Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K151063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Unimed Reusable SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only.

    Device Description

    The subject devices are Unimed Reusable SpO2 Sensors intended for non-invasive measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) in clinical settings. The sensors are designed for compatibility with GE B40i, and are supplied non-sterile.

    Each sensor consists of a connector, a cable, and a reusable patient-contacting sensor element incorporating a light-emitting diode (LED) and photodetector (PD). The sensors are available in multiple configurations, including finger clip, wrap, and soft-tip types, to accommodate various patient needs and anatomical sites.

    The subject devices operate on the same principle and share similar design features, materials, and performance characteristics as the predicate device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Unimed Reusable SpO2 Sensors, based on the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Device Performance

    CriteriaAcceptance CriteriaReported Device Performance
    Saturation Accuracy, No Motion (70-100%)±3%±3%
    Pulse Rate Accuracy, No Motion±2 bpm (30-250 bpm)±2 bpm (30-250 bpm)
    BiocompatibilityPass ISO 10993 cytotoxicity, skin irritation, and skin sensitivity testsPass ISO 10993 cytotoxicity, skin irritation, and skin sensitivity tests

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Twelve human adult volunteers were enrolled for the clinical study. The study contains more than the minimum 200 data points.
      • Data Provenance: The study was conducted on human adult volunteers and includes sufficient darkly pigmented subjects (three dark subjects with Fitzpatrick Type 5-6). It is a prospective clinical study. The country of origin is not explicitly stated but implies testing in a controlled clinical environment, likely linked to the manufacturer's location or a designated clinical trial site.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document implies that arterial oxygen saturation (SaO2) as determined by co-oximetry was used as the ground truth. This is a direct measurement from blood samples. Therefore, typical "experts" in the sense of human readers adjudicating images are not applicable here. The accuracy of co-oximetry itself is the standard.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. The ground truth (SaO2 by co-oximetry) is a direct, objective measurement, not subject to human interpretation or adjudication in the same way as an imaging study.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a sensor (hardware) for SpO2 and pulse rate measurement, not an AI-assisted diagnostic tool or an imaging system that would involve human reader interpretation. No MRMC study was conducted.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device is a sensor that measures physiological parameters. Its performance is inherent to its design and function, not an algorithm's performance. The "clinical test data" section describes the validation of the sensor's accuracy in vivo.
    6. The type of ground truth used:

      • Arterial oxygen saturation (SaO2) as determined by co-oximetry. This is an objective "gold standard" for blood oxygen measurement.
    7. The sample size for the training set:

      • Not applicable. This device is a hardware sensor, not a machine learning model that requires a training set. The clinical study described is for validation/testing, not training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242562

    Validate with FDA (Live)

    Date Cleared
    2025-07-09

    (315 days)

    Regulation Number
    870.1025
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitor B105M, B125M, B155M, B105P and B125P are portable multi-parameter patient monitors intended to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

    The monitor B105M, B125M, B155M, B105P and B125P are intended for use under the direct supervision of a licensed health care practitioner.

    The monitor B105M, B125M, B155M, B105P and B125P are not Apnea monitors (i.e., do not rely on the device for detection or alarm for the cessation of breathing). These devices should not be used for life sustaining/supporting purposes.

    The monitor B105M, B125M, B155M, B105P and B125P are not intended for use during MRI.

    The monitor B105M, B125M, B155M, B105P and B125P can be stand-alone monitors or interfaced to other devices via network.

    The monitor B105M, B125M, B155M, B105P and B125P monitor and display: ECG (including ST segment, arrhythmia detection, ECG diagnostic analysis and measurement), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), Cardiac Output (C.O.), Entropy, neuromuscular transmission (NMT) and Bispectral Index (BIS).

    The monitor B105M, B125M, B155M, B105P and B125P are able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs, Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.

    Device Description

    The proposed monitors B105M, B125M, B155M, B105P and B125P are new version of multi-parameter patient monitors developed based on the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490) to provide additional monitored parameter Bispectral Index (BIS) by supporting the additional optional E-BIS module (K052145) which used in conjunction with Covidien BISx module (K072286).

