    Reference Devices :

    K083627, K061231, K112543, K000281

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WallFlex Biliary RX Fully Covered Stent System RMV is indicated for indwell up to 12 months in the treatment of benign biliary strictures secondary to chronic pancreatitis.

    Device Description

    The WallFlex Biliary RX Fully Covered Stent System RMV consists of 2 components: the implantable fully covered self-expanding metallic biliary stent (FC-SEMBS) and the Rapid Exchange (RX) delivery device. The FC-SEMBS is made out of Nitinol monofilament wire with a radiopaque platinum core, braided in a tubular mesh configuration. The stent has proximal and distal flares at each end to aid in preventing migration. The proximal flare also contains the biliary stent retrieval loop. The biliary stent is covered with a Permalume™ silicone covering along its entire length except for 2 mm on the retrieval loop end. The RX delivery device is a coaxial tube design. The exterior tube is used to constrain the biliary stent before deployment and re-constrain the biliary stent, if biliary stent repositioning is necessary, after partial deployment. The exterior tube has a clear section so that the constrained stent is visible. A yellow transition zone on the inner tube of the delivery system is visible between the stent and the blue outer sheath. There are four radiopaque (RO) markers to aid in the deployment of the stent while using fluoroscopy. There are two RO markers on the inner tube of the delivery system identifying the ends of the constrained biliary stent. Between these RO markers is an additional RO marker that indicates at what point re-constrainment is no longer possible. The fourth RO marker at the leading end of the exterior tube indicates how far the biliary stent has been deployed. There is one visual marker on the interior tube between the handles to aid in the deployment of the biliary stent. The visual marker indicates the point at which re-constrainment is no longer possible. The interior tube has a single central lumen to accommodate a 0.035 inch (0.86 mm) guidewire.

    AI/ML Overview

    The provided text describes the WallFlex Biliary RX Fully Covered Stent System RMV, its design, nonclinical/bench studies, and a clinical study conducted to assess its safety and effectiveness.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria & Reported Device Performance

