K Number
K242753

Validate with FDA (Live)

Date Cleared
2025-01-06

(116 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SD TL Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. Abutments are indicated for screw-retained single restorations or cement-retained single of multi-unit restorations. SD TL Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. All digitally-designed Pre-Milled Abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.

Device Description

An endosseous dental implant is a device made of a material such as Pure Titanium (Conforming to ASTM F67) which will be placed in the alveolar bone to replace the function of the missing tooth. The SD TL Implant System consists of dental implants, abutments for use in one or two-stage dental implant placement and restorations.

The implant-abutment connection is tight and precise fitting with non-submerged external connection. The surface of the fixture is treated with SLA (Sandblasted with Large-grit and Acidetching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone). The SD TL Implants are available with a gingival height of 1.5 mm.

The abutment is made of Ti-6Al-4V Eli (Conforming to ASTM F136) to be used in fabricating patient-specific abutments. The subject devices are indicated for cemented or screw-and cement retained prosthesis (SCRP) restorations. Each patient-specific abutment is individually prescribed by the clinician. The Multi Pre-Milled cylinders come in engaging and non-engaging types. All designed abutments are sent to an ARUM DENTISTRY validated milling center.

The diameters of Multi Pre-Milled Cylinder are 4.8 mm Square.

AI/ML Overview

There were no acceptance criteria or studies on device performance to meet stated acceptance criteria discussed in the provided text. The text describes a 510(k) premarket notification for the "SD TL Implant System," which is a dental implant system. The document focuses on demonstrating substantial equivalence to predicate devices based on material, design, intended use, and non-clinical performance testing (biocompatibility, sterilization, shelf-life, and mechanical testing).

The document states that biocompatibility, sterilization, shelf-life, and mechanical performance testing (according to ISO 14801 for static compression and compression fatigue) were conducted. It concludes that "The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate." However, specific acceptance criteria values for these tests and the detailed results of how the device met them are not provided in the document.

Therefore, for the requested information:

  1. A table of acceptance criteria and the reported device performance: Not provided. The document generally states that tests met "the criteria of the standards" but does not enumerate these criteria or specific performance values.
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not provided. The document mentions mechanical testing according to ISO 14801 but does not specify sample sizes or data provenance.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as this device certification is for a dental implant system based on non-clinical performance, not an AI/diagnostic device requiring expert ground truth for a test set.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of device and testing.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an AI/diagnostic device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of an AI/diagnostic device. For this dental implant, "ground truth" would be the established performance standards (e.g., ISO manufacturing standards, material properties, mechanical strength benchmarks) against which the device's physical properties and performance are measured.
  8. The sample size for the training set: Not applicable, as this is not an AI/machine learning device requiring a training set.
  9. How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 6, 2024

ARUM DENTISTRY Co., Ltd. Choi Won-Yi Official Correspondent 23, Gukjegwahak 11-ro, Yuseong-gu Daejeon, 34002 REPUBLIC OF KOREA

Re: K242753

Trade/Device Name: SD TL Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 12, 2024 Received: December 6, 2024

Dear Choi Won-Yi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242753

Device Name

SD TL Implant System

Indications for Use (Describe)

The SD TL Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. Abutments are indicated for screw-retained single restorations or cement-retained single of multi-unit restorations. SD TL Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. All digitally-designed Pre-Milled Abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Arum Dentistry. The logo consists of a red triangle on top of a smaller, light orange triangle on the left side of the image. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in a bold, dark brown font, with "DENTISTRY" appearing below "ARUM".

7. 510(K) Summary

Submitter

ARUM DENTISTRY Co., Ltd. Won-Yi Choi 23, Gukjeqwahak 11-ro, Yuseong-qu Daejeon, 34002 Republic of Korea Email: arum ra@arumdentistry.com Tel. +82-42-935-3644 Fax. +82-42-935-3633

Device Information

  • Trade Name: SD TL Implant System .
  • . Common Name: Implant, Endosseous, Root-Form
  • . Classification Name: Endosseous Dental Implant
  • Primary Product Code: DZE
  • Secondary Product Code: NHA
  • Panel: Dental
  • . Regulation Number: 21 CFR 872.3640
  • . Device Class: Class II
  • . Date Prepared: 01/06/2025

Predicate Devices

The subject device is substantially equivalent to the following predicate devices:

Primary Predicate

  • K222778, Osstem Implant System by Osstem Implant CO., LTD. .

Reference Device

  • K153268, NR Line Implant System by Dentium Co., Ltd. .
  • K222792, SNUCONE Tissue Level Implant System by SNUCONE Co., Ltd. .
  • K223634, Customized Abutment by ARUM DENTISTRY Co., Ltd.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one beige. The words "ARUM DENTISTRY" are in a bold, brown font, with "ARUM" on top and "DENTISTRY" on the bottom.

