Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151667

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-13

    (147 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K090335, K110483, K133177

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AXS Catalyst™ Distal Access Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The AXS Catalyst Distal Access Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The AXS Catalyst™ Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into the peripheral and neuro vascular system. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostastic Valve (RHV) and Tuohy Borst valve with sideport for flushing, insertion of catheters and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.

    AI/ML Overview

    The medical device in question is the AXS Catalyst™ Distal Access Catheter.

    Here's an analysis of its acceptance criteria and the studies that prove its compliance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly present acceptance criteria in a quantitative table format for all tests. Instead, it states conclusions for each test, indicating whether the device "meets acceptance criteria" or for biocompatibility, "PASS" and a conclusion.

    Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
    Bench PerformanceDimensional VerificationDimensions conform to specifications.Dimensional verification meets acceptance criteria.
    Tip ConfigurationDistal tip is smooth, rounded, tapered, or similarly finished to minimize trauma to vessels during use (per EN ISO 10555-1).Tip configuration meets acceptance criteria.
    Surface IntegrityExternal surface is free from extraneous matter, process and surface defects, and drops of lubricant fluids.Surface integrity meets acceptance criteria.
    Tip BucklingMaximum force required to cause buckling is within specified limits.Tip buckling meets acceptance criteria.
    Catheter lubricity and durabilityFrictional force of the device sample when pulled between two clamped pads is within specified limits after 5 cycles. (Implied: acceptable lubricity and durability as per performance expectations).Coating lubricity and durability meets acceptance criteria.
    TrackabilityPeak tracking force through a neurovascular model over a microcatheter is within specified limits.Track advance force meets acceptance criteria.
    Tensile StrengthTensile force required to induce failure of fused joints, shaft junctions, and marker band is above specified minimums (based on EN ISO 10555-1).Tensile strength meets acceptance criteria.
    Liquid Leak ResistanceMeets freedom from leakage-liquid leak requirement 4.7.1 of EN ISO 10555-1 and catheter hub meets liquid leakage requirement 4.2.1 of EN 1707 (no visible leaks at 300kPa for 30s).Liquid leak resistance of catheter meets acceptance criteria.
    Air Leak ResistanceMeets freedom from leakage-air aspiration requirement 4.7.2 of EN ISO 10555-1 and catheter hub meets air leakage requirement 4.2.2 of EN 1707 (no air bubbles upon withdrawal of syringe plunger to 10cc mark for 15s).Air leak resistance of catheter meets acceptance criteria.
    Catheter Torsional Bond StrengthTorque strength (number of rotations before failure) is above specified minimums.Catheter torsional bond strength meets acceptance criteria.
    Flexural FatigueCatheter shaft shows no functional impact or degradation after specified number of cycles through a tortuous model.Flexural fatigue meets acceptance criteria.
    Catheter Kink RadiusKink radius at distal and specific mid-shaft joint sections is within acceptable limits.Catheter kink radius meets acceptance criteria.
    Catheter Tip and Lumen Patency (Adjunctive Aspiration)No tip or lumen collapse and tip integrity maintained during aspiration with a 60cc syringe in a simulated use model.Catheter tip and lumen patency during adjunctive aspiration meets acceptance criteria.
    Catheter Tip and Lumen Patency (with retrieval device)Ablility to deliver and withdraw retrieval device 3 times in a tortuous path without functional impact and integrity to the tip.Catheter tip and lumen patency meets acceptance criteria.
    Chemical CompatibilityVisual and dimensional integrity of catheter maintained after exposure to saline, non-ionic, and ionic contrast liquids; inner lumen integrity maintained.Chemical compatibility meets acceptance criteria.
    Hub GaugingCatheter hub meets gauging requirement 4.1 of EN 1707.Hub gauging meets acceptance criteria.
    Simulated UsePerformance design attributes are deemed acceptable through user evaluation in an in vitro flow model that replicates neurovasculature tortuosity, diameter, and location in simulated arterial circulation.Simulated use meets acceptance criteria.
    BiocompatibilityMEM Elution Cytotoxicity (ISO 10993-5)No biological activity (e.g., Grade 0 cytotoxicity) in specified mammalian cells at 48 hours post exposure; suitability of test system confirmed by controls.PASS: No cytotoxicity or cell lysis.
    Guinea Pig Maximization Sensitization (ISO 10993-10)No reaction at challenge (0% sensitization) following induction phase with specified extracts.PASS: No evidence of sensitization.
    Intracutaneous Reactivity (ISO 10993-10)Test article sites do not show a significantly greater biological reaction than control sites (e.g., difference of overall mean score = 0.0).PASS: Non-irritant.
    Acute Systemic Injection (ISO 10993-11)Specified extracts do not induce a significantly greater biological reaction than control extracts in test animals (e.g., no mortality or systemic toxicity).PASS: No mortality or evidence of systemic toxicity.
    Rabbit Pyrogen (ISO 10993-11)No rabbit injected with test article extract shows an individual rise in temperature of 0.5°C or more.PASS: Non-pyrogenic.
    Hemolysis Extract/Direct Contact Method (ISO 10993-4)Hemolysis percentage is within acceptable limits (e.g., low percentage above negative control).PASS: Non-hemolytic (0.17% above negative control via direct, 0.12% via indirect).
    In Vitro Hemocompatibility (ISO 10993-4)Results for WBC, RBC, platelets, hematocrit, and hemoglobin are comparable to the Negative Control (e.g., within specific percentage ranges 89%-105%).PASS. (Group 1: 89%-98%, Group 2: 97%-103%, Group 3: 100%-105%, Group 4: 98%-105%).
    Complement Activation (SC5b-9) (ISO 10993-4)Concentration of SC5b-9 in test articles is not statistically higher than negative control; not considered potential activator of complement system.PASS.
    Complement Activation (C3a) (ISO 10993-4)Concentration of C3a in test articles is not statistically higher than negative control; not considered potential activator of complement system.PASS.
    Partial Thromboplastin (PTT) (ISO 10993-4)Clotting time of test sample is comparable to predicate sample and meets requirements of the test (e.g., considered minimal activators with clotting time >= 90.0% of Negative Control).PASS. Results comparable to Negative Control. Test articles considered minimal activators with clotting time being 90.0% (catheter) and 92.1% (tubing) of the Negative Control and, therefore, met the requirements of the test.
    Shelf Life/SterilityShelf Life Testing (Product and Packaging)Meets established criteria for product and packaging integrity and functionality after 6 months.Results met established criteria.
    Sterilization (Ethylene Oxide Residuals)Ethylene Oxide residuals < 4 mg/device average daily dose. Ethylene Chlorohydrin residuals < 9 mg/device average daily dose (per EN ISO 10993-7 for limited contact delivery system - externally communicating).EO Results: 2.29 mg/device extracted residuals. Ethylene Chlorohydrin Results: 0.07 mg/device extracted residuals. These are less than the maximum allowed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The document does not specify the exact sample size for each bench test beyond "Prepare sample for test" or "test sample." However, bench tests are inherently prospective, performed on newly manufactured devices. No country of origin for the data is specified, but as the submitter is Stryker Neurovascular in Fremont, California, the testing likely occurred in the USA or through contracted labs.
    • Animal Study: The document states "An animal study was conducted... to evaluate performance design attributes... in an acute animal model." The sample size (number of animals) is not specified. The study was conducted "in compliance with applicable requirements in the GLP regulation (21 CFR Part 58)," typically indicative of prospective, controlled studies, likely in the USA.
    • Biocompatibility Testing: The document doesn't explicitly state the sample size for each biocompatibility test (e.g., number of replicates for MEM Elution, number of animals for Sensitization tests). However, these are standard biological tests performed on material samples from the device. These tests are prospective. No specific country of origin is listed for the data, but compliance with ISO 10993 implies internationally recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and therefore not provided in the document. The studies conducted are primarily bench (in vitro physical and chemical) and animal (in vivo non-human) performance tests, and biocompatibility tests. These types of studies do not typically involve human expert adjudication of "ground truth" in the same way clinical image interpretation or diagnostic performance studies would. The "ground truth" for these tests is based on objective physical measurements, chemical analyses, or biological reactions against established scientific and regulatory standards (e.g., ISO, EN standards, GLP regulations).