    The acceptance criteria are primarily derived from the "Special Controls" section and the "Summary of Nonclinical Studies" (Table 2). The reported device performance is extracted from the "Summary of Nonclinical/Bench Studies" section and the "Results" section of the clinical study.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    SterilizationThe sterility assurance level (SAL) shall be 10-6.Passed: Sterilization was re-assessed in 2015 to ensure compliance to most recent version of ANSI/AAMI/ISO 10993-7. Testing was conducted on the 10x120mm fully covered stent, considered worst-case.
    Ethylene Oxide ResidualsEtOH residual amounts are below the maximum amount allowed: For the stent, the average daily dose of EO to patient shall not exceed 0.1 mg/day, max EO dose of 20 mg in first 24 hr, 60 mg in first 30 days, 2.5 g in a lifetime. For the delivery catheter, the average daily dose shall not exceed 4 mg.Passed: EtOH residual amounts are below the maximum amount allowed.
    BiocompatibilityCytotoxicity: Grade 0 (non-cytotoxic); Irritation: Non-irritant (based on histopathologic evaluations); Sensitization: No signs of sensitization from polar and non-polar extracts; Acute Systemic Toxicity: No evidence of mortality or systemic toxicity; Genotoxicity: Non-mutagenic; Implantation: Non-irritant (based on histopathologic evaluations); Subchronic Toxicity: No signs of systemic toxicity due to leachable components; Nickel leaching: Acceptable resistance to corrosion and acceptable nickel leach rates; Toxicological Risk: Not likely to lead to systemic toxicity.Passed: Cytotoxicity Grade 0. Irritation non-irritant. No signs of sensitization. No evidence of mortality or systemic toxicity. Non-mutagenic. Non-irritant after implantation. No signs of systemic toxicity. Acceptable nickel leaching demonstrated. Exposure to the stent is not likely to lead to systemic toxicity.
    Deployment TestingThe delivery catheter must safely and accurately deliver the stent to the intended anatomic location. No damage to the stent or simulated stricture model should occur.Passed: Stent sizes tested represent the smallest and largest sizes available and can be used as surrogates. The stent was successfully placed in 100% (127/127) of patients in the clinical study.
    Expansion/Compression Force TestingRadial expansion forces must be at least 85% of specified N values for various stent diameters and compressions. Radial compression forces must be at least 85% of specified N values for various stent diameters and compressions.Passed: The device met the specified radial expansion and compression forces.
    Dimensional TestingAll dimensions should be within design tolerance ranges.Passed: Dimensions of each stent size and delivery catheter were measured and verified to meet the acceptance criteria and labeled dimensions.
    Tensile Strength Testing (Bond Integrity)All bonded components of the delivery catheter should withstand forces that exceed those encountered during clinical use.Passed: Bonded joints of delivery catheters were tested to failure and withstood forces exceeding those encountered clinically.
    Stent IntegrityNo crevice or pitting corrosion; No more than 2 weld breaks; No significant weight loss; Meet specification for radial compression and expansion forces; Wire must meet specification for tensile properties; Stent cover integrity must meet requirements for holes and delamination.Passed: Stents tested represent worst-case scenario. No degradation observed.
    MRI CompatibilityMRI compatibility labeling must be supported by testing. Displacement forces should not damage tissue, and localized temperature increases should not damage tissues when patients are scanned as outlined in the labeling.Evaluated: Mean deflection angle of 3°. Worst case projected temperature rise of 5.5°C for whole body average SAR of 4 W/kg. Artifacts appeared as localized signal voids extending ~10 mm from stent wall perimeter and ~2mm beyond each end. This data supports the MRI Conditional labeling.
    Shelf Life (Package Integrity)Packaging must not be compromised below a minimum pressure threshold. No channels, pinholes, or other barrier breaches allowed. All samples must not have channels visible at 10X magnification.Passed: Package integrity testing demonstrated maintenance of sterility.
    Shelf Life (Functional Performance)Deployment Testing: Delivery catheter should safely and reliably deliver the stent to the intended location without damage to the stent or patient after 25 months real time aging. Expansion/Compression Force Testing: Forces must not be impacted by aging. Flexural Rigidity: Stents must not become rigid after aging. Covering Material Integrity: Maximum of 6 diamonds containing compromised coating integrity. Weld Integrity: No more than two broken wire weld joints observed. Tensile Strength: Bonded components withstand forces after aging.Passed: Deployment testing passed with largest stent size after aging. Expansion/compression forces passed after aging. Flexural rigidity passed after aging. Covering material integrity passed (maximum 6 diamonds with compromised coating). Weld integrity passed (no more than 2 broken welds). Tensile strength testing passed after aging.
    Clinical Performance (Removal)The ability to safely place and subsequently remove the stent after the maximum labeled indwell period.Stent Removability: 72.4% (92/127) ITT, 78.0% (92/118) PP. Removal Success: 84.3% (107/127) ITT, 90.7% (107/118) PP. Stent indwell up to 613 days (~20 months).
    Clinical Performance (Functionality)The stent resolves strictures during the maximum labeled indwell period.Stent Functionality During Stent Indwell: 77.2% (98/127) ITT, 83.1% (98/118) PP (lack of required reintervention). Stricture Resolution: 74.0% (94/127) ITT, 79.7% (94/118) PP (lack of stricture-related re-intervention). Biliary obstructive symptoms reduced from 51.2% (baseline) to 6.3% (Indwell month 6). Bilirubin and alkaline phosphatase levels decreased.
    Clinical Performance (Recurrence)Stricture resolution is maintained post-stent removal.No Stricture Recurrence: 85.1% (80/94) ITT/PP patients who initially achieved stricture resolution did not experience recurrence over median 19.0 months follow-up.
    Clinical Performance (Adverse Events)All adverse event data including bile duct obstruction and trauma to the bile duct should be reported.SAEs (patients with at least one SAE): 36.2% (46/127) ITT, 36.4% (43/118) PP. Common SAEs: Pancreatitis (16%), Cholangitis (9%). No stent or stent removal related deaths.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study (Test Set): 127 patients for the Intent-to-Treat (ITT) cohort, and 118 patients for the Per-Protocol (PP) cohort.
    • Data Provenance:
      • Country of Origin: The clinical study was a "large prospective multinational study utilizing 13 centers in 11 countries outside of the US (OUS)." Specific countries are not listed, but it indicates a broad international representation.
      • Retrospective/Prospective: The clinical study was explicitly prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    The document does not specify the number or qualifications of experts establishing ground truth for the clinical study directly. The "ground truth" for clinical outcomes (e.g., stricture resolution, stent removability, adverse events) would have been established by the clinical investigators at the 13 centers based on patient follow-up, medical records, imaging, and endoscopic findings. These would be qualified medical professionals (e.g., gastroenterologists, interventional radiologists) involved in the care of the patients and the conduct of the trial. However, no specific details on their number or specific qualifications (e.g., "radiologist with 10 years of experience") are provided within this document.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for establishing clinical outcomes. Clinical outcomes would likely have been determined based on predefined criteria in the study protocol and assessed by the attending clinicians at each site. Serious adverse events (SAEs) were reported and summarized, indicating a system for tracking and categorizing these events, but not explicitly an independent adjudication panel.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study involving human readers improving with AI vs. without AI assistance was not done. This device is a medical stent, not an AI diagnostic tool, so such a study would not be applicable. The clinical study assessed the device's performance directly in patients.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone (algorithm only) study was not done. This device is a physical medical device (stent system) that requires human intervention (endoscopic placement and removal). Its performance is inherently linked to human use in a clinical setting.

