Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K163361

    Validate with FDA (Live)

    Date Cleared
    2017-04-07

    (128 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ANTHEM™ Fracture System is indicated for fixation of fractures, arthrodesis and reconstruction of bones for the appropriate size of the device to be used in adult patients, including the clavicle, scapula, humerus, radius, ulpa, small bones (metacarpals, metatarsals, phalanges), wrist, pelvis, fibula, ankle, and foot. Small fragment and distal fibula plates may be used in all pediatric subgroups (except neonates) and small stature adults. Distal radius plates may be used in adolescents (12-21 years of age).

    Device Description

    The ANTHEM™ Fracture System is a family of plates and screws designed to be used for internal bone fixation. The implants are available in various sizes and shapes to accommodate patient anatomy, and may be contoured or straight, sterile and non-sterile, with locking and non-locking screws. ANTHEM™ implants are manufactured from medical grade titanium alloy, cobalt chromium molybdenum alloy, or stainless steel. All implants are for single use only.

    AI/ML Overview

    This document is a 510(k) Summary for the ANTHEM™ Fracture System, a medical device. Based on the provided text, there is no study described that proves the device meets specific acceptance criteria based on AI/algorithm performance, nor is there any mention of an AI device. The document explicitly describes the ANTHEM™ Fracture System as a system of "plates and screws designed to be used for internal bone fixation."

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance related to AI/algorithmic accuracy.
    • Sample size and data provenance for an AI test set.
    • Number of experts and qualifications for AI ground truthing.
    • Adjudication method for an AI test set.
    • MRMC comparative effectiveness study for AI assistance.
    • Standalone AI performance.
    • Type of ground truth used for AI.
    • Sample size for training set for AI.
    • How ground truth for a training set was established for AI.

    The document focuses on the mechanical and material performance of bone fixation devices, not on diagnostic or AI-driven performance.

    Here's a summary of the performance data that is mentioned in the document, which pertains to the physical device itself:

    1. Acceptance Criteria and Reported Device Performance (for the physical device):

    • Acceptance Criteria (Implicit Standard Performance): The device's performance was evaluated in accordance with recognized ASTM standards for bone fixation devices. This implies that the device had to meet performance benchmarks defined by these standards to demonstrate substantial equivalence to predicate devices. While specific numerical acceptance criteria are not explicitly stated (e.g., "minimum bending strength of X N-m"), the adherence to these standards serves as the implicit acceptance criteria for mechanical performance.
    • Reported Device Performance:
      • Tests Conducted: Engineering analysis, bending strength tests (for plates and screws), pullout strength tests (for screws), and insertion/removal torque tests (for screws).
      • Outcome: "Performance data demonstrates substantial equivalence to the predicate devices."
      • Bacterial Endotoxin Testing (BET): Conducted in accordance with ANSI/AAMI ST-72:2011 to ensure sterility or low endotoxin levels.

    2. Sample Size and Data Provenance (for the physical device testing):

    • The document does not specify the sample sizes used for the mechanical performance tests (e.g., how many plates or screws were tested for bending strength).
    • The data provenance is not mentioned (e.g., where the tests were conducted, whether the materials were from specific batches, etc.).

    3. Number of experts used to establish the ground truth... and qualifications of those experts:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm requiring expert review for ground truth.

    4. Adjudication method for the test set:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done...:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For the physical device, "ground truth" would be the objective measurements obtained from standardized mechanical tests (e.g., actual force at failure, actual torque values) against the requirements of the ASTM standards.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140357

    Validate with FDA (Live)

    Date Cleared
    2014-06-09

    (117 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OIC Variable Angle Small Fragment Locking Plate System is indicated for the fixation of fractures, mal-unions, nonunions or osteolomies for the clavicle, humerus, radius, ulna, metacarpal, tibia, fibula, malleolus and metatarsal.

    Device Description

    The OIC Variable Ande Small Fragment Locking Plate System consists of titanium plates for the Distal Radius. Clavice. Proximal Humerus, Tibia and Distal Fibula, bone screws and instruments for implantation. The platety of sizes and are pre-contoured to match the anatomy of the patient and 3.5mm bone screws. The bone screws are available in two diameters (2.5mm) and range in length from 6mm to 130mm. The bone screws are available with both threaded (locking) and non-threaded (non-locking) heads.

    The OIC Variable Angle Small Fragment Locking Plate System implants are made of titanium alloy or cp titanium in compliance with ASTM F136 or ASTM F67.

    AI/ML Overview

    The provided text describes a medical device submission (K140357) for the OIC Variable Angle Small Fragment Locking Plate System. However, it does not contain information regarding traditional acceptance criteria or a study proving the device meets said criteria in the context of clinical performance or diagnostic accuracy.

