510(k) Data Aggregation

    K Number
    K212108

    Validate with FDA (Live)

    Device Name
    Dynamic TiBase
    Date Cleared
    2021-09-24

    (80 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K182246, K034014, K072933, K191919, K162021

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.

    Device Description

    The purpose of this submission is to obtain marketing clearance for Dynamic TiBase, a series of twopiece titanium base abutments that require the fabrication of patient-specific custom superstructures using CAD-CAM technology. In final, finished form, the subject device abutments are intended to be used as a two-piece abutment composed of the base bottom-half (titanium base) with a bonded CAD-CAM zirconia top-half. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Talladium España, SL Dynamic TiBase, which is an endosseous dental implant abutment. The primary purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove clinical effectiveness or safety in the same way clinical trials do. Therefore, the information provided focuses on non-clinical data and comparisons to existing devices, rather than clinical study results with human subjects.

    Based on the provided document, here's a breakdown of the acceptance criteria and the study that proves the device meets them, specifically in the context of a 510(k) submission:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission for a device like this, the "acceptance criteria" are not clinical performance metrics in the traditional sense (e.g., sensitivity, specificity for a diagnostic device). Instead, they revolve around demonstrating substantial equivalence to predicate devices in terms of:

    • Intended Use: The device performs the same function for the same patient population.
    • Technological Characteristics: The device has similar design, materials, and operating principles.
    • Performance (Non-Clinical): The device performs similarly to or better than the predicate in engineering/bench tests relevant to its function and safety. Any differences in technological characteristics must not raise new questions of safety or effectiveness.

    The reported device performance is primarily based on non-clinical (bench) testing and comparison tables, not clinical studies in human subjects.

    Here's a table summarizing the "acceptance criteria" (or areas of equivalence demonstration) and the reported device performance as per the document:

    Acceptance Criterion (Area of Equivalence)Reported Device Performance (as demonstrated in 510(k))
    Intended UseDynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient. Demonstrated Equivalence: Stated to be substantially equivalent to primary predicate (K202026) and reference devices (K182246, K191919, K162021) in indications for use. Differences in wording do not affect intended use.
    Technological CharacteristicsDevice Type: CAD-CAM Titanium Base Abutments (two-piece, titanium base with bonded CAD-CAM zirconia superstructure). Materials: Titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3 for abutments and screws. Zirconia (ISO 13356) for copings. Design/Features: Straight design, 0.7 mm gingival height, cut-out for angled screw channel, specific post heights. Compatible with Thommen SPI® CONTACT dental implant fixtures.
    BiocompatibilityAll subject device abutments are biocompatible. Performance: Biocompatibility testing performed according to ISO 10993-12 (referenced) and confirmatory testing for finished devices (ISO 10993-5 and ISO 10993-12) was done.
    Sterilization ValidationThe device can be effectively sterilized by the end-user. Performance: Moist heat sterilization validation performed according to ISO 17665-1 and ISO 11737-2 (referenced). Device provided non-sterile for end-user sterilization.
    Mechanical Performance (Strength/Fatigue)The device is mechanically robust enough for its intended use. Performance: Static compression and compression fatigue testing performed according to ISO 14801 of worst-case constructs (abutments, zirconia superstructures, abutment screws, compatible implants). The fatigue limit data demonstrated sufficient strength. Differences in superstructure design parameters (wall thickness, gingival height) were mitigated by mechanical testing.
    Manufacturing ProcessThe manufacturing process is consistent and yields safe devices. Performance: Titanium alloy components manufactured from identical materials in identical facilities using identical processes as previously cleared products (K162021). Anodization process is also identical.
    Compatible Implant InterfaceThe device properly interfaces with specified implants. Performance: Established compatibility with Thommen SPI® CONTACT dental implant fixtures (cleared in K034014 and K072933) via business agreement and identical interface connections/platforms.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document explicitly states "No clinical data were included in this submission." The "studies" were non-clinical (bench) tests. While the ISO 14801 standard specifies the number of samples for fatigue and static testing (typically 5-10 for static, 10-12 for fatigue), the exact sample sizes used for each specific test (e.g., number of abutments tested at different loads) are not detailed in this summary. It only says "worst-case constructs" were tested.
    • Data Provenance: The data is non-clinical (bench testing) performed by the manufacturer or a contracted lab. The manufacturer is Talladium España, SL, located in Lleida, Spain. This indicates the testing was likely done in Spain or a European lab adhering to ISO standards. The data is retrospective in the sense that the tests were completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For a 510(k) submission based on non-clinical data, there are no "experts" establishing a clinical "ground truth" through consensus on patient cases. The "ground truth" for non-clinical testing is compliance with established engineering and biocompatibility standards (e.g., ISO 14801, ISO 10993). The "experts" here would be the engineers and scientists who designed and conducted the tests and analyzed the results according to these international standards. Their qualifications are inherent in their ability to perform such testing and demonstrate compliance with these standards.

