Found 5 results

510(k) Data Aggregation

    K Number
    K234106

    Validate with FDA (Live)

    Device Name
    Venue Fit
    Date Cleared
    2024-06-10

    (167 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K231301 Vscan Air, K161047 LOGIQ P9 and LOGIQ P7, K231989 LOGIQ E10s/LOGIQ Fortis, K180374 Voluson S8

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Fit is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. Venue Fit is intended to be used in a hospital or medical clinic. Venue Fit clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Color M Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD. B/Color/CWD.

    Device Description

    The Venue Fit is a general purpose diagnostic ultrasound system. It is a portable system with a touch screen interface. It can be powered through an electrical wall outlet or an internal battery. It utilizes linear, convex, and phased array transducers and supports standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures. The system can display the patient's ECG trace synchronized to the scanned image. A barcode reader and RFID scanner are available as additional input devices. A roller bag is available for transport. It is capable of wired or wireless internet connection and meets DICOM requirements for image storage and archiving.

    AI/ML Overview

    The provided text describes specific features and functionalities of the GE Venue Fit ultrasound system but does not contain information about acceptance criteria, reported device performance, or a specific study proving it meets acceptance criteria as typically outlined for AI/CADe devices.

    The document is a 510(k) Premarket Notification Submission for the Venue Fit ultrasound system. It primarily focuses on demonstrating substantial equivalence to a predicate device (K220848 Venue Fit) regarding general imaging capabilities, technological characteristics, safety, and effectiveness.

    Here's what can be extracted and what is explicitly stated as not applicable:

    1. A table of acceptance criteria and the reported device performance
    Not provided in the document. The document focuses on performance testing in the context of safety and effectiveness, rather than specific diagnostic accuracy metrics.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable. The document states, "The subject of this premarket submission, Venue Fit, did not require clinical studies to support substantial equivalence." Therefore, there is no test set in the context of a clinical study for diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable, as no clinical studies were required.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as no clinical studies were required.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This document does not describe an AI/CADe device that assists human readers, nor does it present results from an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. The Venue Fit is a diagnostic ultrasound system, not a standalone algorithm. Its capabilities, like "Auto Volume Flow (AVF)" or "Bladder Volume Tool", are functionalities of the ultrasound system itself, not standalone AI algorithms with reported diagnostic performance metrics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable, as no clinical studies requiring ground truth establishment were mentioned.

    8. The sample size for the training set
    Not applicable. The document does not describe a machine learning algorithm that would have a training set. The "Venue Fit" is a general-purpose diagnostic ultrasound system. While it has "semi-automated tools," the document does not detail their development or any AI-specific training sets.

    9. How the ground truth for the training set was established
    Not applicable, as there is no mention of a training set for an AI algorithm.


    Summary of what is present:

    The document establishes substantial equivalence through a comparison to a predicate device (K220848 Venue Fit) and a list of reference devices (K231301 Vscan Air, K161047 LOGIQ P9 and LOGIQ P7, K231989 LOGIQ E10s/LOGIQ Fortis, K180374 Voluson S8/ S10/ S10 Expert, K202035 Vscan Air).

    It outlines several non-clinical tests performed to ensure safety and compliance with standards:

    • Acoustic output evaluation
    • Biocompatibility
    • Cleaning and disinfection effectiveness
    • Thermal, electrical, electromagnetic, and mechanical safety

    The system states compliance with various voluntary standards including:

    • Marketing Clearance of Diagnostic Ultrasound Systems and Transducers - Guidance for Industry and Food and Drug Administration Staff (February 21, 2023)
    • AAMI/ANSI ES60601-1 (Medical Electrical Equipment - Part 1: General Requirements for Safety)
    • IEC 60601-1-2 (Electromagnetic Compatibility Requirements and Tests)
    • IEC 60601-2-37 (Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment)
    • IEC 62359 (Ultrasonics Field characterization Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields)
    • ISO 10993-1 (Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process)
    • ISO 14971 (Application of risk management to medical devices)
    • NEMA PS 3.1 3.20e (Digital Imaging and Communications in Medicine (DICOM) Set)
    • AAMI TIR69 (Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems)