    7. Type of Ground Truth Used

    The ground truth for the clinical study was based on clinical outcomes data from patients, including:

    • Endoscopic findings (stent placement, removability, removal success)
    • Patient symptoms (biliary obstructive symptoms)
    • Laboratory test results (bilirubin, alkaline phosphatase levels)
    • Medical records (re-interventions, recurrence, adverse events, surgical outcomes)

    For the non-clinical/bench studies, the ground truth was established through engineering and laboratory test standards, specifications, and physical measurements (e.g., ISO, ASTM standards, pre-defined tolerance ranges, visual inspections, mechanical testing).

    8. Sample Size for the Training Set

    The document describes a single prospective clinical study acting as the primary evidence for the device's clinical performance. It does not mention a separate "training set" in the context of machine learning or AI models. This is a traditional medical device submission, not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of an AI/ML model for this traditional medical device, this question is not applicable. The clinical study data was used to demonstrate the device's safety and effectiveness directly, not to train an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121430

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-01

    (291 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K061231, K081733, K012752, K030107

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is used in palliation of malignant neoplasms in the biliary tree.

    Device Description

    Stent Description: This flexible, self-expanding stent is constructed of nitinol wire with a radiopaque core. Both ends of the stent have an increased diameter called flanges intended to provide resistance to migration. The total length of the stent in its collapsed state is indicated by radiopaque markers on the inner catheter assembly. The stent is provided in either of two body diameters 8mm or 10mm and in the following lengths 4, 6, 8 or 10cm.

    Stent Delivery System Description: The stent is mounted on an inner catheter, which accepts a 0.035" wire guide and is constrained by an outer sheath. A pistol-grip delivery handle allows stent deployment or recapture. The introducer sheath diameter is 8.5Fr and the working length is 200cm.

