510(k) Data Aggregation

    K Number
    K251440

    Validate with FDA (Live)

    Date Cleared
    2025-08-25

    (108 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160276,K150877

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Chromogenic Factor VIII is for clinical laboratory use in the quantitative determination of factor VIII activity in 3.2 % citrated human plasma. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use.

    Device Description

    CRYOcheck Chromogenic Factor VIII is used for determination of FVIII activity and contains the following four components, packaged in glass vials and provided frozen to preserve the integrity of the components:

    • Reagent 1: Bovine FX and a fibrin polymerization inhibitor, with activators and stabilizer.
    • Reagent 2: Human FIIa, bovine FIXa, calcium chloride and phospholipids.
    • Reagent 3: FXa substrate containing EDTA and a thrombin inhibitor.
    • Diluent Buffer: Tris buffer solution containing 1% BSA and a heparin antagonist.

    In the first stage of the chromogenic assay, test plasma (containing an unknown amount of functional FVIII) is added to a reaction mixture comprised of calcium, phospholipids, human purified thrombin and FIXa, and bovine FX (Reagent 1 and Reagent 2). This mixture swiftly activates FVIII to FVIIIa, which works in concert with FIXa to activate FX. When the reaction is stopped, FXa production is assumed to be proportional to the amount of functional FVIII present in the sample. The second stage of the assay is to measure FXa through cleavage of a FXa-specific peptide nitroanilide substrate (FXa Substrate). P-nitroaniline is produced, giving a color that can be measured spectrophotometrically by absorbance at 405 nm.

    AI/ML Overview

    Based on the provided FDA 510(k) Clearance Letter, the device in question is the CRYOcheck Chromogenic Factor VIII. This document details the clearance of a modified version of an existing device, emphasizing the differences from the previous version regarding interference claims and recovery of Factor VIII replacement therapies.

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for each performance claim in a quantified manner (e.g., "Interference must be less than X%"). Instead, it reports the limits of non-interference found in their studies, implying these served as the de facto acceptance criteria. For the Factor VIII replacement therapy recovery, the acceptance criterion appears to be "accurate evaluation" across a range of concentrations, with specific over/under recovery noted.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Interference:
    HemoglobinMust show no interference up to the concentration indicated.No interference observed up to ≤1000 mg/dL (increased from ≤500 mg/dL)
    IntralipidMust show no interference up to the concentration indicated.No interference observed up to ≤830 mg/dL (increased from ≤500 mg/dL)
    Bilirubin (unconjugated)Must show no interference up to the concentration indicated.No interference observed up to ≤40 mg/dL (increased from ≤29 mg/dL)
    Bilirubin (conjugated)Must show no interference up to the concentration indicated.No interference observed up to ≤11 mg/dL (increased from ≤2 mg/dL)
    von Willebrand factorMust show no interference up to the concentration indicated.No interference observed up to ≤20 µg/mL (same)
    Unfractionated heparinMust show no interference up to the concentration indicated.No interference observed up to ≤3.3 IU/mL (increased from ≤2 IU/mL)
    Low molecular weight heparinMust show no interference up to the concentration indicated.No interference observed up to ≤5 IU/mL (increased from ≤2 IU/mL)
    FondaparinuxMust show no interference up to the concentration indicated.No interference observed up to ≤0.2 mg/L (decreased from ≤1.25 mg/L)
    Lupus AnticoagulantMust show no interference up to the concentration indicated.No interference observed up to ≤1.8 dRVVT ratio (same)
    EmicizumabMust show no interference up to the concentration indicated.No interference observed up to ≤150 µg/mL (new claim)
    Mim8Must show no interference up to the concentration indicated.No interference observed up to ≤8 µg/mL (new claim)
    WarfarinMust show no interference up to the concentration indicated.No interference observed up to INR ≤7 (new claim)
    RivaroxabanMust not interfere.Interfered with quantification of FVIII activity.
    DabigatranMust not interfere.Interfered with quantification of FVIII activity.
    Recovery of FVIII Replacement Therapy:Must accurately evaluate potency.Accurately evaluated potency for ADVATE, ADYNOVATE, AFSTYLA, ALTUVIIO, ESPEROCT, HUMATE-P, JIVI, KOVALTRY, Novoeight, Nuwiq, and wilate at 0.05-1.0 IU/mL; ELOCTATE, and XYNTHA at 0.05-0.6 IU/mL (with over recovery at 0.8 & 1.0 IU/mL); Underestimation for OBIZUR.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Interference Studies: Plasma samples were "spiked with possible interferents," and "10 replicates were tested alongside 10 replicates of the corresponding blank matrix control." The total number of individual patient samples from which this plasma was derived is not specified, nor is the country of origin. The study design implies a prospective spiking experiment in a laboratory setting.
    • Recovery of Factor VIII Replacement Therapy: "Congenital FVIII deficient plasma was spiked with 14 FVIII replacement therapies at seven concentrations." The number of individual patient plasma units or lots of deficient plasma used is not specified. The study design implies a prospective spiking experiment in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This is an in vitro diagnostic device for quantitative determination of factor VIII activity, not an AI/imaging device requiring expert human readers for ground truth generation. The ground truth for these studies is established by the known concentrations of spiked interferents or FVIII replacement therapies, and the intrinsic properties of the FVIII deficient plasma.

    4. Adjudication Method for the Test Set

    N/A. As this is a quantitative in vitro diagnostic device, an adjudication method in the context of human expert review of imaging or clinical data is not applicable. The results are measured spectrophotometrically.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done. This type of study is relevant for AI imaging devices where human readers interpret medical images with and without AI assistance. This document describes an in vitro diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Yes, this entire submission describes the standalone performance of the CRYOcheck Chromogenic Factor VIII assay. The device itself performs the quantitative determination of FVIII activity, entirely without a "human-in-the-loop" once the sample is loaded and the assay run according to protocol.

    7. The Type of Ground Truth Used

    • Interference Studies: The ground truth was the known concentration of the spiked interferent (e.g., Hemoglobin, Intralipid, Bilirubin, etc.) added to plasma samples, and the corresponding blank matrix control.
    • Recovery of Factor VIII Replacement Therapy: The ground truth was the known concentration of the spiked FVIII replacement therapy added to congenital FVIII deficient plasma at various concentrations.

    8. The Sample Size for the Training Set

    N/A. This document describes an in vitro diagnostic assay based on chromogenic principles, not an AI/ML algorithm that requires a "training set" in the computational sense. The device's components (reagents, diluent buffer) and their interaction define the assay, which is then validated through performance studies.

    9. How the Ground Truth for the Training Set was Established

    N/A. See point 8. The "ground truth" for developing and optimizing such a chromogenic assay would stem from extensive biochemical research, characterization of reagents, and titrations against known standards, which is inherent in the development of any diagnostic assay, but not referred to as a "training set" or "ground truth establishment" in the AI/ML context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222831

    Validate with FDA (Live)

    Date Cleared
    2023-09-13

    (359 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160276, K150877

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Factor VIII Deficient Plasma with VWF is for clinical laboratory use as a deficient substrate in the quantitative determination of Factor VIII activity in 3.2% citrated human plasma based on the activated partial thromboplastin time (APTT) assay. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use

    Device Description

    CRYOcheck Factor VIII Deficient Plasma with VWF is normal human citrated plasma which has been immunodepleted of factor VIII and to which an exogenous source of human von Willebrand Factor (vWF) has been added and buffered with HEPES. Factor VIII has been assayed at less than 1% of normal activity levels and vWF antigen and activity are >50%. It will be provided to users frozen in small-volume aliquots (25 vials of 1.0 mL, and 25 vials of 1.5 mL). Vials will be packaged into boxes; these will be frozen during the manufacturing process and will be shipped and stored frozen until use to preserve the integrity of the components

    AI/ML Overview

    The provided FDA 510(k) summary for the "CRYOcheck Factor VIII Deficient Plasma with VWF" describes various performance studies and their results. Based on the document, here's a structured breakdown of the acceptance criteria and the study details:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of pre-defined acceptance criteria for each performance characteristic in a pass/fail format. Instead, it presents the results of various studies and implies that these results met internal or regulatory expectations for substantial equivalence to the predicate device. However, we can infer performance targets based on the documented results and common regulatory expectations for in vitro diagnostic devices.

