Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K151638

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-13

    (149 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K021591,K993266,K091559,K862117

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phenom™ Catheters are intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the neuro, peripheral, and coronary vasculatures.

    Device Description

    The Phenom™ Catheters are variable stiffness, single lumen catheters designed to access small, tortuous vasculature. They are available in a variety of lengths, stiffness and inner and outer diameters. The outer surface of the catheter is coated to enhance navigation in the vessel. The catheter also incorporates a liner to facilitate movement of introduction devices passing through its lumen. The distal tip has radiopaque marker(s) to aid visualization and positioning under fluoroscopy.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Phenom™ Catheters. As such, it's a submission to the FDA to demonstrate substantial equivalence to legally marketed predicate devices, not a study proving device meeting specific clinical acceptance criteria in the way a clinical trial would.

    Therefore, many of the requested points related to a specific clinical study (like sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable or directly derivable from this type of regulatory document.

    However, I can extract information regarding the bench testing performed to establish substantial equivalence, which serves as the "study" in this context to prove the device meets certain performance characteristics relative to predicate devices.

    Here's the breakdown based on the provided text, addressing the applicable points and explaining why some are not applicable:


    Description of Acceptance Criteria and Substantiating Study

    The Phenom™ Catheters are submitted as substantially equivalent to predicate devices, meaning their performance characteristics must be comparable to those existing devices and demonstrate no new safety or effectiveness concerns. The "acceptance criteria" are implicitly defined by the safety and performance standards established by the predicate devices and general regulatory requirements for this class of device. The "study" proving the device meets these criteria is a series of bench tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes various bench tests conducted to support the substantial equivalence. The "acceptance criteria" for these tests would typically be defined internally by the manufacturer to ensure the new device performs comparably to or within acceptable limits derived from the predicate devices. Since specific numerical acceptance criteria (e.g., "burst pressure > X bar") are not explicitly stated in this summary, the table will list the performance tests performed, implying that the acceptance was meeting the internal specifications aligned with the predicate performance.

    Acceptance Criteria Category (Performance Test)Reported Device Performance (Implied)
    Dimensional InspectionMet specifications (OD, ID, Length, Distal Tip Configuration) comparable to predicate devices.
    Material VerificationMaterials verified as similar to predicate devices, with extensive clinical history of safe use.
    Accessibility/TrackabilityDemonstrated performance suitable for accessing small, tortuous vasculature, comparable to predicate devices. "Evaluation was performed in the more complex and higher risk neurovascular anatomy, which is the worst case representation of the cardiac and peripheral vascular anatomies."
    Device CompatibilityCompatible with Guide Catheter, Guide Wire, RHV, comparable to predicate devices.
    Shaft StiffnessMet specifications, with slight variations differentiating it for specific applications, comparable to predicate devices.
    Chemical CompatibilityCompatible with Saline, Contrast Medium.
    Tip ShapeabilityMet specifications to aid navigation.
    Kink ResistanceDemonstrated resistance to kinking in tortuous paths.
    Conical Fitting for HubMet specifications for secure connection.
    Dead Space VolumeMet specifications.
    Corrosion ResistanceMet specifications.
    Tensile Strength (Body & Hub Attachment)Met specifications, demonstrating structural integrity.
    Liquid/Air Leakage at HubNo leakage during aspiration.
    Burst PressureMet specifications, demonstrating structural integrity under pressure.
    Particulate TestingMet specifications to ensure minimal particulate release.
    Outer Surface Coating/Lubricity/DurabilityEnhanced navigation as designed, comparable durability to predicate devices.
    Catheter Flow RateMet specifications for infusion of agents.
    Flexural FatigueDemonstrated durability under repeated bending.
    Torque StrengthMet specifications for rotational control.
    Tip Mark Radiopacity TestingRadiopaque markers aid visualization and positioning under fluoroscopy.
    Biocompatibility TestingPassed tests to ensure biological safety.
    Shelf-life TestingValidated for sterility and performance over time.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This refers to the number of devices or components tested for each bench test. This information is not provided in this summary. Bench testing typically involves a statistically significant number of samples for each test, but the exact numbers are not detailed here.
    • Data Provenance: The bench tests are laboratory-based, performed by the manufacturer (Cathera, Inc., Mountain View, CA). This is not human data, so "country of origin" or "retrospective/prospective" does not directly apply in the usual clinical sense. The worst case representation was chosen considering neurovascular anatomy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Bench testing for substantial equivalence does not involve establishing ground truth from human expert interpretation in the way clinical studies with AI algorithms do. Performance is measured against engineering specifications and comparisons to predicate device data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a catheter, not an AI diagnostic algorithm. Therefore, MRMC studies involving human readers and AI assistance are not relevant to its regulatory approval in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a catheter, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the bench tests are engineering specifications, material standards, and performance characteristics derived from predicate devices. For example, "burst pressure" has a defined threshold, and "biocompatibility" is assessed against ISO standards. The evaluation included performance in "more complex and higher risk neurovascular anatomy," implying simulated environments or models representing this.