    Instead, this document focuses on demonstrating substantial equivalence to existing legally marketed devices, primarily through engineering and material testing, rather than clinical trials with patient outcomes or AI-driven performance studies.

    Therefore, many of the requested categories cannot be filled as the information is not present in the provided text.

    Here's an attempt to answer based on the available information, with clear indications where the information is not present:


    Acceptance Criteria and Device Performance Study for K140357: OIC Variable Angle Small Fragment Locking Plate System

    The provided 510(k) summary for the OIC Variable Angle Small Fragment Locking Plate System does not detail specific acceptance criteria or clinical performance studies related to diagnostic accuracy or human-in-the-loop improvements for an AI device. Instead, the document focuses on demonstrating substantial equivalence through mechanical testing and material compliance.

    The "acceptance criteria" can be inferred from the standards the device conforms to and the evaluation methods used to demonstrate substantial equivalence, which are primarily related to mechanical properties and materials.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Inferred from standards/methods)Reported Device Performance
    Material ComplianceConformance to ASTM F136 or ASTM F67Implants made of titanium alloy or cp titanium in compliance with ASTM F136 or ASTM F67.
    Bone Screw StandardsConformance to ASTM F543-07Device conforms to ASTM F543-07.
    Metallic Bone Plate StandardsConformance to ASTM F382Device conforms to ASTM F382.
    Mechanical CharacteristicsAcceptable mechanical characteristics for intended uses, similar to predicate devices.The screws, plates, locking interfaces, and behavior were evaluated using Finite Element Stress/Strain Analyses (FEA). They were "found to have acceptable mechanical characteristics for the intended uses." Each component and construct showed "similar mechanical performance compared to similar devices used for the same indications as identified via a literature review."
    SterilizationProvided non-sterile, suitable for steam sterilization.The device is provided non-sterile and is steam-sterilized by the medical facility prior to implantation.
    Substantial EquivalenceNo significant differences from predicate devices affecting safety and effectiveness."The new device is substantially equivalent to the predicate devices in regards to intended use, materials, and function. There are no significant differences... Any minor differences have no effect on safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable/not provided. The evaluation was primarily based on engineering analyses (FEA) and literature review, not a test set of data in the context of AI or clinical outcomes.
    • Data Provenance: Not applicable/not provided. The document refers to "literature review" for comparative mechanical performance, but does not specify the origin of this literature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Not applicable. The ground truth for mechanical performance was established via engineering analyses and comparisons to published standards and literature, not expert consensus on a test set of cases.

    4. Adjudication Method for the Test Set

    • Not applicable. No test set requiring expert adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices or AI-assisted interpretation, which is not the nature of this orthopedic implant submission.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm performance study was not done. This device is a physical orthopedic implant, not an algorithm.

    7. The Type of Ground Truth Used

    • The 'ground truth' for this 510(k) submission relates to engineering standards, material specifications, and documented mechanical performance characteristics established through finite element analysis and literature review. It is not clinical pathology, outcomes data, or expert consensus on patient cases.

    8. The Sample Size for the Training Set

    • Not applicable. This device is an orthopedic implant, not an AI algorithm requiring a training set. The "training" for its design and evaluation would come from established engineering principles, material science, and the design of predicate devices.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no "training set" in the context of this orthopedic implant. The benchmarks for performance (comparable mechanical properties, material compliance, etc.) were established through recognized ASTM standards, the performance of predicate devices, and engineering analyses (FEA).
    Ask a Question

    Ask a specific question about this device

    K Number
    K132296

    Validate with FDA (Live)

    Device Name
    D-RAD SMART PACK
    Date Cleared
    2014-01-07

    (168 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K993106, K051735, K083032

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D-RAD SMART PACK is intended for the fixation of fractures involving the distal radius.

    Device Description

    The D-RAD SMART PACK is a cost-effective sterile procedural kit that contains an anatomically contoured volar distal radius plate and single-use instruments to treat extra-articular and intra-articular distal radius fractures. The Kit consists of an anatomically shaped distal radius fixation plate, associated locking and non-locking screws and pegs, and instrumentation to assist in implantation of the fixation construct.

    AI/ML Overview

    The provided document, K132296, for the Smith & Nephew D-RAD SMART PACK is a 510(k) Premarket Notification. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than on proving performance against specific acceptance criteria through a clinical study. Therefore, much of the requested information (like acceptance criteria, specific device performance metrics beyond mechanical testing, sample sizes for test/training sets, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, and detailed ground truth establishment) is not typically found or required in a 510(k) submission for this type of device.

    The document explicitly states: "Mechanical testing and analysis has been conducted on the devices to ensure that they meet or exceed predicate standards in terms of fatigue, material strength, and biocompatibility." This indicates that mechanical testing was the primary method used to demonstrate that the device performs equivalently to the predicate devices.