    Here's a table summarizing the reported device performance, with inferred acceptance "criteria" based on the conclusions drawn in the report:

    Performance CharacteristicInferred Acceptance Criteria (Based on reported success)Reported Device Performance
    Precision (Multi-Reagent Lot)<10% CV for controls and high FVIII samples; <0.5 SD for low/very low FVIII samples (implied as 'acceptable')Aggregated Data (Lots 1, 2, and 3): - CRYOcheck Reference Control Normal: 5.7% CV - CRYOcheck Abnormal 1 Reference Control: 6.2% CV - CRYOcheck Abnormal 2 Reference Control: 5.2% CV - High FVIII Plasma Sample: 6.0% CV - Low FVIII Plasma Sample: 0.2 SD (11.1% CV) - Very Low FVIII Plasma Sample: 0.0 SD (24.1% CV) (Note: 24.1% CV for "Very Low" indicates higher variability, but the SD is 0.0, suggesting it met the SD criterion for this range)
    Reproducibility (Multi-Reagent Lot Site-to-Site)<10% CV for controls and high FVIII samples; <0.5 SD for low/very low FVIII samples (implied as 'acceptable')Pooled 3-Site Data: - Reference Control Normal: 6.04% CV - Abnormal 1 Reference Control: 4.75% CV - Abnormal 2 Reference Control: 7.68% CV - High FVIII Plasma Sample: 5.07% CV - Low FVIII Plasma Sample: 0.18 SD (9.39% CV) - Very Low FVIII Plasma Sample: 0.04 SD (38.56% CV) (Note: 38.56% CV for "Very Low" indicates higher variability, but the SD is 0.04, suggesting it met the SD criterion for this range)
    Linearity/Assay Reportable RangeDemonstration of linearity across the claimed reportable range (implied as 'acceptable' based on graphical analysis or statistical fit).Linearity Range: 0 to 230% FVIII activity. The study concluded this range is supported.
    Reference IntervalEstablishment of a clinically meaningful and statistically derived reference interval from a healthy population.Reference Interval: 62 to 163 % FVIII activity (calculated as the non-parametric 95% confidence interval, 2.5th to 97.5th percentiles).
    Shelf-Life StabilityMaintenance of FVIII activity within acceptable limits over the claimed shelf-life.A shelf-life stability claim of 24 months when stored at -40 °C to -80 °C, based on data up to 25 months (real-time study ongoing).
    In-Use StabilityMaintenance of FVIII activity within acceptable limits for the claimed on-board and refrigerated stability periods.A stability claim of 24 hours on board the instrument and at 2-8 °C.
    InterferencesNo significant interference from common substances at specified concentrations.No interference from Hemoglobin (≤1000 mg/dL), Intralipid (≤2000 mg/dL), Bilirubin (conjugated, ≤4.0 mg/dL), Bilirubin (unconjugated, 40 mg/dL), Lupus Anticoagulant (≤1.8 dRVVT ratio), Warfarin (≤ INR ratio 2.72). Certain anticoagulants (Rivaroxaban, fondaparinux, dabigatran, emicizumated heparin, and low molecular weight heparin) were shown to interfere.
    Recovery of FVIII ReplacementsAccurate evaluation (100 ± 25%) of the potency of specified FVIII replacement therapies.Accurately evaluated (100 ± 25%) Advate (92.84%), Afstyla* (97.38% with 2x correction factor), Elocate (94.90%), Jivi (104.49%), Novoeight (113.33%), and Wilate (94.92%) at concentrations 0.05 to 1.0 IU/mL.
    Method ComparisonDemonstrating substantial equivalence to the predicate device via Passing-Bablok regression (e.g., slope close to 1, intercept close to 0, high correlation coefficient).Overall (N=366): Slope = 1.16 (CI: 1.15 to 1.18), Intercept = -0.08 (CI: -0.25 to 0.13), Pearson Correlation Coefficient (r) = 0.96. Predicted biases at medical decision levels were reported (e.g., 0.37% at 1% FVIII, 13.70% at 100% FVIII). The study concluded it performed equivalently.
    Sample IntegritySupport for claimed fresh sample stability and frozen storage stability, including freeze-thaw cycles.Supported a fresh sample stability claim of 2 hours at room temperature and a frozen storage claim of 1.5 months at ≤-70 ℃, including up to two freeze-thaw cycles.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Multi-Reagent Lot Precision: Not explicitly stated as a "test set" sample size for patient plasma, but control samples were tested 80 replicates per sample per lot across 3 lots.
    • Multi-Reagent Lot Site to Site Reproducibility: 270 replicates per sample (control or plasma) for the pooled data, across 3 sites and 3 lots. Patient samples included high, low, and very low FVIII plasma.
    • Linearity/Assay Reportable Range: Fifteen sample dilutions, each tested in quadruplicate.
    • Reference Interval: 136 normal, ostensibly healthy individuals.
    • Shelf-Life Stability: Five replicates of one normal and two abnormal reference controls and two patient plasma samples per timepoint, across 3 lots.
    • In-Use Stability: Five replicates of one normal and two abnormal reference controls and two patient plasma samples per timepoint, across 3 lots.
    • Interferences: 10 replicates of spiked samples and 10 replicates of control for each possible interferent.
    • Recovery of FVIII Replacements: Congenital FVIII deficient plasma spiked with six FVIII replacement therapies at seven concentrations (total 42 conditions).
    • Method Comparison: 366 human plasma samples (from normal ostensibly healthy individuals, patients with congenital or acquired hemophilia A, patients with hemophilia B, patients with von Willebrand disease, hemophilia A patients on FVIII replacement therapies, and patients with other factor deficiencies). These samples were distributed across four sites.
    • Sample Integrity: 94 plasma samples.

    Data Provenance: The document does not specify the country of origin for any of the patient samples. All studies appear to be prospective in design, as they describe the testing of samples under controlled conditions to evaluate device performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is an in vitro diagnostic (IVD) reagent for determining Factor VIII activity in human plasma, which is a quantitative measurement, not an interpretative one like image analysis. Therefore, the concept of "experts establishing ground truth" in the way it applies to diagnostic algorithms (e.g., radiologists for imaging) is not directly relevant here.

    The "ground truth" for the performance studies is established by:

    • The known concentrations/characteristics of reference controls.
    • The actual FVIII activity levels as measured by a legally marketed predicate device (HemosIL Factor VIII Deficient Plasma) in the method comparison study.
    • The carefully prepared spiked samples for linearity and recovery studies.
    • The clinical diagnosis categories of patient samples (e.g., congenital hemophilia A).

    There is no mention of human experts (e.g., pathologists, hematologists) adjudicating "ground truth" for individual cases in these studies beyond ensuring appropriate sample selection and handling. The performance is assessed against quantitative benchmarks and comparison with a known predicate.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    As mentioned above, the device performs a quantitative measurement, not an interpretative diagnosis requiring adjudication. Therefore, no adjudication method (like 2+1 or 3+1 for expert consensus on image interpretation) was used or is applicable for this type of IVD device testing. The "ground truth" is analytical and quantitative.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC comparative effectiveness study is not applicable as this is an in vitro diagnostic reagent, not an AI-powered diagnostic tool that assists human readers. The device itself performs the measurement, and its performance is evaluated directly through analytical studies, not through its impact on human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device itself (the reagent in combination with instrumentation) without human interpretation affecting the measurement result. All the analytical performance studies (precision, reproducibility, linearity, stability, interferences, recovery, method comparison, sample integrity) represent the standalone performance of the CRYOcheck Factor VIII Deficient Plasma with VWF in conjunction with the specified automated APTT assay systems. The "human-in-the-loop" here refers to the user performing the test, but the performance characteristics relate to the analytical accuracy and reliability of the automated test method itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth used in these studies is primarily analytical and quantitative:

    • Known concentrations/values of commercially available reference controls.
    • Known FVIII activity levels measured by a predicate device for method comparison.
    • Expected FVIII activity levels in carefully fabricated samples (e.g., highly concentrated plasma diluted with congenital hemophilia A plasma for linearity, spiked samples for recovery).
    • Clinical classification of patient samples (e.g., "patients with congenital or acquired hemophilia A") as the basis for sample selection in the method comparison and reference interval studies, though specific individual diagnoses are not the "ground truth" for each quantitative measurement, rather the context.

    8. The sample size for the training set

    This document describes a premarket notification (510(k)) for an in vitro diagnostic reagent. IVD reagents are not typically "trained" in the machine learning sense. Therefore, there is no training set in the context of an AI/ML algorithm. The studies detailed are analytical verification and validation studies to demonstrate performance and establish substantial equivalence to a predicate device.

    9. How the ground truth for the training set was established

    Since there is no training set for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K214002

    Validate with FDA (Live)

    Date Cleared
    2022-12-23

    (367 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160276,K150877

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Chromogenic Factor IX is for clinical laboratory use in the quantitative determination of factor IX activity in 3.2% citrated human plasma. It is intended to be used in identifying factor IX deficiency and as an aid in the management of hemophilia B in individuals aged 2 years and older. For in vitro diagnostic use.

    Device Description

    CRYOcheck Chromogenic Factor IX is used for determination of FIX activity and contains the following four components, packaged in vials, and provided frozen to preserve the integrity of the components:
    Reagent 1: Human FVIII, human FX, bovine FV and a fibrin polymerization inhibitor.
    Reagent 2: Human FXIa, human FII, calcium chloride and phospholipids
    Reagent 3: FXa Substrate containing EDTA and a thrombin inhibitor.
    Diluent Buffer: Tris buffer solution containing 1% BSA and a heparin antagonist.

    AI/ML Overview

    The provided text describes the performance of the CRYOcheck Chromogenic Factor IX device. Here's a breakdown of the acceptance criteria and the study that proves the device meets these criteria, based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria with pass/fail thresholds for each performance characteristic. Instead, it presents study results and concludes that the device performed effectively and demonstrates substantial equivalence. However, we can infer some criteria from the performance data presented.