    4. Adjudication method for the test set

    Not applicable. There's no clinical test set requiring adjudication. The non-clinical tests (biocompatibility, sterilization, mechanical) are evaluated against predefined acceptance criteria established by the relevant ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental implant abutment, not an AI-powered diagnostic imaging tool. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this 510(k) submission is compliance with established international engineering standards (e.g., ISO 14801 for mechanical performance, ISO 10993 for biocompatibility, ISO 17665-1 and ISO 11737-2 for sterilization) and the demonstration of substantial equivalence to predicate devices. This is achieved through:

    • Bench Test Results: Quantitative data from static compression and compression fatigue tests.
    • Material Specifications: Conformance to ASTM and ISO material standards.
    • Biocompatibility Test Reports: Results showing no adverse biological reactions.
    • Sterilization Validation Data: Proof of effective sterilization.
    • Direct Comparison: Tables and textual analysis comparing the subject device's features and performance to those of legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051502

    Validate with FDA (Live)

    Date Cleared
    2005-06-21

    (14 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K0030345, K022038, K030689, K034014

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Thommen SPI® Dental Implant System, TST Surface, is intended to be surgically placed, either immediately or delayed, in the bone of the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures. SPI® Dental Implants can be loaded immediately if they are used for crowns, bridges or overdentures on at least four implants in the mandibular arch or six implants in the maxillary arch.

    Contraindications for the use of SPI® CONTACT implant ø 3.5 mm and SPI® ELEMENT implant ø 3.5 mm:
    These implants are not suitable for applications in areas where pronounced rotation and/or bending movements are expected to subject the implant to large bending movements.

    • Restoration of posterior teeth in the upper or lower jaw -
    • Single-tooth restoration of canines and central incisors in the upper jaw -
    • Any application involving retentive anchors -
    Device Description

    The Thommen SPI® Dental Implant System with TST Surface (the modified design) is a self tapping, root form, endosseous dental implant made of commercially pure grade titanium. The implant surface is smooth machined on the transgingival portion and treated with the TST implant surface process, in the area intended to contact bone. The modified TST Surface is available on all implants in the SPI® Dental Implant System. The system is constructed of materials that have a long clinical history of proven acceptance and performance.

    AI/ML Overview

    This 510(k) summary describes a device modification to an existing dental implant system. The core of the submission focuses on demonstrating substantial equivalence to a predicate device, particularly concerning a new surface treatment (TST Surface). As such, a traditional "study" proving acceptance criteria with performance metrics, ground truth, and statistical analysis as one might find for an AI/ML diagnostic device is not applicable here.

    Instead, the submission relies on the established safety and effectiveness of the predicate device and demonstrates that the modified device's characteristics (intended use, operating principle, basic design, materials, packaging, sterilization) are substantially equivalent.

    Therefore, many of the requested categories (2-9) do not have direct answers in this document because they pertain to a different type of device evaluation.

    However, I can extract the closest equivalent information for the acceptance criteria and implied "performance."


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Intended Use Equivalence: Same as predicate device."has the same intended use"
    Operating Principle Equivalence: Same as predicate device."uses the same operating principle"
    Basic Design Equivalence: Same as predicate device."incorporates the same basic design" (modified to include new surface)
    Materials Equivalence: Same as predicate device."incorporates the same materials" (commercially pure grade titanium)
    Packaging & Sterilization Equivalence: Same as predicate."is packaged and sterilized using the same materials and processes" (gamma irradiation at 25 kGy)
    Biocompatibility: Materials have a long clinical history."constructed of materials that have a long clinical history of proven acceptance and performance."
    Conformity to Standards: Meets relevant chemical and mechanical standards."meets the chemical and mechanical requirements of ASTM F67 and ISO 5832-2."
    Sterility Assurance Level (SAL): Meets specified SAL.Sterilization validated by gamma irradiation at 25 kGy (2.5 Mrad) to achieve SAL 10^-6 according to ISO 11137.

    Note: The "acceptance criteria" here are inferred from the requirements for demonstrating substantial equivalence for a device modification.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is not a study assessing device performance on a clinical or imaging dataset. The "test" is the demonstration of substantial equivalence through comparison of device characteristics and compliance with standards.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No ground truth was established by experts in this context.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No adjudication of a test set occurred.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was conducted.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this submission is the established safety and effectiveness of the predicate device, against which the modified device's characteristics are compared for substantial equivalence.


    8. The sample size for the training set

    Not applicable. This device does not involve a training set.


    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a training set.