    In addition to the added parameter, the proposed monitors also offer below several enhancements:

    • Provided data connection with GE HealthCare anesthesia devices to display the parameters measured from anesthesia devices (Applicable for B105M, B125M and B155M).
    • Modified Early Warning Score calculation provided.
    • Separated low priority alarms user configurable settings from the combined High/Medium/Low priority options.
    • Provided additional customized notification tool to allow clinician to configure the specific notification condition of one or more physiological parameters measured by the monitor. (Applicable for B105M, B125M and B155M).
    • Enhanced User Interface in Neuromuscular Transmission (NMT), Respiration Rate and alarm overview.
    • Provided Venous Stasis to assist venous catheterization with NIBP cuff inflation.
    • Supported alarm light brightness adjustment.
    • Supported alarm audio pause by gesture (Not applicable for B105M and B105P).
    • Supported automatic screen brightness adjustment.
    • Supported network laser printing.
    • Continuous improvements in cybersecurity

    The proposed monitors B105M, B125M, B155M, B105P and B125P retain equivalent hardware design based on the predicate monitors and removal of the device Trim-knob to better support cleaning and disinfecting while maintaining the same primary function and operation.

    Same as the predicate device, the five models (B105M, B125M, B155M, B105P and B125P) share the same hardware platform and software platform to support the data acquisition and algorithm modules. The differences between them are the LCD screen size and configuration options. There is no change from the predicate in the display size.

    As with the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P are multi-parameter patient monitors, utilizing an LCD display and pre-configuration basic parameters: ECG, RESP, NIBP, IBP, TEMP, SpO2, and optional parameters which include CO2 and Gas parameters provided by the E-MiniC module (K052582), CARESCAPE Respiratory modules E-sCO and E-sCAiO (K171028), Airway Gas Option module N-CAiO (K151063), Entropy parameter provided by the E-Entropy module (K150298), Cardiac Output parameter provided by the E-COP module (K052976), Neuromuscular Transmission (NMT) parameter provided by E-NMT module (K051635) and thermal recorder B1X5-REC.

    The proposed monitors B105M, B125M, B155M, B105P and B125P are not Apnea monitors (i.e., do not rely on the device for detection or alarm for the cessation of breathing). These devices should not be used for life sustaining/supporting purposes. Do not attempt to use these devices to detect sleep apnea.

    As with the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P also can interface with a variety of existing central station systems via a cabled or wireless network which implemented with identical integrated WiFi module. (WiFi feature is disabled in B125P/B105P).

    Moreover, same as the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P include features and subsystems that are optional or configurable, and it can be mounted in a variety of ways (e.g., shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for K242562 (Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125P) do not contain information about specific acceptance criteria, reported device performance metrics, or details of a study meeting those criteria for any of the listed physiological parameters or functionalities (e.g., ECG or arrhythmia detection).

    Instead, the documentation primarily focuses on demonstrating substantial equivalence to a predicate device (K213490) by comparing features, technology, and compliance with various recognized standards and guidance documents for safety, EMC, software, human factors, and cybersecurity.

    The summary explicitly states: "The subject of this premarket submission, the proposed monitors B105M/B125M/B155M/B105P/B125P did not require clinical studies to support substantial equivalence." This implies that the changes introduced in the new device versions were not considered significant enough to warrant new clinical performance studies or specific quantitative efficacy/accuracy acceptance criteria beyond what is covered by the referenced consensus standards.

    Therefore, I cannot provide the requested information from the given text:

    1. A table of acceptance criteria and the reported device performance: This information is not present. The document lists numerous standards and tests performed, but not specific performance metrics or acceptance thresholds.
    2. Sample size used for the test set and the data provenance: Not explicitly stated for performance evaluation, as clinical studies were not required. The usability testing mentioned a sample size of 16 US clinical users, but this is for human factors, not device performance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as detailed performance studies requiring expert ground truth are not described.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a patient monitor, not an AI-assisted diagnostic tool that would typically involve human readers.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The document describes "Bench testing related to software, hardware and performance including applicable consensus standards," which implies standalone testing against known specifications or simulated data. However, specific results or detailed methodologies for this type of testing are not provided beyond the list of standards.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not explicitly stated for performance assessment. For the various parameters (ECG, NIBP, SpO2, etc.), it would typically involve reference equipment or validated methods as per the relevant IEC/ISO standards mentioned.
    8. The sample size for the training set: Not applicable, as this is not an AI/ML device that would require explicit training data in the context of this submission.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document focuses on demonstrating that the new monitors are substantially equivalent to their predicate through feature comparison, adherence to recognized standards, and various non-clinical bench tests (e.g., hardware, alarms, EMC, environmental, reprocessing, human factors, software, cybersecurity). It does not contain the detailed performance study results and acceptance criteria typically found for novel diagnostic algorithms or AI-driven devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190008