    8. The sample size for the training set

    • Not applicable. As this is a physical medical device (catheter) undergoing bench testing for substantial equivalence, there is no "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143398

    Validate with FDA (Live)

    Date Cleared
    2015-06-26

    (212 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K111490, K091559, K121734

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plato MICROCATH 27B Microcatheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents to the peripheral system. The cather is not intended for use in cither the coronary or neuro vasculature.

    Device Description

    The Plato MICROCATH® 27B Microcatheter is a single lumen microcatheter constructed with a flexible polymer shaft of varying stiffness to aide in accessing vasculature. The catheter is designed to be used with a guide catheter and a steerable guidewire for accessing the vasculature. The proximal end of the catheter has a diameter of 3.4F, tapering to a distal OD of 3.2F, which can be inserted into a 5F guide catheter. The inner diameter is constant throughout the shaft length and accommodates up to a 0.025" guidewire. The catheter is 150 cm in length with a straight tip configuration which can be steam-shaped to the doctor's preferred shape. A steam shaping mandrel is included in the packaging. The microcatheter has hydrophilic coating on the outer distal shaft to reduce friction during manipulation in vessels and has one radiopaque tip marker to facilitate fluoroscopic visualization.

    AI/ML Overview

    The provided document describes the FDA 510(k) summary for the Plato MICROCATH® 27B Microcatheter, focusing on demonstrating substantial equivalence to predicate devices, not on proving effectiveness for a new diagnostic or interventional AI device. Therefore, the traditional acceptance criteria, study design parameters, and concepts like 'human readers improve with AI' are not applicable to this submission.

    Instead, the document details performance data collected from non-clinical testing to show that the new microcatheter is as safe and effective as existing, legally marketed predicate devices. The acceptance criteria essentially derive from established standards for medical devices and comparisons to the predicate devices.