    Performance CharacteristicReported Device Performance (Implicit Acceptance)
    Multi-Reagent Lot PrecisionWithin-Laboratory CV: Reference Control Normal (3.7%), Abnormal 1 (4.5%), Abnormal 2 (7.3%). Very Low FIX Plasma (14.5%), Low FIX Plasma (10.1%), High FIX Plasma (3.7%). These CVs are generally considered acceptable for diagnostic assays.
    Multi-Reagent Lot ReproducibilityAcross-Site CV: Reference Control Normal (5.6%), Abnormal 1 (6.6%), Abnormal 2 (8.6%). Very Low FIX Plasma (15.7%), Low FIX Plasma (13.0%), High FIX Plasma (6.0%). These CVs indicate good reproducibility across different sites and instruments.
    LinearityLinearity Range: 0 to 200% FIX activity. The study results are reported to "support the linearity claim."
    Reference IntervalReference Interval: 79 to 155% FIX activity. Established by calculating the non-parametric 95% confidence interval (2.5th to 97.5th percentiles) from 128 normal individuals.
    Shelf-Life StabilityShelf-Life Stability: At least 18 months when stored at ≤-70 ℃. The study was completed up to 19 months.
    In-Use StabilityOn-Board Stability: 24 hours on board the instrument. Refrigerated Stability: 48 hours at 2-8 °C. Refrozen Stability: One month at ≤-70 ℃ if stored on-board and refrozen within 4 hours, and used within eight hours of next thawing while kept on-board.
    Detection LimitLimit of Blank (LoB): 0.4% FIX activity. Limit of Detection (LoD): 0.5% FIX activity. Limit of Quantitation (LoQ): 0.5% FIX activity. These values indicate the assay's ability to detect and quantify low levels of FIX activity.
    InterferencesNo Interference: Hemoglobin (≤ 1000 mg/dL), Intraplipid (≤ 2000 mg/dL), Bilirubin (unconjugated ≤ 40 mg/dL), Bilirubin (conjugated ≤ 23 mg/dL), Unfractionated heparin (≤ 1.2 IU/mL), Low molecular weight heparin (≤ 1.5 IU/mL), Dabigatran (≤ 0.04 mg/L), Fondaparinux (≤ 0.26 mg/L), Lupus Anticoagulant (≤ 1.8 dRVVT ratio). Interference: Rivaroxaban and warfarin. This indicates the range of substances that do not affect the assay result.
    Recovery of FIX ReplacementsMean Percent Recovery: AlphaNine SD (96%), Alprolix (116%), BeneFIX (93%), Ixinity (82%), Rebinyn (117%), Rixubis (102%). Overestimation: Idelvion (153%). The recoveries demonstrate the device's ability to evaluate the potency of most FIX concentrates.
    Method ComparisonPearson Correlation Coefficient: Ranging from 0.979 to 0.996 across sites, with an overall of 0.992 (r2=0.983). Passing-Bablok Regression: Slopes ranging from 1.05 to 1.21, intercepts from -11.76 to 2.44, and overall slope of 1.10 and intercept of 0.64. The study concludes that the device "performed equivalently to the comparator method."
    Sample IntegrityFresh Sample Stability: 4 hours at room temperature. Frozen Storage Stability: 3 months at ≤-70 °C, including up to two freeze-thaw cycles.
    Overall ConclusionThe performance testing results demonstrate that CRYOcheck Chromogenic FIX is substantially equivalent to the predicate device and the comparator assay, and that the assay is effective for its labeled intended use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Multi-Reagent Lot Precision: 6 samples (1 normal reference control, 2 abnormal reference controls, 3 patient plasma samples).
    • Multi-Reagent Lot Site to Site Reproducibility: 6 samples (1 normal reference control, 2 abnormal reference controls, 3 patient plasma samples).
    • Linearity/Assay Reportable Range: 14 sample dilutions.
    • Reference Interval: 128 normal, ostensibly healthy individuals.
    • Shelf-Life Stability: Not specified, but likely involved multiple reference controls and patient plasma samples at each time point.
    • In-Use Stability: Not specified, but involved multiple reference controls and patient plasma samples.
    • Detection Limit (LoB): 4 blank plasma samples from individuals with severe congenital hemophilia B.
    • Detection Limit (LoD): 4 plasma samples with low FIX activity from congenital hemophilia B donors.
    • Detection Limit (LoQ): 4 plasma samples with low FIX activity from congenital hemophilia B donors.
    • Interferences: Not specified, but likely involved spiked plasma samples.
    • Recovery of FIX Replacements: Congenital FIX deficient plasma spiked with 7 FIX replacement therapies at 7 concentrations.
    • Method Comparison Studies: 368 human plasma samples. These samples were from normal ostensibly healthy individuals, patients with von Willebrand disease, patients with congenital and acquired hemophilia A and B, and patients on recombinant factor IX treatments.
    • Sample Integrity: 65 plasma samples.

    Data Provenance: The document does not explicitly state the country of origin for the patient or donor samples. The studies involve both internal and external sites for reproducibility and method comparison. The mention of "congenital hemophilia B donors" and "patients with von Willebrand disease, and congenital and acquired hemophilia A and B" suggests the use of patient samples, implying a clinical or diagnostic context. The studies are retrospective in nature as they involve testing existing plasma samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For the studies described, the "ground truth" for Factor IX activity is established by:

    • Reference controls: These are quality control materials with known or assigned values for FIX activity.
    • Patient plasma samples: Their FIX activity is determined either by the device itself (for precision and reproducibility) or by a "comparator method" or "validated laboratory developed chromogenic factor IX assay" (for method comparison and LoQ studies).
    • Spiked samples: For linearity, interference, and recovery studies, known amounts of FIX or interfering substances are added to samples to create a controlled "ground truth."
    • Congenital FIX deficient plasma: For LoB, LoD, and recovery studies, plasma from individuals with a known severe deficiency serves as a baseline.

    The document does not mention specific "experts" in the context of radiologists or similar roles establishing ground truth through consensus. Instead, the ground truth is based on:

    • Established assay methods: The comparator method and the validated laboratory developed chromogenic factor IX assay are implicitly considered accurate for determining FIX activity. The qualifications of the personnel running these assays are not individually listed but are assumed to be trained laboratory professionals.
    • Reference materials: Reference controls have assigned values.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1) is typically relevant for studies where human expert interpretation is compared, e.g., in medical image analysis. In this context of an in vitro diagnostic device for quantitative determination in plasma, such an adjudication method is not described or applicable. The "ground truth" is based on the results of the reference methods, reference materials, and defined sample preparations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is not relevant for an in vitro diagnostic device like the CRYOcheck Chromogenic Factor IX, which directly measures a biomarker rather than assisting human readers in interpreting complex data like medical images. The device itself is the "reader" providing a quantitative result.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented are all standalone (algorithm only) performance assessments for the CRYOcheck Chromogenic Factor IX device. The device itself performs the quantitative determination of factor IX activity. The results are generated by the instrument (IL ACL TOP Series or TOP 50 Series Instruments) based on the reagents and the plasma sample, without human interpretative input in the output. Human operators are involved in running the tests, but not in interpreting the quantitative output in a way that typical human-in-the-loop AI studies would assess.

    7. The Type of Ground Truth Used

    The ground truth used in these studies includes:

    • Assigned values of reference controls: For precision and reproducibility.
    • Results from a "comparator device" or "validated laboratory developed chromogenic factor IX assay": For method comparison and limit of quantitation. These are considered gold standards for FIX activity measurement.
    • Known concentrations in spiked samples: For linearity, interference, and recovery studies.
    • Known characteristics of congenital FIX deficient plasma: For detection limits and recovery studies.
    • Plasma samples from "normal, ostensibly healthy individuals": For establishing reference intervals.

    This primarily falls under the category of established assay methods and reference materials.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI. The CRYOcheck Chromogenic Factor IX is an in vitro diagnostic assay, not an AI/ML algorithm that requires training data. Its performance is based on chemical and enzymatic reactions, and its parameters are established through conventional analytical validation studies like those described.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the AI/ML sense for this device, this question is not applicable. The device's operational parameters and performance are intrinsically defined by its design, chemical components, and the analytical validation tests presented.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN200067

    Validate with FDA (Live)

    Device Name
    INNOVANCE VWF Ac
    Date Cleared
    2022-09-29

    (701 days)

    Product Code
    Regulation Number
    864.7293
    Type
    Direct
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    In-vitro diagnostic automated assay for the quantitative determination of the von Willebrand factor-GPIb-binding activity in human plasma collected from venous blood samples in 3.2% sodium citrate tubes on the BCS XP System.

    As an aid used in the evaluation of patients with suspected or confirmed von Willebrand factor disorders.

    Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation and other laboratory findings.

    Device Description

    The INNOVANCE VWF Ac assay is a particle enhanced turbidimetric assay based on binding of VWF to the recombinant GPIb (two gain-of-function mutations included). The assay provides quantitative VWF activity results on 3.2% citrated human plasma when used with Standard Human Plasma Calibrators (K023141).

    The reagent kit consists of three components: Reagent I (containing polystyrene particles coated with anti-GPIb mouse monoclonal antibodies). Reagent II (buffer containing heterophilic blocking reagent) and Reagent III (containing recombinant GPIb).

    AI/ML Overview

    The INNOVANCE VWF Ac assay is a device for the quantitative determination of von Willebrand factor-GPIb-binding activity. The device is intended to aid in the evaluation of patients with suspected or confirmed von Willebrand factor disorders.