    Validate with FDA (Live)

    Device Name
    Carescape One
    Manufacturer
    Date Cleared
    2019-12-20

    (352 days)

    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K051367 DASH 3000/4000/5000 monitoring systems, K900598 TRAMSCOPE SYSTEM, K171580 B105/B125 monitor, K151063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CARESCAPE ONE is a multi-parameter physiological patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.

    The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.

    The CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.

    The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.

    Contraindications for using CARESCAPE ONE:

    The CARESCAPE ONE is not intended for use within a controlled MR environment.

    Device Description

    CARESCAPE ONE is a new patient monitor device based on GE Healthcare predicate devices, the Patient Data Module (K071073) and the CARESCAPE B450 (K132533).

    CARESCAPE ONE, with CARESCAPE Software version 3 belongs to the CARESCAPE V3 patient monitor family. The concept of the CARESCAPE ONE is to provide a flexible bedside monitor that can also be used during intra-hospital transport. The flexibility of the CARESCAPE ONE allows the user to configure the monitor's vital sign acquisition for only the parameters they require. This is achieved using plug and play Active Cable Modules (ACM) that connect via medical grade USB ports on the CARESCAPE ONE monitor. Note that the USB ports are not compatible with commercial USB items on the market due to a custom connector design. Each ACM is dedicated to measuring a particular parameter, including ECG/Respiration, Invasive Blood Pressure, Temperature, SpO2, or CO2. The only exception is the Non-Invasive Blood Pressure (NIBP) measurement which does not require a separate ACM since the capability to measure NIBP is built-in to the CARESCAPE ONE monitor itself. The ACM's are CARESCAPE TEMP, CARESCAPE PRES, CARESCAPE ECG, CARESCAPE SpO2 (TruSignal), CARESCAPE SpO2 -Nellcor, CARESCAPE SpO2 - Masimo and CARESCAPE CO2 - LoFlo. CARESCAPE SpO2 - Nellcor, CARESCAPE SpO2 -Masimo, and CARESCAPE CO2 - LoFlo have been developed by their respective companies/manufacturers (OEM) for use with the CARESCAPE ONE. The technology from each OEM has received 510(k) clearance and is adapted to function with the CARESCAPE ONE. The OEM technologies are not new and are not a part of this submission, only their integration into the Parameters/Active Cable Modules for use with the CARESCAPE ONE is covered in this 510(k).

    CARESCAPE ONE provides the users the acquired display values, waveforms, alarms and status messages in compact footprint monitor that runs on an internal battery as well as AC power when connected to the docking station.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the CARESCAPE ONE device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary focuses on demonstrating substantial equivalence to predicate devices rather than providing a direct table of specific acceptance criteria with corresponding performance values for all features. However, it does highlight performance claims and comparisons for key components, specifically arrhythmia detection and SpO2.

    Parameter/FeatureAcceptance Criteria (Implied by Predicate Equivalence & Standards)Reported Device Performance (CARESCAPE ONE)
    Arrhythmia DetectionPerformance equivalent to EK-Pro V13 (predicate).EK-Pro V14: Substantially equivalent EC57 results and performance values to EK-Pro V13.
    SpO2 AccuracyPerformance as per ISO 80601-2-61:2011 and FDA guidance.TruSignal V3 SpO2: Accuracy validated in controlled desaturation study during motion and non-motion conditions, compared to CO-oximetry.
    Other ParametersEquivalent performance to PDM (K071073) and B450 (K132533)Same parameter acquisition technology as PDM, integrated differently. CO2 measurement technology previously cleared and used by GE Healthcare.
    Electrical SafetyCompliance with applicable consensus standards.Bench testing demonstrated compliance.
    EMCCompliance with applicable consensus standards.Bench testing demonstrated compliance.
    Environmental (Mech.)Compliance with applicable consensus standards.Bench testing demonstrated compliance.
    UsabilitySafe and effective for intended users/uses.Extensive usability work, including critical task identification, formative, and summative testing.
    Software Level of ConcernMajorSoftware design, development, verification, validation, and traceability provided.
    BiocompatibilityApplicable for patient-contacting components.Biocompatibility testing completed for patient cables with intermittent skin contact.
    CybersecurityCompliance with FDA guidance for medical device cybersecurity.Security Risk Assessment, Threat model, Penetration testing, system integrity, access, audit, network, and remote service controls.