    AI/ML Overview

    The provided text describes the Cook Ireland Evolution® Biliary Stent System. Here's an analysis of the provided information concerning acceptance criteria and supporting studies, formatted as per your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit in Bench Testing)Reported Device Performance (Summary from study)
    Mechanical PerformanceDeployment functionalitySuccessfully completed
    Expansion forceSuccessfully completed
    Compression forceSuccessfully completed
    Dimensions ( соответствие спецификациям )Successfully completed
    Joint strengthSuccessfully completed
    Material PropertiesCorrosion resistanceSuccessfully completed
    BiocompatibilityCompliance with ISO 10993-1Supports safety of the Evolution® Biliary Stent System
    Shelf LifeMaintenance of device integrity over timeSuccessfully completed
    MRI CompatibilityCompatibility with MRI environmentsSuccessfully completed
    Equivalence to PredicatesShares many technological characteristics with predicate devices (materials, stent type, preloaded on delivery system, fluoroscopy visualization, dimensions, permanent implantation, expanded stent ends, delivery system functionality, wire guide compatibility, recapture capability, single use, sterile)Substantially equivalent to predicate devices (Boston Scientific Wallflex™ Biliary RX Stent System and Wallstent™ RX Biliary Endoprosthesis Stent System)

    2. Sample Size Used for the Test Set and Data Provenance

    The document refers to "Performance (bench) testing" which was "successfully completed." However, it does not specify the sample size used for the test sets for any of the individual bench tests (deployment, expansion force, etc.).

    The data provenance is laboratory bench testing, presumably conducted by or for Cook Ireland Ltd. The country of origin for the data is not explicitly stated beyond Cook Ireland Ltd.'s address in Limerick, Ireland. The study is a retrospective evaluation of device characteristics against pre-defined engineering and regulatory standards, rather than a prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Bench testing typically relies on engineers and technicians to execute tests and interpret results against established specifications, rather than clinical experts establishing a "ground truth" in the same way it would be done for diagnostic AI.

    4. Adjudication Method for the Test Set

    This information is not provided. Bench testing is usually less about adjudication and more about objective measurements against pre-defined pass/fail criteria.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is a physical medical device (a stent and its delivery system), not an AI algorithm. Therefore, the concept of human readers improving with or without AI assistance is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study in the context of an algorithm or AI was not done. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the bench testing, the "ground truth" was established by predetermined technical specifications, engineering standards (e.g., ASTM standards), and regulatory guidance (FDA's Guidance for the Content of Premarket Notifications for Metal Expandable Biliary Stents). The device's performance was measured against these objective, quantitative criteria.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. There is no mention of a "training set" as this is a physical medical device and not an AI/machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122072

    Validate with FDA (Live)

    Date Cleared
    2012-09-28

    (74 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K061231, K081733, K083374, K083627, K112543

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered are indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.

    Device Description

    The WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex 114 Biliary RX Stent System Fully Covered consist of a flexible delivery system preloaded with a self-expanding biliary metal stent.

    The stent is offered uncovered or covered. The covered stents are offered as fully covered, or partially covered with a Permalume™ stent covering. The stent wires have a radiopaque core to improve radiopacity.

    The stent is preloaded onto the delivery system. which has radiopaque marker bands used to aid in imaging during deployment of the stent. The delivery system accommodates a 0.035 in (0.89 mm) guidewire.

    AI/ML Overview

    The provided text is a 510(k) summary for the WallFlex™ Biliary RX Stent System, which is a medical device. This document does not describe the acceptance criteria and a study proving a device meets those criteria in the context of an Artificial Intelligence (AI) or software-as-a-medical-device (SaMD) study.

    Instead, this document focuses on the substantial equivalence of the WallFlex™ Biliary RX Stent System to its predicate devices for a specific labeling claim regarding MR Conditional safety and compatibility. The performance data mentioned is for in-vitro testing on the biliary metal stent, which is typical for hardware medical devices, not AI/SaMD.

    Therefore, I cannot extract the requested information (acceptance criteria, device performance, sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth types) because the provided text does not describe an AI/SaMD study. It's a regulatory submission for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112543

    Validate with FDA (Live)

    Date Cleared
    2011-12-29

    (119 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WallFlex Biliary Transhepatic Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.

    Device Description

    The WallFlex™ Biliary Transhepatic Stent System is comprised of two components: the implantable WallFlex Biliary Transhepatic stent and the delivery system.