    Here's an analysis of its acceptance criteria and supporting studies:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Precision/Reproducibility- Single Site (internal): Total CVs (Coefficient of Variation) should meet predefined acceptance criteria.- Study 1 (Single Site, Germany): Total CVs ranged from 2.90% to 6.02% (reagent variability) and 2.95% to 5.67% (calibrator variability). All values for single-site precision were within acceptable limits. - Study 2 (Multi-site reproducibility): Total CVs ranged from 3.48% to 6.29%. All values for multi-site reproducibility were within acceptable limits. - Study 3 (Instrument/operator variability): Total CVs ranged from 2.86% to 6.01%. All values were within acceptable limits.
    Linearity- Assay should demonstrate linearity across the claimed reportable range (4 to 300% of norm).- Linearity studies (using spiked and native samples) demonstrated linearity from 4 to 300% of norm on the BCS XP System.
    Analytical Specificity/Interference- No clinically significant interference from specified endogenous or exogenous substances at tested concentrations.- Endogenous Interference: No clinically significant interference from Hemoglobin (up to 1000 mg/dL), Bilirubin (unconjugated) (up to 60 mg/dL), Bilirubin (conjugated) (up to 40 mg/dL), Triglycerides (up to 726 mg/dL), and Rheumatoid Factors (up to 438 IU/mL). - Exogenous Interference: No clinically significant interference from a panel of 30 common drugs tested at specific concentrations. - Heterophilic blocking reagent included to minimize interference from heterophile antibodies.
    High Dose Hook- No high dose hook effect observed up to a specified VWF activity level.- No high dose hook effect observed up to a VWF activity of 656% of norm.
    Traceability- INNOVANCE VWF Ac results should be traceable to the WHO 6th International Standard (IS) Factor VIII / Von Willebrand Factor.- INNOVANCE VWF Ac results are traceable to the WHO 6th International Standard (IS) Factor VIII / Von Willebrand Factor (NIBSC code 07/316), with an observed bias of +2.3% (relative).
    Stability (Shelf-life, In-use, Reconstituted, Sample, Transportation, Freeze/Thaw)- Reagent and calibrator shelf-life stability. - On-board stability. - Once opened stability. - Reconstituted stability (calibrator). - Ambient temperature operating range. - Transportation stability. - Freeze/thaw tolerance for reagent and calibrator. - Sample stability under various storage conditions. - Equivalence between fresh and frozen samples.- Reagent Shelf-Life: 12 months at 2-8°C. - Calibrator Shelf-Life: 12 months at 2-8°C. - On-Board Stability (Reagent): 36 hours at 18-32°C. Accumulated on-board stability: 24 hours for Reagents I & II, 36 hours for Reagent III. - On-Board Stability (Calibrator): 6 hours at 18-32°C (not recommended in labeling due to immediate use intent). - Once Opened Stability (Reagent): 37 days for Reagents I & II, 113 days for Reagent III when stored at 2-8°C. - Reconstituted Stability (Calibrator): 4 hours at 15-25°C; 4 weeks at -20°C; 2 hours at 15-25°C after freeze/thaw. - Ambient Temperature: Correctness assured within 18-32°C operating range. - Transportation Stability: All tested conditions met acceptance criteria. - Freeze/Thaw Tolerance: Stable for one freeze/thaw cycle for both reagent and calibrator. - Sample Stability: 3 months at ≤ -20℃; 12 months at < -74℃; 4 hours at 15-25°C (plasma on/off cells); 4 hours at 15-25°C for once frozen plasma. - Frozen vs. Fresh Samples: Comparability demonstrated, meeting acceptance criteria based on Passing-Bablok regression and Bland-Altman plots.
    Detection Limits (LoB, LoD, LoQ)- Determined following CLSI EP17-A2 guideline.- LoB: 1.131% of norm. - LoD: 1.738% of norm. - LoQ: 3.72% of norm.
    Normal Range / Reference Interval- Established for different subgroups (blood group O, non-O, gender-specific).- Combined ABO: 50.7 to 203.5% of norm. - Blood group O: 49.0 to 178.9% of norm. - Blood group non-O: 60.7 to 214.1% of norm. - Female: 55.4 to 181.3% of norm. - Male: 48.8 to 210.7% of norm. - Pediatric study found values between 37.4 and >300.0% of norm (not for establishing reference interval).
    Carry-Over (Sample & Reagent)- No significant carryover from one sample to the next or from one reagent application to another.- Sample Carryover: No carryover observed. - Reagent Carryover: No cross-contamination observed.
    Diagnostic Accuracy (Method Comparison)- Acceptable comparability to predicate devices (BC von Willebrand Reagent, HemosIL VWF). - Passing-Bablok analysis results (slope, intercept, predicted bias, Pearson correlation coefficient) should meet predefined acceptance criteria.- Vs. BC von Willebrand Reagent (current study): Combined sites (N=102) showed Pearson r = 0.916, Slope = 1.037, Intercept = 3.497. Predicted bias at MDP1 (30% norm) = 4.60% norm; at MDP2 (50% norm) = 10.14% (relative). All met acceptance criteria. - Vs. HemosIL VWF: Combined sites (N=97) showed Pearson r = 0.971, Slope = 0.955, Intercept = -0.552. Predicted bias at MDP1 (30% norm) = -1.90% norm; at MDP2 (50% norm) = -5.76% (relative). All met acceptance criteria. - Vs. BC von Willebrand Reagent (Patzke et al., 2014 re-analysis): Combined sites (N=556, 4-300% of norm) showed Pearson r = 0.985, Slope = 0.955, Intercept = 1.110. Predicted bias at MDP1 (30% norm) = -0.23% norm; at MDP2 (50% norm) = -2.26% (relative). All met acceptance criteria and confirmed acceptable comparability.
    Diagnostic Concordance- Overall percent agreement, positive percent agreement (any type of VWD), and negative percent agreement (VWD excluded) should demonstrate acceptable concordance with existing standard of care (SOC) VWF-activity assays.- Overall Percent Agreement: 83.33% (95% CI: 76.23 - 88.63). - Positive Percent Agreement ('any type' VWD): 74.29% (95% CI: 57.93 - 85.84). - Negative Percent Agreement ('VWD excluded'): 95.77% (95% CI: 88.30 - 98.55). - "Overall percentage agreement of more than 80%… demonstrates that for the majority of patients (> 80%), use of the subject device instead of currently available assays to measure VWF activity will not alter diagnosis." The PPA at low VWF (52.63%) was accepted with clinical justifications.

    2. Sample sizes used for the test set and data provenance:

    • Precision (Single Site):

      • Study 1 (reagent variability): 5 plasma pools + 2 control materials = 7 samples. Each tested 240 times (20 days x 2 runs x 2 replicates x 3 reagent lots or 3 calibrator lots for a total of 240 determinations, 80 per lot).
      • Study 2 (reproducibility multi-site): 5 plasma pools + 2 control materials = 7 samples. Each tested 90 times (3 external sites x 5 days x 2 runs x 3 replicates for a total of 90 determinations, 30 per site).
      • Study 3 (instrument/operator variability): 5 plasma pools + 2 control materials = 7 samples. Each tested 120 times (5 days x 2 runs x 4 replicates x 3 systems for a total of 120 determinations, 40 per system).
      • Provenance: Single site precision studies were conducted in Germany. Multi-site reproducibility and instrument/operator variability studies were conducted across three external sites (locations not specified for multi-site reproducibility study, but internal site for instrument/operator variability implies one location).
    • Linearity (Test Sets):

      • Spiked Sample Linearity: 10 different concentrations of spiked samples.
      • Native Sample Linearity: 12 different concentrations of native samples.
      • Provenance: High pool for spiked samples prepared by spiking normal plasma with VWF concentrate. Low pool with VWF deficient plasma (internally produced). Native samples from contract blood plasma provider. Locations not specified but implied to be part of the testing facilities.
    • Analytical Specificity/Interference (Test Sets):

      • Endogenous Interference: 4 VWF activity levels (low, 2 medical decision levels, high) tested with 5 test samples each. 80 measurements overall (4 VWF levels x 5 test samples x 4 replicates). Each sample prepared with native plasma and dilution/spiking.
      • Exogenous Interference: 4 VWF activity levels for each of the 30 interferent drugs. 32 measurements per interferent (4 VWF levels x 4 aliquots x 2 conditions - spiked/control).
      • Provenance: Samples prepared with native plasma and VWF deficient plasma or VWF concentrate.
    • High Dose Hook (Test Set): 10 different dilution samples from a high VWF Ac activity plasma pool. Measured in 6 replicates.

      • Provenance: High pool prepared by spiking normal plasma with VWF concentrate, diluted with VWF deficient plasma.
    • Traceability (Test Set): Three vials of WHO 6th International Standard for Factor VIII / von Willebrand Factor. Measured in single determination.