General Description

SD TL Implant System consist of below:

Fixture

  • . SD Tissue Level Fixture

Abutment

  • . Multi Pre-Milled Cylinder

Device Description

An endosseous dental implant is a device made of a material such as Pure Titanium (Conforming to ASTM F67) which will be placed in the alveolar bone to replace the function of the missing tooth. The SD TL Implant System consists of dental implants, abutments for use in one or two-stage dental implant placement and restorations.

The implant-abutment connection is tight and precise fitting with non-submerged external connection. The surface of the fixture is treated with SLA (Sandblasted with Large-grit and Acidetching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone). The SD TL Implants are available with a gingival height of 1.5 mm.

The abutment is made of Ti-6Al-4V Eli (Conforming to ASTM F136) to be used in fabricating patient-specific abutments. The subject devices are indicated for cemented or screw-and cement retained prosthesis (SCRP) restorations. Each patient-specific abutment is individually prescribed by the clinician. The Multi Pre-Milled cylinders come in engaging and non-engaging types. All designed abutments are sent to an ARUM DENTISTRY validated milling center.

The diameters of Multi Pre-Milled Cylinder are 4.8 mm Square.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM" and "DENTISTRY" on the right. The geometric shape is made up of a red triangle on top of a smaller, light orange triangle. The words "ARUM" and "DENTISTRY" are written in a bold, dark brown font.

K242753

Patient-specific abutment design parameters:

Image /page/6/Figure/3 description: The image shows a diagram of a dental implant with labels for gingival height, thickness, post height, angle, and diameter. A table below the diagram lists the minimum and maximum values for each parameter. The post height for a single-unit restoration ranges from 4.0 mm to 13.0 mm, the angle ranges from 0° to 30°, the wall thickness ranges from 0.5 mm to 6.0 mm, the gingival height ranges from 0.5 mm to 4.0 mm, and the diameter ranges from a minimum wall thickness to 13.9 mm.

  • Post height for single-unit restoration is defined as the cementable post measured above the gingival collar of the final abutment design

※ When machined at an angle of above 0°

  • Post Height for Single-Unit Restoration: 4.25mm -
  • Wall Thickness:0.5mm -
  • Diameter: 6.25 -
  • Gingival Height: 4.0mm -

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Arum Dentistry. The logo consists of two triangles on the left side, one red and one beige. To the right of the triangles is the text "ARUM" in bold, followed by "DENTISTRY" in bold on the line below.

K242753

Indication for Use

The SD TL Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. Abutments are indicated for screw-retained single restorations or cement-retained single of multi-unit restorations. SD TL Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. All digitally-designed Abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.

Materials

The Fixtures are fabricated from Pure Titanium (Conforming to ASTM F67) All Cylinder and Abutment Screws are fabricated from Ti-6AI-4V Eli (Conforming to ASTM F136).

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Arum Dentistry. The logo consists of a red triangle and a smaller tan triangle on the left side. To the right of the triangles is the text "ARUM" in a bold, dark brown font, with the word "DENTISTRY" underneath it in the same font and color.

