K Number
K120060

Validate with FDA (Live)

Device Name
E-CUBE 9
Date Cleared
2012-03-16

(67 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).

Device Description

E-CUBE 9 product is an ultrasound imaging system for medical diagnosis. This product can be used for the applications of abdominal, obstetrics, gynecology, small parts, cardiology, vascular, etc. The system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality

AI/ML Overview

The provided documentation does not contain detailed acceptance criteria for specific performance metrics of the E-CUBE 9 ultrasound system in a clinical context. Instead, it focuses on demonstrating substantial equivalence to a predicate device (K060993 LOGIQ P5/A5 Diagnostic Ultrasound System) through non-clinical tests and a declaration that clinical studies were not required.

Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for clinical performance cannot be extracted from this document.

The document mainly highlights:

  • Non-Clinical Tests: Evaluation for biocompatibility, acoustic output, thermal, electrical, electromagnetic, and mechanical safety, conforming to applicable medical device safety standards.
  • Voluntary Standards Compliance: The device's application complies with voluntary standards as detailed in the premarket submission.
  • Quality Management System Measures: Mention of Medical Device Risk Management, Requirements Reviews, Design Reviews, Component Verification, Integration Review (System Verification), Performance Testing (System Verification), Safety Testing (Compliance Test), and Design Validation.

The key statement regarding clinical data is:

"The subject of this premarket submission, E-CUBE 9, did not require clinical studies to support substantial equivalence."

This indicates that the FDA granted clearance based on non-clinical data and the device's technical similarity to a previously cleared device, rather than requiring new clinical performance studies with specific acceptance criteria that would typically be evaluated in such studies.

{0}------------------------------------------------

MAR 1 6 2012

(120060)

Page 1 at 2

510(K)-E-CUBE 9

510(k) Summary

In accordance with 21CFR807.92, the following summary of information is provided;

DateJan 5 th 2012
Submitter:ALPINION MEDICAL SYSTEMS Co., Ltd.Address: 1,6 and 7FL, 222-22, Guro-dong, Guro-gu, Seoul, 152-848, Korea Republic
Primary Contact PersonDonghwan KimQARA ManagerAddress: 1,6 and 7FL, 222-22, Guro-dong, Guro-gu, Seoul, 152-848, Korea RepublicPhone: +82 70 7465 2068Fax: +82 2 851 5594Email: donghwan.kim@alpinion.com
Secondary Contact PersonYuchi ChuAddress: Suite 229, 10604 NE 38th Place, Kirkland, WA 98033, United StatesPhone: 425 949 4907Fax: 425 949 4908Email: ychu@alpinionus.com
Device Trade Name:E-CUBE 9
Common/Usual Name:Ultrasonic Pulsed Doppler Imaging System
Classification NamesSystem, Imaging, Pulsed Doppler Ultrasonic
Product Code:Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYNUltrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYODiagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX
Predicate Device(s)K060993 LOGIQ P5/A5 Diagnostic Ultrasound System

ALPINION MEDICAL SYSTEMS Co., Ltd.

.

.

{1}------------------------------------------------

K120060
Page 2 of 2

E-CUBE 9 product is an ultrasound imaging system for medical Device Description: diagnosis. This product can be used for the applications of abdominal, obstetrics, gynecology, small parts, cardiology, vascular, etc.

The system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality

Indications For Use:

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).

Technology:

Determination of Substantial Equivalence:

E-CUBE 9 employs the same fundamental scientific technology as its predicate device.

Summary of Non-Clinical Tests:

E-CUBE 9 has been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. E-CUBE 9 and its application comply with voluntary standards as detailed in this premarket submission. The following quality management system measures were applied to the development of E-CUBE 9:

  • Medical Device Risk Management �
  • Requirements Reviews �
  • Design Reviews �
  • Component Verification �
  • Integration Review (System Verification) �
  • Performance Testing (System Verification) �
  • Safety Testing (Compliance Test) �
  • � Design Validation

Transducer materials and other patient contact materials are biocompatible.

Summary of Clinical Tests:

The subject of this premarket submission, E-CUBE 9, did not require clinical studies to support substantial equivalence.

Alpinion Medical Systems Co., Ltd. Considers E-CUBE 9 to be as Conclusion: safe, as effective, and performance is substantially equivalent to the predicate device.

ALPINION MEDICAL SYSTEMS Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA or the requirements will be published in guidance documents.

