Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K131337

    Validate with FDA (Live)

    Date Cleared
    2013-10-11

    (155 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K060569, K031201

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Balanced Knee® System High Flex Vitamin E PS tibial insert and patella are intended for use in cemented total knee arthroplasty procedures with the following.indications:

    1. Loss of knee joint configuration and joint function.
    2. Osteoarthritis of the knee joint.
    3. Rheumatoid arthritis of the knee joint.
    4. Post-traumatic arthritis of the knee joint.
    5. Valgus, varus, or flexion deformities of the knee joint.
    6. Revision procedures where other treatments or devices have failed.
    Device Description

    The Balanced Knee® System High Flex Vitamin E (High Flex Vit E) PS tibial insert and patella are machined from extensively crosslinked, compression molded, Vitamin E UHMWPE. Both components are single use only. The High Flex Vit E PS tibial insert must be used in conjunction with the High Flex PS femoral component (K123457). Used together, these components are designed to accommodate increased range of motion up to 150° of flexion. The High Flex Vit E patella may be used in conjunction with the Balanced Knee® System (BKS) femoral components (K994370), the BKS modular femoral components (K060569), or the High Flex PS femoral components (K123457). Both the High Flex Vit E PS tibial insert and patella may be used in conjunction with the BKS standard and modular tibial trays, tibial augments, and stems to complete the semi-constrained modular knee prosthesis. The tibial trays, tibial augments, and stems were approved under K994370 and K031201.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Balanced Knee® System High Flex Vitamin E PS Tibial Insert and Patella:

    The provided document is a 510(k) summary for a medical device seeking substantial equivalence to existing predicate devices. It focuses on the technological characteristics and performance data to demonstrate this equivalence, rather than a clinical study with human patients. This is common for devices that are modifications of already approved devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The "Performance Data" section (page 2) lists various properties tested. Since this is a 510(k) summary demonstrating substantial equivalence, the "acceptance criteria" are implicitly that the device's performance is "Substantially Equivalent" or "Substantially equivalent or better than" the predicate device(s).

    PropertyAcceptance Criteria (Implied)Reported Device Performance
    Range of MotionSubstantially Equivalent to predicate device, BKS High FlexUp to 150° flexion; Substantially Equivalent to predicate device, BKS High Flex
    Femorotibial ConstraintSubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent to predicate device, BKS High Flex
    Femorotibial Contact AreaSubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent to predicate device, BKS High Flex
    Patellofemoral ConstraintSubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent to predicate device, BKS High Flex
    Patellofemoral Contact AreaSubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent to predicate device, BKS High Flex
    PS Spine FatigueSubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent
    Insert Assembly/DisassemblySubstantially Equivalent to predicate device, BKS High FlexSubstantially Equivalent
    Crosslinking Characterization of High Flex Vit ESubstantially equivalent or better than BKSSubstantially equivalent or better than BKS
    Vitamin E Characterization of High Flex Vit ESubstantially equivalent or better than BKSSubstantially equivalent or better than BKS
    High Flex Knee WearNot explicitly stated as "equivalent," but improvement is implicitDecrease in wear over BKS

    Note: The last item, "High Flex Knee Wear," is reported as a "Decrease in wear over BKS," which suggests an improvement rather than just equivalence. This would still be considered meeting an implicit acceptance criterion of being at least as good as the predicate (or better).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify sample sizes in terms of number of patients or clinical cases because the studies described are non-clinical (benchtop) tests. The "test set" here refers to the physical samples of the device and predicate devices used in the mechanical and material characterization tests. The provenance of this data is from the manufacturer's internal testing. There is no information about country of origin for data or whether it's retrospective/prospective, as these terms apply to clinical studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This document describes non-clinical (benchtop) testing rather than studies that require expert-established ground truth from medical imaging or clinical observations. Therefore, there were no medical experts (e.g., radiologists) involved in establishing ground truth for the test set in the way this question implies. Ground truth for these types of tests is established by engineering and material science standards and measurements.

    4. Adjudication Method for the Test Set

    Not applicable. As this is non-clinical benchtop testing, there is no "adjudication method" involving human experts in the context of clinical studies (e.g., 2+1, 3+1). The results are based on objective physical measurements and comparisons to predicate devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, as this device is not an AI-assisted diagnostic tool. No effect size of human readers improving with AI vs. without AI assistance is reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This is a medical implant (total knee replacement components), not a software algorithm or AI device. Therefore, a standalone algorithm performance study is not applicable.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Objective mechanical and material property measurements: Standards for flexion, constraint, contact area, fatigue, crosslinking, and Vitamin E content.
    • Comparison to predicate devices: The "ground truth" for substantial equivalence is the established performance characteristics and safety profile of the legally marketed predicate devices. The new device must meet or exceed these established benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103837

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-03-24

    (84 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K994370, K060569, K031201, K020383, K090705

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Balanced Knee System Revision Offset Tibia is indicated for use in total knee arthroplasty procedures, for cemented use only.

    Total knee arthroplasty is indicated for the following conditions:

    1. Loss of joint configuration and joint function.
    2. Osteoarthritis of the knee joint.
    3. Rheumatoid arthritis of the knee joint.
    4. Post-traumatic arthritis of the knee joint.
    5. Valgus, varus, or flexion deformities.
    6. Revision procedures where other treatments or devices have failed.
    Device Description

    The proposed Balanced Knee System Revision Offset Tibia is a modular system intended for primary and revision knee surgery. It consists of an offset tibial tray, adapter, screw, and stem. All components are machined from titanium alloy (Ti-6Al-4V ELI).

