The Stable-C Interbody System is an anterior cervical interbody fusion system. The Stable-C Interbody System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Stable-C Interbody System is to be used with supplemental fixation cleared for use in the cervical spine. The system is intended to be used with autogenous or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The system is to be used in patients who have had six weeks of non-operative treatment.
The Stable-C Interbody System is an anterior cervical interbody device comprised of a cage (lordotic angles of 0 , 6°, and 12°) made from Ti-6Al-4V per ASTM F3001 and two fixation anchors made from Ti-6-Al-4V ELI per ASTM F136. The device is offered in a variety of sizes to accommodate patient anatomy. The devices and instruments are provided clean and non-sterile for steam sterilization at the user's facility.
The provided text is a 510(k) premarket notification FDA letter for a medical device called the "Stable-C Interbody System." This document primarily focuses on demonstrating substantial equivalence to pre-existing, legally marketed devices.
It does NOT describe a study involving an AI/Medical Device that uses machine learning, nor does it provide details about acceptance criteria and performance data in the context of an algorithm's accuracy or effectiveness.
The "Performance Data" section in the document refers to mechanical testing of the physical implant device to prove its structural integrity and safety, not the performance of an AI algorithm.
Therefore, I cannot extract the requested information (acceptance criteria, details of a study proving algorithm performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details) from the provided text because it is not relevant to an AI/ML medical device.
To answer your request, I would need a document describing the validation and performance of an AI/ML-based medical device.
Ask a specific question about this device
(133 days)
The Captiva Spine TirboLOX-C™ Cervical IBFD is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one-disc level. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. The Captiva Spine TirboLOX-C™ Cervical IBFD is used to facilitate fusion in the cervical spine and is placed via an anterior approach at the C3 to C7 disc levels with autograft bone and/or allogenic bone graft composed of cancellous bone graft. Patients should have at least six weeks of non-operative to treatment with an intervertebral body fusion device. The device must be used with supplemental fixation.
The Captiva Spine, Inc. TirboLOX-CTM Cervical IBFD is made from a titanium alloy and is created using 3D printing technologies. The implants are available in various footprints to accommodate a variety of patient anatomies and is provided sterile. The device has a window in the center of device to accept autogenous bone and/or allogenic bone graft. The implant is available in the following configurations: lordotic, anatomically correct and parallel
This is a 510(k) premarket notification for a medical device, not a study evaluating an AI algorithm. Therefore, the requested information about acceptance criteria, study data, ground truth, and AI performance metrics is not applicable here.
The document describes the Captiva Spine TirboLOX-C™ Cervical IBFD, an intervertebral body fusion device. It focuses on demonstrating the device's substantial equivalence to legally marketed predicate devices through a series of non-clinical mechanical tests.
Here's a breakdown of the relevant information present in the document:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table with specific quantitative acceptance criteria alongside performance data. Instead, it states that "The results of this non-clinical testing show that the strength of the TirboLOX-C™ Cervical IBFD is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices." This implies that the device's performance in the listed tests met predetermined criteria acceptable for demonstrating equivalence to the predicate devices, but the exact numerical thresholds are not disclosed.
Device Performance (as reported):
- ASTM F2077-14: Static Axial Compression, Static Shear, Static Torsion, Dynamic Axial Compression, Dynamic Shear, Dynamic Torsion
- ASTM 2267-04: Static Subsidence
- ASTM F1978-12: Abrasion Resistance
- ASTM F1877-16: Wear Debris
- Additionally: Expulsion testing was performed.
The reported performance is qualitative: the device's strength is sufficient for its intended use and substantially equivalent to predicate devices. Specific quantitative results are not provided in this summary document.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided because it's a mechanical device testing, not a clinical study on human data. The "sample" would refer to the number of devices tested, which is not specified. Data provenance like country of origin or retrospective/prospective is not relevant for this type of non-clinical, in-vitro testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Ground truth as typically defined for AI/medical imaging studies (e.g., expert consensus on diagnoses) is not relevant for mechanical device testing. The "ground truth" for these tests are the established standards and specifications of the ASTM methods.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are used in clinical trials or expert reviews to resolve disagreements in interpretations, which is not part of mechanical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a spinal implant, not an AI-powered diagnostic device. No human readers or AI assistance were involved in the testing described in this document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the mechanical performance tests are the established ASTM (American Society for Testing and Materials) standards and test methodologies. These standards define the parameters, procedures, and conditions for evaluating the mechanical properties of medical implants to ensure their safety and effectiveness. The goal is to show the device performs within acceptable limits mandated by these standards and comparably to predicate devices.
