Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222023

    Validate with FDA (Live)

    Device Name
    RayFlow
    Manufacturer
    Date Cleared
    2023-04-06

    (269 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K170544

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RayFlow catheter is intended to be used in adults (patients aged 22 years and older) for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the coronary vasculature. The RayFlow is not intended to be used in the neurovasculature.

    Device Description

    The RAYFLOW infusion catheter is a rapid exchange (RX) double lumen catheter with a series of holes and a radiopaque marker on its distal part. The outer lumen along the entire length of the catheter is used to inject fluids (saline, medication or contrast media) through the holes. The inner RX lumen is dedicated to the passage of a 0.014" quidewire facilitating the progression of the RAYFLOW catheter through the arteries. Four holes are located between the outer lumen and the surface of the catheter, allowing infusion of fluids into the blood resulting in a homogeneous mixing. Two holes located between the outer and inner lumen of the RAYFLOW catheter allow the measurement of the injected solution temperature by a specific pressure/temperature guidewire (Abbott PressureWire™ X Guidewire) at the moment the solution enters the artery. The hub is standard sized and compatible with any automated pump injector used for the continuous infusion of saline/liquid at room temperature. The RAYFLOW is provided sterilized with ethylene oxide. This device is for a single use only and packed in individual unit. A mandrel is inserted into the inner lumen to protect the integrity of the device.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the RayFlow catheter, a medical device. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily through non-clinical bench testing and adherence to recognized standards. It explicitly states that the device does not involve artificial intelligence (AI) and therefore, the acceptance criteria and study design elements typically associated with AI/ML-driven medical devices (such as MRMC studies, expert adjudication for ground truth, sample sizes for training/test sets in machine learning, and human reader performance with AI assistance) are not applicable.

    The "acceptance criteria" for this device are its performance in various bench tests, demonstrating that it meets the requirements of relevant ISO and ASTM standards, and that any differences from predicate devices do not raise new questions of safety or effectiveness.

    Here's an analysis based on the provided text:

    Device: RayFlow Catheter
    Device Type: Continuous flush catheter, intravascular diagnostic catheter, percutaneous catheter.
    Intended Use: Introduction of interventional devices and infusion of diagnostic or therapeutic agents into the coronary vasculature in adults (22 years and older). Not for neurovasculature.

    1. A table of acceptance criteria and the reported device performance:

    Since this is a non-AI device relying on bench testing, the acceptance criteria directly relate to meeting the specifications of recognized standards. The document summarizes the performance by stating the device "showed no sign of leakage," "is able to advance through a coronary anatomical model," "is conform to ISO 10555-1," and that all biocompatibility tests were passed. Precise numerical acceptance criteria and reported values for each test are typically detailed in the full test reports (e.g., TR 20-014, not provided here), which are summarized in this 510(k) document.

