Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K121670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-25

    (233 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medyssey Co. Ltd. Zenius, Iliad and Kora Spinal Systems are intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograff and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Medyssey Co. Ltd., Cannulated Pedicle Screws are intended to be used with their respective (e.g. Zenius) top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Medyssey Co. Ltd., Cannulated Pedicle Screw implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. The Medyssey Co. Ltd., Cannulated Pedicle Screws can be used in the posterior plane providing unilateral and bilateral modes of fixation. The Medyssey Co. Ltd., Cannulated Pedicle Screw Spinal design allows adjustment in both the sagittal and coronal planes permitting screw placement according to the best possible anatomic (spinal) location and orientation. This is accomplished by means of a mobile and non-mobile housing component which houses the screw and the rod which accepts a setscrew which tightens against the rod which tightens head of the pedicle screw upon secure tightening interface of the set screw assembly with the rod. The Cobalt Chrome Rods are for use with the Zenius or Iliad System. Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Medyssey Co. Ltd. Cannulated Pedicle Screw implants.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Medyssey Cannulated Pedicle Screws and Cobalt Chrome Rods:

    1. Acceptance Criteria and Reported Device Performance

    The provided document describes non-clinical (bench) testing for the Medyssey Cannulated Pedicle Screws and Co-Cr-Mo rods. The acceptance criteria are implicitly defined by the chosen ASTM standard and the comparison to predicate devices, which means the new device must perform at least as well as the legally marketed predicates.

    Acceptance Criteria CategorySpecific Criteria (Implicitly from ASTM F1717-04 and Predicate Comparison)Reported Device Performance
    Mechanical Performance- Static Compression Strength"Analysis has shown that the proposed screws and rods are equivalent to the predicate devices." (Implies meeting or exceeding predicate performance for all tests listed below)
    - Static Torsion Strength
    - Static Tension Strength
    - Dynamic Compression Performance (Fatigue)
    Material Compatibility- Use of biocompatible materials (Titanium alloy [Ti-6Al-4V ELI], Co-Cr-Mo) conforming to ASTM F 136."fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136." and "proposed rods are manufactured from Co-Cr-Mo"
    Design Equivalence- Minor differences in design (cannulation of screws, Co-Cr-Mo rods instead of titanium) do not impact safety or effectiveness."Analysis has shown that the proposed screws and rods are equivalent to the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify the exact number of devices (screws/rods) tested for each non-clinical test (static compression, torsion, tension, dynamic compression). Standard test methodologies for medical devices like ASTM F1717-04 typically require a minimum number of samples for statistically valid results, but these specific numbers are not detailed in this summary.
    • Data Provenance: The data is from non-clinical (bench) testing performed by Medyssey Co. Ltd. The country of origin of the testing data is not explicitly stated, but Medyssey is a Korean company with a US-based contact for regulatory purposes.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the study described is non-clinical (bench testing) of physical device properties, not a clinical study involving human patients or expert interpretation of medical data. The "ground truth" here is the measured mechanical performance of the device against established engineering standards and predicate device performance.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study described is non-clinical (bench testing). Adjudication methods are typically employed in clinical studies or studies involving expert review of medical images/data to resolve discrepancies in interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document describes a 510(k) submission for a physical medical device (spinal pedicle screws and rods), not an AI/software device. Therefore, there is no mention of human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical implant, not an algorithm or software.

    7. The Type of Ground Truth Used

    The "ground truth" used for this non-clinical study is:

    • Engineering Standards and Specifications: Adherence to ASTM F1717-04.
    • Predicate Device Performance: The mechanical performance of the new device was compared to that of legally marketed predicate devices. The implicit ground truth is that the predicate devices are safe and effective, and therefore, a device demonstrating equivalent or superior mechanical properties would also be considered safe and effective for the stated indications.

    8. The Sample Size for the Training Set

    This information is not applicable as the study described is non-clinical (bench testing) of a physical device. There is no concept of a "training set" for this type of evaluation.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this non-clinical study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100788

    Validate with FDA (Live)

    Device Name
    COCR RODS
    Date Cleared
    2010-05-14

    (53 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K061202, K081195, K091782, K073172, K092610, K093058

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The REVERE® or BEACON™ Stabilization System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, pseudoarthrosis and failed previous fusion.

    In addition, the REVERE® or BEACON™ Stabilization System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

    When used as a posterior non-pedicle screw fixation system, the REVERE® or BEACON™ Stabilization System is intended for the treatment of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or lordosis, Scheuermann's disease), fracture, pseudarthrosis, tumor resection, and/or failed previous fusion. Overall levels of fixation are T1-sacrum/ilium.