8. The sample size for the training set
Not applicable. There is no AI algorithm being described, so there is no training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set.
Ask a specific question about this device
(286 days)
The Halis™ Lumbar Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S 1 whose condition requires use of interbody fusion combined with supplemental fixation. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have had six months of non-operative treatment prior to surgery. The Halis™ Lumbar Cage System is intended to be used with autogenous bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation. The devices may be implanted through several surgical approaches: transforaminal, posterior, oblique, and direct lateral. When implanted through the direct lateral approach, the Halis™ Lumbar Cage System is only indicated for use in levels L2-L5.
The Caro™ Cervical Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had six weeks of non-operative treatment prior to surgery. Caro™ Cervical Cage System implants are placed via an anterior approach and is to be used with autogenous bone graft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation.
The Lydia™ Anterior Lumbar Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via a laparoscopic or an open anterior approach. The devices are intended to be used with supplemental fixation. This device is not intended for cervical spine use.
The Dica™ Direction Changeable Lumbar Cage System is indicated for interbody fusion with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six months of non-operative treatment prior to surgery. The devices are intended to be used with supplemental fixation.
The Shanghai Sanyou PEEK Cage System consists of four different models of fusion devices:
- Caro™ Cervical Cage System
- Halis™ Lumbar Cage System
- Lydia™ Anterior Lumbar Fusion System
- . Dica™ Direction Changeable Lumbar Cage System
All components are made of polyetheretherketone (PEEK). All fusion devices are intended for stabilization use and to promote bone fusion during the normal healing process following surgical correction of disorders of the cervical or lumbar spine. These Fusion Devices consist of PEEK cages of various heights, widths and lengths, which can be inserted between two lumbar or cervical vertebral bodies to give support and correction during intervertebral body fusion surgeries. All components have serrations on the superior and inferior surfaces of the implant to aid in fixation. The hollow geometry of the implants allows them to be packed with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Radiopaque Markers manufactured from unalloyed tantalum have been embedded within the implant to allow for visualization in radiographic images. Associated instruments are available to facilitate the implantation of device.
The provided document is a 510(k) premarket notification from the FDA for the Shanghai Sanyou PEEK Cage System. This document focuses on demonstrating substantial equivalence of a new medical device to existing predicate devices, rather than proving a device meets specific acceptance criteria through a clinical study or AI performance evaluation.
Therefore, the information you've requested regarding acceptance criteria, study details (sample size, data provenance, expert adjudication, MRMC studies, standalone performance), and ground truth establishment is not typically found in this type of regulatory submission for a PEEK cage system.
A 510(k) submission for an intervertebral body fusion device primarily relies on non-clinical performance testing to demonstrate that the new device is as safe and effective as a legally marketed predicate device.
Here's what can be extracted and what cannot be from the provided text:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated as pass/fail thresholds for clinical endpoints, but rather compliance with recognized standards. The acceptance criteria essentially boil down to meeting or exceeding the performance of predicate devices in mechanical testing and adherence to relevant ASTM standards.
- Reported Device Performance:
- Mechanical Testing: Static and dynamic axial compression, static shear compression, static torsion.
- Subsidence Testing: Completed according to ASTM F2267-04.
- Overall Conclusion: "Performance testing demonstrates that the subject device meets or exceeds performance of predicate devices demonstrating that the subject devices are substantially equivalent to the predicate devices."
This is how the performance is generally reported in this context, focusing on equivalence rather than specific performance metrics against a defined standard (e.g., 90% accuracy, 0.9 AUC).
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not Applicable. This document describes mechanical performance testing of the device itself, not clinical testing on human subjects with a test set of data. The "test set" would be the physical devices subjected to the mechanical tests.