    Here's an attempt to answer the questions based on the available information, noting where information is not present in the document:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Mechanical PerformanceDevice meets or exceeds predicate standards for fatigue, material strength, and biocompatibility.Devices meet or exceed predicate standards in terms of fatigue, material strength, and biocompatibility. (Stated in "Technological Characteristics" section)
    Functional EquivalenceIdentical function to predicate devices.Identical in function to predicate devices. (Stated in "Substantial Equivalence Information" and "Conclusion" sections)
    Intended UseIdentical intended use to predicate devices for distal radius fractures.Identical in intended use for distal radius fractures. (Stated in "Intended Use" and "Conclusion" sections)
    Material/DesignSimilar materials, processing, and design to predicate devices.Similar in materials, processing, design. (Stated in "Conclusion" section)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable or not specified for a clinical study. The "test set" here refers to the actual device components undergoing mechanical testing. The exact number of samples tested is not provided in this summary.
    • Data Provenance: Not applicable for a clinical study. Data would be generated from laboratory mechanical testing. The country of origin for this testing is not specified. It is laboratory data, not patient data (retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth as typically understood in AI/imaging studies (e.g., expert consensus on diagnoses) is not relevant for this type of medical device submission. The "ground truth" for mechanical performance is established by standardized material and mechanical testing protocols, interpreted by engineers and regulatory experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is not a study involving human interpretation or adjudication of cases. Mechanical testing results are typically evaluated against pre-defined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a bone plate and screw system, not an AI-assisted diagnostic or therapeutic device. Therefore, a MRMC comparative effectiveness study involving human readers and AI assistance was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This device is a physical implant and surgical instrumentation, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For mechanical performance: Compliance with established industry standards and engineering specifications for fatigue, material strength, and biocompatibility. These standards themselves are developed from extensive research and clinical experience.
    • For substantial equivalence: Comparison against the design, materials, and mechanical properties of the predicate devices.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. No training set for an algorithm was used.

    Summary of the Study (Mechanical Testing for Substantial Equivalence):

    The "study" referenced in this 510(k) submission is primarily a series of mechanical tests and analyses designed to demonstrate that the D-RAD SMART PACK meets or exceeds the performance standards of its predicate devices in terms of fatigue, material strength, and biocompatibility. This type of testing is standard for orthopedic implants. The document indicates that the device's functional characteristics, intended use, and materials were also compared directly to three predicate devices previously cleared by the FDA (K993106, K051735, K083032). The conclusion is based on this direct comparison and mechanical data rather than a clinical trial or AI performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123055

    Validate with FDA (Live)

    Date Cleared
    2013-01-25

    (116 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Smith & Nephew Plate and Screw Systems Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Plate and Screw Systems and their cleared Indications for Use.

    PERI-LOC Locking Bone Plates, Locking Hole Inserts and Cable Accessories:
    Smith & Nephew PERI-LOC Locking Bone Plates, Locking Hole Inserts and Cable Accessories Instruments are accessory devices and are intended to assist in the implantation of Smith & Nephew Periarticular Locked Plating Systems and their cleared indications for use.
    The PERI-LOC Periarticular Locked Plating System Proximal Femur Bone Plates, Bone Screws, Locking Hole Inserts and Cable Accessories are indicated for:

    • fractures of the trochanteric region including simple intertrochanteric, reverse oblique trochanteric, transverse trochanteric, complex multi-fragmentary, and fractures with the medial cortex instability
    • proximal femur fractures combined with ipsilateral shaft fractures
    • pathological fractures of the proximal femur including metastatic fractures
    • proximal femur osteotomies
    • fixation of fractures in osteopenic bone
    • fixation of nonunions and malunions
    • basi/transcervical femoral neck fractures
    • subcapital femoral neck fractures
    • subtrochanteric femur fractures.
      In addition, the PERI-LOC Periarticular Locked Plating System including Locking Hole Inserts and Cable Accessories can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC bone plates and screws are indicated for:
    • fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus, and clavicle.

    Smith & Nephew VLP FOOT Plating System, Screw System and Accessories:
    Smith & Nephew VLP FOOT Plating System, Screw System and Accessories Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew VLP FOOT Plating System and Screw Systems and their cleared indications for use.
    The Smith & Nephew VLP FOOT Plating System is indicated for:

    • the treatment of fracture fixation, reconstruction or arthrodesis of small bones, including those in the forefoot, midfoot and hindfoot.
      The Smith & Nephew Screw Systems, containing 2.5mm, 3.0mm Cannulated and 3.0mm Headless Compression Screws, are indicated for:
    • fixation of interarticular and extra-articular fractures and non-unions of small bones and small bone fragments;
    • arthrodesis of small joints;
    • bunionectomies and osteotomies;
    • scraphold and other carpal bones;
    • metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.
      The Smith & Nephew 2.0mm QFX Screw is indicated for:
    • osteotomies of the lesser metatarsals, such as Weil osteotomies, osteotomies, fusions and fractures of the phalanges, metacarpals and carpals of the hand.
      Smith & Nephew Accessories, such as pins and wires, are indicated for:
    • pelvic, small and long bone fracture fixation.

