Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140678

    Validate with FDA (Live)

    Date Cleared
    2014-06-19

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical studies were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

    8. The sample size for the training set

    This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1