Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K161689

    Validate with FDA (Live)

    Date Cleared
    2017-05-22

    (336 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSSTEM Implant System - Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

    Device Description

    The OSSTEM Implant System - Abutment is intended for use as an aid in prosthetic restoration. It consists of Abutments, components and Abutment Screws.

    AI/ML Overview

    This document is a 510(k) Premarket Notification of the OSSTEM Implant System - Abutment, indicating it is a dental device and not an AI/ML medical device. Therefore, the information requested in the prompt, which pertains to the acceptance criteria and study proving an AI/ML device meets those criteria, cannot be fully provided from this document.

    The document discusses the substantial equivalence of the OSSTEM Implant System - Abutment to predicate devices. This typically involves demonstrating that the new device is as safe and effective as a legally marketed predicate device, often through material comparisons, design comparisons, and performance testing for mechanical properties and biocompatibility, rather than clinical studies or performance against defined efficacy metrics tied to patient outcomes.

    However, I can extract information related to the device and the type of testing performed that is relevant to its acceptance for market.

    Here's a breakdown based on the provided text, addressing the points of your prompt where applicable, and highlighting what's not present due to the nature of the device:

    Device Description: The OSSTEM Implant System - Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. It consists of Abutments, components, and Abutment Screws.

    Acceptance Criteria and Device Performance (General):
    For this type of device, "acceptance criteria" revolve around demonstrating substantial equivalence to existing, legally marketed predicate devices. This means proving the new device is as safe and effective as the predicate. The performance is assessed through:

    • Material composition: Comparing the materials used (e.g., Titanium Alloy, Zirconia, PEEK, PC, POM).
    • Design and dimensions: Comparing the shapes, diameters, and lengths.
    • Intended use: Confirming the device serves the same purpose.
    • Technological characteristics: Ensuring the fundamental mechanism of action is similar.
    • Nonclinical testing: Biocompatibility, sterilization validation, and mechanical properties.

    Table of Acceptance Criteria and Reported Device Performance (as inferred from a 510(k) for a non-AI/ML device):

    Acceptance Criteria CategorySpecific Criteria (Inferred for a Dental Abutment)Reported Device Performance (as demonstrated by comparison to predicates)
    Intended UseMust be for use with dental implants to support prosthetic restorations.Identical to predicate devices for all components. (e.g., page 7: "Indication for use of subject device is slight differences in phrase with predicate but fundamental Indication is same")
    MaterialsMust use materials with established biocompatibility and mechanical properties for dental implants.Identical to predicate devices (e.g., Titanium Alloy (Ti-6Al-4V), Zirconia, PEEK, PC, POM), or demonstrating equivalence if slight variations exist. (e.g., page 7: "Material: Zirconia, Identical"; page 8: "Material: Titanium Alloy PEEK, Identical")
    Design/DimensionsMust be geometrically compatible with specified implant systems and suitable for intended restorations.Substantially Equivalent or Identical designs and dimensions. Some new dimensions are added but fall within accepted ranges, or minor cosmetic/functional modifications are made that do not raise new safety/effectiveness questions. (e.g., page 7 "Connection structure is different (Hex and Non-Hex) but Design is Substantial Equivalence"; page 8 "Diameter 4.0mm is added"; page 9 "Design is Substantial Equivalence With highlighted predicate with red Box")
    Technological CharacteristicsMust function physically in a manner similar to predicate dental abutments (e.g., cement-retained, screw-retained).Identical to predicate devices (e.g., page 7 "Use for making general cement-type prosthesis. Identical"; page 8 "Cement retained restoration. Capable of altering/removing shape of plastic material. Two piece (Abutment + Screw) Identical")
    BiocompatibilityMust be biocompatible, as per ISO 10993 standards.Relying on identical materials and manufacturing processes as previously cleared predicate devices. "Therefore, no additional testing is necessary." (page 11)
    SterilizationMust be able to be sterilized to an appropriate Sterility Assurance Level (SAL).For Custom Healing Abutment, leveraged data from predicate device (ISO 11137-1, -2, -3 and shelf life testing). For other non-sterile devices, steam sterilization validation conducted per ISO 17665-1 and -2. (page 11)
    Mechanical PerformanceMust withstand normal forces and maintain integrity in the oral cavity. Generally, this would be mechanical testing.The document does not explicitly detail the mechanical performance results in a table format, but states that the device is "substantially equivalent in design, function and intended use" to predicates. This implies mechanical performance is considered equivalent due to material and design similarity, and potentially prior testing on the predicate devices. The listed nonclinical tests (like "tensile test (ASTM F882), seal peel test (ASTM F88/EN868-5), burst test (ASTM F1140), dye penetration (ASTM F1929), bubble test (ASTM F2096)") mentioned under sterilization validation are related to packaging and sterility, not the inherent mechanical strength of the abutment itself, which would be covered by separate mechanical testing for dental implants if required for initial approval. For this 510(k), they rely on the substantial equivalence premise.