    • Stability (Test Sets):

      • Reagent/Calibrator Shelf-Life: 5 plasma pools + 3 control materials. Measured 6-12 replicates at each of 6 time points over 15 months.
      • On-Board Stability: 5 plasma pools + 3 control materials. Measured 6 replicates at each of 5 time points.
      • Once Opened Stability: 5 plasma pools + 3 control materials. Measured 6-12 replicates at multiple time points (2,4,6 weeks for reagent I/II; 6,12,18 weeks for reagent III).
      • Reconstituted Stability (Calibrator): Aliquots of Standard Human Plasma (SHP) measured 6-12 replicates at various time points and conditions.
      • Ambient Temperature: 8 test samples. Each tested in 8 replicates on 3 days.
      • Transportation Study: 5 plasma pools + 3 control materials. Tested 3-6 replicates at various temperature stress conditions.
      • Freeze/Thaw Tolerance: 5 plasma pools + 3 control materials. Tested 3 replicates under stressed and unstressed conditions.
      • Sample Stability: At least 20 samples. Tested in quadruplicate over various time points and storage conditions.
      • Frozen vs. Fresh Samples: 60 fresh samples. Measured one replicate fresh, then one replicate after frozen storage.
      • Provenance: Plasma pools, control materials, patient samples. Specific origins (e.g., country) not detailed for all stability studies, but likely conducted at manufacturer's internal labs or controlled external sites.
    • Detection Limit (Test Sets):

      • LoB: 5 analyte-free samples. n=60 determinations per reagent lot per instrument.
      • LoD: 5 low-analyte samples. n=50 determinations per sample.
      • LoQ: 5 low-analyte samples. n=60 determinations per reagent lot per instrument.
      • Provenance: Samples prepared with VWF deficient plasmas and dilutions of normal plasma pools.
    • Normal Range / Reference Interval (Test Set):

      • Adults: 302 apparently healthy individuals (150 blood group O, 152 blood group non-O) ≥ 18 years of age.
      • Pediatrics: 85 apparently healthy pediatric individuals (44 blood group O, 41 blood group non-O) > 4 weeks to < 18 years of age.
      • Provenance: Samples collected retrospectively/prospectively at three clinical sites in the U.S. at different geographic locations (for adults) and two U.S. clinical sites (for pediatrics).
    • Diagnostic Accuracy (Method Comparison) (Test Set):

      • Vs. BC von Willebrand Reagent (current study): 144 samples (17 fresh, 124 frozen, 3 diluted). 102 samples included in statistical evaluation after filtering for AMI.
      • Vs. HemosIL VWF: 144 samples (17 fresh, 124 frozen, 3 diluted). 97 samples included in statistical evaluation after filtering for AMI.
      • Vs. BC von Willebrand Reagent (Patzke et al., 2014 re-analysis): 618 samples initially; 580 samples (29 diluted, no spiked) included in statistical evaluation.
      • Provenance: Patient samples from the U.S. (for current studies) and Germany, Switzerland, and USA (for Patzke et al. re-analysis). Patient cohort included those with suspected/confirmed VWD, hemophilia A, platelet dysfunction, and without VWD diagnosis.
    • Diagnostic Concordance (Test Set): 138 patients (108 tested against Beckman Coulter vWF Reagent, 30 against HemosIL VWF Activity Reagent).

      • Provenance: Not explicitly stated but implies patient samples from clinical settings. Likely U.S. based on regulatory context.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    The device is an in-vitro diagnostic assay that measures a quantitative biomarker (VWF activity). The "ground truth" for several test sets, such as precision, linearity, interference, and detection limits, is based on the inherent properties of the samples themselves (e.g., known concentrations of VWF, presence of interferents) and measurements against a reference standard (WHO IS).

    For the Reference Interval studies, "apparently healthy individuals" were used. Their healthy status would be determined by clinical criteria, likely involving medical professionals (physicians, nurses) at the collecting sites, but specific "experts" for ground truth adjudication are not mentioned as it's a physiological range determination.

    For Diagnostic Accuracy (Method Comparison), the ground truth is established by comparing the device's performance against another established and cleared VWF activity assay (predicate devices: BC von Willebrand Reagent, HemosIL VWF). The performance of these predicate devices, in turn, has been established through their own regulatory clearance processes and clinical use. No direct "expert consensus" on individual patient cases for ground truth is explicitly described for method comparison. The patient cohort included those with previously diagnosed VWD, suggesting their diagnosis was established by treating physicians using standard clinical and laboratory methods.

    For Diagnostic Concordance, the "ground truth" is defined by the "study site specific standard of care (SOC) VWF-activity assay" results in conjunction with other VWD testing panel results (VWF:Ag, FVIII activity). This is a comparison against established clinical practice rather than a 1:1 expert adjudicated ground truth for each case.

    Therefore, direct explicit statements about "number of experts" and their "qualifications" for establishing individual patient-level ground truth (as might be seen in imaging studies) are generally not applicable or explicitly provided in this type of IVD submission for quantitative assays where the truth is often defined by:

    • Reference to an international standard (WHO IS for VWF).
    • Comparison to a legally marketed predicate device.
    • Known concentrations in spiked/diluted samples.
    • Clinical status (e.g., "apparently healthy individuals").

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Given that this is an in vitro diagnostic quantitative assay for a biomarker, the "adjudication method" in the sense of multiple experts reviewing and reaching consensus on findings (like in an imaging study) is not explicitly detailed or typically performed.

    • Analytical Studies (Precision, Linearity, Interference, Detection Limits, Stability, Carryover): These studies involve objective measurements according to standardized protocols (CLSI guidelines). Data analysis is statistical, not based on expert adjudication of individual results.
    • Reference Interval Studies: These involve collecting data from a "healthy" population. The "ground truth" of their healthy status would be confirmed by clinical screening, not through formal adjudication of individual VWF results by multiple experts.
    • Diagnostic Accuracy (Method Comparison): This involves statistical comparison (e.g., Passing-Bablok regression) of the new device's results against a predicate device. Discrepancies are analyzed statistically, not through expert adjudication of individual patient results.
    • Diagnostic Concordance: This study compares the diagnostic outcome classification (e.g., "any type of VWD," "VWD excluded") from the subject device to the "site-specific standard of care" diagnostic classification. While the SOC classification would have involved clinical expertise, the comparison itself is statistical, not an adjudication process.

    Therefore, for this device, the adjudication method described is none in the traditional sense of expert consensus on individual cases. The ground truth relies on established methodologies, reference standards, and predicate device performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was done. This device is an in vitro diagnostic assay for measuring a biomarker (von Willebrand factor activity) in a blood sample. It is an automated test performed by an instrument, not a device that assists human readers (e.g., radiologists) in interpreting medical images or other complex data. Therefore, the concept of "human readers improving with AI assistance" does not apply to this product.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The INNOVANCE VWF Ac assay is an automated, standalone in-vitro diagnostic device. It provides quantitative VWF activity results on the BCS XP System. This means the analytical and diagnostic performance data presented (precision, linearity, interference, detection limits, method comparison, diagnostic concordance percent agreements) represents the algorithm/device-only performance without direct human intervention in interpreting the VWF activity value generated by the instrument for each sample. Clinicians then interpret these quantitative results in conjunction with other clinical information to make a diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" utilized varies depending on the specific study type:

    • Analytical Performance (Precision, Linearity, Interference, Detection Limits, Stability, Carryover):

      • Known concentrations: For linearity and detection limits, samples were created with known concentrations of VWF by dilution or spiking.
      • Reference materials: For traceability, the WHO 6th International Standard for Factor VIII / von Willebrand Factor was used as the ground truth.
      • Absence of interferents/analytes: For interference and limit of blank, samples were prepared to be analyte-free or with defined amounts of interferents.
      • "Healthy" population status: For reference intervals, individuals identified as "apparently healthy" through clinical screening were used.
    • Diagnostic Accuracy (Method Comparison):

      • Predicate Device Results: The results of established, legally marketed VWF activity assays (BC von Willebrand Reagent and HemosIL VWF) served as the comparator or "ground truth" for evaluating the performance of the new device.
    • Diagnostic Concordance:

      • Site-Specific Standard of Care (SOC) Diagnosis: The ground truth was based on the diagnostic classification ("any type of VWD," "VWD excluded") determined by the study site's existing clinical practice, which involved their SOC VWF-activity assay along with other elements of the initial VWD testing panel (VWF:Ag, FVIII activity). This is a form of clinical consensus based on established standard diagnostic panels.
    • Training Set Ground Truth: (See question 9 below, as this information is not explicitly provided in this document but would be distinct from the test set ground truth).

    8. The sample size for the training set:

    The document does not explicitly state the sample size for a training set. This is common for this type of IVD, where the analytical performance is established through rigorous validation studies against reference methods and spiked samples, rather than relying on machine learning models that require distinct training and test sets in the same way imaging AI does. The device's foundational principle is particle-enhanced turbidimetry, not a machine learning algorithm requiring a separate training pipeline for its core function.

    9. How the ground truth for the training set was established:

    As the document does not explicitly mention a "training set" in the context of machine learning, the concept of establishing ground truth for such a set is not applicable here. The INNOVANCE VWF Ac assay is a biochemical assay. Its development and optimization would rely on established laboratory principles, analytical validation, and calibration against reference standards, rather than a machine learning training process with a ground truth dataset. The calibration process uses "Standard Human Plasma" (SHP) which is traceable to the WHO 6th International Standard. This calibration material, and its assigned values, would be the closest analogue to "ground truth" used in setting up the device for accurate measurements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213426

    Validate with FDA (Live)

    Date Cleared
    2022-08-16

    (299 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160276, K150877

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family and ACL TOP Family 50 Series of analyzers. The product is intended to be used for the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.