    When used as an anterolateral thoracolumbar system, the REVERE® or BEACON™ Stabilization System is intended for anterolateral screw (with or without staple) fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or lordosis), fracture or dislocation of the thoracolumbar spine, pseudoarthrosis, tumor resection, and/or failed previous fusion. Levels of screw fixation are T8-L5.

    Device Description

    CoCr Rods consist of 5.5mm and 6.35mm rods manufactured from cobalt chromium molybdenum alloy. REVERE® and BEACON™ Stabilization System implants mate with these 5.5mm and 6.35mm diameter rods. The REVERE® and BEACON™ implant components can be rigidly locked into a variety of configurations for the individual patient and surgical condition. CoCr Rods are composed of cobalt chromium-molybdenum alloy as specified in F1537.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary for the CoCr Rods, focusing on acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device PerformanceComments
    Mechanical PerformanceCompression-bendingTested in accordance with "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004.The document states testing was conducted in accordance with the guidance, implying the results met the criteria outlined in that guidance. Specific numerical results or pass/fail thresholds are not provided in this summary.
    Mechanical PerformanceTorsionTested in accordance with "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004.Similar to compression-bending, this indicates compliance with the guidance, but specific quantitative results are absent.
    MaterialBiocompatibilityComposed of cobalt chromium-molybdenum alloy as specified in F1537.This implies adherence to a recognized standard for surgical implant materials, which covers biocompatibility. No separate test results are presented.
    Substantial EquivalenceTechnical CharacteristicsSimilar to predicate devices.The submission asserts similarity without specific details.
    Substantial EquivalenceMaterialSimilar to predicate devices (CoCr-Mo alloy).Directly mentioned in the device description.
    Substantial EquivalencePerformanceSimilar to predicate devices (based on mechanical testing).Implied by the statement of "substantial equivalence" supported by mechanical testing.
    Substantial EquivalenceIntended UseSimilar to predicate devices.The stated indications for use are largely consistent with typical spinal stabilization systems.

    2. Sample Size Used for the Test Set and the Data Provenance

    The provided document describes mechanical testing of the CoCr Rods. For such testing, the "sample size" refers to the number of physical rods and corresponding components subjected to the mechanical tests.

    • Sample Size: The document does not explicitly state the number of rods or components used in the compression-bending and torsion tests.
    • Data Provenance: The data provenance is pre-clinical mechanical testing, conducted by the manufacturer, Globus Medical Inc. There is no information suggesting external lab testing or geographical origin of the samples beyond the manufacturer's location in Audubon, PA. This is not clinical data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not applicable to this submission. The "ground truth" in this context is established by engineering standards and physical measurements from mechanical stress tests, not by expert clinical consensus. There are no experts involved in establishing a "ground truth" for clinical outcomes in this 510(k) submission.

    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods are used in clinical studies or expert reviews to resolve discrepancies in observations or diagnoses. Mechanical testing results are objective measurements against predefined engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This 510(k) submission is for a medical device (spinal rods) and not for an AI/CAD (Computer-Aided Detection) system. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable, as this is not an AI/CAD device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is based on established engineering standards and material specifications.

    • For mechanical testing (compression-bending, torsion), the ground truth is defined by the failure criteria outlined in the referenced FDA Guidance document ("Guidance for Spinal System 510(k)s", May 3, 2004), as well as the inherent mechanical properties of the material.
    • For material composition, the ground truth is adherence to ASTM F1537 (Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants).

    8. The sample size for the training set

    This question is not applicable. There is no AI algorithm being trained, so there is no training set.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093294

    Validate with FDA (Live)

    Date Cleared
    2010-02-17

    (119 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K061202,K081195,K091782

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The REVERE® Stabilization System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudoarthrosis and failed previous fusion.

    In addition, the REVERE® Stabilization System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or illum with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

    When used as a posterior non-pedicle screw fixation system, the REVERE® Stabilization System is intended for the treatment of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or lordosis, Scheuermann's disease), fracture, pseudoarthrosis, tumor resection, and/or failed previous fusion. Overall levels of fixation are T1-sacrum/ilium.

    When used as an anterolateral thoracolumbar system, the REVERE® Stabilization System is intended for anterolateral screw (with or without staple) fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or lordosis), fracture or dislocation of the thoracolumbar spine, pseudoarthrosis, tumor resection, and/or failed previous fusion. Levels of screw fixation are T8-L5.

