Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K132304, K043535, K123754

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.

    Device Description

    The proposed EPIQ and Affiniti Diagnostic Ultrasound Systems are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation. The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers). The removable transducers are connected to the system using a standard technology, multi-pin connectors.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the Philips EPIQ and Affiniti Diagnostic Ultrasound Systems based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics for the device itself (e.g., accuracy, sensitivity, specificity for diagnostic tasks). Instead, it focuses on demonstrating compliance with recognized safety and performance standards for ultrasound devices, and equivalence to predicate devices. The "performance" described relates more to meeting these technical and regulatory standards.

    However, based on the Safety Considerations and Nonclinical Performance Data sections, we can infer the following:

    Acceptance Criteria Category (Inferred)Specific Standard/RequirementReported Device Performance
    Acoustic Output LimitsIEC 60601-2-37 Ed 2.0 (Particular requirements for the basic safety and essential performance of ultrasonic medical and monitoring equipment) and IEC 62359, Ed 2.0 (Ultrasonics -Field characterization – Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields), FDA ultrasound guidance documentComplaint Data: - Ispta.3 ≤ 720 MW/cm2 - MI < 1.9 - TI < 6.0 The system and transducers comply with these standards and the FDA ultrasound guidance document. The devices have acoustic output levels within the Track 3 FDA limits.
    Electrical & Physical SafetyIEC 60601-1: Medical electrical equipment. General requirements for basic safety and essential performance, 2005, Amendment 1, 2012 - IEC 60601-1-2 Medical Electrical Equipment Part 1-2. General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2007 - IEC 60601-1-6 Medical Electrical Equipment - Part 1-6, General Requirements for Basic Safety and Essential Performance - Usability, 2010 - IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2007 - CISPR 11 Class A emissionsComplaint Data: The system and transducers are compliant with these standards. The device is manufactured to the same electrical and physical safety standards as the predicate.
    BiocompatibilityISO 10993: Biological evaluation of medical devicesComplaint Data: The device is manufactured of materials with equivalent biosafety (no new material or material change in application).
    Software PerformanceProduct Specifications - Use NeedsComplaint Data: - Software Verification and Validation testing was performed to ensure the system meets specifications and use needs.
    ElastQ Imaging Precision(Implicitly, to ensure proper functioning and accuracy of the new feature)Complaint Data: A bench test for the EPIQ Diagnostic Ultrasound System in ElastQ Imaging mode was performed to verify precision. (Specific precision values are not given in this summary).
    Transducer X8-2t Performance(Implicitly, to meet design specifications for imaging quality and workflow)Complaint Data: - X8-2t Transducer Verification Report confirms that the X8-2t meets its specifications. Its frame rate has been increased to improve imaging quality and workflow.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not detail specific sample sizes for clinical test sets as it states that no clinical study was required. The non-clinical performance data appears to involve bench testing.

    • Test Set Sample Size: Not applicable for clinical data. For non-clinical tests like ElastQ Imaging Measurement, X8-2t Transducer Verification, and Software Verification and Validation, the sample sizes are not specified but would typically refer to the number of tests performed on the physical hardware/software.
    • Data Provenance: The data provenance is non-clinical performance and engineering testing conducted by Philips Ultrasound, Inc. The country of origin for this testing is not explicitly stated, but Philips Ultrasound, Inc. is located in Bothell, WA, USA. The data is prospective in the sense of being generated during the device development and validation process.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided as the submission relies on non-clinical performance data and substantial equivalence to predicate devices, rather than a clinical study requiring expert ground truth for diagnostic accuracy.

    4. Adjudication Method for the Test Set

    This information is not provided, as no clinical study with human readers/interpreters needing adjudication was conducted or required.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned or performed. The device, an ultrasound system, does not appear to incorporate AI for interpretation that would necessitate such a study in this submission. The new features (ElastQ Imaging and the X8-2t transducer) are extensions of existing ultrasound technology, not AI-driven diagnostic aids.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. The device is a diagnostic ultrasound system that requires a human operator and interpretation, not a standalone AI algorithm for diagnosis. The "ElastQ Imaging" is a quantitative measurement tool, not an autonomous diagnostic algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" would be established by:

    • Acoustic Output: Reference measurements against standardized phantoms and measurement equipment.
    • Safety Standards: Compliance with the specified IEC and ISO standards through documented testing procedures.
    • ElastQ Imaging Measurement: Performance against known physical properties of test phantoms.
    • Software V&V: Meeting predefined software requirements and specifications.
      This is a technical, engineering-based "ground truth" rather than a clinical one from expert consensus or pathology.

