Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K252582

    Validate with FDA (Live)

    Date Cleared
    2025-11-18

    (95 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K083192, K160979, K193401, K113779, K153268, K152509

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inclusive® Titanium Abutments are premanufactured prosthetic components connected to endosseous dental implants in edentulous or partially edentulous maxilla or mandible to provide support for cement-retained or screw-retained prosthetic restorations. All digitally designed abutments for use with Inclusive® Titanium Abutments for CAD/CAM are intended to be sent to a Prismatik Dentalcraft validated milling center for manufacture.

    Compatible Implant Systems: Camlog® CONELOG®, Dentium® NR Line

    Compatible Implant Fixtures:
    Camlog® CONELOG®: 3.3 mm/3.3 mm, 3.8 mm/3.8 mm, 4.3 mm/4.3 mm, 5.0 mm/5.0 mm
    Dentium® NR Line: 3.1 mm/3.2 mm, 3.6 mm

    Device Description

    Inclusive® Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. Inclusive® Titanium Abutments are designed and fabricated to be compatible with Camlog® CONELOG® and Dentium® NR Line Implant Systems. The products are made from titanium alloy Ti-6Al-4V ELI, which meets ASTM F136-13(2021), Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). They include Inclusive® Titanium Abutment Blanks intended to be used to fabricate one-piece, all-titanium, patient-specific abutments using CAD/CAM technology and Inclusive® Titanium Abutments 4.5mmH and 6mmH intended to be used for support of fabricated crowns/bridges or a zirconia coping to complete the two-piece abutment. The zirconium coping intended for use with the Inclusive® Titanium Abutments 4.5mmH and 6mmH devices are made from zirconium oxide ceramic, conforming to ISO 13356 Implants for surgery – Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP). Inclusive® Titanium Abutments 4.5mmH and 6mmH devices are a two-piece abutment with a titanium base and a ceramic top-half which when cemented together (Shofu MonoCem K020481) constitutes the final finished abutment. Each patient-specific abutment is prescribed by a clinician and manufactured by an authorized milling center validated by Prismatik Dentalcraft Inc. Inclusive® Titanium Abutments are provided non-sterile and intended for single use and prescription use.

    Inclusive® Multi-Unit Coping is manufactured from titanium alloy, Ti-6Al-4V ELI conforming to ASTM F136-13(2021) and used in conjunction with the Camlog® CONELOG® and Dentium® NR Line multi-unit abutments. Inclusive® Multi-Unit Coping is bonded with the dental restoration prior to being seated on the multi-unit abutment via a multi-unit prosthetic screw. The non-engaging configuration of the multi-unit coping does not have an internal connection feature and seats onto the flat mating surface of the multi-unit abutment. The multi-unit coping is used in combination with screw-retained multi-unit dental prosthetics, e.g. bridges and bars, which are used to reconstruct the function and aesthetics of lost teeth. The multi-unit coping is straight with no angle correction and provided non-sterile. The device is intended for singe use and prescription use.

    The design parameters for the Titanium Abutment Blanks are:

    • the minimum wall thickness is no less than 0.5 mm.
    • the gingival margin diameter is no less than 0.5 mm wider than implant.
    • maximum angle correction is no greater than 30°. *
    • minimum margin height is no less than 0.5 mm.
    • maximum gingival height is no greater than 6 mm, and
    • minimum post height above the gingival collar is no less than 4.0 mm.

    *Angulation up to 30° for Titanium Abutment Blanks only allowed for compatible implant bodies of Camlog® CONELOG® Ø3.3mm Implants and Dentium® NR Line Ø3.1mm Implants for anterior use in the mouth under single-unit loading.

    Inclusive® Titanium Abutments Blanks compatible with Camlog® CONELOG® Ø3.8mm, Ø4.3mm, and Ø5.0mm Implants are not to be used as angled abutments in the posterior region.

