510(k) Data Aggregation

    K Number
    K151209

    Validate with FDA (Live)

    Device Name
    InterOss
    Manufacturer
    Date Cleared
    2015-11-06

    (184 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K113246, K120601

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InterOss® small granules are recommended for:

    • Augmentation or reconstructive treatment of the alveolar ridge .
    • . Filling of infrabony periodontal defects
    • . Filling of defects after root resection, apicocectomy, and cystectomy
    • . Filling of extraction sockets to enhance preservation of the alveolar ridge
    • . Elevation of the maxillary sinus floor
    • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

    InterOss® large granules are recommended for:

    • . Augmentation or reconstructive treatment of the alveolar ridge
    • . Elevation of the maxillary sinus floor
    • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    Device Description

    InterOss® is a hydroxyapatite material derived from Australian bovine bone. The osteoconductive mineral structure is produced from bone through a multi-step purification process. Following placement in bony voids or gaps, InterOss acts as an osteoconductive scaffold for the ingrowth of adjacent viable bone. InterOss® gradually resorbs and is replaced with bone during the healing process.

    InterOss® is available in granule form and is packaged in vials or a syringe-like applicator.

    InterOss®, in the vial form, will be available to the United States market in 8 versions: Filled with 0.25g, 0.5g, 1.0g, 2.0g, or 5.0g of small granules (0.25 - 1.0mm) or filled with 0.5g, 1.0g, or 2.0g of large granules (1.0 - 2.0mm).

    InterOss , in the syringe-like applicator form, will be available to the United States market in 6 versions: Filled with 0.25cc, 0.5cc, or 1.0cc of small granules (0.25 - 1.0mm) or filled with 0.5cc, 1.0cc, or 1.5cc of large granules (1.0mm - 2.0mm).

    The syringe-like applicator was designed to deliver InterOss® granules more precisely to the intended treatment site without having to use other sterile instruments. The InterOss® granules can be wetted with either the patient's blood or sterile physiological saline solution back and then pressing down on the plunger. A removable filter cap prevents granules from falling out of the syringe-like applicator during storage and wetting.

    During the manufacturing process of InterOss®, the granules are placed into a polymer syringe-like applicator or a glass vial, the vial or syringe is then capped with a rubber cap (the vial is also sealed with an aluminum cap), packaged into a polyethylene terephthalate tray, covered with a Tyvek lid, sealed, and then sterilized by gamma irradiation. The sterilized device is place in a protective package (outer box) along with its Instructions for Use and doctors notes.

    All InterOss® products are supplied sterile and are intended for single use only.

    AI/ML Overview

    The provided text describes the regulatory clearance of a bone grafting material called InterOss® and does not contain information about a medical device that relies on AI or offers performance metrics that would typically be described with acceptance criteria like sensitivity, specificity, or AUC, as commonly seen in AI/ML device descriptions.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through various tests and comparisons. The "acceptance criteria" here are implied by the standards for bone grafting materials and the equivalence to the predicate devices.

    Here's an attempt to structure the information based on the request, reinterpreting "acceptance criteria" as the properties and test outcomes required for substantial equivalence in this context:

    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / Acceptance Criteria (Implied by Equivalence to Predicate & Standards)InterOss® Reported Performance
    Physical/Chemical Properties
    Crystallinity ratio106%
    Phase purity100% hydroxyapatite
    Ca/P ratio1.67
    MorphologyGranular and porous
    Average granule size (small)568 µm (0.25 - 1.0mm range)
    Average granule size (large)1625 µm (1.0 - 2.0mm range)
    Pore interconnectivityPositive
    Absence of heavy metalsConfirmed
    Soluble Ca elution (small granules)0.533 ± 0.016 (3 days), 0.322 ± 0.012 (7 days), 0.366 ± 0.024 (14 days)
    Soluble P elution (small granules)3.723 ± 0.004 (3 days), 6.277 ± 0.079 (7 days), 5.865 ± 0.019 (14 days)
    Soluble Ca elution (large granules)0.513 ± 0.018 (3 days), 0.315 ± 0.009 (7 days), 0.351 ± 0.021 (14 days)
    Soluble P elution (large granules)3.640 ± 0.085 (3 days), 6.241 ± 0.041 (7 days), 5.847 ± 0.019 (14 days)
    Pore size (small granules)149.1660 Å
    Pore size (large granules)47.3236 Å
    Surface area (small granules)114.4452 m²/g
    Surface area (large granules)109.0581 m²/g
    Solubility (small granules)74.274 ± 13.599 µg/mm³
    Solubility (large granules)70.247 ± 12.612 µg/mm³
    Average pH (small granules)7.86 ± 0.05
    Average pH (large granules)7.83 ± 0.05
    Compressive strength (small granules)1.00 MPa
    Compressive strength (large granules)1.17 MPa
    Physiochemical properties (comparison to Bio-Oss®)Comparable (pore structure, microstructure, phase structure, chemical composition, residual organic substance)
    Biocompatibility
    Skin SensitizationNon-sensitizing (Stimulation Index = 1.03±0.26)
    MutagenicityNon-mutagenic (Micronucleus and AMES test)
    Intracutaneous ReactivityNo skin reactions
    PyrogenicityNon-pyrogenic
    HemolysisNo hemolytic reaction
    Systemic Toxicity / CytotoxicityNon-toxic and non-cytotoxic (Acute Systemic Injection Test and Cytotoxicity Test)
    Systemic Toxicity (leachables)Negative
    SterilityPassed (SAL 1 X10⁻⁶)
    Bacterial Endotoxin (LAL)Passed (<0.005 EU/mL or <0.232 EU/device)
    Extractables from syringeAcceptable levels (e.g., <0.097 mg/device ethyl ether, no cell lysis, no intraplasmic granules in MEM test)
    In-Vivo Performance (Animal Model)
    Biocompatibility and healingIndistinguishable from predicate (Bio-Oss®) and in some cases superior to untreated defects. Low inflammation. No necrosis.
    Graft resorption and boney integration (radiographic)Similar mean scores to predicate
    Percent mineralized volume and density (micro-CT)No statistical differences between test article and predicate. Statistically greater amounts of bone in defect sites compared to empty controls.
    Percent residual implant and bone marrow values (histomorphometry)No statistically significant differences between test article and predicate.
    Microscopic examination (implant slides)Equivalent performance to predicate; similar average residual implant material; InterOss® had a higher mean percent bone value compared to Bio-Oss® (not statistically significant).

    2. Sample size(s) used for the test set and data provenance:

    • Animal study (in-vivo performance): 4 treatment groups of beagle dogs. Three groups of nine dogs, and one group of six dogs. This refers to the study conducted on the device itself, not an AI test set.
    • Data Provenance: The animal study was conducted by SigmaGraft, Inc. (the manufacturer) or on their behalf, in the USA (presumably, as the FDA submission is from a US company). It is a prospective animal study as described (surgical procedures, observations, termination points).

    3. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Animal study (histology/pathology): Histological analysis was performed by Alizée Pathology, LLC. The specific number and qualifications of experts (e.g., veterinary pathologists with X years of experience) are not detailed in this summary.

    4. Adjudication method for the test set:

    • The document does not describe an adjudication method for human reader interpretations. The evaluation methods for the animal study (micro-CT, semi-quantitative histology scoring, histomorphometry) are objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes a medical device (bone grafting material), not an AI/ML software device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed or is relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This document is not about an AI/ML algorithm.

    7. The type of ground truth used:

    • Animal study (in-vivo performance): A combination of imaging (micro-CT) for mineralized volume/density and histopathology (undecalcified histology with semi-quantitative scoring and histomorphometry) for biocompatibility, tissue response, residual implant, and bone formation. This serves as the "ground truth" for the device's biological performance.

    8. The sample size for the training set:

    • N/A. This is not an AI/ML device where a "training set" would be applicable. The closest analogy would be prior knowledge and predicate device data used in the design and development of InterOss®.

    9. How the ground truth for the training set was established:

    • N/A. As above, this is not an AI/ML device. The "ground truth" for the predicate devices and general scientific understanding of bone grafting materials is based on decades of medical research, clinical experience, and regulatory standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142040

    Validate with FDA (Live)

    Device Name
    OCS-B COLLAGEN
    Manufacturer
    Date Cleared
    2015-07-31

    (368 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113246

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OCS-B Collagen® is recommended for:

    • Augmentation or reconstructive treatment of the alveolar ridge.
    • Filling of infrabony periodontal defects.
    • Filling of defects after root resection, apicoectomy, and cystectomy
    • Filling of extraction sockets to enhance preservation of the alveolar ridge
    • Elevation of maxillary sinus floor
    • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
    Device Description

    OCS-B Collagen® is a combination of purified cancellous bone mineral granules (OCS-B") and 10% collagen in a block form in a blister and cylinderic form in a syringe and blister. It is sterilized by yirradiation.

    AI/ML Overview

    The provided text describes the OCS-B Collagen® device, a bone grafting material, and its substantial equivalence to predicate devices, primarily Bio-Oss Collagen®. The information requested falls under the category of performance testing for medical devices, which typically includes detailed studies to prove efficacy and safety. However, this document is a 510(k) summary, which focuses on demonstrating "substantial equivalence" to existing legally marketed devices rather than performing extensive novel clinical trials as would be required for a novel device or a PMA.