    Here's an adaptation of your requested table and information, tailored to the context of this medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test (Acceptance Criteria Derived From)Acceptance Criteria (Implicit/Explicit from standards; "Comparable to predicate" suggests meeting predicate's performance)Reported Device Performance
    Biocompatibility:
    Cytotoxicity (ISO 10993-5)Non-cytotoxic (Scores = 0)Non-cytotoxic. Test article scores were 0 at 48 hours.
    Sensitization (ISO 10993-10)Negative for dermal sensitization (Scores = 0)Negative for dermal sensitization. Test articles sensitization scores were all 0.
    Irritation/Intracutaneous Reactivity (ISO 10993-10)Non-irritating (No significantly greater biological reaction than control)Non-irritating. Extracts of the test article did not show a significantly greater biological reaction than sites injected with control article.
    Acute Systemic Toxicity (ISO 10993-11)Non-toxic (No toxicity or animal weight loss)Non-toxic. Test articles showed no toxicity or animal weight loss for both cottonseed oil and saline extracts for 72 hour test period.
    Materials-Mediated Pyrogenicity (ISO 10993-11)Non-pyrogenic (No increases in individual temperatures)Non-pyrogenic. No increases in individual temperatures.
    Hemocompatibility – Hemolysis (ASTM 758-08)Non-hemolytic (Corrected hemolysis index below threshold)Non-hemolytic. Corrected hemolysis index was 0.15% by direct method, 0.23% by extract.
    Partial Thromboplastin Time (ASTM F2382)Similar activation profile to predicate (Minimal activators, minimal difference in clotting times)Both test article and predicate were minimal activators. Difference in clotting times between test article and predicate was 3 seconds.
    Complement Activation (ISO 10993-4)Less than or comparable to predicateComplement activation by the test article was less than that of the predicate device.
    Thrombogenicity in Dogs (ISO 10993-4)Comparable to predicate (Thrombosis grade, weight changes)Thrombosis grade comparable to predicate. Weight changes of the implants comparable between test and control articles.
    Functional Testing:
    Visual/Dimensional InspectionNo surface defects, visible droplets, or defects causing trauma; Meets dimensional specificationsNo surface defects or visible droplets of coating on catheters. All catheters met dimensional specifications.
    Air Ingress/Negative Collapse (ISO 10555)No air ingress, no lumen collapseHub fittings do not allow air ingress and no evidence of lumen collapse.
    Kink ResistanceResistant to kinking around small diameter turns per specificationDevice was resistant to kinking around small diameter turns per specification.
    Tensile Strength/Elongation (ISO 10555-1)Meets minimum force breakage based on tube diametersAll catheters met minimum force breakage based on tube diameters specified in ISO 10555-1.
    Liquid Leakage under Pressure/Leakage at Hub (ISO 10555-1)No leakage from hub or catheter body at specified pressuresNo leakage from hub or catheter body.
    Tip Stiffness (ASTM D747-10)Comparable to predicate devicesTip stiffness was comparable that of the predicate devices.
    Pressure vs. Flow CharacterizationFlow rates reported in Instructions for Use at 100 and 300 psiFlow rates reported in Instructions for Use at 1.00 and 300 psi. (Note: "1.00" might be a typo and likely meant 100 psi as per test description.)
    Static Burst Pressure (ISO 10555-1)Maximum peak pressures exceeded 300 psiMaximum peak pressures all exceeded 300 psi.
    Dynamic FlowNo leaks, breaks or occluded lumens at 750 psiNo leaks, breaks or occluded lumens at 750 psi.
    Flexibility Fatigue and ProfileNo signs of cracks or breakage post worst-case simulated useAll catheters showed no signs of cracks or breakage post worst case simulated use.
    Shape RetentionMaintain specified tip angle post shaping/conditioning/use; Tensile strength meets original specs after shapingAll catheters maintained specified tip angle after steam shaping, water bath conditioning and simulated use of guidewires. Tensile testing after tip shaping passed minimum tensile strength requirements.
    Torque to FailureNo signs of breakage, twists or collapsed lumens after specified number of torque turnsAll catheters showed no signs of breakage, twists or collapsed lumens after specified number of torque turns.
    Coating Lubricity and DurabilityMeets specified frictional forcesAll catheters met specified frictional forces.
    Coating IntegrityAcceptable coating coverage post simulated useAll samples showed acceptable coating coverage post simulated use.
    ParticulatesComparable numbers of particles to predicate in each size rangeThe test and predicate catheters had comparable numbers of particles in each size range.
    Chemical CompatibilityNo signs of degradation, corrosion, or physical decomposition after exposure to agentsAll catheters showed no signs of degradation, corrosion or physical decomposition.
    Latex Content (ASTM D6499-07)No detectable traces of latex foundNo detectable traces of latex found.
    Corrosion Resistance (ISO 10555-1)No signs of corrosion on metallic componentsNo signs of corrosion on metallic components of catheters.
    RadiopacityAcceptable radiopacity (evaluated by physicians)Both test and predicate catheters had acceptable radiopacity.
    MRI CompatibilityLabeled MRI Unsafe (as applicable for conducting/magnetic materials)Catheters are labeled MRI Unsafe on IFU.
    Simulated UseComparable Likert scores to predicate in Guidewire tracking and guide catheter movement; Interventional devices successfully deployedCatheters and predicate devices had comparable Likert scores in terms of Guidewire tracking and guide catheter movement. Interventional devices successfully deployed.
    Cadaver TestingComparable performance and access time to predicate in designated target; Ability to deploy/retrieve interventional devicesTest and predicate devices both exhibited comparable performance with similar access time to the designated target.
    Packaging Testing (ASTM F88-00, F2096-04)Minimum seal strength met, no bubbles under leak test conditionsAll sterile barrier pouches met minimum seal strength and showed no bubbles under leak test conditions.
    Shelf Life TestingContinued to meet specifications post accelerated aging and room temperature storageShelf life testing in progress. Expiration date will be advanced as aging data are available to demonstrate package and product continues to meet specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    This document primarily reports on non-clinical bench and lab testing, and some limited biological/simulated use testing. The "test set" here refers to the number of devices or samples tested for each specific non-clinical performance attribute.