    Device Description

    The thromboplastin reagent included in the ReadiPlasTin kit, after mixing with the ReadiPlasTin Diluent, is a liposomal preparation that contains recombinant human tissue factor (RTF), re-lipidated in a synthetic phospholipid blend. In the PT test, the addition of the tissue thromboplastin (ReadiPlasTin reagent) to the patient plasma in the presence of calcium ions initiates the activation of the extrinsic pathway. This results ultimately in the conversion of fibrin, with formation of a solid gel. The fibrinogen is quantitated (PT-based method) by relating the absorbance or light-scatter during clotting to a calibrator.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "HemosIL ReadiPlasTin." This document doesn't describe an AI/ML-based device, but rather an in vitro diagnostic (IVD) reagent used for Prothrombin Time (PT) and Fibrinogen determination. Therefore, many of the requested categories related to AI/ML device testing (e.g., number of experts for ground truth, adjudication method, MRMC study, training set information) are not applicable to this type of submission.

    However, I can extract the relevant information regarding acceptance criteria and performance studies for this IVD device.

    Here's a breakdown of the requested information, adapted for an IVD reagent:

    Device: HemosIL ReadiPlasTin (In vitro diagnostic thromboplastin reagent)
    Purpose: Quantitative determination of Prothrombin Time (PT) and Fibrinogen in human citrated plasma.
    Reason for Submission (K213426): Reformulation of the existing HemosIL ReadiPlasTin by adding EDTA as a stabilizer and removing unnecessary fillers.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this IVD, "acceptance criteria" are typically defined by the method validation standards (e.g., CLSI guidelines) and the equivalence to the predicate device. The performance data presented are the results of meeting these criteria.

    Performance Study TypeAcceptance Criteria (Implied/Standard)Reported Device PerformanceComments
    Precision (Repeatability & Within-Laboratory)CV% within acceptable limits for PT and Fibrinogen based on CLSI EP05-A3 guidelines and clinical utility.PT (Seconds): Repeatability CV 0.5-1.0%; Within Lab CV 0.8-1.7% (ACL TOP Family); Repeatability CV 0.7-1.1%; Within Lab CV 0.9-1.7% (ACL TOP Family 50 Series) PT (INR): Repeatability CV 0.7-1.7%; Within Lab CV 0.9-2.0% (ACL TOP Family); Repeatability CV 0.8-1.5%; Within Lab CV 1.0-2.0% (ACL TOP Family 50 Series) Fibrinogen (mg/dL): Repeatability CV 0.6-1.8%; Within Lab CV 0.8-2.0% (ACL TOP Family); Repeatability CV 0.6-2.0%; Within Lab CV 0.9-2.3% (ACL TOP Family 50 Series)All reported values fall within the expected range for good precision for coagulation assays. The text explicitly states: "The testing below and on the following pages met all acceptance criteria as follows."
    Fibrinogen LinearityResults must support the labeled linearity claim of 60 to 700 mg/dL."The results for all 3 lots on both systems met acceptance criteria, supporting the labeled fibrinogen linearity claim of 60 to 700 mg/dL."Tested across 3 lots on both instrument families per CLSI EP06, 2nd Ed.
    InterferenceNo significant interference from specified substances at given concentrations for PT and Fibrinogen measurements.No interference claimed for: - PT: UFH (1.0 IU/mL), LMWH (1.4 IU/mL), Hemoglobin (500 mg/dL), Triglycerides (1000 mg/dL), Bilirubin (Conjugated & Unconjugated) (50 mg/dL), Daptomycin (100 µg/mL) - Fibrinogen: UFH (1.5 IU/mL), LMWH (1.7 IU/mL), Hemoglobin (500 mg/dL), Triglycerides (600 mg/dL), Bilirubin (Conjugated & Unconjugated) (50 mg/dL), Daptomycin (200 µg/mL)New claims for daptomycin and conjugated bilirubin interference were added. The study used 2 clinical sample levels for both PT and Fibrinogen.
    Method ComparisonStrong correlation and agreement between the subject device and the predicate device (HemosIL RecombiPlasTin 2G). Slope should be near 1, intercept near 0, and correlation coefficient (r) close to 1.ACL TOP Family: - PT (INR): Slope 1.031 (95% CI 1.009, 1.053), Intercept -0.043 (-0.068, -0.018), r 0.997 - Fibrinogen (mg/dL): Slope 0.975 (95% CI 0.963, 0.986), Intercept 7.171 (3.842, 10.50), r 0.995 ACL TOP Family 50 Series: - PT (INR): Slope 1.021 (95% CI 0.999, 1.043), Intercept -0.034 (-0.060, -0.009), r 0.996 - Fibrinogen (mg/dL): Slope 1.015 (95% CI 1.003, 1.027), Intercept -0.811 (-4.148, 2.527), r 0.994All method comparison results demonstrated excellent correlation and agreement, supporting substantial equivalence.
    Open Vial StabilityMaintain performance for 10 days at 2-8°C in closed original vial after preparation."The results support the following labeled open vial stability claim: Once prepared for use, 10 days at 2-8℃ in closed original vial"Tested across 3 lots per CLSI EP25-A.
    On-board Instrument StabilityMaintain performance for 10 days at 15°C on the ACL TOP Family and ACL TOP Family 50 Series after preparation."The results support the following labeled on-board instrument stability claim: Once prepared for use, 10 days at 15°C on the ACL TOP Family and ACL TOP Family 50 Series"Tested across 3 lots per CLSI EP25-A.
    Real-time Shelf-life StabilityDevice maintains stated performance throughout its claimed shelf-life."The study will continue to a point past final claim." (Ongoing assessment to support shelf-life).Tested across 3 lots per CLSI EP25-A.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision:

      • Sample Size: 80 measurements per instrument per lot (n=80/instrument/lot) for PT (controls + 6 native patient samples) and Fibrinogen (controls + 6 fibrinogen sample pools at 3 levels). Tested across 3 lots and representative members of both ACL TOP Family and ACL TOP Family 50 Series.
      • Data Provenance: Not explicitly stated, but "native (unadulterated) patient samples" and "fibrinogen sample pools" suggest human biological samples. Typically, these studies are conducted in a controlled laboratory setting (Likely within the manufacturer's R&D facilities or a contract research organization). The document is submitted to the US FDA, implying relevance to the US market. The retrospective/prospective nature is generally prospective for these types of validation studies.
    • Interference:

      • Sample Size: Two clinical sample levels each for PT (normal pooled plasma and a high INR clinical sample) and Fibrinogen (normal pooled plasma and a low fibrinogen sample). Specific "n" per concentration/sample type is not given, but refers to CLSI EP07, 3rd Ed and CLSI EP37, 1st Ed which define the study design.
      • Data Provenance: Not explicitly stated, but "clinical sample levels" implies human biological samples.
    • Method Comparison:

      • Sample Size:
        • PT (INR): 160 samples (normal and abnormal) for both ACL TOP Family and ACL TOP Family 50 Series.
        • Fibrinogen (mg/dL): 135 samples for ACL TOP Family, 134 samples for ACL TOP Family 50 Series.
      • Data Provenance: Not explicitly stated, but "normal and abnormal samples" implies human biological samples. The study was "in-house."
    • Open Vial and On-board Instrument Stability:

      • Sample Size: For PT, controls and four native (unadulterated) patient samples were tested in eight replicates at each time interval. For Fibrinogen, controls and six fibrinogen sample pools at three levels were tested in eight replicates at each time interval.
      • Data Provenance: Not explicitly stated, but "native (unadulterated) patient samples" and "fibrinogen sample pools" imply human biological samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This is not applicable for this type of IVD device. The ground truth for chemical assays like PT and Fibrinogen is established through precise measurement methods and reference materials, not through expert consensus of visual or diagnostic interpretations. The "truth" is the quantitative value derived from the reference method or calibrator.


    4. Adjudication Method for the Test Set

    Not applicable, as this is an IVD reagent and not an AI/ML-based diagnostic system requiring human interpretation or adjudication.


    5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is an IVD reagent and not an AI/ML-based diagnostic system involving human interpretation.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is an IVD reagent. Its performance is inherent to the chemical reaction and the analytical instrument it is used with.


    7. The Type of Ground Truth Used

    The "ground truth" for this IVD is established by:

    • Reference Methods/Materials: For PT and Fibrinogen, this would rely on internationally recognized standards and calibrators, and the values obtained from a validated reference method (or the predicate device in method comparison studies).
    • Known Concentrations: For linearity and interference studies, samples are often spiked with known concentrations of analytes or interfering substances.
    • Validated Predicate Device: In the method comparison study, the predicate device (HemosIL RecombiPlasTin 2G) serves as the comparator for verifying the new formulation's performance.

    8. The Sample Size for the Training Set

    Not applicable. This is an IVD reagent, not an AI/ML model that requires a "training set." The development of the reagent involves chemical formulation and optimization, not data-driven machine learning.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" for an IVD reagent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193204

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-17

    (240 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K160276, K150877

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Chromogenic Factor VIII is for clinical laboratory use in the quantitative determination of factor VIII activity in 3.2% citrated human plasma. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use.

    Device Description

    CRYOcheck Chromogenic Factor VIII is used for determination of FVIII activity and contains the following four components, packaged in glass vials and provided frozen to preserve the integrity of the components:
    Reagent 1: Bovine FX and a fibrin polymerization inhibitor, with activators and stabilizers.
    Reagent 2: Human FIIa, human FIXa, calcium chloride and phospholipids.
    Reagent 3: FXa substrate containing EDTA and a thrombin inhibitor.
    Diluent Buffer: Tris buffer solution containing 1% BSA and a heparin antagonist.