    Test CategorySpecific Test (Standard)Acceptance Criteria (General)Reported Device Performance (Summary)
    Mechanical/PhysicalDimensional (ISO 10555-1)Meet specified visual and dimensional requirements.Length: 140 cm. Diameters: 2.01F (0.67mm) proximal, 2.5F (0.84mm) distal outer; 0.019 inches (0.49mm) proximal, 0.028 inches (0.71mm) distal inner. (These values are specific to the RayFlow and indicate its actual dimensions, implicitly meeting the design specification criteria).
    Component Dimension Compatibility (ISO 25539-2)Catheter can track through guiding catheter and over guidewire.Compatible with guiding catheter ID ≥ 0.079" (≥ 5Fr) and guidewire 0.014" in diameter; pressure/temperature guidewire 0.014" in diameter. (Verified compatibility).
    Accessibility/Tractability TestAbility to advance through vessel to target site.The RayFlow catheter is able to advance through a coronary anatomical model (advancement without blockage). Results are documented in TR 20-014.
    Freedom from Leakage (ISO 10555-1 Annex C)No liquid leakage detected at hub.RayFlow catheter showed no sign of leakage. Conforms to ISO 10555-1. Results are documented in TR 20-014.
    Flow Rate (ISO 10555-1)Compliance to catheter standard; specific flow rates/pressures.Outer lumen side holes: 4; Inner lumen side holes: 2. Instruction for use: Maximum flowrate ≤ 25 ml/min (350 PSI). Validated via pre-clinical and clinical data (as stated in the comparison, implying meeting the intended flow characteristics). Conforms to ISO 10555-1. Results are documented in TR 20-014.
    Tensile Strength (ISO 10555-1 Annex B)Meet minimal breaking strength requirements for junctions.Conforms to ISO 10555-1. Minimal breaking strength: ≥ 3N for 0.55 < diam < 0.75; ≥ 5N for 0.75 < diam < 1.15. Results are documented in TR 20-014.
    Material Verification (ISO 10993-18)All materials as specified, validated for biological safety.Hub: Polycarbonate/PEBAX; Strain relief: PEBAX; Hypotube: Stainless steel + PTFE coating; Middle tubing: Polyamide; Inner tubing: PEBAX/LLDPE/Polyethylene; Sleeve: PEBAX; Marker: Platinum/Iridium; Soft tip: PEBAX; Coating: Hydrophilic coating. Biocompatibility testing (see below) confirms biological safety.
    Radiopacity (ASTM F640-20)Sufficient visibility for positioning under fluoroscopy."The radiopacity of the subject device RavFlow marker band will be validated during pre-clinical testing following ISO 10555-1." (Implicitly, the testing confirmed it meets the requirement).
    Flexibility/KinkMaintain integrity and function under bending/kinking.Test performed; no specific results provided but included in the list of tests performed to determine performance and efficacy, implying acceptance criteria were met.
    Torsional Bond StrengthMaintain integrity of bonded components under torsional stress.Test performed; no specific results provided but included in the list of tests performed to determine performance and efficacy, implying acceptance criteria were met.
    Simulated UseFunctional as intended in simulated clinical environment.Test performed; no specific results provided but included in the list of tests performed to determine performance and efficacy, implying acceptance criteria were met.
    CorrosionNo significant degradation over time that affects safety/function.Test performed; no specific results provided but included in the list of tests performed to determine performance and efficacy, implying acceptance criteria were met.
    Acute Particulate EvaluationMinimal particulate release during use.Test performed; no specific results provided but included in the list of tests performed to determine performance and efficacy, implying acceptance criteria were met.
    BiocompatibilityCytotoxicity (ISO 10993-5)Not cytotoxic.Tested; devices were not cytotoxic.
    Sensitization (ISO 10993-10)Not sensitizing.Tested; devices were not sensitizing.
    Irritation/Intracutaneous Reactivity (ISO 10993-10)No irritation/intracutaneous reactivity.Tested; did not induce intracutaneous reactivity.
    Acute Systemic Toxicity (ISO 10993-11)Not acutely toxic.Tested; not acutely toxic.
    Hemocompatibility (ISO 10993-4)Hemocompatible.Tested; hemocompatible.
    Pyrogenicity (ISO 10993-11)Non-pyrogenic.Tested; found non-pyrogenic. Also labeled "PYROGEN FREE".
    Chemical Characterization (ISO 10993-18) / Leachables/Extractables (ISO 10993-17)Characterized chemically, within allowable limits for leachables.Tested; analysis of raw materials and leachables/extractables performed. (Implies satisfactory results to support biological safety).
    SterilizationEthylene Oxide Sterilization (ISO 11135:2014)Sterility Assurance Level (SAL) achieved (Overkill approach).Overkill approach with three consecutive experiments achieving total inactivation of biological indicators (population ≥ 10^6). Results in compliance with ISO 11135:2014.
    Shelf Life/PackagingAccelerated Aging (ASTM F1980-16)Support claimed shelf life.Conducted to support a 3-year shelf life.
    Packaging Validation (ISO 11607-1/2)Maintain sterile barrier, protect device, prevent damage.Sealing process validated. Packaging consists of plastic hoop dispenser, Tyvek pouch (sterile barrier), and cardboard box. (Implies successful validation of packaging integrity and protection).
    Risk ManagementRisk Analysis (ISO 14971:2019)All identified risks mitigated to acceptable levels.Complies with ISO 14971:2019. Testing and risk management activities support that the device is as safe and effective as predicate devices.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact number of units tested for each bench test. However, it indicates that testing was conducted according to standards like ISO 10555-1 and ISO 25539-2. These standards typically define the minimum sample sizes required for verification testing. For instance, sampling for inspection is noted to follow ISO 2859-1:1999 / AMD1:2011, which relates to acceptance quality limit (AQL) for lot-by-lot inspection.
    • Data Provenance: The testing was conducted as "in-vitro tests" (bench testing). The applicant (Hexacath) is based in France. The provenance of the "clinical data" mentioned in the flow rate section is implied to be literature articles ("Literature articles based on the clinical studies conducted on the proposed subject device have been included to support the performance of the subject device.") suggesting existing studies rather than a new clinical trial for this specific 510(k) submission. This is a retrospective literature review for clinical data, combined with prospective bench testing performed by the applicant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This submission is for a traditional medical device (catheter) and relies on physical and performance bench testing against established industry standards (ISO, ASTM) and biocompatibility testing. It does not involve expert-labeled data for "ground truth" as would be required for an AI/ML device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As above, no expert-labeled data requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a physical catheter and does not involve AI or human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical catheter and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for this device's performance are the established performance specifications and acceptance criteria defined by international standards (ISO, ASTM) for medical catheters, and the expected biological response in biocompatibility tests. For example, for "Freedom from leakage," the ground truth is "no liquid leakage." For biocompatibility, it's "not cytotoxic," "not sensitizing," etc.

