Found 2 results

510(k) Data Aggregation

    K Number
    K231915

    Validate with FDA (Live)

    Device Name
    Zfx Abutments
    Date Cleared
    2024-03-01

    (246 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zfx Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandbular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments or superstructures and/or hybrid abutment crowns for use with Zfx TiBase or Pre-Milled Blank Abutments are intended to be sent to a Zfx validated milling center for manufacture.

    The Zfx TiBase abutments for the 03.25 mm Certain implant bodies, and 03.1 mm Eztetic implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.

    Device Description

    The subject device, Zfx Abutments, consists of two abutment types, a TiBase and a Pre-milled Abutment Blank, with corresponding retaining screws. Zfx Abutments are offered in a variety of connections compatible with ZimVie dental implants. All abutments and screws are provided non-sterile.

    TiBase Abutments: Subject device TiBases are two-piece abutments. The pre-manufactured titanium base component is the apical part. The coronal part of the two-piece abutment is a CAD/CAM designed and manufactured superstructure. The subject device TiBase abutments are available in hexed/engaging and non-hexed/nonengaging configurations. The engaging TiBase abutments are intended for single and multi-unit restorations and the non-engaging TiBases are intended for multi-unit restorations. The superstructure is intended to be manufactured at a Zfx validated milling center.

    Pre-milled Abutment Blanks: The Pre-milled Abutment Blank is a cylindrical titanium alloy abutment designed for patient-specific abutment fabrication with CAD/CAM technology. The patient-specific abutment milled from a Pre-milled Abutment Blank is secured directly to the implant using a retaining screw. Pre-milled Abutment Blanks are available in an engaging/hexed design only for single-unit and multi-unit restorations.

    Retaining Screws: Corresponding retaining screws are packaged with the abutment and replacement screws are available individually. All screws are compatible with the corresponding ZimVie dental implants.

    Material Composition: All subject device abutments are made of titanium alloy Ti-6Al-4V ELI conforming to ASTM F136. Screws are made of either titanium alloy Ti-6A1-4V ELI conforming to ASTM F136, or Gold-Tite® screws with stainless steel conforming to ASTM F138. Gold-Tite screws have a gold-plating conforming to ASTM B488 and ASTM B571. Zirconia superstructures for use with the TiBase Abutments are made of zirconia conforming to ISO 13356.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for Zfx Abutments, and as such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and supporting data in the way a clinical study report or a formal performance evaluation report would.

    Therefore, the information required to answer many of the questions regarding acceptance criteria, sample sizes, expert involvement, and ground truth establishment is not present in this document. The document primarily outlines the device's technical specifications, indications for use, and a comparison to predicate devices, supported by non-clinical testing data (mechanical testing, biocompatibility, sterilization, etc.).

    However, I can extract the information that is available and explain why other requested details are missing.


    Acceptance Criteria and Device Performance (Based on "Equivalence to Marketed Devices" and "Performance Data" sections):

    This document describes a substantial equivalence submission, meaning the acceptance criteria are primarily demonstrated by showing the device is as safe and effective as a legally marketed predicate device. The performance is assessed through non-clinical testing to ensure it meets established standards comparable to the predicate.

