510(k) Data Aggregation

    K Number
    K211324

    Validate with FDA (Live)

    Date Cleared
    2022-03-14

    (318 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    2.4mm Wise-Lock Distal Radius System:
    2.4mm Wise-Lock Distal Radius System are used for displaced extra-articular distal radius fractures and corrective osteotomies of the distal radius.

    2.4mm Wise-Lock Proximal Radius System:
    2.4mm Wise-Lock Proximal Radius System are used for extra-articular fractures of Proximal radius and multifragmented radial neck fractures.

    2.4mm Variable Angle Distal Radius System:
    2.4mm Variable Angle Distal Radius System are used for displaced extra-articular and intra-articular distal radius fractures and corrective osteotomies of the distal radius.

    Device Description

    The 2.4mm Radius System consists of various types of bone plates, screws for implantation in the radius bone to treat the radius bone fracture.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "2.4mm Radius System." This is a submission to the FDA demonstrating that a new device is substantially equivalent to a legally marketed predicate device, and thus does not require approval of a premarket approval application (PMA).

    Crucially, this document focuses on demonstrating substantial equivalence based on design, materials, and mechanical performance tests against recognized standards, rather than clinical performance or diagnostic accuracy. Therefore, the concepts of acceptance criteria for diagnostic performance, test set characteristics, expert ground truth, MRMC studies, or standalone algorithm performance are not applicable to this type of regulatory submission.

    The "acceptance criteria" in this context refer to compliance with material standards, mechanical performance standards, and sterilization/packaging standards. The "study that proves the device meets the acceptance criteria" refers to the non-clinical bench testing conducted against these standards.

    Here's a breakdown of the requested information based on the provided text, noting where certain categories are not applicable:


    Acceptance Criteria and Study Details for the 2.4mm Radius System (K211324)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance
    Material Standards- ASTM F136/ISO 5832-3: Standard specification for wrought Titanium-6 Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications.- ASTM F899-12: Standard Specification for Wrought Stainless Steels for surgical instruments.- ASTM F138/ISO 5832-1: Standard Specification for Wrought-18 Chromium-14 Nickel-2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants.- Compliance verified with Titanium Alloy (Ti-6Al-4V) Grade 5 as per ISO 5832-1/ASTM F136 for Titanium Implants.- Compliance verified with Stainless Steel (Grade 316L) as per ISO 5832-3/ASTM F138 for Stainless Steel Implants.- Compliance verified with Grade 304 of Stainless steel (SS 304) Material for instruments as per ASTM F899-12b.
    Mechanical Performance- ASTM F382: Standard Specification and Test Method for Metallic Bone Plates.- ASTM F543: Standard Specification and Test Method for Metallic Medical Bone Screws.Plate Testing: - 4-Point Static Test: Conforms - 4-Point Fatigue Test: ConformsScrew Testing: - Driving Torque Test: Conforms - Torsion Test: Conforms - Axial Pull-out Test: Conforms
    Biocompatibility- ISO 10993-3 (General evaluation for raw materials)Raw materials evaluated and found to comply with ISO 10993-3.
    Sterilization & Packaging- ISO 11137-1, -2, -3 (Radiation sterilization)- ISO 17665-1, -2, -3 (Moist heat sterilization)- ISO 11140-1 (Chemical indicators)- ISO 11737-1, -2 (Microbiological methods)- ISO 11607-1, -2 (Packaging for sterile medical devices)- ASTM F1980 (Accelerated Aging of Sterile Barrier Systems)- ASTM F88/F88M (Seal strength)- ASTM F1929 (Seal leaks by dye penetration)Subject device and packaging validated against these standards for both Gamma Sterilization and non-sterile products.
    Bacterial Endotoxin- USP <85> Bacterial Endotoxin Test.- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests.Compliance with USP <85> and <161> for bacterial endotoxin and pyrogen tests.
    Dimensional VerificationComparison to predicate device dimensions.Same dimension as of Predicate device.
    Shelf-life5 Years5 Years (Same as predicate device)
    Use TypeSingle UseSingle Use (Same as predicate device)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document describes bench testing for mechanical integrity and material compliance, not clinical or diagnostic performance on human or animal subjects. The "sample size" would refer to the number of test coupons or devices subjected to mechanical tests (e.g., how many plates for 4-point static test, how many screws for torsion test). This specific number is not explicitly stated in the provided text, but it is implied that a sufficient number was tested to meet the requirements of the ASTM standards.
    • Data Provenance (e.g., country of origin of the data, retrospective or prospective): The data provenance here refers to the origin of the bench test results. These tests were conducted as part of the manufacturer's (Auxein Medical Private Limited, India) submission to the FDA. The tests likely involved physical samples of the manufactured devices. The data itself is prospective in the sense that the tests were performed specifically for this submission against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This device is a bone fixation system (medical hardware), not a diagnostic algorithm that requires expert interpretation to establish a ground truth. The "ground truth" for this device relates to its material properties and mechanical performance, which are established by standardized laboratory testing methods and engineering specifications, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. As this is not a diagnostic performance study, there's no need for an adjudication method for establishing ground truth. The compliance with standards is determined by objective measurements during bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a bone fixation device, not an AI-assisted diagnostic tool. Therefore, MRMC studies and AI assistance comparisons are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device (bone plate and screws), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device is defined by established engineering and material standards (e.g., ASTM F382, ASTM F543, various ISO and ASTM material standards). Compliance with these standards confirms the device's functional and material integrity.

