K Number
K231301

Validate with FDA (Live)

Device Name
Vscan Air
Date Cleared
2023-08-15

(102 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The device is enabling visualization and measurement of anatomical structures and fluid including blood flow.
Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and other environments as described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage.
Vscan Air supports black/white (B-mode), color flow (Color Doppler mode, M-mode, M-mode, combined (B + Color Doppler) and Harmonic Imaging modes with curved, linear and sector array transducers.
With the curved array transducer of the dual headed probe solution, the specific clinical applications and exam types include: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatio, 40 kg and above), vascular/peripheral vascular, musculoskeletal (conventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).
With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam types include: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoraciclung, ophthalmic, pediatics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block, vascular access and biopsy).
With the sector array transducer of the dual headed probe solution, the specific clinical applications and exam types include: cardiac (adult and pediatic, 40 kg and above), abdominal, fetal/obstetnes, gynecological, urology, thoracic/ lung, pediatics, adult cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

Device Description

Vscan Air™ is a battery-operated general-purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. It enables ultrasound imaging guidance, visualization and measurement of anatomical structures and fluid.
Vscan Air consists of an app which can be installed on Android™ or iOS devices, and 2 probes which use wireless technology for communication.
Its pocket-sized portability and simplified user interface enable integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and in other environments. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage assessments for adult, pediatric and neonatal patients. Vscan Air can also be useful for interventional guidance.

AI/ML Overview

The provided text is a 510(k) Summary for the GE Medical Systems Ultrasound and Primary Care Diagnostics Vscan Air, outlining its substantial equivalence to predicate devices. It does not contain information on specific acceptance criteria and the detailed study results that prove the device meets these criteria in terms of clinical performance (e.g., accuracy, sensitivity, specificity) for a particular AI function.

The document primarily focuses on:

  • Regulatory information: Device name, regulation, product codes, predicate devices.
  • Intended Use/Indications for Use: Describes the general applications of the ultrasound system.
  • Technological comparison: Explains how the Vscan Air is similar to its predicate and reference devices in terms of technology, modes, transducers, and software features.
  • Non-Clinical Tests: Mentions evaluation for acoustic output, biocompatibility, cleaning/disinfection, and electrical/electromagnetic/mechanical safety, and adherence to various safety standards.
  • Absence of Clinical Studies: Explicitly states, "The subject of this premarket submission, Vscan Air, did not require clinical studies to support substantial equivalence."

Therefore, based on the provided input, I cannot fulfill the request for:

  1. A table of acceptance criteria and reported device performance (for clinical/AI performance).
  2. Sample size used for the test set and data provenance.
  3. Number of experts and their qualifications for ground truth.
  4. Adjudication method for the test set.
  5. MRMC study details or effect size.
  6. Standalone performance details.
  7. Type of ground truth used.
  8. Sample size for the training set.
  9. How ground truth for the training set was established.

The document confirms that this particular submission for Vscan Air did not require clinical studies, implying that the performance evaluation relied on non-clinical testing and comparison to predicate devices, rather than a prospective clinical trial with a defined test set and established ground truth for an AI algorithm.

{0}------------------------------------------------

August 15, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.

GE Medical Systems Ultrasound and Primary Care Diagnostics % Lee Bush Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226

Re: K231301

Trade/Device Name: Vscan Air Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 1, 2023 Received: August 1, 2023

Dear Lee Bush:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "GE Healthcare" at the top, followed by the text "510(k) Premarket Notification Submission" on the next line. The text is in a simple, sans-serif font and is centered on the image. The text is likely part of a document or presentation related to GE Healthcare's regulatory submissions.

Image /page/2/Picture/6 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are swirling water-like designs around the edge of the circle. The logo is simple and recognizable.

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2023
Indications for UseSee PRA Statement below.
510(k) Number (if known)
K231301
Device Name
Vscan Air
Indications for Use (Describe)
Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified andtrained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or otherlocal municipality in which he or she practices. The users may or may not be working under supervision or authority of aphysician. Users may also include medical students working under the supervision or authority of a physician during theireducation / training. The device is enabling visualization and measurement of anatomical structures and fluid includingblood flow.Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions andexaminations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/airambulance and other environments as described in the user manual. The information can be used for basic/focusedassessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up,and triage.Vscan Air supports black/white (B-mode), color flow (Color Doppler mode, M-mode, M-mode, combined(B + Color Doppler) and Harmonic Imaging modes with curved, linear and sector array transducers.With the curved array transducer of the dual headed probesolution, the specific clinical applications and exam typesinclude: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatio, 40 kg and above),vascular/peripheral vascular, musculoskeletal (conventional guidance (includes free handneedle/catheter placement, fluid drainage, nerve block and biopsy).With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam typesinclude: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoraciclung,ophthalmic, pediatics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluiddrainage, nerve block, vascular access and biopsy).With the sector array transducer of the dual headed probe solution, the specific clinical applications and exam typesinclude: cardiac (adult and pediatic, 40 kg and above), abdominal, fetal/obstetnes, gynecological, urology, thoracic/
lung, pediatics, adult cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage,
nerve block and biopsy).
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection ofinformation unless it displays a currently valid OMB number."
Dago 1 of 1

FORM FDA 3881 (6/20)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, surrounded by a circular design with wave-like elements. The logo is colored in a light blue hue. The wave-like elements give the impression of motion or energy.

