Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K181621

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-19

    (121 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stable-C Interbody System is an anterior cervical interbody fusion system. The Stable-C Interbody System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Stable-C Interbody System is to be used with supplemental fixation cleared for use in the cervical spine. The system is intended to be used with autogenous or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The system is to be used in patients who have had six weeks of non-operative treatment.

    Device Description

    The Stable-C Interbody System is an anterior cervical interbody device comprised of a cage (lordotic angles of 0 , 6°, and 12°) made from Ti-6Al-4V per ASTM F3001 and two fixation anchors made from Ti-6-Al-4V ELI per ASTM F136. The device is offered in a variety of sizes to accommodate patient anatomy. The devices and instruments are provided clean and non-sterile for steam sterilization at the user's facility.

    AI/ML Overview

    The provided text is a 510(k) premarket notification FDA letter for a medical device called the "Stable-C Interbody System." This document primarily focuses on demonstrating substantial equivalence to pre-existing, legally marketed devices.

    It does NOT describe a study involving an AI/Medical Device that uses machine learning, nor does it provide details about acceptance criteria and performance data in the context of an algorithm's accuracy or effectiveness.

    The "Performance Data" section in the document refers to mechanical testing of the physical implant device to prove its structural integrity and safety, not the performance of an AI algorithm.

    Therefore, I cannot extract the requested information (acceptance criteria, details of a study proving algorithm performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details) from the provided text because it is not relevant to an AI/ML medical device.

    To answer your request, I would need a document describing the validation and performance of an AI/ML-based medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181229

    Validate with FDA (Live)

    Date Cleared
    2018-09-19

    (133 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K110585, K180990, K150765, K1251103, K113559, K153250, K120275, K170550

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Captiva Spine TirboLOX-C™ Cervical IBFD is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one-disc level. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. The Captiva Spine TirboLOX-C™ Cervical IBFD is used to facilitate fusion in the cervical spine and is placed via an anterior approach at the C3 to C7 disc levels with autograft bone and/or allogenic bone graft composed of cancellous bone graft. Patients should have at least six weeks of non-operative to treatment with an intervertebral body fusion device. The device must be used with supplemental fixation.

    Device Description

    The Captiva Spine, Inc. TirboLOX-CTM Cervical IBFD is made from a titanium alloy and is created using 3D printing technologies. The implants are available in various footprints to accommodate a variety of patient anatomies and is provided sterile. The device has a window in the center of device to accept autogenous bone and/or allogenic bone graft. The implant is available in the following configurations: lordotic, anatomically correct and parallel

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, not a study evaluating an AI algorithm. Therefore, the requested information about acceptance criteria, study data, ground truth, and AI performance metrics is not applicable here.

    The document describes the Captiva Spine TirboLOX-C™ Cervical IBFD, an intervertebral body fusion device. It focuses on demonstrating the device's substantial equivalence to legally marketed predicate devices through a series of non-clinical mechanical tests.

    Here's a breakdown of the relevant information present in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with specific quantitative acceptance criteria alongside performance data. Instead, it states that "The results of this non-clinical testing show that the strength of the TirboLOX-C™ Cervical IBFD is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices." This implies that the device's performance in the listed tests met predetermined criteria acceptable for demonstrating equivalence to the predicate devices, but the exact numerical thresholds are not disclosed.

    Device Performance (as reported):

    • ASTM F2077-14: Static Axial Compression, Static Shear, Static Torsion, Dynamic Axial Compression, Dynamic Shear, Dynamic Torsion
    • ASTM 2267-04: Static Subsidence
    • ASTM F1978-12: Abrasion Resistance
    • ASTM F1877-16: Wear Debris
    • Additionally: Expulsion testing was performed.