    The design parameters for the zirconia coping for the Titanium Abutment 4.5mmH and 6mmH are:

    • the minimum wall thickness is no less than 0.5mm.
    • the gingival margin diameter is no less than 0.5 mm wider than implant.
    • no angulation correction, straight (0°) only.
    • minimum margin height is no less than 0.5 mm.
    • maximum gingival height is no greater than 6 mm, and
    • minimum post height above the gingival collar is no less than 4.0 mm.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242753

    Validate with FDA (Live)

    Date Cleared
    2025-01-06

    (116 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K153268, K222792, K223634

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SD TL Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. Abutments are indicated for screw-retained single restorations or cement-retained single of multi-unit restorations. SD TL Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. All digitally-designed Pre-Milled Abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.

    Device Description

    An endosseous dental implant is a device made of a material such as Pure Titanium (Conforming to ASTM F67) which will be placed in the alveolar bone to replace the function of the missing tooth. The SD TL Implant System consists of dental implants, abutments for use in one or two-stage dental implant placement and restorations.

    The implant-abutment connection is tight and precise fitting with non-submerged external connection. The surface of the fixture is treated with SLA (Sandblasted with Large-grit and Acidetching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone). The SD TL Implants are available with a gingival height of 1.5 mm.

    The abutment is made of Ti-6Al-4V Eli (Conforming to ASTM F136) to be used in fabricating patient-specific abutments. The subject devices are indicated for cemented or screw-and cement retained prosthesis (SCRP) restorations. Each patient-specific abutment is individually prescribed by the clinician. The Multi Pre-Milled cylinders come in engaging and non-engaging types. All designed abutments are sent to an ARUM DENTISTRY validated milling center.

    The diameters of Multi Pre-Milled Cylinder are 4.8 mm Square.

    AI/ML Overview

    There were no acceptance criteria or studies on device performance to meet stated acceptance criteria discussed in the provided text. The text describes a 510(k) premarket notification for the "SD TL Implant System," which is a dental implant system. The document focuses on demonstrating substantial equivalence to predicate devices based on material, design, intended use, and non-clinical performance testing (biocompatibility, sterilization, shelf-life, and mechanical testing).

    The document states that biocompatibility, sterilization, shelf-life, and mechanical performance testing (according to ISO 14801 for static compression and compression fatigue) were conducted. It concludes that "The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate." However, specific acceptance criteria values for these tests and the detailed results of how the device met them are not provided in the document.

    Therefore, for the requested information:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document generally states that tests met "the criteria of the standards" but does not enumerate these criteria or specific performance values.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not provided. The document mentions mechanical testing according to ISO 14801 but does not specify sample sizes or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as this device certification is for a dental implant system based on non-clinical performance, not an AI/diagnostic device requiring expert ground truth for a test set.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of device and testing.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an AI/diagnostic device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of an AI/diagnostic device. For this dental implant, "ground truth" would be the established performance standards (e.g., ISO manufacturing standards, material properties, mechanical strength benchmarks) against which the device's physical properties and performance are measured.
    8. The sample size for the training set: Not applicable, as this is not an AI/machine learning device requiring a training set.
    9. How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192436

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-23

    (139 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K041368, K112045, K141457, K153268, K172640

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dentium Prosthetics are intended for use as an aid in prosthetic rehabilitation.

    Device Description

    The purpose of this submission is to change the sterilization method of Healing Abutments and Cover Screws. These devices which have been provided non-sterilized by gamma radiation.

    Healing Abutments are used provisionally as an accessory to endosseous dental implant during healing period to prepare gingival tissue for acceptance of a final abutment. Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final abutment. Cover Screws are used provisionally as an accessory to protect the inner features of the implant.

    The Healing Abutments and Cover Screws are prefabricated and made of Ti-6Al-4V ELI (ASTM F136). These devices are sterilized using gamma radiation method and intended for single use only.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for "Healing Abutments and Cover Screws." This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific performance acceptance criteria through the kind of rigorous, large-scale studies typically associated with AI/ML-based medical devices or novel technologies.

    Therefore, many of the requested details, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable to this specific submission. This document describes a change in the sterilization method for existing devices, not a new device with performance metrics that require extensive clinical validation studies beyond sterilization and biocompatibility.