    Device Description

    The REVERE® Stabilization System consists of rods, hooks, monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking caps, t-connectors, offset housing clamps, REVERE® head offset connectors, trans iliac connectors, staples, and associated manual surgical instruments. Screws and rods are available in a variety of sizes to accommodate individual patient anatomy. REVERE® implants mate with 5.5mm diameter rods; REVERE® 6.35 implants mate with 6.35mm diameter rods. Implant components can be rigidly locked into a variety of configurations for the individual patient and surgical condition. Polyaxial screws, hooks, and t-connectors are intended for posterior use only. Staples are intended for anterior use only. Rods and monoaxial screws may be used anteriorly or posteriorly. Locking caps are used to connect screws or hooks to the rod.and trans iliac connectors.

    The most common use of this screw, hook, and rod system in the posterior thoracolumbar and sacral spine is two rods, each positioned and attached lateral to the spinous process via pedicle screws and/or lamina, pedicle or transverse process hooks.

    The most common use of this screw, hook, and rod system in the anterior thoracolumbar spine is one rod, positioned and attached to the vertebral bodies via monoaxial screws through an appropriate size staple.

    Screws and hooks attach to the rods using a locking cap with an inner set screw. The size and number of screws are dependent on the length and location of the rod. Screws are inserted into a pedicle of the thoracolumbar and/or sacral spine. The type and number of hooks are also dependent on the location in the spine needing correction and/or stabilization. Hooks are attached to the laminae, pedicles, or transverse process of the posterior spine.

    T-connectors are modular components designed to connect the two rods of a construct and act as a structural cross member. The rod-clamping set screws secure the t-connectors to the rods. Additional set screws secure the adjustable cross members at the desired length. T-connectors from the PROTEX® system may be used with 6.5mm, 6.0mm or 5.5mm rod systems. REVERE® tconnectors may only be used with 5.5mm rods; REVERE® 6.35 t-connectors may only be used with 6.35mm rods. Additional connectors may be used to connect two rods, and are also secured using set screws.

    REVERE® rods are composed of titanium alloy, commercially pure titanium, or stainless steel, as specified in ASTM F136, F1472, F1295, F67, and F138. All other REVERE® implants are composed of titanium alloy or stainless steel, as specified in ASTM F136, F1472, F1295, and F138. Due to the risk of galvanic corrosion following implantation, stainless steel implants should not be connected to titanium or titanium allov implants.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text for the K093294 510(k) summary for REVERE® Additional Implants.

    Important Note: This 510(k) summary (K093294) primarily focuses on establishing substantial equivalence to existing predicate devices through mechanical testing. It does not describe a clinical study in the typical sense of evaluating AI performance or human reader effectiveness. Therefore, many of the requested fields related to clinical studies, AI performance, ground truth, and expert evaluation will be marked as "Not Applicable" or "Not Provided" because this type of submission doesn't involve those elements for this device.


    Acceptance Criteria and Device Performance

    Acceptance Criteria CategoryReported Device Performance (as implied by the K093294 summary)
    Mechanical PerformanceSubstantial equivalence to predicate devices (K061202, K081195, K091782) as demonstrated by mechanical testing in accordance with "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004.
    Material CompatibilityImplants are composed of titanium alloy, commercially pure titanium, or stainless steel, as specified in ASTM F136, F1472, F1295, F67, and F138. Prohibition against connecting stainless steel to titanium/titanium alloy implants to avoid galvanic corrosion.
    Intended UseThe REVERE® Stabilization System, including the additional implants, is intended for various spinal indications (degenerative disc disease, spondylolisthesis, fracture, scoliosis, etc.) as an adjunct to fusion for posterior pedicle screw, posterior non-pedicle screw, and anterolateral thoracolumbar systems. This implies that the device is fit for these stated clinical purposes. The intended use is identical to the predicate device.

    Study Details (as applicable to a 510(k) for a medical device without AI)

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size for Test Set: Not applicable in the context of a clinical test set for AI. The "test set" here refers to the mechanical testing performed on the device components. The specific number of devices or configurations tested is not provided in this summary but would be detailed in the full 510(k) submission as per the referenced guidance document.
      • Data Provenance: Not applicable in terms of country of origin or retrospective/prospective for a mechanical testing study. The testing would have been conducted by Globus Medical Inc. or a contracted lab.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. This submission relies on mechanical and material standards, not expert-derived ground truth for clinical outcomes or image interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This type of adjudication is for clinical or image interpretation studies, not mechanical testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This is a spinal implant, not an AI-powered diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, a standalone algorithm performance study was not done. This is a medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For mechanical testing, the "ground truth" is adherence to established engineering standards (e.g., ASTM standards) and meeting predetermined pass/fail criteria derived from these standards and the predicate device's performance. There is no biological "ground truth" involved in the summary provided.
    7. The sample size for the training set:

      • Not applicable. There is no AI training set for this device.
    8. How the ground truth for the training set was established:

      • Not applicable. There is no AI training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1