    Here's how the remaining points of your prompt are addressed:

    1. Sample sized used for the test set and the data provenance:

      • Test set sample size: Not applicable in the context of an AI/ML algorithm's test set. The "testing" here refers to nonclinical lab tests and comparison to predicates.
      • Data provenance: Not directly applicable. The "data" are internal company reports demonstrating compliance with standards or comparisons to their own previously cleared devices. The manufacturer is OSSTEM Implant Co., Ltd. and their correspondent is HIOSSEN Inc., based in Republic of Korea and Pennsylvania, USA, respectively. The data are typically generated in a prospective manner for regulatory submission (i.e., tests are performed specifically for the submission).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable as this is a physical medical device, not an AI/ML algorithm requiring ground truth from experts for image interpretation or diagnosis.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept pertains to consolidating expert opinions for ground truth in AI/ML performance studies.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done, as this is not an AI/ML device. The document explicitly states: "No clinical studies are submitted." (page 12)
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable, as this is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable. For this device, "ground truth" is embodied by established engineering specifications, material properties, and performance benchmarks for dental implants, which are verified through nonclinical laboratory testing against international standards (e.g., ISO, ASTM) and comparison to predicate device characteristics.
    7. The sample size for the training set:

      • Not applicable. There is no concept of a "training set" for this type of conventional physical device.
    8. How the ground truth for the training set was established:

      • Not applicable. (See #7).

    Summary of Acceptance:

    The device's acceptance (clearance) by the FDA is based on demonstrating substantial equivalence to existing predicate devices (listed in section 3 and 6) in terms of:

    • Indications for Use: Identical.
    • Technological Characteristics: Very similar, with minor differences in design/dimensions that do not raise new safety/effectiveness concerns.
    • Materials: Identical.
    • Performance (Nonclinical): Biocompatibility is leveraged from predicate devices; sterilization validated according to relevant ISO standards.

    The document explicitly states: "No clinical studies are submitted." This reinforces that the acceptance was based on nonclinical testing and comparison to predicates, as is common for many Class II medical devices in the 510(k) pathway.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120847

    Validate with FDA (Live)

    Date Cleared
    2012-08-10

    (143 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ET/SS Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. The abutment is intended for use with a dental implant fixture to provide support for prosthetic restorations such as crowns, bridges, or overdenture.

    Device Description

    The ET/SS Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated. Abutment is device made of titanium alloy intended for use as an aid in prosthetic restoration. It consists of Abutments and Abutment Screws. Its surfaces are partially Tin coated and uncoated.

    AI/ML Overview

    This 510(k) premarket notification for the ET/SS Implant System does not contain the level of detail typically found in a study describing acceptance criteria and device performance for AI/ML-based medical devices. The submission focuses on demonstrating substantial equivalence to predicate dental implants through a comparison of physical characteristics, intended use, and a summary of nonclinical testing (fatigue testing).

    Therefore, I cannot directly extract the specific information requested in your prompt regarding acceptance criteria, study details, ground truth, sample sizes, or MRMC studies, as the document is for a traditional medical device (dental implant) and not an AI/ML device.

    However, I can extract the information provided about the nonclinical testing for this device and present it in a similar structure:

    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance CriteriaReported Device Performance
    Fatigue TestingCompliance with ISO 14801 and "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" with the worst-case scenario.Results were in compliance and similar to previously cleared predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable as this is a nonclinical, component-level fatigue test, not a performance study on a test set of data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth for fatigue testing is established by engineering standards and measurement.

    4. Adjudication Method

    Not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, this type of study was not done. The device is a physical dental implant, not an AI system that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The "device" is a physical implant. The fatigue testing is a standalone test of the implant's physical properties.

    7. The Type of Ground Truth Used

    Engineering standards and measurements for fatigue life.

    8. The Sample Size for the Training Set

    Not applicable. There is no concept of a "training set" for physical fatigue testing of a dental implant in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.


    Summary of Device Information from the Provided Document:

    • Device Name: ET/SS Implant System
    • Manufacturer: OSSTEM Implant Co., Ltd.
    • Device Type: Endosseous Dental Implant
    • Classification: Class II
    • Key Feature: Made of pure titanium metal, SA (Sandblasting and Acid etching) treated surface.
    • Intended Use: For use in partially or fully edentulous mandibles and maxillae, supporting single or multiple-unit restorations (cemented, screw-retained, or overdenture), and for final or temporary abutment support for fixed bridgework. Intended for delayed loading.
    • Nonclinical Testing: Fatigue testing performed according to ISO 14801 and relevant FDA guidance.
    • Clinical Testing: No clinical studies were submitted.
    • Basis for Equivalence: Substantial equivalence was claimed based on material, indication for use, and similar design and technological characteristics to predicate devices, supported by nonclinical testing results.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1