    The reported performance is qualitative: the device's strength is sufficient for its intended use and substantially equivalent to predicate devices. Specific quantitative results are not provided in this summary document.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided because it's a mechanical device testing, not a clinical study on human data. The "sample" would refer to the number of devices tested, which is not specified. Data provenance like country of origin or retrospective/prospective is not relevant for this type of non-clinical, in-vitro testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth as typically defined for AI/medical imaging studies (e.g., expert consensus on diagnoses) is not relevant for mechanical device testing. The "ground truth" for these tests are the established standards and specifications of the ASTM methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are used in clinical trials or expert reviews to resolve disagreements in interpretations, which is not part of mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a spinal implant, not an AI-powered diagnostic device. No human readers or AI assistance were involved in the testing described in this document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the mechanical performance tests are the established ASTM (American Society for Testing and Materials) standards and test methodologies. These standards define the parameters, procedures, and conditions for evaluating the mechanical properties of medical implants to ensure their safety and effectiveness. The goal is to show the device performs within acceptable limits mandated by these standards and comparably to predicate devices.

    8. The sample size for the training set

    Not applicable. There is no AI algorithm being described, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162496

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-04

    (239 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Foundation™ 3D Interbody Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Foundation™ Cervical implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at one disc level (C2-T1) using autograft bone. Foundation™ 3D Interbody implants are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

    The Foundation™ 3D Interbody Lumbar Cage is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Foundation™ implants are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

    Device Description

    The Foundation™ 3D Interbody Cervical Series Interbody System is an additively manufactured implant comprising of cervical interbody spacers. They are designed to provide mechanical support to the cervical spine while arthrodesis occurs. The cervical line has a partially porous construction and an open architecture with a large variety of footprints and lordosis angles to help optimize implant fit.

    The Foundation™ 3D Interbody Lumbar Series Interbody System is an additively manufactured implant comprising of lumbar interbody spacers. They are designed to provide mechanical support to the lumbar spine while arthrodesis occurs. The lumbar lines feature a wide variety of lordosis and footprint options with fully porous architectures and varying pore sizes to offer increased room for bone growth with mechanical stability.

    The Foundation™ 3D series of intervertebral body fusion devices are made from the titanium alloy Ti-6AL-4V ELI conforming to the ASTM F-136 specifications.

    AI/ML Overview

    This document is a marketing clearance (510(k)) for a medical device, specifically an intervertebral body fusion device. It primarily focuses on demonstrating substantial equivalence to previously cleared devices based on mechanical and material testing, rather than clinical performance or AI algorithm performance.

    Therefore, the requested information regarding acceptance criteria, study design for AI algorithms, human-in-the-loop studies, multi-reader multi-case studies, and ground truth establishment for AI training/testing cannot be extracted from this document. This document describes engineering performance tests, not clinical or AI performance studies.

    Here's what can be extracted related to the device and its testing:

    1. A table of acceptance criteria and the reported device performance:

    The document lists performance tests conducted, but does not provide specific numerical acceptance criteria or the quantitative results of these tests. It only states that the results "show that the strength of the Foundation™ 3D Interbody is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."

    Test ModeAcceptance Criteria (from document)Reported Device Performance (from document)
    Static axial compressionNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    Static torsionNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    Static compressive shearNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    Dynamic axial compressionNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    Dynamic torsionNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    Dynamic compressive shearNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    SubsidenceNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.
    ExpulsionNot specified (implied to meet predicate equivalence)Sufficient for intended use; substantially equivalent to legally marketed predicate devices.

    Regarding the other points, the document does not contain the required information:

    • 2. Sample sized used for the test set and the data provenance: Not applicable for mechanical/material testing.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method for the test set: Not applicable.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not an AI/software device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as this refers to physical/mechanical properties, not diagnostic accuracy.
    • 8. The sample size for the training set: Not applicable (no AI training involved).
    • 9. How the ground truth for the training set was established: Not applicable (no AI training involved).

    In summary, this document is a 510(k) clearance for a physical medical implant, not an AI-powered diagnostic device, and therefore does not contain information about AI acceptance criteria, clinical study designs for AI, or ground truth establishment for patient data. The "performance data" section refers to mechanical engineering tests demonstrating the device's physical strength and equivalence to predicate devices, not its diagnostic or clinical efficacy in a patient population aided by an algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1