510(k) Summary

K231301

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:May 4, 2023
Submitter:GE Medical Systems Ultrasound and Primary Care Diagnostics9900 Innovation DriveWauwatosa, WI 53226
Primary Contact Person:Lee BushRegulatory Affairs DirectorGE HealthcareT:(262)309-9429
Secondary Contact Person:Liwen WeiRegulatory Affairs LeaderGE Healthcare
Trade Name:Vscan Air
Common/Usual Name:Diagnostic Ultrasound Imaging System
Classification Names:Class II
Product Code:Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN;Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Predicate Device:Vscan Air (K202035)
Product Code:Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN;Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. There are also three water droplets or stylized swirls surrounding the circle, adding a dynamic element to the design. The logo is simple, recognizable, and represents the brand identity of General Electric.

Reference Devices:Vscan Extend (K180995)
Product Code:Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN;Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITXVenue Go (K220800)Ultrasonic Pulsed Doppler Imaging System. 21CFR, 892.1550 90-IYN;Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Device Description:Vscan Air™ is a battery-operated general-purpose diagnostic ultrasoundimaging system for use by qualified and trained healthcare professionals.It enables ultrasound imaging guidance, visualization and measurementof anatomical structures and fluid.Vscan Air consists of an app which can be installed on Android™ or iOSdevices, and 2 probes which use wireless technology for communication.Its pocket-sized portability and simplified user interface enableintegration into training sessions and examinations in professionalhealthcare facilities (ex. Hospital, clinic, medical office), homeenvironment, road/air ambulance and in other environments. Theinformation can be used for basic/focused assessments and adjunctivelywith other medical data for clinical diagnosis purposes during routine,periodic follow-up, and triage assessments for adult, pediatric andneonatal patients. Vscan Air can also be useful for interventionalguidance.
Intended Use/ Indicationsfor Use:Vscan Air is a battery-operated software-based general-purposeultrasound imaging system for use by qualified and trained healthcareprofessionals or practitioners that are legally authorized or licensed by lawin the country, state or other local municipality in which he or shepractices. The users may or may not be working under supervision orauthority of a physician. Users may also include medical students workingunder the supervision or authority of a physician during their education /training. The device is enabling visualization and measurement ofanatomical structures and fluid including blood flow.
Vscan Air's pocket-sized portability and simplified user interface enablesintegration into training sessions and examinations in professionalhealthcare facilities (ex. Hospital, clinic, medical office), homeenvironment, road/air ambulance and other environments as describedin the user manual. The information can be used for basic/focusedassessments and adjunctively with other medical data for clinicaldiagnosis purposes during routine, periodic follow-up, and triage.
Vscan Air supports Black/ white (B-mode), Color flow (Color doppler),Pulsed wave Doppler mode, M-mode, combined (B + Color Doppler) andHarmonic Imaging modes with curved, linear and sector arraytransducers.With the curved array transducer of the dual headed probe solution, thespecific clinical applications and exam types include: abdominal,fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult andpediatric, 40 kg and above), vascular/peripheral vascular, musculoskeletal(conventional), pediatrics, interventional guidance (includes free handneedle/catheter placement, fluid drainage, nerve block and biopsy).
With the linear array transducer of the dual headed probe solution, thespecific clinical applications and exam types include: vascular/peripheralvascular, musculoskeletal (conventional and superficial), small organs,thoracic/lung, ophthalmic, pediatrics, neonatal cephalic, interventionalguidance (includes free hand needle/catheter placement, fluid drainage,nerve block, vascular access and biopsy).
With the sector array transducer of the dual headed probe solution, thespecific clinical applications and exam types include: cardiac (adult andpediatric, 40 kg and above), abdominal, fetal/obstetrics, gynecological,urology, thoracic/ lung, pediatrics, adult cephalic, interventional guidance(includes free hand needle/catheter placement, fluid drainage, nerveblock and biopsy).
Technology:The Vscan Air employs the same fundamental scientific technology as itspredicate and reference devices.
Determination ofSubstantial Equivalence:Comparison to Predicate DevicesThe Vscan Air system is substantially equivalent to the predicate andreference devices with regards to intended use, capabilities, technologicalcharacteristics, safety and effectiveness.
Intended Use/Indications for Use
The systems are intended for diagnostic ultrasound imaging and fluid flowanalysisThe proposed Vscan Air and the predicate Vscan Air (K202035) have thesimilar clinical intended use and clinical applications however thefollowing applications are being added to the proposed Vscan Air:• Adult Cephalic, which can be found in reference device VscanExtend (K180995).
Modes and Transducers
The proposed Vscan Air and the predicate Vscan Air (K202035) bothsupport Black/white (B-mode), Color flow (Color doppler), combined (B +Color Doppler) and Harmonic Imaging modes. In addition, the proposedVscan Air also supports Pulsed wave Doppler mode and M-mode, whichare cleared in reference device Venue Go (K220800).
The proposed Vscan Air probes and the predicate Vscan Air (K202035)probe are manufactured with materials which have been evaluated andfound to be safe for the intended use of the device. The proposed VscanAir and the predicate Vscan Air (K202035) use the same linear arraytransducer for shallow scanning and curved array transducer for deepscanning. In addition, by introducing the new Vscan Air SL probe, theproposed Vscan Air adds a sector array for deep scanning, which is similarto the sector array that is cleared in reference device Vscan Extend(K180995).
Software
The proposed Vscan Air uses a software app which can be installed onuser's mobile device that provides some of the processing and uses thescreen of the mobile device as the display, same as the predicate deviceVscan Air (K202035).
The proposed Vscan Air has additional TCD (Transcranial Doppler) presetfor Adult Cephalic application which is cleared in reference device VscanExtend (K180995). The proposed Vscan Air supports same distance and