    The stent is offered uncovered or covered. The covered stents are offered as fully covered, or partially covered with a Permalume™ stent covering. The stent wires have a radiopaque core to improve radiopacity.

    The stent is preloaded onto the delivery system, which has radiopaque marker bands used to aid in imaging during deployment of the stent. The delivery system accommodates a 0.035 in (0.89 mm) guidewire.

    AI/ML Overview

    Here's an analysis of the provided text regarding the WallFlex™ Biliary Transhepatic Stent System, focusing on acceptance criteria and supporting studies:

    Acceptance Criteria and Device Performance for WallFlex™ Biliary Transhepatic Stent System

    This device appears to be a medical stent system, and as such, the "acceptance criteria" and "device performance" are primarily focused on the physical and biological characteristics of the device itself, rather than diagnostic accuracy or algorithmic performance usually seen in AI/software devices. The study conducted is a demonstration of substantial equivalence to existing predicate devices through a combination of biocompatibility and performance bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestsReported Device Performance
    BiocompatibilityCytotoxicityPassed (for delivery system), Stent identical to predicate (K061231, K081733, K083374, K083627) - no further testing.
    SensitizationPassed (for delivery system)
    IrritationPassed (for delivery system)
    Acute Systemic ToxicityPassed (for delivery system)
    USP Physicochemical TestPassed (for delivery system)
    Material Mediated Rabbit Pyrogen TestPassed (for delivery system)
    Performance TestingDeploymentSuccessfully completed, achieved substantial equivalence.
    ReconstrainmentSuccessfully completed, achieved substantial equivalence.
    Guidewire PassageSuccessfully completed, achieved substantial equivalence.
    Trackability/PushabilitySuccessfully completed, achieved substantial equivalence.

    2. Sample Sizes Used for the Test Set and Data Provenance

    Due to the nature of the device (a physical medical stent system) and the type of evaluation performed (bench testing for substantial equivalence), the concept of "test set sample size" and "data provenance" (country, retrospective/prospective) in the context of data-driven AI models does not directly apply here.

    • The "test set" for biocompatibility comprised samples of the delivery system and relied on the stent component being identical to previously cleared devices. The exact number of samples for each biocompatibility test is not specified in the summary but would be standard for EN ISO 10993-1:2009.
    • The "test set" for performance testing involved physical samples of the WallFlex™ Biliary Transhepatic Stent System. The exact number of units tested for each performance metric (deployment, reconstrainment, etc.) is not specified, but the summary states "comparative performance testing was successfully completed."
    • There is no mention of country of origin for data or retrospective/prospective study design as this was a benchtop engineering and materials performance study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device submission. Ground truth, in the context of medical AI/software devices, refers to the definitive determination of a condition or diagnosis, often established by expert consensus or histological examination. For a physical stent system, the "truth" is whether the device meets its engineering and material specifications, which is determined by objective physical and chemical testing rather than expert interpretation of medical data.

    4. Adjudication Method for the Test Set

    Not applicable. The evaluation was based on objective physical and chemical testing, not subjective interpretation requiring adjudication of agreement between experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI/software diagnostic device, thus an MRMC study comparing human readers with and without AI assistance is irrelevant.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this device's evaluation was based on:

    • Established Biocompatibility Standards: EN ISO 10993-1:2009 for the delivery system materials.
    • Engineering and Design Specifications: For the performance characteristics (deployment, reconstrainment, guidewire passage, trackability/pushability). The success of these tests is determined by adherence to predefined, objective performance criteria.
    • Identity to Predicate Device Materials: For the stent component's biocompatibility, the ground truth was "identity to the current WallFlex Biliary RX Stent."

    8. Sample Size for the Training Set

    Not applicable. There is no training set in the context of an AI/machine learning model for this physical medical device. The "training" for such devices involves design, manufacturing, and R&D processes, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set as understood in AI/ML contexts.