    Here's an analysis based on the information available in the document:


    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission concerning a change in sterilization method for mechanical components like healing abutments and cover screws, the "acceptance criteria" revolve around demonstrating that the new sterilization method achieves the required sterility assurance level and does not adversely affect the device's material properties or function, maintaining its substantial equivalence to the predicate.

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance/Evidence
    Sterilization EfficacyAchieve a Sterility Assurance Level (SAL) of 10^-6, in accordance with recognized standards."Gamma radiation sterilization validation according to ISO 111137-1 and ISO 11137-2, demonstrating a sterility assurance level (SAL) of 10^-6." (Page 4)
    BiocompatibilityThe device material (Ti-6Al-4V ELI (ASTM F136)) must be biocompatible and the sterilization process must not negatively impact biocompatibility."Biocompatibility of Ti-6A1-4V ELI (ASTM F136) demonstrated by the referenced Dentium submission, K041368, using the identical materials and manufacturing processes including sterilization as the subject device." (Page 4)
    Material Integrity/Shelf-lifeThe sterilization process should not degrade the material or product function over its intended shelf-life."Accelerated and real time aging studies according to ASTM F1980 demonstrating a shelf life." (Page 4)
    Functional EquivalenceThe design, materials, and intended use of the device must remain substantially equivalent to previously cleared versions (predicates), despite the change in sterilization. This is demonstrated by comparing technological characteristics like material, form, connection type, dimensions, and indication for use. The only noted difference between the subject device and the primary predicate device for both healing abutments and cover screws is the sterilization method (Pages 7, 9).The document states that the subject devices (Healing Abutments and Cover Screws) have the "same characteristics for the followings compared to the primary predicate device: Indication for use, Material, Connection type, Dimension" (for Healing Abutments, Page 7) and "Indication for use, Material, Dimension" (for Cover Screws, Page 9). Dimensions are extensively compared and shown to match various predicates (Pages 6-7, 9).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated as a separate "test set" in the context of clinical performance data typical for AI/ML devices. The "tests" here are primarily engineering and laboratory validations:
      • Sterilization Validation: ISO 11137 standards define the number of units and replicates required for sterility testing (e.g., bioburden determination, dose mapping, sterility testing), which are often small, specific lots. This is standard laboratory validation, not a patient-based test set.
      • Aging Studies: ASTM F1980 dictates sample sizes for accelerated aging, again, laboratory testing of physical samples.
      • Biocompatibility: Relies on a prior submission (K041368) which presumably included its own biocompatibility testing. No new "sample size" for a clinical test set is mentioned for biocompatibility here.
    • Data Provenance: The studies are laboratory-based ("Non-clinical testing data"). The manufacturer is located in the Republic of Korea (Dentium Co., Ltd.). The data would be derived from their internal validation processes. It is neither retrospective nor prospective in the sense of patient studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This submission does not involve image interpretation or clinical diagnosis where expert ground truth would be established. The "ground truth" for sterilization is a defined SAL, verified by microbiological and physical testing per ISO standards. The "ground truth" for material properties is based on ASTM standards and prior biocompatibility data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As there is no clinical "test set" based on human interpretation or an AI algorithm's output requiring adjudication, this concept does not apply.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a traditional medical device (dental implants and accessories), not an AI-assisted diagnostic or therapeutic device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this submission is based on:
      • Standardized Sterility Assurance: Achieving a defined SAL of 10^-6, verified through validated sterilization processes and testing compliant with ISO 11137-1 and ISO 11137-2.
      • Material Standards: Biocompatibility established against Ti-6Al-4V ELI (ASTM F136) as per regulations and previous submissions.
      • Engineering Performance: Shelf-life demonstrated through accelerated and real-time aging studies per ASTM F1980.
      • Substantial Equivalence: Demonstrated by comparing engineering specifications and intended use against legally marketed predicate devices.

    8. The sample size for the training set

    • Not Applicable. This involved physical device manufacturing and validation, not machine learning model training.

    9. How the ground truth for the training set was established

    • Not Applicable. (See point 8)
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1