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Balanced Knee System Revision Offset Tibia) and it details the regulatory process and claims of substantial equivalence. However, it does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the way described in the request.

    The document discusses:

    • Mechanical testing: "Mechanical testing shows that the Balanced Knee System Revision Offset Tibia can withstand anticipated in vivo loading and cycling."
    • Tests for substantial equivalence: "The Tests for substantial equivalence include Tibial Tray fatigue and progressive static load mechanical tests, as well as assembly and disassembly tests."

    These statements indicate that some form of testing was performed, likely to demonstrate the mechanical integrity and durability of the device in comparison to predicate devices. However, the document does not provide quantitative acceptance criteria or the reported device performance against such criteria. It also lacks details on:

    • Sample size used for the test set or data provenance.
    • Number of experts or their qualifications for establishing ground truth.
    • Adjudication method for the test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone performance study.
    • Type of ground truth used.
    • Sample size for the training set (as this is a mechanical device, not an AI/algorithm-based one).
    • How ground truth for the training set was established.

    Therefore, I cannot populate the requested table and information as the provided text does not contain these specific details. The 510(k) process for this type of device relies on demonstrating substantial equivalence to existing legally marketed devices, primarily through engineering principles, material compatibility, and mechanical performance testing, rather than clinical efficacy studies with expert consensus or AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090705

    Validate with FDA (Live)

    Date Cleared
    2009-10-09

    (206 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K994370, K020383, K031201

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ortho Development Balanced Knee® System is indicated for patients suffering from severe knee pain and disability. Specific indications include femoral, tibial and patellar replacement due to:

    1. Loss of joint configuration and joint function.
    2. Osteoarthritis of the knee joint.
    3. Rheumatoid arthritis of the knee joint.
    4. Post-traumatic arthritis of the knee joint.
    5. Valgus, varus, or flexion deformities.
    6. Revision procedures where other treatments or devices have failed.

    The Balanced Knee® System is indicated in the salvage of previously failed surgical attempts where femoral bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.

    Device Description

    The Balanced Knee System Ultracongruent Tibial Insert may be used in conjunction with the Balanced Knee System CR femorals, standard and modular tibial trays, tibial augments, stems, and patellae. The Ultracongruent Tibial Insert has an anterior lip and tighter anterior curvature to stabilize the knee in the anterior direction and are snap fitted into the tibial trays and locked in place by mating features, the circumferential rim of the tray, and tabs on the anterior and posterior sides. The Ultraconqruent Tibial Insert is available in seven sizes to fit the majority of patients encountered. The sizes match each size of the Balanced Knee System tibial trays cleared under Premarket Notifications K994370, K020383, and K031201. The insert is symmetric, not left/right specific. It is supplied in thicknesses of 7, 8, 9, 10, 11, 12, 13. 14. 16. 18. and 20mm.

    Materials: The device is manufactured from Compression Molded Ultra High Molecular Weight Polyethylene (UHMWPE, ASTM F-648).

    Function: The system functions to provide restoration of function as a replacement for diseased and arthritic knees.

    AI/ML Overview

    The provided text does not contain specific acceptance criteria or detailed study results that would allow for the construction of a table with reported device performance or a comprehensive description of a study as requested. The document is a 510(k) summary for a medical device (Balanced Knee System® Ultracongruent Tibial Insert) and focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed performance study with acceptance criteria.

    The key information regarding performance testing is as follows:

    • Performance Testing: "The Food and Drug Administration have established no performance standards applicable to the Ultracongruent Tibial Insert, however biomechanical testing and analysis of the device was performed with results included as part of the submission. Clinical data and conclusion were not needed for this device."
    • Basis for substantial equivalency: "Testing met all acceptance criteria and verifies that the performance of the Ultracongruent Tibial Insert is substantially equivalent in design, function, material and intended use to the predicate devices."

    Based on the available information, I can only provide a limited response:

    Acceptance Criteria and Study Details for K090705: Balanced Knee System® Ultracongruent Tibial Insert

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state specific numerical acceptance criteria or detail the reported device performance. It only states that "Testing met all acceptance criteria." Therefore, a table detailing these cannot be created from the provided text. The submission relied on biomechanical testing and analysis to demonstrate substantial equivalence, rather than specific performance standards established by the FDA for this device type.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "biomechanical testing and analysis," but does not specify the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable as the submission states that "Clinical data and conclusion were not needed for this device." This implies that the evaluation for substantial equivalence did not rely on human expert interpretation of clinical outcomes or images to establish a 'ground truth' in the way a diagnostic AI device would. The evaluation focused on engineering and material characteristics.

    4. Adjudication Method

    This information is not applicable for the same reasons as point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not conducted as this is a knee implant device, not a diagnostic imaging or AI-assisted interpretation device that would involve human readers.

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study of an algorithm was not conducted as this is a physical medical device (knee implant), not a software algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the biomechanical testing would have been based on established engineering and material science principles and physical measurements/simulations of the device's performance against defined parameters (e.g., strength, durability, wear resistance) compared to predicate devices. It was not clinical, pathological, or outcomes data in the traditional sense for a diagnostic device.

    8. Sample Size for the Training Set

    This is not applicable as there is no mention of an algorithm or machine learning model that would require a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no mention of an algorithm or machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1