Summaries of Technology Characteristics

1) SD Tissue Level Fixture

Subject DevicePrimary PredicateReference Device
ManufacturerARUM DENTISTRY Co., Ltd.Osstem Implant CO., LTD.SNUCONE Co., LTD.
Device NameSD Tissue Level FixtureSSIII SA ImplantEF Fixture for SNUCONE TissueLevel Implant System
510(k) NumberN/AK222778K222792
Intended Use/Indications foruseThe SD TL Implant System isindicated for use in partially or fullyedentulous mandibles andmaxillae, in support of single ormultiple-unit restorations including;cemented retained, screw retained,or overdenture restorations, andterminal or intermediate Abutmentsupport for fixed bridgework.Abutments are indicated for screw-retained single restorations orcement-retained single of multi-unitrestorations. SD TL ImplantSystem is dedicated for two stagesurgical procedures and forimmediate loading when there isgood primary stability and anappropriate occlusal load. Alldigitally-designed Pre-MilledAbutments are intended to be sentto an ARUM DENTISTRY validatedmilling center for manufacture.The Osstem Implant System isindicated for use in partially or fullyedentulous mandibles andmaxillae, in support of single ormultiple-units restorationsincluding; cemented retained,screw retained, or overdenturerestorations, and final ortemporary abutment support forfixed bridgework. It is intended fordelayed loading.Ultra wide Implant System isintended to be used in the molarregion.Products with diameter of less than3.25mm should be usedexclusively for the lateral incisor inthe maxilla and a centralor lateral incisor in the mandible.SNUCONE Tissue Level ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unitrestorations including; cementedretained, or overdenturerestorations and terminal orintermediate Abutment support forfixed bridge work. Snucone implantsystem is dedicated for two stagesurgical procedures and forimmediate loading when there isgood primary stability and anappropriate occlusal load. Also,implants with diameters larger than5mm are indicated for molarregions.
MaterialTitanium Grade 4 (ASTM F67)Titanium Grade 4 (ASTM F67)Titanium Grade 4 (ASTM F67)
Range ofDiameters (Ø)
3.7, 4.2, 4.6, 5.04.8, 6.03.7, 3.8, 4.1, 4.3, 4.8, 5.3, 5.5, 5.8
Range ofLengths (mm)7.0, 8.5, 10.0, 11.5, 13.07.0, 8.5, 10.0, 11.5, 13.0, 15.07, 8, 9, 10, 11, 12, 13, 14
SurfacetreatmentSLASLAAcid etching
SterilizationGamma SterilizationGamma SterilizationGamma Sterilization
Principle ofOperationThis product is a root-type fixturewhich is inserted in the alveolarbone. It replaces the functions ofthe missing teeth as a dentalimplant fixture.This product is a root-type fixturewhich is inserted in the alveolarbone. It replaces the functions ofthe missing teeth as a dentalimplant fixture.This product is a root-type fixturewhich is inserted in the alveolarbone. It replaces the functions ofthe missing teeth as a dentalimplant fixture.
SubstantialEquivalentDiscussionSimilaritiesThe SD Tissue Level Fixture and SSIII SA Implant share similar device characteristics with the Reference Device,including indication for use, length, material, functions, structure, and production method. Both devices areintended for use in partially or fully edentulous mandibles and maxillae. While the Primary Predicate device hasdiameters of 4.8 and 6.0, the subject device offers diameters of 3.7, 4.2, 4.6, and 5.0. These dimensionaldifferences are supported by the Reference Device (K222792), which covers a wider range of sizes, ensuringsubstantial equivalence.The abutment intended use is consistent with standard practices and does not introduce new risks. Immediateloading is applicable under specific conditions; if these conditions are not met, delayed loading is used, whichaligns with the Primary Predicate's intended use. The subject device's milling center ensures product quality andmanufacturing control, without affecting the abutment's material, design, or function.In conclusion, the additional details in the subject device do not affect its fundamental operation, intended use,or safety. The SD TL Implant System is therefore substantially equivalent to both the Primary Predicate andReference Device.
Subject DeviceReference DeviceReference Device
ManufacturerARUM DENTISTRY Co., Ltd.ARUM DENTISTRY Co., Ltd.Dentium Co., Ltd.
Device NameMulti Pre-Milled CylinderCustomized AbutmentNR Line Implant System
510(k) No.N/AK223634K153268
Intended Use/Indicationsfor useThe SD TL Implant System isindicated for use in partially or fullyedentulous mandibles and maxillae,in support of single or multiple-unitincluding; cementedrestorationsretained,screw retained,orrestorations,overdentureandterminal or intermediate Abutmentfor fixed bridgework.supportAbutments are indicated for screw-retainedsingle restorationsorcement-retained single of multi-unitrestorations. SD TL Implant Systemis dedicated for two stage surgicalprocedures and for immediateloading when there is good primarystability and an appropriate occlusalload. All digitally-designed Pre-MilledAbutments are intended to be sent toan ARUM DENTISTRY validatedmilling center for manufacture.ARUMDentistry'sCustomizedAbutmentsintendedareforattachment to dental implants inprovidesupportordertoforcustomized prosthetic restorations.Customized Abutments are indicatedfor screw-retained single restorationsor cement-retained single or multi-unit restorations. The CustomizedAbutment will be attached to a dentalimplant using the included ARUMDentistry prosthetic screw.CustomizedAbutmentsarecompatible with the implant systemslisted in the Compatibility Table:lmplantRestorativImplante PlatformPlatformBodycompatibilidiameterdiameter(mm)(mm)tv3.8, 4.0,NB 1 SA3.8, 4.0,4.5, 5.0,4.15, 4.25,Implant5.5, 6.0, 6.54.5, 5.0Systemdigitally-designedCustomizedAllAbutments are intended to be sent toan ARUM Dentistry-validated millingcenter for manufacture.The NR Line Implant System isindicated for use in surgical andrestorativeapplicationsforplacement in the bone of the upperor lower jaw to provide support forprosthetic devices, such as artificialteeth, in order to restore the patient'schewing function. NR Line ImplantSystem is indicated also forimmediate loading when goodprimary stability is achieved and withappropriate occlusal loading.
SterilizationSteam Sterilization by user(Provided Non-Sterile)Steam Sterilization by user(Provided Non-Sterile)Steam Sterilization by user(Provided Non-Sterile)
Type ofRetentionScrew-retained or cement retainedScrew-retained or cement retainedScrew-retained or cement retained
Diameter(mm)CAD/CAM Patient-SpecificAbutment: 4.2, 4.35, 4.6, 4.75, 5.0,5.15, 6.2CAD/CAM Patient-SpecificAbutment: 3.8, 4.0, 4.5, 5.0, 5.5,6.0, 6.53.7, 4.3, 5.5, 6.5
MaterialTi-6Al-4V Eli (Conforming to ASTMF136)Ti-6Al-4V Eli (Conforming to ASTMF136)Ti-6Al-4V Eli (Conforming to ASTMF136)
ConnectiontypeSquareInternal HexSquare
SurfaceTreatmentMachinedMachinedMachined
SubstantialEquivalentDiscussion1. SimilaritiesThe Multi Pre-Milled Cylinder is substantially equivalent in designs, dimensions, material, indications, abutment seat,screw seat, anatomical site, and technological characteristics with the identified primary predicate device. Thepatient-specific abutment is similar in fundamental scientific technology to the predicate. The Indications for Use ofthe subject and primary predicate device are identical.2. DifferencesThe differences between the subject device and reference device are the connection type. To support the squareconnection, K153268 was added as reference device. Although, connection type is slightly different but it doesn'timpact product's safety and effectiveness. Therefore, this difference doesn't impact substantial equivalence.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of two triangles on the left side, one red and one light orange. To the right of the triangles is the text "ARUM" in bold, followed by "DENTISTRY" in bold on the line below.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one beige. The words "ARUM DENTISTRY" are in a bold, sans-serif font, with "ARUM" on top and "DENTISTRY" on the bottom.