- Data Provenance: Not applicable for a clinical study, but the manufacturing facilities are in Shanghai, China.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. No human experts were used for adjudication of a ground truth for a clinical dataset. The "ground truth" here is established by the relevant ASTM standards for mechanical properties.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. There's no adjudication method described as it refers to mechanical testing by standard protocols, not expert consensus on clinical data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a physical medical device (PEEK cage), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies are irrelevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the performance testing, the "ground truth" is defined by the ASTM standards (ASTM F2077-14 for mechanical tests and ASTM F2267-04 for subsidence testing) and the established performance characteristics of the legally marketed predicate devices. The device is considered to meet acceptance criteria if its mechanical properties meet or exceed those of the predicates and comply with the specified ASTM standards.
8. The sample size for the training set:
- Not Applicable. There is no "training set" as this is not an AI/machine learning device. The "training" for the device's design and manufacturing implicitly comes from engineering principles, material science, and the design history of similar predicate devices.
9. How the ground truth for the training set was established:
- Not Applicable. As there's no training set in the context of AI, there's no ground truth establishment for it. The standards and predicate device characteristics guide the design and testing process, but not in the sense of a "training set" ground truth.
In summary, the provided document is a regulatory submission for a physical implantable device (intervertebral body fusion cage), not an AI/software as a medical device. The "acceptance criteria" and "study" are described in terms of compliance with regulatory standards and mechanical performance testing demonstrating substantial equivalence to predicate devices, rather than clinical efficacy studies on patient data or AI algorithm performance tests.
Ask a specific question about this device
(373 days)
SAGICO IBF System - ARIA Lumbar implants are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
SAGICO IBF System - ARIA Lumbar are to be filled with autogenous bone graft material. The SAGICO IBF System -ARIA Lumbar implant is intended to be used with supplemental fixation cleared by the FDA to properly utilize this device.
The SAGICO IBF System - ARION Cervical implant is interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) weeks of non-operative treatment.
The SAGICO IBF System - ARION Cervical implant is to be filled with autogenous bone graft material and placed via open anterior approach. The SAGICO IBF System – ARION Cervical implant must be used with additional internal fixation (e.g. anterior plate or cervical pedicle screws) cleared by the FDA to properly utilize this device.
SAGICO IBF System - ARIA Lumbar implants are interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. SAGICO IBF System - ARIA Lumbar implants are provided in different shapes to accommodate various surgical approaches to the lumbar spine (posterior, transforaminal [posterolateral] or lateral and can expand to the desired height. The SAGICO IBF System - ARIA Lumbar implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. The SAGICO IBF System - ARIA Lumbar implants are to be filled with autogenous bone graft material. Protrusions are located on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion.
SAGICO IBF System - ARION Cervical implants are interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. SAGICO IBF System - ARION Cervical implant is intended to be used between two contiguous levels from C2 to T1 and placed via an open anterior surgical approach. The implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. SAGICO IBF System - ARION Cervical implants are comprised of PEEK Optima LT1 and include large central hollow window to be filled with autogenous bone graft material. SAGICO IBF System - ARION Cervical implants features protrusions located on the top and bottom surfaces of PEEK spacers to engage with superior and inferior endplates of the adjacent vertebrae to resist rotational and expulsion. SAGICO IBF System - ARION Cervical includes internal titanium spin blade offering internal fixation and expansion of 1mm in height allowing low profile insertion. SAGICO IBF System - ARION Cervical implant is intended to be used with FDA cleared supplemental fixation to properly utilize the device.
The SAGICO IBF System ARIA Lumbar and ARION Cervical implants are manufactured from medical Grade PEEK (Polyetheretherketone) OPTIMAT LT I (Invibio"") per ISO 10993-1 USP Class VI, and ASTM F2026, Titanium Alloy per ASTM F136 and Tantalum beads /rods to be Grade IINS R05200, IINS R05400 according to ASTM F560.
The SAGICO IBF System ARIA Lumbar and ARION Cervical implants are intervertebral body fusion devices to help restore integrity to the spine in the cervical and lumbar regions.