    8. The sample size for the training set

    • Not Applicable. This device is not an AI/machine learning algorithm, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARMAR™ / ARTIS™ Small Fragment and Calcaneal Plates & MBOSS™/FIXION™ Screws are indicated for temporary internal fixation and stabilization of osteotomies and fractures, such as:

    • comminuted fractures
    • supracondylar fractures
    • extra-articular fractures
    • fractures in osteopenic bone
    • nonunions
    • malunions
      Smaller-sized ARMAR™ / ARTISTM plates are used for small bones and small fragments of the hands and feet.
      ARMARTM / ARTIS™ Calcaneal plates are indicated for temporary internal fixation and stabilization of osteotomies and fractures of the calcaneous.

    ARMAR™ / ARTIS™ Humerus and Olecranon plates & MBOSSTM/FIXIONT™ Screws are intended for fractures and fracture dislocations, osteotomnies, and non-unions of the distal and proximal humerus, olecranon, and ulna in adult, particularly in osteopenic bone.
    Specifically, distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, osteotomies, malunions and non-unions of the distalhumerus. Olecranon and Proximal ulna plates are indicated for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna.

    ARMARTM / ARTISTM Tibia and Buttress Plates & MBOSS™/FIXION™ Screws are intended for fixation of the ankle in adults. Specifically,

    • Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia.
    • Medial Distal and Proximal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
    • L Buttress and T Buttress Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
    • Proximal Lateral Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia, a Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia.

    ARMAR™ / ARTISTM 2.4mm Plates & MBOSS™/FIXION™ Screws are intended for Fixation of complex intra- and extra-articular fractures and osteotomnies of the distal radius and other small bones in adults.

    ARMARTM / ARTISTM Clavical Hook Plate & MBOSS™/FIXION™ Screws is intended for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.

    ARMARTM / ARTISTM Metaphyseal Plate & MBOSS™/FIXION™ Screws indicated for fractures, osteotomies, and non-unions of the radius and other small bones. Additionally, the Metaphyseal Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal humerus, clavicle, and scapula.

    ARMARTM / ARTISTM Calcaneal Plate & MBOSS™/FIXION™ Screws is intended for fixation of complex extra-articular and intra-articular fractures and osteotomies of the calcaneus.

    Device Description

    The Meril Healthcare Trauma System utilizes screws and plates of various lengths and designs to accommodate a variety of patient anatomies. The Trauma System utilizes medical grade stainless steel and titanium. Trauma System implant are provided clean and non-sterile with instructions for steam sterilization. The instruments are provided non-sterile and are reusable. The implants are provided nonsterile and are single use only.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding a medical device, specifically bone plates and screws. It does not contain information about an AI/ML device, AI acceptance criteria, or a study proving an AI device meets acceptance criteria.

    Therefore, I cannot extract the requested information regarding AI device performance, acceptance criteria, sample sizes, expert ground truth, MRMC studies, or training/test set details from this document. The document describes the substantial equivalence of bone fixation appliances to existing predicate devices based on technological characteristics and mechanical performance data (e.g., static four-point bending, dynamic four-point bending, static torsion, static axial pullout).