Part 2: Establishing the Sterilization Dose."

These products are labeled "Sterile, Nonpyrogenic." Package Verification testing is based on Visual Inspection, ASTM F88 Seal Strength, and ASTM F2096 Bubble Leak.

Shelf Life

Baxter has performed aging testing to support a shelf-life claim of 2 (two) years.

CONCLUSION:

The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate devices that are currently marketed.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.