    Quality assurance measures applied during development include:

    • Risk Analysis
    • Requirements Reviews
    • Design Reviews
    • Testing on unit level (Module verification)
    • Integration testing (System verification)
    • Performance testing (Verification & Validation)
    • Safety testing (Verification)

    The conclusion states that based on equipment design similarities, conformance to recognized performance standards, and performance testing, the proposed Venue Fit is considered substantially equivalent in safety, effectiveness, and performance to the predicate device.

    Crucially, the document explicitly states: "The subject of this premarket submission, Venue Fit, did not require clinical studies to support substantial equivalence." This means that the information requested regarding acceptance criteria, device performance, sample sizes, ground truth, experts, adjudication, and MRMC/standalone studies in a diagnostic performance context is not available within this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240053

    Validate with FDA (Live)

    Device Name
    Venue Go
    Date Cleared
    2024-06-10

    (154 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Go is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. Venue Go is intended to be used in a hospital or medical clinic. Venue Go clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis.

    The Venue Go is a compact, portable system with a small footprint. The system can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or wall mounted. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time with full functionality and scanning.

    The Venue Go utilizes a variety of linear, convex, and phased array transducers which provide high imaging performance and support standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures.

    The system is capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    A barcode reader and RFID scanner are available as additional input devices. A roller bag will also be available for the customer to use when transporting the system.
    Venue Go is capable of wireless communication through a built-in Wireless LAN device. The system meets DICOM requirements to support image storage and archiving (local PACS or products such as Q-Path) and allows for output to printing devices.

    AI/ML Overview

    The provided FDA 510(k) summary for the GE Venue Go ultrasound system does not include acceptance criteria for artificial intelligence (AI) performance or the study details to prove the device meets such criteria.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Venue Go K220800) and other reference devices, based on:

    • Identical fundamental scientific technology.
    • Similar imaging capabilities, technological characteristics, safety, and effectiveness.
    • Minor updates to the Indications for Use statement for clarity regarding operator qualifications.
    • Addition of new transducers already cleared on other GE devices.
    • New features/functionality such as Auto Volume Flow (AVF), Bladder Volume Tool, expanded MSK diagrams, Venue Coach, electronic software delivery, expanded Auto Inferior Vena Cava (IVC), Auto B-Lines, Real-Time ejection fraction (RT-EF), cNerve for Vscan Air CL and SL probes (which are already cleared with these AI features on predicate Vscan Air K231301 and reference LOGIQ E10s/LOGIQ Fortis K231989), and AppAPI functionality.
    • Conformity to recognized performance standards for acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, electromagnetic, and mechanical safety.
    • Quality assurance measures like risk analysis, requirements/design reviews, and various levels of testing.

    Specifically, the document states: "The subject of this premarket submission, Venue Go, did not require clinical studies to support substantial equivalence." This indicates that any evaluation of "AI performance" as typically understood (e.g., using a test set with ground truth, performance metrics, and acceptance criteria) was not part of this 510(k) submission.

    The AI-related features mentioned (Auto IVC, Auto B-Lines, RT-EF, cNerve) are noted as being "expanded" or already "cleared" on predicate/reference devices (K231301 Vscan Air, K231989 LOGIQ E10s/LOGIQ Fortis). This implies their performance was likely evaluated as part of those previous 510(k) submissions, and the current submission is leveraging that prior clearance for their inclusion in the Venue Go system when used with the particular probes.