    8. The Sample Size for the Training Set

    Not applicable. This submission is for a diagnostic ultrasound system, not an AI/machine learning algorithm requiring a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set for an algorithm was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153480

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-12-16

    (14 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132304, K141261

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearVue850 Diagnostic Ultrasound system is a general purpose, portable, cart based ultrasound system, intended for use by customers in various clinical settings (private offices, clinics, small hospitals, primary and community healthcare centers) by different users (General Practitioners, Paraprofessionals, physician specialists including OBG's, Ultrasound Lab physicians, and nurse practitioners, etc.,

    The System is intended for diagnostic ultrasound imaging in the following modes: B (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Tissue Harmonics, iSCAN, X-Res, Angio, 3D (freehand), 4D and SonoCT, Combined modes includes FloVue, Elastography (strain).

    The system is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications, as listed in FDA 's Diagnostic Ultrasound Indications for Use Form section 1.3 which includes Fetal / OB, Abdominal, Pediatric, Small Organ (breast, thyroid, testicle), Neonatal Cephalic, Trans-rectal, Trans-vaginal, Musculo-skel. (Conventional), Musculo-skel. (Superficial), Other: GYN, Other: Urology, Cardiac Pediatric, Transesophageal (Cardiac), Other (Fetal Echo), Peripheral vessel and Cerebral Vascular.

    Device Description

    The ClearVue850 Diagnostic Ultrasound System is a general purpose, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data in various modes of operation.

    The ClearVue 850 System is substantially equivalent to the currently marketed and predicate ClearVue350/550 system(K120321) in terms of design and fundamental scientific technology. The ClearVue850 Model is provided with additional transducers D2CWC and L12-5 and software features and additional modes. The software feathres of the ClearVue 850 System include -Live Panoramic Imaging, FloVue, Strain Elastography Imaging, Curved ROI Tool, QLAB plug-ins IMT and GI3DQ, Stress Echo protocol.

    The ClearVue 850 System supports CW Doppler transducer with external CW adapter module which gets attached to the SAM transducer connector, an Up/down mechanism for control panel. The system is designed to be highly reliable and easily serviceable.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Philips ClearVue 850 Diagnostic Ultrasound System. It aims to demonstrate substantial equivalence to previously cleared devices.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria or report specific device performance metrics in a pass/fail manner. Instead, the submission focuses on demonstrating substantial equivalence to predicate devices. The "performance" is implicitly demonstrated by stating that the ClearVue 850 shares the same fundamental scientific technology, has similar intended uses and technological characteristics, and has undergone non-clinical testing for compliance with relevant safety and EMC standards.

    The closest to "acceptance criteria" are the Indications for Use outlined in the tables (pages 4-11), which detail the clinical applications and modes of operation for the system and its various transducers. The device "meets" these by being listed for these applications. The "reported device performance" is a general statement of equivalency rather than specific metrics.

    Table of Acceptance Criteria and Reported Device Performance:

    Since explicit quantitative acceptance criteria are not presented, this table lists the stated "Indications for Use" as the de facto "acceptance criteria" that the device is intended to meet, and the "Reported Device Performance" as the claim of substantial equivalence for these indications.

    Acceptance Criteria (Indications for Use)Reported Device Performance (Claim of Substantial Equivalence)
    Diagnostic ultrasound imaging and fluid flow analysis for: Fetal / OB, Abdominal, Pediatric, Small Organ (breast, thyroid, testicle), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skel. (Conventional), Musculo-skel. (Superficial), Other: GYN, Other: Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac), Other (Fetal Echo), Peripheral vessel and Cerebral Vascular. The document also lists specific applications for each transducer, some of which are 'new indications' (N) and some 'previously cleared' (P).The ClearVue 850 System is substantially equivalent to the currently marketed and predicate ClearVue350/550 system (K120321) in terms of design and fundamental scientific technology. The ClearVue 850 is demonstrated to meet these indications for use through substantial equivalence to predicate devices (K120321, K132304, K141261). The document repeatedly states that the device introduces "no new indications for use, features, or technologies as compared to the currently marketed and predicate devices (EPIQ/ClearVue) that require clinical testing." Specific additions/equivalences mentioned: - L12-5 and D2CWC transducers: Added with equivalent additional indication for use to K132304.- FloVue: New indication, substantially equivalent to the coded pulse mode of GE K141261 (both help in simultaneous imaging of tissue and blood flow).- Elastography: Already cleared with EPIQ K132304.- Other software features: Live Panoramic Imaging, Strain Elastography Imaging, Curved ROI Tool, QLAB plug-ins IMT and GI3DQ, Stress Echo protocol, CW Doppler transducer with external CW adapter module.
    Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37 standards (Non-clinical performance data).Non-clinical tests were performed to demonstrate compliance with these standards.

