Search Results
Found 1 results
510(k) Data Aggregation
(147 days)
The Osstem Implant is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Ultra Wide Fixture System is intended to be used in the molar region.
The SS SA Fixture is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated.
This document is a 510(k) Premarket Notification from Osstem Implant Co., Ltd. to the FDA for their "SS SA Fixture" dental implant. It aims to demonstrate substantial equivalence to legally marketed predicate devices rather than proving the device meets specific acceptance criteria through a study with performance metrics as one might see for a diagnostic AI device.
Therefore, the requested information elements (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC study, standalone performance, ground truth types, training set size, and training ground truth establishment) are generally not applicable in the context of this 510(k) submission for a non-AI dental implant.
However, I can extract information related to the device's technical specifications and the non-clinical testing performed to establish substantial equivalence.
Here's an adaptation of the requested format based on the available information:
1. A table of acceptance criteria and the reported device performance
For a dental implant, "acceptance criteria" usually refers to meeting established standards (e.g., ISO, ASTM) and demonstrating equivalence to predicate devices, rather than specific performance metrics like sensitivity/specificity for an AI algorithm. The device performance is primarily assessed through non-clinical testing to ensure its physical and mechanical properties are comparable or superior to the predicate.
| Acceptance Criterion (Based on predicate equivalence and applicable standards) | Reported Device Performance (SS SA Fixture) |
|---|---|
| Material Composition (Pure Titanium Grade 4 per ASTM F67) | Same as Predicate: Pure Titanium Grade 4 (ASTM F67) |
| Surface Treatment (SA - Sandblasted and Acid etched) | Same as Predicate: SA surface treatment. No change from predicate. |
| Sterilization (Radiation Sterile) | Same as Predicate: Radiation Sterile. No change from predicate. |
| Shelf Life (Compared to predicate) | Same as Reference Predicate: 8 years (Primary Predicate was 5 years, Reference Predicate was 8 years). |
| Biocompatibility (As established for predicate material and processes) | No additional testing required: Material of construction and manufacturing processes are the same as predicate devices (ET/SS Implant System, K120847 and OSSTEM Implant System, K161604). |
| Bone to Implant Contact (BIC) Value (Comparable to predicate device with similar lengths) | Substantial equivalence confirmed: BIC analysis conducted for implants < 7mm. Compared BIC value and confirmed substantial equivalence with ET/SS Implant System (K120847). |
| Surface Area Analysis (Comparable to predicate device with similar lengths) | Substantial equivalence confirmed: Surface area analysis conducted for implants < 7mm. Compared surface area and confirmed substantial equivalence with ET/SS Implant System (K120847). |
| Pullout Force Value (Comparable to predicate device with similar lengths) | Substantial equivalence confirmed: Pullout force test conducted for implants < 7mm. Compared pullout force value and confirmed substantial equivalence with ET/SS Implant System (K120847). |
| Fatigue Performance (Meet ISO 14801; considered against worst-case scenario and predicate) | Deemed superior/not requiring additional testing: SS SA Fixture has bigger diameter than primary predicate (ET/SS Implant System, K120847). The result of static and dynamic compression-bending test is expected to be higher than this predicate, thus no additional fatigue test was considered necessary. Testing was considered according to "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" and ISO 14801. |
| Indications for Use (Comparable to predicate devices) | Same as Predicate: The indications for use are substantially equivalent to the predicate devices, with the addition of "Ultra Wide Fixture System is intended to be used in the molar region." |
| Structural Characteristics (E.g., Non Submerged, Self-tapping, Internal Octagonal connection, Taper Body) | Same as Predicate: Has same connection structure, material, manufacturing process, principles of operation, and technological characteristics as ET/SS Implant System (SSII/III SA Fixture). Some dimensions (diameter) are larger than the primary predicate but are included in the range of the reference predicate. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set (for non-clinical studies): Specific sample sizes for each non-clinical test (BIC, Surface Area, Pullout, Fatigue) are not explicitly stated in the summary, only that these tests were conducted "with SS SA Fixture containing implants of lengths less than 7mm." The data provenance is implied to be from the manufacturer's internal testing. This is not patient data, but material/device performance data.
- Data Provenance: Not applicable in the sense of patient data origin as these are non-clinical (laboratory/bench) tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable as this is a 510(k) for a physical medical device, not a diagnostic AI device requiring expert-established ground truth from images or clinical data. The "ground truth", in this context, refers to established engineering and material science standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable for non-clinical bench testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI-based device, and no clinical studies were submitted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable as this is not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's evaluation relies on:
- Regulatory Standards: Compliance with FDA regulations (21 CFR Part 872.3640), Class II Special Controls, and international standards (e.g., ISO 14801 for fatigue testing, ASTM F67 for material).
- Predicate Device Equivalence: The established safety and effectiveness of the legally marketed predicate devices (K120847, K161604).
- Scientific Principles: Established principles of biocompatibility, mechanical engineering, and material science.
8. The sample size for the training set
- Not applicable. This is a physical device, not an AI model, and therefore does not have a "training set."
9. How the ground truth for the training set was established
- Not applicable. No training set as it's not an AI model.
Ask a specific question about this device
Page 1 of 1