2) Multi Pre-Milled Cylinder

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains a logo for "ARUM DENTISTRY". The logo consists of two triangles, one red and one light orange, placed next to each other. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in bold, dark brown letters, with "ARUM" on top of "DENTISTRY".

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains a logo for "ARUM DENTISTRY". The logo consists of two triangles, one red and one light orange, on the left side of the text. The text "ARUM" is on the top line, and "DENTISTRY" is on the bottom line, with both lines in a bold, dark brown font.

Performance Data

Non-clinical testing data submitted, referenced or relied on in this submission support demonstrating substantial equivalence.

Biocompatibility

Biocompatibility of Ti-6Al-4V Eli (ASTM F136) demonstrated by the reference ARUM DENTISTRY submission, K213506, using the same materials and manufacturing processes as the subject device.

Sterilization validation

Sterilization validating testing has been performed in accordance with ISO 11137-1 and ISO 11137-2 to verify the sterility assurance level (10°) by selecting and substantiating a 25 kGy dose using method VDmax25. (Referenced from K213506);

End User Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ISO 17665-1, -2, ISO 11737-1,-2, and ISO 11138-1 referenced in K213506;

LAL endotoxin testing according to AAMI / ANSI ST72:2011/(R)2016:

Shelf-Life

The tests to validate the Shelf-Life of the device through the proposed Shelf-Life were conducted using the accelerated aging method in accordance to ASTM F1980 and test results validated 5 years Shelf-Life. Also, the following guidance documents were referred to

  • . Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile. (referenced from K213506);

Non-Clinical Data

Mechanical performance testing was performed according to ISO 14801. For each compatible implant line, worst-case constructs were subjected to static compression and compression fatigue testing. Minor differences in the designs, dimensions, sizes, or compatible implant lines among the subject device, the primary predicate devices, and the reference devices do not affect substantial equivalence. These minor differences do not impact substantial equivalence because these differences are related to the compatible implant designs, or are mitigated by the mechanical performance testing.

Non-clinical performance data submitted to demonstrate substantial equivalence included:

  • . Static and fatigue testing according to ISO 14801.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains a logo for Arum Dentistry. The logo consists of two triangles on the left side, one red and one light yellow. To the right of the triangles is the text "ARUM" in bold, followed by "DENTISTRY" in bold on the line below.

The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate.

MR Environment Condition

Non-Clinical worst-case MRI review was performed to evaluated the SD TL Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan. (2019). Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

Conclusion

The Indications for Use statements are highly similar, differing only in the list of compatible implant systems. Overall, the Technological Characteristics of the Subject device are highly similar to the Predicate device. The Subject device, the Predicate device, and the Reference devices have the same intended use, have similar technological characteristics, and are made of the same materials. The Subject device, the Predicate, and Reference devices encompass the same range of physical dimensions, and are to be sterilized using similar methods. The data included in this premarket notification demonstrate substantial equivalence to the Predicate device listed above. Overall, the Subject device is substantially equivalent to the Predicate device.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.