    Acceptance Criterion (Implicit for Substantial Equivalence)Reported Device Performance (Non-Clinical/Design)
    Indications for Use Equivalence"Zfx Abutments (subject device) are substantially equivalent in design, function, material, and Indications for Use to the Primary Predicate, DESS Dental Smart Solutions abutments cleared in K222288. All are intended for use with endosseous dental implants in the maxilla and mandible to provide support for single and multi-unit restorations."
    Material Equivalence"All subject device abutments are made of titanium alloy Ti-6Al-4V ELI conforming to ASTM F136." "Screws are made of either titanium alloy Ti-6A1-4V ELI conforming to ASTM F136, or Gold-Tite® screws with stainless steel conforming to ASTM F138." "Zirconia superstructures for use with the TiBase Abutments are made of zirconia conforming to ISO 13356." This is compared to comparable materials in the predicate.
    Product Design and Manufacturing Process Equivalence"All digitally designed subject device abutments and primary predicate abutments are to a validated milling center for manufacture. Subject device abutments are similar in range of sizes, connections, and technological characteristics to the DESS Dental Smart Solutions, K22288." "Design parameters for the TiBase CAD/CAM zirconia superstructure are listed in the following tables (Tables 2-5)." "Design parameters for the CAD/CAM patient-specific abutment using a Pre-milled Abutment Blank are included in Table 6."
    Sterilization Efficacy (Non-clinical)"sterilization validation according to ISO 17665-1"
    Biocompatibility (Non-clinical)"biocompatibility testing according to ISO 10993-1 Table A-1, ISO 10993-5, and ISO 10993-10"
    Reprocessing Validation (Non-clinical)"reprocessing validation according to ISO 17665-2"
    Mechanical Strength/Fatigue (Non-clinical)"mechanical testing according to ISO 14801 to determine that the subject device has sufficient strength for its intended use."
    MR Safety (Non-clinical)"MR Safety testing was conducted according to ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 for a determination of MR Conditional."
    Compatibility with Existing Systems"reverse engineering analysis to confirm compatibility with the Sirona inCoris Meso Blocks" (for TiBase).

    Missing Information and Explanations:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not provided. This document details a 510(k) submission based on non-clinical testing for substantial equivalence, not a clinical trial or performance study with human data. The "Performance Data" section explicitly states: "No clinical data were included in this submission." Therefore, there is no "test set" in the context of clinical images or patient data to analyze. The "samples" used were physical samples for mechanical, sterilization, and biocompatibility testing. The data provenance would be from laboratory tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not applicable/Not provided. As no clinical data or clinical "test set" was used, there was no need for expert adjudication to establish ground truth from patient data. The ground truth for mechanical and material properties is established by engineering specifications and industry standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable/None. No clinical test set to adjudicate.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device (Zfx Abutments) is a physical dental implant component, not an AI software/device. Therefore, an MRMC study or AI assistance is not relevant to its performance evaluation for this submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. As above, this is a physical medical device, not an algorithm or AI.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for demonstrating substantial equivalence for this device relies on engineering specifications, material standards (e.g., ASTM, ISO), mechanical test results (e.g., forces to fracture), and comparative analysis with the predicate device's established safe and effective performance. There is no clinical outcomes ground truth cited here.
    7. The sample size for the training set:

      • Not applicable. This refers to a dataset for training an AI model. This device is not an AI/software.
    8. How the ground truth for the training set was established:

      • Not applicable. As above, no training set for an AI model.

    In summary, this document is a regulatory clearance letter focused on substantial equivalence for a physical dental device. The "study" proving it meets acceptance criteria consists of various non-clinical bench tests and a comparison to a predicate device's specifications and performance, rather than a clinical study with patients or an AI model's performance evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173374

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-13

    (47 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K061410, K133339, K113753, K132258

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

    Device Description

    The purpose of this submission is to obtain 510(k) premarket clearance for the TSVIM BellaTek® Encode® Healing Abutments compatible with Tapered Screw-Vent® and Trabecular Metal implant systems. Abutments sizing is based on implant platform diameter.

    TSV™ BellaTek® Encode® Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. They have the added design feature of machined markings for identification when taking an abutment level impression or an intraoral scan/digital impression. The occlusal surface of the device include markings that provide information about the mating implant's position and orientation.

    The principal of operation and Encode "Coding Scheme of the same as the primary predicate device. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent", and Trabecular Metal implant lines.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria. Instead, it is a 510(k) summary for a medical device (TSV™ BellaTek® Encode® Healing Abutments) demonstrating substantial equivalence to a predicate device.

    The document explicitly states: "No clinical data were included in this submission." and relies on non-clinical testing data and equivalence to a predicate device.

    Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth, expert opinions, or MRMC studies, as this information is not present in the provided text.