ALPINION MEDICAL SYSTEMS Co., Ltd.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Alpinion Medical Systems Co., Ltd. % Mr. Yuchi Chu Correspondent AUS Co., Inc. 10604 NE 38th Place, Suite 229 KIRKLAND WA 98033

Re: K120060

Trade/Device Name: E-CUBE 9 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 9, 2012 Received: January 9, 2012

Dear Mr. Chu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the E-CUBE 9, as described in your premarket notification:

Transducer Model Number

SC1-6L3-12H
C1-6SP1-5
SVC1-6E3-10
VC1-6L3-8
L3-12SP3-8

MAR. 1 6 2012

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.

Sincerely Yours,

Janine M. Morris

Acting Directd Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K111864

Device Name: E-CUBE 9

Indications for Use:

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)
Division Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K120060

ALPINION MEDICAL SYSTEMS Co., Ltd.

{5}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
FetalPPPPPPP
AbdominalPPPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNN
Small Organ(breast, testes, thyroid)PPPPPNP
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPP
Trans-vaginalPPPPPP
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPNP
Musculo-skeletal(Superficial)PPPPPNP
Intravascular
Cardiac AdultPPPNPPPP
Cardiac PediatricNNNNNNNN
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPPPPPNP
Urology (including prostate)NNNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

E-2

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.

sion Sign Division of Radiologica! Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K =

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with SC1-6 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationCombined*(Specify)Other**(Specify)
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImaging
Ophthalmic
FetalPPPPPPP
AbdominalPPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNN
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)NNNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
K120060

{7}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with C1-6 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
FetalPPPPPPP
AbdominalPPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNN
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)NNNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

The even do not write below this line-continue on another bage in meeded) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological DevicesOffice of In Vitro Diagnostic Device Evaluation and Safety
510KK120868

{8}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with SVC1-6 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
FetalPPPPPPPP
AbdominalPPPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNN
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)NNNNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

E-2

(Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices

ામદિક વા on and Satery

ETOK

{9}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with VC1-6 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
FetalPPPPPPPP
AbdominalPPPPPPPP
Intra-operative (Specify).
Intra-operative (Neuro)
Laparoscopic
PediatricNNN.NNNNN
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)NNNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)
----------------------------------------

ALPINION MEDICAL SYSTEMS Co., Ltd.

(Division Sign-Off)
Division of Radiological Devices
Office of Na Vitro Diagnostic Device Evaluation and Safety

510K. K120060

{10}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with L3-12 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNN
Small OrganPPPPPPP
(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPNP
Musculo-skeletal(Superficial)PPPPPNP
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPPPPPNP
Urology (including prostate)

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K

{11}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with L3-12H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationTissueHarmonicImagingCombined*(Specify)Other**(Specify)
BMPWDCWDColorDopplerPowerDoppler
Ophthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNN
Small Organ(breast, testes, thyroid)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPP
Musculo-skeletal(Superficial)PPPPPP
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPPPPPP
Urology (including prostate)

N = new indication, P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF MEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-OH)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

610K.

{12}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with SP1-5 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
Fetal
AbdominalPPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNN
Small Organ
(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethrai.
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeietal
(Superficial)
Intravascular
Cardiac AdultPPPNPPPP
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANQTHER PAGE IF NEEDED) Concurrence of CDRH; Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.

(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K. K120068

E-d

{13}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with E3-10 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColor DopplerPower DopplerTissue Harmonic ImagingCombined* (Specify)Other** (Specify)
Ophthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPP
Trans-vaginalPPPPPP
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)NNNNNN

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.

Division of Radiological Devices Office of In Vitro Diagnostic luation and Safety

510K Kianna66

{14}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with L3-8 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNN
Small Organ(breast, testes, thyroid)NNNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)NNNNNN
Musculo-skeletal(Superficial)NNNNNN
Intravascular
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselNNNNNN
Urology (including prostate)

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
510KOffice of in Vitro Diagnostic Device Evaluation and Safety
K120068

{15}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

E-CUBE 9 with SP3-8 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerPowerDopplerTissueHarmonicImagingCombined*(Specify)Other**(Specify)
Ophthalmic
Fetal
AbdominalNNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNN
Small Organ(breast, testes, thyroid)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Cardiac Adult
Cardiac PediatricNNNNNNN
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vessel
Urology (including prostate)

N = new indication; P = previously cleared by FDA; E = added under appendix

  • Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(Please do not write below this line-continue on another Page if needed) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.

(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K120068

N/A