Here's a breakdown of the acceptance criteria and study information for the SAGICO IBF System - ARIA Lumbar and ARION Cervical devices, based on the provided document:
This document describes a 510(k) premarket notification for intervertebral body fusion devices, which means the approval process focuses on demonstrating substantial equivalence to already legally marketed devices, rather than establishing a completely new safety and effectiveness profile through extensive clinical trials. Therefore, the "study" referred to is a non-clinical performance study comparing the device to predicates, not a clinical trial with human subjects.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Test/Standard | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|---|
| Mechanical/Physical | ASTM F2077 (Standard Test Methods for Intervertebral Body Fusion Devices) | Meet mechanical performance requirements for intervertebral body fusion devices. | Results demonstrate performance substantially equivalent to legally marketed predicate devices. |
| - Static and Dynamic Compression Test | Withstand static and dynamic compressive forces without failure. | Demonstrated substantial equivalence to predicate devices. | |
| - Static and Dynamic Compression Shear Test | Withstand static and dynamic shear forces without failure. | Demonstrated substantial equivalence to predicate devices. | |
| - Static and Dynamic Torsion Test | Withstand static and dynamic torsional forces without failure. | Demonstrated substantial equivalence to predicate devices. | |
| ASTM F2267-04 (Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device under Static Axial Compression) | Exhibit resistance to subsidence under static axial compression within acceptable limits. | Demonstrated substantial equivalence to predicate devices. | |
| ASTM F04.25.02.02 (Static Push-out Test) | Resist expulsion forces from the vertebral body. | Demonstrated substantial equivalence to predicate devices. | |
| Biocompatibility | ISO 10993-1 USP Class VI, and ASTM F2026 (for PEEK Optima LT1) | Biocompatible and safe for implantation. | Manufactured from medical Grade PEEK OPTIMAT LT I (Invibio™) per specified standards, implying compliance. |
| ASTM F136 (for Titanium Alloy) | Biocompatible and safe for implantation. | Manufactured from Titanium Alloy per specified standard, implying compliance. | |
| ASTM F560 (for Tantalum beads/rods) | Biocompatible and safe for implantation. | Manufactured from Tantalum beads/rods per specified standard, implying compliance. |
Note on Acceptance Criteria: For a 510(k), explicit numerical acceptance criteria for each test are often not provided in the summary document but are implicitly understood to be meeting or exceeding the performance of the chosen predicate devices according to recognized standards. The "reported device performance" is essentially that the device performed substantially equivalently to the predicates in these tests.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify exact sample sizes for the non-clinical tests (e.g., how many devices were tested for compression, shear, etc.). It refers to "non-clinical performance data testing."
- Data Provenance: This is a non-clinical, in-vitro study. Therefore, there is no country of origin for patient data. The data is generated from laboratory testing of the devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This question is not applicable in the context of this 510(k) submission. "Ground truth" usually refers to clinical diagnoses or outcomes established by medical experts for a diagnostic or AI-driven system. Here, the "truth" is established by adherence to engineering and biocompatibility standards.
4. Adjudication Method for the Test Set
- Not applicable. This is a non-clinical engineering study, not a clinical study requiring adjudication of expert opinions.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a mechanical intervertebral body fusion device, not a diagnostic or AI-driven imaging interpretation system. MRMC studies are not relevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical implant device, not an algorithm.
7. The Type of Ground Truth Used
- The "ground truth" for these types of devices is established by engineering standards and material specifications. The device's performance is compared against the requirements outlined in standards like ASTM F2077, F2267-04, F04.25.02.02, and material standards (ISO 10993-1, ASTM F2026, F136, F560). The "truth" is that the device meets these established performance and material safety benchmarks, demonstrating substantial equivalence to predicate devices that also meet these benchmarks.
8. The Sample Size for the Training Set
- Not applicable. There is no concept of a "training set" for a physical medical implant device in this context. This is not an AI/machine learning application.
9. How the Ground Truth for the Training Set was Established
- Not applicable, as there is no training set for a physical implant device.