    Smith & Nephew InterTAN CHS Limited Collapse Set Screw:
    Smith & Nephew InterTAN CHS Limited Collapse Set Screw Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew InterTAN CHS Limited Collapse Set Screw Systems and their cleared indications for use.
    InterTAN CHS Limited Collapse Set Screw is indicated for:

    • Intracapsular fractures of the proximal femur (For certain high subcapsular fractures, it may be more prudent to select a prosthesis in lieu of internal fixation to reduce the risk of nonunion or AVN of the femoral head).
    • Intertrochanteric fractures.
    • Stable and unstable fractuers of the proximal femur in which medial cortex stability can be restored.
    • Hip osteotomy

    PERI-LOC Periarticular Locked Plating System- VLP Locking Bone Plates and Locking/Non-Locking Bone Screws:
    Smith & Nephew VLP Locking Bone Plates and Locking/Non-Locking Bone Screws Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew PERI-LOC Periarticular Locked Plating Systems and their cleared indications for use.
    PERI-LOC contoured VLP Plates and Screws are indicated for:

    • partial articular fractures (AO/OTA Fracture Classification Type B) of the distal and proximal tibia and for fracture fixation of the fibula
      PERI-LOC VLP One-Third Tubular Locking Plates are indicated for:
    • fixation of fractures, non-unions, and osteotomies of the medial malleolus, fibula, distal ulna, olecranon, calcaneus and metatarsals.

    Smith & Nephew 6.5mm and 8.0mm Cannulated Screws:
    Smith & Nephew 6.5mm and 8.0mm Cannulated Screw Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew 6.5mm and 8.0mm Cannulated Screw Systems and their cleared indications for use.
    Smith & Nephew 6.5mm and 8.0mm Cannulated Screws are indicated for:

    • fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones;
    • treatment of the calcaneal
    • hip arthrodesis
    • provisional bone fixation.

    Smith & Nephew Bone Plate System:
    Smith & Nephew Bone Plate System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Bone Plate Systems and their cleared indications for use.
    Smith & Nephew Bone Plates, Bone Screws, and Accessories are indicated for:

    • pelvic fracture fixation
    • small bone fracture fixation
    • long bone fracture fixation.

    Titanium Classic Compression Hip Screw System:
    Smith & Nephew Titanium Classic Compression Hip Screw System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Titanium Classic Compression Hip Screw Systems and their cleared indications for use.
    The Titanium Classic Compression Hip Screw System is indicated for:

    • Pelvic Radius Plate- general fractures of the pelvis and acetabulum
    • Tibial Plateau Plate- fractures of the proximal tibia with or without diaphyseal involvement
    • One Third Tubular Plate- fractures of the fibula, lateral malleolus, metacarpals, metatarsals, olecranon, and distal ulna
    • Small Auto-Compression Plate- fracture fixation of metatarsals, metacarpals, radius, and ulna
    • Titanium Cortical Bone Screw- small and long bone fracture fixation
    Device Description

    Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. Plate and Screw Systems Instruments. The subject devices are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Plates and Screws and their cleared Indications for Use. Smith & Nephew Plate and Screw Systems Instruments can be organized into instrument families which are categorized as follows: Templates, Reamers, Handles, Bases, Pins, Taps, Drill Guides, Insertion or Removal, and Other Guides.

    AI/ML Overview

    This document describes a 510(k) premarket notification for "Plate and Screw Instruments" by Smith & Nephew, Inc. It does not contain information about acceptance criteria or a study proving the device meets said criteria in the context of an AI/ML medical device.

    The document pertains to orthopedic surgical instrumentation, which are physical tools (templates, reamers, handles, pins, taps, drill guides, etc.) used to assist in the implantation of bone plates and screws. These are accessory devices to existing cleared bone fixation systems.

    The submission is for substantial equivalence to previously cleared predicate devices, not for a novel device requiring performance studies or clinical trials to establish new safety and effectiveness claims based on specific acceptance criteria. The basis for substantial equivalence is listed as:

    • Sharing the same raw materials
    • Manufactured through the same processes
    • Utilizing the same sterilization procedures
    • Having a similar nature of body contact
    • Similar in design and function to competing surgical instrumentation on the market.

    Therefore, the requested information points, which are typically relevant for AI/ML device submissions involving performance benchmarks and clinical studies, are not applicable to this document. There are no acceptance criteria related to a model's performance, no data sets (training or test), no ground truth establishment, and no mention of human-in-the-loop studies or standalone algorithm performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1