    Therefore, the study described here is primarily a comparative study against predicate devices, supported by biocompatibility testing and animal studies, rather than a standalone clinical efficacy study with specific acceptance criteria in terms of performance metrics like sensitivity, specificity, or improvement with AI assistance.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in terms of quantitative performance metrics (e.g., specific percentages for success rates, bone growth, etc.). Instead, the acceptance criteria are implicitly that the OCS-B Collagen® device is "substantially equivalent" to its predicate devices (Bio-Oss® Collagen and OCS-B®) in terms of:

    Acceptance Criterion (Implicit)Reported Device Performance
    Same Indications for UseOCS-B Collagen® has the same indications for use as Bio-Oss® Collagen, including augmentation/reconstructive treatment of alveolar ridge, filling of infrabony periodontal defects, defects after root resection/apicoectomy/cystectomy, extraction sockets, elevation of maxillary sinus floor, and filling of periodontal/peri-implant defects in conjunction with GTR/GBR.
    Same Intended UseOCS-B Collagen® has the same intended use as Bio-Oss® Collagen and OCS-B® (adjective therapy in restoring bony defects).
    Similar Physical and Chemical StructureBoth OCS-B Collagen® and Bio-Oss® Collagen are porous, biocompatible bone grafts facilitating new bone formation. They have the same component ratio of bone mineral granules and collagen, same source of bone (bovine bone), and collagen (porcine skin). Physical and chemical characteristics were found to be comparable.
    Comparable BiocompatibilityOCS-B Collagen® was subjected to a full range of biocompatibility tests (ISO 10993, cytotoxicity, hemolysis, acute systemic injection, intracutaneous reactivity, skin sensitization, genotoxicity, oral mucosa irritation, implantation) and confirmed equivalence to predicates.
    Similar Performance in Animal Studies (Bone Growth and Resorption)In several animal studies, both OCS-B Collagen® and predicate devices were found to grow new bone and be subsequently resorbed at similar rates.
    Successful Clinical Outcomes (Case Series)A clinical case series using OCS-B Collagen® resulted in defect healing and formation of new bone of sufficient quality to allow dental implant placement.
    Same Sterilization ProcessSterile by Gamma Irradiation (SAL 1 x 10⁻⁶).
    Same Shelf-Life36 Months.
    Compliance with Relevant Standards (ASTM F1581-99, FDA "Medical Device Materials Derived from Animal Sources", ISO 11137, ASTM F1980-99)OCS-B Collagen® was tested in accordance with these standards. (Compliance implies meeting the requirements set forth in these standards, but specific results against numerical criteria are not detailed in this summary).
    Controlled RisksOCS-B Collagen® presents the same types of potential risks as the predicate and controls these risks in a similar manner (e.g., manufacturing processes for removal of organic impurities, protein content limits, TSE/virus inactivation studies).

    2. Sample Size Used for the Test Set and Data Provenance

    • Animal Studies: "several animal studies" are mentioned, but no specific sample sizes or species are provided in this summary document.
    • Clinical Case Series: "a clinical case series" is mentioned, but no specific number of patients (sample size) is provided.
    • Data Provenance: Not explicitly stated for animal studies or the clinical case series (e.g., country of origin, retrospective/prospective). However, the sponsor is NIBEC Co., Ltd. from Korea. Given the K-number, the submission is to the U.S. FDA, so the data would have been submitted to the U.S. regulatory body. The document does not specify if these studies are retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This type of information is generally not applicable to the studies described for this device. The "ground truth" here is the biological and clinical outcome (bone formation, resorption, defect healing, suitability for implant placement), which would be assessed by standard histological, imaging, and clinical follow-up methods by trained professionals (e.g., veterinarians for animal studies, dentists/oral surgeons for clinical case series). The document does not specify an "expert panel" for establishing ground truth in the context of, for example, image interpretation.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this 510(k) summary. Adjudication methods (like 2+1, 3+1) are typically used for studies where multiple human readers interpret data (e.g., images) and consensus is needed, often in AI comparative effectiveness studies. The studies mentioned here focus on biological outcomes rather than interpretive tasks.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not conducted as described in this summary. The comparison is between the device and predicate devices based on physical/chemical properties, biocompatibility, and general performance in animal studies and a clinical case series, not a comparative study of human reader performance with and without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an AI algorithm's performance, which is not relevant for this bone graft material. The device itself is a physical implant, not a diagnostic algorithm.

    7. The Type of Ground Truth Used

    The ground truth used in the animal studies and clinical case series would be based on:

    • Histology: Direct examination of tissue to confirm new bone formation and resorption.
    • Imaging: X-rays, CT scans, or other imaging modalities to assess defect healing and bone density/volume.
    • Clinical Outcomes: Assessment of defect healing, absence of adverse events, and functional outcomes, such as successful dental implant placement.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical bone graft, not a machine learning algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a machine learning algorithm.