    8. The sample size for the training set:

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221470

    Validate with FDA (Live)

    Date Cleared
    2022-06-17

    (28 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K170544

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Each Langston dual lumen catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.

    Device Description

    The Langston dual lumen catheters have two lumens, both with distal sideholes, to allow simultaneous pressure measurements from two sites. The high-pressure inner lumen, which extends the entire length of the catheter, can be used for pressurement and rapid delivery of contrast medium. The outer lumen ends proximal to the distal end of the catheter and is used for pressure measurement only. A side port fitted with an extension tube and a stopcock assembly is used for fluid flow and pressure measurement within the outer lumen. The Langston dual lumen catheter will accommodate a standard ≤ 0.038" diameter guidewire and is supplied with a single use pigtail straightener. The Langston dual lumen catheter is sterilized with ethylene oxide.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the Langston dual lumen catheter (K221470). It's a medical device submission, not a study proving the device meets acceptance criteria for an AI or diagnostic algorithm. Therefore, the questions related to AI/algorithm performance, ground truth, expert consensus, and comparative effectiveness studies are not applicable to this document.

    However, I can extract the acceptance criteria and performance data for the physical device based on the provided text.

    Here's the breakdown of what can be found and what cannot:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Bench Testing:
    AspirationMet specified acceptance criteria
    Liquid LeakMet specified acceptance criteria
    Hub to Shaft TensileMet specified acceptance criteria
    Static BurstMet specified acceptance criteria
    Flow RateMet specified acceptance criteria
    Biocompatibility:Complies with ISO 10993-1, leveraging previous testing

    Study that proves the device meets the acceptance criteria:

    The study was a series of bench tests and biocompatibility evaluation to demonstrate substantial equivalence to a predicate device, focusing on a design modification to the strain relief to hub bond.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the document for any of the bench tests.
    • Data Provenance: The tests were performed internally by Vascular Solutions LLC ("Performance Data" section). The country of origin and retrospective/prospective nature of data collection for these specific tests are not detailed, but they are generally considered prospective bench tests performed for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This document concerns a physical medical device (catheter) and its design validation via bench testing and biocompatibility, not an AI or diagnostic algorithm requiring expert "ground truth" for interpretative accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As above, this is about physical device performance, not diagnostic interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For Bench Testing: The "ground truth" was defined by the specified acceptance criteria for each individual physical performance test (e.g., a minimum tensile strength, a maximum leak rate). These criteria are objective measurements, not subjective expert interpretations or biological outcomes.
    • For Biocompatibility: Compliance with ISO 10993-1 is the standard used, which involves a series of tests to assess biological safety.

    8. The sample size for the training set:

    • Not Applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not Applicable. There is no "training set."

    In summary, this document is a regulatory clearance for a physical medical device. The "acceptance criteria" and "performance data" provided relate to the manufacturing and physical characteristics of the catheter, not to the diagnostic accuracy of an algorithm or the improvement of human readers with AI assistance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1