    AI/ML Overview

    The CRYOcheck Chromogenic Factor VIII device is intended for the quantitative determination of Factor VIII (FVIII) activity in human plasma to identify FVIII deficiency and aid in Hemophilia A management.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful outcomes of various studies, aligning with CLSI guidelines and demonstrating performance comparable to the predicate device. Specific numerical acceptance criteria were not explicitly stated as pass/fail thresholds in the document, but the reported performance values are presented as the fulfillment of these criteria.

    Performance CharacteristicAcceptance Criteria (Implied by Study Design & Comparator Performance)Reported Device Performance (CRYOcheck Chromogenic Factor VIII)
    PrecisionCV < 10% for FVIII > 1%; SD $\leq$ 0.1% for very low FVIIIPooled precision: <10% CV for FVIII > 1%; $\leq$0.1% SD for very low FVIII plasma (Table in document)
    ReproducibilityCV < 10% for FVIII > 1%; SD $\leq$ 0.1% for very low FVIIIPooled reproducibility: <10% CV for FVIII > 1%; $\leq$0.1% SD for very low FVIII plasma (Table in document)
    Linearity RangeClinically relevant range (e.g., 0-200% FVIII activity)0 to 200% FVIII activity
    Reference IntervalEstablished from healthy population43.2-159.3% FVIII activity (95% CI)
    Shelf-Life StabilityAt least 12 months at recommended storage conditionsAt least 12 months at $\leq$-70°C (study completed up to 13 months)
    In-Use StabilityDefined operational stability durations8 hours on-board instrument; 5 days at 2-8°C; 1 month refrozen storage at $\leq$-70°C (if refrozen within 4 hours of initial thaw)
    Limit of Detection (LoD)Clinically relevant low detection limit0.5% FVIII activity
    Limit of Blank (LoB)Clinically relevant lower blank limit0.4% FVIII activity
    Limit of Quantitation (LoQ)Clinically relevant low quantitation limit0.5% FVIII activity
    InterferencesNo significant interference from common substancesNo interference from tested: Hemoglobin $\leq$ 500 mg/dL, Intralipid $\leq$ 500 mg/dL, Bilirubin (unconjugated) $\leq$ 29 mg/dL, vWF $\leq$ 20 µg/mL, Unfractionated heparin $\leq$ 2 IU/mL, Low molecular weight heparin $\leq$ 2 IU/mL, Fondaparinux $\leq$ 1.25 mg/L, Lupus Anticoagulant $\leq$ 1.8 dRVVT ratio. Interference from Rivaroxaban and Dabigatran noted.
    Method ComparisonEquivalent performance to predicate devicePassing-Bablok regression: Slope ~1, Intercept ~0, Pearson Correlation Coefficient > 0.99 (Overall: Slope 1.038, Intercept 0.473, r=0.994)
    Sample IntegrityDefined sample stability durations2 hours at room temperature; 3 months at $\leq$-70°C (including up to two freeze-thaw cycles)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Study: One normal and two abnormal reference controls, and five patient plasma samples (representing very low, mid, normal, and high FVIII activity). Each sample was measured in duplicate, twice a day for 20 days (total 80 replicates per sample per lot) with 3 lots of the device. Data provenance is internal (as it's an internal study).
    • Reproducibility Study: One normal and two abnormal reference controls, and three patient plasma samples (representing very low, normal, and high FVIII activity). Each sample was measured in triplicate, twice a day for 5 days at each of 3 sites with 3 lots of the device. Data provenance includes one internal and two external sites, but specific countries are not mentioned beyond "Canada" for the submitter. This appears to be prospective data collection for the study.
    • Linearity/Assay Reportable Range Study: Fifteen sample dilutions created by combining high FVIII plasma (260%) with congenital FVIII deficient plasma (0%). Each level was tested in quadruplicate. Data provenance is internal (as it implies an internal study).
    • Reference Interval Study: One hundred and twenty ostensibly healthy individuals $\geq$ 18 years. Data provenance is not specified beyond "citrated plasma samples collected from". This appears to be prospective data collection.
    • Shelf-Life Stability Study: Six plasma samples representing low to normal FVIII activity levels. Three lots of the device were tested at various time points up to 37 months (13 months completed). Data provenance is internal.
    • In-Use Stability Study: Six plasma samples representing low to normal FVIII activity levels. Three lots of the device were tested at various time points. Data provenance is internal.
    • Detection Limit (LoB/LoD/LoQ) Studies:
      • LoB: Four blank plasma samples from individuals with severe congenital hemophilia A.
      • LoD: Four plasma samples with low FVIII activity from congenital hemophilia A donors.
      • LoQ: Aliquots of four plasma samples with low FVIII activity from congenital hemophilia A donors.
        All samples for LoB/LoD were measured in triplicate over five days using three lots of the device. For LoQ, samples were tested in triplicate on five different days at an external laboratory. Data provenance for LoB/LoD/LoQ samples is from individuals with hemophilia A.
    • Interference Studies: Plasma samples spiked with possible interferents. Ten replicates of each spiked sample and 10 replicates of corresponding blank matrix control were tested. Data provenance is internal.
    • Method Comparison Studies: Three hundred and eighteen human plasma samples from normal individuals, patients with congenital or acquired hemophilia A, and various types of von Willebrand disease. Samples were distributed across three sites. Data provenance spans "normal ostensibly healthy individuals and from patients". This appears to be prospective data collection.
    • Sample Integrity Study: Forty-six plasma samples. Data provenance is not further specified. This appears to be prospective data collection at two external sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For this in vitro diagnostic device, "ground truth" is typically established by reference methods or accepted clinical classifications rather than expert consensus on individual images or cases.

    • For studies like Precision, Reproducibility, Linearity, Stability, and Detection Limits, the "ground truth" for the FVIII activity levels in controls and calibrators would be established by the manufacturer's internal characterization methods, traceable to international standards if applicable (though not explicitly stated as such for FVIII activity). The expertise would lie in clinical chemistry, hematology, and laboratory medicine for the development and validation of these reference materials and methods. No specific number or qualification of experts is mentioned for these.
    • For the Reference Interval Study, the ground truth is derived from the statistical distribution of FVIII activity in a population of ostensibly healthy individuals. This doesn't involve individual expert ground-truthing but rather statistical analysis.
    • For the Method Comparison Study, the ground truth for FVIII activity was established by a comparator device, Coatest SP FVIII (K042576), used at a central reference laboratory. The "experts" in this context are the trained laboratory personnel performing the reference method.
    • For sample selection in studies involving "congenital hemophilia A donors" or "patients with congenital or acquired hemophilia A and various types of von Willebrand disease," the diagnosis (which forms part of the "ground truth" for classifying these samples) would have been made by medical professionals (e.g., hematologists) based on standard clinical and laboratory diagnostic criteria. No specific number or qualifications are given.

    4. Adjudication Method for the Test Set

    Adjudication methods like "2+1" or "3+1" are typically used for subjective diagnostic tasks, often involving image interpretation or clinical reviews. Since the CRYOcheck Chromogenic Factor VIII is a quantitative assay for FVIII activity, the "adjudication method" is the quantitative measurement itself, often with replicates and statistical analysis, as detailed in the study designs (e.g., "in duplicate," "in triplicate"). No expert adjudication in the traditional sense is described or expected for this type of device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. MRMC studies are primarily for evaluating the impact of a new diagnostic tool on the diagnostic performance of human readers, typically with subjective assessments. This device is a quantitative in vitro diagnostic, not a tool for human reading in the conventional sense.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the device operates as a standalone algorithm (assay) measuring FVIII activity. The performance studies described (Precision, Reproducibility, Linearity, Reference Interval, Stability, Detection Limits, Interferences, Method Comparison, Sample Integrity) all represent the performance of the device itself, without human-in-the-loop diagnostic interpretation as part of its core function, other than trained laboratory personnel operating the instrument and interpreting numerical output.

    7. The Type of Ground Truth Used

    The ground truth used for these studies varies by the specific test:

    • Reference materials and controls: For precision, linearity, and detection limit studies, the ground truth for FVIII activity levels is based on established values of these materials, likely traceable to international standards or well-characterized internal standards.
    • Healthy population data: For the reference interval, the ground truth for "normal" FVIII activity is derived from a statistically representative healthy population.
    • Comparator device: For the method comparison study, the ground truth was established by the predicate device, Coatest SP FVIII (K042576).
    • Clinical diagnosis: For studies involving patient samples (e.g., hemophilia A, von Willebrand disease), the ground truth for their disease status would be based on clinical diagnosis by medical professionals.

    8. The Sample Size for the Training Set

    The document describes performance studies (validation tests) for the device. It does not provide information on a "training set" in the context of machine learning. This is because the device is a chemical assay, not an AI/ML-based diagnostic system that would require a training set to develop its model.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or applicable for this type of chemical assay device, this question is not relevant based on the provided information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183390

    Validate with FDA (Live)

    Device Name
    QNext and DG-PT
    Date Cleared
    2019-08-21

    (258 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    ACL Top 700 (K160276), HemosIL PT Fibrinogen HS Plus (K060931)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QNext is a fully-automated, random-access instrument, intended for in vitro diagnostic use in clinical laboratories to perform hemostasis testing by detecting the changes in optical density.