    Therefore, since this 510(k) does not present new AI performance study data for the Venue Go, I cannot extract the requested information. The document focuses on demonstrating that the Venue Go, with its included features, is substantially equivalent to existing cleared devices, rather than proving novel AI algorithm performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240111

    Validate with FDA (Live)

    Device Name
    Venue
    Date Cleared
    2024-06-10

    (146 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K231301 Vscan Air, K161047 LOGIQ P9 and LOGIQ P7, K231989 LOGIQ E10s/LOGIQ Fortis, K180374 Voluson S8

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country. state or other local municipality in which he or she practices, for ultrasound imaging, measurement. display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training.

    Venue is intended to be used in a hospital or medical clinic. Venue clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic. Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B. M. PW Doppler, CW Doppler, Color M Doppler, Color M Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD,B/CWD, B/Color/CWD.

    Device Description

    The proposed Venue system is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement and analysis of the human body and fluid that provides digital acquisition, processing and display capabilities. Venue can be used in offices, clinics and hospitals.

    The Venue is a mobile system with a small footprint that easily fits into tight spaces and positioned to accommodate the sometimes-awkward work settings of the point of care user.

    The Venue has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The single surface screen provides easy cleanability. Articulated monitor arm enables flexible display positions in order to be accessible and clearly visible in both user-standing and sitting positions.

    The system is capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    Barcode reader and RFID scanner are available as additional input devices.

    The Venue has a battery that allows for scanning without the need to plug in to an electrical outlet.

    System meets DICOM requirements to support users image storage and archiving needs and allows for output to printing devices.

    The Venue utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures.

    The system includes several automated tools designed to simplify and shorten the workflow time of the healthcare professional for some common assessments.

    AI/ML Overview

    The provided document is a 510(k) Summary for a diagnostic ultrasound system (Venue). It does not contain information about acceptance criteria and a study that proves the device meets those criteria for software performance.

    The document states:

    • "The subject of this premarket submission, Venue, did not require clinical studies to support substantial equivalence." (Page 10)
    • The comparison is primarily focused on hardware, transducers, and general features, and states that the Venue system is substantially equivalent to a predicate device regarding "imaging capabilities, technological characteristics and safety and effectiveness." (Page 6)

    Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, data provenance, expert qualifications, ground truth establishment, or multi-reader multi-case studies because this document does not present such a study. The submission focuses on demonstrating substantial equivalence to a predicate device, which often relies on non-clinical tests and comparisons rather than new clinical performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210438

    Validate with FDA (Live)

    Device Name
    Versana Premier
    Date Cleared
    2021-09-10

    (210 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Versana Premier is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Premier clinical applications include : Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, VascularPeripheral Vascular, Musculoskeletal Conventional, Musculosketal Superficial, Thoracic/Pleural, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular access). Modes of operation include : B. M. PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD. Versana Premier is intended to be used in a hospital or medical clinic.

    Device Description

    The Versana Premier is a general purpose, Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals. The system is a mobile console that includes an operator control panel, display monitor and transducers. The console provides digital acquisition, processing and display capability. The system has an internal battery to allow for acquisition while the system is not plugged into a power source. The operator control panel includes function keys, trackball, an alfa-numeric keyboard and a touch panel as input sources of the device. The variety of transducers include convex, linear, sector, dual and mechanical 4D transducers. The access types include trans- body surface, transrectal, transvaginal and transcranial. Data can be imported or exported by DVD, USB, LAN or WiFi if the USB wireless adapter is connected to the system. An external ECG module has been verified to use as input for gating during scanning. The system has a HDMI port, VGA connection port, S-Video port, and a Composite Out port connection. The system has an external AC outlet to allow connection of a printer box of the console and an option for external Printer USB Isolator for other printers to connect. The system supports one way, Bluetooth communication capability from the system to a personal device to allow for sharing of the patient's data/images when the external Bluetooth USB adapter is connected to the system.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Submission for the GE Versana Premier ultrasound system. This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving that a new AI-powered diagnostic device meets specific performance acceptance criteria through a dedicated clinical study.