Summary of the Study:
The study described is a non-clinical performance evaluation focusing on the mechanical and material characteristics of the SAGICO IBF System - ARIA Lumbar and ARION Cervical devices. The goal was to demonstrate substantial equivalence to legally marketed predicate devices, a requirement for 510(k) clearance by the FDA. The testing included:
- Mechanical Testing: According to ASTM F2077 (Static and Dynamic Compression, Compression Shear, Torsion) and ASTM F2267-04 (Subsidence), and ASTM F04.25.02.02 (Static Push-out Test).
- Material Biocompatibility: Assessed by the use of medical-grade materials (PEEK Optima LT1, Titanium Alloy, Tantalum beads/rods) that comply with established international and ASTM standards (ISO 10993-1 USP Class VI, ASTM F2026, ASTM F136, ASTM F560).
The conclusion of this non-clinical testing was that the SAGICO IBF System devices demonstrated performance substantially equivalent to legally marketed predicate devices, thus supporting their clearance for market.
Ask a specific question about this device
(239 days)
The Foundation™ 3D Interbody Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Foundation™ Cervical implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at one disc level (C2-T1) using autograft bone. Foundation™ 3D Interbody implants are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
The Foundation™ 3D Interbody Lumbar Cage is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Foundation™ implants are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
The Foundation™ 3D Interbody Cervical Series Interbody System is an additively manufactured implant comprising of cervical interbody spacers. They are designed to provide mechanical support to the cervical spine while arthrodesis occurs. The cervical line has a partially porous construction and an open architecture with a large variety of footprints and lordosis angles to help optimize implant fit.
The Foundation™ 3D Interbody Lumbar Series Interbody System is an additively manufactured implant comprising of lumbar interbody spacers. They are designed to provide mechanical support to the lumbar spine while arthrodesis occurs. The lumbar lines feature a wide variety of lordosis and footprint options with fully porous architectures and varying pore sizes to offer increased room for bone growth with mechanical stability.
The Foundation™ 3D series of intervertebral body fusion devices are made from the titanium alloy Ti-6AL-4V ELI conforming to the ASTM F-136 specifications.
This document is a marketing clearance (510(k)) for a medical device, specifically an intervertebral body fusion device. It primarily focuses on demonstrating substantial equivalence to previously cleared devices based on mechanical and material testing, rather than clinical performance or AI algorithm performance.
Therefore, the requested information regarding acceptance criteria, study design for AI algorithms, human-in-the-loop studies, multi-reader multi-case studies, and ground truth establishment for AI training/testing cannot be extracted from this document. This document describes engineering performance tests, not clinical or AI performance studies.
Here's what can be extracted related to the device and its testing:
1. A table of acceptance criteria and the reported device performance:
The document lists performance tests conducted, but does not provide specific numerical acceptance criteria or the quantitative results of these tests. It only states that the results "show that the strength of the Foundation™ 3D Interbody is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."
| Test Mode | Acceptance Criteria (from document) | Reported Device Performance (from document) |
|---|---|---|
| Static axial compression | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Static torsion | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Static compressive shear | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Dynamic axial compression | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Dynamic torsion | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Dynamic compressive shear | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Subsidence | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
| Expulsion | Not specified (implied to meet predicate equivalence) | Sufficient for intended use; substantially equivalent to legally marketed predicate devices. |
Regarding the other points, the document does not contain the required information:
- 2. Sample sized used for the test set and the data provenance: Not applicable for mechanical/material testing.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- 4. Adjudication method for the test set: Not applicable.
- 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not an AI/software device.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as this refers to physical/mechanical properties, not diagnostic accuracy.
- 8. The sample size for the training set: Not applicable (no AI training involved).
- 9. How the ground truth for the training set was established: Not applicable (no AI training involved).
In summary, this document is a 510(k) clearance for a physical medical implant, not an AI-powered diagnostic device, and therefore does not contain information about AI acceptance criteria, clinical study designs for AI, or ground truth establishment for patient data. The "performance data" section refers to mechanical engineering tests demonstrating the device's physical strength and equivalence to predicate devices, not its diagnostic or clinical efficacy in a patient population aided by an algorithm.