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with a stylized "GE" in the center. The "GE" is surrounded by swirling lines, which give the impression of movement and energy. The logo is simple, yet recognizable, and it represents the company's long history of innovation and technology.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative, swirling lines around the outer edge of the circle, giving the impression of movement or energy. The logo is simple, recognizable, and associated with a large, multinational corporation.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle is surrounded by a series of white, teardrop-shaped elements, resembling water droplets or stylized flames. The overall design is clean and recognizable, representing the brand identity of General Electric.

time and angle measurements and OB measurement display, which are cleared in reference device Venue Go (K220800) The proposed Vscan Air adds an annotation function to software app to add/delete/modify free text as labels on image, which is used in reference device Venue Go (K220800). The proposed Vscan Air and the predicate Vscan Air (K202035) both have a transducer element check function, however, the proposed Vscan Air made improvements to quickly assess health of transducers as soon as the probe starts up. The proposed Vscan Air software app allows probes to be on low-powerconnected mode while probe is in freeze mode or charging, for power saving and convenience of continuous scanning after probe charging. The proposed Vscan Air adds more basic patient information fields, configuration settings, and basic calculations, which are used in reference device Venue Go (K220800). Hardware The proposed Vscan Air uses Wi-Fi technology (Wi-Fi direct) to communicate between the probe and the mobile device with the installed Vscan Air app, same as predicate device Vscan Air (K202035). The proposed Vscan Air uses wireless charging technology to charge the Vscan Air CL and Vscan Air SL Probe batteries, same as predicate device Vscan Air (K202035). The proposed Vscan Air and predicate Vscan Air (K202035) have been designed in compliance with approved electrical and physical safety standards. Summary of Non-Clinical Tests The Vscan Air device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection as well as thermal, electrical, electromagnetic, mechanical safety and wireless, and has been found to conform to applicable medical device safety standards. The Vscan Air and its applications comply with voluntary standards: ● AAMI/ANSI ES60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance, 2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text)

ellipse measurement as predicate Vscan Air (K202035) and adds velocity,

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The logo is colored in blue, with the letters and border appearing in a lighter shade of blue against a darker blue background.

IEC 60601-1-2, Medical Electrical Equipment – Part 1-2: GeneralRequirements for Safety and Essential Performance - CollateralStandard: Electromagnetic disturbance - Requirements and Tests,Edition 4.1, 2020 IEC 60601-2-37, Medical Electrical Equipment – Part 2-37:Particular Requirements for the Safety of Ultrasonic MedicalDiagnostic and Monitoring Equipment, 2015 ISO 10993-1, Biological Evaluation of Medical Devices- Part 1:Evaluation and Testing- Fifth edition 2018-08 ISO 14971, Application of risk management to medical devices,2019 Edition 2.1 IEC 62359, Ultrasonics – Field characterization – Test methods forthe determination of thermal and mechanical indices related tomedical diagnostic ultrasonic fields, Edition 2.1, 2017 IEC 60601-1-11, Medical Electrical Equipment - Part 1-11: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Requirements For Medical ElectricalEquipment And Medical Electrical Systems Used In The HomeHealthcare Environment, Edition 2.1 2020-07 CONSOLIDATEDVERSION IEC 60601-1-12, Medical Electrical Equipment - Part 1-12: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Requirements For Medical ElectricalEquipment And Medical Electrical Systems Intended For Use In TheEmergency Medical Services Environment, Edition 1.1 2020-07CONSOLIDATED VERSION
The following quality assurance measures are applied to the developmentof the system: Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification & Validation) Safety testing (Verification)
Probe materials are biocompatible.
Summary of Clinical Tests:The subject of this premarket submission, Vscan Air, did not requireclinical studies to support substantial equivalence.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized, cursive font in the center. The circle is surrounded by a series of curved lines that resemble water droplets or waves. The logo is simple and recognizable, and it is often used to represent the company's products and services.

Conclusion:Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE Healthcare considers the proposed Vscan Air to be as safe, effective, and performs in a substantially equivalent manner as the predicate device Vscan Air (K202035).
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

N/A