    • Sample Sizes: Not explicitly stated for each test, but implied by the nature of bench testing where multiple samples are typically used to ensure reliability. For example, "all catheters" met specifications implies multiple catheters were tested. For animal studies (thrombogenicity) and cadaver studies, the sample size would be small (e.g., number of dogs, cadavers). For Radiopacity and Simulated Use, "physicians," plural, suggests more than one device per observer.
    • Data Provenance: The data is primarily from bench-top laboratory testing and in vitro studies, in accordance with ISO and ASTM standards. Some tests involve animal models (e.g., thrombogenicity in dogs) and human cadavers. The context implies these studies were conducted by the manufacturer or their contracted labs. No specific country of origin for the data is mentioned, but the submission is to the U.S. FDA, and the company is US-based. This is all prospective testing, specifically designed to demonstrate compliance for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Experts: The document mentions "physicians" evaluated radiopacity and participated in cadaver testing. For simulated use, "physicians" also provided Likert scores.
    • Number: Not explicitly stated how many physicians were involved in these evaluations. The plural "physicians" indicates more than one.
    • Qualifications: Not explicitly stated in terms of years of experience or specific subspecialty (e.g., interventional radiologist), but given the context of medical device evaluation, it is implied that these are qualified medical professionals with expertise relevant to microcatheters and their use.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly detailed. For tests involving "physicians" (Radiopacity, Simulated Use, Cadaver Testing), it is stated that comparisons were made or scores obtained. The document does not specify if there was a formal adjudication process (e.g., 2+1, 3+1), rather it presents the aggregate findings or an overall determination of "comparable" or "acceptable."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study (often used for diagnostic imaging AI) is not applicable to a premarket notification for a physical medical device like a microcatheter. The comparison is primarily against established performance standards and a predicate device's existing characteristics. The "simulated use" and "cadaver testing" involve human readers/users (physicians) providing feedback, but this is to assess usability and performance in a simulated environment, not an MRMC study comparing AI with and without human assistance.
    • Effect Size of Human readers with/without AI: Not applicable, as this is not an AI device.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This submission is for a physical medical device (microcatheter), not an algorithm or AI product. The performance evaluations are intrinsic to the device itself and its interaction with biological and simulated environments.

    7. The Type of Ground Truth Used

    The concept of "ground truth" as it applies to diagnostic or AI studies is not directly applicable here. Instead, the basis for comparison and acceptance is:

    • Established Industry Standards: ISO 10993 (biocompatibility), ISO 10555 (intravascular catheters), ASTM standards.
    • FDA Guidance: FDA guidance for Short-Term and Long-Term Intravascular Catheters, FDA Special Controls Document for PTCA Catheters.
    • Predicate Device Performance: Direct comparison to the ev3 Marksman™ Catheter (K111490, K091559) and the Scientia Vascular, LLC Plato MICROCATH 27 (K121734). The goal is to demonstrate that the new device is "substantially equivalent" in safety and effectiveness.
    • Expert Observational Consensus: For aspects like radiopacity and simulated use feedback from physicians.

    8. The Sample Size for the Training Set

    • Not applicable. This is not a machine learning or AI device, so there is no "training set." The device design and materials are based on engineering principles and prior device knowledge.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI model, this question does not apply. The "ground truth" for the device's design and expected performance comes from established engineering principles, regulatory standards, and the performance characteristics of predicate devices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1