    Therefore, the information required to populate the fields related to acceptance criteria, ground truth, sample size, expert adjudication, and comparative effectiveness studies is not present in this document. The document explicitly states: "The subject of this premarket submission, Versana Premier, did not require clinical studies to support substantial equivalence."

    The only mention of an "Artificial Intelligence (AI) feature" is "Whizz Label," and no performance data or study details are provided for this feature. The document indicates that the Versana Premier employs "the same fundamental scientific technology as its predicate device and reference devices," and that it is "substantially equivalent ... with regard to intended use, imaging capabilities, technological characteristics, imaging modes, hardware, and safety effectiveness."

    Here's what can be extracted from the document regarding the AI feature, acknowledging the absence of the requested study details:

    • AI Feature Mentioned: Whizz Label is an Artificial Intelligence (AI) feature that is being added.

    All other requested information cannot be found in the provided document:

    1. A table of acceptance criteria and the reported device performance: Not provided.
    2. Sample sized used for the test set and the data provenance: Not provided.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not provided.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Explicitly stated that clinical studies were not required.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not provided.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not provided.
    8. The sample size for the training set: Not provided.
    9. How the ground truth for the training set was established: Not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201828

    Validate with FDA (Live)

    Date Cleared
    2020-10-11

    (101 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K180374 Voluson S series Diagnostic Ultrasound System, K192159 Voluson E10 Diagnostic Ultrasound System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voluson SWIFT, Voluson SWIFT+ is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Voluson SWIFT, Voluson SWIFT+ clinical applications include: Fetal/Obstetrics; Abdominal (including renal and GYN/Pelvic); Pediatric; Small Organ (Breast, Thyroid, etc.); Neonatal Cephalic; Adult Cephalic; Cardiaς (Adult and Pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal(including Urology/ Prostate) (TR): Transvaginal (TV).

    Mode of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography and Combined modes: B/M, B/Color, B/PWD, B/ Power/PWD. The Voluson SWIFT / Voluson™ SWIFT+ are intended to be used in a hospital or medical clinic.

    Device Description

    The subject device is a Track 3 device, primarily intended for general-purpose radiology evaluation and specialized for OB/GYN with particular features for real-time 3D/4D acquisition. The Voluson SWIFT, Voluson SWIFT+ provide digital acquisition, processing and display capability. Voluson SWIFT, Voluson SWIFT+ consist of a mobile console with control panel, full touch monitor, optional image storage and printing devices. It provides high-performance ultrasound imaging and analysis and have comprehensive networking and DICOM capability. It utilizes a variety of linear, curved linear, matrix phased array transducer including mechanical and electronic scanning transducers which provide accurate real-time three-dimensional imaging supporting all standard acquisition modes.

    AI/ML Overview

    The provided text is a 510(k) summary for the GE Voluson SWIFT, Voluson SWIFT+ diagnostic ultrasound system. It outlines the device's technical characteristics, intended use, and comparison to predicate devices to establish substantial equivalence.

    However, the document explicitly states that clinical studies were NOT required or performed to support the substantial equivalence of the Voluson SWIFT, Voluson SWIFT+.

    Therefore, I cannot provide the detailed information requested regarding acceptance criteria and performance studies, including:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes and data provenance for test sets.
    • Number and qualifications of experts for ground truth.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study results.
    • Standalone performance.
    • Type of ground truth used.
    • Sample size for the training set.
    • Method for establishing ground truth for the training set.

    The submission focuses entirely on non-clinical tests to demonstrate safety and effectiveness, based on compliance with electrical, mechanical, and safety standards, and by showing substantial equivalence to previously cleared predicate devices through similar technology, intended use, imaging modes, and acoustic power levels.

    The relevant section in the document states:
    "Summary of Clinical Tests: The subject of this premarket submission. Voluson SWIFT, Voluson SWIFT+, did not require clinical studies to support substantial equivalence."

    Ask a Question

    Ask a specific question about this device