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 refer to consensus building among human experts, usually radiologists or clinicians, to establish ground truth in diagnostic studies. The studies presented here (bench, animal, biocompatibility) rely on objective measurements and established protocols rather than human expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is a physical medical device (catheter), not an AI-powered diagnostic or assistive tool for human readers. Therefore, this question is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No standalone (algorithm only) performance study was done. This question is relevant to AI/software as a medical device, which this device is not.

    7. The Type of Ground Truth Used

    • Bench Performance Tests: The ground truth is based on objective physical and mechanical measurements, as compared to pre-defined engineering specifications and relevant international standards (e.g., EN ISO 10555-1, EN 1707). For instance, for Dimensional Verification, the truth is the actual measured dimension compared to a design specification. For Tensile Strength, the truth is the force at which failure occurs compared to a minimum required force.
    • Animal Study: The ground truth is derived from in vivo physiological observations and assessments of performance design attributes in an acute animal model, conducted under GLP regulations. This includes direct observation of the device's behavior and effects within the animal's neurovascular system.
    • Biocompatibility Tests: The ground truth is based on biological reactions and analytical measurements as defined by the specific protocols in ISO 10993 series of standards. For example, for cytotoxicity, the ground truth is the observed cellular response (e.g., cell lysis grade); for pyrogenicity, it's the temperature rise in rabbits; for hemolysis, it's the percentage of red blood cell lysis.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical medical catheter, not a machine learning model. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1