    DG-PT is a thromboplastin reagent for the quantitative determination of Prothrombin Time on human plasma samples collected in 3.2% sodium citrate.

    The product is used for the evaluation of the extrinsic and common coagulation pathways in seconds and for the monitoring Oral Anticoagulant Therapy with warfarin in International Normalized Ratio (INR).

    For use with ONext.

    For clinical professional laboratory and prescription use only.

    For in vitro diagnostic use.

    The performance of this device has not been established in neonate and pediatric patient populations.

    Device Description

    QNext is designed to automatically perform all stages of the procedures associated to hemostasis tests allowing the operator to:

    • Absorb the workload involved in running hemostasis laboratory tests profiles, optimizing the execution of these profiles in the shortest time possible, and ensuring the maximum possible precision and accuracy in the results.
    • Increase the reliability of the analytical process, eliminating any possible errors in the identification ● and treatment of samples and products and in the revision and transcription of the results.
    • Reduce the risk of Operator contamination by minimizing interaction between the Operator and ● samples and products during the analytical process.

    To perform the operations for which it has been designed, QNext automatically follows the steps listed below:

    • Sample management: loading, positive identification, dilution (if required) and dispensation into cuvettes.
    • Reagent management: loading, positive identification, cooling, stirring, aspiration and dispensation into cuvettes.
    • Cuvette management: loading, transport, incubation during the reactions and management of used cuvettes.
    • Management of test requests. ●
    • Execution of test procedures.
    • Result management: optical measurement of the reactions, algorithm calculation of analytical parameters from reaction curves, validation of results, traceability, bi-directional transmission of requests and results to the LIS.
    • Management of disposable components. ●

    The data analyzed can be stored, displayed and printed. Additionally, the analyzer allows conducting integrated functions, such as the analysis of urgent samples or the Quality Control module.

    DG-PT consists of a glass vial containing lyophilized thromboplastin (tissue factor and phospholipids) from rabbit brain tissue, buffer, calcium ions and preservative. The closure system includes a stopper and a screw сар.

    DG-PT reagent is used to perform PT tests for:

    • the evaluation of the extrinsic and common coagulation pathways.
    • The monitoring Oral Anticoagulant Therapy with warfarin.

    The assay is based on the activation of the extrinsic coagulation pathway by the addition of the reagent to the plasma sample. The thromboplastin interacts with FVII and calcium ions activating a series of specific enzymes that comprise the extrinsic and common pathways of the coagulation cascade ultimately leading to the formation of a fibrin clot. The QNext reader measures the light change produced during the reaction.

    AI/ML Overview

    The provided text describes the performance testing of the ONext instrument and DG-PT reagent for Prothrombin Time (PT) determination, but it does not contain information related to an AI/ML powered device. Therefore, many of the requested criteria regarding AI/ML device performance (e.g., multi-reader multi-case studies, human-in-the-loop performance, training set details, expert ground truth adjudication) are not applicable to this document.

    However, I can extract and present the acceptance criteria and performance data for the described in-vitro diagnostic device:

    Device: ONext (fully-automated hemostasis instrument) and DG-PT (thromboplastin reagent for PT determination).

    Study Proving Device Meets Acceptance Criteria:
    The submission details several performance studies conducted according to CLSI guidelines to demonstrate the substantial equivalence of the ONext and DG-PT to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined quantitative acceptance criteria thresholds for each study (e.g., "Precision CV must be less than X%"). Instead, it presents the results of various performance studies (Precision, Reference Interval, Sensitivity, Specificity, Method Comparison) which are implicitly understood to meet the necessary performance characteristics for substantial equivalence. The "performance" column therefore reflects the observed results from these studies.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (ONext and DG-PT)
    PrecisionRepeatability and Within-laboratory precision should demonstrate acceptable variability.Repeatability (PT seconds / INR): Varies by sample level. For PT (seconds), 1.4-2.1% CV; for INR, 1.6-2.4% CV. Upper One-sided 95% Limits slightly higher. Within-laboratory Precision (PT seconds / INR): Varies by sample level. For PT (seconds), 2.4-5.7% CV; for INR, 2.2-8.2% CV. Upper One-sided 95% Limits higher. Reproducibility (PT seconds / INR): Varies by sample level. For PT (seconds), Repeatability 1.6-3.1% CV, Reproducibility 3.1-6.0% CV; for INR, Repeatability 1.9-3.7% CV, Reproducibility 5.5-9.2% CV. Upper One-sided 95% Limits higher. (See Tables 1 & 2 in source for full details)
    Reference IntervalEstablish a clear reference interval for healthy individuals.PT (seconds): 12.1 to 16.1 seconds PT (INR): 0.83 to 1.17 INR
    Sensitivity (Factor)Demonstrate ability to detect deficiencies in key coagulation factors.FII: 29 IU/dL FV: 45 IU/dL FVII: 44 IU/dL FX: 43 IU/dL
    Specificity (Interference)Demonstrate resistance to interference from common endogenous and exogenous substances.Maximum Concentration without Significant Interference (Normal/Abnormal samples): Hemoglobin: 10.0 g/L Conjugated bilirubin: 43.3 mg/dL Triglycerides: 3000 mg/dL Citrate: 0.6% UFH: 1.1 IU/mL (Normal), 0.7 IU/mL (Abnormal) LMWH: 2.3 IU/mL (Normal), 1.8 IU/mL (Abnormal) Observed Interference at All Concentrations Tested for: Rivaroxaban (e.g., 313, 625, 938, 1250 ng/mL for normal sample) Argatroban (e.g., 375, 750, 1125, 1500 ng/mL for normal sample) Dabigatran (e.g., 150, 300, 450, 600 ng/mL for normal sample)
    Method ComparisonResults should correlate well with a legally marketed predicate device.Results from two US sites (n=360): PT (s): Slope 0.913 (95% CI: 0.893-0.938), Intercept 1.118 (95% CI: 0.706-1.477), r=0.983 PT (INR): Slope 1.034 (95% CI: 1.010-1.060), Intercept -0.024 (95% CI: -0.054-0.009), r=0.980 Results from ex-US site (n=271): PT (s): Slope 0.985 (95% CI: 0.962-1.003), Intercept 0.272 (95% CI: -0.078-0.716), r=0.988 PT (INR): Slope 1.000 (95% CI: 0.978-1.022), Intercept -0.070 (95% CI: -0.099--0.042), r=0.990

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Studies:
      • Repeatability and Within-laboratory: 7 patient-derived samples. Data provenance not explicitly stated (e.g., country of origin), assumed to be clinical samples. Retrospective/prospective not specified, but typically for precision studies, these would be collected prospectively or obtained from a biobank.
      • Reproducibility: 7 patient-derived samples. Data provenance not explicitly stated (e.g., country of origin), assumed to be clinical samples. Retrospective/prospective not specified.
    • Reference Interval Study: 243 reference samples. Data provenance not explicitly stated.
    • Sensitivity Study: Not specified, but involved a set of samples spanning from 100% to <1% of specific coagulation factors.
    • Specificity Study: Two sample levels (pooled normal and pooled abnormal samples) with various interfering substances added. Specific total sample count not given.
    • Method Comparison Study:
      • US Sites: 360 patient-derived samples covering the claimed analytical measurement range (AMR). Data provenance: from two US clinical sites.
      • Ex-US Site: 271 patient-derived samples covering the claimed AMR. Data provenance: from one ex-US clinical site.
      • Retrospective/Prospective: Not explicitly stated, but clinical validation studies often use prospectively collected samples or samples from a well-characterized biobank.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Applicable. This is an in-vitro diagnostic device for measuring Prothrombin Time. The "ground truth" is established by direct measurement of the blood coagulation time or the INR, not by expert interpretation of images or clinical data. Therefore, expert consensus for ground truth is not relevant to this type of device.

    4. Adjudication Method for the Test Set

    Not Applicable. As per point 3, the ground truth for this device is a direct quantitative measurement, not a subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not Applicable. This document describes a medical device for in-vitro diagnostics (measuring blood coagulation), not an AI/ML powered device for image interpretation or diagnosis. There are no "human readers" or "AI assistance" components mentioned in the context of this device's intended use or validation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not Applicable. This is not an AI/ML algorithm. The device, ONext, is a fully-automated instrument, meaning its performance is inherently "standalone" in terms of its operation and measurement of PT/INR. The "human-in-the-loop" would be the laboratory professional using the instrument, but their role is in operating the device and interpreting its quantitative output, not in assisting or being assisted by an algorithm in a diagnostic task.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance studies of the ONext and DG-PT device is:

    • Direct Measurement: The actual PT/INR values obtained from patient samples using the predicate device or a reference method.
    • Clinical Samples with Known Characteristics: For sensitivity studies, samples with known factor deficiencies were used. For specificity studies, samples with known concentrations of interfering substances were used.
    • Reference Intervals: Established from healthy donor samples.

    8. The Sample Size for the Training Set

    Not Applicable. As this is not an AI/ML powered device, there is no "training set" in the context of machine learning model development. The device's operational parameters and algorithms are determined through traditional engineering and chemistry principles, not data-driven machine learning training.

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. See point 8.