    2. Sample Size Used for the Test Set and Data Provenance

    • Arrhythmia (EK-Pro V14): The document references ANSI/AAMI EC57:1998 (R)2012. This standard specifies databases (e.g., AHA, MIT-BIH) that are used for testing arrhythmia algorithms. The specific sample size from these databases used for EK-Pro V14 testing is not explicitly stated in this 510(k) summary, but it would be derived from the standard's requirements.
      • Data Provenance: The databases mentioned in EC57 are typically retrospective, internationally recognized ECG databases. The exact country of origin for the data used for this specific EK-Pro V14 validation is not specified, but the standard's databases often compile data from various global sources.
    • SpO2 (TruSignal V3): The document states a "controlled desaturation study." The sample size is not explicitly stated in this summary.
      • Data Provenance: A controlled desaturation study is a prospective clinical study designed to test pulse oximeter accuracy. The country of origin for this study is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Arrhythmia (EK-Pro V14): When using standard databases like those referenced by ANSI/AAMI EC57, the ground truth (arrhythmia annotations) is typically established by multiple, highly experienced cardiologists or cardiac electrophysiologists. The exact number and qualifications for the original annotation of the standard databases are not specified in this document, as the standard itself defines these.
    • SpO2 (TruSignal V3): For controlled desaturation studies, the "ground truth" for SpO2 is provided by a CO-oximeter measurement, which itself is a highly accurate laboratory method for oxygen saturation. This is an objective measurement rather than expert consensus. Therefore, "experts" in the traditional sense for establishing ground truth are not directly applicable here.

    4. Adjudication Method for the Test Set

    • Arrhythmia (EK-Pro V14): For standard databases, ground truth annotation often involves multiple experts reviewing and adjudicating discrepancies, sometimes with a "majority rules" or senior expert decision process. The specific adjudication method for the creation of the databases referenced by EC57 is not detailed in this summary, but would adhere to the methodologies defined by the creators of those databases.
    • SpO2 (TruSignal V3): As the ground truth comes from CO-oximetry, an objective measurement, adjudication by experts is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was NOT done. The document explicitly states: "Clinical studies of the CARESCAPE ONE device performance were not required to establish substantial equivalence." This indicates that human-in-the-loop performance with or without AI assistance was not assessed in this submission for primary equivalence.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Yes, standalone performance was assessed for key algorithms:
      • EK-Pro V14 Arrhythmia Detection algorithm: Its performance was evaluated against the ANSI/AAMI EC57 standard, comparing it to the predicate's EK-Pro V13 algorithm. This is a standalone algorithm performance evaluation.
      • TruSignal V3 SpO2 algorithm: Its accuracy was validated in a controlled desaturation study against CO-oximetry. This is also a standalone algorithm performance evaluation.

    7. The Type of Ground Truth Used

    • Arrhythmia (EK-Pro V14): The ground truth would be expert consensus annotations of ECG waveforms from standard arrhythmia databases (e.g., those specified in ANSI/AAMI EC57).
    • SpO2 (TruSignal V3): The ground truth was objective laboratory measurement using CO-oximetry in a controlled study. This is considered a gold standard for blood oxygen saturation.

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size used for training the EK-Pro V14 arrhythmia detection algorithm or the TruSignal V3 SpO2 algorithm. In 510(k) submissions, training set details (especially for existing, updated algorithms) are often not required to be as extensively disclosed as test set performance, particularly when demonstrating equivalence to a predicate.

    9. How the Ground Truth for the Training Set was Established

    • The document does not explicitly state how the ground truth for the training sets of the EK-Pro V14 or TruSignal V3 SpO2 algorithms was established. For complex algorithms like these, training data is typically meticulously annotated by a combination of clinical experts (for arrhythmia) or using highly accurate reference methods (for SpO2) and then verified. Again, in a 510(k) for an updated algorithm or component, the full developmental history and training data specifics may not be detailed if the focus is on demonstrating equivalent performance post-modifications.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1