    Study Information:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      The document explicitly states: "Summary of Clinical Tests: The ClearVue850 System introduces no new indications for use, features, or technologies as compared to the currently marketed and predicate devices (EPIQ/ClearVue) that require clinical testing." This indicates that no new primary clinical testing (test set) was performed specifically for this 510(k) submission. The substantial equivalence relies on the performance of the predicate devices. Therefore, there is no information about a test set sample size or its provenance in this document.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      As no new clinical testing was performed for this submission, this information is not provided. The ground truth for the predicate devices' clearances would have involved relevant medical experts, but details are not in this document.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable, as no new clinical test set was used for this submission.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No MRMC study was performed or is mentioned. The device is a diagnostic ultrasound system and not an AI-assisted diagnostic tool in the sense of providing automated interpretations to readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This is a diagnostic ultrasound system, which inherently involves a human operator and interpreter. It is not an algorithm that provides standalone diagnostic outputs without human interaction.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the purposes of this 510(k) submission, the "ground truth" relies on the established safety and effectiveness of the predicate devices. The document implies that the clinical safety and effectiveness of ultrasound systems with these characteristics are "well accepted." For the predicate devices, their ground truth would likely have been established through a combination of expert clinical interpretation, correlation with other imaging modalities, and potentially pathology or surgical outcomes, depending on the specific application. However, these details are not provided for the ClearVue 850 in this submission.

    7. The sample size for the training set:
      Not applicable. This document describes a traditional 510(k) for an ultrasound system, not an AI/Machine Learning device that requires a distinct "training set" in the computational sense. The "training" for the device's development would be through extensive engineering, design, and internal validation processes based on established ultrasound principles and medical knowledge.

    8. How the ground truth for the training set was established:
      Not applicable for the same reasons as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150773

    Validate with FDA (Live)

    Date Cleared
    2015-07-23

    (121 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K120060, K132304

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications: Fetal: Abdominal (renal & GYN/pelvic); Pediatric, Small Organ (breast, testes, thyroid); Adult Cephalic; Trans-rectal, Trans-vaginal, Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult& pediatric); Peripheral Vascular (PV): and Urology (including prostate).

    Device Description

    E-CUBE 15 product is an ultrasound imaging system for medical diagnosis. The system platform provides optimal patient diagnosis workflow with the 18.5" (or 19.5") wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.
    Modes of operation:

    1. Signal Mode:
      B(2D) mode, M mode, Color Flow(CF) mode, Power Doppler(PD) mode, Pulsed Wave Doppler(PWD) mode, Tissue Harmonic Imaging (THI)
    2. Combination Mode:
      B/M, B/CF, B/PD, B/PWD, B/CF/PWD, B/PD/PWD, B/CF/M
      Acoustic output track: Track 3
    AI/ML Overview

    The provided text describes the "E-CUBE 15 Diagnostic Ultrasound System" and its various transducers but does not contain information about specific acceptance criteria, studies proving performance, sample sizes, data provenance, expert ground truth establishment, or multi-reader multi-case (MRMC) comparative effectiveness studies related to any AI components. The document is a 510(k) premarket notification to the FDA, focusing on establishing substantial equivalence to previously cleared predicate devices for various clinical applications and operating modes.

    The document lists the intended uses and technical specifications of the E-CUBE 15 system and its transducers, comparing it to predicate devices (E-CUBE 9 and iU22). The "Summary of Non-Clinical Tests" mentions evaluation for biocompatibility, acoustic output, and thermal, electrical, electromagnetic, and mechanical safety, conforming to applicable medical device safety standards. It also lists quality management system measures applied during development.

    Crucially, the "Summary of Clinical Tests" section explicitly states:
    "The subject of this premarket submission, E-CUBE 15, did not require clinical studies to support substantial equivalence."

    This means that for the purposes of this 510(k) submission, the device's substantial equivalence was established through non-clinical testing and comparison to predicate devices, rather than through clinical trials with defined acceptance criteria for AI performance, independent test sets, expert ground truth, or MRMC studies.

    Therefore, the requested information regarding acceptance criteria, specific performance metrics, sample sizes, data provenance, expert adjudication, MRMC studies, or standalone algorithm performance for AI components cannot be extracted from this document, as it appears the submission did not rely on such data for device clearance. The document focuses on hardware aspects and basic functional modes, not advanced AI features requiring such rigorous evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150533

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-05-09

    (68 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K112953, K132304, K070792, K110482, K140773, K082098

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Teratech Corporation Terason™ uSmart3200T is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Ophthalmic, Fetal, Abdominal, Intra-operative (Spec. and Neuro.), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esophageal (non-Card.) Musculo-skel. (Convent.), Musculo-skel. (Superfic); Cardiac (Adult & Pediatric); Trans-esophageal (Cardiac) Peripheral Vascular.

    Device Description

    The Terason uSmart3200T ultrasound system is a portable tablet-style, fullfeature, general purpose diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. The Terason uSmart3200T Ultrasound System is equivalent to the previously cleared versions of the uSmart3200T Ultrasound Systems. The modification includes the addition of 5 transducers (9MC3, 8TE3, PDOF, 8L2, 8V3A), the Trans-esophageal (non-Cardiac and Cardiac) IFUs and the Ophthalmic IFU associated with the 12L5A and 15L4 transducers with no change to the tablet-style computer form factor.

    AI/ML Overview

    The document provided is a 510(k) premarket notification for the Terason uSmart3200T Ultrasound System, detailing its substantial equivalence to predicate devices. It focuses on the safety and performance aspects of the device and its transducers.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in a consolidated manner for the entire device. Instead, it describes various tests and standards met. However, it does provide specific acoustic output limits and the measured values for different transducers.

    Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance (Worst Case for each Transducer)
    Acoustic OutputNEMA UD 2-2004, NEMA UD 3-2004, IEC 60601-2-37
    Global Maximum ISPTA.3- 12L5A: 597 mW/cm²
    - 4V2A: 603 mW/cm²
    - 5C2A: 660 mW/cm²
    - 8EC4A: 633 mW/cm²
    - 16HL7: 554 mW/cm²
    - 15L4: 563 mW/cm²
    - 8L2: 598 mW/cm²
    - 8V3A: 560 mW/cm²
    - 9MC3: 577 mW/cm²
    - 8TE3: 245 mW/cm²
    - PDOF: 506 mW/cm²
    TI (Thermal Index)- 12L5A: 3.3
    - 4V2A: 5.6
    - 5C2A: 4.7
    - 8EC4A: 2.29
    - 16HL7: 1.22
    - 15L4: 5.8
    - 8L2: 2.8
    - 8V3A: 4.7
    - 9MC3: 2.8
    - 8TE3: 1.0
    - PDOF: 4.2
    MI (Mechanical Index)- 12L5A: 1.8
    - 4V2A: 1.5
    - 5C2A: 0.7
    - 8EC4A: 1.8
    - 16HL7: 1.6
    - 15L4: 1.7
    - 8L2: 1.7
    - 8V3A: 1.7
    - 9MC3: 1.3
    - 8TE3: 1.3
    - PDOF: 0.1
    B-Mode AccuracyAIUM Quality Assurance Manual for Gray-Scale Ultrasound ScannersEvaluated (details not specified)
    Doppler AccuracyAIUM Performance Criteria and Measurements for Doppler Ultrasound DevicesEvaluated (details not specified)
    Doppler SensitivityAIUM Performance Criteria and Measurements for Doppler Ultrasound DevicesEvaluated (details not specified)
    Electrical SafetyIEC60601-1Compliance verified
    EMCIEC 60601-1-2, CISPR11 Class BCompliance verified
    UsabilityIEC60601-1-6Compliance verified
    BiocompatibilityISO 10993 Part 5 and Part 10Biocompatibility reports fulfilled
    SoftwareIEC 62304 (Software Life Cycle Process), IEC 62366 (Application for Usability)Undergone Quality Assurance testing

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "sample size" for a specific test set of patient data. The studies performed are primarily non-clinical bench tests on the ultrasound system and its transducers.

    • Data Provenance: Not applicable as the documented tests are primarily for physical and technical performance and safety, not diagnostic accuracy on patient data. The document mentions acoustic tank evaluations for transducers.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The reported tests are not clinical studies that require expert-established ground truth on patient cases. They are technical performance evaluations against engineering standards and specifications.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical test set with patient data requiring expert consensus or adjudication is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a diagnostic ultrasound system, not an AI-powered image analysis tool. The document describes the system's technical and safety characteristics, not its comparative effectiveness in a clinical reading environment with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical imaging device, not a standalone algorithm. The "performance" discussed relates to the physical and technical capabilities of the ultrasound system and its transducers.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The type of "ground truth" for the tests performed is primarily based on technical standards and specifications (e.g., NEMA UD 2, NEMA UD 3, IEC 60601-2-37 for acoustic output; AIUM guidelines for B-Mode and Doppler accuracy/sensitivity; ISO 10993 for biocompatibility). For a diagnostic device, the "ground truth" is that it performs according to these established and recognized technical and safety benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware-based diagnostic imaging system, not an AI model that requires a training set of data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no AI model or training set described in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1